  9289   Fri Oct 25 04:03:40 2013 MasayukiUpdateLSC'scope and spectrum analyser for REFL165

As Jenne's Elog we want to see Spectrum and time series of REFL 165 (our PRMI LSC locking PD) to see if the signal is saturated while bring the arms into resonance.
I started to connect the spectrum analyser and the 'scope to REFL165 output.

Directional coupler (Mini=-circuits ZMDC-10-2 ZMDC-20-3) was connected just before the dimod boad input. The main output of coupler is plugged into demod board's input.The other output of the coupler is connected to AG4395A using BNC cable.

The spectrum analyser output can be read using netgpibdata in control room. The IP address is 192.168.113.108 and the GPIB address is 17. For this I dissconected the network hub from another AG4395A, which is at the front of 1X2 lack.

I didn't connected the 300 MHz 'scope right now, but tomorrow it will be connected using power splitter and also be able to get data by internet. For connect 'scope to network, I disconected the network hub from SR785.

  9290   Fri Oct 25 04:54:21 2013 MasayukiUpdateSUSETM violin mode

Summary

When PRMI + 2arms are locked yesterday, we heard the noise from suspension violin mode. For attenuation of that noise, we should design the resonant filter at that frequency and put into the ALS servo. I tried to measure the violin mode of ETMs SUS.

What I did

 1.The arms were locked by IR PDH. I used awggui to excite the suspention. I injected the Normal waveform, 10 Hz of bandwidth wave into C1:SUS-ETMs_ULCOIL_EXC. I put cheby filter in the FIlter of awggui. The order of that filter was 4, that has same bandwidth as that of injection wave and ripple was 4dB. I increase the injection gain with some ramp time(5sec). I swept from 600 Hz to 700 Hz. During that injection I saw the PDH error signal (POX11I and POY11I) in order to find resonance peak of violin mode.
 In ETMX resonances were easily found. That were at 631 Hz and 691 Hz. the 631 Hz peak was seen ALS error signal yesterday. On the other hand, I couldn't find ETMY violin mode. No peaks appeared any frequency.

2. For find the ETMY violin mode, I used dtt swept sine measurement. The excitation channel was C1:SUS-ETMs_ULCOIL_EXC. I measured the TF from excitation channel to POX11I and POY11I error signal. The measurement range was above 400 Hz and below 1000Hz,. The number of point is 600. I attached that result.
In ETMX curve, the coherence become bad near the resonant frequency of violin mode and also the TF is large. Although ETMX violin modes are obvious, ETMY violin modes are not visible. At 660 Hz, 780 Hz, 900 Hz the coherence is not good. That is because 60 Hz comb noise.

Discussion

 I attached the spectrum of the POX and POY error signal. Black and red curve is measured different time. I didn't inject any signal in both measurement, but the violin mode excitation has huge difference. Also there are peaks at beat frequency between violin mode and bounce mode(16 Hz), yaw motion(3 Hz). In ALS in-loop noise or XARM in-loop measurement, sometimes this region had big spikes. That was because of this resonance. And also that resonance peak couples to POY11I.

 I will measure the Q and design the resonant filter for ALS.

Attachment 1: violin1.pdf
violin1.pdf
Attachment 2: violin2.pdf
violin2.pdf
  9294   Fri Oct 25 21:28:49 2013 MasayukiUpdateLSCREFL PDs spectrum

 I measured the spectrum of the REFL165 output using AG4395A. As this entry we put the directional coupler between REFL165 output and demod board input, so I measure the signal from the coupler during the PRMI was locked.

 After measure REFL165, I also measured REFL55 output in order to make sure that the signal is not smaller than noise because of coupler. I terminated the couple output of coupler on the REFL165, and take signal from REFL55 output port directly. Both plots seems same except for around the resonant frequency of each PDs. From this plot we cannot say that the coupler reduce signal to spectrum analyser too much.

 After this measurement I reconnected the REFL165 to analyser and reconnected the REFL55 output to demod board.

Attachment 1: REFL.png
REFL.png
Attachment 2: REFLspe.zip
  9295   Fri Oct 25 21:36:51 2013 MasayukiUpdateLSC'scope and spectrum analyser for REFL165

Quote:

As Jenne's Elog we want to see Spectrum and time series of REFL 165 (our PRMI LSC locking PD) to see if the signal is saturated while bring the arms into resonance.
I started to connect the spectrum analyser and the 'scope to REFL165 output.

Directional coupler (Mini=-circuits ZMDC-10-2 ZMDC-20-3) was connected just before the dimod boad input. The main output of coupler is plugged into demod board's input.The other output of the coupler is connected to AG4395A using BNC cable.

The spectrum analyser output can be read using netgpibdata in control room. The IP address is 192.168.113.108 and the GPIB address is 17. For this I dissconected the network hub from another AG4395A, which is at the front of 1X2 lack.

I didn't connected the 300 MHz 'scope right now, but tomorrow it will be connected using power splitter and also be able to get data by internet. For connect 'scope to network, I disconected the network hub from SR785.

[Jenne, Masayuki]

We changed the Directional coupler from ZMDC-20-3 to ZMDC-20-5-S+ because that coupler seemed to introduce some high frequency noise.

  9304   Mon Oct 28 14:24:01 2013 MasayukiUpdateLSC'scope and spectrum analyser for REFL165

 

 I connected the 'scope between REFL165 output and demod board input. I split the signal from coupler using the splitter (Mini-Circuits ZFSC-2-5). One signal is going to 'scope CH1 and the other is going to spectrum analyzer. I connected the 'scope to 40MARS. The IP adress is 192.168.113.25. I connected that by cabling from 1X2.

 

  9305   Mon Oct 28 18:57:27 2013 MasayukiHowToLSCread 'scope and spectrum analyser datas

 

The command to get the data from spectrum analyzer right now

From command line, put ./netgpibdata -i 192.168.113.108 -d AG4395A -a 17 -f meas01

(EDIT JCD:  You must first be in the correct folder:  /opt/rtcds/caltech/c1/scripts/general/netgpibdata/)

(EDIT JCD again: "meas01" in the command line instruction will be the name of the filename.  Also, the output file meas01.dat has a comment in the first line that must be deleted before you can plot the data.  This sucks, and we should write a script to strip that line, then make nice plots.)

Please take notice that although IP address of AG4395A is same as written in the help of netgpibdata, the GPIB address is not same. It's 17.

 

How to use  'scope from control room.

Open the browser. Put the IP adress of 'scope (192.168.113.25) into adrress bar of the browser. If it's on the network, below screen will open.

You can control 'scope, get the data, and so on from control room.

Please take notice that Google Chrome cannot connect the 'scope. So you have to use the Firefox or other browser.

  9307   Tue Oct 29 10:51:16 2013 MasayukiUpdateSUSETMY sensors compared to ETMX

[Steve, Masayuki]

We lowered the c1iscey machine to make space upside of the computer for heat flow. 

First we turned off the computer. And then we droped the computer down by 1  notches in the rack. Now the upside and downside spaces are almost same. We restarted the computer after that and we leave the door open.

 


  5045   Wed Jul 27 12:31:47 2011 Manuel, IshwitaSummaryPEMWeekly summary

We kept reading about digital filtering

We tested the seismometer last friday

Jan came and tested again the seismometer last monday

We wrote a simulation of the stacks transfer functions, and of the distance between the mirrors.

 

  5063   Fri Jul 29 18:43:02 2011 Manuel, IshwitaUpdatePEMplugging seismometers to ADC

[Manuel, Ishwita, Jenne, Jamie]

We changed the C1PEM model and the names of the C1:PEM channels.

We reinstalled the blue breakout box, since the purple one still didn't work.

So, now the AA board channels are connected as follows...

C1 = C1:PEM-SEIS_GUR1_X

C2 = C1:PEM-SEIS_GUR1_Y

C3 = C1:PEM-SEIS_GUR1_Z

C4 = C1:PEM-SEIS_GUR2_X

C5 = C1:PEM-SEIS_GUR2_Y

C6 = C1:PEM-SEIS_GUR2_Z

C7 = C1:PEM-SEIS_STS_1_X

C8 = C1:PEM-SEIS_STS_1_Y

C9 = C1:PEM-SEIS_STS_1_Z

C11 = C1:PEM-SEIS_STS_2_X

C12 = C1:PEM-SEIS_STS_2_Y

C13 = C1:PEM-SEIS_STS_2_Z

C14 = C1:PEM-SEIS_STS_3_X

C15 = C1:PEM-SEIS_STS_3_Y

C16 = C1:PEM-SEIS_STS_3_Z

C17 = C1:PEM-ACC_MC1_X

C18 = C1:PEM-ACC_MC1_Y

C19 = C1:PEM-ACC_MC1_Z

C20 = C1:PEM-ACC_MC2_X

C21 = C1:PEM-ACC_MC2_Y

C22 = C1:PEM-ACC_MC2_Z

Although the channels for all 3 STS-2 seismometers are made but only one is installed. So only Channels C1 to C9 are now in operation...

We checked the data from the plugged channels with the Dataviewer. We could see the peak whenever someone jumped in the lab. Even Kiwamu jumped and saw his signal.

  5087   Mon Aug 1 23:29:24 2011 Manuel, IshwitaUpdateWienerFilteringGetting Data by matlab

We tried to acquire data from the seismometers and the mode cleaner using the Matlab function

datalist = NDS2_GetData({'C1:PEM-SEIS_GUR1_X_IN1_DQ'}, 996258376 , 10, CONFIG.nds.C)

and encountered the following error

Warning: daq_request_data failed
 
??? Error using ==> NDS2_GetData
Fatal Error getting channel data.

The same error was obtained with the following other channels

C1:PEM-SEIS_GUR2_X_IN1_DQ

C1:PEM-SEIS_STS_1_X_IN1_DQ

But we are able to get data from channel

C1:LSC-MC_OUT_DQ

for the same gps time.

We checked with Dataviewer that the data are saved (we viewed data of last 24h) for every channel.

  5102   Wed Aug 3 02:28:08 2011 Manuel, IshwitaUpdateWienerFilteringWiener Filtering in X-arm

Wiener Filtering was applied on the data collected from the X-arm during the time: GPS time-996380715 (Aug 02, 2011. 21:25:00. PDT) to GPS time-996382215 (Aug 02, 2011. 21:50:00. PDT) for a duration of 1500 seconds. During this time the X-arm was locked, we checked it by acquiring data from channel C1:LSC-TRX_OUT_DQ .

The seismometers were near the beam splitter (guralp2) and near MC2 (guralp1).

Target data was obtained from channel C1:LSC-XARM_IN1_DQ.

Schermata-6.png

Schermata-7.png

Following graphs were obtained after applying the Wiener filter:

 

      1.Seismic data acquired from Guralp1 (X and Y) and Guralp2 (X and Y)                              2.Seismic data acquired from Guralp2 X                                                              3.Seismic data acquired from Guralp2 Y 

WFgur1X1Y2X2YN20000srate2048.pngWFgur2XN20000srate2048.pngWFgur2YN20000srate2048.png

These graphs were obtained with srate = 2048 (sample rate) and N = 20000 (order of the filter).

Graph 1 is the best because the black (residual) line is below the red (target) line for low frequencies since we used seismic data from 4 channels. Graph 3 is the worst because we used seismic data from only one Y channel (Y axis of Guralp2) that is less related with the X-arm mirrors' motion since they are oriented orthogonally.

  5106   Wed Aug 3 12:24:08 2011 Manuel, IshwitaUpdateWienerFilteringWeekly summary

Last Friday (Jul 29) we reinstalled the blue breakout box, and changed the names of the C1:PEM channels. Elog Reference

We continued the work on the simulation ad applied wiener filter on the simulated ground motion, but the result is unsatisfactory, yet. We will post reasonable results soon.

We did wiener filtering for the first time on real data from the Xarm while it was locked. Elog Reference

  5157   Tue Aug 9 16:21:59 2011 Manuel, IshwitaUpdateWienerFilteringFirst results of our simulations

We did the simulation of the stacks by defining a transfer function for one stack (green plot) and another similar transfer function for the other stack.

We simulated the ground motion by filtering a white noise with a low pass filter with a cutoff frequency at 10Hz. (blue plot) (the ground motion for the 2 stacks are completely uncorrelated)

We simulated the electronic white noise for the seismic measurements. (black plot)

We filtered the ground motion (without the measurements electronic noise) with the stack's transfer function and subtracted them to find the mirror response (red plot), which is the target signal for the wiener filter.

We computed the static wiener filter with the target signal (distance between the mirrors) and the input data (seismic measurements = ground motion + electronic noise).

We filtered the input and plotted the output (light blue plot).

We subtracted the target and the output to find the residual (magenta plot).

We didn't figure out why the residual is above the electronic noise only under ~6hz. We tried to increase and decrease the electronic noise and the residual follows the noise still only under ~6Hz.

It also shows that the residues are above the target at frequencies over 20Hz. This means that we are injecting noise here.

simu1.png

We tried to whiten the target and the input (using an high pass filter) to make the wiener filter to care even of higher frequencies.

The residues are more omogeneously following the target.

We also plotted the Wiener filter transfer function without making whitening and with making whitening. It shows that if we do whitening we inject no noise at high frequency. But we loose efficency at low frequencies.

simu2w.png

TFwhitened2.png

We shouldn't care about high frequency, because the seismometers response is not good over 50Hz. So, instead of whitening, we should simply apply a low pass filter to the filter output to do not inject noise and keep a good reduction at low frequencies.

 

  5170   Wed Aug 10 12:33:34 2011 Manuel, IshwitaSummaryPEMWeekly summary

We got the results of the wiener filtering simulations (Elog Entry)

We got the power spectra and coherence of the seismic noise measurements from GURALPs and STS seismometers (Elog Entry)

We tried to whiten the target and the input signal for the computation of the wiener filter for the real data, but the results are unsatisfactory. We should not care about high frequencies in wiener filter computation so we will just filter them off in the filter output with a low pass filter.

We just found the right gain for the system seismometer-AAboard-ADC (Elog Entry)

  5213   Fri Aug 12 17:05:22 2011 Manuel, IshwitaConfigurationPEMSTS2 Cable configuration

The WWF_M connector is the end of the STS2 seismometer orange cable and the S1 connector is the end of the gray 26-pin-cable

01050901.PDF

  5171   Wed Aug 10 13:52:23 2011 Manuel , IshwitaUpdatePEMMoving Seismometers

We turned off the power of the seismometers and moved the Guralp1 close to the STS. Both are now situated below the center of the mode cleaner vacuum tube.

We oriented the X axis of the STS & Guralp1 along the X axis of the interferometer. Then we turned on the power again, but the STS channels don't give any signal. We think this is, because we didn't push the auto zero button.

  5187   Thu Aug 11 11:50:56 2011 Manuel , IshwitaUpdatePEMCalibration of Guralp and STS2

 

 We just checked with a function generator the calibration of the ADC. We set a square wave with amplitude 1V. We measured the voltage with the oscilloscope and we found on the data viewer that one volt is 3208 counts. That's what we expected (+/- 10V for 16bits) but now we are more sure.

 

  5033   Mon Jul 25 18:51:38 2011 ManuelUpdatePEMSTS-2 seismometer hardware testing with Jan

[Jan, Manuel, Jenne]

Jenne called Jan to check and figure out why the Streckeisen seismometer (SN #100151) doesn't work, hence we checked the output of the seismometer boxes as we did last friday. (This is the problem of seeing the X and Y channels saturated, when we look at them on a floating 'scope, as in the linked elog entry.)

Jan unplugged and plugged again the orange cable into the seismometer and nothing happened.   Well, what Jan was listening for was "clicks" inside the seismometer indicating that it was receiving power.  We heard these, and moved on to examining the breakout boxes.  Also, we checked that we could hear the "clicks" (one per mass) when we pushed the mass-centering button on the little green companion box.

We weren't sure that the purple box was working properly, so since we had seen the blue box work last time, we changed the purple box with the blue box in rack 1X6.

The Z-channel of the purple box returns a correct signal, that means that all the masses in the seismometer work (because the Z-signal is a linear combination of the three masses U, V, W), the X and Y channel have a DC component of about 10 Volts, Jan said that the recentering of the seismometer masses could need all the night, so we keep the power of the box on. If tomorrow morning the X and Y signal won't  both be zero mean, we will open and check the box.

The power of the box is still on so that the masses can recenter overnight.

Edits by JD

  5034   Mon Jul 25 23:43:20 2011 ManuelHowToElectronicsManual for 1201 Low Noise Preamplifier

I found the manual for the Low Noise Preamplifier Model 1201 at this link and I attached it.

The one we have in the lab (S/N 48332) miss the battery packs and miss also the remote programming options input/output. Its inside battery compartment is empty and I found 2 unscrewed screws with washers and nuts inside the preamplifier box. The battery cable are disconnected and they have 2 green tape labels (-) and 2 red tape label (+).

 

 

Attachment 1: ITHACO_1201_Instruction_&_Maintenance.pdf
ITHACO_1201_Instruction_&_Maintenance.pdf ITHACO_1201_Instruction_&_Maintenance.pdf ITHACO_1201_Instruction_&_Maintenance.pdf ITHACO_1201_Instruction_&_Maintenance.pdf ITHACO_1201_Instruction_&_Maintenance.pdf ITHACO_1201_Instruction_&_Maintenance.pdf ITHACO_1201_Instruction_&_Maintenance.pdf ITHACO_1201_Instruction_&_Maintenance.pdf
Attachment 2: DSC_3249.png
DSC_3249.png
  5318   Mon Aug 29 16:27:34 2011 ManuelConfigurationSUSSUS Summary Screen

I edited the C1SUS_SUMMARY.adl file and set the channels in alarm mode to show the values in green, yellow and red according to the values of the thresholds (LOLO, LOW, HIGH, HIHI)

I wrote a script in python, which call the command ezcawrite and ezcaread, to change the thresholds one by one.

You can call this program with a button named "Change Thresholds one by one" in the menu come down when you click the  button.

I'm going to write another program to change the thresholds all together.

  5428   Thu Sep 15 22:31:44 2011 ManuelUpdateSUSSummary screen

I changed some colors on the Summary of Suspension Sensor  using my italian creativity.

I wrote a script in Python to change the thresholds for the "alarm mode" of the screen.

The script takes a GPS-format start time as the 1st argument and a duration time as the second argument.

For every channel shown in the screen, it compute the mean value during this time.

The 3rd argument is the ratio between the mean and the LOW threshold. The 4th argument is the ratio between the mean and the LOLO threshold.

Then it sets the thresholds simmetrycally for HIGH and HIHI threshold.

It does that for all channels skipping the Gains and the Off Sets because this data are not stored.

For example is ratio are 0.9 and 0.7 and the mean is 10, thresholds will be LOLO=7, LOW=9, HIGH=11, HIHI=13.

You can run the script on pianosa writing on a terminal '/opt/rtcds/caltech/c1/scripts/SUS/set_thresholds.py' and the arguments.

I already run my program with those arguments: 1000123215 600 0.9 0.7

The time is of this morning at 5:00 for 10 minutes

 

This is the help I wrote

HELP: This program set the thresholds for the "alarm mode" of the C1SUS_SUMMARY.adl medm screen.

 Written by Manuel Marchio`, visiting student from University of Pisa - INFN for the 2011 summer at Ligo-Caltech. Thrusday, 15th September 2011.

The 1st argument is the time in gps format when you want to START the mean

The 2nd argument is the DURATION

The 3rd argument is the ratio of the LOW and the HIGH thresholds. It must be in the range [0,1]

The 4th argument is the ratio of the LOLO and the HIHI thresholds. It must be in the range [0,1]

Example: path/set_thresholds.py 1000123215 600 0.9 0.7

and if the the mean is 10, thresholds will be set as LOLO=7, LOW=9, HIGH=11, HIHI=13

 

Attachment 1: sussum.png
sussum.png
  8085   Fri Feb 15 01:41:02 2013 Manasa, YutaSummaryAlignmentIFO aligned and ready for PRMI locking

[Yuta, Manasa, Jenne, Jamie, Steve]

IFO aligned and ready for PRMI locking

Alignment procedure

0. Measured MC centering (off by 5mrad) before getting the doors off.

1. Got the TTs to 0.0 in pitch and yaw.

2. Using the MMTs, the beam was centered on the TTs.

3. TT1 was adjusted such that the incident beam was centered at PRM (with target).

4. TT2 was adjusted such that the beam passed through the center of BS (with target).

5. Centered the beam on PR2 by sliding it on the table.

6. Moved PR2 and tweaked TT2 to center the beam on ITMY and BS respectively.

7. Using TTs, we got the beam centered on ETMY while still checking the centering on ITMY.

8. ITMY was adjusted such that it retro-reflected at the BS.

9. ETMY was aligned to get a few bounces in the arm cavity.

10. Centered on ITMX by adjusting BS and then tweaked ITMX such that we retro-reflected at BS.

11. At this point we were able to see the MI fringes at the AS port.

12. Tweaked ITMX to obtain reflected MI fringes in front of MMT2.

13. By fine adjustments of the ITMs, we were able to get the reflected MI to go through the faraday  while still checking that we were retro-reflecting at the BS.

14. Tweaked the PRM, such that the PRM reflected beam which was already visible on the 'front face back face of faraday' camera went through the faraday and made fine adjustments to see it fringing with the reflected MI that was already aligned.

15. At this point we saw the REFL (flashing PRMI) coming out of vacuum unclipped and on the camera.

16. Started with alignment to get the AS beam out of vacuum. We tweaked OM1 and OM2 (steering mirrors in the ITMY chamber) to center the beams on OM4 and OM3 (steering mirrors in the BSC) respectively.

17. We then adjusted steering mirrors OM5 and OM6 (in the OMC chamber) such that the beam went unclipped out of vacuum.

18. Note that we took out the last steering mirror (on the AS table) in front of the AS camera, so that we can find the AS beam easily. This can be fixed after we pump down.

 

 

Tomorrow

  0. REFL still looks like an egg, but leave it .

  1. Align PRMI (no more in-vac!) .

  2. Align POP/REFL/AS cameras and PDs.

  3. Setup PRM/BS/ITMX/ITMY oplevs.

  4. Balance the coils on these mirrors.

  5. Lock PRMI.

  7694   Fri Nov 9 17:15:05 2012 Manasa, Steve, AyakaUpdateGeneralWe're closed! Pumping down monday morning

Quote:

After a brief look this morning, I called it and declared that we were ok to close up.  The access connector is almost all buttoned up, and both ETM doors are on.

Basically nothing moved since last night, which is good.  Jenne and I were a little bit worried about how the input pointing might have been effected by our moving of the green periscope in the MC chamber.

First thing this morning I went into the BS chamber to check out the alignment situation there.  I put the targets on the PRM and BS cages.  We were basically clear through the PRM aperture, and in retro-reflection.

The BS was not quite so clear.  There is a little bit of clipping through the exit aperture on the X arm side.  However, it didn't seem to me like it was enough to warrant retouching all the input alignment again, as that would have set us back another couple of days at least.

Both arm green beams are cleaning coming out, and are nicely overlapping with the IR beams at the BS (we even have a clean ~04 mode from the Y arm).  The AS and REFL spots look good.  IPANG and IPPOS are centered and haven't moved much since last night.  We're ready to go.

The rest of the vertex doors will go on after lunch.

Jamie and Steve got the ETM doors on this morning.

We got the other heavy doors including the ITMs, BS and the access connector in place.

If nobody raises any concerns in reply to this elog, Steve will assume it as a green signal and will start pumping down first thing Monday morning after the final check on the access connector bellow screws.

 

Steve! 

Ayaka and I got the ITMY and BS door closed at 45foot pounds just now. 

  8282   Wed Mar 13 03:12:47 2013 Manasa, JenneUpdateLockingTWO arms TWO colors

[Jenne, Manasa]

2 colors 2 arms realized!

1. Spot centering:

We spot centered the IR in both arms.
- Use TT1 and TT2 to center in Y arm (I visually center the spots on the ITM and ETM and then use TTs iteratively)
- Use BS-ETM to center in X arm

Spot positions after centering
               X arm            Y arm
         itmx    etmx        itmy    etmy
pitch    -0.86    0.37        1.51    0.05
yaw      0.01    -0.1        0.08    0.10


2. TT1 drifting in pitch (Bistable)
During the arm alignment routine for spot centering, we observed that TRY dropped (from TRY = 0.9 until the arm lost lock) every 40minutes or so. The arm was relocked by moving TT1 in pitch. The (locking - unlocking due to drift - relocking) cycle was monitored and we observed that it was bistable i.e. if TT1 was moved up in pitch (0.2 on the slider) to relock for the first time ; the next time it lost lock, TT1 had to be moved down by nearly the same distance to relock the arm.
Moving TT2 or the testmasses did not help with relocking the arms; so TT1 seems to be the one causing all th trouble atleast for today.

3. ALS - green alignment

We then moved on to Ygreen.  We used the out of vac steering mirrors to center the beam on the 2 irises that are in place on the table, which was a good starting place.  After doing that, and tweaking a small amount to overlap the incident and reflected beams on the green steering mirrors, we saw some mode lock.  We adjusted the end table steering mirrors until the Ygreen locked on TEM00.  We then followed Rana's suggestion of locking the IR to keep the cavity rigid while we optimized the green transmission.  Yuta, while adjusting ITMY and ETMY (rather than the out of vac mirrors) had been able to achieve a green transmission for the Yarm of ~2700 counts using the GTRX DC PD that's on the table. We were only able to get ~2200, with brief flashes up to 2500.

After that, we moved on to the X arm.  Since there are no irises on the table, we used the shutter as a reference, and the ETM optic itself.  Jenne looked through the viewport at the back of the ETM, while Manasa steered mirrors such that we were on the center of the ETM and the shutter.  After some tweaking, we saw some higher order modes lock.  We had a very hard time getting TEM00 to stay locked for more than ~1 second, even if the IR beam was locked.  It looks like we need to translate the beam up in pitch.  The leakage of the locked cavity mode is not overlapped with the incident beam or the promptly reflected beam.  This indicates that we're pretty far from optimally aligned.  Manasa was able to get up to ~2000 counts using the same GTRX PD though (with the Ygreen shutter closed, to avoid confusion).  Tomorrow we will get the Xarm resonating green in the 00 mode.

We need to do a little cleanup on the PSL green setup.  Yuta installed a shutter (I forget which unused one he took, but it was already connected to the computers.), so we can use it to block the PSL green beam.  The idea here is to use the 4th port of the combining beam splitters that are just before each beat PD, and place a PD and camera for each arm.  We already have 2 PDs on the table connected to channels, and one camera, so we're almost there. Jenne will work on this tommorrow during the day, so that we can try to get some beat signals and do some handoffs in the evening.

  7196   Wed Aug 15 17:17:58 2012 Manasa, JanUpdateIOORingdown measurements

Finally ringdown at IMC conquered and oopsie that came out so clean!

The finesse of the cavity from the current ringdown measurement, F= 453, differs from the measurements made in the document dated 10/1/02 on dcc...not sure if things have changed since then.

While I thought that the bumps observed at the end of the ringdown might be because of the cavity trying to lock itself, Jan commented that they have always existed in these measurements and their source is not known yet.

Ringdown_815.jpg

  7675   Tue Nov 6 17:22:51 2012 Manasa, JamieUpdateAlignmentAlignment- POY and oplevs

Right now, Manasa, Jamie and Ayaka are doing some finishing touches work, checking that POY isn't clipping on OM2, the second steering mirror after the SRM, and they'll confirm that POX comes out of the chamber nicely, and that POP is also still coming out (by putting the green laser pointer back on that table, and making sure the green beam is co-aligned with the beam from PR2-PR3.  Also on the list is checking the vertex oplevs.  Steve and Manasa did some stuff with the ETM oplevs yesterday, but haven't had a chance to write about it yet.

We were trying to check POY alignment using the green laser in the reverse direction (outside vacuum to in-vac) . The green laser was installed along with a steering mirror to steer it into the ITMY chamber pointing at POY.

We found that the green laser did follow the path back into the chamber perfectly; it was clipping at the edge of POY. To align it to the center of POY (get a narrower angle of incidence at the ITMY), the green laser had to be steered in at a wider angle of incidence from the table. This is now being limited by the oplev steering optics on the table. We were not able to figure out the oplev path on the table perfectly; but we think we can find a way to move the oplev steering mirrors that are now restricting the POY alignment.

The oplev optics will be moved once we confirm with Jenne or Steve.

 

[Steve, Manasa]

We aligned the ETM oplevs yesterday. We confirmed that the oplev beam hit the ETMs. We checked for centering of the beam coming back at the oplev PDs and the QPDsums matched the values they followed before the vent.

Sadly, they have to be checked once again tomorrow because the alignment was messed up all over again yesterday.

  7690   Thu Nov 8 20:54:08 2012 Manasa, AyakaUpdateAlignmentReconfirming on IPPOS, IPANG and oplevs centering

Quote:

" We found that IPANG was not on its photodiode, but determined that it was centered on all of the in-vac mirrors, and that it was just a little bit of steering on the ETMY end out-of-vac table that needed to be done."

Manasa took photos of all test mass chambers and the BS chamber, so we can keep up-to-date CAD drawings. 

Oplevs and IPPOS/IPANG are being centered as I type.  Manasa and Ayaka are moving the lens in front of IPANG such that we have a slightly larger beam on the QPD.

 

The lens in front of IPANG on the out-of-vac table was moved to get a larger beam giving reasonable signals at the QPD.

IPPOS did not need much adjustment and was happy at the center of the QPD.

All oplevs but the ETMY were close to the center. I had to move the first steering mirror about half an inch on the out-of-vac table to catch the returning oplev beam from ETMY and direct it to the oplev PD.

* We have taken reasonable amount of in-vac pictures of ETM, ITM and BS chambers to update the CAD drawing.

 

  7810   Tue Dec 11 11:40:07 2012 Manasa, AyakaUpdatePSLPMC drift

[Manasa, Ayaka]

I found that MC got unstable this morning. This is caused by the drift of PMC. The transmission of PMC was going down and eventually unlocked PMC.

PMCdrift121211.pdf

We adjusted 'Slow Actuator Adjust' in FSS and now the PMC is locked with transmission of ~ 0.735.
Also we aligned the MC to be locked. Now it is locked with transmission of ~ 0.5 with WFS and MCL on.

  7090   Mon Aug 6 11:07:06 2012 ManasaUpdate40m UpgradingOptical layout updated

ACAD files of the 40m optical layout have been updated as per the vent in Aug 2011.

Files are available at the 40m svn docs-->Upgrade12-->Opt_Layout2011.

 

  7122   Wed Aug 8 19:54:06 2012 ManasaConfigurationIOOMC trans optics configured

Jan and I wanted to measure the ringdown at the IMC. Since the QPD at the MC trans is not fast enough for ringdown measurements, we decided to install a pickoff to include a faster PD while not disturbing much of the current MC trans configuration. The initial configuration had very little space to accommodate the pickoff. So the collimating lens along with the QPD were moved 2 inches closer to the incoming beam. A 50-50 BS was put in front of the QPD and the steering mirror was moved behind to reflect MC trans output to the new PD. The current configuration is shown below with the MC autolocker threshold mentioned in Jenne's elog

Pic1.png

The hunt for a faster PD wasn't satisfactory and we found a couple of PDs that were good for measurements actually didn't work after installing them. The one currently installed is also not satisfactorily fast enough for ringdown measurements. We'll hunt for faster PDs at Bridge tomorrow and replace PDA400. Also the IMC unlocked from time to time....may be we were noisy and didn't master the 'interferometer walk' very well.

 

 

  7125   Wed Aug 8 20:51:56 2012 ManasaUpdate40m UpgradingOptical layout updated

Quote:

ACAD files of the 40m optical layout have been updated as per the vent in Aug 2011.

Files are available at the 40m svn docs-->Upgrade12-->Opt_Layout2011.

 

 To ease the pain of hunting files, optical layout ACAD files have been moved to a new directory 40M_Optical Layout in the repository. Relevant files from directories Upgrade12 and upgrade 08 will be moved to "40M_Optical Layout" very soon and eventually these old directories will be removed. 

  7127   Wed Aug 8 22:17:43 2012 ManasaConfigurationIOOMC trans optics configured

Quote:

  The PDA255 is a good ringdown detector - Steve can find one in the 40m if you ask him nicely.

 We found a PDA255 but it doesn't seem to work. I am not sure if that is one you are mentioning...but I'll ask Steve tomorrow!

  7140   Fri Aug 10 09:54:51 2012 ManasaConfigurationIOOMC trans optics configured

Quote:

Quote:

  The PDA255 is a good ringdown detector - Steve can find one in the 40m if you ask him nicely.

 We found a PDA255 but it doesn't seem to work. I am not sure if that is one you are mentioning...but I'll ask Steve tomorrow!

 I double checked the PDA255 found at the 40m and it is broken/bad. Also there was no success hunting PDs at Bridge. So the MC trans is still in the same configuration. Nothing has changed. I'll try doing ringdown measurements with PDA400 today.

  7144   Fri Aug 10 15:05:52 2012 ManasaConfigurationIOOMC trans optics configured

Quote:

Quote:

Quote:

Quote:

  The PDA255 is a good ringdown detector - Steve can find one in the 40m if you ask him nicely.

 We found a PDA255 but it doesn't seem to work. I am not sure if that is one you are mentioning...but I'll ask Steve tomorrow!

 I double checked the PDA255 found at the 40m and it is broken/bad. Also there was no success hunting PDs at Bridge. So the MC trans is still in the same configuration. Nothing has changed. I'll try doing ringdown measurements with PDA400 today.

Can you explain more what "broken/bad" means?  Is there no signal?  Is it noisy?  Glitch?  etc.

 The PD saturates the oscilloscope just by switching on the power; with no real signal at all. But Steve helped locating a PD that is not being used at the AP table. So I will check it and replace the current one if it works!

  7159   Mon Aug 13 12:17:41 2012 ManasaConfigurationIOOPD from AP table removed

The PD (pda255) at the AP table, close to the MC refl , which Steve mentioned to be not in use, has been removed from the table for testing.

  7164   Mon Aug 13 19:29:10 2012 ManasaSummary Ringdown measurements

I tried to make ringdown measurements at the IMC using the DC falling edge as the trigger. Input to the MC was switched off by changing the polarity of the MC servo. But it does not seem to give the needed data as there seem to be several DC falling edges as soon as the polarity is switched. We should think about a better trigger or try to setup data recording from the oscilloscope seamlessly.

Also an ethernet cable has been connected to the router from the oscilloscope at the MC trans, but has not been set up to record data yet.

  7200   Wed Aug 15 20:53:48 2012 ManasaUpdateIOORingdown measurements

Quote:

Quote:

While I thought that the bumps observed at the end of the ringdown might be because of the cavity trying to lock itself, Jan commented that they have always existed in these measurements and their source is not known yet.

What I meant to say was that in all ringdown measurements that we observed today, the bumps were consistently part the ringdown, and that I have no explanation for the bumps. It should also be mentioned that fitting the bumpy part of the ringdown instead (we used the clean first 10us), the ringdown time is about twice as high. In either case, the ringdown time is significantly smaller than we have seen in documents about previous measurements.

We (basically I) also made one error when producing the plots. The yaxis label of the semi-logarithmic plot should be log(...), not log10(...).

 I thought about  why we do not find any bumps beyond the exponential fall. Could it be because we neglected fluctuations of voltage in the negative direction and plotted the absolute values?

  7205   Thu Aug 16 16:44:55 2012 ManasaConfigurationIOOPD from AP table removed

Quote:

The PD (pda255) at the AP table, close to the MC refl , which Steve mentioned to be not in use, has been removed from the table for testing.

 The PD installed at MC trans to make ringdown measurements has been replaced with the above PDA255. 

  7206   Thu Aug 16 17:28:51 2012 ManasaConfigurationIOOMC trans optics configured

Quote:

Quote:

Quote:

Quote:

Quote:

  The PDA255 is a good ringdown detector - Steve can find one in the 40m if you ask him nicely.

 We found a PDA255 but it doesn't seem to work. I am not sure if that is one you are mentioning...but I'll ask Steve tomorrow!

 I double checked the PDA255 found at the 40m and it is broken/bad. Also there was no success hunting PDs at Bridge. So the MC trans is still in the same configuration. Nothing has changed. I'll try doing ringdown measurements with PDA400 today.

Can you explain more what "broken/bad" means?  Is there no signal?  Is it noisy?  Glitch?  etc.

 The PD saturates the oscilloscope just by switching on the power; with no real signal at all. But Steve helped locating a PD that is not being used at the AP table. So I will check it and replace the current one if it works!

Koji opened up the PD and found that the screw connecting the PD to the pole was doing an additional job as well; connecting the power cable to the PD output in the inside. The PD is now fixed! Yippie...we have two PDA255 s at 40m now!!

  7222   Fri Aug 17 18:49:55 2012 ManasaUpdate40m UpgradingOptical layout updated

Quote:

Quote:

ACAD files of the 40m optical layout have been updated as per the vent in Aug 2011.

Files are available at the 40m svn docs-->Upgrade12-->Opt_Layout2011.

 

 To ease the pain of hunting files, optical layout ACAD files have been moved to a new directory 40M_Optical Layout in the repository. Relevant files from directories Upgrade12 and upgrade 08 will be moved to "40M_Optical Layout" very soon and eventually these old directories will be removed. 

Changes mentioned by Koji and Steve have been updated to the files (except for the cable connector which have been added but whose part number has to be found to match accurately with the current layout). The file in the directory should now match the current setup after the last vent Aug 2011.

Let me know if you find any mismatch between the current setup and the layout.

Plans about new installations/reconfiguration during the new vent will be carried out in a separate file.

  7245   Tue Aug 21 18:23:58 2012 ManasaUpdate40m UpgradingETMX table layout

Optical layout of the current endtable at ETMX has been updated in the svn repository (directory: 40M_Optical Layout). This layout will help in redesigning the table for the proposed replacement.

Some part numbers of mounts/optics are missing and will be updated once I find them. If you find anything wrong with the layout, do let me know.

 

  7256   Thu Aug 23 12:17:39 2012 ManasaUpdate IMC Ringdown

The ringdown measurements are in progress. But it seems that the MC mirrors are getting kicked everytime the cavity is unlocked by either changing the frequency at the MC servo or by shutting down the input to the MC. This means what we've been observing is not the ringdown of the IMC alone. Attached are MC sus sensor data and the observed ringdown on the oscilloscope.  I think we need to find a way to unlock the cavity without the mirrors getting kicked....in which case we should think about including an AOM or using a fast shutter before the IMC.

P.S. The origin of the ripples at the end of the ringdown still are of unknown origin. As of now, I don't think it is because of the mirrors moving but something else that should figured out.

Attachment 1: mozilla.pdf
mozilla.pdf
Attachment 2: MC_sus.pdf
MC_sus.pdf
  7260   Thu Aug 23 17:51:25 2012 ManasaUpdate IMC Ringdown

Quote:

 

 It is HIGHLY unlikely that the IMC mirrors are having any effect on the ringdown. The ringdowns take ~20 usec to happen. The mirrors are 0.25 kg and you can calculate that its very hard to get enough force to move them any appreciable distance in that time.

The huge kick observed in the MC sus sensors seem to last for ~10usec; almost matching the observed ringdown decay time. We should find a way to record the ringdown and the MC sus sensor data simultaneously to know when the mirrors are exactly moving during the measurement process. It could also be that the moving mirrors were responsible for the ripples observed later during the ringdown as well.

* How fast do the WFS respond to the frequency switching (time taken by WFS to turn off)? I think this information will help in narrowing down the many possible explanations to a few.

  7330   Fri Aug 31 17:44:21 2012 ManasaUpdateRingdownData

Quote:

Ok, so the whole idea that mirror motion can explain the ripples is nonsense. At least, when you think off the ringdown with "pump off". The phase shifts that I tried to estimate from longitudinal and tilt mirror motion are defined against a non-existing reference. So I guess that I have to click on the link that Koji posted...

Just to mention, for the tilt phase shift (yes, there is one, but the exact expression has two more factors in the equation I posted), it does not matter, which mirror tilts. So even for a lower bound on the ripple time, my equation was incorrect. It should have the sum over all three initial tilt angles not only the two "shooting into the long arms" of the MC.

Quote:

Laser frequency shift = longitudinal motion of the mirrors

Ringing: http://www.opticsinfobase.org/ol/abstract.cfm?uri=ol-20-24-2463

Quote:

Hmm. I don't know what ringing really is. Ok, let's assume it has to do with the pump... I don't see how the pump laser could produce these ripples. They have large amplitudes and so I always suspected something happening to the intracavity field. Therefore I was looking for effects that would change resonance conditions of the intracavity field during ringdown. Tilt motion seemed to be one explanation to me, but it may be a bit too slow (not sure yet). Longitudinal mirror motion is certainly too slow. What else could there be?

 

 

It is essential we take a look at the ringdown data for all measurements made so far to figure out what must be done to track the source of these notorious ripples. I've attached the plot for the same showing the decay time to be the same in all cases. About the ripples; it seems unlikely to both Jan and me that the ripples are some electronic noise because the ripples do not follow any common pattern or time constant. We have discussed with Koji about monitoring the frequency shift, the input power to the MC and also try other methods of shutting down the pump to track their source as the next steps.

 

cum_plot.png 

  7336   Tue Sep 4 13:44:17 2012 ManasaUpdateGeneralThe Plan

Quote:

We need a plan for the rest of the week.  I want to be closing the heavy doors on Friday at the latest.  Please add to / comment on this list!

 

Tues

* Lock MICH to get BS, ITMs aligned well

* Check if beam is hitting center of ITMs. 

* Check for clipping around BS

     - Use Watek in-vac to look at beam at all 4 BS ports - make sure no clipping going into BS, after BS in the michelson, or the AS port

* Try to get arms to flash??

* Prepare glass beam dumps??

Wed

* IPPOS / IPANG - make sure beam gets out of chambers (this may require opening ETMY)

* Jan take photos of ETMX scattering setup

* Manasa take in-vac photos of all tables, for table layouts

* Install glass beam dumps?

* If ETMY open, install glass baffle

* ????

Thurs

* ????

* Check table levelling one last time on all tables. 

Fri

* Close all heavy doors.  (Access connector, ITMX, ITMY, BS, ETMX, ETMY? )

* Start at ~10am?

Mon (if not Fri)

* Start pumping

Wed

* Jan/Manasa - Measure transmission of viewport at ETMX

 

 

 

 

 

  7354   Thu Sep 6 19:21:58 2012 ManasaConfiguration40m UpgradingBaffle problem

For the current baffle (dia. 40mm) centered along the beamline place at 1.77" from the test mass, the baffle will allow ~8.6mm visibility on the camera from the center of the test mass (in case of ETMY).

*assuming the pick off mirror is placed at the edge of the tunnel

Attachment 1: bfl.png
bfl.png
  7359   Fri Sep 7 11:58:12 2012 ManasaConfiguration40m UpgradingBaffle problem

Quote:

The required diameter for the baffle if it sits on the cage at 1.77" from the test masses: the current baffle (dia. 40mm) centered along the beamline, will allow ~8.6mm visibility from the center of the test mass (in case of ETMY).

*assuming the pick off mirror is placed at the edge of the tunnel

Estimations of the visibility region (r1 on the test mass) with baffle (aperture size 40mm).

The baffle is installed on the cage at 1.125" from the test mass (distance changed from the previous elog after a double check).

The 40mm aperture is in no way going to help get clear view of the ITMs; 

Attachment 1: bfl.png
bfl.png
  7361   Fri Sep 7 13:01:53 2012 ManasaConfiguration40m UpgradingBaffle problem

Quote:

Quote:

The required diameter for the baffle if it sits on the cage at 1.77" from the test masses: the current baffle (dia. 40mm) centered along the beamline, will allow ~8.6mm visibility from the center of the test mass (in case of ETMY).

*assuming the pick off mirror is placed at the edge of the tunnel

Estimations of the visibility region (r1 on the test mass) with baffle (aperture size 40mm).

The baffle is installed on the cage at 1.125" from the test mass (distance changed from the previous elog after a double check).

The 40mm aperture is in no way going to help get clear view of the ITMs; 

Required baffle diameter to have a visibility region r1 = 3 times the beam diameter

Picture1.png

  7403   Tue Sep 18 20:32:42 2012 ManasaConfigurationPSLAOM installation

 {Jan, Manasa}

We tried towards calibrating the RF driver of the AOM. We decided to use the normal power supply for both the driver control voltage and the ALC voltage.  But we could not figure out the type of the ALC port to find a compatible mating connector...it did not match with SMA, SMB or SMP. Finally I wrote to the company and got to know it is a filtered feed through. Now that we know how to control the ALC voltage, we will try looking at the signal for varying ALC voltage and see how that goes. 

But when we tried to see the 2W RF signal through the RF scope, with ALC open, we found that the RF signal was distorted and did not measure 80MHz.  It was lame that we did not get a snapshot 

P.S. The AOM has been left disconnected from the RF driver. 

  7409   Wed Sep 19 11:39:37 2012 ManasaConfigurationPSLAOM installation

Quote:

 {Jan, Manasa}

We tried towards calibrating the RF driver of the AOM. We decided to use the normal power supply for both the driver control voltage and the ALC voltage.  But we could not figure out the type of the ALC port to find a compatible mating connector...it did not match with SMA, SMB or SMP. Finally I wrote to the company and got to know it is a filtered feed through. Now that we know how to control the ALC voltage, we will try looking at the signal for varying ALC voltage and see how that goes. 

But when we tried to see the 2W RF signal through the RF scope, with ALC open, we found that the RF signal was distorted and did not measure 80MHz.  It was lame that we did not get a snapshot 

P.S. The AOM has been left disconnected from the RF driver. 

 {Jan, Manasa}

We started again to calibrate the RF driver. We connected the ALC to the power supply and observed the output RF power on the scope. The RF power did change with ALC voltage, but the RF signal still seems not to be operating at 80MHz 

There is some kind of additional disturbance to the waveform at 80MHz (the frequency of just the waveform with tall peaks or small peaks alone). We made sure we get a snapshot this time!! I am not sure if it will be safe to feed this RF signal to the AOM as such

ALC_25.png