40m QIL Cryo_Lab CTN SUS_Lab CAML OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 279 of 355  Not logged in ELOG logo
ID Date Author Type Categoryup Subject
  4496   Thu Apr 7 11:38:56 2011 steveUpdatePSLenclosure windows on the east side of the PSL

The PSL enclosure now have 4 windows on each side. The bottom rail guides on the east side will be replaced by  one  U-channel for smoother, more gentle sliding.

Door position indicator- interlock switches are not wired yet.

Attachment 1: P1070538.JPG
P1070538.JPG
  4633   Thu May 5 10:19:22 2011 steveUpdatePSLPSL-FSS_RMTEMP is back

Valera and I installed the the temp sensor and the interface box that Rana fixed. This may help with diagnosing the PSL drift.

Attachment 1: P1070637.JPG
P1070637.JPG
  4635   Thu May 5 13:12:06 2011 JenneUpdatePSLTemperature drift, trying to shield PMC with foil

eeek.  I've been running around all day, so this is an incomplete elog.  I'll fill in more stuff in the next hour or so, but just to let people know what's going on:

[Valera, Jenne]

Valera noticed that lots of things in and around the PSL table are drifting with temperature.  This is why he and Steve installed a temp sensor on the table earlier today.

PSL_Temp_drift_5May2011.png

Since the alignment into the PMC, and also the alignment downstream of the PMC have been drifting in angle, we supposed that it might be the PMC itself which is changing somehow with temperature.  We don't have a good idea of how exactly it is sensitive to temperature, but we're working on figuring it out.

Round 1 of testing:  We put a foil hat over the PMC to shield it from the HEPA air blowing directly down on top of it.  I made sure that the foil is also covering the PZT and the metal ring at the end of the PMC, because this could potentially be the problem (metal is usually more temperature sensitive than glass, or the PZT itself could be changing, either of which could make the end mirror twist, and change the alignment of the PMC).  We'll see later if this did anything useful or not. 

I have photos of the aluminum foil setup, which I will post later when I get back to the lab after teaching. 

P5050092small.jpg

  4641   Thu May 5 15:05:06 2011 steveUpdatePSLPSL-FSS_RMTEMP is not back

Quote:

Valera and I installed the the temp sensor and the interface box that Rana fixed. This may help with diagnosing the PSL drift.

 I was wrong. Rana did not fix the interface box. I removed the interface box and turned down the HEPA flow  from 100 to 20% on the Variac.

Attachment 1: rtnfxd.jpg
rtnfxd.jpg
  4648   Thu May 5 20:47:41 2011 KojiUpdatePSLPMC aligned

The PMC exhibited the reduction of the transmission, so it was aligned.

The misalignment was not the angle of the beam but the translation of the beam in the vertical direction
as I had no improvement by moving the pitch of one mirror and had to move those two differentially.

This will give us the information what is moving by the temperature fluctuation or whatever.

  4662   Sun May 8 22:59:40 2011 ranaUpdatePSLPSL reference cavity temperature box modifications

I looked at the PSL temperature box. It started out as D980400-B-C. Then it was revised by Peter King as per the LHO mods E020247.

There are some more things to do to it to make it useful for us:

  1. R3, 4, 7, 8, 12, & 13 should be changed from 1k to 0 Ohms, I think. I cannot figure out their purpose.
  2. All resistors should be made metal-film. Right now, its kind of a mish-mash.
  3. The active filter U6B has a corner frequency of ~50 Hz. This seems not useful for keeping the 4116 DAC noise out of the temperature. We should lower this to ~30 mHz to take advantage of the stability of the LT1021 which was put in.

** Frank reminds me that we don't use the TIdal or VME external inputs anymore since we moved to the EPICS/Perl PID control. So all we have to do is make sure these inputs are hardware disabled/disconnected.

  4663   Mon May 9 09:37:51 2011 valeraUpdatePSLPSL and MC trends

The attached plot shows 7 day trends of the MC and PMC power levels, PSL QPDs, and temperature. The MC stayed locked for ~40 hours over the weekend. The temperature swings were somewhat smaller over the past couple of days but one should remember to turn the PSL HEPA down after working on the table. Steve turned the HEPA flow from 100% down to 20% on Thursday and posted the reminder signs on the PSL enclosure.

Attachment 1: pslmcdrift2.pdf
pslmcdrift2.pdf
  4666   Mon May 9 15:21:36 2011 josephbUpdatePSLFixed channel names for PSL QPDs, fixed saturation, changed signs

[Valera, Joe]

Software Changes:

First we changed all the C1:IOO-QPD_*_* channels to C1:PSL-QPD_*_* channels in the /cvs/cds/caltech/target/c1iool0/c1ioo.db file, as well as the /opt/rtcds/caltech/c1/chans/daq/C0EDCU.ini file.  We then rebooted the frame builder via "telnet fb 8087" and then "shutdown".

This change breaks continuity for these channels prior to today.

The C1:PSL-QPD_POS_HOR and C1:PSL-QPD_POS_VERT channels were found to be backwards as well.  So we modified the /cvs/cds/caltech/target/c1iool0/c1ioo.db file to switch them.

Lastly, we changed the ASLO and AOFF values for the C1:PSL-QPD_POS_SUM and the C1:PSL-QPD_ANG_SUM so as to provide positive numbers.  This was done by flipping the sign for each entry.

ASLO went from 0.004883 to -0.004883, and AOFF when from -10 to 10 for both channels.

Hardware Changes:

The C1:PSL-QPD_ANG_SUM channel had been saturated at -10V.  Valera reduced the power on the QPD to drop it to about 4V by placing an ND attenuator in the ANG QPD path.

  4672   Mon May 9 20:30:20 2011 ranaUpdatePSLPSL reference cavity temperature box modifications

I re-installed the box (@ ~8:15) after reflowing some of the solder joints. I will observe it over night and then remove the 1K resistors. Attached is a 8 hour minute-trend.

Attachment 1: Untitled.png
Untitled.png
  4677   Tue May 10 10:06:23 2011 josephbUpdatePSLFixed channel names for PSL QPDs, fixed saturation, changed signs

I added calculation entries to the /cvs/cds/caltech/target/c1iool0/c1ioo.db file which are named C1:IOO-QPD_*_*, as the channels were originally named.  These calculation channels have the identical data to the C1:PSL-QPD_*_* channels.  I then added the channels to the C0EDCU.ini file, so as to once again have continuity for the channels, in addition to having the newer, better named channels.

The c1iool0 machine ("telnet c1iool0", "reboot") and the framebuilder process ("telnet fb 8087", "shutdown") were both restarted after these changes.

These channels were brought up in dataviewer and compared.  The approriate channels were identical.

Quote:

[Valera, Joe]

Software Changes:

First we changed all the C1:IOO-QPD_*_* channels to C1:PSL-QPD_*_* channels in the /cvs/cds/caltech/target/c1iool0/c1ioo.db file, as well as the /opt/rtcds/caltech/c1/chans/daq/C0EDCU.ini file.  We then rebooted the frame builder via "telnet fb 8087" and then "shutdown".

This change breaks continuity for these channels prior to today.

 

  4678   Tue May 10 11:14:44 2011 steveUpdatePSLPSL-QPD_ANG qpd optimized

Valera and I placed F 572.7 mm lens ~15 cm away  from the ang  qpd (in the same mount with ND filter) so that two qpds see different combination of ang and pos motion - there was no lenses prior to this change. The beam diameter is reduced to ~half .

Attachment 1: qpdangl.jpg
qpdangl.jpg
  4681   Tue May 10 20:57:05 2011 ranaUpdatePSLPSL reference cavity temperature box modifications : after 24 hours still OK

Quote:

I re-installed the box (@ ~8:15) after reflowing some of the solder joints. I will observe it over night and then remove the 1K resistors. Attached is a 8 hour minute-trend.

 I compared this 24 hour trend with the one from this day exactly one year ago. Seems the same, so now I can make the resistor change.

Attachment 1: Untitled.png
Untitled.png
  4697   Thu May 12 00:59:45 2011 ryan, ranaUpdatePSLReturn of the PSL temperature box

The PSL temperature box has returned to service, with some circuit modifications. The 1k resistors on all the temp. sensor inputs (R3, R4, R7, R8, R12, R12) were changed to 0 Ohm. Also, the 10k resistors R26, R28, R29, and R30 were changed to 10.2k metal film. The DCC document will be updated shortly. There is now an offset in the MINCOMEAS channel compared to the others, which will be corrected in the morning after looking at the overnight trend.

  4798   Thu Jun 9 10:38:30 2011 steveUpdatePSLPMC needs help

The PMC is losing power.

Attachment 1: pmcnothp.jpg
pmcnothp.jpg
  4799   Thu Jun 9 12:21:07 2011 kiwamuUpdatePSLRe : PMC needs help

Nulling the slow actuation offset fixed the issue. Now PMC is back to normal.

Untitled.png

The reflected beam on the CCD was quite symmetric (it looked very TEM00 mode !) for some reasons, I somehow suspected the mode matching to PMC.

One possibility I thought of was the laser temperature because it could change the laser spatial mode.

So I looked at the slow actuation offset on the FSS screen and found it was at -4.0 which sounds somewhat big.

Then I zeroed the offset by the slider and relocked PMC.

Then the spatial pattern of the reflected beam became usual (i.e. junk light looking) and the transmitted light wet up to 0.83 which is normal.

Quote from #4798

The PMC is losing power.

 

  4817   Tue Jun 14 16:38:31 2011 JamieUpdatePSLTweaked input pointing to PMC

The PMC trans power was a little low (0.77V or so).  I tweaked up the input pointing and now we're getting about 0.875V transmitted.

  4865   Thu Jun 23 10:17:49 2011 steveUpdatePSLAP table is open to PSL again

Access to the north side of the PSL table is blocked by the    8" beam guard. This opens the beam pathways between them.

Attachment 1: P1070911.JPG
P1070911.JPG
Attachment 2: P1070910.JPG
P1070910.JPG
  4870   Thu Jun 23 22:39:34 2011 JenneUpdatePSLPMC found unlocked

I found the PMC unlocked.  Koji noticed that the FSS Slow Actuator Adjust was railed at the positive end of the slider.  I set it close to zero, and relocked the PMC.  The FSS slow loop servo is doing its thing, and the PMC and MC are now locked. 

  4880   Fri Jun 24 21:21:46 2011 SureshUpdatePSLPlaced labels on the zig-zag mirrors on PSL table

I put labels on the pair of beam steering mirrors which are at the output end of the PSL table.  I had changed one of these mirrors (elog) and Jenne had changed the other (elog).  This was at about 3PM today

I just learned from Kiwamu that this has messed up the MC alignment.

 

  4980   Sun Jul 17 18:23:23 2011 JenneUpdatePSLFound the PMC unlocked

It was unlocked since ~4:30am.  No idea why.  It's relocked so I can try round N of measuring the PRC length.

Attachment 1: PMCunlocked_17July2011.png
PMCunlocked_17July2011.png
  4985   Mon Jul 18 21:06:32 2011 PSL Table GuardianOmnistructurePSLDon't leave the PSL table open, unattended!!!!!!!!!!!!11111

I found the PSL table left open, and unattended again. 

As far as I know, Jamie and Jenne (working on the LSC rack, so no lasers / optics work involved) have been the only ones in the IFO room for several hours now. 

I'm going to start taking laser keys, or finding other suitable punishments.  Like a day of lab cleanup chores or something.  Seriously, don't leave the PSL table open if you're not actively working on it.

  5043   Wed Jul 27 10:10:12 2011 steveUpdatePSLlocking status

 80 days: PMC is drifting

Attachment 1: 80dpmcMC.jpg
80dpmcMC.jpg
  5057   Thu Jul 28 19:49:12 2011 SonaliUpdatePSLPMC trans beam aligned.

Kiwamu and I aligned the PMC transmitted beam the incident beam going to PMC today.

I learnt how to lock the PMC using the digital controls.

  5064   Sat Jul 30 00:33:33 2011 KojiUpdatePSLABSL Laser crystal temp left largely excited & left unattended for more than 3hours

I found that the ref cav trans CCD view was blinking with 30-50 fringe amplitudes. This meant the laser freq was swinging ~50GHz.

I checked the ABSL laser and the SG out of a lock-in amplifier was connected to the slow input.

This was shaking the laser temp from 29degC to 46degC. This was the cause of the fringe swinging.
This big excitation changing the output power too as the temp was changed across it mode-hop region.

I have disconnected the excitation from the laser no matter how useful experiments were took place as there was no e-log entry about this.

I need the explanations

1. Why our precious laser is exposed to such a large swing of temperature?

2. Why the excitation is left like that without any attendance?

3. Why there was no elogging about this activity?

  5065   Sat Jul 30 02:47:43 2011 ranaUpdatePSLABSL Laser crystal temp left largely excited & left unattended for more than 3hours

 Hmm. Should have only been +/- 1 GHz. Some setting got changed apparently...

This is a part of the RefCav temperature measurement setup. You'll get an elog from Jenny very soon.

  5069   Sat Jul 30 10:01:35 2011 JennyUpdatePSLPSL table work

I've been working on the PSL table to put together a setup so that I can measure the reference cavity's response to a temperature step increase at the can surrounding it. My first step was to mode match the beam coming from the AP table to the cavity.

I implemented my mode matching solution. I ended up using a different one from the one I last elogged about. Here is the solution I used:

Two lenses: f = 1016.7.6 mm at -0.96 m and f = 687.5 mm at -0.658 m. (I set my origin at the polarizing beam splitter--the spot where I want my beam to match the beam coming from the PMC, so all waists are behind that point). Below is what it should look like.

modematchpic.pngmodematchinfo.png

What I did on the table:

  • Before placing lenses I aligned the beam and added a 1/2-wave plate between the two polarizing beam splitters to change the polarization of the beam from S to P.
  • I aligned the beam so that it reflected off of the cavity opening (monitoring the reflected power with a photodetector connected to an oscilloscope and tweaking the alignment to maximize the reflected signal). 
  • I then placed the lenses at -0.93 and -0.64 mm because the exact spots were blocked by optics being used in another setup.
  • I reasoned that since the fitting for the initial waist is so uncertain, the lens position being off by a few cm will not produce the dominating source of error. I am now driving the laser frequency using a lock-in as a function generator to drive the laser temperature at ~1 Hz. I'm then monitoring the power transmitted by the reference cavity with a camera connected to a TV monitor. I will use this setup to improve my mode matching.

Here's a picture of the PSL table with the lenses and mirror I added. The beam is redirected by a mirror and then a polarizing beam splitter. Past the beam splitter is another lens (f=286.5 mm), which was already in place from the mode matching of the beam from the PMC to the reference cavity.

modematch_setup_pic.png

Here is a block diagram of my intended experimental setup:

LIGO_block_diagram.png

I am going to try to lock the laser to the cavity given my preliminary mode matching and then go back and improve it later. My next step is to find a frequency range for dithering the voltage sent to the PZT. To do this I will:

  • Measure the transfer function (amplitude response) of the PZT using a photodiode. The power outputted by the laser varies with driving frequency.
  • Find a frequency region in which the amplitude response is low.
Attachment 3: LIGO_block_diagram.png
LIGO_block_diagram.png
  5071   Sat Jul 30 19:06:25 2011 ryan, ranaUpdatePSLReturn of the PSL temperature box

Quote:

The PSL temperature box has returned to service, with some circuit modifications. The 1k resistors on all the temp. sensor inputs (R3, R4, R7, R8, R12, R12) were changed to 0 Ohm. Also, the 10k resistors R26, R28, R29, and R30 were changed to 10.2k metal film. The DCC document will be updated shortly. There is now an offset in the MINCOMEAS channel compared to the others, which will be corrected in the morning after looking at the overnight trend.

 Forgot to do this in May. Have just changed the values in the psl.db file now as well as updating them live via Probe.

To make the appropriate change, I took the measured offset (5.31 deg) and added 2x this to the EGUF and EGUL field for the MINCO_MEAS channel. (see instructions here)

Committed the .db file to the SVN.

attached plot shows 8 days of trend with 5.31 degC added to the black trace using the XMGRACE Data Set Transformations

Attachment 1: rctempbox.png
rctempbox.png
  5072   Sat Jul 30 20:41:50 2011 ranaUpdatePSLRefCav Stabilization back on

 Untitled.png

I turned the RefCav heater and servo back on a couple days ago. At first it was stabilizing again at a low setpoint, but in reality the right temperature (~40 C). After fixing the in-loop signal offsets, the setpoint now correctly reflects the actual temperature.

Jenny is going to calibrate the sensors using some kind of dunking cannister next week.

  5075   Sun Jul 31 00:37:57 2011 kiwamuUpdatePSLRe : PSL table work

I think you made a simple mistake in your diagram -- the mixer must be replaced by a summer circuit. Otherwise you cannot do the PDH lock.

Quote from #5069

LIGO_block_diagram.png

  5096   Tue Aug 2 17:40:04 2011 JennyUpdatePSLReducing beam intensity incident on photodiode

I am using a PDA255 photodiode to measure the power outputted by the NPRO beam on the PSL table. (I'm going to then use a network analyzer to measure the amplitude response of the PZT to being driven at a range of frequencies. I'll detect the variation in in response to changing the driving frequency using this PDA255.)

The PDA255 has an active area of 0.8mm^2 and a maximum intensity for which the response is linear of 10mW/cm^2. This means that a beam I focus on the PD must have a power less than 0.08 mW (and even less if the spot size is smaller than the window size).

I used a power meter to measure the beam power and found it was 0.381 mW.

The second polarizing beam splitter in the setup transmits most of the beam power, but reflects 0.04 mW (according to the power meter). I'm going to place the photodiode there in the path of the reflected beam.

  5114   Thu Aug 4 00:04:52 2011 JennyUpdatePSLNetwork analyzer and PD set up to measure amplitude response of PZT

Today I placed the PDA255 photodiode on the PSL table to catch the small amount of beam power reflected by the second polarizing beam splitter in my setup. I plugged the PD output to the oscilloscope to measure the voltage output and positioned the PD such that the voltage output was maximized. At best I was able to achieve a 300 mV DC output voltage from the PD, (which seems a bit low, as the PD is specified to go from 0 to 5 V and the specifications say that the response becomes nonlinear after 10 mW/cm^2 and my beam has an intensity of approximately 5 mw/cm^2. I would therefore expect to get more beam power but after over an hour of maneuvering, 300 mV was the highest voltage output I could get).

I am planning, tomorrow afternoon, to take a measurement of the amplitude response of the PZT driving the NPRO laser. I moved the 4395 spectrum/network analyzer to near the PSL table and connected the RF output to an RF splitter. I fed one output of that into the PZT and the other output into the R port on the network analyzer. I fed the PD output into the A port. I plan to measure A/R as a function of driving frequency, sweeping from 10 Hz to 30 mHz.

I also worked to improve the mode matching of the NPRO beam coming from the AP table to the reference cavity. I drove the temperature of the NPRO at 0.100 Hz with an amplitude of 0.300 V, which Koji told me corresponds to a 1GHz change in the laser frequency. The transmission from the cavity is being monitored by a camera connected to a TV monitor, and also by a PD connected to an oscilloscope. I then repositioned the second lens in my mode matching setup in an attempt to increase the transmission peaks from the zeroth order spacial mode and decrease the transmission peaks from higher order modes. I may have improved the mode matching slightly but I was unable to improve it significantly.

  5126   Fri Aug 5 18:29:35 2011 JennyUpdatePSLNetwork analyzer and PD set up to measure amplitude response of PZT

Quote:

Today I placed the PDA255 photodiode on the PSL table to catch the small amount of beam power reflected by the second polarizing beam splitter in my setup. I plugged the PD output to the oscilloscope to measure the voltage output and positioned the PD such that the voltage output was maximized. At best I was able to achieve a 300 mV DC output voltage from the PD, (which seems a bit low, as the PD is specified to go from 0 to 5 V and the specifications say that the response becomes nonlinear after 10 mW/cm^2 and my beam has an intensity of approximately 5 mw/cm^2. I would therefore expect to get more beam power but after over an hour of maneuvering, 300 mV was the highest voltage output I could get).

I am planning, tomorrow afternoon, to take a measurement of the amplitude response of the PZT driving the NPRO laser. I moved the 4395 spectrum/network analyzer to near the PSL table and connected the RF output to an RF splitter. I fed one output of that into the PZT and the other output into the R port on the network analyzer. I fed the PD output into the A port. I plan to measure A/R as a function of driving frequency, sweeping from 10 Hz to 30 mHz.

I also worked to improve the mode matching of the NPRO beam coming from the AP table to the reference cavity. I drove the temperature of the NPRO at 0.100 Hz with an amplitude of 0.300 V, which Koji told me corresponds to a 1GHz change in the laser frequency. The transmission from the cavity is being monitored by a camera connected to a TV monitor, and also by a PD connected to an oscilloscope. I then repositioned the second lens in my mode matching setup in an attempt to increase the transmission peaks from the zeroth order spacial mode and decrease the transmission peaks from higher order modes. I may have improved the mode matching slightly but I was unable to improve it significantly.

The ABSL beam had been blocked so that it wouldn't enter the interferometer. I moved the block so that the beam I've been using is unblocked by the beam going to the interferometer is still blocked.

I positioned a fast lens (f=28.7mm) a little over an inch in front of the PDA255 in order to decrease the spot size incident on the PD. I adjusted the rotation angle of the half wave plate to maximize the transmitted power through the PBS to the cavity and minimize the power reflected to my PD. I then adjusted the lens potion to fix the beam on the PD. The voltage output of the PD is now 150mW, but I have the ability to increase the incident power by rotating the wave plate slightly.

Now all I need is to set up the network analyzer again to record the amplitude response to modulating the PZT from 10 Hz to 30 MHz, reduce the input voltage into the analyzer using a DC block.

  5144   Mon Aug 8 20:23:14 2011 JennyUpdatePSLNetwork analyzer and PD set up to measure amplitude response of PZT

Quote:

Quote:

Today I placed the PDA255 photodiode on the PSL table to catch the small amount of beam power reflected by the second polarizing beam splitter in my setup. I plugged the PD output to the oscilloscope to measure the voltage output and positioned the PD such that the voltage output was maximized. At best I was able to achieve a 300 mV DC output voltage from the PD, (which seems a bit low, as the PD is specified to go from 0 to 5 V and the specifications say that the response becomes nonlinear after 10 mW/cm^2 and my beam has an intensity of approximately 5 mw/cm^2. I would therefore expect to get more beam power but after over an hour of maneuvering, 300 mV was the highest voltage output I could get).

I am planning, tomorrow afternoon, to take a measurement of the amplitude response of the PZT driving the NPRO laser. I moved the 4395 spectrum/network analyzer to near the PSL table and connected the RF output to an RF splitter. I fed one output of that into the PZT and the other output into the R port on the network analyzer. I fed the PD output into the A port. I plan to measure A/R as a function of driving frequency, sweeping from 10 Hz to 30 mHz.

I also worked to improve the mode matching of the NPRO beam coming from the AP table to the reference cavity. I drove the temperature of the NPRO at 0.100 Hz with an amplitude of 0.300 V, which Koji told me corresponds to a 1GHz change in the laser frequency. The transmission from the cavity is being monitored by a camera connected to a TV monitor, and also by a PD connected to an oscilloscope. I then repositioned the second lens in my mode matching setup in an attempt to increase the transmission peaks from the zeroth order spacial mode and decrease the transmission peaks from higher order modes. I may have improved the mode matching slightly but I was unable to improve it significantly.

The ABSL beam had been blocked so that it wouldn't enter the interferometer. I moved the block so that the beam I've been using is unblocked by the beam going to the interferometer is still blocked.

I positioned a fast lens (f=28.7mm) a little over an inch in front of the PDA255 in order to decrease the spot size incident on the PD. I adjusted the rotation angle of the half wave plate to maximize the transmitted power through the PBS to the cavity and minimize the power reflected to my PD. I then adjusted the lens potion to fix the beam on the PD. The voltage output of the PD is now 150mW, but I have the ability to increase the incident power by rotating the wave plate slightly.

Now all I need is to set up the network analyzer again to record the amplitude response to modulating the PZT from 10 Hz to 30 MHz, reduce the input voltage into the analyzer using a DC block.

 I rolled the network analyzer over to the PSL table (on the south side). I'm borrowing the DC block from Kiwamu's green locking setup. I'm going to first measure the amplitude response of a low pass filter to made sure that the analyzer is outputting what I expect. Then I will measure the laser PZT amplitude response. I plan to finish the measurement and return the network analyzer to it's usual location tonight.

  5149   Tue Aug 9 02:34:26 2011 JennyUpdatePSLPZT transfer function measurement

Using a PDA255 on the PSL table, I measured the amplitude response of the NPRO PZT, sweeping from 10kHz to 5 MHz.

I took a run with the laser beam blocked. I then took three runs with the beam unblocked, changing the temperature of the laser by 10 mK between the first two runs and by 100mK between the second and third runs.

At the end of the night I turned off the network analyzer and unplugged the inputs. I'm leaving it near the PSL table, because I'd like to take more measurements tomorrow, probing a narrow bandwidth where the amplitude response is low.

On the PSL table, I'm still monitoring the reflected light from the cavity and the transmitted light through the cavity on the oscilloscope. I'm no longer driving the NPRO temperature with the lock-in.

I closed the shutter on the NPRO laser at the end of the night.

I'll log more details on the data tomorrow morning.

  5156   Tue Aug 9 16:00:58 2011 JennyUpdatePSLAmplitude response of PZT

AMresponsePZT.png

The top plot shows a sweep from 10 kHz to 5 MHz of the ratio of the voltage output of the PD detecting power from the NPRO laser beam and the RF source voltage (the magnitude of the complex transfer function). The black trace was taken with the laser beam blocked. For runs 2 and 3 I changed the laser temperature set point by 10 mK and 100 mK respectively to see if there was a significant change in the AM response. The bottom plots shows runs 2 and 3 compared to run 1 plotted in dB (to be explicit, i'm plotting 10 times the base 10 log of the magnitude of the ratio of two complex transfer functions). Changing the temperature seems to have only a minor effect on the output except at around 450kHz, where the response has a large peak in run 1 and much smaller peaks in runs 2 and 3. 

The traces in the top plot consist of 16 averages taken with a 300Hz IF bandwidth, 15 dBm source power (attenuated with a 6 dB attenuator) and with 20dB attenuation of the input power from the PD.

Next I'm going to probe a narrow band region where the response is low (2.0MHz or 2.4MHz perhaps) and choose a bandwidth for the dither frequency for the PDH locking.

Attachment 1: AMresponsePZT.png
AMresponsePZT.png
  5165   Wed Aug 10 02:40:40 2011 JennyUpdatePSLDither freq for PZT chosen: 2.418 MHz

I've finished using the network analyzer to characterize find a dither frequency for driving the PZT to use in my PDH locking. I found a region in which the amplitude response of the PZT is low: The dip is centered at 2.418 MHz. Changing the NPRO laser temperature by 100mK has no significant effect on the transfer function in that region. I will post plots tomorrow.

I'm finished with the network analyzer. It is unplugged, and the cart is still near the PSL table. (I'll roll it back tomorrow when it won't disturb interferometer locking).

I closed the shutter on the NPRO at the end of the night.

Tomorrow I plan to put together the fast locking setup. I'll drive the PZT at 2.418 MHz. More details to come tomorrow.

  5179   Wed Aug 10 20:40:17 2011 JennyUpdatePSLPDH locking: got an error signal

I ended up choosing a different dither frequency for driving the NPRO PZT: 230 kHz, because the phase modulation response in that region is higher according to other data taken on an NPRO laser (see this entry). At 230 there is a dip in the AM response of the PZT.

I am driving the PZT at 230 kHz and 13 dBm using a function generator. I am then monitoring the RF output of a PD that is detecting light reflected off the cavity. (The dither frequency was below the RF cutoff frequency of the PD, but it was appearing in the "DC output", so I am actually taking the "DC output" of the PD, which has my RF signal in it, blocking the real DC part of it with a DC block, and then mixing the signal with the 230kHz sine wave being sent to the PZT.

I am monitoring the mixer output on an oscilloscope, as well as the transmission through the cavity. I am sweeping the laser temperature using a lock in as a function generator sending out a sine wave at 0.2 V and 5 mHz. When there is a peak in the transmission, the error signal coming from the mixer passes through zero.

My next step is to find or build a low pass filter with a pole somewhere less than 100 kHz to cut out the unwanted higher frequency signal so that I have a demodulated error signal that I can use to lock the laser to the cavity.

 

  5202   Fri Aug 12 03:49:45 2011 JennySummaryPSLNPRO PDH-Locked to Ref Cav

DMass and I locked the NPRO laser (Model M126-1064-700, S/N 238) on the AP table to the reference cavity on the PSL table using the PDH locking setup shown in the block diagram below (the part with the blue background):

 

LIGO_block_diagram_2.png

 

A Marconi IFR 2023A signal generator outputs a sine wave at 230 kHz and 13 dBm, which is split. One output of the splitter drives the laser PZT while the other is sent to a 7dBm mixer. Also sent to the mixer is the output of a photodiode that is detecting the reflected power from off the cavity. (A DC block is used so that only RF signal from the PD is sent to the mixer). The output of the mixer goes through an SR560 low-noise preamp, which is set to act as a low pass filter with a gain of 5 and a pole at 30 kHz. That error signal is then sent to the –B port of the LB1005 PDH servo, which has the following settings: PI corner at 10kHz, LF gain limit of 50 dB, and gain of 2.7 (1.74 corresponds to a decade, so the signal is multiplied by 35). The output signal from the LB1005 is added to the 230 kHz dither using another SR560 preamp, and the sum of the signals drive the PZT.

 

I am monitoring the transmission through the cavity on a digital oscilloscope (not shown in the diagram) and with a camera connected to a TV monitor. I sweep the NPRO laser temperature set point manually until the 0,0 mode of the carrier frequency resonates in the cavity and is visible on the monitor. Then I close the loop and turn on the integrator on the LB1005.

 

The laser locks to the cavity both when the error signal is sent into the A port and when it is sent into the –B port of the PDH servo. I determined that –B is the right sign by comparing the transmission through the cavity on the oscilloscope for both ways.

 

When using the A port, the transmission when it was locked swept from ~50 to ~200 mV (over ~10 second intervals) but had large high frequency fluctuations of around +/- 50 mV. Looking at the error signal on the oscilloscope as well, the RMS fluctuations of the error signal were at best ~40 mV peak to peak, which was at a gain of 2.9 on the LB1005.

 

Using the –B port yielded a transmission that swept from 50 to 250 mV but had smaller high frequency fluctuations of around +/- 20 mV. The error signal RMS was at best 10mV peak to peak, which was at a gain of 2.7. (Although over the course of 10 minutes the gain for which the error signal RMS was smallest would drift up or down by ~0.1).

 

 

The open loop error signal peak-to-peak voltage was 180 mV, which is more than an order of magnitude larger than the RMS error signal fluctuations when the loop is closed, indicating that it is staying in the range in which the response is linear.

openlooperror.jpg

 

In the above plot the transmission signal is offset by 0.1 V for clarity.

Below is the closed loop error signal. The inset plot shows the signal viewed over a 1.6 ms time period. You can see ~60 microsecond fluctuations in the signal (~17 kHz)

closedlooperror.jpg

The system remained locked for ~45 minutes, and may have stayed locked for much longer, but I stopped it by opening the loop and turning off the function generator. Below is a picture of the transmitted light showing up on a monitor, the electronics I'm using, and a semi-ridiculous mess of wires.

 

IMG_3034.JPG

 

I determined that it’s not dangerous to leave the system locked and leave for a while. The maximum voltage that the SR560 will output to the PZT is 10Vpp. This means that it will not drive the PZT at more than +/-5 V DC. At low modulation rates, the PZT can take a voltage on the order of 30 Vpp, according to the Lightwave Series 125-126 user’s manual, so the control signal will not push the PZT too hard such that it’s harmful to the laser.

 

 

  5217   Fri Aug 12 20:33:57 2011 DmassSummaryPSLNPRO PDH-Locked to Ref Cav

To aid Jenny's valiant attempt to finish her SURF project, I did some things with the front end system over the last couple days, largely tricking Jamie into doing things for me lest I ruin the 40m RCG system. Several tribulations have been omitted.

We stole a channel in the frontend, in the proccess:

  1. Modified the C1GFD simulink model (now analog) to be "ADC -> TMP -> DAC" where TMP is a filter bank
    • C1GFD_TMP.adl (in /opt/rtcds/caltech/c1/medm/c1gfd) is the relevant part which connects the ADC to the DAC in the frontend
  2. Confirmed that the ADC was working by putting a signal in and seeing it in the frontend
  3. Could not get a signal out of the anti aliasing board
  4. Looked sad until Kiwamu found a breakout board for the SCSI cable coming from the DAC
  5. Used SR560 to buffer DAC output
    • drove a triangle wave with AWG into the TMP EXC channel (100 counts 1 Hz) and looked at it after the ~25 ft of BNC cable running between the DAC and the NRPO driver
    • wave looked funny (not like a triangle wave), maybe the DAC is not meant to push a signal so far, so added buffer
  6. Took the control signal going to the fast input of the NPRO driver (using the 500 Ohm SR560 output - see Jenny's diagram) and put it into the anti aliasing board of the ADC
  7. Added switchable integrator to filter bank with Foton
    • I couldn't get the names to display in the filter bank, so I looked sad again
    • Jamie and Koji both poked at the "no name displayed" problem but had no conclusions, so I decided to ignore it
    • I confirm that when the two filters were toggled "on" that the transfer function was as expected: simple integrator with a unity gain at ~10mHz - agrees with what Foton's Bode Plot tool says it should be (see attached DTT plot)
  8. I got Jamie to manually add the two epics channels from the TMP model to the appropriate .ini file so they would be recorded
    • C1:GFD-TMP_OUTPUT  (16 Hz)
    • C1:GFD-TMP_INMON    (16 Hz)
  9. RefCav heater servo seems to still be set up, so we can use existing channels:
    • C1:PSL-FSS_RCPID_SETPOINT (temp setpoint - will do +/-1C steps about 35 C)
    • C1:PSL-FSS_MINCOMEAS (In loop temp sensor - in C)
    • C1:PSL-FSS_RCTEMP (out of loop temp sensor - in C)
    • C1:PSL-FSS_TIDALSET (Voltage to heater - rails @ +/- 2V)
  10.  Closed loop on the control signal for the NPRO driver with an integrator, saw error signal go to zero
    • Turned up gain a little bit, saw some oscillations, then turned gain down to stop them, final gain = 2
  11. Left system on for Jenny to come in and do step responses
Attachment 1: TMP_INT_TF.pdf
TMP_INT_TF.pdf
  5228   Sun Aug 14 04:12:37 2011 JennyUpdatePSLTemperature steps and slow actuator railing

Below are some plots from dataviewer of temperature-step data taken over the past 32 hours. (They show minute trends). I am looking at the thermal coupling from the can surrounding the reference cavity on the PSL table to the cavity itself, and trying to measure the cavity temperature response via the control signal sent to heat the NPRO laser, which is locked to the cavity.

Picture_6.png

Picture_7.png

  • Top left: out-of-loop temperature sensor on can surrounding ref cav (RCTEMP)
  • Top right: control signal sent to slow drive of laser (laser heater), which is supposed to follow the cavity temperature (TMP_OUTPUT)
  • Bottom left: in-loop can temperature sensors (MINCOMEAS)
  • Bottom right: room temperature reading (RMTEMP)

 

I stepped the temperature set point from 35 to 36 deg. C for the can at 12:30am last night. Then I waited to see the cavity temperature change and the slow actuator (laser heater: TMP_OUTPUT) follow that change.

I was a bit worried about the oscillations that were occuring in the TMP_OUTPUT signal even long after this temperature step was made, but I figured that they were simply room-temperature changes propagating into the cavity, since they seemed to have a similar pattern to the room-temperature variations, and since it is clear that the out-of-loop temperature sensor on the can (RCTEMP) experiences variations, even when the in-loop sensors are recording no variation.

At 8:46pm tonight I stepped the temperature down 2 degrees to 34 deg. C. The step had a clear effect on TMP_OUTPUT. The voltage to the heater dropped and eventually railed at its lowest output. I'm worried that the loop is unstable, although I haven't ruled out other possibilities, such as that a 2 deg. C temperature step is too large for the loop. I will investigate further in the morning.

  5230   Sun Aug 14 15:37:39 2011 JennyUpdatePSLTemperature steps and slow actuator railing

Quote:

Below are some plots from dataviewer of temperature-step data taken over the past 32 hours. (They show minute trends). I am looking at the thermal coupling from the can surrounding the reference cavity on the PSL table to the cavity itself, and trying to measure the cavity temperature response via the control signal sent to heat the NPRO laser, which is locked to the cavity.

Picture_6.png

Picture_7.png

  • Top left: out-of-loop temperature sensor on can surrounding ref cav (RCTEMP)
  • Top right: control signal sent to slow drive of laser (laser heater), which is supposed to follow the cavity temperature (TMP_OUTPUT)
  • Bottom left: in-loop can temperature sensors (MINCOMEAS)
  • Bottom right: room temperature reading (RMTEMP)

 

I stepped the temperature set point from 35 to 36 deg. C for the can at 12:30am last night. Then I waited to see the cavity temperature change and the slow actuator (laser heater: TMP_OUTPUT) follow that change.

I was a bit worried about the oscillations that were occuring in the TMP_OUTPUT signal even long after this temperature step was made, but I figured that they were simply room-temperature changes propagating into the cavity, since they seemed to have a similar pattern to the room-temperature variations, and since it is clear that the out-of-loop temperature sensor on the can (RCTEMP) experiences variations, even when the in-loop sensors are recording no variation.

At 8:46pm tonight I stepped the temperature down 2 degrees to 34 deg. C. The step had a clear effect on TMP_OUTPUT. The voltage to the heater dropped and eventually railed at its lowest output. I'm worried that the loop is unstable, although I haven't ruled out other possibilities, such as that a 2 deg. C temperature step is too large for the loop. I will investigate further in the morning.

 The lock was lost when I came in around noon today to check on it. The slow actuator was still railing.

1) I got lock back for a few minutes, by varying the laser temperature set point manually. TMP_OUTPUT was still reading -30000 cts (minimum allowed) and the transmission was not as high as it had been.

2) I toggled the second filter button off. The TMP_OUTPUT started rising up to ~2000 cts. I then toggled the second filter back on, and TMP_OUTPUT jumped the positive maximum number of counts allowed.

3) I lost the lock again. I turned off the digital output to the slow actuator.

4) I have so far failed at getting the lock back. My main problem is that when the BNC cable to the slow port is plugged in, even when I'm not sending anything to that port, it makes it so that changing the temperature set point manually has almost no effect on the transmission (it looks as though changing the setpoint is not actually changing the temperature, because the monitor shows the same higher order mode even when with +-degree temperature setpoint changes).

  5271   Fri Aug 19 19:08:40 2011 JennyUpdatePSLRelocking NPRO to reference cavity.

I am trying again to measure a temperature step response on the reference cavity on the PSL table.

I have been working to relock the NPRO to the cavity. I unblocked the laser beam, reassembled the setup described in my previous elog entry: 5202. I then did the following:

1) Monitored error signal (from LB1005 PDH servo), transmitted signal, and control signal sent to drive PZT on oscilloscope.

2) With loop open, swept through 0,0-mode resonance, saw a peak in the transmission, saw an accompanying error signal similar to the signal shown in 5202.

3) Tried to lock. Swept the gain on the LB1005 and could not find a gain that would make it lock. Tried changing the PI-corner freq. from 10 kHz to 30 kHz and back and still could not lock.

4) Noticed that the open loop error signal displayed on the scope was DC-offset from zero. Changed the offset to zero the error signal.

5) Tried to lock again and succeeded.

6) Noticed that upon closing the loop, the error signal became offset from zero again. Turning on the integrator on the LB1005 increased DC-offset.

7) Reduced the gain on the SR560 being used as a low pass filter from 5 to 1. Readjusted the open loop error signal offset on the LB1005.

8) Closed the loop and locked. Closing the loop then caused a much smaller DC change in the signal than I had seen with the larger gain (now around 3mV). RMS fluctuations in error signal are now 1 mV (well within the linear region of the error signal).

9) Noticed transmission has a strange distorted harmonic oscillation in it a 1MHz. (Modulation freq is 230kHz, so it's not that). Checked reflected signal and also saw a strange oscillation--in a sawtooth-like pattern.

 

I intend to

1) Post oscilloscope traces here showing transmitted and reflected signal when locked.

2) Look upstream to see if the sawtooth-like oscillation is in the laser beam before it enters the cavity:

  • Sweep the temperature of the laser so that the beam is no longer resonating in the cavity.
  • Compare the reflected signal off the cavity to the signal detected before being directed into the cavity (using the PDA255 that I used for measuring the AM response of the PZT) both with and and without the frequency modulation.

3) At some point, try to close the slow digital loop, perhaps readjusting the gain.

4) Try to measure a temperature step response.

  5272   Fri Aug 19 23:41:20 2011 JennyUpdatePSLRelocking NPRO to reference cavity.

Quote:

I am trying again to measure a temperature step response on the reference cavity on the PSL table.

I have been working to relock the NPRO to the cavity. I unblocked the laser beam, reassembled the setup described in my previous elog entry: 5202. I then did the following:

1) Monitored error signal (from LB1005 PDH servo), transmitted signal, and control signal sent to drive PZT on oscilloscope.

2) With loop open, swept through 0,0-mode resonance, saw a peak in the transmission, saw an accompanying error signal similar to the signal shown in 5202.

3) Tried to lock. Swept the gain on the LB1005 and could not find a gain that would make it lock. Tried changing the PI-corner freq. from 10 kHz to 30 kHz and back and still could not lock.

4) Noticed that the open loop error signal displayed on the scope was DC-offset from zero. Changed the offset to zero the error signal.

5) Tried to lock again and succeeded.

6) Noticed that upon closing the loop, the error signal became offset from zero again. Turning on the integrator on the LB1005 increased DC-offset.

7) Reduced the gain on the SR560 being used as a low pass filter from 5 to 1. Readjusted the open loop error signal offset on the LB1005.

8) Closed the loop and locked. Closing the loop then caused a much smaller DC change in the signal than I had seen with the larger gain (now around 3mV). RMS fluctuations in error signal are now 1 mV (well within the linear region of the error signal).

9) Noticed transmission has a strange distorted harmonic oscillation in it a 1MHz. (Modulation freq is 230kHz, so it's not that). Checked reflected signal and also saw a strange oscillation--in a sawtooth-like pattern.

 

I intend to

1) Post oscilloscope traces here showing transmitted and reflected signal when locked.

2) Look upstream to see if the sawtooth-like oscillation is in the laser beam before it enters the cavity:

  • Sweep the temperature of the laser so that the beam is no longer resonating in the cavity.
  • Compare the reflected signal off the cavity to the signal detected before being directed into the cavity (using the PDA255 that I used for measuring the AM response of the PZT) both with and and without the frequency modulation.

3) At some point, try to close the slow digital loop, perhaps readjusting the gain.

4) Try to measure a temperature step response.

I decided to go forward and try to close the digital loop in spite of the unexplained oscillations in the transmission.

1) Plugged the 20dB attenuator into the slow port on the laser drive. This pushed the laser out of lock and, for some reason, made the laser temperature stop responding to sweeping the set point manually with the knob.

2) Plugged the output from the digital system into the slow port (with the attenuator still in place).

3) Displayed the beam seen by the camera on a monitor in the control room

4) Stepped the laser temperature using MEDM until finding the 0,1 mode. Locked to that mode.

5) Closed the digital loop (input to slow laser drive attenuated 20dB attenuator). Gain 0.010

6) Loop appeared stable for 30 minutes, then temperature began shooting off. I opened the loop, cleared history, reduced the gain to 0.008, and started it again. Loop appears stable after 15 minutes of watching. I'm going to leave it for a few hours, then come back to check on it and, if it's stable, step the can temperature.

  5274   Sat Aug 20 23:01:39 2011 JennyUpdatePSLTaking temperature step-response data: successes and tribulations

After finishing my last elog entry, I monitored the digital loop's error signal (the control signal for the fast loop) and the output to the laser heater remotely, (from West Bridge), using dataviewer. The ref cav surrounding can temperature was set to 36 degrees C.

With the loop closed and a gain of 0.008, after seeing the output voltage to the laser heater (TMP_OUTPUT) remain fairly constant and the error signal (TMP_INMON) stay close to zero for ~3 hours, I tried to step the temperature. (This was at 2am last night). I was working remotely from West Bridge. To step the temperature I used the following command:

ezcawrite C1:PSL-FSS_RCPID_SETPOINT 35.5

 

Rather than change the can temperature to 35.5 C, it outputted:

C1:PSL-FSS_RCPID_SETPOINT=0.

 

It had set the setpoint to 0 degrees C, which was essentially turning the heater off. I tried resetting it back to 36 and had no luck. I tried changing the syntax slightly.: ezcawrite C1:PSL-FSS_RCPID_SETPOINT=36 and ezcawrite C1:PSL-FSS_RCPID_SETPOINT (36). No success.

I ran over to the 40m and changed the temperature back to 36 manually. The in-loop temp sensor had decreased to 31.5 degrees C before I was able to step the setpoint back up. The system seems to have recovered from this large impulse though, and the laser has remained locked.

5hourwbigimpulse.jpg

 5hourwbigimpulse2.jpg

(5 hours of minute-trend data)

From left to right: 

Top: Out-of-loop can temp sensor; Voltage sent to heat can

Middle: signal sent to heat the laser (TMP_OUTPUT); room temp

Bottom: Error signal for slow loop (sampled control signal from fast loop); In-loop can temp sensor

 

At 9:30 this morning (7 and a half hours after accidentally setting the setpoint to zero), I came in to the 40m. TMP_OUTPUT was still decreasing but was slowing somewhat, so I decided to step the can temperature up half a decree to 36.5 C.

TMP_OUTPUT responded to the step, but it is also oscillating slowly with room-temperature changes, and these oscillations are on the same order as the step response. The oscillations look like the room-temp oscilations, but inverted. (TMP_OUTPUT reaches maxima when RMTEMP reaches minima). Oddly, there does not appear to be much of a time delay between the room temperature and TMP_OUTPUT signals. I would expect a time delay since there's a time constant for a room-temperature change to propagate into the cavity. Perhaps the laser itself is susceptible to room-temperature changes and those propagate into the laser cavity on a much faster time scale. I don't know the thermal coupling of ambient temperature changes into the laser.

23hoursbefore920pm.jpg

23hoursbefore920pm2.jpg

(24-hours of second-trend data)

 

Options are:

--If the system can handle it, do a larger temperature step (3 degrees, say), so that I can more clearly distinguish the oscillations with room temp from the step response.

--Insulate the cavity with foam (will in principle make the temperature over the can surrounding the ref cav more uniform and less affected by room temperature changes).

--Insulate the laser? Is this possible?

--Leave the system as is and, as a first approximation, fit the room-temp data to a sine wave and subtract it off somehow from my data to just see the step response.

--Don't bother with steps and just try to get the transfer function from out-of-loop temperature (RCTEMP, which is affected by temperature noise from the room) to TMP_OUTPUT via taking the Fourier transforms of both signals.

 

I'm flying out tomorrow morning, so I'll either need to figure out how to step the temperature set point of the can remotely, successfully, or I'll need someone to manually enter in the temperature steps for me in the control room.

  5549   Mon Sep 26 17:49:51 2011 KojiUpdatePSLc1psl

[Koji Suresh]

c1psl has got frozen during our ezcaread/write business.
After the target was rebooted and we lost the previous setting as there was no burt snapshot for the slow targets since Dec 13, 2010.

It seems that burtrestore is essential for the bootstrapping of the MC servo, as the auto locker script refers the locking parameters
from the PSL setting values (C1PSL_SETTINGS_SET.adl).

Jenne is working on the recovery of the snap-shotting for the slow targets.

  5551   Mon Sep 26 20:04:03 2011 KojiUpdatePSLMC lock has been recovered

[Kiwamu Suresh Koji]

Some main parameters of the PSL has been recovered using Dataviewer and some screen snapshots, as seen in the screenshots below.

Attachment 1: snapshot1.png
snapshot1.png
Attachment 2: snapshot2.png
snapshot2.png
Attachment 3: snapshot3.png
snapshot3.png
  5564   Wed Sep 28 13:30:01 2011 JenneUpdatePSLPMC was unlocked

Relocked the PMC.  MC came back immediately.

  5606   Mon Oct 3 20:02:59 2011 SureshUpdatePSLAM / PM ratio

[Koji, Suresh]

In the previous measurement, the PDA 255 had most probably saturated at DC, since the maximum ouput voltage of PDA255 is 5V when it is driving a 50 Ohm load.  It has a bandwidth of 0 to 50MHz and so can be reliably used to measure only the 11 MHz AM peak.  In this band it has a conversion efficiency of 7000 V per Watt (optical power at 1064nm).  [Conversion efficiency:  From the data sheet we get 0.7 A/W of photo-current at 1064nm and 10^4 V/A of transimpedance]  The transimpedance at 55 MHz is not given in the data sheet.  Even if PDA255 is driving a high impedance load, at high incident power levels the bandwidth will be reduced due to finite gain x bandwidth product of the opamps involved, so the conversion efficiency at 11 MHz would not be equal to that at DC.

So Koji repeated the measurement with a lower incident light level:

**********************************

V_DC = 1.07 V  with 50 Ohm termination on the multimeter.

Peak height at 11 MHz on the spectrum analyzer (50 Ohm input termination) = -48.54 dBm

***********************************

Calculation: 

a) RF_Power at 11 MHz :  -48.45 dBm = 1.4 x 10^(-8) W

b) RF_Power = [(V_rms)^2] / 50_ohm  ==> V_rms = 8.4 x 10^(-4) V

c) Optical Power at 11 MHz: [V_rms / 7000] = 1.2 x 10^(-7) W

d) Optical Power at DC =  [V_DC / 7000] = 1.46 x 10^(-4) W

e) Intensity ratio:  I_AM / I_c = 7.9 x 10^(-4) . AM:Carrier amplitude ratio is half of the intensity ratio = 4.0 x 10^(-4)

f) PM amplitude ratio from Mirko's measurement is 0.2

g) The PM to AM amplitude ratio is 506

_________________________________

As the AM peak is highly dependent upon the drifting EOM position in yaw, it is quite likely that a higher PM/AM ratio could occur.  But this measurement shows how small it could get if the current situation is allowed to continue.

 

Quote:

[Mirko / Kiwamu]

 We have reviewed the AM issue and confirmed the ratio of AM vs. PM had been about 6 x103.

The ratio sounds reasonably big, but in reality we still have some amount of offsets in the LSC demod signals.

Next week, Mirko will estimate the effect from a mismatch in the MC absolute length and the modulation frequency.

 


(Details)

 Please correct us if something is wrong in the calculations.

 According to the measurement done by Keiko (#5502):

        DC = 5.2 V

        AM @ 11 and 55 MHz = - 56 dBm = 0.35 mV (in 50 Ohm system)

Therefore the intensity modulation is 0.35 mV / 5.2 V = 6.7 x 10-5

Since the AM index is half of the intensity modulation index, our AM index is now about 3.4 x 10-5

According to Mirko's OSA measurement, the PM index have been about 0.2.

As a result,  PM/AM = 6 x 103

Quote from #5502

Measured values;

* DC power = 5.2V which is assumed to be 0.74mW according to the PDA255 manual.

*AM_f1 and AM_f2 power = -55.9 dBm = 2.5 * 10^(-9) W.

 

 

  5616   Tue Oct 4 16:58:45 2011 SureshUpdatePSLAM / PM ratio

Correction: Koji noted that Mirko actually reports a PM modulation index of 0.17 for the 11 MHz sideband (elog: http://nodus.ligo.caltech.edu:8080/40m/5462. This means

f) the amplitude ratio of the PM side-band to carrier is half of that = 0.084

g)  the PM to AM amplitude ratio as 0.084 / [4.0 x 10^(-4)]  = 209.

  5644   Mon Oct 10 15:41:56 2011 KojiUpdatePSLPMC aligned

[Koji Suresh]

The steering mirrors for PMC were aligned. The transmission went up from 0.779 to 0.852.

ELOG V3.1.3-