  4055   Tue Dec 14 22:42:56 2010 kiwamuUpdateSUSinstalled the last TT . And some more progresses

  [Koji, Oasmu and Kiwamu]

We made the following progresses :

   (1) installation of the last TT called SR2.

 (2) fixing an earthquake stop issue on the BS

  (3) fixing a clipping of beams at the dark port


(earthquake stops on the BS)

 - When we were aligning the BS, we found that the BS showed funny behaviors.

For example a kick on LR didn't shake LRSEN, and a big DC angular offset (~ 8 in the medm screen ) was needed to keep a horizontal beam axis in the reflection.

We checked all the earthquake stops and found two suspicious earthquake stops at right bottom side.

They looked like slightly touching the BS. So we moved them further away from the BS.

Then the problem had gone.  We doublechecked the health by kicking, applying an angular offset  and so on.

 

(beam clipping)

We found transmitted/reflected beams from the BS to the dark port was clipped at the BS tower.

We moved the BS tower by ~1cm and realigned the rotation of the BS tower.

We also found the beam spots on the two TTs, PR2 and PR3, were offcentered. Especially the beam spot on PR3 was almost on the edge of the mirror.

Probably this was because we touched PZT1 when aligning the beam to the Y end.

So it means, it is not a good idea to align the beam to the end only by steering PZT1. We should use PR2 and PR3 as well when we align the beam to the Y end.

We realigned them such that the beam hits the center of PR2, PR3 and ETMY.

  4063   Thu Dec 16 01:23:43 2010 kiwamuUpdateSUSSRM OSEMs installed

  [Zach and Kiwamu]

 We worked on some more vacuum businesses.  Today we finished did the following works:

 - alignment of the POX mirrors 

 - alignment of the POP1 and POP2 mirrors

 - installation of OSEMs onto SRM

 - alignment of the SRM tower

 

 (alignment of POP mirrors)

 Since a beam on the POP path was quite too weak to see even by IR viewers, we used a He-Ne laser to imitate the real beam instead.

We injected the He-Ne beam from an optical bench to the chamber, and made it go through the PRM and PR2 by using some steering mirrors.

 

(OSEM cables)

 The pin assignment was flipped in a way of mirror image due to the extension cables which cause a mirroring.

 So we made mirroring connectors to flipp them back to the correct pin assignment, and plugged the mirroring connectors in between the feedthrough of the BS chamber and the SRM satellite box.

This is a picture showing how they are connected now.

DSC_2757_ss.jpg

  4069   Fri Dec 17 03:37:47 2010 ranaUpdateSUSITMY / SRM / BS / PRM OPLEVs aligned

Quote:

The sole thing that has been deviated from the optical layout was that the SRM returning beam had to be reroute
as the SRM has better reflectivity on the AR surface in stead of the HR one.

 I suppose that if we were really clever we would intentionally choose either the AR or HR surface so as to minimize the effect of the thermal lensing and/or thermal expansion from the locked interferometer absorption.

  4079   Mon Dec 20 23:10:25 2010 JenneUpdateSUSPretty much ready for pump-down. A few final things....

[Kiwamu, Jenne, Koji, Osamu]

We have mostly prepared the IFO for pump down. 

After lunch [Steve, Bob, Koji, Kiwamu, Jenne, Joe, Joon Ho, Vladimir, Osamu] put the access connector back in place.  Hooray!  Steve still has to check the Jam Nuts before we pump down.  Kiwamu checked the leveling of the IOO table, and fixed all of the weights to the table.

For all 4 test masses, bars (upside-down dog clamps) were placed to mark the alignment of 2 sides of the suspension tower.  All test mass tables were re-leveled, and the weights fixed to the tables.  

For ETMY, PRM, BS, SRM, we confirmed that the OSEMs were close to their half-range.  ETMX was already fine.  ITMY (the screens and the optics wiki are still old-convention, so this is listed as ITMX! No good!) OSEMs are pretty much fine, but ITMX desperately needs to be adjusted.  Unfortunately, no one can find the standard screwdriver (looks like a minus), to adjust the ITM OSEMs.  All the other towers had hex-key set screws, but the ITMs need a screwdriver.  We will ask Bob to sonicate a screwdriver in the morning. 

 

 

  4086   Wed Dec 22 11:24:23 2010 haixingUpdateSUSmeasurement of imbalance in quadrant maglev protope

Yesterday, a sequence of force and gain measurement was made to determine the imbalance in the

quadrant, magnetic-levitation prototype. This was the reason why it failed to achieve a stable levitation.

The configuration is shown schematically by the figure below:

 

config_2x2.png

 

Specifically, the following measurements have been made:

(1) DC force measurement among four pairs of magnets at fixed distance with current of the coils on and off

From this measurement,  the DC force between pair of magnets is determined and is around 1.6 N at with a

separation of 1 cm. This measurement also lets us know the gain from voltage to force near the working point.

 

The force between pair "2" is about 13% stronger than other pairs which are nearly identical. The force by the

coil is around 0.017 N per Volt (levitation of 5 g per 3 Volt); therefore, we need around 12 volt DC compensation

of pair "2" in order to counterbalance such an imbalance.  Given the resistence of the coil equal to 26 Om, this

requires almost 500 mA DC compensation. Koji suggested that we need a high-current buffer, instead of what

has been used now.

 

(2) DC force measurement among four pairs of magnets (with current of the coils off) as a function of distance

From this measurement, we can determine the stiffness of the system. In this case, the stiffness or the

effective spring constant is negative, and we need to compensate it by using a feedback control. This is

one of the most important parameters for designing the feedback control. The data is still in processing.

 

(3) Gain measurement of the OSEM from the displacement to voltage.

This measurement is a little bit tricky due to the difficulty to determine the displacement of the flag.

After several measurements, it gave approximately 2 V/cm.

 

Plan for the next few days:

From the those measurements, all the parameters for the plant and sensor that need to determine the

feedback control are known. They should be plugged into the simulink model and to see whether the

old design is appropriate or not. Concerning the experimental part, we will first try to levitate the configuration

with 2 pairs of magnets, instead of 4 pairs, as the first step, which is easier to control but still interesting.

 

 

 

 

 

 

  4100   Sat Jan 1 16:36:16 2011 ranaConfigurationSUSrampdown.pl was not running

The crontab for op340m which runs various IFO maintenance activities has been set to the wrong path for the watchdog rampdown script since the CDS changeover.

This is a dangerous situation. With the watchdog thresholds set as high as 1000, the magnets can be broken if the new CDS has some mental problems. This kind of thing happened before at LHO (which is why Stan invented the watchdogs). Please try to make sure that the watchdog thresholds are not set that way - its our only defense against bad CDS.

I've now corrected the crontab. The watchdog thresholds are being stepped down every 30 minutes as before.

However, out test points are gone again, so who knows how things are behaving.

  4204   Wed Jan 26 02:18:12 2011 KojiUpdateSUSETMX length to angle matrix

I have put an offset of 1000 counts to C1:SUS-ETMX_ALS_OFFSET. This actually misalign the mirror a lot.

While the offset is applied. I adjusted the balance of the coil matrix.
UL 1.580 UR 0.620
LL 0.420 LR 1.380

> ezcaread C1:SUS-ETMX_TO_COIL_0_0_GAIN
C1:SUS-ETMX_TO_COIL_0_0_GAIN = 1.58
> ezcaread C1:SUS-ETMX_TO_COIL_0_1_GAIN
C1:SUS-ETMX_TO_COIL_0_1_GAIN = 0.62
> ezcaread C1:SUS-ETMX_TO_COIL_0_2_GAIN
C1:SUS-ETMX_TO_COIL_0_2_GAIN = 0.42
> ezcaread C1:SUS-ETMX_TO_COIL_0_3_GAIN
C1:SUS-ETMX_TO_COIL_0_3_GAIN = 1.38

Now, we can keep TEM00 for green with +/-1000counts of push although the fast step of the offset make the lock lost.

It turned out that the step longitudinal input temporary misalign the mirror in pitch because the length and pitch are coupled.
I guess that we don't excite pitch if we push the mirror slowly. Eventually, we need f2p transfer function adjusted in the output matrix.

Kiwamu told us that:
(2)  Length to Alignment coupling. Pushing ETMX causes a misalignment.

 

  4236   Tue Feb 1 17:34:21 2011 JenneUpdateSUSETMX and PRM watchdogs tripped

I sat down in the control room to find that ETMX and PRM's watchdogs had been tripped.  I don't know how long they've been crazy, but there was a big something that showed up in the seismometers around 16:30UTC, or ~11:30 this morning.  I don't find any significant earthquakes on the USGS site for that time though, so it might be more local, i.e. work next door or trucks or whatever.

I take back the suggestion that it was that seismic event.  Clearly the PRM and the ETMX were kicked at different times, neither of which is the same as the seismic action. Mystery.  You can see they have been ringing down for a while though, which is neat. 

Attachment 1: Seis_1Feb2011.png
Seis_1Feb2011.png
Attachment 2: Seis_SUS_1Feb2011.png
Seis_SUS_1Feb2011.png
  4293   Mon Feb 14 23:29:04 2011 ranaSummarySUSETMX Fitlers moved around

All of the SUS used to have only 1 filter module for SIDE. They now have 3 filter modules for SIDE just like the other DOFs.

Today I moved the filters around so that the sensor filters are in SDSEN, the servo filters are in SUSSIDE, and the dewhitening for the coil is in SDCOIL.

I noticed along the way that the bounce/roll mode notches for all of the suspensions are still set for the frequencies of the previous suspensions. Suresh has 'volunteered' to find the new frequencies and make the new bandstop filters by looking up the seminal work on this by Dan Busby / Sam Waldman.

  4326   Fri Feb 18 18:46:08 2011 kiwamuSummarySUSf2p done on ETMX and ITMX

The f2p measurements are done on ETMX and ITMX with the real time lockin systems.

I don't explain what is the f2p measurement in this entry, but people who are interested in it can find some details on an old elog entry here or somewhere on DCC.

So far the resultant filters looked reasonable compared with the previous SRM f2p filters.

 

- backgrounds -

  Some times ago I found that the coils on ETMX had not  been nicely balanced, and it made a POS to angle coupling when I tried green locking (see here).

In addition to that, accuracy of A2L kind of measurement including the dithering techniques depend on how well the coils are balanced.  Therefore we need to balance the coils basically at all the suspended optics.

There used to be a script for this particular purpose, called f2praio.sh. This script does measure the imbalances and then balance the coils.

However this time I used the realtime lockin system to measure the imbalances instead of using the old f2p script.

One of the reasons using the real time system is that,  some of the ezca and tds commands for the old script don't work for some reasons.

Therefore we decided to move on to the real time system like Yuta did for the A2L measurement a couple of months ago.

The f2p measurement finally gives us parameters to generate a proper set of filters for POS and also the coil gains. We apply those filters and the gains in order to eliminate the POS to angle coupling and to balance the coils.

 

- results -

The followers are the resultant filters and coil gains.

The plots below show new f2p filters according to the measurement.

f2p_ITMX.png       f2p_ETMX.png

 

ITMX (assuming pendulum POS has f0 = 1 Hz and Q = 1)

ULPOS  fz = 1.009612   Qz = 1.009612 

URPOS fz = 1.125965   Qz = 1.125965  

LLPOS  fz = 0.873725   Qz = 0.873725    

LRPOS  fz = 0.974418   Qz = 0.974418  

C1:SUS-ITMY_ULCOIL_GAIN      -1.103044

C1:SUS-ITMY_URCOIL_GAIN      0.884970

C1:SUS-ITMY_LLCOIL_GAIN      0.950650

C1:SUS-ITMY_LRCOIL_GAIN      -1.060326

 

 

ETMX (assuming pendulum POS has f0 = 1 Hz and Q = 1)

ULPOS  fz = 1.055445   Qz = 1.055445   

URPOS  fz = 1.052735   Qz = 1.052735   

LLPOS  fz = 0.944023   Qz = 0.944023   

LRPOS  fz = 0.941600   Qz = 0.941600   

C1:SUS-ETMX_ULCOIL_GAIN      -0.887550

C1:SUS-ETMX_URCOIL_GAIN      1.106585

C1:SUS-ETMX_LLCOIL_GAIN = 1.07233

C1:SUS-ETMX_LRCOIL_GAIN      -0.931013

  

The precision of the coil gains looked something like 1% because every time I ran the measurement script, the measured imbalances fluctuated at this level.

The precision of the filter gain at DC (0.01 Hz) could be worse, because the integration cycles for the measurement are fewer than that of the coil gains done at high frequency (8.5 Hz).

Of course we can make the precisions by increasing the integration cycles and the excitation amplitudes, if we want to.

  4327   Fri Feb 18 20:06:59 2011 kiwamuSummarySUScheck f2p function on ETMX

 The plot below shows how the f2p filters work.

At -2 min I turned on the f2p filters.

 f2p_ETMX.png

  4349   Thu Feb 24 11:18:08 2011 steveUpdateSUSSRM sus-cables

Cheater cables   for SRM sus tied up. They were dangling aimlessly on the floor.

Attachment 1: P1070428.JPG
P1070428.JPG
  4592   Sat Apr 30 00:15:26 2011 kiwamuUpdateSUSDRMI prep : suspension diagnostic

As a part of the DRMI preparation,

I leave all the suspensions free from the watchdogs for 5 hours from now.

Please DO NOT touch them.

I will check the spectra and the mechanical resonant frequencies on Monday.

Also I will renew all the input matrices of the local dampings based on these free swinging spectra.

  4594   Mon May 2 11:14:27 2011 kiwamuUpdateSUSscript : opticshutdown

Just FYI.

There is a useful script for this particular job : shutting down all the suspensions and bringing it back to operation after 5 hrs.

It is called opticshudown, which resides in /cvs/cds/rtcds/caltech/c1/scripts/SUS/.

 

Also I added this script on the list in the wiki where all the scripts will be listed.

http://blue.ligo-wa.caltech.edu:8000/40m/Computers_and_Scripts/All_Scripts#opticshutdown

If you find any other useful scripts, please add them on the wiki.

Quote from #4592

I leave all the suspensions free from the watchdogs for 5 hours from now.

 

  4596   Mon May 2 13:30:13 2011 kiwamuUpdateSUSPRMI locking prep: PRM oplev

I had a quick look at PRM optical lever.

The He-Ne beam is still successfully coming out from the chamber and I could guide it to the QPD by using steering mirrors.

But the beam size looks too big for the QPD. We should slide the lens which is standing before the injection to get a moderately smaller beam size at the QPD.

Quote from #4595

 - activation of PRM oplev

  4599   Mon May 2 17:10:55 2011 steveUpdateSUSPRM oplev

The returning spot diameter on the qpd ~10 mm. In order to reduce the spot size I moved the f 1145 mm lens toward the PRM ~ 25 cm. The spot size was reduced to ~8 mm, 3200 counts.

I'll try to find an other lens tomorrow.

  4611   Tue May 3 13:22:13 2011 LeoUpdateSUSRe: DRMI prep : suspension diagnostic

Here are the free-swinging spectra for the BS, ETMX, ETMY, ITMX, ITMY, MC1, MC2, MC3, and PRM chambers.  Kiwamu left the suspensions free for 5 hours this weekend, starting at Sat Apr 30 00:15:26 2011.

This is GPS time 988 182 941.  Quick tip: you can do local to GPS time conversions using lalapps_tconvert, which is a lot like tconvert but with special powers.  It is installed on pianosa.

$ lalapps_tconvert Sat Apr 30 00:15:26 2011

988182941

I generated these figures with the attached Python script, measure.py.

Notice that the C1:SUS-ITMX_SENSOR_UL and C1:SUS-MC3_SENSOR_UL spectra fall as 1/f.  Jenne suggested that this might indicate that there is a loose electrical connection.

Also, notice that C1:SUS-ETMY_SENSOR_LR, C1:SUS-ITMY_SENSOR_LL, and C1:SUS-PRM_SENSOR_SIDE are a lot noisier above 10 Hz.

Attachment 1: BS.png
BS.png
Attachment 2: ETMX.png
ETMX.png
Attachment 3: ETMY.png
ETMY.png
Attachment 4: ITMX.png
ITMX.png
Attachment 5: ITMY.png
ITMY.png
Attachment 6: MC1.png
MC1.png
Attachment 7: MC2.png
MC2.png
Attachment 8: MC3.png
MC3.png
Attachment 9: PRM.png
PRM.png
Attachment 10: SRM.png
SRM.png
  4613   Tue May 3 15:04:20 2011 kiwamuUpdateSUSRe: DRMI prep : suspension diagnostic

Jenne went through all the suspension racks and pushed all the connectors.

After pushing them, we had a quick look at those spectra and found no funny noise spectrum except for C1:PRM-SENSOR_UL.

We then checked connection around the SCSI cables and eventually found the connection between ADC_card_0 and a SCSI was loose.

We put short standoffs on the ADC card so that the screws from the SCSI can nicely reach to the ADC card. Now everything looks fine.

SUS diagnostic is quite useful !

Quote from #4611

Notice that the C1:SUS-ITMX_SENSOR_UL and C1:SUS-MC3_SENSOR_UL spectra fall as 1/f.  Jenne suggested that this might indicate that there is a loose electrical connection.

  4616   Tue May 3 16:20:13 2011 steveUpdateSUSPRM & BS oplevs are ready

Quote:

The returning spot diameter on the qpd ~10 mm. In order to reduce the spot size I moved the f 1145 mm lens toward the PRM ~ 25 cm. The spot size was reduced to ~8 mm, 3200 counts.

I'll try to find an other lens tomorrow.

 Atm 1,  PRM oplev inward path with 2 lens solution: 14 cm gap between F 1145 and F 1545 mm lenses. 

Atm 2,   The PRM beam size 3 mm and the beam  quality is still bad. The BS path only needed alignment.

Attachment 1: P1070630.JPG
P1070630.JPG
Attachment 2: P1070632.JPG
P1070632.JPG
  4626   Wed May 4 13:57:04 2011 kiwamuUpdateSUSmechanical resonances updated

[Leo  w/ a little help from Kiwamu]

Leo summarized the mechanical resonances of all the suspensions, based on the free-swinging spectra taken on Sat Apr 30.

Since Leo doesn't have the wiki account I helped him putting the information on the wiki.

Good work, Leo !

http://blue.ligo-wa.caltech.edu:8000/40m/Mechanical_Resonances

Quote from

Here are the free-swinging spectra for the BS, ETMX, ETMY, ITMX, ITMY, MC1, MC2, MC3, and PRM chambers.  Kiwamu left the suspensions free for 5 hours this weekend, starting at Sat Apr 30 00:15:26 2011.

 

  4646   Thu May 5 17:19:21 2011 Leo SingerConfigurationSUSTuning notch filters for bounce mode suspensions

I am tuning the notch filters for the bounce modes in the suspensions, starting with the ITMs and ETMs.  I'll do the MCs, the PRMs, and the SRMs next.

 

I noticed that the filter for ITMX (in the file C1SUS.txt, the module ITMX_SUSPOS, the selection BounceRoll) that the filter was composed of two bandstops (and a constant gain).  It looked like this:

 

ellip("BandStop",4,1,40,11.4,12.2)ellip("BandStop",4,1,40,16.7,17.5)gain(1.25872)

 

Valera said that one of these was for the roll mode and the other for the bounce mode.  However, looking at the spectra that Kiwamu and I made this week, I don't perceive a resonance between 11.4 and 12.2 Hz.  So, we're taking a guess that this was for a mode that has moved due to new pendulum designs.  For many of the suspensions, in the free swinging test we noticed a line around 23 Hz; we thought we might as well re-use one of these elliptical filters to avoid exciting this line.  Of course, if this line does *not* result from excitation of an uncontrolled degree of freedom, this will not help and could be detrimental.  When we talk to Valera again, we can review this decision and at that point we might decide just to take out that bandstop.

 

ITMX is done.  I'll continue tomorrow.  I've attached closed-loop spectra for before the tuning (itmx-before.pdf) and after (itmx-after.pdf).

 

(Update: the following day, I took closed loop spectra with (itmx-withbounceroll.pdf) and without (itmx-nobounceroll.pdf) the bandstops.  It looks like the bandstops made the bounce mode slightly worse, but the roll mode slightly better.)

 

 

Attachment 1: itmx-before.pdf
itmx-before.pdf
Attachment 2: itmx-before.pdf
itmx-before.pdf
Attachment 3: itmx-withbounceroll.pdf
itmx-withbounceroll.pdf
Attachment 4: timx-nobounceroll.pdf
timx-nobounceroll.pdf
  4652   Fri May 6 14:59:36 2011 Leo SingerConfigurationSUSTuning ITMY bandstop

I tuned the ITMY bandstops -- 'before' and 'after' spectra attached.  Note that the after the tuning, the bounce mode at ~16 Hz is about twice as quiet!

 

However, notice that in the 'before' plot the roll mode at about 23.5 Hz did not show up at all, whereas it is quite prominent in the 'after' plot.  I was concerned that this line could have been a result of placing the bandstop there, so I made another plot with the BounceRoll filter turned off.  Sure enough, the 23.5 Hz line is still there.  So I'm not crazy: the roll mode did start acting up at some time between my 'before' and 'after' plot, but not as a result of the tuning.

Attachment 1: itmy-before.pdf
itmy-before.pdf
Attachment 2: itmy-after.pdf
itmy-after.pdf
Attachment 3: itmy-nobounceroll.pdf
itmy-nobounceroll.pdf
  4655   Fri May 6 17:11:55 2011 steveUpdateSUSSRM-oplev cable

The SRM qpd cable was removed from the BS-table. It's path was changed from 1x4 to ITMY-table following the inner cable tray.

  4682   Tue May 10 22:56:17 2011 kiwamuUpdateSUSf2p filters installed on ETMY

New f2p filters were installed on ETMY. 

The statistical error of the coil gain settings are now about 0.8% at high frequency (i.e. above the resonant freq of the pendulum mode)

 What I did :

 -   measured and corrected the coil imbalances on ETMY using a script called F2P_LOCKIN.py

 -   made the new f2p filters based on the measurements and installed them.

  Next step :

 -  do the same adjustment for all the suspensions including PRM, SRM, BS, ITMs and ETMs

 


(Notes on F2P_LOCKIN.py)

 F2P_LOCKIN.py is a script that I've made in python. This is basically the same as the old script, f2pratio, but uses the realtime LOCKINs instead of ezcademods.

The script automatically measures the coil imbalances on an optics of interest by driving the local LOCKIN oscillators.

In the first step the script automatically balances the coil gains at high frequency (8.5Hz).

In the next step it gives some coefficients, which basically represent the coil imbalances at low frequency (0.01Hz)

Then with those coefficients one will be able to design the f2p filters.

It is not well polished yet, so I will spend some more times to make it user-friendly and readable.

Example usage : F2P_LCKIN.py -o ETMX

It currently resides in /cvs/cds/rtcds/caltech/c1/scriptss/SUS/

 

(new f2p filters)

The plot below shows the new f2p filters. Note that they are already installed.

f2p_etmy.png

Pendulum mode = 0.982 Hz (according to the wiki)
Q of the pendulum mode = 1 (to avoid ringing of IIR filters)
C1:SUS-ETMY_ULCOIL_GAIN = -0.793435
C1:SUS-ETMY_URCOIL_GAIN = 1.16877
C1:SUS-ETMY_LLCOIL_GAIN = 1.25028
C1:SUS-ETMY_LRCOIL_GAIN = -0.730704
UL_fz = 0.965125
UR_fz = 1.029517
LL_fz = 0.934231
LR_fz = 0.996562
UL_Qz = 0.982816
UR_Qz = 1.048388
LL_Qz = 0.951356
LR_Qz = 1.014829
  4683   Tue May 10 23:14:01 2011 kiwamuUpdateSUSETMX_OPLEV : He-Ne laser dead ?

I found that a He-Ne laser which has been used for ETMX_OPLEV was NOT giving the light.

Since I didn't find the switch key for it I have no idea if the laser is simply off or dead.

  4686   Wed May 11 11:05:45 2011 steveUpdateSUSETMX_OPLEV : He-Ne replaced

Quote:

I found that a He-Ne laser which has been used for ETMX_OPLEV was NOT giving the light.

Since I didn't find the switch key for it I have no idea if the laser is simply off or dead.

 The dead laser was replaced by new JDSU 1103P of  2.6mW. The return beam is big ~5 mm diameter of 0.3 mW,  1400 counts

Attachment 1: P1070672.JPG
P1070672.JPG
  4687   Wed May 11 11:24:59 2011 steveUpdateSUSSRM-oplev qpd is back

Quote:

The SRM qpd cable was removed from the BS-table. It's path was changed from 1x4 to ITMY-table following the inner cable tray.

 Laser diode oplev SRM is working. Qpd matrix values were reset like others.

In 0.44mW, returning 0.1mW,  -500 counts.

Attachment 1: srm-prm_oplevs.jpg
srm-prm_oplevs.jpg
  4688   Wed May 11 11:44:52 2011 kiwamuUpdateSUSf2p filters installed on ITMY

New f2p filters were installed on ITMY this morning. The statistical error of the coil gain setttigs are about 0.6 % at high frequency.

NEXT : PRM, SRM, BS, ITMX and ETMX.

 

f2p_itmy.png

 Pendulum mode = 0.988 Hz, Q = 1

C1:SUS-ITMY_ULCOIL_GAIN = 1.02482

C1:SUS-ITMY_URCOIL_GAIN = -1.06831

C1:SUS-ITMY_LLCOIL_GAIN = -0.996671

C1:SUS-ITMY_LRCOIL_GAIN = 0.91079

UL: fz = 1.014824 Hz, Q = 1.027150

UR: fz = 0.975038 Hz, Q = 0.98688

LL: fz = 1.000229 Hz, Q = 1.012378

LR: fz = 0.972688 Hz, Q = 0.946116

Quote from #4682

New f2p filters were installed on ETMY. 

The coil imbalances are now about 0.8% at high frequency (i.e. above the resonant freq of the pendulum mode)

  4693   Wed May 11 21:58:30 2011 ranaUpdateSUSETMX_OPLEV : He-Ne replaced

Quote:

Quote:

I found that a He-Ne laser which has been used for ETMX_OPLEV was NOT giving the light.

Since I didn't find the switch key for it I have no idea if the laser is simply off or dead.

 The dead laser was replaced by new JDSU 1104P of  2.6mW. The return beam is big ~5 mm diameter of 0.3 mW,  1400 counts

 Whenever replacing any Oplev laser, please also put into the ELOG when it was installed so that we have an electronic record of the laser lifetime.

  4713   Fri May 13 17:20:48 2011 Leo SingerConfigurationSUSTuned bounce and roll mode of ETMY suspension

I tuned the bounce and roll mode bandstops for ETMY, although it was difficult for me to tell if there was improvement with the bandstops on relative to the bandstops off because it seemed like the bounce and roll modes were being excited intermittently.  I'll take spectra with the filters both on and off during an evening next week.

  4719   Sun May 15 12:42:29 2011 kiwamuUpdateSUSf2p ratio adjutment done for all the suspensions

The f2p adjustment for all the suspensions are done (except for MC1,2,3)

Attachment 1: f2p_summary.pdf
f2p_summary.pdf f2p_summary.pdf f2p_summary.pdf f2p_summary.pdf f2p_summary.pdf f2p_summary.pdf f2p_summary.pdf
  4731   Tue May 17 16:53:22 2011 kiwamuUpdateSUSoplev sign was wrong on ETMY

Last night I found that the sign of the oplev control of PITCH on ETMY was wrong. I flipped it to the correct sign.

We've been locking the Y arm by feeding a signal back to ITMY  because pushing ETMY somehow made the lock unstable in the angular motion.

After the correction of the oplev contol sign, I was able to keep the lock robustly by pushing ETMY.

  4734   Tue May 17 19:38:32 2011 kiwamuUpdateSUSwrong connection on 1X5

Today Steve was working around the 1X5 rack to strain relief the cable jungles and the jungle is now getting less jungle.

During the work he disconnected and reconnected some cables.

So for a doublecheck I checked all the suspensions to see if the suspensions are still healthy or not.

Aha, then I found a mistake.

 

See the pictures below. It's a very subtle difference. This wrong connection prevented MC1 and MC3 from damping.

wrong.png correct.png

  4752   Fri May 20 01:02:50 2011 loose connection hunterUpdateSUSloose connection on ETMY rack

The UL signal of the shadow sensor on ETMY went to zero this evening.

This was due to a loose connection on the cross connection board on the 1Y4 rack.

In order to make them tighten, a combination of stand-offs and screws were installed on the connectors. They won't be loose any more.

ETMY_loose.png

  4775   Tue May 31 17:30:42 2011 steveUpdateSUSETMY damping restored

ETMY's watch dogs were found tripped. They were restored.

  4780   Thu Jun 2 16:23:42 2011 JamieUpdateSUSSUS control models updated to use new sus_single_control library part

A new library part was made for the single suspension controller (it was originally made from the c1scx controller), using the following procedure:

  1. Opened c1scx model (userapps/trunk/sus/c1/models/c1scx)
  2. Cut ETMX subsystem block out of SUS subsystem
  3. Pasted ETMX block into new empty library, and renamed it C1_SUS_SINGLE_CONTROL
  4. Tweaked names of inputs, and generally cleaned up internals (cosmetically)
  5. Saved library to: userapps/trunk/sus/c1/models/lib/sus_single_control.mdl

Once the new sus_single_control library part was made and the library was committed to the cds_user_apps repo, I replaced all sus controller subsystems with this new part, in:

  • c1scx
  • c1scy
  • c1sus (x5 for each vertex mass)

All models were rebuild, installed, and tested, and everything seems to be working fine.

  4791   Mon Jun 6 22:41:22 2011 ranaUpdateSUSSwitching problem in SUS models

Some weeks ago, Joe, Jamie, and I reworked the ETMY controls.

Today we found that the model rebuilds and BURT restores have conspired to put the SUS damping into a bad state.

1) The FM1 files in the XXSEN modules should switch the analog shadow sensor whitening. I found today that, at least on ETMY and ETMX, they do nothing. This needs to be fixed before we can use the suspensions.

2) I found all of the 3:30 and cts2um buttons OFF AGAIN. There's something certainly wrong with the way the models are being built or BURTed. All of our suspension tuning work is being lost as a consequence. We (Joe and Jamie) need to learn to use CONLOG and check that the system is not in a nonsense state after rebuilds. Just because the monitors have lights and the MEDM values are fluctuating doesn't mean that "ITS WORKING". As a rule, when someone says "it seems to work", that basically means that they have no idea if anything is working.

3) We need a way to test that the CDS system is working...

Attachment 1: a.pdf
a.pdf
  4792   Mon Jun 6 23:56:16 2011 rana, valeraConfigurationSUSETMX/ETMY OSEM whitening

 We measured the OSEM PD whitening transfer function of the ETMX OSEM UL whitening stage (D000210) by comparing the input signal to the whitening amplifier (single pin LEMO monitor) to the output signal - both were piped into the DAQ. The transfer function was close to constant 0 dB/180 deg independent of the whitening switch selection (FM1 filter engaged/disengaged)  up to ~20 Hz where we run out of coherence. All other ETMX and ETMY spectra at the input of the digital whitening compensation don't change when the whitening is switched on/off so by induction we conclude that all the ETMX/ETMY OSEM PD hardware whitening filters are not on.

  4793   Tue Jun 7 11:39:27 2011 JamieUpdateSUSNo binary output module in ETMY
Quote:

1) The FM1 files in the XXSEN modules should switch the analog shadow sensor whitening. I found today that, at least on ETMY and ETMX, they do nothing. This needs to be fixed before we can use the suspensions.

Joe discovered today that ETMY in fact has no binary output module at all, so there is actually no digital control of the whitening filters at ETMY.

We suspect that the ETMY binary output module was maybe harvested to put in the LSC rack, but we're not sure.

We found a spare binary output adapter pcb, which I will try to assemble into a module to install in ETMY.

This does not explain what's going on with ETMX, though.  ETMX has a binary output module, that appears to be properly hooked up.  I'll try to debug what's going on there as well.

In the mean time, I've removed the ETMX binary output module to use as a reference for putting together another identical module for ETMY.

  4802   Thu Jun 9 20:10:38 2011 kiwamuUpdateSUSETMY whitening filter : all off

I checked the state of the whitening filters for the ETMY shadow sensors.

Result : They've been OFF  (i.e. flat response).

 

(measurement and setup)

 I measured the transfer functions of the whitening board (D000210) by looking at the signal before and after the whitening stage.

 The whitening board handles five signals; UL, UR, LR, LL and SD, and there are five single-pin lemo outputs for each signal on the front panel.

A good thing on those lemo monitors is that their signals are monitored before the whitening stages.

Rana suggested me to use these signals for the denominator of the transfer functions and consider the sensor signals as excitation signals.

So I plugged those signals into extra ADC channels via an AA-board and measured the transfer functions.

In the measurement the coherence above 4 Hz was quite small while the suspension was freely swinging.

Therefore I had to excite the ETMY suspension by putting random noise in a frequency band from 5 Hz to 35 Hz to obtain better coherence.

 

(results)

 The response is flat over frequency range from ~ 0.2 Hz to ~40 Hz, see the plot below. 

According to the spectrum of each signal the measurements above 10 Hz are just disturbed by the ADC noise.

If the whitening filters are ON, a pole and zero are expected to appear at 30 Hz and 3 Hz respectively according to the schematic, but no such features.

ETMY_WF2.png

 

  4827   Thu Jun 16 00:43:36 2011 KojiUpdateSUSVertex SUS Binary Output Boxes were turned off / need investigation

- I was investigating the SUS whitening issue.

- I could not find any suspension which can handle the input whitening switch correctly.

- I went to 1X5 rack and found that both of the two binary output boxes were turned off.
As far as I know they are pulling up the lines which are switched by the open collector outputs.

- I tried to turn on the switch. Immediately I noticed the power lamps did not work. So I need an isolated setup to investigate the situation.

- The cables are labelled. I will ask steve to remove the boxes from the rack.

  4828   Thu Jun 16 08:45:14 2011 steveUpdateSUSVertex SUS Binary Output Boxes removed

Quote:

- I was investigating the SUS whitening issue.

- I could not find any suspension which can handle the input whitening switch correctly.

- I went to 1X5 rack and found that both of the two binary output boxes were turned off.
As far as I know they are pulling up the lines which are switched by the open collector outputs.

- I tried to turn on the switch. Immediately I noticed the power lamps did not work. So I need an isolated setup to investigate the situation.

- The cables are labelled. I will ask steve to remove the boxes from the rack.

 I shut down damping to the Vertex optics and removed Binary IO  Adapter chassy BO0 and BO1

About a week ago I discussed the BO0's power indicator lights with Kiwamu. They were  not on or they were blinking on-off.

I put screws into ps connectors in the back, but it did not helped.

Attachment 1: P1070894.JPG
P1070894.JPG
  4829   Thu Jun 16 23:19:09 2011 KojiUpdateSUSVertex SUS Binary Output Boxes removed

[Jamie, Koji]

- We found the reason why some of the LEDs had no light. It was because the LEDs were blown as they were directly connected to the power supply.
The LEDs are presumably designed to be connected to a 5V supply (with internal current-limiting resistor of ~500Ohm). The too much current
with the 15V (~30mA) made the LED blown, or the life-time of them shorter.

- Jamie removed all of the BO modules and I put 800Ohm additional resister such that the resultant current is to be 12mA.
The LEDs were tested and are fine now.

- The four BO boxes for C1SUS were restored on the rack. I personally got confused what should be connected where
even though I had labeled for BO0 and BO1. I just have connected CH1-16 for BO0. The power supplies have been connected only to BO0 and BO1.

- I tested the whitening of PRM UL sensor by exciting PRM UL sensor. The transfer function told us that the pendulum response can be seen
up to 10-15Hz. When the whitening is on, I could see the change of the transfer function in that freq band. This is good.
So the main reason why I could not see theis was that the power supply for the BOs were not turned on.

- I suppose Jamie/Joe will restore all of the BO boxes on the racks tomorrow. I am going to make a test script for checking the PD whitenings.

  4831   Fri Jun 17 08:03:48 2011 steveUpdateSUSITMY sus damping restored

ITMY sus damping restored.

 

  4833   Fri Jun 17 17:02:15 2011 JamieUpdateSUSETMX/ETMY binary output modules (re)installed, not yet tested

I have installed a new binary output module in ETMY, where there was none previously.  It is installed, powered (with working LEDs), hooked up (to the binary output card and the cross connect), but it hasn't been fully tested yet.

I also re-installed the binary output module in ETMX, with newly modified power-indicator LEDs.

Both modules are fully installed, but they have not yet been fully tested to confirm that they are indeed switching the whitening and de-whitening filters.

  4836   Mon Jun 20 09:04:13 2011 steveUpdateSUSETMX damping restored

ETMX sus damping restored

  4839   Mon Jun 20 11:04:03 2011 NicoleUpdateSUSWork Plan for Week 2

Here is my work plan for this week:

Current Week Plan (Week 2) (As of 6/17/11)

 

Setting Up for Horizontal Displacement Measurements

1) Help Steve clean small table for experiment

2) Remove aluminum base from TT suspension

3) Mount shaker onto table base

4) Mount horizontal slider onto table base

5) Connect TT suspension, shaker, and horizontal slider

Begin Assembly of Sensors

1) Begin building circuit for displacement photosensors

2) Calibrate photosensor using linear regions of power versus distance curves

3) Circuit box for photosensors?

  4844   Mon Jun 20 18:12:20 2011 NicoleUpdateSUSSmall Table Cleaned and Levelled

P6220198.JPG

The small optical bench (next to the MC-2 Chamber and the tool box tower) has been cleared of the misc. object previously on it, cleaned, and leveled (after much calibration X___X).

PLEASE, PLEASE, PLEASE do NOT MOVE OR HIT THE TABLE! It was incredibly painful to level.

This is how leveling the table made me feel...

P6220199.JPG

VERY SAD...so do not move please!

The shaker has already been moved to the table and the amplifier for my shaking experiment is located behind the table (not on the table, as to prevent scratching).

 

 

  4853   Wed Jun 22 12:24:44 2011 NicoleSummarySUSMidweek 2 Work Summary

I have made my transfer function model and posted it to the suspension wiki. Here is the link to my model!

Bode Plot Model

Please let me know if there need to be any adjustments, but I have posted the bode plots, a model image, and an explanation of why I think it's right! ^ ___^ V

I am currently working on the photo sensor circuit for the displacement detector. So far, I have gotten the infared LED to light up! ^ ___^ V

I am now trying to get a plot of forward voltage versus current for the LED. HOPEFULLY it will match the curve provided in the LED datasheet.

I'm using the bread board circuit box and when I'm not working at the bench, I have signs posted. PLEASE DO NOT REMOVE THE CONNECTIONS! It is

fine to move the bread board circuit box, but please do not disturb the connections > ____<

Here is a photo of the workspace

P6220200.JPG

  4858   Wed Jun 22 18:41:23 2011 NicoleSummarySUSBROKEN bread board circuit box and L9337 LED Current Versus Voltage Curve

NOTE: The potentiometers on the bread board circuit box (the one I have been using with the signal generator, DC power, LED displays, and pulse switches) is BROKEN!

The potential across terminals 1 and 2 (also 2&3) fluctuates wildly and there dial does not affect the potential for the second potentiometer (the one with terminals 4, 5, and 6).

This has been confirmed by Koji and Jaimie.  PS I didn't break it! >____<

 

NEVERTHELESS, using individual resistors and the 500 ohm trim resistor, I have managed to get the current versus forward voltage plot for the Hamamatsu L9337 Infared LED

LED_I_vs_V_exp_plot.png