ID |
Date |
Author |
Type |
Category |
Subject |
3893
|
Thu Nov 11 07:26:03 2010 |
Alberto | Update | Electronics | REFL11 Photodiode Not Working |
Quote: |
[Koji and Kevin]
I was trying to characterize the REFL11 photodiode by shining a flashlight on the photodiode and measuring the DC voltage with an oscilloscope and the RF voltage with a spectrum analyzer. At first, I had the photodiode voltage supplied incorrectly with 15V between the +15 and -15 terminals. After correcting this error, and checking that the power was supplied correctly to the board, no voltage could be seen when light was incident on the photodiode.
We looked at the REFL55 photodiode and could see ~200 mV of DC voltage when shining a light on it but could not see any signal at 55 MHz. If the value of 50 ohm DC transimpedance is correct, this should be enough to see an RF signal. Tomorrow, we will look into fixing the REFL11 photodiode.
|
I just wanted to remind you that the most up to date resource about the RF system upgrade, including photodiodes, is the SVN.
https://nodus.ligo.caltech.edu:30889/svn/trunk/alberto/40mUpgrade/RFsystem/
There you can find everything: measurements, schematics, matlab scripts to plot and fit, etc. Poke around it to find what you need.
For instance, the schematic of the modified REFL11 photodiode is at:
https://nodus.ligo.caltech.edu:30889/svn/trunk/alberto/40mUpgrade/RFsystem/RFPDs/REFL11/REFL11Schematics/40mUpgradeREFL11schematic.pdf
Because I was doing new things all the time, the wiki is not up to date. But the SVN has all I've got. |
3896
|
Thu Nov 11 13:54:05 2010 |
kiwamu | Update | Green Locking | 80MHz VCO : about PLL hold-in range |
The hold-in range of the PLL must be greater than +/- 4MHz in order to bring the arm cavity to its resonance.
(Hold-in range is the range of frequencies over which the PLL can track the input signal.)
However as I mentioned in the past elog (see this entry), the PLL showed a small hold-in range of about +/- 1MHz which is insufficient.
In this entry I explain what is the limitation factor for the hold-in range and how to enlarge the range.
(Requirement for hold-in range )
We have to track the frequency of the green beat signal and finally bring it to a certain frequency by controlling the cavity length of the arm.
For this purpose we must be able to track the beat signal at least over the frequency range of 2*FSR ~ +/- 4MHz.
Then we will be able to have more than two resonances, in which both the end green and the PSL green are able to resonate to the arm at the same time.
And if we have just two resonances in the range, either one of two resonances gives a resonance for both IR and green. At this phase we just bring it to that frequency while tracking it.
Theoretically this requirement can be cleared by using our VCO because the VCO can drive the frequency up to approximately +/- 5MHz (see this entry)
The figure below is an example of resonant condition of green and IR. The VCO range should contain at least one resonance for IR.
(In the plot L=38.4m is assumed)

(an issue)
However the measured hold-in range was about +/- 1MHz or less. This is obviously not large enough.
According to a textbook[1], this fact is easily understandable.
The hold-in range is actually limited by gains of all the components such as a phase detector's, a control filter's and a VCO's gain.
Finally it is going to be expressed by,
[hold-in range] = G_pd * G_filter * G_vco

At the PD (Phase Detector which is a mixer in our case) the signal does not exceed G_pd [V] because it appears as G_pd * sin(phi).
When the input signal is at the edge of the hold-in range, the PD gives its maximum voltage of G_pd to maintain the lock.
Consequently the voltage G_pd [V] goes through to G_filter [V/V] and G_vco [Hz/V].
This chain results the maximum pushable frequency, that is, hold-in range given above equation.
In our case, the estimated hold-in range was
[hold-in range] ~ 0.4 [V] * 3 [V/V] * 1 [MHz/V]
= 1.2 [MHz]
This number reasonably explains what I saw.
In order to enlarge the hold-in range, increase the gain by more than factor of 5. That's it.
* reference [1] "Phase-Locked Loops 6th edition" Rolan E. Best |
3898
|
Thu Nov 11 17:47:36 2010 |
kiwamu | Update | Green Locking | 80MHz VCO : improve PLL hold-in range and put a boost |
In order to enlarge the hold-in range I modified the control filter and increased the gain by factor of 25 in the PLL.
It successfully enlarged the range, however the lock was easily broken by a small frequency change.
So I put a low frequency boost (LFB) and it successfully engaged the PLL stiffer.
Now it can maintain the lock even when the frequency disturbance of about 1MHz/s is applied.
(enlargement of the hold-in range)
I modified the control filter by replacing some resistors in the circuit to increase the gain by factor of 25.
- R18 390 [Ohm] => 200 [Ohm]
- R20 1000 [Ohm] => 5000 [Ohm]
- R41 39 [Ohm] => 10 [Ohm]
This replacement also changes the location of the pole and the zero
- pole 1.5 [Hz] => 0.3 [Hz]
- zero 40 [Hz] => 159 [Hz]
Note that this replacement doesn't so much change the UGF which was about 20 kHz before.
It becomes able to track the input frequency range of +/- 5MHz if I slowly changes the frequency of the input signal.
However the PLL is not so strong enough to track ~ 1 kHz / 0.1s frequency step.
(make the PLL stiffer : a low frequency boost)
One of the solution to make the PLL stiffer is to put a boost filter in the loop.
I used another channel to more drive the VCO at low frequency. See the figure below.

The 80MHz VCO box originally has two input channels, one of these inputs was usually disabled by MAX333A.
This time I activated both two input channels and put the input signal to each of them.
Before signals go to the box, one of the signal path is filtered by SR560. The filter has G=20000, pole=0.3Hz. So it gives a big low frequency boost.

Once the PLL was achieved without the boost, I increased the filter gain of SR560 to 20000 because locking with the boost is difficult as usual.
|
3899
|
Thu Nov 11 18:05:55 2010 |
valera | Update | PSL | PMC mode matching at full laser power |
The PMC mode matching was initially done at low power ~150 mW. It was expected and found that at full power ~2 W (injection current 2.1 A) the mode matching got much worse:
the visibility degraded from 80% to 50% (1 - refl locked/refl unlocked) . The thermal lensing could be in the laser, EOM, or FI.
The first attached plot shows the scan of the beam after the EOM at low and full laser power. At full power the waist position is 10 mm after the turning mirror after the EOM and the waist size is 310 um.
The second plot shows the ABCD calculation for the mode matching solution. 
I removed the MM lens PLCX-25.4-77.3-C and placed the PLCX-25.4-180.3-UV about 20 mm after the first PMC periscope mirror (the second mirror after the EOM).
The PMC visibility improved to 94% and the power through the PMC, as measured by the PMC transmission PD, went up by a factor of 2. |
Attachment 1: scan.pdf
|
|
Attachment 2: pmc2-abcd.png
|
|
3900
|
Thu Nov 11 21:07:49 2010 |
josephb | Update | CDS | Plugged c1iscex into DAQ network - still causes network slowdown |
I connected the c1iscex computer to the dedicated DAQ network switch (located in 1X7).
This does not seem to have helped c1iscex stop spewing out "OMX: Failed to find peer index of board 00:00:00:00:00:00 (Peer Not Found in the Table)" at the rate of ~1 Gigabyte per minute.
c1iscex is currently off until a solution can be found. |
3902
|
Fri Nov 12 00:13:34 2010 |
Suresh | Update | SUS | ETM assembly started |
[Jenne, Suresh]
Selection of ETMs
Of the four ETMs (5,6,7 and 8) that are with us Koji gave us two (nos. 5 and 7) for use in the current assembly. This decision is based on the Radius of Curvature (RoC) measurements from the manufacturer (Advanced Thin Films). As per their measurements the four ETMs are divided into two pairs such that each pair has nearly equal RoC. In the current case, RoCs are listed below:
Radii of Curvature of ETMs
ETM # |
RoC from Coastline Optics (m) |
RoC from Advanced Thin Films (m) |
5 |
57.6 |
60.26 |
6 |
57.4 |
54.58 |
7 |
57.1 |
59.48 |
8 |
57.9 |
54.8 |
The discrepancy between the measurements from these two companies leaves us in some doubt as to the actual radius of curvature. However we based our current decision on the measurement of Advanced Thin Films.
Assembly of ETMs
We drag wiped both the ETMs (5 and 7) and placed them in the Small Optic Gluing Fixture. The optics are positioned with the High Reflectace side facing downwards and with the arrow-mark on the Wire Standoff side (big clamp). We then used the microscope to position the Guide Rod and the Wire Standoff in the tangential direction on the ETMs (step 4 of the procedure specified in E010171-00-D)
We will continue with the rest of the assembly tomorrow.
|
3903
|
Fri Nov 12 00:42:11 2010 |
rana, koji | Update | IOO | MC alignment |
We decided to ignore the computer script outputs for the beam positions and use instead the eyeball method to get the beam into the MC:
- Adjust PSL launch beam to get the beam centered on IM1.
- Eyeball the beam to hit the center of MC1. We can get this pretty good by using the brackets to get the vertical and using the centering of the input/refl beams to center it horizontally.
- Use MC3 suspension to hit the center of MC2. We did this by hitting each of the 3 EQ stop screw heads and triangulating the MC3 bias settings.
- Use MC2 bias to hit the center of MC1.
- Use MC1 to get good flashes.
- Use all 3 MC sus biases to maximize the transmitted light and minimize the REFL DC.
With this rough alignment in place, we leave it to Yuta to finish the coil balancing and the A2L. We will have an aligned MC in the morning and will start the BS chamber alignment. |
3904
|
Fri Nov 12 02:51:20 2010 |
Kevin | Update | Electronics | Photodiode Testing |
[Jenne and Kevin]
I started testing the REFL55 photodiode. With a light bulb, I saw ~270 mV of DC voltage from the photodiode but still could not see any RF signal. I connected the RF out from the spectrum analyzer to the test input and verified that the circuit was working.
I then set up the AM laser and looked at the laser light with REFL11 and an 1811 photodiode. I was able to see an RF signal and verified that the resonant frequency is 55 MHz.
The current setup is not very reliable because the laser is not mounted rigidly. Next, I will work on making this mounting more reliable and will continue to work on finding an RF signal with a flashlight. |
3905
|
Fri Nov 12 06:10:24 2010 |
yuta | Update | IOO | MC aligned (without coil balancing) |
Background:
Last night, we found that one of DAC channels are poorly connected, so we fixed the connectors.
Rana and Koji used their incredible eyeballs to roughly align MC.
Next thing to do is to balance the coils, but it takes some time for the setup.
So, we decided to do A2L anyway.
What I did:
Using the last steering mirror at PSL table and IM1, changed the incident beam direction to align MC.
Result:

I was amazed by their eyeballs.
I turned the nobs of SM@PSL and IM1 in small increments, so I never lost TEM00.
Is it enough?:
The length of the whole faraday is about 20cm and aperture diameter is about 12mm. (I couldn't measure the aperture size of the core)
The beam is about 9mm diameter at 6w.
So, if the beam is vertically tilted at more than ~3/200rad, it(6w) cannot go through.
3/200 rad is about 20% difference in position at MC1 and MC3.
So, the result meets the requirement.
Also, assuming that coils have 5% imbalance, the beam position I measured have ~3% error.
So, to do more precise beam centering, we need to balance the coils. |
3906
|
Fri Nov 12 10:49:34 2010 |
josephb, valera | Update | CDS | Test of ADC noise |
Test:
Look at the effects of the ADC voltage range on the ADC noise floor.
ADC input was terminated with 50 ohms. We then looked at the channel with DTT. This was at +/- 10 V range. We used C1:SUS-PRM_SDSEN_IN1 as the test channel.
The map.c file (in /opt/rtcds/caltech/c1/core/advLigoRTS/src/fe/ ) then had two lines added at line 766.
//JCB temporary 2.5V test, remove me
adcPtr[devNum]->BCR &= 0x84240;
This hard coded the 2.5 V range (we default to the 10 V range at the moment).
We then rebuilt the c1x02 model and reran the test.
Finally, we reverted the code change to map.c and rebuilt c1x02.
Results:
I've attached the DTT output of the two tests.
It appears the ADC is limited by 1.6 uV/rtHz. Hence the increase in noise in counts by a factor of 4 when we drop to +/- 2.5 V from +/- 10 V. |
Attachment 1: ADC_noise.pdf
|
|
3909
|
Fri Nov 12 13:12:55 2010 |
kiwamu | Update | PSL | increased NPRO power |
I maximized the laser power by rotating the HWP after the NPRO.
If someone works on the MC locking, one should decrease it again.
|
3910
|
Fri Nov 12 19:24:56 2010 |
Koji | Update | CDS | Test of ADC noise |
[Koji Yuta]
We found one of the ADC cables were left unconnected. This left the MC suspensions uncontrollable through the whole afternoon.
Please keep the status updated and don't forget to revert the configuration...
Quote: |
Test:
Look at the effects of the ADC voltage range on the ADC noise floor.
ADC input was terminated with 50 ohms. We then looked at the channel with DTT. This was at +/- 10 V range. We used C1:SUS-PRM_SDSEN_IN1 as the test channel.
The map.c file (in /opt/rtcds/caltech/c1/core/advLigoRTS/src/fe/ ) then had two lines added at line 766.
//JCB temporary 2.5V test, remove me
adcPtr[devNum]->BCR &= 0x84240;
This hard coded the 2.5 V range (we default to the 10 V range at the moment).
We then rebuilt the c1x02 model and reran the test.
Finally, we reverted the code change to map.c and rebuilt c1x02.
Results:
I've attached the DTT output of the two tests.
It appears the ADC is limited by 1.6 uV/rtHz. Hence the increase in noise in counts by a factor of 4 when we drop to +/- 2.5 V from +/- 10 V.
|
|
3912
|
Sat Nov 13 15:53:05 2010 |
yuta | Update | CDS | diagonalization of MC input matrix |
Motivation:
MC is aligned from the A2L measurement, but to do the beam centering more precisely, we need coils to be balanced.
There are several ways to balance the coils, like using oplev or WFS QPD RF channels.
But oplev takes time to setup, especially for MC3. Also, c1ioo WFS channels were newly setup and haven't been checked yet.
So, I decided to use OSEM sensors.
An OSEM sensor itself is sensitive to every DOF of an optic motion, but we can diagonalize them using 4 OSEM sensors and proper input matrix.
Method:
1. Measure transfer functions between
ULSEN and URSEN (H_UR(f))
ULSEN and LRSEN (H_LR(f))
ULSEN and LLSEN (H_LL(f))
2. Make a matrix A.
A = [[ 1 1 1 ]
[ H_UR(f_pos) H_UR(f_pit) H_UR(f_yaw)]
[ H_LR(f_pos) H_LR(f_pit) H_LR(f_yaw)]
[ H_LL(f_pos) H_LL(f_pit) H_LL(f_yaw)]]
where f_dof are resonant frequencies.
3. A is
s = Ad
where vectors s^T=[ULSEN URSEN LRSEN LLSEN] and d^T=[POS PIT YAW].
So,
d = Bs = (A^TA)^(-1)A^Ts
where A^T is transpose of A.
B is the input matrix that diagonalizes 3 DOFs.
What I did:
1. Measured the TFs using diaggui and exported as ASCII.
2. Made a script that reads that TF file, calculates and sets a new input matrix B.
/cvs/cds/caltech/users/yuta/scripts/inputmatrixoptimizer.py
You need to set resonant frequencies to use the script.
New input matrices for MCs are;
C1:SUS-MC1_INMATRIX
[[ 1.17649712 0.94315611 0.85065054 1.02969624]
[ 0.55939288 1.28066594 -0.85235358 -1.3075876 ]
[ 1.23467139 -0.74521928 -1.29394051 0.72616882]]
C1:SUS-MC2_INMATRIX
[[ 1.12630748 1.01451545 0.9013457 0.95783137]
[ 1.03043025 0.67826036 -1.37270598 -0.91860341]
[ 0.83546271 -1.26311029 -0.6456881 1.2557389 ]]
C1:SUS-MC3_INMATRIX
[[ 1.18212117 1.26419447 0.77744155 0.77624281]
[ 0.79344415 0.84959646 -1.10946339 -1.247496 ]
[ 1.00225331 -0.84807863 -1.21772132 0.93194674]]
I ignored SIDE this time.
Result:
Spectra of each SUSDOF_IN1_DAQ before diagonalization (INMATRIX elements all 1 or -1) were

After diagonalization, spectra are

As you can see, each SUSDOF has only single peak (and SIDE peak) after the diagonalization.
SUSSIDE still has 4 peaks because SIDE is not included this time.
For MC2, POS to SUSPIT and POS to SUSYAW got worse. I have to look into them.
Effect of resonant frequency drift:
As you can compare and see from the spectra above, resonant frequencies of MC1 are somehow drifted(~0.5%) from Nov 9 to Nov 13.
If resonant frequency you expected was wrong, calculated input matrix will be also wrong.
The effect of 0.5% drift and wrong input matrix can be seen from this spectra. DOFs are not clearly separated.

Plan:
- learn how to use diaggui from command line and fully automate this process
- balance the coils using these diagonalized SUSPOS, SUSPIT, SUSYAW |
3915
|
Sun Nov 14 11:56:59 2010 |
valera | Update | CDS | Test of ADC noise |
We missed a factor of 2 in the ADC calibration: the differential 16 bit ADC with +/-10 V input has 20 V per 32768 counts (1 bit is for the sign). I confirmed this calibration by directly measuring ADC counts per V.
So the ADC input voltage noise with +/-10V range around 100 Hz is 6.5e-3 cts/rtHz x 20V/32768cts = 4.0 uV/rtHz. Bummer. 
The ADC quantization noise limit is 1/sqrt(12 fs/2)=1.6e-3 cts/rtHz. Where the ADC internal sampling frequency is fs=64 kHz. If this would be the limiting digitization noise source then the equivalent ADC input voltage noise would be 1 uV/rtHz with +/-10 V range. |
3916
|
Sun Nov 14 16:26:31 2010 |
jenne, valera | Update | Electronics | SRM side OSEM noise with no magnet |
We realized that the SRM sensors are connected to the readout but just sitting on the BS in vacuum table with no magnets and therefore no shadows in them. We swapped the inputs to the SRM and PRM satellite boxes to use the higher transimpedance gain of the PRM side sensor. The attached plot shows the current spectrum in this configuration. The PD readback voltage was 9.5 V. Since this is close to the rail we put a slightly higher voltage into the AA of this channel to test that we can read out more ADC counts to make sure we are not saturating. The margin was 15800 vs 15400 counts with p-p of 5 counts on the dataviewer 1 second trend. We returned all cables to nominal configuration.
The calibration from A to m is 59 uA/1 mm. |
Attachment 1: SRM-SD-Current-NoMagnet.pdf
|
|
3917
|
Sun Nov 14 16:40:46 2010 |
Jenne | Update | Treasure | Activities related to OSEM measurement |
[Valera, Jenne]
We pondered the idea of clamping the PRM optic to measure the OSEM noise. So we opened up the BS tank to give this a try. We rediscovered that Jenne is too short to reach the other side of the PRM tower, so we couldn't fully clamp the optic (when is Jaime coming again? He's kind of tall...) If we only did the back 2 EQ stops, the optic would still be able to rock, and thus defeat the purpose of clamping anyway. So we didn't go for it.
While we were in there we saw that the SRM OSEMs were just hanging out on the table, and decided to go with them. See Valera's elog for details on our measurement. We closed up the tank without making any changes to anything.
In other news, we still need to figure out how to change up the connectors to get those OSEMs over to the ITM table. This needs to happen pretty soonish. |
3918
|
Mon Nov 15 04:57:10 2010 |
rana | Update | Electronics | SRM side OSEM noise with no magnet |
IF I believe this calibration and IF I believe that the noise is the same with no magnet in there, then its almost 1 nm/rHz @ 1 Hz.
I am guessing that Jenne's calculation will show that this is an unacceptably high level of OSEM sensor noise, OAF-wise. |
3919
|
Mon Nov 15 11:13:12 2010 |
josephb | Update | CDS | Modified rc.local to not start mx_streams automatically |
Problem:
c1iscex floods the network with about 1 gigabyte of error messages in a few seconds, writing to a log file in /opt/rtcds/caltech/c1/target/fb/logs/
Temporary change:
I commented out the following line in the rc.local file on the fb machine in the /diskless/root/etc/ directory:
#nice --20 ./mx_stream -s "$SYSTEMS" -d fb:0 >& logs/$HOSTNAME.log&
This disables the automatic start up of the mx_streams code on all the front ends. This will prevent the network being brought to its knees by c1iscex while we debug the problem.
It also means on a reboot of the front ends, the mx_stream process needs to be started by hand until this change is reverted.
To do this, log into the front end and then change directory to /opt/rtcds/caltech/c1/target/fb
For c1sus, run:
./mx_stream -s c1x02 c1sus c1mcs c1rms c1rfm -d fb:0
For c1ioo, run:
./mx_stream -s c1x03 c1ioo -d fb:0
|
3920
|
Mon Nov 15 11:52:22 2010 |
kiwamu | Update | Green Locking | PLL with real green signal |

Stabilizing the beat note frequency using Yuta's temperature servo (see this entry)
I was able to acquire the PLL of 80MHz VCO to the real green signal.
Some more details will be posted later. |
3921
|
Mon Nov 15 14:36:37 2010 |
Koji | Update | PSL | C1PSL rebooted? |
Has C1PSL rebooted? Has burtrestore been forgotten? Even without elog?
We found some settings are wrong and the PMC has pretty low gain. |
3922
|
Mon Nov 15 14:42:01 2010 |
Aidan | Update | PSL | C1PSL rebooted? |
Yeah. Joe and I rebooted c1psl a couple of times this morning. I didn't realize the burtrestore wasn't automatic.
Quote: |
Has C1PSL rebooted? Has burtrestore been forgotten? Even without elog?
We found some settings are wrong and the PMC has pretty low gain.
|
|
3923
|
Mon Nov 15 15:01:51 2010 |
kiwamu | Update | IOO | realigned the wideband EOM |
Since we are going to lock the MC today, I aligned it back to the default place.
Quote: #3888 |
For Yuta's business, I intentionally misaligned the wideband EOM slightly to Yaw direction.
|
|
3925
|
Mon Nov 15 16:09:44 2010 |
Jenne | Update | VAC | ETMX, ETMY, ITMY chambers opened |
[Steve, Jenne, Suresh, Koji]
The remaining test mass chambers have been opened, and have light doors in place. Now we can do all of the rest of the IFO alignment, and then (hopefully) button up before the New Year. |
3926
|
Mon Nov 15 16:26:46 2010 |
josephb | Update | CDS | c1iscex is now running and the network hasn't died |
Problem:
c1iscex was spamming the network with error messages.
Solution:
Updated the front end codes to current standards (they were on the order of months out of date). After fixing them up and rebuilding the codes on c1iscex, it no longer had problems connecting to the frame builder.\
Status:
I can look at test points for ETMX. It is not currently damping however.
To Do:
Move filters for ETMX into the correct files.
Need to add a Binary output blue and gold box to the end rack, and plug it into the binary output card. Confirm the binary output logic is correct for the OSEM whitening, coil dewhitening, and QPD whitening boards.
Get ETMX damped.
Figure out what we're going to do with the aux crate which is currently running y-end code at the new x-end. Koji suggested simply swapping auxilliary crates - this may be the easiest. Other option would be to change the IP address, so that when it PXE boots it grabs the x-end code instead of the y-end code.
Current CDS status:
MC damp |
dataviewer |
diaggui |
AWG |
c1ioo |
c1sus |
c1iscex |
RFM |
Sim.Plant |
Frame builder |
TDS |
|
|
|
|
|
|
|
|
|
|
|
|
3927
|
Mon Nov 15 17:10:59 2010 |
kiwamu | Update | Green Locking | PLL with real green signal |
I checked the slow servo and the PLL of 80MHz VCO using the real green beat note signal.
The end laser is not locked to the cavity, so basically the beat signal represents just the frequency fluctuation of the two freely running lasers.
The PLL was happily locked to the green beat note although I haven't fedback the VCO signal to ETMX (or the temperature of the end laser).
It looks like we still need some more efforts for the frequency counter's slow servo because it increases the frequency fluctuation around 20-30mHz.
(slow servo using frequency counter)
As Yuta did before (see his entry), I plugged the output of the frequency counter to an ADC and fedback the signal to the end laser temperature via ezcaservo.
The peak height of the beat note is bigger than before due to the improvement of the PMC mode matching.
The peak height shown on the spectrum analyzer 8591E is now about -39dBm which is 9dB improvement.
The figure below is a spectra of the frequency counter's readout taken by the spectrum analyzer SR785.

When the slow temperature servo is locked, the noise around 20-30 mHz increased.
I think this is true, because I was able to see the peak slowly wobbling for a timescale of ~ 1min. when it's locked.
But this servo is still useful because it drifts by ~5MHz in ~10-20min without the servo.
Next time we will work on this slow servo using Aidan's PID control (see this entry) in order to optimize the performance.
In addition to that, I will take the same spectra by using the phase locked VCO, which provides cleaner signal.
(acquisition of the PLL)
In order to extract a frequency information more precisely than the frequency counter, we are going to employ 80MHz VCO box.
While the beat note was locked at ~ 79MHz by the slow servo, I successfully acquired the PLL to the beat signal.
However at the beginning, the PLL was easily broken by a sudden frequency step of about 5MHz/s (!!).
I turned off the low noise amplifier which currently drives the NPRO via a high-voltage amplifier, then the sudden frequency steps disappeared.
After this modification the PLL was able to keep tracking the beat signal for more than 5min.
(I was not patient enough, so I couldn't stand watching the signal more than 5min... I will hook this to an ADC)
Quote: #3920
|
Some more details will be posted later.
|
|
3928
|
Mon Nov 15 22:24:28 2010 |
Suresh | Update | SUS | Selection of Magnets |
I have selected a set of 16 magnets which have a B field between 900 to 950 Gauss (5% variation) when measured in the following fashion.
I took a Petri-dish, of the type which we usually use for mixing the glue, and I placed a magnet on its end. I then brought the tip of the Hall-probe into contact with the Petri-dish from the opposite side and adjusted the location (and orientation) of the probe to maximise the reading on the Gauss meter.
The distribution of magnets observed is listed below
Range of B Field (Gauss) |
# of Magnets |
800-849 |
2 |
850-899 |
6 |
900-950 |
|
The set of sixteen has been have been placed inside two test tubes and left on the optical bench (right-side) in the clean room.
|
3929
|
Tue Nov 16 03:33:22 2010 |
yuta | Update | IOO | aligned Faraday, beam reached SM just before PRM |
(Koji, Yuta)
We aligned the Faraday after MC and we are now ready to install PRM.
Background:
MC was roughly aligned (beam spot ~0.7mm from the actuation center).
So, we started aligning in-vac optics.
First thing to align was the Faraday after MC3.
What we did:
1. Ran A2L.py for confirmation.(Second from the last measurement point on the A2L result plot)
2. Aligned the Faraday so that MC3 trans can go through it. We moved the Faraday itself, while we didn't touch IM2.
We turned the pitch nob of the last steering mirror at PSL table in CCW slightly in order to lower the beam at the Faraday by ~1mm.
3. During the alignment, we found that the polarization of the incident beam was wrong. It should have been S but it was P.
As there is the HWP right before the EOM, Rana rotated it so as to have the correct polarization of S on the EOM and the MC.
Note that the PMC and the main interferometer are configured to have P-pol while the MC is to have S-pol.
4. Setup the video camera to monitor the entrance aperture of the Faraday. It required 4 steering mirrors to convey the image to the CCD.
5. Moved all of the OSEMs for MC1 and MC3 so that the sensor output can have roughly half of their maxima.
6. Ran A2L.py. (The last measurement point on the A2L result plot)
7. Aligned the IO optics so that the beam goes Faraday -> MMT1 -> MMT2 -> SM3.
Result:
1. OSEM sensor outputs for MC1 and MC3 are;
(V) |
MC1 |
MC3 |
max |
current value |
max |
current value |
ULSEN |
1.3 |
0.708 |
1.37 |
0.699 |
URSEN |
1.4 |
0.845 |
1.71 |
0.796 |
LRSEN |
1.45 |
0.743 |
1.77 |
0.640 |
LLSEN |
1.56 |
0.762 |
1.56 |
0.650 |
SDSEN |
1.67 |
0.801 |
1.59 |
0.821 |
2. A2L result is;

The beam position slightly got lower(~0.2mm), because we touched SM at PSL table.
Alignment slider values changed because we moved MC1 and MC3 OSEMs.
3. Now, MC_RFPD_DCMON is ~0.39 when MC unlocked and ~0.083 when locked.
So, the visibility of MC is ~79% (for S-pol).
4. Now the incident beam to the MC has S polarization, the cavity has higher finesse. This results the increased MC trans power.
It was ~8e2 when the polarization was P, now it is ~4.2e3 when the MC is locked.
5. The beam reached SM3 at BS table. The alignment of the SM2, MMT1, MMT2 were confirmed and adjusted.
6. All pieces of the leftover pizza reached my stomach.
Plan:
- Install PRM to the BS chamber.
- Align PRM and get IFO reflection beam out to the AP table
|
3930
|
Tue Nov 16 09:02:54 2010 |
Aidan | Update | Green Locking | PID loop - calibration of SR620 output |
[Aidan, Kiwamu]
Kiwamu and I roughly calibrated the analogue output from the SR620 frequency counter yesterday. The input channel, intuitively named C1:PSL-126MOPA_126MON, now reads the measured frequency in MHz with an error of about 0.1MHz - this is, I think, due to the bit noise on the D/A conversion that Kiwamu discovered earlier. That is, the output range of the SR620 corresponds to around 100MHz and is digitized at 10-bit resolution, and ...
100MHz/(10^2) ~= 0.098MHz. [Sad Face]
Calibration:
We set the EPICS range to [-100, 100] (corresponding to [-5V, 5V]), connected a Marconi to the Freq Counter, input a variety of different frequencies and measured the counts on the EPICS channel.
The linear fit to the calibration data was F = 2.006*EPICScount - 0.2942. From this we worked out the maximum and the minimum for the range settings that give the channel in MHz: EGUF = -200.8942 and EGUL = 200.3058. The previous range was [-410, 410]
Calibration of SR620 analogue output
Input Frequency (MHz) |
Measured EPICS Value |
10 |
5.191 |
20 |
9.98 |
30 |
15.21 |
40 |
20.00 |
50 |
25.18 |
60 |
29.99 |
70 |
35.18 |
71 |
35.565 |
72 |
35.9894 |
73 |
36.3861 |
74 |
37.17 |
75 |
37.576 |
76 |
37.9669 |
77 |
38.3575 |
78 |
39.166 |
79 |
39.5691 |
80 |
39.978 |
|
|
|
3931
|
Tue Nov 16 10:47:45 2010 |
Aidan | Update | Green Locking | Rebooted c1psl - added new GRNBEAT_FREQ channel |
I restored C1:PSL-126MOPA_126MON to its original settings (EGUF = -410, EGUL = 410) and added a new calc channel called C1:LSC-EX_GRNBEAT_FREQ that is derived from C1:PSL-126MOPA_126MON. The calibration in the new channel converts the input to MHz.
grecord(calc, "C1:LSC-EX_GRNBEAT_FREQ")
{
field(DESC,"EX-PSL Green Beat Note Frequency")
field(SCAN, ".1 second")
field(INPA,"C1:PSL-126MOPA_126MON")
field(PREC,"4")
field(CALC,"0.4878*A")
}
I rebooted c1psl and burtrestored. |
3932
|
Tue Nov 16 12:47:30 2010 |
Aidan | Update | Green Locking | PID loop but no green |
The PID loop is ready to be run on the green beat note but, since the tanks are open, there is no green transmission from the end getting to the PSL table. Nevertheless, here's the screen for the PID loop. The loop script is still in my directory /cvs/cds/caltech/users/abrooks/GRNXSlowServo
The medm screen is attached. It shows the current beat note frequency in MHz ()
In c1auxey/ETMYaux.db I added a couple of channels. These are all displayed on the MEDM screen. I added them to autoBurt.req as well.
- C1:LSC-EX_GRNLSR_TEMP_NOM: the zero-volt setpoint temperature of the end laser (as set on the front panel of the Mephisto controller). This must be entered manually in EPICS as there is no way to read it remotely. [Sad Face]
- C1:LSC-EX_GRNLSR_TEMP_CALC: the sum of the zero-volt set point temperature and the offset temperature set by the input voltage from C1:LSC-EX_GREENLASER_TEMP
I rebooted c1auxey to get these to work.
Once we get the green beat back again, the PID loop should servo on the end laser temperature to drive the Beat Frequency to the Frequency Setpoint, C1:LSC-EX_GRN_PID_SETPT, which can be set by the pink slider.
RA: All MEDM screens must be in the proper MEDM directory!! Also, all perl scripts must have a .pl extension!!! Also, all scripts must be in the scripts directory even if they are in development!!! And all scripts should use 'env' rather than have absolute pathnames for the location of perl, csh, tcsh, python, etc. |
Attachment 1: Screenshot-C1LSC_EX_GRN_SLOW.adl.png
|
|
3933
|
Tue Nov 16 15:32:18 2010 |
valera | Update | Electronics | OSEM noise at the output of the satellite box |
I measured the SRM OSEM (no magnets at the moment) noise out of the satellite box with a SRS785 spectrum analyzer. I inserted a break out board into the cable going from the satellite box to the whitening board. The transimpedances of the SRM OSEMs are still 29.2 kOhm. The DC voltages out of the SRM satellite box are about 1.7 V. The signal was AC coupled using SR560 with two poles at 0.03 Hz and a gain of 10.
The noise is consistent with the one measured by the ADC except for the 3 Hz peak which does not show up in the ADC spectrum from Sunday. The peak appears in several channels I looked at. The instrument noise floor was measured by terminating the SR560 with 50 Ohm.
I recommend to change all OSEM transimpedance gains from 29 to 161 kV/A. Beyond this gain one will rail the AA filter module when the magnet is fully out of the OSEM.
The OSEM noise at 1 Hz is about factor of 10 above the shot noise. The damping loops impress this noise on the optics around the pendulum resonance frequency. Also the total contribution to the MC cavity length is sqrt(12) time the single sensor as there are 12 OSEMs contributing to MC length. The ADC noise is currently close but never the less not limiting the OSEM noise below 100 Hz. It can be further reduced by getting an extra factor of 2-3 in whitening gain above ~0.3 Hz. The rms of the ADC input of the modified PRM SD (R64 = 161 kOhm) channel is 10-20 cts during the day with damping loop off and whitening on.
The transimpedance amplifier LT1125CS is also not supposed to be limiting the noise. At 1 Hz the 1/f part of the noise: In<1pA/rtHz and Vn<20nV/rtHz. |
Attachment 1: osemnoise.pdf
|
|
3934
|
Tue Nov 16 16:00:26 2010 |
Aidan | Update | Green Locking | PID loop but no green |
Quote: |
RA: All MEDM screens must be in the proper MEDM directory!! Also, all perl scripts must have a .pl extension!!! Also, all scripts must be in the scripts directory even if they are in development!!! And all scripts should use 'env' rather than have absolute pathnames for the location of perl, csh, tcsh, python, etc.
|
That's not unreasonable. But if we try
grep "perl" /cvs/cds/rtcds/caltech/c1/scripts/*/*
you can see that we've got a fair amount of housekeeping to attend to. We might want to think about tidying up the scripts directory as part of the cds upgrade.
|
3935
|
Tue Nov 16 21:42:31 2010 |
rana | Update | CDS | Screen Time Fix |
I learned today that the following python code will do a find/replace to fix the TIME string on any MEDM screen which has a whited out time field.
Previously, this field was sourced from the c1dscepics of c1losepics process. Now we have to get it from the IOO or SUS front ends
Here's the python code:
import re
o = open("output.adl","w")
data = open("test.adl").read()
o.write( re.sub("C0:TIM-PACIFIC_STRING","C1:FEC-34_TIME_STRING",data) )
o.close()
Where 'output.adl' could be the same name as 'test.adl' if you want to
replace the existing file. Also FEC-34 just refers to which FE you're running.
It could, in principle, be any one of them.
The next step is to figure out how to apply this to all the files in a directory. |
3936
|
Tue Nov 16 23:36:29 2010 |
Suresh, Jenne | Update | SUS | Assembly of ETMs |
[Jenne, Suresh]
The ETM assembly has moved forward a couple of steps. We have completed the following:
1) Positioning the guide rod and wire stand-off on both the ETMs (5 and 7)
2) The magnets had to be cleaned with an acetone wash as they had touched the plastic Petri-dish (not cleaned for vacuum).
3) The magnets and the Al dumb-bells have been glued together and left to cure in the gluing fixture.
4) The guide-rod and wire stand-offs have also been glued to the optic and left to cure for 24 hrs.
JD: As you can see in my nifty status table, we are nearing the end of the suspension story.

We are going to try (but can't guarantee) to get ETMX to Bob for baking by Friday at lunchtime, that way we can re-suspend it on ~Monday, and place it in the chamber. Then we could potentially begin Green arm locking next week. Steve has (hopefully!!) ordered the spring plungers for ETMY. The receiving and baking of the spring plungers is the only current delay that I can foresee, and that only is relevant for one of the optics.
We (who is going to be in charge of this?) still need to move the SRM OSEMs & cables & connectors to the ITMY chamber from the BS chamber.
|
3937
|
Wed Nov 17 02:53:41 2010 |
yuta | Update | IOO | placed new PRM to BS table |
(Kiwamu, Yuta)
Background:
Yesterday, we aligned the Faraday and the beam reached SM2 at BS table.
Today, we placed a new PRM tower to BS table.
What we did:
1. Moved IPPO, IPPOSSM1, IPPOSSM3, IPANGSM1, IPANGSM2 out from the BS chamber.
2. Moved SRM tower(at PRM's place) to the ITMX chamber.
3. Placed the new PRM tower at the BS chamber.
4. Adjusted positions of the OSEMs for PRM and BS so that the sensor output can have roughly half of their maximum.
5. Checked damping servo for PRM and BS. They were working and helped us when adjusting OSEM positions.
6. Placed IPPO back and using SM2, made the beam hit PR2 at ITMX table.
7. Aligned the PRM so that the reflected beam path overlaps the incident beam.
We checked it by looking at MMT1.
For the alignment, we used IFO align sliders(C1:SUS-PRM_PIT_COMM, YAW_COMM).
To use them, we rebooted c1susaux.
Result:
1. The new PRM tower is placed.
2. OSEM sensor outputs for PRM and BS are;
(V) |
PRM |
BS |
max |
current value |
max |
current value |
ULSEN |
1.72 |
1.006 |
1.50 |
0.757 |
URSEN |
1.66 |
0.918 |
1.57 |
0.821 |
LRSEN |
1.92 |
1.304 |
1.57 |
0.821 |
LLSEN |
2.06 |
1.031 |
1.38 |
0.704 |
SDSEN |
9.21 |
4.366 |
1.57 |
0.821 |
We changed PRM aligning slider values, and they changed OSEM sensor outputs. We set the slider values to 0 when adjusting OSEM positions. |
3938
|
Wed Nov 17 10:39:20 2010 |
josephb | Update | CDS | Screen Time Fix |
An improved python code to apply a replacement to all *.adl files in a directory would be:
import re, os
files = os.listdir("./")
for file in files:
if ".adl" in str(file):
data = open(file).read()
o = open(file,"w")
o.write( re.sub("C0:TIM-PACIFIC_STRING","C1:FEC-34_TIME_STRING",data) )
o.close()
Of course, this entire python script can be replaced with a single sed command:
sed -i 's/C0:TIM-PACIFIC_STRING/C1:FEC-34_TIME_STRING/g' *
A more complicated script could be written which looks for key identifiers either in the file header or inside the file to determine which front end is appropriate, using a dictionary like:
dcuid_dict = {"BS":21,"PRM":37,"SRM":37,"ITMX":21,"ITMY":21,"MC1":36,"MC2":36,"MC3":36,"ETMX":24,"ETMY":26}
and then using for loops and if statements.
|
3939
|
Wed Nov 17 15:49:53 2010 |
rana | Update | DAQ | Ole Channel Names |
The following channels should be named as below to keep in line with their names pre-upgrade rather than use _DAQ in the name.
C1:SUS-{OPT}_{POS,PIT,YAW}
|
SUS{POS,PIT,YAW}_IN1
|
C1:SUS-{OPT}_OPLEV_{P,Y}ERROR
|
OL{PIT,YAW}_IN1
|
C1:SUS-{OPT}_SENSOR_{UL,UR,LL,LR,SIDE}
|
{UL,UR,LL,LR,SD}SEN_OUT
|
C1:SUS-{OPT}_OPLEV_{P,Y}OUT
|
OL{PIT,YAW}_OUT
|
C1:IOO-MC_TRANSPD
|
MC2_OLSUM_IN1
|
|
3940
|
Wed Nov 17 16:02:30 2010 |
josephb | Update | CDS | Modified feCodeGen.pl to fix filtMuxMatrix name generation |
Problem:
Sometime in the last 3 weeks, probably when Alex brought his latest changes from Hanford to the 40m and did an SVN update, the code which generates the names of the filter .adl files links for the overall matrix view broke.
Fix:
I modified FE code gen to use $basename instead of the base name after the top name transform (this changes _ to - after the first 3 letters
@@ -3520,11 +3522,11 @@
my $tn = top_name_transform($basename);
my $basename1 = $usite . ":" . $tn . "_";
- my $filtername1 = $usite . $tn;
+ my $filtername1 = $usite . $basename;
Still having problems:
The filter modules built with the matrix of filter modules run (offests/gains work), but will not load filter coefficients/filter names. All the other filter modules outside the matrix seem to load fine. At this point, doing a rebuild of any of the front end machines may cause the A2L filter banks to be unloadable.
|
3942
|
Wed Nov 17 23:45:20 2010 |
Jenne | Update | SUS | A bad day for suspensions |
[Jenne, Suresh]
Today has been a downright miserable day in the world of suspension work. Thumbs down to that: 
Yesterday, we had glued 2 full sets of magnets to dumbbells. Today, half of those broke. I think I put too thin of a layer of glue on the magnets when gluing them to the dumbbells. All magnet/dumbbell assemblies should pass the test of being picked up by the dumbbell while the magnet is stuck to the optical table or a razor blade. 6 of the 12 magnets failed this test. Suresh soaked the dumbbells that had been used in acetone, and scrubbed them, so we can reuse them when we reglue things tomorrow. By some miracle, we have exactly one full set intact (for each set of 6, we need 4 of one direction and 2 of the other). This was frustrating, but not yet a deal-breaker. That part comes next....
I got ETMU05 nicely aligned in the magnet gluing fixture, and then was on my last check of whether the side magnets would be glued in the correct place when I realized that the fixture is all wrong for the ETMs. This final check was added to the procedure after the drama with the ITMs of having the side magnets glued incorrectly as a result of the fixture being specific to the wedge angle of the optic. Kiwamu and I had set the fixture to be just right for the ~1deg wedge corner station optics, but the ETMs have a 2.35deg wedge (according to the Coastline spec sheet, which is consistent with our measurements when placing the guiderod and standoffs). Suresh and I need to reset the height of the optic in the fixture using more teflon sheets, but we don't have a whole lot of options ready in the cleanroom. We're going to cut some more pieces and ask Bob to clean them tomorrow. Since the way the fixture holds the teflon is a little hoaky, Suresh suggested just resting the optic on teflon pads, rather than screwing the teflon to the fixture, and then putting the optic on the pads. We'll try Suresh's method tomorrow, and hopefully it will be pretty easy.
At least the guiderods and standoffs were successfully glued to the optics....
Here's the updated Status Table. I don't think we're going to be able to have an ETM ready for the chambers early next week, but we should still be able to have both ready for the Monday after Thanksgiving. The spring plungers arrived today, and were given immediately to Bob and Daphen for cleaning.

|
3943
|
Thu Nov 18 00:40:31 2010 |
yuta | Update | IOO | PRM reflected beam reached AP table |
(Kiwamu, Yuta)
Summary:
Yesterday, we placed the new PRM to BS chamber and the beam reached PR2 at ITMX chamber.
Today, we lead the PRM reflected beam back to AP table.
Also, we aligned PRs so that the beam hits ITMX and ITMY.
What we did:
1. Aligned PR2 at ITMX chamber and PR3 at BS chamber so that the beam hits ITMY.
2. Aligned ITMX using IFO_ALIGN sliders so that the reflected beam overlaps at BS.
3. Aligned BS using IFO_ALIGN sliders so that the splitted beam to ITMX overlaps the green beam from the X-end.
4. Roughly aligned ITMY using IFO_ALIGN sliders so that the reflected green goes to far x-end.
5. From yesterdays in-vac work, the reflected beam from PRM reached the Faraday.
Aligned 2 steering mirrors in MC chamber so that the beam reaches AP table.
6. Found the beam is double-spotted by a steering mirror at just after the Faraday symmetric port.
The mirror is Y1-2037-45S. The beam is hitting it in ~10deg, so we have to replace it.
Plan:
- replace the steering mirror right next to the Faraday symmetric port.
- recyled Michealson
Note:
We had to use "ITMX" channels to align ITMY. We have to fix and check X-Y confusion.
Also, damping servo for ITMs does not seem to work. We have to check this. |
3944
|
Thu Nov 18 01:52:58 2010 |
Kevin | Update | Electronics | REFL55 Transfer Functions |
I measured the optical and electrical transfer functions for REFL55 and calculated the RF transimpedance. To measure the optical transfer function, I used the light from an AM laser to simultaneously measure the transfer functions of REFL55 and a New Focus 1611 photodiode. I combined these two transfer functions to get the RF transimpedance for REFL55. I also measured the electrical transfer function by putting the RF signal from the network analyzer in the test input of the photodiode.
I put all of the plots on the wiki at http://lhocds.ligo-wa.caltech.edu:8000/40m/Electronics/REFL55. |
3945
|
Thu Nov 18 11:06:20 2010 |
josephb | Update | CDS | c1sus and ADCs |
Problem:
ADCs are timing out on c1sus when we have more than 3.
Talked with Rolf:
Alex will be back tomorrow (he took yesterday and today off), so I talked with Rolf.
He said ordering shouldn't make a difference and he's not sure why would be having a problem. However, when he loads the chassis, he tends to put all the ADCs on the same PCI bus (the back plane apparently contains multiples). Slot 1 is its own bus, Slots 2-9 should be the same bus, and 10-17should be the same bus.
He also mentioned that when you use dmesg and see a line like "ADC TIMEOUT # ##### ######", the first number should be the ADC number, which is useful for determining which one is reporting back slow.
Plan:
Disconnect c1sus IO chassis completely, pull it out, pull out all cards, check connectors, and repopulate with Rolf's suggestions and keeping this elog in mind.
In regards to the RFM, it looks like one of the fibers had been disconnected from the c1sus chassis RFM card (its plugged in in the middle of the chassis so its hard to see) during all the plugging in and out of the cables and cards last night. |
3946
|
Thu Nov 18 14:05:06 2010 |
josephb, yuta | Update | CDS | c1sus is alive! |
Problem:
We broke c1sus by moving ADC cards around.
Solution:
We pulled all the cards out, examined all contacts (which looked fine), found 1 poorly connected cable internally, going between an ADC and ADC timing interface card (that probably happened last night), and one of the two RFM fiber cables pulled out of its RFM card.
We then placed all of the cards back in with a new ordering, tightened down everything, and triple checked all connections were on and well fit.
Gotcha!
Joe forgot that slot 1 and slot 2 of the timing interface boards have their last channels reserved for duotone signals. Thus, they shouldn't be used for any ADCs or DACs that need their last channel (such as MC3_LR sensor input). We saw a perfect timing signal come in through the MC3_LR sensor input, which prevented damping.
We moved the ADC timing interface card out of the 1st slot of the timing interface board and into slot 6 of the timing interface board, which resolved the problem.
Final Configuration:
Timing Interface Board
Timing Interface Slot |
1 (Duotone) |
2 (Duotone) |
3 |
4 |
5 |
6 |
7 |
8 |
9 |
10 |
11 |
12 |
13 |
Card |
None |
DAC interface (can't use last channel) |
ADC Interface |
ADC interface |
ADC interface |
ADC
interface
|
None |
None |
None |
DAC interface |
DAC interface |
None |
None |
PCIe Chassis
Slot |
0 |
1 |
2 |
3 |
4 |
5 |
6 |
7 |
8 |
9 |
10 |
11 |
12 |
13 |
14 |
15 |
16 |
17 |
PCIe Number |
Do Not Use |
1 |
6 |
5 |
4 |
9 |
8 |
7 |
3 |
2 |
14 |
13 |
12 |
17 |
16 |
15 |
11 |
10 |
Card |
None |
ADC |
DAC |
ADC |
ADC |
ADC |
BO |
BO |
BO |
BO |
DAC |
DAC |
BIO |
RFM |
None |
None |
None |
None |
Still having Issues with:
ITM West damps. ITM South damps, but the coil gains are opposite to the other optics in order to damp properly.
We also need to look into switching the channel names for the watchdogs on ITMX/Y in addition to the front end code changes. |
3947
|
Thu Nov 18 14:19:01 2010 |
josephb | Update | CDS | Swapped c1auxex and c1auxey codes |
Problem:
We had not switched the c1aux crates when we renamed the arms, thus the watchdogs labeled ETMX were really watching ETMY and vice-versa.
Solution:
I used telnet to connect to c1auxey, and then c1auxex.
I used the bootChange command to change the IP address of c1auxey to 192.168.113.59 (c1auxex's IP), and its startup script. Similarly c1auxex was changed to c1auxey and then both were rebooted.
c1auxey > bootChange
'.' = clear field; '-' = go to previous field; ^D = quit
boot device : ei
processor number : 0
host name : linux1
file name : /cvs/cds/vw/mv162-262-16M/vxWorks
inet on ethernet (e) : 192.168.113.60:ffffff00 192.168.113.59:ffffff00
inet on backplane (b):
host inet (h) : 192.168.113.20
gateway inet (g) :
user (u) : controls
ftp password (pw) (blank = use rsh):
flags (f) : 0x0
target name (tn) : c1auxey c1auxex
startup script (s) : /cvs/cds/caltech/target/c1auxey/startup.cmd /cvs/cds/caltech/target/c1auxex/startup.cmd
other (o) :
value = 0 = 0x0
c1auxex > bootChange
'.' = clear field; '-' = go to previous field; ^D = quit
boot device : ei
processor number : 0
host name : linux1
file name : /cvs/cds/vw/mv162-262-16M/vxWorks
inet on ethernet (e) : 192.168.113.59:ffffff00 192.168.113.60:ffffff00
inet on backplane (b):
host inet (h) : 192.168.113.20
gateway inet (g) :
user (u) : controls
ftp password (pw) (blank = use rsh):
flags (f) : 0x0
target name (tn) : c1auxex c1auxey
startup script (s) : /cvs/cds/caltech/target/c1auxex/startup.cmd /cvs/cds/caltech/target/c1auxey/startup.cmd
other (o) :
value = 0 = 0x0
|
3952
|
Fri Nov 19 03:43:33 2010 |
Kevin | Update | Electronics | REFL55 Characterizations |
[Koji, Rana, and Kevin]
I have been trying to measure the shot noise of REFL55 by shining a light bulb on the photodiode and measuring the noise with a spectrum analyzer. The measured dark noise of REFL55 is 35 nV/rtHz. I have been able to get 4 mA of DC current on the photodiode but have not been able to see any shot noise.
I previously measured the RF transimpedance of REFL55 by simultaneously measuring the transfer functions of REFL55 and a new focus 1611 photodiode with light from an AM laser. By combining these two transfer functions I calculated that the RF transimpedance at 55 MHz is ~ 200 ohms. With this transimpedance the shot noise at 4 mA is only ~ 7 nV/rtHz and would not be detectable above the dark noise.
The value of 200 ohms for the transimpedance seems low but it agrees with Alberto's previous measurements. By modeling the photodiode circuit as an RLC circuit at resonance with the approximate values of REFL55 (a photodiode capacitance of 100 pF and resistance of 10 ohms and an inductance of 40 nH), I calculated that the transimpedance should be ~ 230 ohms at 55 MHz. Doing the same analysis for the values of REFL11 shows that the transimpedance at 11 MHz should be ~ 2100 ohms. A more careful analysis should include the notch filters but this should be approximately correct at resonance and suggests that the 200 ohm measurement is correct for the current REFL55 circuit. |
3953
|
Fri Nov 19 04:23:12 2010 |
Suresh | Update | SUS | Cleaned and stuck magnets to dumbbells |
To clean the glue off the magnets and dumbbells I soaked them in Acetone for about an hour and then scrubbed the ends clean with a lint free tissue soaked in Acetone.
I then examined the ends under a microscope and found that while the flat faces were clean some of the grooves were still filled with glue.
While examining the magnets I found some small magnetic fibers stuck to the magnets. Rana had mentioned these before as potential trouble makers which could degrade the high frequency performance of the OSEMs.

To try and get the glue out of the grooves I put the dumbells through an ultrasonic bath for ten mins. Most of the glue has been removed from the grooves. Pics below
I proceeded try and recover the lost time by sticking the magnets back to the dumbbells. Increased the quantity of the glue to a slightly larger amount than usual. It should definitely squish out a bit now. We will know tomorrow when we open the gluing fixture.
|
3954
|
Fri Nov 19 12:53:50 2010 |
josephb | Update | CDS | Testpoints on c1iscex now working |
Problem:
c1iscex did not have test points working last night.
Solution:
The diag -i command indicated that :
awg 19 0 192.168.113.80 822095891 1 192.168.113.80
awg 45 0 192.168.113.80 822095917 1 192.168.113.80
The first number after the awg should be the DCUID number. The IP address 192.168.113.80 corresponds to c1iscex. So we had awg and testpoints setup for DCUI 19 and 45 on c1iscex. DCUID 19 is c1x01 (the IOP), but 45 was used for a test awhile back.
Turns out that in the testpoint.par file located in /cvs/cds/rtcds/caltech/c1/target/gds/param, there were two entries for c1scx, one with DCUID 24 and also DCUID 45. The model at the time was running with DCUID 24.
So I changed the model DCUID to 45, deleted the [C-node24] entry in the testpoint.par file, and restarted the machine, and also did a "telnet fb 8088" and "shutdown" to restart the frame builder. |
3955
|
Fri Nov 19 15:51:50 2010 |
Koji | Update | Electronics | REFL55 Characterizations |
RF Transimpedance of 200Ohm means the residual impedance at the resonance (R_res) of 40,
if you consider the amplifier gain (G_amp) of 10 and the voltage division by the 50Ohm termination,
this corresponds to the thermal noise level of Sqrt(4 kB T R_res)*G_amp/2 = 4nV/rtHz at the analyzer, while you observed 35nV/rtHz.
35nV/rtHz corresponds to 7nV/rtHz for the input noise of the preamp. That sounds too big if you consider the voltage noise of opamp MAX4107 that is 0.75nV/rtHz.
What is the measurement noise level of the RF analyzer?
Quote: |
[Koji, Rana, and Kevin]
I have been trying to measure the shot noise of REFL55 by shining a light bulb on the photodiode and measuring the noise with a spectrum analyzer. The measured dark noise of REFL55 is 35 nV/rtHz. I have been able to get 4 mA of DC current on the photodiode but have not been able to see any shot noise.
I previously measured the RF transimpedance of REFL55 by simultaneously measuring the transfer functions of REFL55 and a new focus 1611 photodiode with light from an AM laser. By combining these two transfer functions I calculated that the RF transimpedance at 55 MHz is ~ 200 ohms. With this transimpedance the shot noise at 4 mA is only ~ 7 nV/rtHz and would not be detectable above the dark noise.
The value of 200 ohms for the transimpedance seems low but it agrees with Alberto's previous measurements. By modeling the photodiode circuit as an RLC circuit at resonance with the approximate values of REFL55 (a photodiode capacitance of 100 pF and resistance of 10 ohms and an inductance of 40 nH), I calculated that the transimpedance should be ~ 230 ohms at 55 MHz. Doing the same analysis for the values of REFL11 shows that the transimpedance at 11 MHz should be ~ 2100 ohms. A more careful analysis should include the notch filters but this should be approximately correct at resonance and suggests that the 200 ohm measurement is correct for the current REFL55 circuit.
|
|
3956
|
Fri Nov 19 16:13:09 2010 |
Jenne | Update | SUS | ETMU05: magnets glued to optic |
[Jenne, Suresh]
Suresh and I glued the intact-from-the-first-round magnets to ETMU05. I accidentally got too much glue on one of the dumbbells (the glue was connecting the dumbbell to the gripper - bad news if we let that dry), and while I was cleaning it, the magnet broke off. So I used one of the ones that Suresh had re-glued last night, and he is putting that one back together after some cleaning.
To set the fixture, Suresh had the great idea of using small pieces of foil underneath the teflon pads to set the height of the optic in the fixture. The optic still rests on the teflon pads, but with the foil we have finer control over how the optic sits. Neat. Since both ETMs are the same, we shouldn't have to do any more adjustment for the other ETM.
The updated Status Table:

|
3957
|
Fri Nov 19 17:12:22 2010 |
yuta | Update | CDS | ETMX damped, but with weird TO_COIL matrix |
Background:
c1iscex machine is currently being setup and RT model c1scx is running.
But ETMX(south) didn't seem to be damped, so I checked it.
What I did:
1. Checked the wiring. It seemed to be OK.
Looked LEMO monitor output of SUS PD Whitening Board(D000210) with oscilloscope and they seemed to be getting some sensor signal except SDSEN.
SDSEN is funny. C1:SUS-ETMX_SPDMon decreases slowly when PD input cable is disconnected, and increases slowly when connected.
There might be some problem in the circuits.
Looked LEMO monitor output of SOS Coil Driver Module(D010001) with oscilloscope and they seemed to be receiving correct signal from DAC.
When ULCOIL offset is added, ch1 increased and so on.
2. Checked the direction of SUSDOF motion when kicked with one coil.
The result was;
kick (+) |
POS |
PIT |
YAW |
ULCOIL |
+ |
+ |
+ |
URCOIL |
+ |
- |
+ |
LRCOIL |
+ |
- |
- |
LLCOIL |
+ |
+ |
- |
This table tells you, when ULCOIL_OFFSET increases, SUSPOS increases and so on.
If URCOIL and LLCOIL are swapped, they look correct.
Also, they have opposite sign to the usual optics(e.g. MCs, BS, PRM).
3. Changed TO_COIL matrix according to the table above(see Attachment #1). Changed signs of XXCOIL_GAINs.
4. ETMX damped!
Plan:
- Check the wiring after SOS Coil Driver Module and circuit around SDSEN
- Check whitening and dewhitening filters. We connected a binary output cable, but didn't checked them yet.
- Make a script for step 2
- Activate new DAQ channels for ETMX (what is the current new fresh up-to-date latest fb restart procedure?)
|
Attachment 1: ETMXdamping.png
|
|