40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 103 of 341  Not logged in ELOG logo
ID Date Author Typeup Category Subject
  2483   Thu Jan 7 14:08:46 2010 JenneUpdateComputersWe haven't had a bootfest yet this week.....so today's the day

All the DAQ screens are bright red.  Thumbs down to that.

  2484   Thu Jan 7 14:55:36 2010 JenneUpdateComputersWe haven't had a bootfest yet this week.....so today's the day

Quote:

All the DAQ screens are bright red.  Thumbs down to that.

 All better now. 

  2486   Fri Jan 8 10:38:35 2010 josephb, kojiUpdateComputersRFM and Megatron

Last night, we installed the VMI 5565 RFM card into Megatron.  After turning off the watchdogs for the ETMY optic, we disconnected the RFM fiber, and connected it to megatron, then powered it up. 

We modified the RCG code to have 3 rfmio blocks, which were reading 0x11a1c0 (ascPit), 0x11a1c4 (ascYaw), and 0x11a1c8 (lscPos).  These were connected to the approriate filter module inputs, and we also added grounds to the front of the rfmio blocks (we looked at the ass code which was setup that way, so we just did the same thing).  When we started it however, it didn't read properly.  If we turned off the input and set and offset, it calculated the output of the filter module correctly, (i.e. just the offset value), but as soon as we turned on the input, it was set to 0, no matter the offset value, which indicated it was reading something correctly.

After this test, the RFM fibers were reconnected to c1iscey, we rebooted c1iscey, and we confirmed that the system was working properly again.  We turned the watch dogs back on for ETMY.

 

 

  2487   Fri Jan 8 11:43:22 2010 josephb, alexUpdateComputersRFM and Megatron

Alex came over with a short RFM cable this morning.  We used it to connect the rfm card in c1iscey to the rfm card megatron

Alex renamed startup.cmd in /cvs/cds/caltech/target/c1iscey/ to startup.cmd.sav, so it doesn't come up automatically.  At the end we moved it back.

Alex used the vxworks command d to look at memory locations on c1iscey.  Such as d 0xf0000000, which is the start of the rfm code location.  So to look at 0x11a1c8 (lscPos) in the rfm memory, he typed "d 0xf011a1c8".  After doing some poking around, we look at the raw tst front end code (in /home/controls/cds/advLigo/src/fe/tst), and realized it was trying to read doubles.  The old rts code uses floats, so the code was reading incorrectly.

As a quick fix, we changed the code to floats for that part.  They looked like:

etmy_lsc = filterModuleD(dsp_ptr,dspCoeff,ETMY_LSC,cdsPciModules.pci_rfm[0]? *(\
(double *)(((void *)cdsPciModules.pci_rfm[0]) + 0x11a1c8)) : 0.0,0);

And we simply changed the double to float in each case.  In addition we changed the RCG scripts locally as well (if we do a update at some point, it'll get overwritten).  The file we updated was /home/controls/cds/advLigo/src/epics/util/lib/RfmIO.pm

Line 57 and Line 84 were changed, with double replaced with float.

return "cdsPciModules.pci_rfm[0]? *((float *)(((void *)cdsPciModules.pci
_rfm[$card_num]) + $rfmAddressString)) : 0.0";

. "  *((float *)(((char *)cdsPciModules.pci_rfm[$card_nu
m]) + $rfmAddressString)) = $::fromExp[0];\n"

This fixed our ability to read the RFM card, which now can read the LSC POS channel, for example.

Unfortunately, when we were putting everything the way it was with RFM fibers and so forth, the c1iscey started to get garbage (all the RFM memory locations were reading ffff).  We eventually removed the VME board, removed the RFM card, looked at it, put the RFM card back in a different slot on the board, and returned c1iscey to the rack.  After this it started working properly.  Its possible in all the plugging and unplugging that the card somehow had become loose.

The next step is to add all the channels that need to be read into the .mdl file, as well as testing and adding the channel which need to be written.

 

  2488   Fri Jan 8 15:40:14 2010 josephb, alexUpdateComputersRFM and RCG

Alex added a new module to the RCG, for generating RFMIO using floats.  This has been commited to CVS.

  2491   Sat Jan 9 09:47:03 2010 AlbertoUpdateLSCProblems trying to lock the arms

This morning I've been having problems in trying to lock the X arm.

The X arm's filter FM6 in the LSC screen starts blinking as it was halfloaded. Then the transmitted power drops from 1 to ~0.5 and eventually the arm loses lock.
To me it looked like a computer related issue. So I decide to reboot C1ISCEX by powercycling it.

That doesn't seem to have solved the problem. The X arm can get locked but TRX slowly moves between 0.2 and 1.

  2492   Sat Jan 9 11:07:30 2010 AlbertoUpdateLSCProblems trying to lock the arms

Quote:

This morning I've been having problems in trying to lock the X arm.

The X arm's filter FM6 in the LSC screen starts blinking as it was halfloaded. Then the transmitted power drops from 1 to ~0.5 and eventually the arm loses lock.
To me it looked like a computer related issue. So I decide to reboot C1ISCEX by powercycling it.

That doesn't seem to have solved the problem. The X arm can get locked but TRX slowly moves between 0.2 and 1.

The X arm is now locked with TRX stable at ~1.

I think earlier on today I was having problems with running the alignment scripts from op540. Now I'm controlling the IFO from Rosalba and I can easily and stably lock all degrees of freedom.

I needed the X arm to be locked to align the auxiliary beam of the AbsL experiment to the IFO. To further stabilize TRX I increased the loop gain from 1 to 1.5.

Now the auxiliary beam is well aligned to the IFO and the beat is going through the PRC. I'm finally ready to scan the recycling cavity.

I also changed the gain of the PRC loop from -0.1 to -0.5.

  2493   Sat Jan 9 15:02:01 2010 AlbertoUpdateABSLPRC scanning
I scanned the PRC in the frequency range of 30-60 MHz, untill the PLL lost lock. But everything is working fine.
The PRC remained lock for all time, with SPOB at ~1000.
I'm leaving the lab now, planning to come back tomorrow.
I turned the flipping mirrors down and closed the mechanical shutter of the auxiliary NPRO.
  2494   Sun Jan 10 13:32:09 2010 HaixingUpdateSUStransfer function measurement of the quadrant maglev circuit

I have assembled the circuit and the control box for the quadrant magnetic levitation yesterday. The final setup is shown

in the figure below:

Quad_maglev_ctrl_box_in.JPGQuad_maglev_ctrl_box_front.JPG

 

Due to my carelessness, I I connected the wrong ends of the power supply. I damaged 4 op-amp and one voltage 

regulator during this assembly. This stupid mistake spent me several hours to fix, and I got a bitter lesson;-(

 

Afterwards, I replaced those op-amps and reconnected the power supply . Kiwamu helped me and we measured

the transfer function of this circuit.  The transfer function agrees with  the specification in the schematics which

has a integrator below 1 Hz and a differentiator from 5 to 20 Hz. The bode plot for the measured transfer function

is the following:

quad_mag_tf_amp.png

 quad_mag_tf_phs.png

Today I tested the photodetector parts and found that there is a mysterious oscillation. Whenever I connect the

photodector input A of the circuit (as indicated in the figure below),

PD.PNG

the output of the op-amp has a 500k Hz oscillation shown up in the oscilloscope.This happens even A is disconnected from

the photodetector and connected to an open end wire. I don't know how to eliminate it, and its amplitude is so large (peak to

peak is around 2.5 V) which completely dominates the photodetector output. Does anybody has some ideas? Thanks.

 

Quad_mag_lev_osc.JPG

  2495   Sun Jan 10 15:47:26 2010 KojiUpdateSUStransfer function measurement of the quadrant maglev circuit

1. Why do all of the BNCs have no GND connection? Each should have the individual cables to the ground. Each signal line and the corresponding ground line should be twisted together.

2. This looks the (usual) oscillation of the V-I conversion stage but I can't tell anything as I don't have the circuit diagram of the whole circuit.

3. In a certain case, putting some capacitance at the feedback may help. Read P.11 of the data sheet of LT1125. Try to put some capacitors from 20pF to some larger one whether it changes the situation or not. I suppose the bandwidth of your sensor can be ~1kHz. So putting a capacitance less than 10nF still has no effect to the servo.

  2497   Sun Jan 10 16:50:34 2010 HaixingUpdateSUStransfer function measurement of the quadrant maglev circuit

Quote:

1. Why do all of the BNCs have no GND connection? Each should have the individual cables to the ground. Each signal line and the corresponding ground line should be twisted together.

2. This looks the (usual) oscillation of the V-I conversion stage but I can't tell anything as I don't have the circuit diagram of the whole circuit.

3. In a certain case, putting some capacitance at the feedback may help. Read P.11 of the data sheet of LT1125. Try to put some capacitors from 20pF to some larger one whether it changes the situation or not. I suppose the bandwidth of your sensor can be ~1kHz. So putting a capacitance less than 10nF still has no effect to the servo.

 1. They are all connected to the box which has a single connection to the board ground. If I connect each of them to the ground, there would be many small loops

of ground. Of course, I should have replaced all the connectors such that the they are disconnected to the box as point out by Robert.

2. The oscillation disappears after I add 5 nF capacitor in parallel to the existing resistor. Thank you very much for pointing this out.

  2498   Sun Jan 10 17:15:25 2010 KojiUpdateSUStransfer function measurement of the quadrant maglev circuit

1. Yes. That is the bad. You should eventually replace the BNCs to the isolated ones.

2. OK. I like to emphasize again that everyone works on electronics should read data sheets more carefully and seriously because they have many important practical instructions to exploit full performance of the components. 

Quote:

Quote:

1. Why do all of the BNCs have no GND connection? Each should have the individual cables to the ground. Each signal line and the corresponding ground line should be twisted together.

2. This looks the (usual) oscillation of the V-I conversion stage but I can't tell anything as I don't have the circuit diagram of the whole circuit.

3. In a certain case, putting some capacitance at the feedback may help. Read P.11 of the data sheet of LT1125. Try to put some capacitors from 20pF to some larger one whether it changes the situation or not. I suppose the bandwidth of your sensor can be ~1kHz. So putting a capacitance less than 10nF still has no effect to the servo.

 1. They are all connected to the box which has a single connection to the board ground. If I connect each of them to the ground, there would be many small loops

of ground. Of course, I should have replaced all the connectors such that the they are disconnected to the box as point out by Robert.

2. The oscillation disappears after I add 5 nF capacitor in parallel to the existing resistor. Thank you very much for pointing this out.

 

  2500   Mon Jan 11 09:18:44 2010 AlbertoUpdateGeneralMeasurement running

I'm working on the AbsL experiment. A measurement which involved the PRC locked is running at the moment.

Please make sure of not interfering with the interferometer until it is done. Thank you.

  2502   Mon Jan 11 11:06:53 2010 AlbertoUpdateABSLMeasurement running

Quote:

I'm working on the AbsL experiment. A measurement which involved the PRC locked is running at the moment.

Please make sure of not interfering with the interferometer until it is done. Thank you.

 I'm done for the moement.

I realized that I need to take into account somehow the DC power from the photodiode. By now the measurement of the transmitted beat's power is affected by the total power circulating inside of the PRC and thus it depends on the cavity alignment.

I closed the laser shutter and turned down the flipping mirrors.

I'm planning to restart measuring by 2.30pm today. Till then the interferometer is available.

  2503   Mon Jan 11 14:16:59 2010 JenneUpdatePEMGur2 cables reconnected to the PEM ADCU

So that we can use both Guralps for Adaptive stuff, and so that I can look at the differential ground motion spectra, I've reconnected the Guralp Seismometers to the PEM ADCU, instead of where they've been sitting for a while connected to the ASS ADC.  I redid the ASS.mdl file, so that the PEM and PEMIIR matricies know where to look for the Gur2 data.  I followed the 'make ass' procedure in the wiki.  The spectra of the Gur1 and Gur2 seismometers look pretty much the same, so everything should be all good.

There's a problem with DataViewer though:  After selecting signals to plot, whenever I hit the "Start" button for the realtime plots, DataViewer closes abruptly. 

When I open dataviewer in terminal, I get the following output:

allegra:~>dataviewer
Warning: communication protocol revision mismatch: expected 11.3, received 11.4
Warning: Not all children have same parent in XtManageChildren
Warning: Not all children have same parent in XtManageChildren
Warning: Not all children have same parent in XtManageChildren
Warning: Not all children have same parent in XtManageChildren
Warning: Not all children have same parent in XtManageChildren
Warning: communication protocol revision mismatch: expected 11.3, received 11.4
msgget: No space left on device
allegra:~>framer4 msgget:msqid: No space left on device

Does anyone have any inspiration for why this is, or what the story is?  I have GR class, but I'll try to follow up later this afternoon.

  2504   Mon Jan 11 16:59:14 2010 AlbertoUpdateABSLInterferometer Busy

I'm currently running a measurement on the PRC.

Please don't interfere with the IFO until it is done. Talk with Alberto if you've been intending to work inside the lab.

Thank you.

  2505   Mon Jan 11 19:36:13 2010 AlbertoUpdateABSLMeasurement running

Leaving for dinner. Back in ~1hr.

I left a measurement running. Please don't interfere with it till I'm back. Thanks.

  2506   Mon Jan 11 21:49:17 2010 JenneUpdateABSLMeasurement running

Quote:

Leaving for dinner. Back in ~1hr.

I left a measurement running. Please don't interfere with it till I'm back. Thanks.

 Per Alberto's instructions, I have closed the shutter on his laser so that the Adaptive Team can play with the Mode Cleaner.

  2508   Tue Jan 12 09:37:05 2010 AlbertoUpdateABSLIFO available

I finished measuring the AbsL for this morning. The IFO is again available.

Please don't mess up with the interferometer though. I'll be back in a couple of ours

  2509   Tue Jan 12 11:34:26 2010 josephbUpdatePEMAllegra dataviewer

Quote:

So that we can use both Guralps for Adaptive stuff, and so that I can look at the differential ground motion spectra, I've reconnected the Guralp Seismometers to the PEM ADCU, instead of where they've been sitting for a while connected to the ASS ADC.  I redid the ASS.mdl file, so that the PEM and PEMIIR matricies know where to look for the Gur2 data.  I followed the 'make ass' procedure in the wiki.  The spectra of the Gur1 and Gur2 seismometers look pretty much the same, so everything should be all good.

There's a problem with DataViewer though:  After selecting signals to plot, whenever I hit the "Start" button for the realtime plots, DataViewer closes abruptly. 

When I open dataviewer in terminal, I get the following output:

allegra:~>dataviewer
Warning: communication protocol revision mismatch: expected 11.3, received 11.4
Warning: Not all children have same parent in XtManageChildren
Warning: Not all children have same parent in XtManageChildren
Warning: Not all children have same parent in XtManageChildren
Warning: Not all children have same parent in XtManageChildren
Warning: Not all children have same parent in XtManageChildren
Warning: communication protocol revision mismatch: expected 11.3, received 11.4
msgget: No space left on device
allegra:~>framer4 msgget:msqid: No space left on device

Does anyone have any inspiration for why this is, or what the story is?  I have GR class, but I'll try to follow up later this afternoon.

 This problem seems to be restricted to allegra.  Dataviewer works fine on Rosalba  and op440m, as well as using ssh -X into rosalba to run dataviewer remotely.  DTT seems to work fine on allegra.  The disk usage seems no where near full on allegra.  Without knowing which "device" its refering to (although it must be a device local to allegra), I'm not sure what to look at now. 

I'm going to do a reboot of allegra and see if that helps.

Update:  The reboot seems to have fixed the issue.

 

 

 

 

  2510   Tue Jan 12 13:24:50 2010 HaixingUpdateSUSQuadrant Magnetic Levitation

I have tried to make the quadrant magnetic levitation work but unfortunately I did not succeed. During the experiment, I have made

the following changes to the circuit and setup:

 

(1) I added small resistors (6K) in parallel to R11, R23, R35 and R47 (indicated in the following schematics)  to increase

the control bandwidth from 20 Hz  to 80 Hz.

 

 

schmematic.png

 

(2) I changed RLED1, RLED2, RLED3, RLED4 from 2.2K to 1.5K  in the LED driver such that the current of the

LED increases and in turn the displacement sensitivity increases.

 

(3) I changed R51 and R51 (in the differencing block for PD1 and PD2) from 5K to 1 K. Correspondingly,

I increased R52 and R54 from 5K to 50K. This changes increase the gain in the differential control by a

factor of 50, which compensates the gain loss after increasing the control bandwidth.

 

 (4) The trim pots in the coil drive block have the following values: 100K for pot1 and pot4, 1K fro pot 2 and pot3.

To increase the gain, I replaced R17, R30, R31, R41 by 102 Om resistors (we run out of 100 Om chip resistors.)

 

(5) I wrapped the OSEM flags by plastic tubes to block the light from the LED more efficiently. This also removed

the changes of the photocurrent in the photodetector when the levitated plate moves horizontally.

 

Possible issues that cause the setup not working at the moment:

 

(1) The feedback gain could be probably still not enough. During the experiment, I can't feel any force changes when the

flags crossing the zero point. The error signals and control signal has the right sign.

 

(2) The levitated weight may be not enough and the working point is below the extremum of the DC attracting force.

This could give rise to a large negative spring which requires unreasonable feedback gain to compensate.

 

(3) The DC attracting force between the magnets are differing each other too much (mismatch) and can't

be compensated by the coil driving force.

 

(4) The control bandwidth may be still not  large enough. Initially my design was 100 Hz, but I forgot to divide

the angular frequency by 2pi and the resulting circuit has a bandwidth of 20 Hz. Later I increase it up to 80 Hz

by changing the resistors as mentioned before.

 

(5) The polarization of the coil could have a wrong sign. I have checked with Gauss meter, but still the absence

of zero-point crossing force change makes me worry about this.

 

For continuation of this work, I will finish writing my document and summarize all the results and outline what

needs to be done in the future. If everything goes well, I will be back in June and can spend more time on this

experiment. I can also work with the summer students together on this project.

  2512   Wed Jan 13 12:01:06 2010 AlbertoUpdateComputersc1dcuepics, c1lsc rebooted this morning
Since last night the alignemtn scripts couldn't work.
c1lsc wasn't working properly because attempts to lock the X arm would try to control ETMY and attempts of locking the Y arm, wouldn't actuate any optics.
Also, another sign of a malfunctioning c1lsc was that one of the LSC filter modules, FM6, couldn't get loaded properly. It looked like only half loaded on the LSC MEDM screen.
On the other hand, plotting the trend of the last month, c1lsc's CPU didn't look more loaded that usual.
 
Rebooting and restarting C1lsc wasn't enough and I also had to reboot c1dcuepics a couple of times beforse getting things back to work.
  2513   Wed Jan 13 12:03:00 2010 AlbertoUpdateABSLMeasurement now running. Please be careful

At the moment I'm running a measurement on the PRC and I'm planning to leave it going for the time we'll be at the 40m meeting.

Please be careful in the lab. Thank you.

  2515   Fri Jan 15 11:21:05 2010 josephbUpdateComputersMegatron and tst model for ETMY

The tst model wasn't compiling this morning due to some incorrect lines in the RfmIOfloat.pm file located in /home/controls/cds/advLIgo/src/epics/util/lib file on megatron. 

The error was "Undefined subroutine &CDS::RfmIOfloat::partType called at lib/Parser2.pm line 354, <IN> line 3363."

I changed RfmIO to RfmIOfloat on lines 1 and 6.
Basically the first 6 lines are now

package CDS::RfmIOfloat;
use Exporter;
@ISA = ('Exporter');

sub partType {
        return RfmIOfloat;
}

The tst now compiles.  At the moment, I believe we should be able to switch megatron in for ETMY and attempt to lock the arm.  The whitening/dewhitening filters are still not automatically synced in software and hardware, but I don't think that should prevent locking.

  2516   Fri Jan 15 12:04:26 2010 Sanjit, mevansUpdateAdaptive FilteringCanceling noise again!

 

OAF is successfully canceling noise again, thanks to Matt!

Here is a plot showing more than a factor of 10 noise reduction around 3Hz (similar to what we saw in the simulations)

The changes that has made it work are:

  • use of RANGER channel (with ACC_MC1_X and/or ACC_MC2_X)
  • mu = 0.01, tau = 1.0e-6, ntaps = 2000, nDown = 16
  • nDelay = 5 and nDelay = 7 both work (may not be so sensitive on delay at low frequencies)
  • Main changes: filter bank on the PEM channels (ASS_TOP_PEM_## filters: 0.1:0, 1:, Notch24, AA32)
  • Added the AI800 filter for upsampling in MC1 (should not matter)

 Matt suggested playing with the emphasis (EMPH) filters to cancel noise in different frequency bands.

 

Attachment 1: OAF_15JAN2010.png
OAF_15JAN2010.png
  2517   Fri Jan 15 18:53:05 2010 josephb,peter,kojiUpdateComputersAttempted locking with Megatron replacing ETMY front end

This afternoon we attempted to lock the interferometer using Megatron insteady of the usual ETMY front end.

We attempted it once, found the alignment seemed particularly bad, then realized we had forgotten to add the QPDs.  In the process of adding the QPDs to the tst.mdl file, we realized I (Joe) should have been looking at the iscNetDsc.h file, not the iscNetDs40m.h file for the RFM memory structure.  The only major difference was the memory location of where the qpd information gets passed.

We added QPDs to the tst.mdl file, writing to the RFM network with the QPD pitch, yaw and sum values.

We also added normalization to the oplev, by dividing the OL pitch and yaw values by the OL sum value in the tst.mdl file.

The lscPos, ascPit, ascYaw were working properly, although we did have to poke at the ascPit and ascYaw values once before they started reading properly at Megatron (they started at -1e20).  We think the RFM card may have started in an odd state, and without something causing the ascPit and ascYaw values to change, it never updated.

At the end of the day, the interferometer was locked for a few seconds.  There is are still some issues to be worked on, but its progress.

Koji returned everything back to normal operations after the test.

  2519   Sun Jan 17 19:18:55 2010 AlbertoUpdateComputersBoot fest to restart the computer and c1iscey not responding.

Thi afternoon I found that the RFM network in trouble. The frontends sync counters had railed to 16384 counts and some of the computers were not responding. I went for a bootfest, but before I rebooted c1dcu epics. I did it twice. Eventually it worked and I could get the frontends back to green.

Although trying to burtrestore to snapshots taken at any time after last wednesday till today would make the RFM crash again. Weird.

Also, c1iscey seems in a coma and doesn't want to come back. Power cycling it didn't work. I don't know how to be more persuasive with it.

  2521   Mon Jan 18 18:34:01 2010 AlbertoUpdateABSLMeasurement in progress

I started a long measurement of the PRC's transmissivity. I'm leaving the lab and I'm going to be back at about 8 tonight. Please do not disturb the interferometer. it is important that the MC and the PRC stay locked all the time.

  2522   Mon Jan 18 20:58:40 2010 AlbertoUpdateABSLMeasurement in progress

Quote:

I started a long measurement of the PRC's transmissivity. I'm leaving the lab and I'm going to be back at about 8 tonight. Please do not disturb the interferometer. it is important that the MC and the PRC stay locked all the time.

 That measurement is finished. I'm now going to start another one that will take another hour or so. I'm leaving it running for the night. If you want to work on the IFO, it should be definitely done by 11pm.

  2523   Mon Jan 18 23:44:19 2010 kiwamuUpdateElectronicstriple resonant circuit for EOM

The first design of the triple resonant EOM circuit has been done.

If only EOM has a loss of 4 Ohm, the gain of the circuit is expected to be 11 at 55MHz

So far I've worked on investigation of the single resonant circuit and accumulated the knowledge about resonant circuits.

Then I started the next step, designing the triple resonant circuit.

Here I report the first design of the circuit and the expected gain.

 


( What I did )

At first in order to determine the parameters, such as inductors and capacitors, I have solved some equations with numerical ways (see the past entry).

In the calculation I put 6 boundary conditions as followers;

(first peak=11MHz, second peak=29.5MHz, third peak=55MHz, first valley=22MHz, second valley=33MHz, Cp=18pF)

The valley frequencies of 22MHz and 33MHz are chosen in order to eliminate the higher harmonics of 11MHz, and Cp of 18pF represents the capacitance of the EOM.

Basically the number of parameters to be determined is 6 ( L1, L2, ...,), therefore it is completely solved under 6 boundary conditions. And in this case, only one solution exists.

The point is calculation does not include losses because the loss does not change the resonant frequency.

 

whole_circuit.png

( results )

As a result I obtained the 6 parameters for each components shown in the table below.

Cp [pF] 18.1
C1 [pF]  45.5
C2 [pF] 10.0
Lp [uH] 2.33
L1 [uH] 1.15
L2 [uH] 2.33

Then I inserted the loss into the EOM to see how the impedance looks like. The loss is 4 Ohm and inserted in series to the EOM. This number is based on the past measurement .

Let us recall that the gain of the impedance-matched circuit with a transformer is proportional to square-root of the peak impedance.

It is represented by G = sqrt(Zres/50) where Zres is the impedance at the resonance.

 As you can see in the figure, Zres = 6.4 kOhm at 55MHz, therefore the gain will be G=11 at 55MHz.

Essentially this gain is the same as that of the single resonant circuit that I've been worked with, because its performance was also limited mainly by the EOM loss.

 An interesting thing is that all three peaks are exactly on the EOM limited line (black dash line), which is represented by Zres = L/CR = 1/ (2pi f Cp)^2 R. Where R = 4 Ohm.

 designed_circuit.png

( next plan )

There are other solutions to create the same peaks and valleys because of the similar solution.

 It is easy to understand if you put Cp' = Cp x N, the solutions must be scaled like L1'=L1/N, C1'=C1 x N, ...,  Finally such scaling gives the same resonant frequencies.

So the next plan is to study the effect of losses in each components while changing the similar solution.

After the study of the loss I will select an optimum similar solution.

  2524   Tue Jan 19 00:10:44 2010 ranaUpdateElectronicstriple resonant circuit for EOM

Very cool.         

  2525   Tue Jan 19 02:39:57 2010 kiwamuUpdateElectronicsdesign complete --- triple resonant circuit for EOM ---

The design of the triple resonant circuit has been fixed.

I found the optimum configuration, whose gain is still 11 at 55MHz even if there are realistic losses.

As I mentioned in the last entry, there are infinite number of the similar solutions to create the same resonant frequencies.

However owing to the effect of the losses, the resultant gain varies if the similar solution changes

The aim of this study is to select the optimum solution which has a maximum gain ( = the highest impedance at the resonance ).

In order to handle the losses in the calculation, I modeled the loss for both inductors and the capacitors.

Then I put them into the circuit, and calculated the impedance while changing the solutions.

 


 

(method)

1). put the scaling parameter as k in order to create the similar solution.

2). scale the all electrical parameters (L1, L2,...) by using k, so that C1'=C1 x k, L1'=L1/k ,...

3). Insert the losses into all the electrical components

4). Draw the impedance curve in frequency domain.

5). See how the height of the impedance at the resonance change

6). Repeat many time this procedure with another k.

7). Find and select the optimum k

scaling.png

There is a trick in the calculation.

I put a capacitor named Cpp in parallel to the EOM in order to scale the capacitance of the EOM (see the schematic).

For example if we choose k=2, this means all the capacitor has to be 2-times larger.

For the EOM, we have to put Cpp with the same capacitance as Cp (EOM). As a result these two capacitors can be dealt together as 2 x Cp.

So that Cpp should be Cpp = (k-1) Cp, and Cpp vanishes when we choose k=1.

 

The important point is that the scaling parameter k must be greater than unity, that is k > 1.

This restriction directly comes from Cp, the capacitance of the EOM, because we can not go to less than Cp.

If you want to put k < 1, it means you have to reduce the capacitance of the EOM somehow (like cutting the EO crystal ??)

 

(loss model)

I've modeled the loss for both the inductors and the capacitors in order to calculate the realistic impedance.

The model is based on the past measurements I've performed and the data sheet.

   Loss for Capacitor :  R(C) = 0.5 (C / 10pF)^{-0.3} Ohm

   Loss for Inductor :    R(L) = 0.1 ( L / 1uH) Ohm

Of course this seems to be dirty and rough treatment.

But I think it's enough to express the tendency that the loss  increase / decrease monotonically as  L / C get increased.

These losses are inserted in series to every electrical components.

( Note that: this model depends on both the company and the product model. Here I assume use of Coilcraft inductors and mica capacitors scattered around 40m )

 

( results )

The optimum configuration is found when k=1, there is no scaling. This is the same configuration listed in last entry

Therefore we don't need to insert the parallel capacitor Cpp in order to achieve the optimum gain.

The figure below shows the some examples of the calculated impedance. You can see the peak height decrease by increasing the scale factor k.

realistic.png

The black dash line represents the EOM-loss limit, which only contains the loss of the EOM.

The impedance at the resonance of 55MHz is 6.2 kOhm, which decreased by 3% from the EOM-loss limit. This corresponds to gain of G = 11.

The other two peaks, 11MHz and 29.5MHz dramatically get decreased from EOM-loss limit.

I guess this is because the structure below 50MHz is mainly composed by L1, L2, C1, C2.

In fact these components have big inductance and small capacitance, so that it makes lossy.

 

( next step )

The next step is to choose the appropriate transformer and to solder the circuit.

  2526   Tue Jan 19 02:40:38 2010 KojiUpdateElectronicstriple resonant circuit for EOM

The design looks very good. I have some questions.

1. As far as I remember, you've got the gain of slightly worse than 10 for a 55MHz single resonant case. Why your expectation of the gain (G=11) for the highest resonance better than this?

Supposing the loss exists only on the EOM, the other part of the LC network for the triple work as an inductor at the resonant frequency. This is just equivalent as the single resonant case. So the expected gain at 55MHz should coincides with what we already have. Probably, the resistance of 4 Ohm that is used here had too rough precision???

2. How can you adjust the resonances precisely?

Do we need any variable components for Cs and Ls, that may have worse quality than the fixed one, generally speaking.
I myself has no experience that I had to tune the commercial EOM because of a drift or whatever. I hope if you can adjust the resonance with a fixed component it should be fine.

3. Changing Cp. What does it mean?

Do you put additional cap for Cp?

4. The resonances for the lower two look very narrow. Is that fine?

This will show up in a better shape if we look at the transfer function for the gain. Is this right?

If we have BW>100kHz, it is sufficient.

5. Impedance matching for the lower two resonances.

Yep. You know this problem already.

 

  2527   Tue Jan 19 03:04:14 2010 KojiUpdateElectronicstriple resonant circuit for EOM

Self-follow:

I got the answer of Q3 from the follow-up entry.

For Q4, once you get the impedance of the LC network lower than n^2*50, the EOM gain will be quite low. This means that the resonance is anyway narrow.
I did some simple calculation and it shows that the width of the resonance will be 100kHz~500kHz. So it maybe OK.

Quote:

The design looks very good. I have some questions.

1. As far as I remember, you've got the gain of slightly worse than 10 for a 55MHz single resonant case. Why your expectation of the gain (G=11) for the highest resonance better than this?

Supposing the loss exists only on the EOM, the other part of the LC network for the triple work as an inductor at the resonant frequency. This is just equivalent as the single resonant case. So the expected gain at 55MHz should coincides with what we already have. Probably, the resistance of 4 Ohm that is used here had too rough precision???

2. How can you adjust the resonances precisely?

Do we need any variable components for Cs and Ls, that may have worse quality than the fixed one, generally speaking.
I myself has no experience that I had to tune the commercial EOM because of a drift or whatever. I hope if you can adjust the resonance with a fixed component it should be fine.

3. Changing Cp. What does it mean?

Do you put additional cap for Cp?

4. The resonances for the lower two look very narrow. Is that fine?

This will show up in a better shape if we look at the transfer function for the gain. Is this right?

If we have BW>100kHz, it is sufficient.

5. Impedance matching for the lower two resonances.

Yep. You know this problem already. 

 

  2528   Tue Jan 19 03:20:28 2010 KojiUpdateElectronicsdesign complete --- triple resonant circuit for EOM ---

First I was confused, but now I think I understood.

My confusion:
If the k get bigger, L get smaller, C get bigger. This makes R(L) smaller and R(C) smaller. This sounds very nice. But why smaller k is preferable in the Kiwamu's result?

Explanation:
The resultant impedance of the network at a resonance is determined by Zres = L/(R C) or something like that. Here R = R(L)+R(C). (I hope this is right.)

Here larger Zres is preferable. So smaller R is nice.

But If the speed of reduction for R is slower than that of L/C (which is proportional to k^-2), increasing k does not help us to increase of Zres. And that's the case.

This means "if we can put the LC network in the box of EOM, we can do better job!" as we can reduce Cp.

Quote:

scaling.png


   Loss for Capacitor :  R(C) = 0.5 (C / 10pF)^{-0.3} Ohm

   Loss for Inductor :    R(L) = 0.1 ( L / 1uH) Ohm

  2529   Tue Jan 19 03:27:47 2010 kiwamuUpdateElectronicsRe: triple resonant circuit for EOM

1. You are right, the gain for the single resonant circuit was about 9.3 in my measurement.

But the reason why the triple is better than the single resonant circuit comes from the transformer.

The impedance can be degraded by a loss of the transformer, because it got worse after applying the transformer in the past measurement.

Also I definitely confirmed that the circuit had the impedance of 7.2 kOhm at the resonance of 52.9MHz without the transformer.

So it shall give the gain of 12, but did not after applying the transformer.

 

2.  Yes, I think we need some variable components just in case.

 

5.  For the impedance matching, I will select a transformer so that 55MHz is matched. In contrast I will leave two lower resonances as they are.

This is because 11MHz and 29.5MHz usually tend to have higher impedance than 55MHz. In this case, even if the impedance is mismatched, the gain for these can be kept higher than 11.

I will post the detail for this mismatched case tomorrow.

 

Quote:

The design looks very good. I have some questions.

1. As far as I remember, you've got the gain of slightly worse than 10 for a 55MHz single resonant case. Why your expectation of the gain (G=11) for the highest resonance better than this?

Supposing the loss exists only on the EOM, the other part of the LC network for the triple work as an inductor at the resonant frequency. This is just equivalent as the single resonant case. So the expected gain at 55MHz should coincides with what we already have. Probably, the resistance of 4 Ohm that is used here had too rough precision???

2. How can you adjust the resonances precisely?

Do we need any variable components for Cs and Ls, that may have worse quality than the fixed one, generally speaking.
I myself has no experience that I had to tune the commercial EOM because of a drift or whatever. I hope if you can adjust the resonance with a fixed component it should be fine.

3. Changing Cp. What does it mean?

Do you put additional cap for Cp?

4. The resonances for the lower two look very narrow. Is that fine?

This will show up in a better shape if we look at the transfer function for the gain. Is this right?

If we have BW>100kHz, it is sufficient.

5. Impedance matching for the lower two resonances.

Yep. You know this problem already.

 

 

  2530   Tue Jan 19 10:30:29 2010 josephbUpdateComputersBoot fest to restart the computer and c1iscey not responding.

Quote:

Thi afternoon I found that the RFM network in trouble. The frontends sync counters had railed to 16384 counts and some of the computers were not responding. I went for a bootfest, but before I rebooted c1dcu epics. I did it twice. Eventually it worked and I could get the frontends back to green.

Although trying to burtrestore to snapshots taken at any time after last wednesday till today would make the RFM crash again. Weird.

Also, c1iscey seems in a coma and doesn't want to come back. Power cycling it didn't work. I don't know how to be more persuasive with it.

During the testing of Megatron as the controller for ETMY, c1iscey had been disconnected from the ethernet hub.  Apparently we forgot to reconnect it after the test.  This prevented it from mounting the nfs directory from linux1, and thus prevented it from coming up after being shutdown.  It has been reconnected, restarted, and is working properly now.

  2531   Tue Jan 19 12:54:39 2010 AlbertoUpdateABSLMeasurement in progress

A measurement will be running for the next hour. Please be careful.

  2532   Tue Jan 19 16:21:18 2010 AlbertoUpdateABSLWatchdogs not working and then fixed

This afternoon the watchdogs stopped working: they didn't trip when the suspension positions crossed the threshold values.

I rebooted c1susaux (aka c1dscl1epics0 in the 1Y5 rack), which is the computer that runs the watchdog processes.

The reboot fixed the problem.

  2533   Tue Jan 19 23:26:07 2010 kiwamuUpdateElectronicsRe:Re: triple resonant circuit for EOM

Quote:

5.  For the impedance matching, I will select a transformer so that 55MHz is matched. In contrast I will leave two lower resonances as they are.

This is because 11MHz and 29.5MHz usually tend to have higher impedance than 55MHz. In this case, even if the impedance is mismatched, the gain for these can be kept higher than 11.

I will post the detail for this mismatched case tomorrow.

 

Here the technique of the impedance matching for the triple resonant circuit are explained.

In our case, the transformer should be chosen so that the impedance of the resonance at 55MHz is matched.

We are going to use the transformer to step up the voltage applied onto the EOM.

To obtain the maximum step-up-gain, it is better to think about the behavior of the transformer.

When using the transformer there are two different cases practically. And each case requires different optimization technique. This is the key point.

By considering these two cases, we can finally select the most appropriate transformer and obtain the maximum gain.

 

 


( how to maximize the gain ?)

Let us consider about the transformer. The gain of the circuit by using the transformer is represented by

eq1.png         (1)

Where ZL is the impedance of the load (i.e. impedance of the circuit without the transformer ) and n is the turn ratio.

It is apparent that G is the function of two parameters, ZL and n.  This leads to two different solutions for maximizing the gain practically. 

 

matching_edit.png

 

  - case.1 : The turn ratio n is fixed.

In this case, the tunable parameter is the impedance ZL.  The gain as a function of ZL is shown in the left figure above.

In order to maximize the gain, Z must be as high as possible.  The gain G get close to 2n when the impedance ZL goes to infinity.

There also is another important thing; If the impedance ZL is bigger than the matched impedance (i.e. ZL = 50 * n^2 ), the gain can get higher than n.

 

  - case.2 : The impedance ZL is fixed.

In contrast to case1, once the impedance ZL is fixed, the tunable parameter is n. The gain as a function of n is shown in the right figure above.

In this case the impedance matched condition is the best solution, where ZL=50*n^2. ( indicated as red arrow in the figure )

The gain can not go higher than n somehow. This is clearly different from case1.
 

 

( Application to the triple resonant circuit )

Here we can define the goal as "all three resonances have gain of more than n, while n is set to be as high as possible"

According to consideration of case1, if each resonance has an impedance of greater than 50*n^2 (matched condition) it looks fine, but not enough in fact.

For example if we choose n=2, it corresponds to the matched impedance of 50*n^2 = 200 Ohm. Typically every three resonance has several kOhm which is clearly bigger than the matched impedance successfully.

However no matter how big impedance we try to make,  the gains can not be greater than G=2n=4 for all the three resonance. This is ridiculous.

What we have to do is to choose n so that it matches the impedance of the resonance which has the smallest impedance.

In our case, usually the resonance at 55MHz tends to have the smallest impedance in those three. According to this if we choose n correctly, the other two is mismatched.

However they can still have the gain of more than n, because their impedance is bigger than the matching impedance. This can be easily understand by recalling the case1.

 

(expected optimum gain of designed circuit)

 By using the equation (1), the expected gain of the triple resonant circuit including the losses is calculated. The parameters can be found in last entry.

designed_response.png

The turn ratio is set as n=11, which matches the impedance of the resonance at 55MHz. Therefore 55MHz has the gain of 11.

The gain at 11MHz is bigger than n=11, this corresponds to the case1. Thus the impedance at 11MHz can go close to gain of 22, if we can make the impedance much big.

 

  2534   Wed Jan 20 20:11:56 2010 AlbertoUpdateABSLSome preliminary results from measuring PRC's transmissivity for an amplitude modulated beam
Here I'm posting a plot showing a set of measurements that I made in the past few days to determine the absolute length of the PRC cavity.
As in my other AbsL measurements, I inject an auxiliary laser beam into the cavity and look at the transmission. In the PRC case, the beam is injected through the dark port and I look at a pick-off of the REFL beam. The aux laser is phase locked to the PSL beam and I control the differential frequency between the two. The PSL and the aux beam interfere and beat at their differential frequency.
 
The attached plot shows the transmitted power as a function of the beat frequency.
 
Fitting the data with the model would let me determine the cavity length. 
By now I can estimate the length of the PRC at about 2.257m, but it's still a rather approximate value.
I can't provide accurate error bars yet. I need to optimize the measurements to get a more precise value.
 
I will go more through the details of the measurement technique and of the fitting function as soon as I have more definitive results.
 
The data points shown here were taken at different times and not always in optimal alignment condition of the PRC. 
To get a good fit of the data I should have fewer frequency segments, taken in a shorter period of time, and in which the power circulating inside of the cavity (ie SPOB) fluctuates as little as possible.
 
For what regards the time needed for a measurements, I already significantly sped up the measurements (i.e. optimizing the scanning and acquisition GPIB scripts, and fixing a couple of problems with the PDH box used in the PLL), and finally now I can scan several tens of MHz in few minutes.
About the frequency segments, so far they have been determined by two factors
1) Tthe frequency generator in the PLL: the Marconi works as a continuous wave generator only in limited ranges. Switching from one to another brakes the wave in a way that causes the PLL to lose lock.
2) Getting below 18 MHz a series of other beats appear on the PLL photodiode and make the PLL lose lock.
 
For the first problem, I'm thinking of using two Marconis and to mix their signals. I would keep one at 300MHz and I would scan the other from 300MHz to 500MHz. In fat, in that frequency range the Marconi has not discontinuity.
 
To try to avoid the other beats at low frequency, I'm not entirely sure about what to do yet. 
 
To be continued...
Attachment 1: 2010-01-19_PRCtransmissivityVsModel.png
2010-01-19_PRCtransmissivityVsModel.png
  2536   Thu Jan 21 10:31:13 2010 KojiUpdateABSLSome preliminary results from measuring PRC's transmissivity for an amplitude modulated beam

Nice and interesting plot.

I suppose slight decrease of the Schnupp asymmetry (in your model) adjusts the discrepancy in the high freq region.
At the same time, it will make the resonance narrower. So you need to put some loss at the recombination (=on the BS)?

...of course these depends on the flatness of the calibration.

  2538   Thu Jan 21 11:08:30 2010 kiwamu, steveUpdateVACDry Pump replacement

 This morning I and Steve replaced  the dry fore pump of TP3, which is located under the y-arm.

After replacing it we confirmed vacuum normal condition. The fore line pressure of TP3 went down to 11 mTorr from 750 mTorr

  Attached picture is new pump after setup.

Attachment 1: DSCN0428.JPG
DSCN0428.JPG
  2541   Fri Jan 22 02:54:06 2010 AlbertoUpdateABSLOvernight measurement

I'm leaving a measurement running overnight. It should be done in about one hour.

Tomorrow morning, If you need to use the interferometer, and you don't want to have the auxiliary beam going onto the dark port, you can turn down the flipping mirror and close the NPRO's mechanical shutter.

  2542   Fri Jan 22 12:33:37 2010 josephb, alexUpdateComputersModified CDS_PARTS for Binary output

Turns out the CDSO32 part (representing the Contec BO-32L-PE binary output) rquires two inputs. One for the first 16 bits, and one for the second set of 16 bits.  So Alex added another input to the part in the library.  Its still a bit strange, as it seems the In1 represents the second set of 16 bits, and the In2 represents the first set of 16 bits.

I added two sliders on the CustomAdls/C1TST_ETMY.adl control screen (upper left), along with a bit readout display, which shows the bitwise and of the two slider channels. For the moment, I still can't see any output voltage on any of the DO pins, no matter what output I set.

 

  2543   Fri Jan 22 14:40:49 2010 AlbertoUpdateABSLOvernight measurement

Quote:

I'm leaving a measurement running overnight. It should be done in about one hour.

Tomorrow morning, If you need to use the interferometer, and you don't want to have the auxiliary beam going onto the dark port, you can turn down the flipping mirror and close the NPRO's mechanical shutter.

 This is what I measured last night:

2010-01-21_PRCtransmissivityVsModel.png

 This is not a fit. It's just a comparison of the model with the data.

  2544   Mon Jan 25 11:42:24 2010 josephbUpdateComputersMegatron and BO board

I was talking with Vladimir on Friday discussing the Binary Output board, we looked at the manual for it (Contec DO-32L-PE) and we realized its an opto-coupler isolated open-collector output.  He mentioned they had the right kind of 50 channel breakout board for testing in Riccardo's lab.

This morning I borrowed the 50 channel breakout board from Riccardo's lab, and along with some resistor loads, test the BO board.  It seems to be working, and I can control the output channel on/off state.

  2545   Mon Jan 25 16:30:37 2010 AlbertoUpdateABSL166 MHz sideband turned off

I turned off the modulation at 166MHZ becasue I don't need it if I'm only locking the PRC.

It was introducing extra amplitude modulations of the beam which interfered with the AbsL's PLL photodiode.

I'm going to turn it back on later on.

  2546   Mon Jan 25 16:46:33 2010 AlbertoUpdateABSL166 MHz sideband turned off

Quote:

I turned off the modulation at 166MHZ becasue I don't need it if I'm only locking the PRC.

It was introducing extra amplitude modulations of the beam which interfered with the AbsL's PLL photodiode.

I'm going to turn it back on later on.

 I turned back on the 166MHz modulation just a bit. I set the slider at 4.156.

When it was totally off the MZ seemd quite unhappy.

  2547   Tue Jan 26 03:28:56 2010 ranaUpdateABSL166 MHz sideband turned off

 

 You can turn the 166 off if you want. MZ is unhappy after its turned off, but that's just the thermal transient from removing the RF heat. After a several minutes, the heat goes away and the MZ can be relocked.

One of these days we should evaluate the beam distortion we get in EOMs because of the RF heat induced dn/dT. Beam steering, beam size, etc.

ELOG V3.1.3-