40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 73 of 341  Not logged in ELOG logo
ID Date Author Type Category Subjectup
  16958   Tue Jun 28 18:19:09 2022 TomislavUpdateElectronicsElectronics noise

I measured electronics noise of WFSs and QPD (of the WFS/QPD, whitening, ADC...) by closing PSL and measuring the error signal. It was needed to put the offset in C1:IOO-MC_TRANS_SUMFILT_OFFSET to 14000 cts (without offset the sum of quadrants would give zero, and 14000 cts is the value when the cavity is locked). For WFS that are RF, if there is intensity noise at low frequencies, it is not affecting the measurement.

In the attachment please find the power spectrum of the error signal when the PSL shutter is on and off.

Attachment 1: electronics_noise_spectra.png
electronics_noise_spectra.png
Attachment 2: error_signal.png
error_signal.png
  16963   Wed Jun 29 18:53:38 2022 ranaUpdateElectronicsElectronics noise

this is just the CDS error signal, but is not the electronics noise. You have to go into the lab and measure the noise at several points. It can't be done from the control room. You must measure before and afte the whitening.

Quote:

I measured electronics noise of WFSs and QPD (of the WFS/QPD, whitening, ADC...) by closing PSL and measuring the error signal. It was needed to put the offset in C1:IOO-MC_TRANS_SUMFILT_OFFSET to 14000 cts (without offset the sum of quadrants would give zero, and 14000 cts is the value when the cavity is locked). For WFS that are RF, if there is intensity noise at low frequencies, it is not affecting the measurement.

In the attachment please find the power spectrum of the error signal when the PSL shutter is on and off.

 

  16992   Tue Jul 12 14:56:17 2022 TomislavSummaryElectronicsElectronics noise measurements

[Paco, Tomislav]

We measured the electronics noise of the demodulation board, whitening board, and ADC for WFSs, and OPLEV board and ADC for DC QPD in MC2 transmission. We were using SR785.

Regarding the demodulation board, we did 2 series of measurements. For the first series of measurements, we were blocking WFS (attachment 1) and measuring noise at the output of the demod board (attachment 2a). This measurement includes dark noise of the WFS, electronics noise of demod board, and phase noise from LO. For the second series of the measurements, we were unplugging input to the demod board (attachment 2b & 2c is how they looked like before unplugging) (the mistake we made here is not putting 50-ohm terminator) and again measuring at the output of the demod board. This measurement doesn't include the dark noise of the WFS. We were measuring it for all 8 segments (I1, I2, I3, I4, Q1, Q2, Q3, Q4). The dark noise contribution is negligible with respect to demod board noise. In attachments 3 & 4 please find plots that include detection and demodulation contributions for both WFSs.

For whitening board electronics noise measurement, we were terminating the inputs (attachment 5) and measuring the outputs (attachment 6). Electronics noise of the whitening board is in the attachments 7 & 8.

For ADC electronics noise we terminated ADC input and measured noise using diaggui (attachments 9 & 10). Please find these spectra for WFS1, WFS2, and MC TRANS in attachments 11, 12 & 13.

For MC2 TRANS we measured OPLEV board noise. We did two sets of measurements, as for demod board of WFSs (with and without QPD dark noise) (attachments 14, 15 & 16). In the case of OPLEV board noise without dark noise, we were terminating the OPLEV input. Please find the electronics noise of OPLEV's segment 1 (including dark noise which is again much smaller with respect to the OPLEV's electronics noise) in attachment 17.

For the transfer functions, demod board has flat tf, whitening board tf please find in attachment 18, ADC tf is flat and it is (2**16 - 1)/20 [cts/V], and dewhitening tf please find in attachment 19. Also please find the ASD of the spectral analyzer noise (attachment_20).

Measurements for WFS1 demod and whitening were done on 5th of July between 15h and 18h local time. Measurements for WFS2 demod and whitening were done on 6th of July between 15h and 17h local time. All the rest were done on July 7th between 14h and 19h. In attachment 21 also find the comparison between electronics noise for WFSs and cds error signal (taken on the 28th of June between 17h and 18h). Sorry for bad quality of some pictures.

Attachment 1: attachment_1.jpg
attachment_1.jpg
Attachment 2: attachment_2a.jpg
attachment_2a.jpg
Attachment 3: attachment_2b.jpg
attachment_2b.jpg
Attachment 4: attachment_2c.jpg
attachment_2c.jpg
Attachment 5: attachment_3.png
attachment_3.png
Attachment 6: attachment_4.png
attachment_4.png
Attachment 7: attachment_5.jpg
attachment_5.jpg
Attachment 8: attachment_6.jpg
attachment_6.jpg
Attachment 9: attachment_7.png
attachment_7.png
Attachment 10: attachment_8.png
attachment_8.png
Attachment 11: attachment_9.jpg
attachment_9.jpg
Attachment 12: attachment_10.jpg
attachment_10.jpg
Attachment 13: attachment_11.png
attachment_11.png
Attachment 14: attachment_12.png
attachment_12.png
Attachment 15: attachment_13.png
attachment_13.png
Attachment 16: attachment_14.jpg
attachment_14.jpg
Attachment 17: attachment_15.jpg
attachment_15.jpg
Attachment 18: attachment_16.jpg
attachment_16.jpg
Attachment 19: attachment_17.png
attachment_17.png
Attachment 20: attachment_18.png
attachment_18.png
Attachment 21: attachment_19.png
attachment_19.png
Attachment 22: attachment_20.png
attachment_20.png
Attachment 23: attachment_21.png
attachment_21.png
  16998   Wed Jul 13 13:26:44 2022 ranaSummaryElectronicsElectronics noise measurements

as I said to you yesterday, I don't think image 2a shows the output of the demod board. The output of the demod board is actually the output connector ON the demod board. What you are showing in 2a, is the signal that goes from the whitening board to the ADC I believe. I may be msitaken, so please check with Tega for the signal chain.

  4584   Thu Apr 28 22:38:38 2011 AidanUpdateGreen LockingElectronics schematic for vertex beatbox

 With some assistance from Kiwamu and Koji, I've drawn up the electronics design for the Beat Box for the vertex green locking. The Omingraffle schematic is posted on the Green Locking Wiki page. It's also attached below. Some final touches are necessary before we can Altium this up.

 Attachment 1: Schematic of beatbox

Attachment 2: Front and back panel designs.

Attachment 1: Canvas_1.png
Canvas_1.png
Attachment 2: Canvas_2.png
Canvas_2.png
  4593   Sat Apr 30 05:14:33 2011 KojiUpdateGreen LockingElectronics schematic for vertex beatbox

- AC coupling for the comparator circuit of the green locking

In order to relieve the power consumption of the RF buffer, ac coupling circuits have been added.

The ac coupling before the buffer amp helps to relieve the power consumption in the chip.
But because of the distortion of the signal (and the limitation of the bandwidth), the output still has some DC (~0.6V).
Therefore, the output is also AC coupled.

Note that the BW pin of BUF634P should be directly connected to -15V in order to keep the bandwidth of the buffer.

The drawings are also uploaded on the green electronics wiki

Attachment 1: freq_divider.png
freq_divider.png
  1036   Wed Oct 8 22:23:43 2008 YoichiConfigurationElectronicsElectronics work bench cleanup
Yesterday, I cleaned up the electronics work bench. I figured that keeping the work bench
in order has larger effect on the work efficiency than buying expensive soldering stations.
Whoever works there should clean up the table after the work to the state shown on
the right side of the picture (at least).
If you leave the bench for a while (more than 30min) but intend to return later and
resume the work, please write your name and time on a piece of paper and put it on the bench.
Otherwise, your stuff can be taken away anytime.
Attachment 1: Cleanup.jpg
Cleanup.jpg
  14405   Fri Jan 18 15:34:37 2019 gautamUpdateThermal CompensationElliptical reflector part number

Nobody documented this, but here is the part number with mechanical drawings of the elliptical reflector installed at EY: Optiforms E180. Heater is from Induceramics, but I can't find the exact part which matches the dimensions of the heater we have (diameter = 3.8mm, length = 30mm), perhaps it's a custom part?

The geometry dictates that if we want the heater to be at one focus and the ETM to be at the other, the separation has to be 7.1 inches. It certainly wasn't arranged this way before. It seems unrealistic to do this without clipping the main beam, I propose we leave sufficient clearane between the main beam and the reflector, and accept the reduced heating efficiency. 

Thanks to Steve for digging this up from his secret stash.

  3452   Fri Aug 20 20:36:05 2010 AlastairUpdateGeneralElog

 Elog was down, I restarted it.

  5739   Tue Oct 25 21:23:17 2011 DmassBureaucracyelogElog Restarted

Elog went nonreponsive. SSH'ed into nodus to run restart script. Elog came back ~15 minutes later.

  10323   Fri Aug 1 15:32:07 2014 ericqUpdateComputer Scripts / ProgramsElog and svn backups

Koji and Evan have both brought up a good point that we may not be backing up the svn and ELOG properly.

I have modified the rsync.backup script that nodus' cron runs every night that backs up /cvs/cds to what I presume are the tape backups at ldas-cit.ligo.caltech.edu.

Specifically, I added two rsync commands that grab the svn and elog directories from /export/home and copy them to their old locations in /cvs/cds/caltech. This way, the old locations are updated, and the tape backups stay current.

  2302   Thu Nov 19 16:04:48 2009 AlbertoConfigurationelogElog debugging output - Down time programmed today to make changes

We want the elog process to run in verbose mode so that we can see what's going. The idea is to track the events that trigger the elog crashes.

Following an entry on the Elog Help Forum, I added this line to the elog starting script start-elog-nodus:

./elogd -p 8080 -c /cvs/cds/caltech/elog/elog-2.7.5/elogd.cfg -D -v > elogd.log 2>&1

which replaces the old one without the part with the -v argument.

The -v argument should make the verbose output to be written into a file called elogd.log in the same directory as the elog's on Nodus.

I haven't restarted the elog yet because someone might be using it. I'm planning to do it later on today.

So be aware that:

We'll be restarting the elog today at 6.00pm PT. During this time the elog might not be accessible for a few minutes.

  2303   Thu Nov 19 18:49:55 2009 AlbertoConfigurationelogElog debugging output - Down time programmed today to make changes

Quote:

We want the elog process to run in verbose mode so that we can see what's going. The idea is to track the events that trigger the elog crashes.

Following an entry on the Elog Help Forum, I added this line to the elog starting script start-elog-nodus:

./elogd -p 8080 -c /cvs/cds/caltech/elog/elog-2.7.5/elogd.cfg -D -v > elogd.log 2>&1

which replaces the old one without the part with the -v argument.

The -v argument should make the verbose output to be written into a file called elogd.log in the same directory as the elog's on Nodus.

I haven't restarted the elog yet because someone might be using it. I'm planning to do it later on today.

So be aware that:

We'll be restarting the elog today at 6.00pm PT. During this time the elog might not be accessible for a few minutes.

 I tried applying the changes but they didn't work. It seems that nodus doesn't like the command syntax.

I have to go through the problem...

The elog is up again.

  3441   Thu Aug 19 09:52:51 2010 AlastairUpdateComputersElog down

 I restarted it using start-elog-nodus and this worked out fine - even though I did it from Pete's on my phone ;-)

  3464   Tue Aug 24 14:29:18 2010 josephbUpdateelogElog down for 1 minute

I'm going to take the elog down for one minute and restart it under gdb (using a copy of gdb stolen from fb40m since I couldn't figure out how to install an old enough version on nodus from source).  The terminal with information is running on Rosalba under the "Phase Noise" panel, so please don't close it.  Ideally, the next time the elog crashes, I'll have some output indicating why or at least the line in the code.  I can then look at the raw source code or send the line back to the developer and see if he has any ideas.

  5556   Tue Sep 27 11:43:59 2011 JenneUpdateelogElog has been dying a lot lately...

WTF?

  2187   Fri Nov 6 00:23:34 2009 AlbertoConfigurationComputersElog just rebooted

The elog just crashed and I rebooted it

  1181   Fri Dec 5 20:40:38 2008 YoichiHowToComputersElog multi-keyword search
The current Elog search allows you to look for only one keyword in the text.
You cannot search for two keywords by simply separating them with a white space.
That is, a search term "abc def" matches a literal "abc def", not a text containing "abc" and "def".
This is extremely annoying. However, there are still some ways to search for multiple keywords.
The Elog search fields are treated as regular expressions.
In order to match a text containing "abc" and "def", you can use a search term "abc.*def".
A period (.) means "any character", and an asterisk (*) means "any number of repetition of the preceding character".
Therefore, ".*" matches "any number of any character" i.e. anything.
The search term "abc.*def" works fine when you know "abc" appears first in the text you are looking for.
If you don't know the order of appearance of the keywords, you have two choices: either to use,
"(abc.*def)|(def.*abc)"
or
"(abc|def).*(abc|def)"
The vertical bar (|) means "or". Parentheses are used for grouping.
The first example does exactly what you want. However, you have to list all the permutations of your keywords
separated by |. If you have more than two keywords, it can be a very very long search word.
(The length of the search word is O(n!), where n is the number of keywords).
In the second example, the length of the keyword is O(n). However, it can also match a text containing two "abc".
This means the search result may contain some garbages (entries containing only "abc").
I guess in most cases we can tolerate this.

To automatically construct a multiple keyword search term for the Elog, I wrote a bash script called elogkeywd
and it is installed in the control room machines.
You can type
elogkeywd keyword1 keyword2 keyword3
to generate a regular expression for searching a text containing "keyword1", "keyword2" and "keyword3".
The generated expression is of the second type shown above. You can then copy-and-paste the result to
the Elog search field.
The script takes any number of keywords. However, there seems to be a limit on the number of characters you can type
into the search field of the Elog. I found the practical limit is about 3 keywords.
  3795   Wed Oct 27 11:52:45 2010 josephbUpdateelogElog needed to be restarted

I had to restart the elog on Nodus because it was no longer responding.

  2739   Wed Mar 31 10:34:02 2010 josephbUpdateelogElog not responding this morning

When I went to use the elog this morning, it wasn't responding.  I killed the process on nodus, and then restarted, per the 40m wiki instructions.

  2196   Fri Nov 6 18:02:22 2009 josephbUpdateComputersElog restarted

While I was writing up an elog entry, the elog died again, and I restarted it.  Not sure what caused it to die since no one was uploading to it at the time.

  11058   Mon Feb 23 15:04:10 2015 ZachUpdateGeneralElog restarted

The elog crashed while I was creating an entry to the Cryo log. I restarted it with the start-elog.csh script.

  3994   Tue Nov 30 12:10:44 2010 josephbUpdateelogElog restarted again

The elog seemed to be down at around 12:05pm.  I waited a few minutes to see if the browser would connect, but it did not.

I used the existing script in /cvs/cds/caltech/elog/ (as opposed to Zach's new on in elog/elog-2.8.0/) which also seems to have worked fine.

  4168   Wed Jan 19 10:31:24 2011 josephbUpdateelogElog restarted again

Elog wasn't responding at around 10 am this morning.  I killed the elogd process, then used the restart script.

  2562   Tue Feb 2 18:15:47 2010 AlbertoUpdateelogElog restarted it

 Zach made me notice that the elog had crashed earlier on this afternoon. 

I just restarted it with the restarting script.

Instructions on how to run the last one are now in the wiki page. Look on the "How To" section, under "How to restart the elog".

  3461   Tue Aug 24 06:34:01 2010 AlastairUpdateGeneralElog...

...was down.  Have restarted it.

  2473   Mon Jan 4 17:21:30 2010 JenneConfigurationIOOElusive Mode Matching Solution found!

I think I have finally found a Mode Matching solution for our new Input Mode Matching Telescope!  And after looking at the layout diagram with Koji and Raffaele, it seems like all of the optics will fit into the chambers / onto the tables (not true as of last week). 

3. RoCMMT1 is -5m
   RoCMMT2 is 8m,
   with the MMTs 1.89m apart.
   This is a 1.6x telescope.
   MMT2 is 2.2641m from the PRM
   MMT1 is 2m from MC3.
   The Condition Number for this optical chain is 89219047.5781.

This layout is very similar to the one that Koji posted on the wiki yesterday:  Upgrade09/Optical Layout.  The difference is that I want to move MMT1 ~20cm closer to the MC13 table, so just on the other side of the main red beam that goes directly to PRM.  There is plenty of space there, so it should be all good.  The tricky bit is that the flat steering mirrors fit into things now while they are piezos, but they will be trickier to fit if we make them into Tip Tilts.  But I have full faith in Koji's amazing optical table layout skills, that he can make it happen. 

Unless there are major objections, I think this is the MMT that we're going to go with. (So speak now or forever hold your peace.)  The angle between tilt and translation isn't quite what we'd like it to be (at ~18deg), but it's not too terrible.  And we still have 99.5% overlap which is very important.

Attachment 1: Awesome_MM_Solution.png
Awesome_MM_Solution.png
  2481   Wed Jan 6 03:44:41 2010 KojiConfigurationIOOElusive Mode Matching Solution found!

I am in the way to get a reasonable optical layout.
Please calculate the final results with the following conditions.

"Result" =
- mode overlapping with astigmatism
- alignment matrix (m/rad, rad/rad) for Pitch and Yaw
- alignment orthogonality
- sensitivity of the mode overlapping to the perturbations
  * histgram
  * individual scan of the optic positions

Optics chain: MC3 - SM1(flat) - MMT1(f=-5m) - MMT2(f=+8m) - SM2(flat) - PRM

Incident angles: SM1 24deg, MMT1 3deg, MMT2 1deg, SM2 44.5deg

Distances:
MC3 HR - SM1: 884mm
SM1 - MMT1: 1058.2mm
MMT1 - MMT2: 1890mm
MMT2 - SM2: 2007.9mm
SM2 - PRM HR: 495.6mm

It has ~200mm deviation from the solution. I can move only MMT1 for final optimization.
Give us the numbers if it can improve the performance.
Note that this move changes SM1-MMT1 and MMT1-MMT2 simultaneously.

Quote:

I think I have finally found a Mode Matching solution for our new Input Mode Matching Telescope!  And after looking at the layout diagram with Koji and Raffaele, it seems like all of the optics will fit into the chambers / onto the tables (not true as of last week). 

3. RoCMMT1 is -5m
   RoCMMT2 is 8m,
   with the MMTs 1.89m apart.
   This is a 1.6x telescope.
   MMT2 is 2.2641m from the PRM
   MMT1 is 2m from MC3.
   The Condition Number for this optical chain is 89219047.5781.

This layout is very similar to the one that Koji posted on the wiki yesterday:  Upgrade09/Optical Layout.  The difference is that I want to move MMT1 ~20cm closer to the MC13 table, so just on the other side of the main red beam that goes directly to PRM.  There is plenty of space there, so it should be all good.  The tricky bit is that the flat steering mirrors fit into things now while they are piezos, but they will be trickier to fit if we make them into Tip Tilts.  But I have full faith in Koji's amazing optical table layout skills, that he can make it happen. 

Unless there are major objections, I think this is the MMT that we're going to go with. (So speak now or forever hold your peace.)  The angle between tilt and translation isn't quite what we'd like it to be (at ~18deg), but it's not too terrible.  And we still have 99.5% overlap which is very important.

 

  7207   Thu Aug 16 18:59:18 2012 JenneUpdateSAFETYEmergency Exit door unlocked again

Koji just found the emergency exit door unlocked again.  NOT GOOD.

We have determined that if you use the emergency door to enter the lab, it leaves the door unlocked, unless you go back outside and deliberately lock it.  This means that someone has been using the emergency exit as a regular entrance. 

It's fine to leave by that door, but you should make a habit of entering through the regular door.  Using the back door as an entrance is a special case situation, when they have the main door blocked.

  15485   Wed Jul 15 19:23:44 2020 gautamUpdateGeneralEmergency light on in control room

The emergency lamps above the exit sign on the NW entrance to the control room are on. I tried opening and closing the door, but it remains on. Probably nothing to worry about, but noting here anyway.

  15486   Wed Jul 15 19:51:51 2020 KojiUpdateGeneralEmergency light on in control room

It happened before too. Doesn't it say it has occasional self-testing or something?

  15487   Wed Jul 15 20:58:40 2020 gautamUpdateGeneralEmergency light on in control room

True - it is now not on anymore.

Quote:

It happened before too. Doesn't it say it has occasional self-testing or something?

  16047   Mon Apr 19 09:17:51 2021 JordanUpdateVACEmpty N2 Tanks

When I came into the lab this morning, I noticed that both N2 tanks were empty. I had swapped one on Friday (4-16-21) before I left the lab. Looking at the logs, the right tank (T2) sprung a leak shortly shortly after install. I leak checked the tank coupling after install but did not see a leak. There could a leak further down the line, possibly at the pressure transducer.

The left tank (T1) emptied normally over the weekend, and I quickly swapped the left tank for a full one, and is curently at ~2700 psi. It was my understanding that if both tanks emptied, V1 would close automatically and a mailer would be sent out to the 40m group. I did not receive an email over the weekend, and I checked the Vac status just now and V1 was still open.

I will keep an eye on the tank pressure throughout the day, and will try to leak check the T2 line this afternoon, but someone should check the vacuum interlocks and verify.

 

Attachment 1: N2_Pressure.PNG
N2_Pressure.PNG
  16064   Wed Apr 21 12:56:00 2021 JordanUpdateVACEmpty N2 Tanks

Installed T2 today, and leaked checked the entire line. No issues found. It could have been a bad valve on the tank itself. Monitored T2 pressure for ~2 hours to see if there was any change. All seems ok.

Quote:

When I came into the lab this morning, I noticed that both N2 tanks were empty. I had swapped one on Friday (4-16-21) before I left the lab. Looking at the logs, the right tank (T2) sprung a leak shortly shortly after install. I leak checked the tank coupling after install but did not see a leak. There could a leak further down the line, possibly at the pressure transducer.

The left tank (T1) emptied normally over the weekend, and I quickly swapped the left tank for a full one, and is curently at ~2700 psi. It was my understanding that if both tanks emptied, V1 would close automatically and a mailer would be sent out to the 40m group. I did not receive an email over the weekend, and I checked the Vac status just now and V1 was still open.

I will keep an eye on the tank pressure throughout the day, and will try to leak check the T2 line this afternoon, but someone should check the vacuum interlocks and verify.

 

 

Attachment 1: Screenshot_2021-04-21_12-53-26.png
Screenshot_2021-04-21_12-53-26.png
  16090   Wed Apr 28 11:31:40 2021 JonUpdateVACEmpty N2 Tanks

I checked out what happened on c1vac. There are actually two independent monitoring codes running:

  1. The interlock service, which monitors the line directly connected to the valves.
  2. A seaparate convenience mailer, running as a cronjob, that monitors the tanks.

The interlocks did not trip because the low-pressure delivery line, downstream of the dual-tank regulator, never fell below the minimum pressure to operate the valves (65 PSI). This would have eventually occurred, had Jordan been slower to replace the tanks. So I see no problem with the interlocks.

On the other hand, the N2 mailer should have sent an email at 2021-04-18 15:00, which was the first time C1:Vac-N2T1_pressure dropped below the 600 PSI threshold. N2check.log shows these pressures were recorded at this time, but does not log that an email was sent. Why did this fail? Not sure, but I found two problems which I did fix:

  • One was that the code was checking the sensor on the low-pressure side (C1:Vac-N2_pressure; nominally 75 PSI) against the same 600 PSI threshold as the tanks. This channel should either be removed or a separate threshold (65 PSI) defined just for it. I just removed it from the list because monitoring of this channel is redundant with the interlock service. This does not explain the failure to send an email.
  • The second issue was that the pyN2check.sh script appeared to be calling Python 3 to run a Python 2 script. At least that was the situation when I tested it, and this was causing it to fail partway through. This might well explain the problem with no email. I explicitly set python --> python2 in the shell script.

The code then ran fine for me when I retested it. I don't see any further issues.

Quote:

Installed T2 today, and leaked checked the entire line. No issues found. It could have been a bad valve on the tank itself. Monitored T2 pressure for ~2 hours to see if there was any change. All seems ok.

Quote:

When I came into the lab this morning, I noticed that both N2 tanks were empty. I had swapped one on Friday (4-16-21) before I left the lab. Looking at the logs, the right tank (T2) sprung a leak shortly shortly after install. I leak checked the tank coupling after install but did not see a leak. There could a leak further down the line, possibly at the pressure transducer.

The left tank (T1) emptied normally over the weekend, and I quickly swapped the left tank for a full one, and is curently at ~2700 psi. It was my understanding that if both tanks emptied, V1 would close automatically and a mailer would be sent out to the 40m group. I did not receive an email over the weekend, and I checked the Vac status just now and V1 was still open.

I will keep an eye on the tank pressure throughout the day, and will try to leak check the T2 line this afternoon, but someone should check the vacuum interlocks and verify.

 

  16305   Wed Sep 1 14:16:21 2021 JordanUpdateVACEmpty N2 Tanks

The right N2 tank had a bad/loose valve and did not fully open. This morning the left tank was just about empty and the right tank showed 2000+ psi on the gauge. Once the changeover happened the copper line emptied but the valve to the N2 tank was not fully opened. I noticed the gauges were both reading zero at ~1pm just before the meeting. I swapped the left tank, but not in time. The vacuum interlocks tripped at 1:04 pm today when the N2 pressure to the vacuum valves fell below 65psi. After the meeting, Chub tightened the valve, fully opened it and refilled the lines. I will monitor the tank pressures today and make sure all is ok.

There used to be a mailer that was sent out when the sum pressure of the two tanks fell <600 psi, telling you to swap tanks. Does this no longer exist?

  16344   Mon Sep 20 14:11:40 2021 KojiUpdateBHDEnd DAC Adapter Unit D2100647

I've uploaded the schematic and PCB PDF for End DAC Adapter Unit D2100647.

Please review the design.

  • CH1-8 SUS actuation channels.
    • 5CHs out of 8CHs are going to be used, but for future extensions, all the 8CHs are going to be filled.
    • It involves diff-SE conversion / dewhitening / SE-diff conversion. Does this make sense?
  • CH9-12 PZT actuation channels. It is designed to send out 4x SE channels for compatibility. The channels have the jumpers to convert it to pass through the diff signals.
  • CH13-16 are general purpose DIFF/SE channels. CH13 is going to be used for ALS Laser Slow control. The other 3CHs are spares.

The internal assembly drawing & BOM are still coming.

Attachment 1: D2100647_End_DAC_Adapter.pdf
D2100647_End_DAC_Adapter.pdf D2100647_End_DAC_Adapter.pdf D2100647_End_DAC_Adapter.pdf D2100647_End_DAC_Adapter.pdf D2100647_End_DAC_Adapter.pdf D2100647_End_DAC_Adapter.pdf D2100647_End_DAC_Adapter.pdf D2100647_End_DAC_Adapter.pdf
  11335   Fri May 29 02:05:08 2015 ericqUpdateLSCEnd Laser temperatures set

Both green beatnotes have been found with nominal amplitudes. (X: -30dBm Y: -20dBm), at temperatures which don't seem to be prone to mode hopping. 

X = 42.64, Y ~40.15

Both arms can lock on ALS, but as Koji mentioned in ELOG 11334, the ALS noise seems anomalously high. 


Details

The temperatures I posted in the previous log ended up not being so useful. To find the right end laser temperatures, I looked at the IR beatnotes on the control room analyzer out to 1GHz, and swept around the SLOW_SERVO2_OFFSET channels to change the laser temperature. During this time, the end green shutters were closed, the PSL shutter was closed, the FSS slow servo was off, and the FSS_SLOWDC was set to 0.0. (The green PSL shutter had to be open, because the IR beat fiber is coupled after it.)

For each arm, I found three temperatures where an IR beat could be observed; as Koji mentioned on Wednesday, we should use the middle mode. For each of the arms, scanning around from the middle to move the beat by +-1GHz did not cause a mode hop - the beat stayed visible on the scope. Once I found a real IR beat for the X arm, I took at look at the RF output of the X BBPD, and found my alignment from the other night was actually pretty good; I made a minor touch up to maximize the green beat. 

For the AUX X innolight, I was able to find the actual temperature of the laser crystal when at the correct point, remove the digital offset, and return to the same temperature with the controller front panel dial. This temperature is 42.64 degrees Celcius. There is no diode temperature control on the unit, as far as I could tell, but the maximum green transmitted power through the X arm is about the same as it's ever been, around GTRX=0.5. 

My motivation for doing this was that I always have problems remembering the historical typical values for the digital temperature offset. It seems much cleaner to me to set things up such that a beat is visible "at the origin" (i.e. FSS_SLOWDC=0 and SLOW_SERVO2_OFFSET=0) I suppose that this will also depend on what mode of the PMC we're locked to. During my time working today, its been locked near the middle of its range, currently sitting at 145V. 

However, for the AUX Y lightwave, I was a little perplexed to find that moving the the laser crystal setpoint around does not apear to change the real laser temperature at all. Thus I could not offload the digital offset in the same way. Aditionally, the lightwave controller does not have the same temperature measuring accuracy as the innolight, even with the back panel voltage readout that is hooked up to the multimeter that lives under the optical table. The best I can tell is around 40.1-40.2 degrees C. Y_SLOW_SERVO2_OFFSET of -10690 counts gives a beat <100MHz at FSS_SLOWDC=0. This is actually very close to the previous operating point, so the same mode seems to still work. 

The arms now easily lock on ALS, albeit with higher noise. With arms locked on ALS, POX and POY show >1kHz rms noise. frown 


I gave the PRFPMI locking script a few tries, but it's having problems keeping the PRMI locked. The ALS is a few times noisier than usual, and I haven't revisited the validity of the PRC angular feedforward, so I'm not so surprised. 

  9248   Wed Oct 16 19:19:14 2013 MasayukiUpdateGreen LockingEnd PDH 60 Hz comb noise in YARM

[Manasa, Masayuki]

- Motivation

For PRMI + 2arm, we tried to make the ALS control noise better. As this entry we had huge 60 Hz comb noise in PDH loop of YARM.

So we tried to figure out the problem and fix it.

- What we did

We checked which power supply the staff in Y-end are connected to, and change some of them to connect to 1Y4 AC power supply from wall AC. What we changed was
1.Main end laser
2.He-Ne laser
3.Green REFL PD

We checked error signal of PDH control and compared before and after. The 60 Hz peak get better from -80 dBVpk to -90 dBVpk. Also I attached the plot of XARM, privious YARM (the data of Yesterday night), and current YARM ALS in-loop noises. The RMS of ALS in-loop noise of Y-arm get better by factor of 2. However, even the 60 Hz comb noise get better than before, RMS get worse by comb noise. 

We would like to make these noise better at least until these noises don't affect to RMS, so we should continue to check.

Attachment 1: comb_noise.pdf
comb_noise.pdf
  9273   Thu Oct 24 04:07:32 2013 MasayukiUpdateGreen LockingEnd PDH control signal, X-end PDH servo gain optimization

Control signal measurement of end PDH control

The Yarm ALS wasn't robust. Yesterdays night, we found that suspension kicked by something and that was the reason why the end PDH control lost lock. To make sure that the PDH loop itself is robust, I measured control signals of End PDH loops. When the gain inclease, the peak at UGF appeared and become unstable. Both arms does not seems unstable before the peaks appear.

controlsignal.png

 Xarm PDH servo gain optimization

I optimized the x end PDH servo gain with measuring OLTF. Now the servo gain is 5.0. UGF is around 10 kHz and phase margin is 40 degree.

OLTF.png

Also I measured out of loop noise. I locked the arm using IR PDH, and measure the ALS error signal. The high frequency noise become better.

outojloop.png

  16538   Sun Jan 2 20:46:46 2022 KojiUpdateSUSEnd SUS Electronics building

19:00~ Start working on the electronics bench

The following units were tested and ready to be installed. These are the last SUS electronics units and we are now ready to upgrade the end SUS electronics too.

40m End ADC Adapter Unit D2100016 / 2 Units (S2200001 S2200002)

40m End DAC Adapter Unit D2100647/ 2 Units (S2200003 S2200004)

These are placed on Tega's desk together with the vertex DAC adapters

0:30 End work

Attachment 1: PXL_20220103_081133119.jpg
PXL_20220103_081133119.jpg
  10629   Tue Oct 21 18:40:46 2014 manasaUpdateGeneralEnd laser fiber setup

 [Manasa, Diego]

This is our first time touching tables for Frequency Offset Locking. 

The goal was to couple the 1064nm that leaks after the SHG crystal and couple it into the fiber before we run it along the length of the arm. 

The fiber has been mounted at the end but there is no light coupled into the fiber as yet.

In the process, the following were done:

1. ETMY oplev servo disabled. This was enabled after the work.
2. NPRO laser power was reduced so that nothing gets burnt accidently while putting things on the table. This was also reset after the work.

The arms could be locked to green and IR after the work. So I am hoping today's work will not affect locking.

  47   Thu Nov 1 16:42:48 2007 Andrey RodionovSummaryEnvironmentEnd of Daylight Saving Time this weekend
Useful information for everyone, as a friendly reminder:

According to the web-page

http://www.energy.ca.gov/daylightsaving.html,

this coming weekend there will be the end of Daylight Saving Time.

Clocks will be adjusted backward one hour.
  16869   Mon May 23 13:16:59 2022 PacoUpdateBHDEnd of vent - checks

[Paco, Yuta]

Prep for closing and pump down.

  • Aligned IFO to maximize DC levels.
    • YARM (flashing peak 0.05 with PRM misaligned), XARM (flashing peak 0.06 with PRM misaligned), PRC (PRY flashing -30 @ POPDC, offset -70 and REFL DC 270), SRC (SRY flashing -30 @ POPDC, offset -70), BHD.
  • GTRY clipping
    • We tried moving the alignment of PR3, PR2, ETMY, ITMY to reduce clipping and retain IR flashing. We found it kind of difficult, so we only used the unclipped GTRY temporarily to improve the input YAUX injection after which the YAUX locked. We then restored the clipping in favor of the IR beam alignment.
  • PR3 position
    • PR3 seems to be +1 inch away towards East, nominally placed along North-South, and offset in YAW.
  • Aligned OPLEVs to center at around Mon May 23 13:20:32 2022
  • Snapshot of all cameras in the control room around Mon May 23 13:24:51 2022

[Chub, JC, Jordan, Yuta, Yehonathan, Paco]

Closed in the following order:

  • IMC chamber
  • OMC chamber
  • BS chamber
  • ITMY chamber
  • ITMX chamber

[Yuta, Paco]

After closing the heavy doors, we tried to have GTRY less clipped using PR2, PR3, ITMY and ETMY. During this adventure, we also aligned GRY injection beam by hand. Rotating a waveplate for GRY injection made GRY locking stably at GTRY of ~0.3.

Attachment 1: OSEMs20220523.txt
2022-05-23 13:15:59.531109
 =====BS=====
C1:SUS-BS_ULSEN_OUT16 =   600 
C1:SUS-BS_LLSEN_OUT16 =   575 
C1:SUS-BS_URSEN_OUT16 =   600 
C1:SUS-BS_LRSEN_OUT16 =   635 
C1:SUS-BS_SDSEN_OUT16 =   670 
 =====ITMX=====
C1:SUS-ITMX_ULSEN_OUT16 =   404 
C1:SUS-ITMX_LLSEN_OUT16 =   608 
... 76 more lines ...
Attachment 2: Screenshot_2022-05-23_13-24-06_IFOaligned_GTRY_still_clipping_but_less.png
Screenshot_2022-05-23_13-24-06_IFOaligned_GTRY_still_clipping_but_less.png
Attachment 3: PR3position.JPG
PR3position.JPG
  3432   Wed Aug 18 11:40:59 2010 josephb,kiwamu,yoichiUpdateCDSEnd station working with latest RCG code

We were able to get the latest SVN checkout (revision 2009), working with the c1x00 (IOP) and c1vgl models at the new X end on the c1iscex machine.

We were unable to get it working yesterday, and as far as we can tell, the only significant change was a reboot of the c1sus machine.  Before the reboot, it did not look like there were any conflicting models running on the c1sus machine, but apparently something was preventing the models on c1iscex from running properly.

Other simulated plant models still need to have their shared memory location blocks updated to the new type.

Now that we have proved we still can get the end model running, we are moving onto the c1sus machine.

  8892   Mon Jul 22 17:17:30 2013 annalisaUpdateendtable upgradeEnd table picture
Attachment 1: YendTable.jpg
YendTable.jpg
  8896   Tue Jul 23 00:51:46 2013 KojiUpdateendtable upgradeEnd table picture

The spot on the IPANG QPD was checked. The spot is higher than the center and South side of the lens.
Some photos are found below.

The spot on the IPANG steering mirrors in the ETMY chamber was also checked.
It is clipped at the top of the steering mirror. (See attachment 4)
So basically the spot is about 1" above the center of the mirror.

Attachment 1: P7224222.JPG
P7224222.JPG
Attachment 2: P7224223.JPG
P7224223.JPG
Attachment 3: P7224224.JPG
P7224224.JPG
Attachment 4: P7224238.JPG
P7224238.JPG
  8910   Tue Jul 23 19:37:08 2013 AnnalisaUpdateendtable upgradeEnd table picture

Quote:

The spot on the IPANG QPD was checked. The spot is higher than the center and South side of the lens.
Some photos are found below.

The spot on the IPANG steering mirrors in the ETMY chamber was also checked.
It is clipped at the top of the steering mirror. (See attachment 4)
So basically the spot is about 1" above the center of the mirror.

 After the vent, the IPang spot position on the steering mirrors on the Yend table moved approximately by 1 inch down.

Inside the chamber, the spot position is in the center of the steering mirror. (difficult to take a picture because the PSL beam power has been reduced)

Attachment 1: IMG950073.jpg
IMG950073.jpg
Attachment 2: IMG950072.jpg
IMG950072.jpg
  7731   Tue Nov 20 11:40:19 2012 ranaConfigurationGreen LockingEnd table upgrade for auxiliary green laser : ETMX layout on new table

 

 Mounts:

  1. No more mounts using the 1" dia. pedestal / fork technology.
  2. No more mounts using the 1/2" post / post holder technology. Both of these are loose, weak, and cause noise.
  3. All steerable mirror mounts which carry the important sensing beams should use steel mounts (e.g. Polaris from Thorlabs). Aluminum mirror mounts are not to be used.
  4. The mounts must be mounted to a 3/4" steel post (these are the custom ones we used in the PSL; Steve should get some more of them made).
  5. The post is then mounted on an aluminum base (The BA2 or BA3 (2" x 3" aluminum) from Thorlabs is OK. The 1" x 3" ones are not). These must be fastened to the table using 2 screws, each with a SS washer.

 

  9345   Tue Nov 5 16:47:09 2013 JenneUpdateLSCEnd transmission QPDs

I think Steve is trying to align the end transmission QPDs, since the arms are locked nicely right now.  I noticed that the QPDX pitch and yaw signals were digital zeros.  A quick look determined that the QPDX matrix to go from 4 quadrants to 3 degrees of freedom had been filled in for the POS row, but not pitch and yaw.  So, I copied the QPDY matrix over to QPDX (so the ordering of the rows and columns is assumed to be the same). 

Hopefully this will get us close to centered, but I suppose we ought to check really which quadrant is which, by shining a laser pointer at each quad at each end.

ELOG V3.1.3-