Mon Dec 3 19:44:58 2007 |
Accelerometers on new mounts | Configuration | PEM | Andrey |
I (Andrey) continued today working with new accelerometer mounting. (see entry #151 about my Friday work).
I bought screws/washers and attached those mounts with accelerometers to metallic frames which are firmly cemented to the floor.
One such mount with three accelerometers (in X-, Y-, Z-directions) is installed near the ITMX (in the previous location, but NOT on top of the unused stack as before Friday), the other mount with three accelerometers in three orthogonal directions is installed near ETMX in the east end of the room (this set of accelerometers was installed between MC and BS before Friday). I uncoiled the cables, put them into the cable tray towards the ETMX, and hooked-up the three accelerometers near ETMX in the east end of the room.
Now all six accelerometers are hooked-up (that is, connected to power supply board with cables).
We decided with Steve Vass to put red cones (similar to those that are on highways in the road construction zones) in order to prevent people from bumping into accelerometers. Please use caution when walking along the X-arm.
I took several pictures of the new accelerometer setup. Picture "DSC_0194.JPG" shows the mount with accelerometers near the the ITMX and the beamsplitter chamber,
picture "DSC_0195.JPG" is the "zoomed-in" view of the same accelerometers, while picture "DSC_0196.JPG" shows the mount with accelerometers near ETMX in the east end of the room.
Many thanks to Mr. Steve Vass for his thorough explanation/showing me how to drill the metal and put threads in the holes. |
Attachment 1: DSC_0194.JPG
|
|
Attachment 2: DSC_0195.JPG
|
|
Attachment 3: DSC_0196.JPG
|
|
162
|
Mon Dec 3 22:20:09 2007 |
tobin | Configuration | PSL | ISS |
I replaced the painfully short 1' cables on the ISS photodiodes with luxurious five foot cables, made by chopping a ten foot Amphenol cable (P/N:CS-DSPMDB09MM-010) in half and using each half for one of the diodes. All of the ISS GND connections are wired to the PD GND, as is the PD- differential signal. The diodes are installed on the PSL table, but I have not tested them beyond looking at the DC values as I blocked/unblocked the beam. |
172
|
Wed Dec 5 23:19:03 2007 |
Andrey | Configuration | PEM | Accelerometers are turned on |
All accelerometers have been turned on, as Alan asked during Wednesday meeting.
Typical power spectra and coherence plots are attached below.
"East" in the name means that the previous location of accelerometrs was to the east from "Beamsplitter" (the location for "east" accelerometers was not changed, actually, it is still near ITMX), while "west" means that previously accelerometers were to the west from the BS, but now their new location is near the ETMX.
I will change the names of the channels tomorrow (Thursday) when someone (Tobin?) will show to me how to do it.
P.S. (addition made on Dec. 19th, 2007, by Andrey) I intended to change the names of accelerometers the next day, Thursday Dec. 06,
but I did not do it that day (did not understand how to do it), then I fell ill, and eventually
I changed the names of accelerometers on December 19th, see entry to ELOG #204) |
Attachment 1: Power_Sp_and_Coh_XY-EAST.pdf
|
|
Attachment 2: Coherence-ZX_East.pdf
|
|
Attachment 3: Coherence-ZY_East.pdf
|
|
Attachment 4: Power_Sp_WEST.pdf
|
|
Attachment 5: Coherence-ZX_West.pdf
|
|
Attachment 6: Coherence-XY_West.pdf
|
|
Attachment 7: Coherence-YZ_West.pdf
|
|
176
|
Thu Dec 6 19:19:47 2007 |
Andrey | Configuration | SUS | Suspension damping Gain was restored |
Suspension damping gain was disabled for some reason (all the indicators in the most right part of the screen C1SUS_ETMX.adl were red), it is now restored. |
186
|
Mon Dec 10 19:08:03 2007 |
tobin | Configuration | PSL | MZ |
The MZ seems finicky today--it keeps unlocking and relocking.
I've temporarily blocked one of the MZ arms while I work on the ISS. |
187
|
Mon Dec 10 20:35:59 2007 |
tobin | Configuration | Computer Scripts / Programs | autolocking scripts |
I added this tidbit of csh code to the MZ autolocker to prevent multiple copies from running (on one computer):
if (`pgrep lockMZ | wc -l` > 1) then
echo lockMZ is already running!
exit
endif Similarly, here's some bash code that does something similar; I'll add it to the other autolocker scripts:
if
pgrep `basename $0` | grep -v $$ > /dev/null
then
echo Another copy of this program is already running. Exiting!
exit 1
fi This code searches for all processes with the same name as this script ($0) and then use grep to exclude (-v) the current process ID ($$). |
191
|
Thu Dec 13 23:56:02 2007 |
Andrey | Configuration | Computer Scripts / Programs | Overnight measurements |
After my disease (fever, vomitting, nose problem, overall weakness) I returned to LIGO today for the first time after the weekend, and I am running the script for the XARM-measurements over this night.
So, suspension dumping gains should undergo changes in the interval from 1 to 10 in both ITMX and ETMX.
XARM has been of course locked.
I started running the script for the first time at about 10PM, but I realized after an hour and a half that my step of gain increase 0.2 was too shallow, too small to execute my program during one night. Therefore, I needed to terminate the program, change my program so that it increases the gain with increment 0.5, not 0.2, and started it again around midnight.
Going home.
P.S. The red pump that I borrowed from the lab (Steve's pump?) is back at its previous place. The tire-worker tells me that I absolutely need to change all four tires for almost 500 dollars. I regret a lot about that huge money loss. |
194
|
Mon Dec 17 23:42:08 2007 |
Andrey | Configuration | Computer Scripts / Programs | Overnight measurements in X-arm |
I am making overnight measurements this night (from Monday to Tuesday) in XARM.
The X-arm is now locked, and the values for suspension damping gain will be changed in the interval from 1 to 7 with the step 0.5 in both ITMX and ETMX.
This is the second, repeated measurement. The results of the first measurement from Saturday to Sunday night will be reported in the separate ELOG entry (sorry, I did not make an ELOG entry on Saturday evening about running the program overnight).
The very first attempt to run the script in the night from Thursday to Friday was not successful. |
198
|
Tue Dec 18 23:27:36 2007 |
Andrey | Configuration | Computer Scripts / Programs | New overnight measurements (this night from Tue to Wed) |
I am making overnight measurements in XARM tonight.
This is the third night of measurements in XARM, but tonight I am scanning the narrower region between values of damping gain 1.00 and 4.50 with the smaller step 0.25. (for comparison, during two previous measurements the region was between 1.0 and 7.0 with the step 0.5).
I have relocked the XARM before the start of the measurements.
I started running the program at 9.30PM, and it should collect all the data by 9.00AM wednesday morning.
Below are explanations why I chose these different parameters for the interval and step:
I am going to put the results of previous night measurements into the next ELOG entry, and it we be pretty obvious from those graphs that results in XARM from the two previous (different) nights agree well with each other, and the approximate positions of minima and areas of "big growth" of the surfaces are pretty obvious from those graphs. It is clear that RMS are too big for the values of the damping gain bigger than 4.0, and that minima are somewhere near the values of 2.0. But those graphs were too rough to locate a somewhat precise value for the minima. Therefore, I am studying tonight the interval of gains between 1.00 and 4.50 with a smaller step.
A short note how I estimate time that is necessary to collect the experimental data.
there are 15 experimental points for each ETMX and ITMX suspension gains in the interval between 1.00 and 4.50 with the step 0.25. They are: 1.00, 1.25, 1.50, 1.75, 2.00, ..., 3.75, 4.00, 4.25, 4.50. As I am changing both ETMX and ITMX gains, I have an array of 15*15=225 elements.
It takes 3 minutes for each point to collect the data (I wrote the program that way). Therefore, the total time it takes to run the program is: 225*3=675 minutes, or 675/60=11.25 hours, almost 11 and a half hours. |
211
|
Sat Dec 22 00:52:57 2007 |
tobin | Configuration | PSL | ISS surgery |
In an attempt to quell oscillations in the (unused) outer loop portion of the ISS, I shorted the "PD+" and "PD-" signals from the (nonexistent) outer-loop diodes, and soldered in 47pf compensation capacitors in C92 and C220. This seems to have eliminated oscillations seen at TP41 and TP42. There's still something amiss at TP30 and maybe TP20. Otherwise, the ISS seems happy. I can turn the gain slider to +15dB without saturation (with the HEPA off), though there seems to be less light on the diode (~3.9V) than a week or two ago. |
238
|
Mon Jan 14 23:11:26 2008 |
tobin | Configuration | General | fiber |
John and I removed the fiber that ran from the SP table to the cleanroom. We plan to build a MZ interferometer with this fiber inserted into one of the arms, for the purpose of measuring its phase noise. |
243
|
Wed Jan 16 19:57:49 2008 |
tobin | Configuration | Photos | ISCT_EX |
Here's a photo of the ISCT_EX table, for the purpose of planning our auxiliary laser arm locking scheme. Note the (undumped!) beam from the beamsplitter before QPDX (the leftmost gold-colored box); perhaps we could inject there. |
Attachment 1: trx-annotated-small.jpg
|
|
250
|
Fri Jan 18 20:53:56 2008 |
tobin | Configuration | General | ETMY oplev |
I monkeyed around with the ETMY oplev, adding a folding mirror and moving the HeNe so that John, Sam, and I have more room for our auxiliary laser setup. (The ISCT-EY has more room than ISCT-EX; the latter has an extra photodiode for IP ANG.) I believe I successfully recommissioned the oplev, though it might not be up to the SV standard. I verified that wiggling the ETMY alignment sliders showed corresponding wiggles in the oplev signals. However, it seems poorly diagonalized.
Our current plan is to have an NPRO, EOM, and fiber coupler on the SP table. This fiber will take light to ISCT-EY where we'll have a mode-matching telescope and inject light to the Y arm via a polarized beamsplitter. This auxiliary beam will have polarization orthogonal to the beam from the PSL. |
260
|
Thu Jan 24 20:03:40 2008 |
Andrey | Configuration | SUS | Changes to Dataviewer channels (XARM) |
1) Good news. I added a chanel "C1:SUS-ETMX_POS" to Dataviewer.
I followed the instructions from WIKI-40:
modify the file "C1SUS_EX.ini" in /cvs/cds/caltech/chans/daq,
then telnet to fb40m,
then "click the appropriate blue button on the DAQ MEDM screen".
So, I can now read a signal from the channel "C1:SUS-ETMX_POS" in Dataviewer,
and this allows me to measure Q-factors of ETMX this evening (make similar work for what I did on Tuesday for ITMX).
2) BAD NEWS. While "clicking the appropriate blue button" on the DAQ MEDM screen,
namely CODAQ_DETAIL,adl screen, I obviously clicked some blue button that I should not have clicked,
and as a result the signal in Dataviewer from the channel "C1:SUS-ITMX_POS" has disappeared (it is now a straight line).
Description of what has happened and of my wrong actions:
I had two channels opened in Dataviewer simultaneously (both "C1:SUS-ETMX_POS" and "C1:SUS-ITMX_POS"),
and after clicking some blue button on CODAQ_DETAIL,adl screen, the signal from "C1:SUS-ITMX_POS" became
a straight line, while signal from "C1:SUS_ETMX_POS" continued to be a random noise.
I was scared that I made worse for the channels and for Dataviewer, and I started clicking random blue buttons chaotically hoping that it will restore the signal from "C1:SUS-ITMX_POS". Random clicking on arbitrary blue buttons did not return the signal.
As the channel "C1:SUS-ETMX_POS" works normally, I will be measuring Q-factors of ETMX tonight,
but it is obvious that someone else (Rana, Robert,Steve?) needs to restore the correct settings for "C1:SUS-ITMX_POS".
Moreover, as I was clicking chaotically all the blue buttons on CODAQ_DETAIL,adl screen, someone else (Rana, Robert, Steve?) will need to check somehow that I did not destroy signals from some other channels.
I apologize for the negative consequences of my channel adding,
but Rana asked me in the very beginning in September to let others know if I spoil something, so that others would be aware of it and could fix the problem.
Again, I apologize and hope that the problem is not very serious. |
261
|
Thu Jan 24 22:10:49 2008 |
Andrey | Configuration | Computer Scripts / Programs | Problem with channels - help of Rana, Robert or Steve is needed |
I definitely spoiled something (changes some settings) by chaotically clicking those blue buttons (see my previous entry # 260).
Unfortunately, I cannot use standard library of functions for reading from channels from mDV directory.
Although I see the curve of a noise in the Dataviewer from the channel "Ch1:SUS_ETMX_POS", when I try to read data from the channel using the program "get_data" from MDV directory, I get the error message
"Warning: No data available for [numbers representing "gps_start_time" and "gps_end_time"].
In new_readframedata at 136
In new_fetch_shourov at 71
In get_data at 98"
I checked, both gps-times are in the past from now, so as far I understand, nothing is recorded into the channels.
Of course, I added two hours ago to the directory "mDV", that is I used addpath(pwd) in that directory.
And I also cannot run the program that I used on Tuesday evening which takes data from "C1:SUS_ITMX_POS" (no data from that channel), which worked perfectly on Tuesday.
I again apologize for clicking the wrong blue button (see my explanation in my previous message #260). I ask someone who knows how to return normal working of channels (normal interaction of computer and channel memory) to do that.
Before that I cannot take data. I do not know how to restore the initial settings which existed before I started adding the channel to Dataviewer.
Andrey. |
263
|
Fri Jan 25 08:55:26 2008 |
rob | Configuration | General | Changes to Dataviewer channels (XARM) |
As a general rule,
Quote: | clicking random blue buttons chaotically |
is not a good problem solving technique. It is thus now explicitly discouraged as an option in the LIGO 40m Lab. |
265
|
Fri Jan 25 10:14:35 2008 |
rob | Configuration | SUS | Changes to Dataviewer channels (XARM) |
Quote: |
2) BAD NEWS. While "clicking the appropriate blue button" on the DAQ MEDM screen,
namely CODAQ_DETAIL,adl screen, I obviously clicked some blue button that I should not have clicked,
and as a result the signal in Dataviewer from the channel "C1:SUS-ITMX_POS" has disappeared (it is now a straight line).
Description of what has happened and of my wrong actions:
I had two channels opened in Dataviewer simultaneously (both "C1:SUS-ETMX_POS" and "C1:SUS-ITMX_POS"),
and after clicking some blue button on CODAQ_DETAIL,adl screen, the signal from "C1:SUS-ITMX_POS" became
a straight line, while signal from "C1:SUS_ETMX_POS" continued to be a random noise.
I was scared that I made worse for the channels and for Dataviewer, and I started clicking random blue buttons chaotically hoping that it will restore the signal from "C1:SUS-ITMX_POS". Random clicking on arbitrary blue buttons did not return the signal.
As the channel "C1:SUS-ETMX_POS" works normally, I will be measuring Q-factors of ETMX tonight,
but it is obvious that someone else (Rana, Robert,Steve?) needs to restore the correct settings for "C1:SUS-ITMX_POS".
Moreover, as I was clicking chaotically all the blue buttons on CODAQ_DETAIL,adl screen, someone else (Rana, Robert, Steve?) will need to check somehow that I did not destroy signals from some other channels.
I apologize for the negative consequences of my channel adding,
but Rana asked me in the very beginning in September to let others know if I spoil something, so that others would be aware of it and could fix the problem.
|
I eventually resolved the situation by restarting the c1susvme1 processor, which had somehow got confused by the clicking random blue buttons chaotically. The data acquisition should be working again. |
266
|
Fri Jan 25 11:38:16 2008 |
josephb | Configuration | Cameras | Working GiGE video on Linux - sort of |
1)I have been able to compile the SampleViewer program which can stream the video from the Prosilica 750C camera. This was accomplished on my 64-bit laptop running Ubuntu, after about 3 hours of explicitly converting strings to wxStrings and back again within the C++ code. (There was probably an easier way to simply overload the functions that were being called, but I wasn't sure how to go about doing so). By connecting it to the CDS network, I was able to immediately detect the camera and display the images.
Unfortunately, I have not yet been able to get it to compile on Mafalda with the x86 architecture. This may be do the fact that it has wxWidgets version 2.8.7 while my laptop has 2.8.4. Certainly the failure at compile time looks different from the errors earlier, and seem to be within the wxWidget code rather than the SampleViewer code. I may simply need to uninstall 2.8.7 and install 2.8.4 of wxWidgets.
The modified code that will compile on my machine has been copied to /cvs/cds/caltech/target/Prosilica/examples/SampleViewer2b.
2)The Snap program (under /cvs/cds/caltech/target/Prosilica/examples/Snap) also will now take full resolution images even on Mafalda. This was achieved by reducing the packet size to 1000 and also increasing the wait until timeout time up to 400 ms, which originally was at 100. Apparently, it takes on the order of 1 ms per packet as far as I can tell. So full resolution at 752x480 required something of order 360 packets.
To Do:
1) Get sample viewer to compile on Mafalda, and then statically compile it so it can be run from any Linux based machine.
2) Get a user friendly version of Snap up and running, statically compiled, with options for a continuous loop every X seconds and also to set desired parameters (such as height, width, file name to save to, save format, etc).
3) Figure out data analysis with the images in Matlab and an after the fact image viewer.
Attached is an example .tiff image from the Snap program. |
Attachment 1: snap.tiff
|
267
|
Fri Jan 25 13:36:13 2008 |
josephb | Configuration | Cameras | Working GiGE video on Mafalda |
Finally got the GiGE camera sample viewer video running on Mafalda by updating to the latest API (version 1.16 from Dec 16, 2007) from Prosilica and then using the modified Sample Viewer code I had written. The API version previously in cvs was 1.14.
It can currently be run by ssh -X into Mafalda and going to /cvs/cds/caltech/target/Prosilica/bin-pc/x86 and running the SampleViewer executable found there. |
269
|
Fri Jan 25 17:11:07 2008 |
Max , Andrey | Configuration | General | NEW_FETCH_SHOUROV and GET_DATA do not work |
The problem which started yesterday after Andrey's framebuilder restart still persists.
It is still impossible to read data in the past from the channels using "get_data" which in turn uses "new_fetch_shourov".
Max was trying to read data from the channel
"C1:LSC-DARM_CTRL",
and he got the same error messages as Andrey.
Andrey tried earlier today to read data from "C1:SUS-ITMS_SUS" or "C1:SUS-ETMX_SUS" with the error meassge
Error in ==> new_fetch_shourov at 22
at (start_time+duration) > stops(end)
So, it seems that Robert Ward fixed just one problem out of two problems.
Robert revived the realtime signals in Dataviewer,
but did not revive the memory of channels for new_fetch_shourov.
To be more precise, channels have memory (it is possible to see the "Playback" curves in Dataviewer"),
but "get_data" and "new_fetch_shourov" do not see the data from those channels. The problem appeared immediately after Andrey's clicking on blue buttons to restart the framebuilder.
Andrey again apologizes. |
281
|
Mon Jan 28 17:16:54 2008 |
Andrey | Configuration | Computers | Matlab libraries DO NOT WORK properly sometimes |
Working in Matlab, I encountered at two different times today the license distribution problem:
??? License checkout failed.
License Manager Error -4
Maximum number of users for Curve_Fitting_Toolbox reached.
Try again later.
To see a list of current users use the lmstat utility or contact your License Administrator.
Troubleshoot this issue by visiting:
http://www.mathworks.com/support/lme4a |
288
|
Thu Jan 31 12:39:14 2008 |
John | Configuration | General | Y arm test mass cameras |
I've adjusted the test mass cameras on the y arm to make the beam injected through ETMY more visible. |
289
|
Thu Jan 31 16:53:41 2008 |
josephb | Configuration | Cameras | Improving camera user interface |
There's a new and improved version of Snap program at the moment people are free to play with.
Located in /cvs/cds/caltech/target/Prosilica/40mCode/
The program Snap now has a -h or --help option which describes some basic command line arguments. The height (in pixels), width (in pixels), exposure time (in micro seconds), file name to be saved to (in .tiff format), and packet size can all be set. The format type (i.e. pixel format such as Mono8 or Mono16) doesn't work at the moment.
At the moment, it only runs on mafalda.
Currently in the process of adding a loop option which will take images every X seconds, saving them to a given file name and then appending the time of capture to the file name.
After that need to add the ability to identify and choose the camera you want (as opposed to the first one it finds).
Lastly, I've been finding on occassion that the frame fails to save. However if you try again a few seconds later with the exact same parameters, it generally does save the second time. Not sure whats causing this, whether on the camera or network side of things.
I've attached two images, the first at default exposure time (15,000 microseconds) and the second at 1/5th that time (3,000 microseconds).
The command line used was "./Snap -E 3000 -F 'Camera_exp_3000.tiff' " |
Attachment 1: Camera_exp_15000.tiff
|
Attachment 2: Camera_exp_3000.tiff
|
292
|
Fri Feb 1 15:04:54 2008 |
josephb | Configuration | Cameras | Snap with looping functionality available |
New GiGE camera code is available in /cvs/cds/caltech/target/Prosilica/40mCode/. Currently only runs on Mafalda.
Snap has expanded functionality to continuously loop infinitely or for a maximum number of images set by the user. File names generated with the loop option have the current Unix time and .tiff appended to them. So -f './test' will produce tiff files with format "test1234567.tiff". The -l option sets the number of seconds between images.
"./Snap -l 5 -i -f './test' " will cause the program to infinitely loop, saving images every 5 seconds. Using "-m 10" instead of "-i" will take a series of 10 images every 5 seconds (so taking a total of 50 seconds to run).
It also now defaults to 16-bit (in reality only 10 bit) output instead of 8 bit output. You can select between the two with -F 'Mono8' or -F 'Mono16'.
Use --help for a full list of options.
Note that if you ctrl-c out of the loop, you may need to run ./ResetCamera 131.215.113.104 (or whatever the IP is - use ./ListCameras to determine IP if necessary) in order to reset the camera because it doesn't close out elegantly at the moment. |
297
|
Tue Feb 5 15:32:29 2008 |
josephb | Configuration | Cameras | PMC and the GigE Camera |
The PMC transmission video camera has been removed and replaced with the GigE GC750 camera for the moment.
A ND4.0 filter has been added in the path to that camera to reduce saturation for the moment.
The old camera has been placed on the elevated section inside the enclosure, and the cable for it is still on the table proper.
The Gige camera is currently running the Snap code on Linux3 with the following command line:
./Snap -E 2000 -l 60 -m 1440 -f './pmc_trans/pmc_trans'
So its going to be taking tiff images every minute for the next 24 hours into the cvs/cds/caltech/target/Prosilica/40mCode/pmc_trans/ directory.
Attached is an example image with exposure set to 2000, loaded into matlab and plotted with the surf command. 2500 microseconds looked like it was still saturating, but this seems to be a good level (with a max of 58560 out of 65535). |
Attachment 1: pmc_trans.jpg
|
|
298
|
Tue Feb 5 17:39:05 2008 |
jweiner | Configuration | PEM | PEM-AS_MIC taken down from AS table, will put in PSL enclosure soon |
I took down the microphone that Andrey hung above the AS table his first week in lab. I want to hang the microphone above the PMC to check the effect of acoustic noise on the PMC. The cables were a little more tangled than I thought so I've only taken the microphone down and haven't hung it back up yet, but on Thursday I'll have enough time to carefully put it up inside the PSL and see what I can find out about acoustic noise inside the PSL. I think the microphone should be sensitive enough for the frequencies we're interested in, and I'll hopefully find out if it's sufficient once I put it up in the PSL. The microphone cable and microphone are on top of the PSL for now. |
300
|
Wed Feb 6 16:50:47 2008 |
josephb | Configuration | Cameras | Regions of Interest and max frame rate |
The Snap code has once again been modified such that setting the -l option to 0 will take images as fast as possible. Also, the -H and -W options set the height and width, while in principle the -Y and -X options set the position in pixels of the top edge and left edge of the image. It also seems possible to set these values such that the saved image wraps around. I'll be adding some command checking so that the user can't do this in the near future.
Doing some timed runs, using a -H 350 and -W 350 (as opposed to the full 752x480), 100 images can be saved in roughly 8 seconds, and 1000 images took about 73 seconds. This corresponds to a frame rate of about 12-13 frames per second (or a 12-13 Hz display). The size of this area was sufficient to cover the current PMC transmission beam.
The command line I used was
time ./Snap -l 0 -m 1000 -f 'test' -W 350 -H 350 -Y 50 -X 350 -E 2000
Interestingly enough, there would be bursts of failed frame saves if I executed commands in another terminal (such as using ls on the directory where the files were being stored).
As always, this code is available in /cvs/cds/caltech/target/Prosilica/40mCode/. |
301
|
Wed Feb 6 19:39:11 2008 |
rana | Configuration | Cameras | Regions of Interest and max frame rate |
We really need to look into making the 40m CDS network have an all GigE backbone so that we can have cooler cameras as well as collect multiple datastreams... |
303
|
Fri Feb 8 17:55:53 2008 |
josh | Configuration | PEM | PEM-AS_MIC now in PSL enclosure |
I have moved the microphone to the PSL enclosure, hanging near the south (Y) side from a support rod for the overhanging storage area so that it's reasonably close to the PMC. I've fastened it in many places using cable ties to make sure that it won't fall.
Alberto helped me solder together a female BNC-female 3.5 mm stereo adapter so that I can use the DAQ to output through BNC to PC speakers. My plan is do sweep sine output through PC speakers to find the transfer function of sound from outside the enclosure to inside the enclosure and by moving the microphone more centrally over the PSL table, check if there are any strong resonances. Hopefully I can use this technique at other places around the interferometer or measure the effect of installing acoustic foam. |
307
|
Sun Feb 10 21:43:16 2008 |
rob | Configuration | IOO | MC alignment tweaked |
I adjusted the alignment of the free hanging mode cleaner to best transmit the PSL beam. |
309
|
Mon Feb 11 22:44:29 2008 |
rob | Configuration | DAQ | Change in channel trending on C1SUS2 |
I removed the MC2 optical lever related channels from trends, and the SRM POUT and YOUT (as these are redundant if we're also trending PERROR and YERROR). I did this because the c1susvme2 processor was having bursts of un-syncy lateness every ~15 seconds or so, and I suspected this might interfere with locking activities. This behaviour appears to have been happening for a month or so, and has been getting steadily worse. Rebooting did not fix the issue, but it appears that removing some trends has actually helped. Attached is a 50 day trend of the c1susvme2 sync-fe monitor. |
Attachment 1: srm_sync.png
|
|
311
|
Tue Feb 12 16:18:29 2008 |
rob | Configuration | Computer Scripts / Programs | autoburt cron moved to op340m |
I moved the autoburt cron job from op440m to op340m. For some reason, burtrb requires gcc to run (I gather it uses the C-preprocessor to parse input files), so I had to install that on op340m to get it to work properly.
There are no more cron jobs running on op440m now. |
345
|
Thu Feb 28 16:19:37 2008 |
josephb | Configuration | Computers | Mafalda rewired and multiple cameras running |
1) Mafalda is now connected via an orange Cat5E ethernet cord to the gigabit ethernet switch in rack in the office space. It has been labeled at both ends with "mafalda".
2) Both the GC650M camera (from MIT) and the GC750M are working. I can run the sampleviewer code and get images simultaneously. Unforutnately, the fps on both cameras seems to drop roughly in half (not an exact measurement) when displaying both simultaneously at full resolution.
3)Discovered the Gigabit ethernet card in Mafalda doesn't support jumbo packets (packets of up to 9k bytes), which is what they recommend for optimum speed.
4)However, connecting the cameras through only gigabit switches to Mafalda did seem to increase the data rate anyways, roughly by a factor of 2. (Used to take about 80 seconds to get 1000 frames saved, now takes roughly 40 seconds).
5)Need to determine the bottleneck on the cameras. It may be the ethernet card, although its possible to connect multiple gigabit cards to single computer (depending on the number of PCI slots it has). Given the ethernet cards are cheap ($300 for 20) compared to even a single camera (~$800-1500), it might be worth while outfitting a computer with multiple. |
346
|
Thu Feb 28 19:37:41 2008 |
rob | Configuration | Computers | multiple cameras running and seisBLRMS |
Quote: | 1) Mafalda is now connected via an orange Cat5E ethernet cord to the gigabit ethernet switch in rack in the office space. It has been labeled at both ends with "mafalda".
2) Both the GC650M camera (from MIT) and the GC750M are working. I can run the sampleviewer code and get images simultaneously. Unforutnately, the fps on both cameras seems to drop roughly in half (not an exact measurement) when displaying both simultaneously at full resolution.
3)Discovered the Gigabit ethernet card in Mafalda doesn't support jumbo packets (packets of up to 9k bytes), which is what they recommend for optimum speed.
4)However, connecting the cameras through only gigabit switches to Mafalda did seem to increase the data rate anyways, roughly by a factor of 2. (Used to take about 80 seconds to get 1000 frames saved, now takes roughly 40 seconds).
5)Need to determine the bottleneck on the cameras. It may be the ethernet card, although its possible to connect multiple gigabit cards to single computer (depending on the number of PCI slots it has). Given the ethernet cards are cheap ($300 for 20) compared to even a single camera (~$800-1500), it might be worth while outfitting a computer with multiple. |
I found the SampleViewer running and displaying images from the two cameras. This kept mafalda's network so busy that the seisBLRMS program fell behind by a half-hour from its nominal delay (so 45 minutes instead of 12), and was probably getting steadily further behind. I killed the SampleViewer display on linux2, and seisBLRMS is catching up. |
356
|
Tue Mar 4 19:14:09 2008 |
rana | Configuration | CDS | TDS & SVN |
Matt, Rob, Rana
Today we added the TDS software to the 40m SVN repo.
First we rationalized things by deleting all the old TDS directories and taking
the tds_mevans dir and making it be the main one (apps/linux/tds).
We also deleted all of the TDS directories in the project area. It is now very
likely that several scripts will not work. We're going to have the teething
problems of repointing everything to the nominal paths (in the apps areas).
Finally we did:svn import tds https://40m.ligo.caltech.edu/svn/40m/tds --username rana
to stick it in. To check it out do:
svn checkout https://40m.ligo.caltech.edu/svn/40m/tds --username rana
We'll get a couple of the O'Reilly SVN books as well to supplement our verion control knowledge.
Unitl then you can use the SVN cheat sheets available at:http://www.digilife.be/quickreferences/quickrefs.htm |
357
|
Tue Mar 4 20:14:02 2008 |
rana | Configuration | IOO | MC Alignment |
The MC alignment was pretty far off. We were getting TEM01 mode locks only.
Rather than inspect what happened I just aligned the MC suspensions to get
the transmission higher. Now Matt should be able to lock the X arm and collect
adaptive filter data. |
363
|
Fri Mar 7 00:47:54 2008 |
rana | Configuration | PEM | Ranger SS-1 |
Yesterday evening around 7:30 PM, I changed the Ranger seismometer from a
vertical to a horizontal seismometer. To do this I followed the instructions
in the manual.
1) Lock it down.
2) Turn it sideways. Use the leveling screws to center the bubble level.
3) Carefully loosen the hanger rod and release slowly the tension to allow
the mass to recenter.
4) Look through the little viewhole next to the rod to make the white lines
line up. This means the mass is centered.
5) Look at the output on a scope. It should be freely moving with a ~1 sec.
period.
The attached plot shows the before and after spectra. |
Attachment 1: ss1.pdf
|
|
367
|
Mon Mar 10 20:46:41 2008 |
John | Configuration | LSC | ETMY Trans PD & QPD |
I've placed a 10% reflector in the path from ETMY to the trans and quadrant photodiodes. |
368
|
Tue Mar 11 23:14:01 2008 |
rana | Configuration | PEM | Accelerometer and Seismometer movements |
Steve and Matt moved the accelerometers and seismometers today.
The accelerometers are now placed around the MC and the seismometer is in-between MC1 & MC2.
We have changed the names of the acc channels to reflect whether they are close to MC1/MC3
or MC2. We tested the accelerometer to channel name mapping by switching gains at the wilcoxon
breakout box and also by tapping. It seems now that the previous setup near the ITMX/ETMX had
some few channels mislabeled which would have given some confusing results.
Alex, Jay, and Rolf came over today and installed, then de-installed some of the hardware for
sending the PEM channels over to the C1ASS machine where the adaptive filter front end will go.
Everything should be back to the way it was...hopefully, the guys will modify the ADCU PEM
code to send the signals to the new FE over the reflective memory net and then send them to the
MCL inputs of the suspensions. So the first incarnation should use the accelerometers and seismometer
to drive MC1 and/or MC3. |
Attachment 1: Acc.pdf
|
|
369
|
Wed Mar 12 00:36:52 2008 |
rana | Configuration | PEM | Accelerometer and Seismometer movements |
I used the MISO FIR Wiener matlab code to see how well we might do in principle.
The attached 3 page PDF file shows the MC_L control signal (force on MC2) and the residual
after subtracting off the accelerometer and seismometer using a 32 Hz sample rate and
512 taps (page 1), 1024 taps (page 2), and 2048 taps (page 3). As Matt smarmily points out,
there's not a lot to win by going beyond 512; maybe a factor of sqrt(2) for a factor of 4
tap number. |
Attachment 1: finished.pdf
|
|
370
|
Wed Mar 12 00:40:35 2008 |
rana | Configuration | PEM | Accelerometer and Seismometer movements |
Same as above but with 2048 taps and a 128 Hz sample rate. Does much better at the 16 Hz bounce mode. |
Attachment 1: mc2048-128.pdf
|
|
371
|
Wed Mar 12 00:47:26 2008 |
rana | Configuration | PEM | Accelerometer and Seismometer movements |
And this is a cool snapshot showing how this operation used 16 cores on menkar ! |
Attachment 1: Screenshot.png
|
|
373
|
Thu Mar 13 02:52:06 2008 |
Lisa | Configuration | LSC | Locking with 3f |
Today we have tried to use the reflected signal demodulated at 3*f1 ~ 99 MHz (REFL31) for length control.
This signal is cool because it is generated by the beating of sidebands, so it is not very sensitive to what the carrier does inside the IFO.
In particular, its gain and the demodulation phase shouldn't change much while changing the CARM offset during the locking sequence.
The idea is therefore to use REFL31_I and REFL31_Q for controlling MICH and PRCL, with the goal of making the lock acquisition sequence more robust.
We minimized hardware changes by using the 199MHz demodulation board, changing the local oscillator to 99.586317 MHz, with an amplitude of +10 dbm (the 3f signals are therefore acquired as LSC-PD6_I and LSC-PD6-Q).
We locked both the PRM and the DRM in a stable way using the REFL31_I and REFL31_Q, after tuning the demodulation phase (50) and removing their offsets.
On the other hand, we weren't able to acquire the lock in the DRM configuration directly by using the 3f signals. We needed instead to use the f signals first, and switch to the 3f signals once the lock was already acquired, otherwise ending up locking DRM at a different working point.
One explanation for that might be the fact that the beam impinging upon the 3f diode is too big compared with the diode size (only 1 mm, half of the size of the f1 diode).
For these reason, in presence of misalignments, some of the reflected light goes in high order modes, which can be partially (or all) off the diode, thereby generating multi-zero crossing in the demodulated error signal.
The next step before making the test with the whole IFO is therefore to modify the telescope in front of the 3f diode in order to reduce the beam size and repeat the tests we did tonight in DRM configuration.
P.S.: We made a test by changing the frequency of the local oscillator by a little bit and then coming back to the original value. We observed that the phase of the signal can change, so every time this frequency is moved the 3f demod phase need to be retuned.
John, Rob, Rana, Lisa |
378
|
Fri Mar 14 12:06:29 2008 |
josephb | Configuration | Cameras | GC750 looking at ETMX while locked |
The GC750 (CMOS) is currently looking at the front of ETMX. Unfortunately, its being routed through a 10Mbit connection (which I will be purchasing a replacement for today), so getting it to send images to Mafalda/Linux 2 or 3 isn't working well, but by using a local gigabit switch and a laptop I can get sufficient speed for full images with the sample viewer.
The attached image is from a full 752x480 reslution with 10,000 microsecond exposure with the X-arm locked. Although it looks like I still need to work on the focusing. Will be switching the GC750 with the GC 650 (CCD) later today and comparing the resulting images. |
Attachment 1: ETMX_zoom_exp_10000_750.tiff
|
379
|
Fri Mar 14 14:59:51 2008 |
josephb | Configuration | Cameras | Comparison between GC650 (CCD) and GC750 (CMOS) looking at ETMX |
Attached are images taken of ETMX while locked.
The first two are 300,000 microsecond exposure time, with approximately the same focusing/zoom. (The 750 is slightly more zoomed in than the 650 in these images). The second are 30,000 microsecond exposures. The la
The CMOS appears to be more sensitive to the 1064 nm reflected light (resulting in bright images for the same exposure time). This may make a difference in applications where images are desired to be taken quickly and repeatedly.
Both seem to be resolving individual specks on the optic reasonably well.
Next test is to place both camera on a Gaussian beam (in a couple different modes say 00, 11, and so forth), probably using the PMC. |
Attachment 1: ETMX_z2_exp_300000_650.tiff
|
Attachment 2: ETMX_z2_exp_300000_750.tiff
|
Attachment 3: ETMX_z2_exp_30000_650.tiff
|
Attachment 4: ETMX_z2_exp_30000_750.tiff
|
381
|
Fri Mar 14 15:52:07 2008 |
rob | Configuration | LSC | LSC code change |
I've edited the LSC code to send different signals to the ASS box. Now, instead of the previously selected error signals deemed to be acceptable for the Alignment Sensing and Stabalization system, it sends the LSC control signals for each suspension to the ASS box (in its new incarnation as the Adaptive Susurration Subtraction system). These are the signals after the output matrix, and also after the LSC-[SUS] filter modules. |
383
|
Sun Mar 16 17:03:32 2008 |
rob | Configuration | CDS | ASS code change |
I've updated the ass.mdl file in the directory:
/cvs/cds/caltech/users/alex/cds/advLigo/src/epics/simLink/
to get us started in the adaptive PEM noise subtraction.
After several iterations of remote help from Alex, the code compiles and runs, receives signals from the LSC, PEM, and MC2, and communicates with the suspension controllers. I've also adapted the .par file from the code generator, but haven't got the testpoints working with the new ASS code. There are no MEDM screens yet, and Matt's adaptive filter code has not been installed (there's a matrix as a placeholder).
Putting in the adaptive code should be simple, building the MEDM screens tedious, and getting the testpoints working uncertain. I noticed that the new testpoint.par file starts at a different channel number than the previous (working) version, which is strange. I probably have a script somewhere to change all these numbers by a constant offset, but I don't know if that's the actual problem--maybe stuff just needs to be rebooted.
The code receives as input the first 24 channels from the PEM ADCU, the eight suspension control signals from the LSC, and the output of the MCL filter from MC2. It outputs to the MCL filter input of each suspension (except MC2). |
384
|
Mon Mar 17 18:30:48 2008 |
mevans | Configuration | PEM | Adaptive Filtering |
It seems that adaptive filtering can achieve results similar to those of the MISO FIR Wiener (entry 369). The adaptive code simulates real-time operation, but uses the same data used by Rana for the Wiener filter. I ran the adaptive filter over the data 100 times to ensure that it was well trained... maybe too well. |
Attachment 1: mcacc_adaptive.png
|
|
386
|
Thu Mar 20 16:06:27 2008 |
rob | Configuration | LSC | LSC code change |
I changed the LSC code again. I noticed that when turning off the LSC (e.g., going from LA to OFF), the cpu time would jump from ~50 to ~80, and irrevocably de-sync all the SUS controllers. This was because turning off the LSC would suddenly zero the inputs to the decimation filters that send information to the ASS box, which for some reason greatly increases the computation time of the iir filter function call. I changed the code so that these inputs are never zeroed. The ASS receives inputs from the LSC all the time now.
I also noticed that the ASS machine was running in ~2400 usec. Yes, 2,400 microseconds. I don't know how long it's been doing that, but I restarted it. Immediately after restart, it ran at 1700 microseconds. After using the "RESET" field in the adaptOnline code, that dropped to ~100 usec. Now it's not doing any adaptive filtering, as I don't know what the good settings are and no-one has been elogging their IFO work the last few days. |
390
|
Fri Mar 21 17:01:21 2008 |
rana | Configuration | DMF | Locale change on Mafalda & seisBLRMS restart |
Ever since we moved the accelerometers to be around the MC and changed their names, the seisBLRMS
has not been working. I tried to restart it today after fixing the channel names in the par file
but I ran into a PERL / UBUNTU bug.
perl: warning: Setting locale failed.
perl: warning: Please check that your locale settings:
LANGUAGE = (unset),
LC_ALL = (unset),
LC_TIME = "en_US.ISO8859-1",
LC_MONETARY = "en_US.ISO8859-1",
LC_CTYPE = "en_US.ISO8859-1",
LC_COLLATE = "en_US.ISO8859-1",
LC_MESSAGES = "C",
LC_NUMERIC = "en_US.ISO8859-1",
LANG = "en_US.UTF-8"
are supported and installed on your system.
perl: warning: Falling back to the standard locale ("C").
I don't know how this crept up or when it started. There were a bunch of fixes on the Ubuntu
forums which didn't work.
In the end I just set the 'unset' environment variables via our cshrc.40m file and this seemed
to make ligotools/perl happy. Lets hope this lasts...I love Linux. |