  15189   Wed Feb 5 21:04:10 2020 YehonathanUpdatePSLBench testing of PSL ai channels

{Yehonathan, Jon}

We retested the failed ai channels. Most of them got fixed by applying the inverse calibration in the yaml file.

We still find some anomalous channels, mostly in the DB25 connector. Turns out, their limits were ill-defined in the EPICS database. Specifying the right limit fixed the issue.

We find one miswired channel (BNC4). We connected the BNC to the right channel on the Acromag unit which fixed the issue.

Overall all the ai channels were successfully bench-tested.

Quote:

I checked the failed channels against the EPICS database definitions and the yaml file inputted to IFOTest. The channels where the readings are something other than +10/0 V, but the high/low values do change, I think can be attributed to one of two things:

  • An incorrect gain and/or offset conversion parameter in the yaml file
  • The EPICS SMOO parameter (smoothing) is set to some long value

I fixed the channel gains/offsets in the master yaml file (PSL.yaml). I also disabled smoothing in the EPICS defintions of the new PSL channels for the purpose of testing. We can uncomment those lines after installing the new chassis if noise is a problem. Please go ahead and re-test the channels again.

Quote:

I marked in red the failed channels. Some of them are probably calibration issues, but the ones that show the same voltage for high and low are probably disconnected wires.

  15194   Thu Feb 6 21:54:13 2020 JonUpdatePSLc1psl bench testing complete

Today I engineered the last piece of the new c1psl system: the multi-bit binary output (mbbo) channels that control the MC servo board gains. These 6-bit channels have to be split across two 4-bit Acromag registers. To enforce synchronous switching, I adapted the latch.py script developed by Gautam to address this problem in c1iscaux. Analogously to the c1iscaux implementation, I scripted the code to automatically run as a systemd service which is launched by the main modbusIOC service. I tested this all using the DB37 LED test board and confirmed it to work.

This now completes the electronics bench testing.

There are still several DB37 connectors to be wired, which carry only spare channels for future use and are not interfaced with the EPICS IOC. Jordan and I discussed this today and he or Chub will complete it shortly. To allow time for the spare channel wiring to be completed (as well as for more locking progress before interruption), Gautam and I think Monday/Tuesday next week would be the earliest possible window to install the new system.

  15202   Mon Feb 10 10:07:20 2020 gautamUpdatePSLPMC re-locked

I found the PMC unlocked this morning. It was re-locked using the usual procedure. I feel like this has been happening more frequently in the last month than before. In the past, the cause seems to have been the PZT voltage drifting too close to one of the rails - however, in this case, it looks like an IMC unlock event is what triggered the PMC lockloss (admittedly the PZT voltage was somewhat close to the rail). It would be good if someone can re-connect the PMC Transmission photodiode, it was a useful diagnostic channel we had working fine before the ringdowns started.

I also tweaked the input pointing into the PMC and ran the WFS DC offset relief script.

Attachment 1: PMCunlock.png
PMCunlock.png
  15204   Mon Feb 10 15:54:47 2020 JordanUpdatePSLCompleted Acromag Wiring

All spare channels on the PSL acromag chassis are connected with ~12in of spare wiring for future use.

  15205   Mon Feb 10 15:55:46 2020 JordanUpdatePSLPMCTRANSPD

[Gautam, Jordan]

Gautam showed me how the PMCTRANSPD signal was reading zero, and he suspected it might have to do with the acromag wiring. Disconnected the acromag box underneath the PSL table and checked the ADC wiring. Side note: When benchtesting the c1psl acromag chassis there was excess noise in the AI channels, and grounding the minus pin of the ADC channel eliminates the noise.

So I grounded the (-) pins on the ADC1 (192.168.113.122), which PMCTRANSPD is connected to and that seemed to fix the problem. As of right now PMCTRANSPD is reading ~.75 V.

See attached pictures

gautam: While this fix seems to have worked, I wonder why this became necessary only in the last month. Note that the problem was a noisy readback on the PMC transmission PD, which also made the FSS_RMTEMP channel noisy, leading me to suspect some kind of ground loop issue.

Attachment 1: ADC1.jpg
ADC1.jpg
  15231   Thu Feb 27 17:50:36 2020 gautamUpdatePSLc1psl setup setup

[many people]

in prep for the install tomorrow, we did the following:

  • Install the c1psl Supermicro in the 1X2 rack (Attachment 1). To make room we removed the anti-image filter and mounted it on the OMC rack.
  • Set up a local workstation (monitor+mouse+keyboard) for the Supermicro so we can do some local testing (Attachment 2).
  • Clear up the immediate area around the 1X1/1X2 rack, setup a cart for the Acromag.
  • Make sure there are sufficient adaptor boards cables (DB37, DB15, DB9, DB25, ethernet) etc available at the cart.
  • Label cables, connect on Acromag chassis end (Attachment 3).
  • Keep some large (A3) printouts of the channel mapping handy by the cart.
  • made sure we have open fuse-able DIN rail connectors for +/-15 V DC and +/-24 V DC for the Acromag box (we are waiting on some thinner gauge cabling for the 24V supply, once that arrives, we will power the box from the Sorensens. For now, they are powered by bench supplies on the cart).
  • made sure c1psl1 (still this name for the Supermicro) is ssh-able.

Barring objections, tomorrow (Friday 28 Feb 2020) morning I will commence the switch (I still want to work on the IFO tonight).

Attachment 1: 20200227_173535.jpg
20200227_173535.jpg
Attachment 2: 20200227_173454_HDR.jpg
20200227_173454_HDR.jpg
Attachment 3: 20200227_172659.jpg
20200227_172659.jpg
  15234   Fri Feb 28 08:05:22 2020 gautamUpdatePSLc1psl setup setup

And so it begins.

Quote:

Barring objections, tomorrow (Friday 28 Feb 2020) morning I will commence the switch

  15235   Fri Feb 28 10:04:41 2020 gautamUpdatePSLc1psl setup setup

Summary:

There are several problems evident already.

  1. Several EPICS database entries were missing. WTF.
  2. After fixing the missing entries, the PMC could be locked. However, the IMC could not be locked.
  3. I think the FSS Interface card is not configured correctly.

For now, I've returned the old c1psl connections, the PMC and IMC are both locked. Need to do some debugging on the bench.

  15236   Fri Feb 28 19:37:18 2020 gautamUpdatePSLNew c1psl installed
  1. The new c1psl Acromag crate is now interfaced to the Eurocrate electronics in 1X1 (formerly VME c1psl) and 1X2 (formerly c1iool0).
  2. The PMC and IMC can be locked. We will investigate stability / duty cycle over the weekend.
  3. There were a few issues with the wiring - specifically, the worng kind of Acromag BIO unit (sourcing, whereas we want sinking) was used for the FSS board switches. Once Jordan fixed this issue, the IMC could be locked.
  4. I began to do the detailed tests of the IMC Servo card channels - there may be some issues with the boost stages, but I ran out of time yesterday, so tbc Monday.

On Monday, we will remove the old c1psl and c1iool0 machines from the electronics rack and install the Acromag crate in a more permanent way. We will also clean up some of the old cabling and cross connects, althoug the situation seems so complicated (some cross connects are also used by the rtcds c1ioo expansion chassis) that I am inclined not to remove any cables.

The area around 1X1/1X2 has a lot of dangling cables and general detritus. Be careful if you are walking around there. We will clean up on monday.

  15247   Wed Mar 4 11:16:37 2020 gautamUpdatePSLPMC realignment

I realigned the input pointing into the PMC this morning. Usually, the way I do this is to minimize any discernible mode structure in the PMC reflection CCD image. Today, I noticed that making the DC reflection go down also makes the DC transmission go down. Possibilities:

  • we are sampling slightly different spots inside the PMC cavity which change the buildup by ~2-3%.
  • we are misaligned on the transmission/reflection photodiode.
  • ??
Attachment 1: PMCrealignment.png
PMCrealignment.png
  15253   Wed Mar 4 22:38:31 2020 JonUpdatePSLc1psl communications problem resolved

I investigated the problem reported earlier today with the BIO1 channels. By logging the systemd messages generated when the IOC starts, I was immediately able to determine that the problem was not limited to BIO1. The modbus communications were failing for several other units as well.

Because some in-situ rewiring of a handful of channels had recently been done (more on this soon), I initially suspected that one of the Acromags had been damaged in the process. However, removing BIO1 (or other non-communicating modules) did not restore communications with the rest of the modules. To test whether the chassis was the source of the problem at all, we set up a fresh ADC (new out of the package) and directly connected it to the secondary Ethernet interface of c1psl. With only the one new ADC connected, the modbus IOC failed in exactly the same way.

To confirm that the new ADC did in fact work, we connected it to c1auxex in the same configuration. The unit worked fine connected to c1auxex. This established that the source of the problem was the c1psl host. After some extensive debugging, I traced the problem to a pre-execution script (part of the modbus IOC systemd service) which resets the secondary network interface (the one connected to the Acromag chassis) prior to launching the IOC. This was to ensure the secondary interface always had the correct IP address. It appears this reset was somehow creating a race condition that allowed the modbus initializations (first communications with the Acromags) to sometimes start before the network interface had actually come back up.

I still don't understand how this was happening, or why the pre script worked just fine up until yesterday, but eliminating the network interface reset fixes the problem in 100% of the trials we ran. Unfortunately we lost the entire day to debugging this problem, so the final round of testing is still to be completed. We plan to pick it back up tomorrow afternoon.

  15256   Thu Mar 5 19:45:23 2020 JonSummaryPSLC1PSL in-situ test results

We've completed almost all of the in-situ testing of the c1psl channels. During this process, we identified several channels which needed to be rewired to different Acromags (BIO sinking v. sourcing). We also elected to change the connector type of a few channels for practical advantages. Those modifications and other issues found during testing are detailed below. Also attached are the updated channel assignments, with a column indicating the in-situ testing status of each channel.

Post-installation modifications

  • All four channels connected to the sourcing BIO module were found to in fact require sinking I/O. They were reassigned to sinking BIO modules. Affected channels:
    • C1:PSL-FSS_FASTSWEEP
    • C1:PSL-FSS_SW1
    • C1:PSL-FSS_SW2
    • C1:PSL-PSL_Shutter
  • Added a new AI channel:
    • C1:PSL-FSS_MIXERM
  • Removed an unneccessary AI channel:
    • C1:PSL-FSS_LODET
  • Moved two AI channels from BNC connectors to a new Dsub connector (labelled DB25M-2 in the spreadsheet).
    • C1:PSL-FSS_RCTEMP
    • C1:PSL-FSS_RMTEMP_VOLTS

Issues identified during testing

  • Digital calibration. The following channels work, but we need to verify their EPICS calibration parameters (EGUF/EGUL):
    • C1:PSL-FSS_FASTGAIN
    • C1:PSL-FSS_FAST
    • C1:PSL-PMC_RFADJ
    • C1:PSL-PMC_MODET
  • IMC servo board. The Acromag channels themselves were found to work, but the linearity of the mbbo gain stages are in question (i.e., a potential problem with the board). GV is currently testing the servo board.
  • PSL QPD board apears to be dead. We connected a scope directly to the test points on the board and measured a high level of noise and no signal (for all four of the QPD channels). I understand this QPD has not been used in some time, so it may not have been noticed before.
  • WFS DC channels are saturating when the IMC is unlocked. The acceptance range of the Acromag ADC is only +/-10 V, but we measured sensor voltages as high as ~14 V. It appears that the old ADCs were somehow accepting a range of 0 to +20 V instead of -10 to +10 V. However, the Acromags do not support the input range 0-20 V. Since SNR is not critical for these channels (they're used only for initial alignment), I propose we simply install a voltage divider inside the chassis, just before the Acromag, for each of these signals.
Attachment 1: c1psl_feedthrough_wiring_-_By_Connector_(3).pdf
c1psl_feedthrough_wiring_-_By_Connector_(3).pdf c1psl_feedthrough_wiring_-_By_Connector_(3).pdf c1psl_feedthrough_wiring_-_By_Connector_(3).pdf c1psl_feedthrough_wiring_-_By_Connector_(3).pdf
  15266   Wed Mar 11 18:12:53 2020 gautamSummaryPSLWFS Demod board modifications

[koji, gautam]

Attachment #1 shows the relevant parts of the schematic of the WFS demod board (not whitening board). 

  • The basic problem was that the switchable gain channels were not accounted for in the Acromag channel list 😒.
  • What this meant was that the DC gain was set to the default x100 (since the two DG211s that provide the switchable x10 and x1 gain options had their control logic pins pulled up to +5V because these pins weren't connected to any sinking BIO channel).
  • Rather than set up new connections to Acromags inside the chassis (though we have plenty of spares), Koji and I decided to make these fixed to x1 gain.
  • The actual fix was implemented as shown in the annotated schematic. There are some pictures 📷 of the PCB in the DCC entry linked above.
  • Amusingly, this board will now require a sourcing BIO unit if we want to still have the capability of switching gains.

Before removing the boards from the eurocrate: 

  • I dialled down the Kepco HV supplies
  • disconnected all the cabling to these boards after noting down cable numbers etc.

After Koji effected the fix, the boards were re-installed, HV supplies were dialled back up to nominal voltage/currents, and the PMC/IMC were re-locked. The WFS DC channels now no longer saturate even when the IMC is unlocked 👏 👏 . I leave it to Yehonathan / Jon to calibrate these EPICS channels into physical units of mW of power. We should also fix the MEDM screen and remove the un-necessary EPICS channels.

Later in the evening, I took advantage of the non-saturated readbacks to center the beams better on the WFS heads. Then, with the WFS servos disabled, I manually aligned the IMC mirrors till REFLDC was minimized. Then I centered the beam on the MC2 transmission QPD (looking at individual quadrants), and set the WFS1/2 RF offsets and MC2 Trans QPD offsets in this condition.

Quote:

WFS DC channels are saturating when the IMC is unlocked.

Attachment 1: D980233-B_Mar2020Mods.pdf
D980233-B_Mar2020Mods.pdf
  15272   Thu Mar 12 16:13:16 2020 gautamUpdatePSLTemperature sensors connected to Acromag

[jordan, gautam]

the long DB25 cable to connect the Acromag chassis to the temperature sensor interface box arrived. We laid it out today. This cable does the following:

  1. Supply the box with +/- 24 V DC from the chassis. The pin arrangement here is rather unfortunate (the +/-24 V DC and GND pins are in close proximity), so if you're not careful, you'll create a short as you plug this cable in (as we found out today). So the best practise is to power down the crate before connecting/disconnecting this cable. Jordan will label this accordingly tomorrow.
  2. Pipe the "TAMB" and "TCAV" signals, corresponding to temperature sensors mounted to the PSL table-top and reference cavity exterior respectively, to the Acromags. We found during some initial testing that the "TAMB" signal was reaching the DB25 connector on the Acromag chassis, but wasn't going to any ADC channel - this was rectified.

Both signals now show up in the EPICS channels, but are noisy - I suspect this is because the return pin of the Acromag is not shorted to ground (this is a problem I've seen on the bench before). We will rectify this tomorrow as well.

We took this opportunity to remvoe the bench supply and temporary Acromag crate (formerly known as c1psl2) from under the PSL table. While trying to find some space to store the bench supply, we came across a damaged Oscilloscope in the second "Electronics" cabinet along the Y-arm, see Attachment #1. 

After this work, I found that the IMC autolocker was reliably failing to run the mcup script at the stage where the FSS gains are ramped up to their final values. I was, however, able to smoothly transition to the low-noise locked state if I was manipulating the EPICS sliders by hand. So I added an extra 2 seconds of sleep time between the increasing of the VCO gain to the final value and the ramping of the FSS gains in the mcup script (where previously there was none). Now the autolocker performs reliably.

Attachment 1: P_20200317_153736_vHDR_On_2.jpg
P_20200317_153736_vHDR_On_2.jpg
  15275   Fri Mar 13 14:28:39 2020 gautamUpdatePSLAdded tees to PMC Trans and REFL

I want to monitor the PMC TRANS and REFL levels on the PSL table - previously there were some cables going to the oscilloscope on the shelf but someone had removed these. I re-installed them just now. While there, I disconnected the drive to the AOM - there must've been some DC signal going to it because when I removed the cable, the PMC and IMC transmission were recovered to their nominal levels.

  15310   Wed Apr 22 17:29:14 2020 gautamUpdatePSLFSS debugging attempts

Summary:

On Monday, I hooked up an AG4395 to the PMC error point (using the active probe). The idea was to take a spectrum of the PMC error point every time the FSS PC drive RMS channel indicated an excursion from the nominal value. An initial look at the results don't suggest that this technique is particularly informative. I'll have to think more about a workaround, but please share your ideas/thoughts if you have some.

Also, the feature in the spectrum at ~110 kHz makes me suspect some kind of loop instability. I'll measure the IMC loop OLG at the next opportunity.

Details:

  • The PMC servo bandwidth is ~2 kHz, so above this, the PMC error point should be a faithful monitor of the PSL frequency noise, provided the sensing noise is low enough.
  • The PMC error point sensing noise is ~100nV/rtHz (I'm monitoring this straight after the Minicircuits mixer+bandpass filter that we are using as a demodulator). This corresponds to ~2 Hz/rtHz, using the ~10 MHz/V PDH discriminant calibration from January. Seems consistent with this elog.
  • I was hoping to see if there was a particular frequency band in which the noise gets elevated, and if the crossover frequency is a few kHz and the IMC servo BW is ~110 kHz, I would have expected this to be in the 10-100 kHz region. Possibly my frequency resolution isn't good enough? But with the Agilent, doing a finer grid would mean a longer measurement time, in which case the IMC might lose lock before the measurement is done.
  • But, as shown in Attachment #1, there isn't any clear evidence, from the ~20 excursions that were recorded last night. The color of the line is meant to be indicative of the average value of the PC drive RMS channel in the measurement time.
  • A significant bottleneck in this whole process is that it takes ~1 minute to initiate the GPIB measurement, and download the data. The pseudo-code I used is:
    • While the IMC is locked, watch PCdrive RMS EPICS channel's "ALARM" state, which becomes non-zero when the PCdrive RMS exceeds 1 V (this is how it is defined in the EPICS db record right now).
    • Make sure this isn't a transient feature - I do this by waiting 5 seconds and checking that the ALARM flag is still flagged.
    • Initiate a AG4395 measurement over GPIB - I use the measurement span of 1 kHz - 1 MHz with a BW/span ratio of 0.1%, 5 averages.
    • Check that the IMC is still locked (if it got unlocked while the measurement was made, presumably the measurement is garbage).
  • Is there a better monitor of the laser frequency noise? I can imagine using POX/POY which I think have a lower electronics noise floor but I'm not sure if that's true at 100 kHz and having the arms locked in addition to the IMC seems more complicated...
  • Since we are planning a laser upgrade, is this worth spending more time on? I may leave the measurement running on pianosa in a tmux session while I'm not in the lab...
Attachment 1: FSSspec.pdf
FSSspec.pdf
  15312   Thu Apr 23 10:42:02 2020 ranaUpdatePSLFSS debugging attempts

I had set up the 4395 to do this automatically a few years ago, but it looked at the FSS/IMC instead. When the PCDRIVE goes high there is this excess around ~500 kHz in a broad hump.

But the IMC loop gain changes sometimes with alignment, so I don't know if its a loop instability or if its laser noise. However, I think we have observed PCDRIVE to go up without IMC power dropping so my guess is that it was true laser noise.

This works since the IMC is much more sensitive than PMC. Perhaps one way to diagnose would be to lock IMC at a low UGF without any boosts. Then the UGF would be far away from that noise making frequency. However, the PCDRIVE also wouldn't have much activity.

  15570   Tue Sep 15 10:57:30 2020 gautamUpdatePSLPMC re-locked, RGA re-enabled

The PMC has been unlocked since September 11 sometime (summary pages are flaky again). I re-locked it just now. I didn't mess with the HEPA settings for now as I'm not using the IFO at the moment, so everything should be running in the configuration reported here. The particulate count numbers (both 0.3um and 0.5um) reported is ~x5-8 of what was reported on Thursday, September 10, after the HEPA filters were turned on. We don't have logging of this information in any automated way so it's hard to correlate things with the conditions in Pasadena. We also don't have a working gauge of the pressure of the vacuum envelope.

The RGA scanning was NOT enabled for whatever reason after the vacuum work. I re-enabled it, and opened VM1 to expose the RGA to the main volume. The unit may still be warming up but this initial scan doesn't look completely crazy compared to the reference trace which is supposedly from a normal time based on my elog scanning (the timestamp is inherited from the c0rga machine whose clock is a bit off).


Update 1500: I checked the particle count on the PSL table and it barely registers on the unit (between 0-20 between multiple trials) so I don't know if we need a better particle coutner or if there is negligible danger of damage to optics due to particulate matter.

Attachment 1: RGAscan.pdf
RGAscan.pdf
  15592   Mon Sep 21 16:40:52 2020 gautamUpdatePSLHEPAs are no longer running

The HEPA filters on the PSL enclosure are no longer running. I tried cycling the power switch on the NW corner of the enclosure, and also turned the dial on the Variac down to zero and back up to maximum, no effect. Judging by the indicator LED on it, the power is making it to the Variac - bypassing the Variac and directly connecting the HEPA to the AC power seems to have no effect either.

I can't be sure, but I'm almost certain I heard them running at max speed an hour ago while I was walking around inside the VEA. Probably any damage that can happen has already been done, but I dialled down the Innolight injection current, and closed its shutter till the issue can be resolved.

  15662   Fri Nov 6 14:08:44 2020 gautamUpdatePSLPMC re-locked

The PMC servo railed and so I re-locked it at ~half range. I've been noticing that the diurnal drift of the PZT control voltage has been larger than usual - not sure if it's entirely correlated with temperature on the PSL table. Anyway the cavity is locked again so all is good.

  16016   Mon Apr 12 08:32:54 2021 Anchal, PacoSummaryPSLPMC unlocked at 2pm on Sunday; ~ Restored

PMC lost lock between 21:00 and 22:00 UTC on April 11th as seen in the summary pages:

https://nodus.ligo.caltech.edu:30889/detcharsummary/day/20210411/psl/#gallery-4

That's between 2pm and 3pm on Sunday for us. We're not sure what caused it. We will attempt to lock it back.


Mon Apr 12 08:45:53 2021: we used milind's python script in scripts/PSL/PMC/pmc_autolocker.py. It locked the PMC in about a minute and then IMC catched lock succefully.

However, the PMC transmission PD shows voltage level of about 0.7V. On medm, it is set to turn red below 0.7 and yellow above. In Summary pages in the past, it seems like this value has typically been around 0.74V. Simil;arly, the reflection RFPD DC voltage is around 0.063 V right now while it is supposed to be around 0.04 nominally So the lock is not so healthy.

We tried running this script and the bashscript version too (scripts/PSL/PMC/PMCAutolocker) a couple of times but it was unable to acquire lock.

Then we manually tried to acquire lock by varying the C1:PSL-PMC_RAMP (with gain set to -10 dB) and resetting PZT position by toggling Blank. After a few attempts, we were able to find the lock with transmission PD value around 0.73V and reflection RFPD value around 0.043. PZT control voltage was 30V and shown in red in medm to begin with. So we adjusted the output ramp again to let it come to above 50V (or maybe it just drifted to that value by itself as we could se some slow drift too). At Mon Apr 12 09:50:12 2021 , the PZT voltage was around 58V and shown in green.

We assume this is a good enough point for PMC lock and move on.

  16019   Mon Apr 12 18:34:26 2021 YehonathanSummaryPSLPMC unlocked at 2pm on Sunday; ~ Restored

PMC lost lock again at around 16:00 April 12. I was able to lock it again but the transmission is only 0.6 now and REFL is 0.14.

Rana came in and realigned the PMC stirring mirrors. Now the transmission is 0.757V, and the REFL is 0.03V.

I noticed that the PZT was around 250V. Given that the PMC got unlocked at 16:00, which is around the peak temperature time in the lab (lagging behind the outside weather), due to the PZT voltage going down to 0V, I figured that the PZT voltage would go up during the night when the lab gets cold and therefore will likely go out of range again.

I found a different working point at 150V and relocked the PMC.

 

  16023   Tue Apr 13 19:24:45 2021 gautamUpdatePSLHigh power operations

We (rana, yehonathan and i) briefly talked about having high power going into the IFO. I worked on some calcs a couple of years ago, that are summarized here. There is some discussion in the linked page about how much power we even need. In summary, if we can have

  • T_PMC ~85% which is what I measured it to be back in 2019
  • T_IMC * T_inputFaraday ~60% which is what I estimate it to be now
  • 98% mode matching into the IMC
  • power recycling gain of 40-45 once we improve the folding mirror situation in the recycling cavities
  • and a gain of 270-280 in the arm cavities (20-30ppm round trip loss)

then we can have an overall gain of ~2400 from laser to each arm cavity (since the BS divides the power equally between the two arms). The easiest place to get some improvement is to improve T_IMC * T_inputFaraday. If we can get that up to ~90%, then we can have an overall gain of ~4000, which is I think the limit of what is possible with what we have.

We also talked about the EOM. At the same time, I had also looked into the damage threshold as well as clipping losses associated with the finite aperture of our EOM, which is a NewFocus 4064 (KTP is the Pockel medium). The results are summarized in Attachments #1 and #2 respectively. Rana thinks the EOM can handle factor of ~3 greater power than the rated damage threshold of 20W/mm^2.

Attachment 1: intensityDist.pdf
intensityDist.pdf
Attachment 2: clippingLoss.pdf
clippingLoss.pdf
  16024   Tue Apr 13 20:45:16 2021 YehonathanUpdatePSLLaser amplifier

{Yehonathan, Rana}

We unpacked the content of the amplifier crate in front of the water fountain (see attachments). Inside we found:

1. Amplifier head. (attachment 1)
2. Amplifier electronics and pump diodes (attachment 2).
3. Optical fiber (attachment 3).
4. 2 Long water hoses (~2m) and 2 short ones.
5. Network cable.
6. A wooden plate.
7. Cable sleeve (attachment 2)?
8. Some manuals will be uploaded to the wiki soon.

Please don't move/touch any of that stuff

Things that we need to consider/obtain:
1. A suitable power cable (attachment 4) with suitable power ratings (800W according to the amplifier specs). The connector head is C19 I believe.
2. A chiller. Rana says Aidan knows where to find one. Should we chill the amplifier head as well?
3. A mounting plate for the amplifier head with good thermal conductivity.
4. The pump wavelength is 808nm, we need to get suitable safety goggles.
5. Where to put the big electronics box. Preferably on 1X1 or 1X2.
6. How do we arrange the different components on the table? We also need to mode match the beam into the amplifier.

 

Attachment 1: 20210413_204721.jpg
20210413_204721.jpg
Attachment 2: 20210413_203300.jpg
20210413_203300.jpg
Attachment 3: 20210413_204940.jpg
20210413_204940.jpg
Attachment 4: 20210413_205549.jpg
20210413_205549.jpg
  16032   Wed Apr 14 19:48:18 2021 gautamUpdatePSLLaser amplifier

A couple of years ago, I got some info about the amplifier setup at the sites from Terra - sharing here in case there is some useful info in there (our setup will be rather different, but it looked to me like our Amp is a 2017 vintage and it may be that the performance is not the same as reported in the 2019 paper).

collection of docs (table layout in 'Proposed....setup') : https://dcc.ligo.org/LIGO-T1700046

LVC 70W presentation: https://dcc.ligo.org/LIGO-G1800538 

I guess we should double check that the beam size everywhere (in vacuum and on the PSL table) is such that we don't exceed any damage thresholds for the mirrors used. 

  16034   Thu Apr 15 09:46:24 2021 YehonathanUpdatePSLLaser amplifier

Some more relevant documents provided by Matt:

Phase III:70W amplifier integration at LIGO

70W amplifier External Shutter

aLIGO PSL high power attenuator

 

  16044   Fri Apr 16 18:21:36 2021 YehonathanUpdatePSLLaser amplifier

I surveyed a bit the 1X1/2 area to plan for the installation of the laser amplifier.

There is a vacancy at the bottom of 1X2 (attachment 1). I measured the dimensions of the diode box (DB) and it should fit. The optical fiber bundle is 75m long and should reach the amplifier head on the table easily.

According to the specs, the maximum power consumption of the DB is 800W (typically 600W), it should probably have its own circuit breaker. It can easily draw more than a few amps. The rack power strips are connected to this 4 socket box (attachment 2), is this just another power strip? It is connected to a circuit breaker with a 30A rating. How do we proceed from here?

In any case, we will need at least 2 meters of power cable.

I also tried to find a suitable place for a water chiller. A few suggestions are in the attachments. Basically either between the electronics shelves and the small rack next to 1X2 or next to the small rack close to the optical table. Maybe put it where the ladder sits and find another place for the ladder. Other options?

We would also need a windows machine running the Beckhoff software. The idea is that all the different laser components (DB, chillers, interlocks, switches) are connected to the EtherCat (over the ethernet infrastructure) so that the Beckhoff code can recognize a failure and switch off everything.

The things that are monitored:

1. Is the NPRO on?

2. Is the flow rate from the chillers enough?

3. Is the temperature of the diodes in the normal range?

4. Is one of the interlocks open?

5. Was one of the emergency buttons pushed?

6. Was the key switch on the DB turned to OFF?

The DB is EtherCat ready but the rest of the signals need to be interfaced somehow. Do we have to buy these EtherCAT terminals?

 

 

Attachment 1: 20210416_143642.jpg
20210416_143642.jpg
Attachment 2: 20210416_145408.jpg
20210416_145408.jpg
Attachment 3: 20210416_145448.jpg
20210416_145448.jpg
Attachment 4: 20210416_181324.jpg
20210416_181324.jpg
  16046   Sun Apr 18 21:29:55 2021 ranaUpdatePSLLaser amplifier
  • Ideally, we put the chiller outside of the interferometer area. The PSL chiller used to be in the control room near the door by IMC REFL. We could also put it in the drill press room.
  • Once we figure out a couple of places where the Diode Box can go, we can ask facilities to make the appropriate power connections. They will have to eval the situation to figure out if the main power to the lab needs to be shut down.
  • Can we put the laser diode box in the drill press room too? Then the hoses can be short. Perhaps less EMI getting into our sensitive places.
  16056   Wed Apr 21 00:08:15 2021 KojiUpdatePSLPSL Table (sort of) covered / HEPA "chimney"

Shutdown Procedure:
PSL Shutter closed / MC Autolocker disabled / PSL mechanical shutter closed / Laser injection current turned to zero / Laser turn off (red button) / Laser key turned off

The laser stat before the shutdown:
- LD Temp A: Set 22.07 (Untouched)
- LD Temp B: Set 21.03(Untouched)
- Laser Injection Current: Dial 9.53, Actual 2.100 -> Dial was moved to zero upon shutting down
- Laser Crystal Temp: Dial 3.34 (untouched)  Set 30.57 Actual 30.60 (Untouched)

PSL Table covering

- Because of the so many cables going up and down, sealing the PSL table with the metalized sheet was not easy. Therefore, the sheets have been just softly laid above the optics. (Attachment 1)
- The largest sheet which covers the east half of the table was taped to the table at the bottom, so that the air from the chimneys (see below) does not come up to the table

- The large dust could come from the opening of the enclosure during the filter replacement. So it was considered to be easier to seal the openings. (Attachment 2)
- Of course, the HEPAs are going to be tested after the maintenance work. It means that vent paths were needed so that the seals do not explode with the pressure (together with dust).
- Thus, the tubes of the sheets are attached to the seals to form "chimneys" for guiding the airflow beneath the table. (Attachment 2/3/4)
- This configuration was not meant to be sufficiently strong for a continuous run of the fans. Long running of the HEPAs may cause the failure of the seal tapes.
  Therefore the HEPA test should be done with a low flow rate and/or a short period of high flow.

- Once the work has been done, all the sheets should be carefully removed without scattering the fallouts onto the optics.

 

Attachment 1: 20210420235324_IMG_0560.jpeg
20210420235324_IMG_0560.jpeg
Attachment 2: 20210420235304_IMG_0559.jpeg
20210420235304_IMG_0559.jpeg
Attachment 3: 20210420235243_IMG_0558.jpeg
20210420235243_IMG_0558.jpeg
Attachment 4: 20210420235344_IMG_0562.jpeg
20210420235344_IMG_0562.jpeg
  16057   Wed Apr 21 01:14:03 2021 KojiUpdatePSLPSL Table (sort of) covered / HEPA "chimney"

I also located the (possible) HEPA filters in the lab. (Attachments 1~3)

Oh! This is NO-NO! We can't place anything in front of the mains breakers. (Attachment 2)
I relocated the objects (Attachment 3)

 

Attachment 1: P_20210421_005056.jpg
P_20210421_005056.jpg
Attachment 2: P_20210421_005114.jpg
P_20210421_005114.jpg
Attachment 3: P_20210421_005302.jpg
P_20210421_005302.jpg
  16062   Wed Apr 21 11:09:57 2021 yehonathanUpdatePSLLaser amplifier

I went to the TCS lab to take a look at the chillers lying around. I spotted two chillers:

1. Thermoflex1400 (attachment 1,2). Spec sheet.

2. Polyscience Recirculator 6000 series (attachment 3,4). Manual.

The Thermoflex has various communication ports. The Recirculator doesn't have any communication ports, but it is connected to a flow meter with what seems to be an electronic readout (attachment 5). Manual.

Both chillers have similar capacity ~ 4 gallons/minute. Thermoflex has 2 times more reservoir capacity than the Recirculator.

None of them seem to be Bechkoff-ready.

I guess we can have interlock code handling mixed signals Beckhoff+Non beckhoffs?

Attachment 1: 20210420_171606.jpg
20210420_171606.jpg
Attachment 2: 20210420_171621.jpg
20210420_171621.jpg
Attachment 3: 20210420_171611.jpg
20210420_171611.jpg
Attachment 4: 20210420_171629.jpg
20210420_171629.jpg
Attachment 5: 20210420_171702.jpg
20210420_171702.jpg
  16068   Wed Apr 21 19:28:03 2021 AnchalUpdatePSLPSL/IFO recovery

[Anchal, Koji]

Removed the top sheet

  • Opened first from the door side so that any dust would spill outside.
  • Then rolled the sheet inward to meet in the middle.
  • Repeated this twice for the 2 HEPA filters.

Removed the sheets on the table

  • Lifted sheet up making sure the top side face outside always.
  • Rolled it sideways halfway through.
  • Cut down the sheet vertically.
  • Slided the doors to the other side and rolled the remaining half.
  • On the door side, the sheets above the ALS optics were simply lifted off.

Restarting PSL

  • Turned on the HEPAs at the max speed
  • Switched on laser to jsut above the threshold
    • Before the 1st eom, power was 20mW 
    • After the EOM/AOM, 18mW. So about 90% transmission through all polarizing optics.
    • We saw the resonances of the PMC but could not lock it even with highest gain available (30 dB).
  • Increased the input power to PMC to 100mW
    • Locked the PMC at 30dB gain
    • The transmitted power was ~50-60 mW. (Had to use power meter suspended by hand only.
    • The right before the IMC (after the 2nd EOM) 48mW. So none of the alignment was lost.
  • Opened the PSL shutter.
  • We were able to see IMC reflection signal.
  • We were also able to see IMC catching lock as the servo was left ON earlier.
  • Switched off the servo.
  • Decided to increase the power while watching PMC Trans/Refl and IMC REFL
  • Injection diode current to innolight was increased slowly to 2.10A. Saw a mod hopping region aroun 1.8A.
  • We recovered the PMC Trans >0.7 V.
  • PZT was near the edge, so moved by one FSR.
  • The PMC refelction signal is still shown in red at around 48 mV.

Back to control room

  • IMC was locked almost immediately by manually finding the lock while keeping IMC WFS off to preserve the offsets from yesterday.
  • Then switch on IMC WFS. Working good.
  • Then unlocked the servo and switched on IMC Autolocker. Lock was caught immediately.

Decided to start locking the arms

  • The arm transmissions were flashing but at 0.2~0.3 level.
  • Decided to adjust TT1 and TT2 Pitch and Yaw to align the light going into the arms.
  • This made TRY ~0.6 / TRX ~0.8 at the peak of the flashing
  • Locked the arms. (By switching on C1:LSC-MODE_SELECT which engages all servos).
  • Used ASS to align Yarm then align Xarm. Procedure:
    • Sitemap > ASC > c1ass
    • Open striptool to look at progress. ! Scripts YARM > striptool.
    • Switch on ASS. ! More Scripts > ON
    • Wait for the TRY to reach to around 0.97.
    • Freeze the outputs. ! Scripts > Freeze Outputs.
    • Offload the offsets to preserve the output. ! More Scripts > OFFLOAD OFFSETS.
    • Switch off ASS. ! More Scripts > OFF
    • Repeted this for XARM.
  • At the end, both XARM and YARM were locked with TRX ~ 0.97 and TRY ~ 0.96.
  16069   Wed Apr 21 19:43:20 2021 KojiUpdatePSLNew HEPA speed control

The new HEPA speed controllers are attached at the middle of the HEPA unit (not at the edge of the unit)... (Attachment 1)
You still need a step./stool to touch the knob and need a ladder for a more precise setting.

We still don't know the optimal speed of the nominal IFO operation. For now, the HEPAs are running at the max speed (Attachment 2).
Once we know the optimal setting, we mark the knobs so that we can see them only with the step.

Attachment 1: P_20210421_193637.jpg
P_20210421_193637.jpg
Attachment 2: P_20210421_193627.jpg
P_20210421_193627.jpg
  16074   Thu Apr 22 14:41:55 2021 ChubUpdatePSLNew HEPA speed control

When adjusting the blower speed, give the blower at least 30 seconds to speed up or slow down to the set speed.  The flywheel effect of the big motor armature and blower mass requires time to follow the control current.  Note the taller Flanders HEPA filters.  These and the new intake filters should keep the PSL air clean for a long time!

Quote:

The new HEPA speed controllers are attached at the middle of the HEPA unit (not at the edge of the unit)... (Attachment 1)
You still need a step./stool to touch the knob and need a ladder for a more precise setting.

We still don't know the optimal speed of the nominal IFO operation. For now, the HEPAs are running at the max speed (Attachment 2).
Once we know the optimal setting, we mark the knobs so that we can see them only with the step.

 

Attachment 1: 40M_PSL_HEPA.jpg
40M_PSL_HEPA.jpg
  16076   Thu Apr 22 15:15:26 2021 gautamUpdatePSLPMC transmission

I was a bit surprised by these numbers suggesting the PMC transmission is only 50-60%. I went to the table today and confirmed that it is more like 85% (1.3 W in, 1.1 W transmitted, both numbers from with the FieldMate power meter), as I claimed in 2019. Even being conservative with the power meter errors, I think we can be confident T_PMC will be >80% (modulo any thermal effects with higher power degrading the MM). There isn't any reliable record of what the specs of the PMC mirrors are, but assuming the IO couplers have T=4000ppm and the end mirror has T=500ppm as per Alan's plot, this is consistent with a loss of something like 300ppm loss per mirror - seems very high given the small beam spots, but maybe these mirrors just aren't as high quality as the test masses?

It's kind of unfortunate that we will lose ~20% of the amplifier output through the first filter, but I don't see an easy way to clean these mirrors. It's also not clear to me if there is anything to be gained by attempting a cleaning - isn't the inside of the cavity supposed to be completely isolated from the outside? Maybe some epoxy vaporization events degraded the loss?

Quote:

The transmitted power was ~50-60 mW. (Had to use power meter suspended by hand only.

  16077   Thu Apr 22 15:34:54 2021 AnchalUpdatePSLPMC transmission

Koji mentioned that the mode of the laser is different for lower diode currents. So that might be the reason why we got less transmission at the low input power but more afterward.

  16080   Thu Apr 22 17:28:34 2021 YehonathanUpdatePSLLaser amplifier

According to the aLIGO 70W amplifier interlock concept the flow rate of the chiller should be between 5 and 40 l/min. The chillers I found in the TCS lab both have around 15 l/min flow rate so we should be fine in that regard.

Assuming that the power consumption of the diode box is ~800W and that the optical output power of the diode is ~ 300W of optical power, the chillers need to be able to remove the remaining power. At room temperature, they both have enough cooling capacity according to their specs.

As for the idea to put the chiller and diode box in the drill room: There are not a lot of options here. The only viable place is the SW corner (attachment 1). I was told this place is used sometimes for liquid nitrogen dewar. Alternatively, if possible, we can move the fire extinguishers to the SW corner and use the NW corner. In that way, we don't have to clear all the junk from the SW corner, as long as the extinguishers are still accessible.

I made a sketch (attachment 2) showing a possible setup for a diode box + chiller rack. The fiber and network cable can go through the hole in the wall that already exists for the N2. It will have to get bigger though (attachment 3). The rack would also need to host some Acromag unit to convert the communication channel of the chiller/flow meter to Ethernet. The Acromag on 1X7 has no spare channels.

The only power socket in the room, to which the drill is connected, is circuit #36 which is connected to panel L in the lab. The breaker's ampacity seems to be 20A if I'm reading the number on the breaker correctly.

 

Attachment 1: 20210422_124940.jpg
20210422_124940.jpg
Attachment 2: DrillRoomSchematics.pdf
DrillRoomSchematics.pdf
Attachment 3: 20210422_125240_1.png
20210422_125240_1.png
  16082   Fri Apr 23 18:00:02 2021 gautamUpdatePSLHEPA speed lowered

I will upload some plots later - but in summary, I set the HEPA speed to ~40%. I used (i)IMC transmission RIN, (ii) Arm cavity transmission RIN and (iii) ALS beat noise as 3 diagnostics, to see how noise in various frequency bands for these signals change as a function of the HEPA speed. The MC2T RIN shows elevated noise between 1-10Hz at even the lowest speed I tried, ~20% of the max on each blower. The elevated noise extended to ~50-70 Hz for HEPA speeds >40% of the maximum, and the arm cavity RIN and ALS signals also start to become noisy for speeds >60% of the maximum. So I think 40% is a fine speed to run at - for squeezing measurement we may have to turn off the HEPA for 10mins but for the usual single arm / PRMI / DRMI locking, this should be just fine. For the elevated ALS noise - I'm not sure if the coupling is happening over the top of the enclosure where the fiber bringing light from EX comes close to the HEPA filters, or if it is happening inside the PSL enclosure itself, near the beat mouth - but anyways, at the 40% speed, I don't see any effect on the ALS noise.

I checked with a particle counter at the SW corner of the PSL table (which is the furthest away we can be on the table from the HEPA blowers) after leaving the blowers on for ~30mins and it registered 0 for both 0.3um and 0.5um sized particles (if the blowers are off, the respective numbers are 43 and 9 but I forgot what the units were, and I believe they have to be multiplied by 10). 

I have not yet marked the speed control units yet in case there is some other HEPA science that needs to be done before deciding what is the correct setting. But I think I can get the PRFPMI lock without much issue with this lower speed, which is what I will try later today evening.

Attachment 1: HEPAdiag.pdf
HEPAdiag.pdf
  16083   Fri Apr 23 19:26:58 2021 KojiUpdatePSLHEPA speed lowered

I believe that there is an internal setting for the minimum flow, so the flow is not linear ("0%" is not zero), but we should mark this flow speed once you find this is sufficiently low for the locking too.

  16141   Fri May 14 17:45:05 2021 ranaUpdatePSLHEPA speed raised

The PSL was too hot, so I turned on the south HEPA on the PSL. The north one was on and the south one was off (or so slow as to be inaudible and no vibration, unlike the north one). Lets watch the trend over the weekend and see if the temperature comes down and if the PMC / WFS variations get less. Fri May 14 17:46:26 2021

  16142   Sat May 15 12:39:54 2021 gautamUpdatePSLNPRO tripped/switched off

The NPRO has been off since ~1AM this morning it looks like. Is this intentional? Can I turn it back on (or at least try to)? The interlock signal we are recording doesn't report getting tripped but I think this has been the case in the past too.


After getting the go ahead from Koji, I turned the NPRO back on, following the usual procedure of diode current ramping. PMC and IMC locked. Let's see if this was a one-off or something chronic.

Attachment 1: NPRO.png
NPRO.png
  16144   Tue May 18 00:52:38 2021 ranaUpdatePSLHEPA speed raised

Looks like the fan lowered the temperature as expected. Need to get a few more days of data to see if its stabilized, or if that's just a fluke.

The vertical line at 00:00 UTC May 18 is about when I turned the fans up/on.

Attachment 1: Untitled.png
Untitled.png
  16145   Tue May 18 20:26:11 2021 ranaUpdatePSLHEPA speed raised

Fluke. Temp fluctuations are as usual, but the overall temperature is still lower. We ought to put some temperature sensors at the X & Y ends to see what's happening there too.

  16400   Thu Oct 14 09:28:46 2021 YehonathanUpdatePSLPMC unlocked

PMC has been unlocked since ~ 2:30 AM. Seems like the PZT got saturated. I moved the DC output adjuster and the PMC locked immidiatly although with a low transmission of 0.62V (>0.7V is the usual case) and high REFL.

IMC locked immidiately but IFO seems to be completely misaligned. The beams on the AS monitor are moving quite alot syncronously. BS watchdog tripped. I enabled the coil outputs. Waiting for the RMS motion to relax...

Its not relaxing. RMS motion is still high. I disabled the coils again and reenabled them. This seems to have worked. Arms were locked quite easily but the ETMs oplevs were way off and the ASS couldn't get the TRX and TRY more than 0.7. I align the ETMs to center the oplev. I realign everything else and lock the arms. Maximium TR is still < 0.8.

 

 

  16401   Thu Oct 14 11:25:49 2021 YehonathanUpdatePSLPMC unlocked

{Yehonathan, Anchal}

I went to get a sandwich around 10:20 AM and when I came back BS was moving like crazy. We shutdown the watchdog.

We look at the spectra of the OSEMs (attachment 1). Clearly, the UR sensing is bad.

We took the BS sattelite box out. Anchal opened the box and nothing seemed wrong visually. We returned the box and connected it to the fake OSEM box. The sensor spectra seemed normal.

We connected the box to the vacuum chamber and the spectra is still normal (attachment 2).

We turn on the coils and the motion got damped very quickly (RMS <0.5mV).

Either the problem was solved by disconnecting and connecting the cables or it will come back to haunt us.

 

 

 

Attachment 1: BS_OSEM_Sensor_PSD.pdf
BS_OSEM_Sensor_PSD.pdf
Attachment 2: BS_OSEM_Sensor_PSD_AfterReconnectingCables.pdf
BS_OSEM_Sensor_PSD_AfterReconnectingCables.pdf
  113   Fri Nov 16 18:46:49 2007 steveBureaucracyPSL MOPA was turned off & on
The "Mohana" boys scouts and their parents visited the 40m lab today.
The laser was turned off for their safety.
It is back on !
  2951   Wed May 19 14:36:46 2010 AidanHowToPhase CameraPhase Camera algorithm and stuff

 I had a think about the algorithm we might use for the phase camera measurement. MATLAB has an fft function that will allow us to extract the data that we need with a single command.

We record a series of images from a camera and put them into a 3D array or movie, image_arr, where the array parameters are [x-position, y-position, time], i.e. a 2D slice is a single frame from the camera. Then we can do an FFT on that object with the syntax, f3D = fft(image_arr, [ ], 3), which only does the FFT on the temporal components. The resulting object is a 3D array where each 2D slice is an 2D array of amplitude and phase information across the image for a single temporal frequency of the movie.

So if we recorded a movie for 1s where the sample rate is 58Hz, then the 1st frame of f3D is just a DC image of the movie, the 2nd frame are the complex 1Hz components of the movie, etc all the way up to 29Hz. 

Suppose then that we have a image, part of which is being modulated, e.g. a chopper wheel rotating at 20 or 24Hz, or a laser beam profile which contains a 1kHz beat between a sideband and a reference beam. All we have to do is sample at at least twice that modulation frequency, run the command in MATLAB, and then we immediately get an image which contains the phase and magnitude information that we're interested in (in the appropriate 2D slice o the FFT).

As an example, I recorded 58 frames of data from a camera, sampling at 58Hz, which was looking at a spinning chopper wheel. There was a white sheet of paper behind the wheel which was illuminated from behind by a flashlight. The outer ring was chopping at 24Hz and the inner ring was chopping at 20Hz. I stuck all the images into the 3D array in MATLAB, did the transformation and picked out the DC, 20Hz and 24Hz signals. The results are shown in the attached PDFs which are:

  1. phase_camera_DC_comp.pdf - a single image from the camera and the DC component (zoomed in) of the FFT
  2. phase_camera_F1_comp.pdf - the magnitude and phase information of the 20Hz component of the FFT
  3. phase_camera_F2_comp.pdf - the magnitude and phase information of the 24Hz component of the FFT (this PDF contains a typo that says 25Hz).
  4. load_raw_data.m - the MATLAB routine that loads the saved data from the camera and does the FFT

You can, and I have, run the MATLAB engine from C directly. This will allow you to transfer the data from the camera to MATLAB directly in memory, rather than via the disk, but it does need proper memory allocation to avoid segmentation faults - that was too frustrating for me in the short term. In this case, the 58 frames were recorded to a file as a contiguous block of data which I then loaded into MATLAB, so it was slower than it might've otherwise been. Also the computer I was running this on was a bit of a clunker so it took a bit of time to do the FFT.

The data rate from the camera was 58fps x (1024 x 1024) pixels per frame x 2 bytes per pixel = 116MB per second. If we were to use this technique in a LIGO phase camera, where we want to measure a modulation which is around 1kHz, then we'd need a sample rate of at least 2kHz, so we're looking at at least a 30x reduction in the resolution. This is okay though - the original phase camera had only ~4000 spatial samples. So we could use, for instance, the Dalsa Falcon VGA300 HG which can give 2000 frames per second when the region of interest is limited to 64 pixels high.

Attachment 1: phase_camera_DC_comp.pdf
phase_camera_DC_comp.pdf
Attachment 2: phase_camera_F1_comp.pdf
phase_camera_F1_comp.pdf
Attachment 3: phase_camera_F2_comp.pdf
phase_camera_F2_comp.pdf
Attachment 4: load_raw_data.m
% load a raw data file into MATLAB

fid = fopen('phase_camera_data.dat');
n1 = 750;
A3D = ones(n1, n1, 58);

for jj = 1:58
    A = fread(fid, [1024, 1024], 'uint16');
    A3D(:,:,jj) = A((512-floor(n1/2)):(512-floor(n1/2))+n1-1, ...
                    (512-floor(n1/2)):(512-floor(n1/2))+n1-1);
... 64 more lines ...
  2955   Thu May 20 10:06:56 2010 AidanHowToPhase CameraPhase Camera- raw data video


 

  3096   Tue Jun 22 09:45:21 2010 Aidan, Joe, RazibUpdatePhase CameraCurrent phase camera setup. Seeing Acoustic beat

 We've set up a preliminary test bed for the phase camera. It simply uses a HeNe that is split into two beams. One is frequency shifted by an AOM by -40 MHz - df, where df is some acoustic frequency. The second beam is transmitted through a 40MHz EOM to get sidebands. The two beams are recombined and are, currently, incident on a photodetector, but this can be replaced by the phasecamera.

We turned everything on with df = 1kHz and confirmed that a 1kHz signal is visible on the output from the photodetector (PD). The signal looks to be about 1:300 of the DC level from the PD.

Attachment 1: 2010-06-22_Phase_camera_layout_version_1.pdf
2010-06-22_Phase_camera_layout_version_1.pdf
  3097   Tue Jun 22 13:38:13 2010 KojiUpdatePhase CameraCurrent phase camera setup. Seeing Acoustic beat

1. In terms of the AOM:

How much beam power is incident on the AOM? How much is the deflection efficiency?
i.e. How much is the power lost by the crystal, deflected in the 1st order, and remaining in the oth order?

I am curious because I assume the AOM (which vender?) is designed for 1064nm and the setup uses 632nm.

2. In terms of the EOM:

How much sidebands do you expect to have?

I assume the EOM is designed for 1064nm, the only difference is the coating at the end. Is this right? 

3. Beating

How much beating strength do you expect?

Is your beating level as expected?

How much is the contrast between the PM sideband and the frequency shifted carrier?
This must include the consideration on the presence of the carrier and the other sidebands.

Quote:

 We've set up a preliminary test bed for the phase camera. It simply uses a HeNe that is split into two beams. One is frequency shifted by an AOM by -40 MHz - df, where df is some acoustic frequency. The second beam is transmitted through a 40MHz EOM to get sidebands. The two beams are recombined and are, currently, incident on a photodetector, but this can be replaced by the phasecamera.

We turned everything on with df = 1kHz and confirmed that a 1kHz signal is visible on the output from the photodetector (PD). The signal looks to be about 1:300 of the DC level from the PD.

 

ELOG V3.1.3-