ID |
Date |
Author |
Type |
Category |
Subject |
15207
|
Tue Feb 11 19:11:35 2020 |
shruti | Update | Computer Scripts / Programs | MATLAB on donatella |
Tried to open MATLAB on Donatella and found the error:
MATLAB is selecting SOFTWARE OPENGL rendering.
License checkout failed.
License Manager Error -9
This error may occur when:
-The hostid of this computer does not match the hostid in the license file.
-A Designated Computer installation is in use by another user.
If no other user is currently running MATLAB, you may need to activate.
Troubleshoot this issue by visiting:
http://www.mathworks.com/support/lme/R2015b/9
Diagnostic Information:
Feature: MATLAB
License path: /home/controls/.matlab/R2015b_licenses/license_donatella_865865_R2015b.lic:/cvs/cds/caltech/apps/lin
ux64/matlab15b/licenses/license.dat:/cvs/cds/caltech/apps/linux64/matlab15b/licenses/license_chiara_
865865_R2015b.lic:/cvs/cds/caltech/apps/linux64/matlab15b/licenses/license_pianosa_865865_R2015b.lic
Licensing error: -9,57.
So I used my caltech credentials to get an activation key for the computer. I could not find the option for a campus license so I used the individual single machine license.
Now it can be run by going to the location:
/cvs/cds/caltech/apps/matlab17b/bin
and running
./matlab
On opening MATLAB, there were a whole bunch of other errors including a low-level graphics error when we tried to plot something. |
15206
|
Tue Feb 11 16:39:00 2020 |
shruti | Update | ALS | AM/PM |
The results of the AM/PM measurements:
- Attachment 1: Traces of 9 AM TFs overlaid on top of each other, calibrated by measuring the voltage at the ‘GREEN_REFL’ output where the TF was measured (described in elog 40m:15197). This was almost exactly 2 V.
- Attachment 2: Traces of 9 PM TFs also overlaid measured using DLFD (as described in elog 40m:15180). Calibrated using the measured ~600 mV pk-pk voltage. The phase plots were unwrapped (shifted by 180 deg if needed) so that each started from roughly 0 deg.
Both the AM and PM TFs were scaled to make them have the same average value. Manually adjusting the delay line offset for each measurement using the oscilloscope was probably not accurate enough and therefore resulted in different scaling which this should somewhat compensate.
Attachment 3:
- The orange and green lines are the averages of the PM and AM values of Attachments 1 and 2 respectively.
- The solid red line is at 230 kHz, which was the previously chosen value for PDH locking. The peak seems to have shifted to the left from previous measurements (elog 40m:12077).
- A horizontal black dashed line is drawn to show where the ratio is 10^5.
- The red regions correspond to frequencies where PM/AM > 10^5 [only shown for frequencies greater than 200kHz], these are roughly (in kHz):
- 211.4-213.9
- 221.4-230.7 (peak at 225.642)
- 240.8-257.9
- ~748.3
- 753.3-799.8, two largest peaks at 763.673 and 770.237
- 809.6-829.3, peak at 819.472
- 839.2-842.4
- 881.8-891.7
Updated Calibration
Attachment 2 and 3 were miscalibrated due to an error in my understanding of the delay line, but the net result of the change in factors is qualitatively almost the same and the position of the major peaks remain predominantly unchanged.
The new plot is in Attachment 5.
The new calibration factor used: 5 MHz/V at the output of the mixer to obtain the frequency modulation and then division by the mod. freq. to obtain PM.
5 MHz/V because changing the PZT voltage by 0.01 V=> change in beat frequency by 0.1 MHz, which was seen as a 20 mV change in the delay line mixer output.
Again, the calibration is not very precise and I will probably repeat this experiment at some point more precisely. |
Attachment 1: AM.pdf
|
|
Attachment 2: PM.pdf
|
|
Attachment 3: Ratio_all.pdf
|
|
Attachment 4: Ratios_FM_PM.pdf
|
|
Attachment 5: Ratio_all_new.pdf
|
|
15205
|
Mon Feb 10 15:55:46 2020 |
Jordan | Update | PSL | PMCTRANSPD |
[Gautam, Jordan]
Gautam showed me how the PMCTRANSPD signal was reading zero, and he suspected it might have to do with the acromag wiring. Disconnected the acromag box underneath the PSL table and checked the ADC wiring. Side note: When benchtesting the c1psl acromag chassis there was excess noise in the AI channels, and grounding the minus pin of the ADC channel eliminates the noise.
So I grounded the (-) pins on the ADC1 (192.168.113.122), which PMCTRANSPD is connected to and that seemed to fix the problem. As of right now PMCTRANSPD is reading ~.75 V.
See attached pictures
gautam: While this fix seems to have worked, I wonder why this became necessary only in the last month. Note that the problem was a noisy readback on the PMC transmission PD, which also made the FSS_RMTEMP channel noisy, leading me to suspect some kind of ground loop issue. |
Attachment 1: ADC1.jpg
|
|
15204
|
Mon Feb 10 15:54:47 2020 |
Jordan | Update | PSL | Completed Acromag Wiring |
All spare channels on the PSL acromag chassis are connected with ~12in of spare wiring for future use. |
15203
|
Mon Feb 10 15:04:42 2020 |
Jordan | Update | General | HDMI Routing for new tv |
Ran HDMI to the new tv mounted on the north wall of control room. |
15202
|
Mon Feb 10 10:07:20 2020 |
gautam | Update | PSL | PMC re-locked |
I found the PMC unlocked this morning. It was re-locked using the usual procedure. I feel like this has been happening more frequently in the last month than before. In the past, the cause seems to have been the PZT voltage drifting too close to one of the rails - however, in this case, it looks like an IMC unlock event is what triggered the PMC lockloss (admittedly the PZT voltage was somewhat close to the rail). It would be good if someone can re-connect the PMC Transmission photodiode, it was a useful diagnostic channel we had working fine before the ringdowns started.
I also tweaked the input pointing into the PMC and ran the WFS DC offset relief script. |
Attachment 1: PMCunlock.png
|
|
15201
|
Mon Feb 10 09:40:54 2020 |
Larry Wallace | Summary | General | SolidWorks Computer Upgrade and Printer repair |
On February 5, 2020, the Dell engineering workstation located in the 40M lab, was replaced with a newer Engineering workstation, per a request from Koji . The new workstation should perform a good deal better over the older unit. It has more cores, more memory and a better video card. Since this unit is being used by the 40M group, the Comsol s/w pkg. was also installed on the unit.
During the computer swap, Koji had a problem with a print job and it was discovered the bottom tray of the HP5550 printer was broken. The broken tray was replaced from another unit that was being disposed of. |
15200
|
Fri Feb 7 19:39:10 2020 |
Koji | Update | LSC | More high BW POY experiments |
This measurement tells you how the gain balance between the SLOW_CM and AO paths should be. Basically, what you need is to adjust the overall gain before the branch of the paths.
Except for the presence of the additional pole-zero in the optical gain because of the power recycling.
You have compensated this with a filter (z=120Hz, p=5kHz) for the CM path. However, AO path still don't know about it. Does this change the behavior of the cross over?
If the servo is not unconditionally stable when the AO gain is set low, can we just turn on the AO path at the nominal gain? This causes some glitch but if the servo is stable, you have a chance to recover the CARM control before everything explodes, maybe? |
15199
|
Fri Feb 7 15:00:16 2020 |
gautam | Update | LSC | More high BW POY experiments |
To study the evilution of the AO path TFs a bit more, I've hooked up POY11_Q Mon to IN1 of the CM board. I will revert the usual setup later in the evening.
Update 1730: I've returned the cabling at 1Y2 to the nominal config, and also reverted all EPICS settings that I modified for this test. Y-arm POY locking works. Attachment #1 shows the summary of the results of this test - note that the AO gain was kept fixed at +5dB throughout the test. I have arbitrarily trimmed the length of the frequency vector for some of these traces so that the noisy measurement doesn't impede visual interpretation of the plots so much. At first glance, the performance is as advertised. I basically followed the settings I had here to get started, and then ramped up various gains to check if the measured OLTF evolved in the way that I expected it to. The phase lead due to the AO path is clearly visible.
Some important differences between this test and the REFL11 blending is (i) in the latter case, there will also be a parallel loop, CARM_A, which is effecting some control, and (ii) the optical gain of CARM-->REFL11_I is much higher than L_Y-->POY. So the initial gain settings will have to be different. But I hope to get some insight into what the correct settings should be from this test. I think IMC servo IN2 gain and AO gain slider on the CM board are degenerate in the effect they have, modulo subtle effects like saturation.
One possibility is that the gain allocation I used yesterday was wrong for the dynamic range of the CARM error signal. In some initial trials today, when I set the CM board IN1 gain to -32dB (as in the case of attempting the CARM RF handoff) and compensated for the reduced POY PDH fringe amplitude by increasing the digital gain for the CM_Slow path, I found that there was no phase advance visible even when I ramped up the IMC IN2 gain to +10dB. So, for the CARM handoff too, I might have to start with a higher CM board IN_1 gain, compensate by reducing the CM_Slow digital gain even more, and then try upping the IMC IN2 gain.
P.S. When the excitation input to the CM board was enabled in order to make TF measurements, I saw significant increase in the RMS of the error signal. Probably some kind of ground loop issue. |
Attachment 1: AO_TFs.pdf
|
|
15198
|
Fri Feb 7 12:58:25 2020 |
Yehonathan | Update | General | Metal PMC parts |
I took the metal PMC box and examined its content and find the following items:
Name |
Quantity |
Picture (Attachment #) |
Metal PMC body (PMC1) |
1 |
1-3 |
Metal PMC body with two mounted 41 deg mirrors (PMC2) |
1 |
4-6 |
"Baked PZT Caps" |
3 |
7 |
PZT Caps |
2 |
8 |
Flat mirror mounts |
2 |
9 |
Bar clamps |
4 |
10 |
Clamp studs |
8 |
10 |
PZTs |
4 |
11 |
ORings |
INF |
12 |
Ball bearings |
INF |
13 |
6.8 deg AoI curved mirrors (r=-1000mm) |
6 |
14 |
41 deg AoI flat mirrors, R=99% @ 1064nm (1 Damaged) |
11 |
15 |
There seem to be enough parts to build 2 PMCs + spares.
I find several problems in the metal PMCs:
PMC1 has a broken screw in one of its flat mirror mounts (Attachment 16). We need to get it out in the machine shop.
PMC2 one of the flat mirrors has a scratch on the AR coating and its ORing is failing (Attachment 17). Mirror and ORing need to be replaced.
I measure the physical dimensions of the PMC with the help of https://dcc.ligo.org/LIGO-E1400332. The roundtrip is found to be 24cm which gives an FSR of 1.25GHz.
I use Evan Hall's Python script for calculating the mode spectrum as a function of the cavity length of the metal PMC and overlay the RF sidebands (Green dashed lines) on it (Attachment 18) to check for any HOM coincidence. The width of the lines is the mode splitting due to the cavity astigmatism.
It seems like the only issue might come from a 10th order modes (green ribbon) which are hopefully small enough in reality. |
Attachment 1: PMC1Side.jpg
|
|
Attachment 2: PMC1Front.jpg
|
|
Attachment 3: PMC1Back.jpg
|
|
Attachment 4: PMC2Side.jpg
|
|
Attachment 5: PMC2Back.jpg
|
|
Attachment 6: PMC2Front.jpg
|
|
Attachment 7: 20200207_123118.jpg
|
|
Attachment 8: 20200207_123055.jpg
|
|
Attachment 9: 20200207_122448.jpg
|
|
Attachment 10: 20200207_122400.jpg
|
|
Attachment 11: 20200207_122040.jpg
|
|
Attachment 12: 20200207_122227.jpg
|
|
Attachment 13: 20200207_122149.jpg
|
|
Attachment 14: 20200207_123328.jpg
|
|
Attachment 15: 20200207_123405_HDR.jpg
|
|
Attachment 16: PMC1Screw.jpg
|
|
Attachment 17: PMC2BadMirror.jpg
|
|
Attachment 18: homVersusLength.pdf
|
|
15197
|
Fri Feb 7 09:45:03 2020 |
shruti | Update | General | AM at X end |
I took a few AM TF measurements at the X end for which I:
- Misaligned the ITMX (then re-aligned it)
- Opened the X green shutter during the measurements and closed it at the end
- Moved the Agilent from the PSL area to the X end, the delay line and mixer still remains near the PSL area (will move it soon)
- Took a bunch of TFs
I will post the data soon. |
15196
|
Fri Feb 7 02:41:28 2020 |
Koji | Update | General | office area temperature |
Not sure what's wrong, but the workstation desk is freezing cold again and the room temp is 18degC (64degF). |
15195
|
Fri Feb 7 02:24:24 2020 |
gautam | Update | LSC | Some short notes |
[koji, gautam]
Plots + interpretation tomorrow.
- CM_Slow path can be used to stabilize the arm powers somewhat but the AO crossover remains out of reach.
- The REFL11 (=CARM_B) path offset has to be manually determined - we found that it can change by ~20% depending on the alignment, which maybe isn't surprising given that the mode shapes seen at POP, REFL and AS look like Rorschach inkblots.
- We saw TRX/TRY regularly hit ~150, and at times even 200 (= recycling gain of ~10). Though any conclusive statement about the PRG can only be made once the lock is stabilized.
- I was able to take a few CARM loop TFs with an SR785 hooked up at 1Y2. Despite ramping up the AO gain, we saw no effect at high frequencies in the TF shape (the phase bubble continued to roll off at ~100 Hz and there was no visible phase lead even as the AO gain was increased). It has to be estimated what the expected crossover gain is from the experiment with the high BW POY locking (taking into account the net difference in optical gain between POY for single arm and REFL for the full IFO).
- The fact that I was able to hold the high BW POY lock makes me think that the IMC servo board's IN2 input (and indeed the rest of the IMC locking loop) is functioning as expected. But maybe this board will benefit from a detailed checkout like Koji did for the CM board.
Getting closer... To facilitate this work, I made some convenience scripts that can be run from the CM MEDM screen. |
15194
|
Thu Feb 6 21:54:13 2020 |
Jon | Update | PSL | c1psl bench testing complete |
Today I engineered the last piece of the new c1psl system: the multi-bit binary output (mbbo) channels that control the MC servo board gains. These 6-bit channels have to be split across two 4-bit Acromag registers. To enforce synchronous switching, I adapted the latch.py script developed by Gautam to address this problem in c1iscaux. Analogously to the c1iscaux implementation, I scripted the code to automatically run as a systemd service which is launched by the main modbusIOC service. I tested this all using the DB37 LED test board and confirmed it to work.
This now completes the electronics bench testing.
There are still several DB37 connectors to be wired, which carry only spare channels for future use and are not interfaced with the EPICS IOC. Jordan and I discussed this today and he or Chub will complete it shortly. To allow time for the spare channel wiring to be completed (as well as for more locking progress before interruption), Gautam and I think Monday/Tuesday next week would be the earliest possible window to install the new system. |
15193
|
Thu Feb 6 16:14:44 2020 |
rana | Update | General | office area temperature |
I changed the office area thermostate near Steve's desk from 68F to 73F today. Please do not change it.
If anyone from facilities comes to adjust something, please put the details in the elog on the same day so that we can know to undo that change rather than chase down other drifts in the system. |
15192
|
Thu Feb 6 01:25:50 2020 |
gautam | Update | LSC | Locking updates |
Summary:
I managed to partially stabilize the arm citculating powers - they stay in a region in which the REFL 11 signal is hopefully approximately linear and so I can now measure some loop TFs and tweak the transition appropriately.
Procedure:
The main change I made tonight was to look at the REFL11 signal as I swept the ALS CARM offset through 0. I found that the maximum arm powers coincided with a non-zero REFL11 signal value (i.e. a small CARM offset was required at the input to the CARM_B filter bank). Not so long ago, I had measured the PM/AM ratio for 11 MHz to be ~10^5 - so it's not entirely clear to me where this offset is coming from. Then, I was able to turn on the integrator (z:p = 20:0) in the CARM_B filter bank while maintaining high POP_DC. At this point, I ramped up the IN2 gain on the IMC servo board (= AO path), and was able to further stabilize the power.
Attachment #1 shows this sequence from earlier in the evening. Note that in this state, both ALS and IR control of CARM is in effect. The circulating power is fluctuating wildly - partly this is probably the noisy ALS control path, but there is also the issue of the (lack of) angular control - although looking at the transmon QPDs and the POP QPD signals, they seem pretty stable.
The next step will be to try and turn off the ALS control path. Eventually, I hope to transition DARM control to AS55 as well. But at this point, I can at least begin to make sense of some of the time series signals, and get some insight into how to improve the lock.
Quote: |
No systematic diagnosis scheme comes to mind.
|
|
Attachment 1: semiStableArms.png
|
|
Attachment 2: armAngStability.png
|
|
15191
|
Thu Feb 6 01:16:58 2020 |
gautam | Update | LSC | Diagnosis results |
Summary:
I did some more detailed tests to see if I could isolate where the excess ALS noise at low CARM offset is coming from, by measuring the spectrum of the IMC error point (in loop). The results, shown in Attachment #1 and #2, are inconclusive.
Details:
Since MC_F didn't show any signatures of elevated noise, I decided to hook up an SR785 to the A excitation bank TEST1 input of the IMC servo board to monitor the in-loop error signal. I initially took a few measurements spanning 800 Hz in frequency, and to my surprise, I found that there was elevated noise in the frequency band we see an increase in the ALS noise, even when the CARM feedback goes to the ETMs (so the IMC cavity is in principle isolated from the main interferometer). This is Attachment #1. So I re-took a couple of measurements (this time only for the case of CARM feedback to the ETMs), with a 200 Hz frequency span, and found no significant noise elevation. This is Attachment #2. I am led to conclude that the IMC error point level changes over time for reasons other than the CARM offset - it'd be nice to have a spectrogram of the IMC error point and compare excursions relative to the median level over a few 10s of minutes, but we don't have this data stream digitized by the CDS system - maybe I will hijack the MC_L channel temporarily to record this data stream. It seems a waste that we're not able to take full advantage of the measured <10pm RMS noise of the IR ALS system. |
Attachment 1: IMCspec_ALS.pdf
|
|
Attachment 2: IMCspec_ALS_smallSpan.pdf
|
|
15190
|
Wed Feb 5 21:13:17 2020 |
Yehonathan | Update | IOO | IMC Ringdowns extended data analysis |
I translate the results obtained in the previous elog to the IMC 3 mirror cavity. I assume the loss in each mirror in the IMC is equal and that M2 has a negligible transmission.
I find that to a very good approximation the loss per IMC mirror is 2/3 the loss per mirror in the 2 mirror cavity model. That is the loss per mirror in the IMC is 56 ppm. The transmission per mirror in the IMC is the same as in the 2 mirror model, which is 1980 ppm.
The total transmission is the same as in the 2 mirror model and is given by:

where is the coupling efficiency to the TEM00 mode. |
15189
|
Wed Feb 5 21:04:10 2020 |
Yehonathan | Update | PSL | Bench testing of PSL ai channels |
{Yehonathan, Jon}
We retested the failed ai channels. Most of them got fixed by applying the inverse calibration in the yaml file.
We still find some anomalous channels, mostly in the DB25 connector. Turns out, their limits were ill-defined in the EPICS database. Specifying the right limit fixed the issue.
We find one miswired channel (BNC4). We connected the BNC to the right channel on the Acromag unit which fixed the issue.
Overall all the ai channels were successfully bench-tested.
Quote: |
I checked the failed channels against the EPICS database definitions and the yaml file inputted to IFOTest. The channels where the readings are something other than +10/0 V, but the high/low values do change, I think can be attributed to one of two things:
- An incorrect gain and/or offset conversion parameter in the yaml file
- The EPICS SMOO parameter (smoothing) is set to some long value
I fixed the channel gains/offsets in the master yaml file (PSL.yaml). I also disabled smoothing in the EPICS defintions of the new PSL channels for the purpose of testing. We can uncomment those lines after installing the new chassis if noise is a problem. Please go ahead and re-test the channels again.
Quote: |
I marked in red the failed channels. Some of them are probably calibration issues, but the ones that show the same voltage for high and low are probably disconnected wires.
|
|
|
15188
|
Wed Feb 5 16:35:12 2020 |
gautam | Update | LSC | Diagnosis plan |
The goal is to try and identify the source of the excess ALS noise as the CARM offset is reduced. The idea is to look at the MC_F spectrum (or the IMC error point) in a few conditions:
- Regular CARM --> MC2 actuation scheme, PRMI locked on 3f signals, CARM held off resonance.
- Regular CARM --> MC2 actuation scheme, PRMI locked on 3f signals, CARM held on resonance.
- Alternate CARM --> 7.5*ETMX + 1.5*ETMY, PRMI locked on 3f signals, CARM held on resonance.
- Control arms in X/Y basis, lock PRMI on 3f signals and bring the arms into resonance individually, look for excess ALS noise.
#1 vs #2 is like a control experiment, we expect to see the excess noise imprinted on the MC length and hence in MC_F (provided the sensing noise is low enough). #2 vs #3 will be informative of something like backscatter to the PSL increasing the frequency noise. #2/3 vs #4 will help isolate the problem to an individual arm's AUX PDH loop or some optomechanical effect.
I was looking back at some spectra from the last couple of nights but I don't really have an apple-to-apple comparison in the various actuation schemes (some ALS loops were engaged/disengaged), so I'll do a more systematic test tonight. Already, it looks like MC_F is not a good candidate to look for the excess frequency noise, I don't really see a big difference between conditions #1 and #2. According to this, we are looking for an increase at the level of a few 100Hz/rtHz @ ~40 Hz, wheras MC_F is much noisier. |
Attachment 1: ALSnoise.pdf
|
|
15187
|
Wed Feb 5 08:57:11 2020 |
Yehonathan | Update | PSL | Bench testing of PSL ai channels |
I checked the failed channels against the EPICS database definitions and the yaml file inputted to IFOTest. The channels where the readings are something other than +10/0 V, but the high/low values do change, I think can be attributed to one of two things:
- An incorrect gain and/or offset conversion parameter in the yaml file
- The EPICS SMOO parameter (smoothing) is set to some long value
I fixed the channel gains/offsets in the master yaml file (PSL.yaml). I also disabled smoothing in the EPICS defintions of the new PSL channels for the purpose of testing. We can uncomment those lines after installing the new chassis if noise is a problem. Please go ahead and re-test the channels again.
Quote: |
I marked in red the failed channels. Some of them are probably calibration issues, but the ones that show the same voltage for high and low are probably disconnected wires.
|
|
15186
|
Tue Feb 4 18:13:01 2020 |
Yehonathan | Update | PSL | Bench testing of PSL ai channels |
{Yehonathan, Jon, Jordan}
I tested the ai channels of the new PSL Acromag by looping an already-tested ao channel (C2:PSL-FSS-INOFFSET) back to the different ai channels.
I use Jon's IFOTest with /users/jon/ifotest/PSL.yaml.
I created a spreadsheet for the testing based on the current wiring spreadsheet. I added two columns for the high and low readings for each ai channel (attached pdf).
I marked in red the failed channels. Some of them are probably calibration issues, but the ones that show the same voltage for high and low are probably disconnected wires.
I redid the test on the channel that seemed disconnected to confirm.
I created a yaml file with all the failed channels for retesting called /users/jon/ifotest/PSL_failed_channels.yaml. |
Attachment 1: c1psl_wire_testing_-_By_Connector.pdf
|
|
15185
|
Tue Feb 4 02:13:02 2020 |
gautam | Update | LSC | Locking updates |
Summary:
The CARM-->RF transition remains out of reach. No systematic diagnosis scheme comes to mind.
Details:
- Config is PRFPMI, SRM is misaligned macroscopically.
- PRMI can easily be locked with 3f signals while CARM is offset from resonance. Aided by DAFI, I turned on the PR violin filter in the BS output section to prevent it from ringing up, making the lock much more robust.
- When the CARM offset is reduced
- POP22 level dips and sometimes goes negative - i don't see this in my simple simulations. POP22 is the trigger signal for MICH/PRCL loops, so to prevent the PRMI lockloss, I mix in some POPDC into the trigger matrix element.
- Once the circulating power exceeds ~10, the ALS noise apparently increases.
- The arms "buzz" through resonance, but the power fluctuation is nearly 0-200 in TRX/TRY, corresponding to several CARM linewidths, but all the out-of-loop ALS noise measurements have me believe that we are close to the CARM linewidth in noise. So we should only see ~factor of 2 fluctuation in power.
- The RF error signal for CARM (=REFL 11) doesn't show any features that i can use to aid the transition / diagnose what is going on systematically.
- Koji suggested changing the actuation for CARM from MC2 to the ETMs, and check if the MC OSEMs witness the excess motion at small CARM offsets
- The ALS transition is scripted, so I had to make a modified version that accommodates this changed actuation scheme.
- The usual CARM-->MC2 matrix element is -1.
- The frequency actuation strength of MC2 is ~3x that of the ETMs. Additionally, ETMX has 5x the series resistance of ETMY. So I used the output matrix elements shown in Attachment #1 so as to get the same loop UGF with the same loop gains elsewhere in the chain. Confirmed the actuation strength is the same using the sensing matrix infrastructure and comparing line heights.
- Attachment #2 shows the measured UGF - both CARM and DARM look okay to me.
- With this new ALS output matrix actuation scheme, I was able to make it to PRMI + arms on zero offset a couple of times tonight, but the drifting input alignment makes the PRMI lock not so robust anymore.
TBC. Mercifully at least the shaker stayed still tonight. |
Attachment 1: modifiedOutMat.png
|
|
Attachment 2: OLTFs.pdf
|
|
15184
|
Mon Feb 3 15:22:39 2020 |
Jon | Update | PSL | c1psl progress/Acromag ADC grounding |
I tested the c1psl AO channels on the electronics bench on Friday. While I found all the wiring to be correct, some of the channels exhibited excess noise with all appearances of a grounding problem.
Today Jordan, Gautam, and I investigated this further. It is indeed a grounding problem, but actually with the Acromag ADCs. The Acromag DAC outputs are single-ended (return is grounded), so (for the purpose of a loopback test) I would expect to leave the ADC inputs ungrounded. This is the configuration I tested Friday. Today we also tested driving the ADC with a floating source. The ADC noise behavior is exactly the same, whether the source end is grounded or not.
However, grounding the minus pin of the ADC channel eliminates the noise. We don't understand why this seems to be required irrespective of the driving source, so there something we're missing about the ADC design. As it turns out, this same fix was made to the AI channels of the previously-upgraded Acromag machines. I know Chub and I had to do this for the AI channels of c1vac, but at the time we thought the source grounding was causing the issue. However, today Jordan and I looked inside c1iscaux, which Chub wired, and confirmed that its AI channels are wired in the same way.
So in any case, Jordan is grounding the c1psl AI channels in the same way as c1iscaux. Once this is done, we'll continue with the bench testing tomorrow.
gautam: here are my notes about this issue when i was doing the c1iscaux testing. As I note there, "previously-upgraded Acromag machines" in the plural may be a bit of a stretch - I have no idea what the grounding situation is in c1susaux / c1auxex for example. |
15183
|
Mon Feb 3 13:54:10 2020 |
Yehonathan | Update | IOO | IMC Ringdowns extended data analysis |
I extended the ringdown data analysis to the reflected beam following Isogai et al.
The idea is that measuring the cavity's reflected light one can use known relationships to extract the transmission of the cavity mirrors and not only the finesse.
The finesse calculated from the transmission ringdown shown in the previous elog is 1520 according to the Zucker model, 1680 according to the first exponential and 1728 according to the second exponential.
Attachment 1 shows the measured reflected light during an IMC ringdown in and out of resonance and the values that are read off it to compute the transmission.
The equations for m1 and m3 are the same as in Isogai's paper because they describe a steady-state that doesn't care about the extinction ratio of the light.
The equation for m2, however, is modified due to the finite extinction present in our zeroth-order ringdown.
Modelling the IMC as a critically coupled 2 mirror cavity one can verify that:
![m_2=P_0KR\left[T-\alpha\left(1-R\right)\right]^2+\alpha^2 P_1](https://latex.codecogs.com/gif.latex?m_2%3DP_0KR%5Cleft%5BT-%5Calpha%5Cleft%281-R%5Cright%29%5Cright%5D%5E2+%5Calpha%5E2%20P_1)
Where is the coupled light power
is the power rejected from the cavity (higher-order modes, sidebands)
is the cavity gain.
and are the power reflectivity and transmissivity per mirror, respectively.
is the power attenuation factor. For perfect extinction, this is 0.
Solving the equations (m1 and m3 + modified m2), using Zucker model's finesse, gives the following information:
Loss per mirror = 84.99 ppm
Transmission per mirror = 1980.77 ppm
Coupling efficiency (to TEM00) = 97.94% |
Attachment 1: IMCTransReflAnalysis_anotated.pdf
|
|
15182
|
Fri Jan 31 16:57:09 2020 |
gautam | Update | General | Metal PMC parts |
Jon brought over a box of parts for constructing the metal PMCs. I have stored it along the Y-arm, on top of the green optics cabinet.
I didn't do an exhaustive inventory check, but the following are the rough contents of the box:
- 41 deg AoI flat mirrors, R=99% @ 1064nm --- 11 pcs
- 6.8 deg AoI curved mirrors --- 5 pcs
- PZTs --- 3pcs
- Metal PMC body --- 2 pcs
- "Baked PZT endcaps" --- 3 pcs
- Ball bearings, clamps, misc hardware
I didn't inspect the optics but since we have so many, I am hoping we can find 3 good quality ones for one cavity at least. We should check that the geometry is suitable for our RF sideband frequencies. |
15181
|
Fri Jan 31 16:04:30 2020 |
gautam | Update | IOO | Input pointing drift |
One factor which hampers locking efforts is the apparent drift of the input beam into the IFO. Over timescales of ~1 hour, I have noticed that the spot on the AS camera drifts significantly (~1 spot size) in pitch. The IPPOS QPD bears out this observation, see Attachment #1. The IMC WFS control signals do not show a correlated drift, hence my claim that the TTs are to blame.
I am able to correct this misalignment by moving TT1 in pitch (see Attachment #2, which shows some signals from a ~1 hour PRMI lock, during which time the pointing drifted, and I corrected it by moving TT1 pitch). Assuming the problem is purely TT1 pitch drifting, this corresponds to 3mm / 6m ~500urad of shift in 1 hour - seems very large. The fact that the drift is only present in pitch and doesn't really show up in yaw makes me think the problem is likely mechanical (unless the voltage to the top two coils is drifting relative to the bottom, but no LR drift, which would be very coincidental). At the moment, this is just an annoyance, but it'd be good for this problem to be fixed.
I haven't yet figured out how to make ndscope export these plots to SVG preserving the dark color theme, hence the weird light axes... |
Attachment 1: IPdrift.pdf
|
|
Attachment 2: IPdrift_PRMI.pdf
|
|
15180
|
Thu Jan 30 22:02:42 2020 |
shruti | Update | General | delay line frequency discriminator for PM |
I could not find any level 3 mixers, but by adjusting the beat frequency the power in each of the delay line channels rose to almost 6.5 dBm.
In short: Delay line seems to work
Things I did earlier today:
- Played with the slow servo on the FSS screen, but then reset the parameters to what was there before (Later found out that this was to lock the PSL freq to the IMC when the IMC power is significant.)
- Connected the AG 4395A to the X PZT
- Closed the PSL shutter
Transfer function measurement: (Refer Attachment 1)
Everything about the setup remained as I had left it earlier: described in elog 15174
except
- SR560 gain set to 10, DC coupled
- DC block at channel A of Agilent (The measurement was A/R)
I did not use a slow servo, but took individual sweeps adjusting the PSL temperature each time to bring the error voltage between +/-25 mV. The beat frequency was over 100 MHz.
For the plot posted in Attachment 1, the measurement paramters are the following. Will do further measurements/analysis tomorrow.
# AG4395A Measurement - Timestamp: Jan 30 2020 - 21:58:00
# Parameter File: TFAG4395Atemplate.yml
#---------- Measurement Parameters ------------
# Start Frequency (Hz): 50000.0, 50000.0
# Stop Frequency (Hz): 1000000.0, 1000000.0
# Frequency Points: 801, 801
# Measurement Format: LOGM, PHAS
# Measuremed Input: AR, AR
#---------- Analyzer Settings ----------
# Number of Averages: 1
# Auto Bandwidth: Off, Off
# IF Bandwidth: 1000.0, 1000.0
# Input Attenuators (R,A,B): 0dB 0dB 0dB
# Excitation amplitude = -20.0dBm
Quote: |
yes, its fine to use this with a level 3 or level 7 mixer; let's see some PM transfer functions !
Quote: |
Is this sufficient enough for the mixer to work?
|
|
|
Attachment 1: Figure_2.png
|
|
15179
|
Thu Jan 30 17:41:10 2020 |
gautam | Update | PSL | Errant FSS_INOFFSET change |
You can trend the data for the past few hours and see what the appropriate value. I think these tests should only be done when whoever is running a test is in the lab.
P.S. I was surprised that the IMC didn't lose lock when this step was applied. I manually stepped this voltage between +/- 10 V and didn't see any response in the FSS readbacks. Either the channel doesn't work, or there is a divide by 40 in the physical circuit or something...
Quote: |
A script I was testing errantly set C1:PSL-FSS_INOFFSET => 10 V at about 5:30 pm. I manually reverted the channel value to 0, but I don't know what the value was initially. Someone please check this value if there are problems locking the FSS.
|
|
15178
|
Thu Jan 30 17:31:28 2020 |
Jon | Update | PSL | Errant FSS_INOFFSET change |
A script I was testing errantly set C1:PSL-FSS_INOFFSET => 10 V at about 5:30 pm. I manually reverted the channel value to 0, but I don't know what the value was initially. Someone please check this value if there are problems locking the FSS. |
15177
|
Thu Jan 30 15:24:10 2020 |
? | Update | General | delay line frequency discriminator for PM |
yes, its fine to use this with a level 3 or level 7 mixer; let's see some PM transfer functions !
Quote: |
Is this sufficient enough for the mixer to work?
|
|
15176
|
Thu Jan 30 12:52:10 2020 |
Jon | Update | BHD | Metal OMCs procured |
Last night Yehonathan and I located the two steel PMCs in the QIL, with help from Anchal. They are currently sitting on my desk in Bridge, inside a box that also contains optics and other OMC parts. I will bring them over to the 40m the next time I come. |
15175
|
Wed Jan 29 12:40:24 2020 |
Yehonathan | Update | IOO | IMC Ringdowns preliminary data analysis |
I analyze the IMC ringdown data from last night.
Attachment 1 shows the normalized raw data. Oscillations come in much later than in Gautam's measurement. Probably because the IMC stays locked.
Attachment 2 shows fits of the transmitted PD to unconstrained double exponential and the Zucker model.
Zucker model gives time constant of 21.6us
Unconstrained exponentials give time constants of 23.99us and 46.7us which is nice because it converges close to the Zucker model. |
Attachment 1: IMCRingdownNormalizedRawdata.pdf
|
|
Attachment 2: IMCTransPDFits.pdf
|
|
15174
|
Wed Jan 29 12:29:33 2020 |
shruti | Update | General | delay line frequency discriminator for PM |
Today I began working on a TF measurement based on the delay line frequency discriminator setup in elog 4254 using a single mixer (without the 'I' and 'Q' readout).
For this, I re-organised the setup for the PLL measurement of the transfer function (elog 15148), increasing the HEPA for the initial changes while the PSL door was open, and then reverting it back to ~30%:
- I removed the 20dB coupler and connected the splitter directly after the amplifer to split the beat note signal into two coaxial cables one of which was ~1.5m longer than the other
- The recombined signals were combined in a mixer outside the PSL enclosure. I also replaced the 1.9 MHz LPF with a 5 MHz LPF.
- I used an SR 560 to amplify the signal after the LPF.
With the above setup the power that was seen at each channel of the delay line was <1dBm, which is not ideal for the any of the available mixers.
After the group meeting, I changed the amplifer to ZHL-3A (that is near the beat mouth) instead of a ZFL-500HLN because it had a higher gain (~28dB as opposed to ~19dB of the latter). The power seen at each of the delay line channels is over 5.5 dBm. This is consistent with the estimation 0 dBm beat -> -20 dBm after 20dB coupler -> 8 dBm after amplifier -> 5 dBm after splitter with insertion loss of 3 dB.
Is this sufficient enough for the mixer to work? In Attachment 1: A shows the mixer output (point B in Attachment 2) when the IMC is locked, in B the IMC is unlocked at the middle of the spectrum, and each of the dips show the DC voltage being sent to the PSL temperature servo being decreased by 0.01 V.
Gautam pointed me to the location of a few other RF amplifiers (ZHL-32A+, ZHL-1A) which don't possess a higher gain but can be used without disrupting the ALS related work (I was told).
For shorter duration changes that I made later, I opened and closed the PSL enclosure doors without changing the HEPA.
Attachment 2 shows the current setup as is, but I might add a PSL servo tomorrow to stabilise its frequency corresponding to a null mixer output without driving anything else. |
Attachment 1: 20200128.png
|
|
Attachment 2: IMG_BB01C068495A-1.jpeg
|
|
15173
|
Wed Jan 29 03:05:47 2020 |
rana, gautam | Update | SUS | MC misalignments / sat box games |
In the last couple days, as the IMC ringdowns have been going on, we have noticed that the MC is behaving bad. Misaligning, drifting, etc.
Gautam told me a horror story about him, Koji, and melted wires inside the sat boxes.
I said, "Its getting too hot in there. So let's take the lids off!"
So then we:
- Removed the lid (only 4 screws were still there)
- cut off some of the shield - ground wires and insulated them with electrical tape
- squished the IDC connectors on tightly
- left it this way to see if MC would get better - certainly the painfully hot heatinks inside the box were now just 110 F or so
After some minutes, we saw no drifting. So maybe my theory of "hot heatsink partially shorting a coil current to GND through partially melted ribbon cable" makes sense? IF this seems better after a month, lets de-lid all the optics.
Let's look at some longer trends and be very careful next to MC2 for the next 3 days! I have put a dangerous mousetrap there to catch anyone who walks near the vacuum chamber.
gautam: the grounding situation per my assessment is that the shield of all the IDC cables are connected to a common metal strip at 1X5 - but in my survey, I didn't see any grounding of this strip to a common ground. |
Attachment 1: IMG_8366.JPG
|
|
15172
|
Wed Jan 29 00:29:43 2020 |
gautam | Update | LSC | locking 2020 |
The goal tonight was to go through the locking scripts to see if I could recover the state from November 2019, when I could have the arm lengths controlled by ALS, and sit at zero CARM offset with the PRMI remaining locked and the arm powers fluctuating between 0-300. The IR-->ALS transitions went smoothly tonight, and the PRMI locking was also fairly robust when the CARM offset was large, but was less good when reduced to 0. Nevertheless, it is good to know that the system can be restored to the state it was late last year. Next step is to figure out how to keep the PRMI locked and get the AO path engaged, this was what I was struggling with before the new year. |
Attachment 1: PRFPMI_2020Jan.png
|
|
15171
|
Wed Jan 29 00:27:13 2020 |
gautam | Update | Computer Scripts / Programs | mcup / mcdown modified |
To fix the apparent slowness of execution of the caput commands on megatron, I changed the "ewrite" macro in the mcup and mcdown scripts to use ezcawrite instead of caput. The old lines are simply commented out, and can be reverted to at any point if we so desire. After these changes, we saw that both scripts complete execution much faster. |
15170
|
Tue Jan 28 20:51:37 2020 |
Yehonathan | Update | IOO | IMC WFS servos stable again |
I resume my IMC ringdown activities now that the IMC is aligned again.
To avoid any accidental misalignments Gautam turned off all the inputs to the WFS servo.
I set up a PD and a lens as in attachment 1 (following Gautam's setup).
I connect the REFL, TRANS and INPut PDs to the oscilloscope.
I connect a Siglent function generator to the AOM driver. I try to shut off the light to the IMC using 1V DC waveform and pressing the output button manually. However, it produced heavily distorted step function in the PMC trans PD.
I use a square wave with a frequency of 20mHz instead with an amplitude of 0.5V offset of 0.25V and dutycycle of 1% so there will be minimal wasted time in the off state. I get nice ringdowns (attachment 2) - forgot to take pictures. The autolocker slightly misaligns the M2 every time it is acting, so I manually align it everytime the IMC gets unlocked.
Data analysis will come later.
I remove the PD and lens and reenable the WFS servo inputs. The IMC locks easily. The WFS outputs are very different than 0 now though. |
15169
|
Tue Jan 28 19:40:15 2020 |
shruti | Update | General | PLL / PM measurement of Xend NPRO PZT |
Over the past few days, I have been trying to make measurements of the phase modulation transfer function by modulating the X end laser PZT via PLL.
The setup was modified every time during the experiment in the same manner as mentioned in elog 15148.
I could not make the PLL lock for long enough to take a proper TF measurement, resulting in TFs that look like Attachment 1. The next step would be to use the method of a delay line frequency discriminator instead of the PLL.
Comments about locking with LB1005 PI controller:
- I do not understand why the high PI corner frequency of 1kHz or 3kHz was required to lock.
- The rms level of the error signal when locked was ~100 mV, which is 25% of the total mixer range (~400 mVpp). Decreasing the gain only caused the loop to go out of lock and did not decrease this noise in the error signal.
- The setup was also partly inside the PSL enclosure, with the HEPA turned to 100%, which is probably a noisy environment for this measurement. Closing and opening the shutters or any disturbance near the enclosure resulted in movement of the beat note up to 5 MHz.
- It may have been a better idea to actuate the PSL laser instead of the X NPRO because of its larger range, but would this solve the issue with the noise?
|
Attachment 1: PMTF.pdf
|
|
Attachment 2: BeatSpectrum.pdf
|
|
15168
|
Tue Jan 28 19:12:30 2020 |
Jon | Configuration | PSL | Spare channels added to c1psl chassis |
After some discussion with Gautam, I decided to build more spare channels into the new c1psl machine. This is anticipation of adding new laser and ISS channels in the near future, to avoid having to disconnect the installed chassis and pull it out of the rack. The spare channels will be wired to DB37M feedthroughs on the front side of the chassis, with enough wire length to be able to pull the breakout boards out of the front to reconfigure their wiring as needed (e.g., split off channels onto a separate connector).
To have enough overhead, this will require installing 1 additional ADC unit (XT1221) and 1 additional DAC (XT1541). We have enough spare BIO channels among the existing units (both sinking and sourcing). This will give us:
- 13 spare ADC channels
- 14 spare DAC channels
- 16 spare sinking BIO channels
- 12 spare sourcing BIO channels
The updated c1psl chassis wiring assignments are attached. It adds 4 new DB37M connectors for the spare channels (highlighted in yellow) and fixes one typo Jordan found while wiring today. The most current spreadsheet is available here. |
Attachment 1: c1psl_feedthrough_wiring_v2.pdf
|
|
15167
|
Tue Jan 28 17:36:45 2020 |
gautam | Configuration | Computers | Local EPICS7.0 installed on megatron |
[Jon, gautam]
We found that the caput commands were taking much longer to execute on megatron than on pianosa (for example). Suspecting that this had something to do with the fact that megatron was using EPICS binaries from the shared NFS drive which were compiled for a much older OS, I installed the latest stable release of EPICS on megatron. The new caput commands execute much faster. I also added the local EPICS directory to the head of the $PATH variable used by the MC autolocker and FSS Slow scripts, so that they use the new caput command. But mcup is still slow - maybe my new path definition isn't picked up and it is still using the NFS binaries? To be looked into...
Quote: |
There were a bunch of medm processes stalled on megatron (connected with screenshot taking). To see if they were interfering with the other scripts, I killed all of the medm processes, and commented out the line in the crontab that runs the screenshots every 10 mins. Let's see if this improves stability.
|
|
15165
|
Tue Jan 28 16:01:17 2020 |
gautam | Update | IOO | IMC WFS servos stable again |
With all of the shaking (man-made and divine), it was a hard to debug this problem. Summary of fixes:
- The beam was misaligned on the WFS 1 and 2 heads, as well as the MC2 trans QPD. I re-aligned the former with the IMC unlocked, the latter (see Attachment) with the IMC locked (but the MC2 spot centering loops disabled).
- I reset the WFS DC and RF offsets, as well as the QPD offsets (once I had hand-aligned the IMC mirrors to obtain good transmission).
At least the DC indicators are telling me that the IMC locking is back to a somewhat stable state. I have not yet checked the frequency noise / RIN. |
Attachment 1: QPD_recenter.png
|
|
15164
|
Tue Jan 28 15:39:04 2020 |
gautam | Configuration | Computers | Sluggish megatron? |
There were a bunch of medm processes stalled on megatron (connected with screenshot taking). To see if they were interfering with the other scripts, I killed all of the medm processes, and commented out the line in the crontab that runs the screenshots every 10 mins. Let's see if this improves stability. |
15163
|
Tue Jan 28 14:33:24 2020 |
gautam | Update | PSL | Inferred free-running frequency noise |
To conclude my PMC noise investigations: Attachment #1 shows the PMC noise inferred from the calibrations earlier in this thread and the fitted OLTF for the PMC loop. Attachment #2 compares the frequency noise (inferred from the error point of the PMC servo) when the IMC is locked / unlocked. I don't know what to make of the fact that the PMC suggests improvement from ~20 Hz onwards already - does this mean that the NPRO noise model is wrong by 1 order of magnitude at 30 Hz?
- The IMC was locked for the measurement shown in Attachment #1.
- The in-loop spectra of the error (at the I/F output of the mixer) and control (at TP3) signals were measured with the SR785.
- The control signal voltage monitors don't seem to work - neither the front panel LEMO nor the signals hooked up to the CDS system show me sensible shapes for the spectra between 1-3 Hz.
- To convert in loop to free-running, I multiplied the measured error (control) signal spectra by
( ), where L is the OLTF. THe control signal was pre-processed by multiplying by a pole at 11.3 Hz, corresponding to the LPF formed by the 63.3 kohm series resistor and the 225 nF PZT capacitance.
- The "NPRO noise model" curve is 10^4/f Hz/rtHz.
While I initially thought the 1/f^2 rise below ~100 Hz is attributable to the IMC cavity length fluctuations, I found that this profile is present even in the measurement with the PSL shutter closed. I am not embarking on a detailed PMC noise budgeting project for now. Note however that we are not shot noise limited anywhere in this measurement band.
- The measured TF (dots in Attachment #5) was fit with LISO (solid lines in Attachment #5) to allow inferring the out-of-loop servo noise by monitoring the in-loop noise (that plot to follow).
|
|
Attachment 1: inLoopNoise_IMClocked.pdf
|
|
Attachment 2: freqNoiseComparison.pdf
|
|
15162
|
Tue Jan 28 08:26:53 2020 |
rana | Frogs | PEM | shaking |
https://breakthrough.caltech.edu/magazine/2019-aug/#article-Listening-with-Light

|
15161
|
Mon Jan 27 21:48:49 2020 |
gautam | Update | PSL | Ringdown measurements |
It's fine to block the WFS while doing ringdowns but please return the config to normal so I don't have to spend time every night recovering the interferometer before doing the locking. As I mention in that post, it is possible to do this in a non-invasive way without having to run any extra cables / permanently block any beams. If there is some issue with the data quality, then we can consider a new setup. But I see no reason to re-invent the wheel.
The IMC was also massively misaligned. I had to re-align both MC1 and MC2 to recover the lock. I took this opportunity to reset the WFS offsets. Please do not disturb the alignment of the existing optical layout unless you verify that everything is working as it should be after your changes.
And for whatever reason, ITMX was misaligned. If you do something with the interferometer, no matter how minor it seems, please leave a note on the ELOG. It will save many painful debugging hours.
As I fix these, the seismic activity has gone up . I'll wait around for an hour, but not an encouraging restart to the locking 😢
Quote: |
Zeroth order IMC ringdown
Following Gautam's IMC ringdown setup, I took the the REFL PD form the PMC ringdown experiment and installed it in the IMC REFL path blocking WFS2 (Attachment 1).
|
|
Attachment 1: elevatedSeis.pdf
|
|
15160
|
Mon Jan 27 21:35:06 2020 |
Yehonathan | Update | PSL | Ringdown measurements |
Zeroth order IMC ringdown setup
Following Gautam's IMC ringdown setup, I took the REFL PD form the PMC ringdown experiment and installed it in the IMC REFL path blocking WFS2 (Attachment 1).
I also ran a BNC cable from the transmission PD that Gautam installed on the IMC table to the vertex where the signals are measured on the scope.
I offloaded the WFS servo output values onto the MC alignment (using the WFS servo relief script) so that its dc values would be correct when the servo is off.
Unfortunately, it seems like the script severely misaligned the MC mirrors at some point when the MC got unlocked. We should fix the script such that it stops when the offloading is complete.
We got the MC realigned but left it in a state where it is not locking easily. |
Attachment 1: IMC_REFL_Beam_Path.jpg
|
|
15159
|
Mon Jan 27 18:16:30 2020 |
gautam | Configuration | Computers | Sluggish megatron? |
I've also been noticing that the IMC Autolocker scripts are running rather sluggishly on Megatron recently. Some evidence - on Feb 11 2019, the time between the mcup script starting and finishing is ~10 seconds (I don't post the raw log output here to keep the elog short). However, post upgrade, the mean time is more like ~45-50 seconds. Rana mentioned he didn't install any of the modern LIGO software tools post upgrade, so maybe we are using some ancient EPICS binaries. I suspect the cron job for the burt snapshot is also just timing out due to the high latency in channel access. Rana is doing the software install on the new rossa, and once he verifies things are working, we will try implementing the same solution on megatron. The machine is an old Sun Microsystems one, but the system diagnostics don't signal any CPU timeouts or memory overflows, so I'm thinking the problem is software related...
Quote: |
The burt snapshotting is still not so reliable - for whatever reason, the number of snapshot files that actually get written looks random. For example, the 14:19 backup today got all the snaps, but 15:19 did not. There are no obvious red flags in either the cron job logs or the autoburt log files. I also don't see any clues when I run the script in a shell. It'll be good if someone can take a look at this.
|
|
15158
|
Mon Jan 27 14:01:01 2020 |
Jordan | Configuration | General | Repurposed Sorenson Power Supply |
The 24 V Sorenson (2nd from bottom) in the small rack west of 1x2 was repurposed to 12V 600 mA, and was run to a terminal block on the north side of 1X1. Cables were routed underneath 1X1 and 1X2 to the terminal blocks. 12V was then routed to the PSL table and banana clip terminals were added. |
15157
|
Sun Jan 26 14:40:55 2020 |
gautam | Update | ALS | ALS OOL noise |
In preparation for resuming IFO locking activities, I measured the ALS noise with the arm lengths locked to IR, AUX laser frequencies locked to the arm lengths. Looks promising (y-axis units are Hz/rtHz). |
Attachment 1: ALSnoise_20200126.pdf
|
|