40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 295 of 335  Not logged in ELOG logo
ID Date Author Type Category Subjectup
  9292   Fri Oct 25 19:56:58 2013 ranaUpdatePSLlaser drift monitor set up

 

 I went to re-align the beam into the PMC just now. I also tapped all the components between the laser and the PMC; nothing seems suspicious or loose.

The only problem was that someone (probably Steve or Valera) had closed down the iris just downstream of the AOM to ~1-2 mm diameter. This is much too tight! Don't leave irises closed down after aligning. An iris is not to b used as a beam dump. Getting it within a factor of 5-10 of the beam size will certainly make extra noise from clipping/scattering. After opening the iris, the reflected beam onto the PMC REFL camera is notably changed.

Not sure if this will have any effect on our worsening transmission drift, but let's see over the weekend.

I took pictures of this clipping as well as the beam position on Steve's new Retro Position Sensor, but I can't find the cable for the Olympus 570UZ. Steve, please buy a couple more USB data cables of this particular kind so that we don't have to hunt so much if one of the cryo (?) people borrows a cable.

Attachment shows PMC power levels before and after alignment. After alignment, you can see spikes from where I was tapping the mounts in the beamline. We ought to replace the U-100 mount ahead of the AOM with a Polanski

EDIT: Cryo team returns cable - receives punishments. Picture added.

Attachment 1: PMC-IRSISSS.png
PMC-IRSISSS.png
Attachment 2: PA250052.JPG
PA250052.JPG
Attachment 3: PA280044.JPG
PA280044.JPG
  9547   Fri Jan 10 15:33:02 2014 SteveUpdatePSLlaser drift monitor set up idea

this locationQuote:

Quote:

Quote:

I wonder what's drifting between the laser and the PMC? And why is it getting worse lately?

 The PMC refl is bad in pitch today, and the transmission is only 0.76, rather than our usual 0.83ish.

I did a quick, rough tweak-up of the alignment, and now we're at 0.825 in transmission.

 The PMC transmission continuously degrades. In order to see what is really drifting the laser output after PBS was sampled as shown.

 IOO pointing is drifting in pitch. I'd like to use a QPD instead of the paper target to see if the Innolite output is stable. The idea is to move temporarily IOO-QPD_POS to  this location

Attachment 1: 2daysDrift.png
2daysDrift.png
  9552   Tue Jan 14 10:12:12 2014 SteveUpdatePSLlaser drift monitor set up idea

Quote:

this locationQuote:

Quote:

Quote:

I wonder what's drifting between the laser and the PMC? And why is it getting worse lately?

 The PMC refl is bad in pitch today, and the transmission is only 0.76, rather than our usual 0.83ish.

I did a quick, rough tweak-up of the alignment, and now we're at 0.825 in transmission.

 The PMC transmission continuously degrades. In order to see what is really drifting the laser output after PBS was sampled as shown.

 IOO pointing is drifting in pitch. I'd like to use a QPD instead of the paper target to see if the Innolite output is stable. The idea is to move temporarily IOO-QPD_POS to  this location

 I do like to move IOO-QPD_POS temporarily to see that the feedback has anything to do with with the pointing.

Attachment 1: bad4thday.png
bad4thday.png
  6395   Fri Mar 9 16:00:46 2012 steveUpdateGreen Lockinglaser emergency shut down switch replaced at the south end

Over-sized local laser emergency switch was held by large C clamp at the south end. This was replaced by a smaller one and it is mounted with magnets.

The Innolight laser was turned off, while the interlock was wired.

  3108   Wed Jun 23 17:48:16 2010 steveUpdateMOPAlaser head temp

The laser chiller temp is fluctuating and the power output is decreasing. See 120 days plot.

Yesterday I removed ~300cc water from the overflowing chiller tank.

Attachment 1: htemp120d.jpg
htemp120d.jpg
  325   Wed Feb 20 11:34:17 2008 steveUpdatePSLlaser head temp is up
MOPA head temp is running at 20.3C now
Nomally it is at 18.5C
Attachment 1: htempup.jpg
htempup.jpg
  12   Wed Oct 24 08:58:09 2007 steveOtherPSLlaser headtemp is up
C1:PSL-126MOPA_HTEMP is 19.3C

Half of the chiller's air intake was covered by loose paper
Attachment 1: htempup.jpg
htempup.jpg
  3033   Wed Jun 2 07:54:55 2010 steveUpdateMOPAlaser headtemp is up

Is the cooling line clogged? The chiller temp is 21C See 1 and 20 days plots

Attachment 1: htemp.jpg
htemp.jpg
Attachment 2: htemp20d.jpg
htemp20d.jpg
  3035   Wed Jun 2 11:28:31 2010 KojiUpdateMOPAlaser headtemp is up

Last night we stopped the air conditioning. It made HDTEMP increase.
Later we restored them and the temperature slowly recovered. I don't know why the recovery was so slow.

Quote:

Is the cooling line clogged? The chiller temp is 21C See 1 and 20 days plots

 

  6245   Fri Feb 3 14:47:51 2012 steveUpdatePEMlaser interlock drawing

 Rough draft of      updated interlock drawing by Ben is here.

 

  1280   Fri Feb 6 14:49:31 2009 steveSummarySAFETYlaser inventory
40M Laser Inventory at Feb 05, 2009
 

1, Lightwave PA#102 @ 77,910 hrs 1064nm of 2.8W @ 27.65A
                   NPRO#206 @ 2.4A                    at PSL enclosure............"Big Boy" is waiting for to be retired but not now.

2, Lightwave  NPRO 1064nm of 700mW  #415   at AP table.......cavity length measurements of Alberto

3, CrystaLaser  IRCL-100-1064S, 1064nmS of 100mW  ,sn#IR81132 at east arm cabinet

4, CrystaLaser 1064nm of 180mW # -1274 flq at scattering setup.........flashlight quality

5, RF-PD tester 1064nm of 1.2mW @20mA at SP table

6, Lumix 800-1100nm of 500mW at east arm cabinet

7, JDS-Uniphase 633nmP of 4mW oplev sus laser at 5 places,
    plus four spares in east arm cabinet
 
 
The same information is posted at the 40M WIKI also
 
 
 
  3514   Thu Sep 2 16:41:32 2010 steveConfigurationSAFETYlaser is ON: safety glasses required!

I hooked up interlock to the Innolight 2W 1064 nm in the enclosure. The manual shutter  is closed on this unit.

SAFETY GLASSES REQUIRED !

  6242   Wed Feb 1 17:00:57 2012 steveUpdateIOOlaser is back ON

Quote:

 

The 2W PSL laser is turned off.  The danger laser lights are not illuminated at the entry doors because of malfunctioning electronic circuit!!!

Laser safety glasses are still required!  Other lasers are in operation!

 BEN fixed the interlock.  The laser is turned ON. Thanks for all, Rich and Sam who came over to help. Atm1

All emergency shut- off switches, lights and door indicators are working at this moment. More about this tomorrow.

Atm2, PSL enclosure interlock jungle without REAL schematic drawing.....at this point.... We all agreed it is easier to redo the hole thing than find the problem

Atm3, Emergency shut off switches and illuminated signs from  entry doors to AC on-off box  ( Use this switches in emergency ONLY,  otherwise leave alone , even it is labeled obsolete !)

Summery: I still do not really know what was wrong.

 

Attachment 1: P1080525.JPG
P1080525.JPG
Attachment 2: P1080514.JPG
P1080514.JPG
Attachment 3: P1080518.JPG
P1080518.JPG
  7172   Tue Aug 14 08:43:42 2012 SteveUpdateIOOlaser off and on

The janitor accidentally hit the laser emergency kill switch at room 103  entry door. It did shut down the PSL laser. The laser was turned back on.

Attachment 1: 1day.png
1day.png
  365   Fri Mar 7 19:04:39 2008 steveOmnistructurePSLlaser pointer
Green laser pointer was found in my desk.
I blamed Rana for not returning it to me after a conference talk.
It is surprisingly bright still.
I will bring sweets for Wednesday meeting.
  712   Tue Jul 22 09:24:17 2008 steveUpdatePSLlaser power
Laser power reality of 120 days
Attachment 1: power120d.jpg
power120d.jpg
  1547   Tue May 5 10:42:18 2009 steveUpdateMOPAlaser power is back

Quote:

As PSL-126MOPA_DTEC went up, the power out put went down yesterday

 The NPRO cooling water was clogged at the needle valve. The heat sink temp was around ~37C

The flow-regulator  needle valve position is locked with a nut and it is frozen. It is not adjustable. However Jeenne's tapping and pushing down on the plastic hardware cleared the way for the water flow.

We have to remember to replace this needle valve when the new NPRO will be swapped in. I checked on the heat sink temp this morning. It is ~18C

There is condensation on the south end of the NPRO body, I wish that the DTEC value would just a little higher like 0.5V

The wavelenght of the diode is temp dependent: 0.3 nm/C. The fine tuning of this diode is done by thermo-electric cooler ( TEC )

To keep the diode precisely tuned to the absorption of the laser gain material the diode temp is held constant using electronic feedback control.

This value is zero now.

 

Attachment 1: uncloged.jpg
uncloged.jpg
  2142   Mon Oct 26 15:40:01 2009 steveUpdatePSLlaser power is down

The laser power is down 5-6%

Attachment 1: laserpowerdown.jpg
laserpowerdown.jpg
  2147   Mon Oct 26 23:14:08 2009 KojiUpdatePSLlaser power is down

I adjusted the steerings to the PMC and gained 7%. Now the MC_TRANS 7.0 has been recovered.

Actually I need another 7% to get MC_TRANS 7.5.
But I couldn't find how I can recover 126MOPA-AMPMON to 2.8ish.

Quote:

The laser power is down 5-6%

 

Attachment 1: PSL091026.png
PSL091026.png
  1542   Mon May 4 10:38:52 2009 steveUpdateMOPAlaser power is dropped

As PSL-126MOPA_DTEC went up, the power out put went down yesterday

Attachment 1: dtecup.jpg
dtecup.jpg
  1543   Mon May 4 16:49:56 2009 AlbertoUpdateMOPAlaser power is dropped

Quote:

As PSL-126MOPA_DTEC went up, the power out put went down yesterday

Alberto, Jenne, Rob, Steve,
 
later on in the afternoon, we realized that the power from the MOPA was not recovering and we decided to hack the chiller's pipe that cools the box.
 
Without unlocking the safety nut on the water valve inside the box, Jenne performed some Voodoo and twisted a bit the screw that opens it with a screw driver. All the sudden some devilish bubbling was heard coming from the pipes.
The exorcism must have freed some Sumerian ghost stuck in our MOPA's chilling pipes (we have strong reasons to believe it might have looked like this) because then the NPRO's radiator started getting cooler.
I also jiggled a bit with the valve while I was trying to unlock the safety nut, but I stopped when I noticed that the nut was stuck to the plastic support it is mounted on.
 
We're now watching the MOPA power's monitor to see if eventually all the tinkering succeeded.

 

[From Jenne:  When we first opened up the MOPA box, the NPRO's cooling fins were HOT.  This is a clear sign of something badbadbad.  They should be COLD to the touch (cooler than room temp).  After jiggling the needle valve, and hearing the water-rushing sounds, the NPRO radiator fins started getting cooler.  After ~10min or so, they were once again cool to the touch.  Good news.  It was a little worrisome however that just after our needle-valve machinations, the DTEC was going down (good), but the HTEMP started to rise again (bad).  It wasn't until after Alberto's tinkering that the HTEMP actually started to go down, and the power started to go up.  This is probably a lot to do with the fact that these temperature things have a fairly long time constant. 

Also, when we first went out to check on things, there was a lot more condensation on the water tubes/connections than I have seen before.  On the outside of the MOPA box, at the metal connectors where the water pipes are connected to the box, there was actually a little puddle, ~1cm diameter, of water. Steve didn't seem concerned, and we dried it off.  It's probably just more humid than usual today, but it might be something to check up on later.]

  3202   Tue Jul 13 10:02:30 2010 steveUpdateMOPAlaser power is dropping slowly

I have just removed an other 400 cc of water from the chiller.  I have been doing this since the HTEMP started fluctuating.

The Neslab bath temp is 20.7C, control room temp 71F

 

Attachment 1: power100d.jpg
power100d.jpg
  335   Fri Feb 22 14:45:06 2008 steveUpdateMOPAlaser power levels

At the beginning of this 1000 days plot shows the laser that was running at 22C head temp
and it was send to LLO

The laser from LHO PA#102 with NPRO#206 were installed at Nov. 29, 2005 @ 49,943 hrs
Now,almost 20,000 hrs later we have 50% less PSL-126MOPA_AMPMON power
Attachment 1: lpower1000d.jpg
lpower1000d.jpg
  865   Thu Aug 21 10:24:20 2008 steveConfigurationSAFETYlaser safe mode condition
The MOPA and PSL shutters are closed.
Manual beam blocks are in place.
Enclosure interlock is enabled.
No other high power laser is in operation.

We are in laser safe of operation for visiting students from Japan

NO safety glasses required
  6347   Fri Mar 2 16:05:52 2012 DenUpdateSAFETYlaser safety

Today I've attended the laser safety seminar.

  12921   Fri Mar 31 10:16:07 2017 SteveUpdatesafetylaser safety glasses annual inspection

Laser safety glasses cleaned in " Dawn Ultra " mild soap - water solution and measured for 1064 nm transmission at 150 mW
 

Attachment 1: march2017.jpg
march2017.jpg
Attachment 2: 2017march.jpg
2017march.jpg
  8287   Wed Mar 13 16:04:24 2013 steveUpdateSAFETYlaser safety glasses checked

 All safety glasses were cleaned in soapy water by Bob. I measured their transmission at 1064 nm, 150 mW,  beam diameter 1.5 mm  They are in working order, no transmission.

 Inventory:

10 pieces of KG-5, fit over, from Laser Safety

 4 pieces of KG-5, std size, from Drever Lab, best visibility

 1 piece of KG-5 coated for visible, std size, from Kentek

15 pieces of green-plastic LOTG-YAG, fit over, from UVEX

 7 pieces of green-plastic B-D+S 137, std areo fit, from Sperian

 3 pieces of green-plastic, old Thorlab, fit over

 2 pieces of green-plastic, fit over, from Laservision

 8 pieces of braun- plastic, fit over, for green & IR protection, from UVEX & Thorlabs

Attachment 1: safetyglasseschecked.jpg
safetyglasseschecked.jpg
  203   Wed Dec 19 16:40:12 2007 steveUpdateSAFETYlaser safety glasses measured
I measured the coarse transission at 1064nm of the 40m safety glasses today.

12 pieces of UVEX # LOTG-YAG/CO2 light green, all plastic construction, ADSORBANT

3 pieces of 6KG5, Scott colored filter glass type,

individual prescription glasses: alan, bob, ben, jay and steve

7 pieces of dual waveleght glasses

These glasses showed 0.00mW transmission out of 170mW Crysta Laser 1064
  6239   Tue Jan 31 08:44:10 2012 steveUpdateIOOlaser shuts down

 

 The 2W Innilight shutdown shut when I opened side door for safety scan. This was not a repeatable by opening -closing side doors later on. Turned laser on, locked PMC and MC locked instantly. The MC was not locked this moring and it seemed that the MC2 spot was still some high order mode

like yesterday. MC lock was lost when the janitor bumped something around the MC.

Attachment 1: laseroff.png
laseroff.png
  6240   Tue Jan 31 14:58:30 2012 kiwamuUpdateIOOlaser shuts down

[Steve/ Kiwamu]

 We found that the laser had completely shut off for ~ 4 hours even with all the PSL doors closed.

We are guessing it is related to the interlock system and Steve is working on it to fix it.

Quote from #6239

 The 2W Innilight shutdown shut when I opened side door for safety scan. This was not a repeatable by opening -closing side doors later on. Turned laser on, locked PMC and MC locked instantly. The MC was not locked this moring and it seemed that the MC2 spot was still some high order mode

like yesterday. MC lock was lost when the janitor bumped something around the MC.

 

  4614   Tue May 3 15:48:26 2011 Larisa ThorneUpdateElectronicslaser temperature control LPF, final version!

This is a continuation of this

 The low pass filter is finally acceptable, and its Bode graph is below (on a ~3Hz frequency span that shows the cutoff frequency is at 0.1Hz)

Attachment 1: LPF100k.2.jpg
LPF100k.2.jpg
  9814   Tue Apr 15 13:24:42 2014 SteveUpdatePSLlaser turned on

The 2W Innolight was off for 4 hours.

Attachment 1: laserOff4hrs.png
laserOff4hrs.png
Attachment 2: laserShutsDown.png
laserShutsDown.png
  11373   Wed Jun 24 08:05:43 2015 SteveUpdatePSLlaser turned on

The laser went off around 11am yesterday. It was turned on

Attachment 1: laserWasOff.png
laserWasOff.png
  7853   Tue Dec 18 16:37:40 2012 SteveUpdateVAClast RGA scan before vent

 

 

Attachment 1: pd74m36d.png
pd74m36d.png
Attachment 2: 36d_vacuum_normal.png
36d_vacuum_normal.png
  1009   Tue Sep 30 13:43:43 2008 robUpdateLockinglast night
Steady progress again in locking again last night. Initial acquisition of DRMI+2ARMs was working well.
Short DOF handoff, CARM->MCL, AO on PO_DC, and power ramping all worked repeatedly, in the cm_step script.
This takes us to the point where the common mode servo is handed off to an RF signal and the CARM offset
is reduced to zero. This last step didn't work, but it should just require some tweaking of the gains
during the handoff.
  1011   Wed Oct 1 00:24:54 2008 ranaUpdateLockinglast night
I had mistakenly left the MC boost off during my FAST investigations. The script is now restored.

The ISS is still saturating with gains higher than -5 dB. We need to request a PeterK / Stefan consult in the morning.

Also found the MZ gain down at -10 dB around midnight - need an alarm on that value.
  1024   Fri Oct 3 15:57:05 2008 robUpdateLockinglast night, again
Last night was basically a repeat of the night before--marginally better locking with the DRMI resonating the +f2
sideband. Several stable locks were achieved, and several control handoffs to DDM signals worked, but never from
lock to lock--that is, a given DD handoff strategy would only work once. This really needs to work smoothly before
more progress can be made.

Also, a 24Hz mode got rung up in one/several of the suspensions--this can also impede the stability of locks.
  6703   Tue May 29 15:29:16 2012 JamieUpdateComputerslatest pynds installed on all new control room machines

The DASWG lscsoft package repositories have a lot of useful analysis software.  It is all maintained for Debian "sqeeze", but it's mostly installable without modification on Ubuntu 10.04 "lucid" (which is based on Debian squeeze).  Basically the only thing that needs to access the lscsoft repositories is to add the following repository file:

controls@rossa:~ 0$ cat /etc/apt/sources.list.d/lscsoft.list 
deb http://www.lsc-group.phys.uwm.edu/daswg/download/software/debian/ squeeze contrib
deb-src http://www.lsc-group.phys.uwm.edu/daswg/download/software/debian/ squeeze contrib

deb http://www.lsc-group.phys.uwm.edu/daswg/download/software/debian/ squeeze-proposed contrib
deb-src http://www.lsc-group.phys.uwm.edu/daswg/download/software/debian/ squeeze-proposed contrib
controls@rossa:~ 0$ 

A simple "apt-get update" then makes all the lscsoft packages available.

lscsoft includes the nds2 client packages (nds2-client-lib) and pynds (python-pynds).  Unfortunately the python-pynds debian squeeze package currently depends on libboost-python1.42, which is not available in Ubuntu lucid.  Fortunately, pynds itself does not require the latest version and can use what's in lucid.  I therefore rebuilt the pynds package on one of the control room machines:

$ apt-get install dpkg-dev devscripts debhelper            # these are packages needed to build a debian/ubuntu package
$ apt-get source python-pynds                              # this downloads the source of the package, and prepares it for a package build
$ cd python-pynds-0.7
$ debuild -uc -us                                          # this actually builds the package
$ ls -al ../python-pynds_0.7-lscsoft1+squeeze1_amd64.deb
-rw-r--r-- 1 controls controls 69210 2012-05-29 11:57 python-pynds_0.7-lscsoft1+squeeze1_amd64.deb

I then copied the package into a common place:

/ligo/apps/debs/python-pynds_0.7-lscsoft1+squeeze1_amd64.deb

I then installed it on all the control room machines as such:

$ sudo apt-get install libboost-python1.40.0 nds2-client-lib python-numpy   # these are the dependencies of python-pynds
$ sudo dpkg -i /ligo/apps/debs/python-pynds_0.7-lscsoft1+squeeze1_amd64.deb

I did this on all the control room machines.

It looks like the next version of pynds won't require us to jump through these extra hoops and should "just work".

  3024   Tue Jun 1 11:47:14 2010 steveUpdatePEMlead balls on concrete

Quote:

Valera and I put the 2 Guralps and the Ranger onto the big granite slab and then put the new big yellow foam box on top of it.

There is a problem with the setup. I believe that the lead balls under the slab are not sitting right. We need to cut out the tile so the thing sits directly on some steel inserts.

You can see from the dataviewer trend that the horizontal directions got a lot noisier as soon as we put the things on the slab.

 The tiles were cut out in 1.5" ID circle to insure that the 7/16" OD lead balls would not touch the tiles on Wednesday, May 26, 2010

Granite surface plate specifications: grade B, 18" x 24" x 3" , 139 lbs

These balls and granite plate were removed by  Rana in entry log #3018 at 5-31-2010

Attachment 1: P1060269.JPG
P1060269.JPG
  3060   Wed Jun 9 19:47:08 2010 nancyUpdatePEMlead balls on concrete

Quote:

Quote:

Valera and I put the 2 Guralps and the Ranger onto the big granite slab and then put the new big yellow foam box on top of it.

There is a problem with the setup. I believe that the lead balls under the slab are not sitting right. We need to cut out the tile so the thing sits directly on some steel inserts.

You can see from the dataviewer trend that the horizontal directions got a lot noisier as soon as we put the things on the slab.

 The tiles were cut out in 1.5" ID circle to insure that the 7/16" OD lead balls would not touch the tiles on Wednesday, May 26, 2010

Granite surface plate specifications: grade B, 18" x 24" x 3" , 139 lbs

These balls and granite plate were removed by  Rana in entry log #3018 at 5-31-2010

 I tried to calculate the frequency of resonance using Rayleigh's method.  approximated the geometry of lead to be that of a perfect cylinder, and  the deformation in the lead by the deflection in a cantilever under  a shear strain.

this rough calculation gives an answer of 170Hz and depends on the dimensions of each lead, number of leads, and mass of the granite. But the flaw pointed out is that this calculation doesnot depend on the dimension of the granite slab, nor on the exact placing of the lead spheres with respect toteh COM of the slab.

I will put up the calculations details later, and also try to do a FEM analysis of the problem.

 

BTW, latex launched this new thing for writing pdfs. doesnot require any installations.  check  http://docs.latexlab.org

  14434   Tue Feb 5 10:11:30 2019 gautamUpdateVACleak tests complete, pumpdown 83 resumed

I guess we forgot to close V5, so we were indeed pumping on the ITMY and ETMY annuli, but the other three were isolated suggest a leak rate of ~200-300 mtorr/day, see Attachment #1 (consistent with my earlier post).

As for the main volume - according to CC1, the pressure saturates at ~250 uTorr and is stable, while the Pirani P1a reports ~100x that pressure. I guess the cold-cathode gauge is supposed to be more accurate at low pressures, but how well do we believe the calibration on either gauge? Either ways, based on last night's test (see Attachment #2), we can set an upper limit of 12 mtorr/day. This is 2-3x the number Steve said is normal, but perhaps this is down to the fact that the outgassing from the main volume is higher immediately after a vent and in-chamber work. It is also 5x lower rate of pressure increase than what was observed on Feb 2.

I am resuming the pumping down with the turbo-pumps, let's see how long we take to get down to the nominal operating pressure of 8e-6 torr, it ususally takes ~ 1 week. V1, VASV, VASE and VABS were opened at 1030am PST. Per Chub's request (see #14435), I ran RP1 and RP3 for ~30 seconds, he will check if the oil level has changed.

Quote:
 

Let's leave things in this state overnight - V1 and V5 closed so that neither the main volume nor the annuli are being pumped, and get some baseline numbers for what the outgassing rate is.

Attachment 1: Annuli.png
Annuli.png
Attachment 2: MainVol.png
MainVol.png
  5759   Fri Oct 28 18:33:59 2011 steveUpdateVACleaking nitrogen line fixed

I was lucky to notice that the nitrogen supply line to the vacuum valves was leaking. Closed ALL valves. Open supply line to atm. Fixed leak. 

This was done fast so the pumps did not have to be shut down. Pressurized supply line and open valves to

"Vac Normal" condition in the right sequence.

Attachment 1: allvalvesclosed.png
allvalvesclosed.png
  10790   Fri Dec 12 10:38:22 2014 SteveUpdatePEMleaky roof

The first real rain of this year finds only one leak at the 40m

Attachment 1: leakyroof.jpg
leakyroof.jpg
  12031   Fri Mar 11 16:52:53 2016 SteveUpdatePEMleaky roof

Johannes found dripping water at the vac rack. It is safe. It is not catching anything. Actual precipitation was only 0.62"

Attachment 1: leakyRoof.jpg
leakyRoof.jpg
Attachment 2: leakyRoofA.jpg
leakyRoofA.jpg
  12087   Fri Apr 22 13:58:13 2016 SteveUpdatePEMleaky roof is fixed

Dan sealed the leak today.

 

Attachment 1: leakyRoof_(2).jpg
leakyRoof_(2).jpg
  2706   Wed Mar 24 03:58:18 2010 kiwamu, matt, kojiUpdateGreen Lockingleave PLL locked

We are leaving the PLL as it is locked in order to see the long term stability. And we will check the results in early morning of tomorrow.

DO NOT disturb our PLL !!

  


(what we did)

After Mott left, Matt and I started to put feedback signals to the temperature control of NPRO.

During doing some trials Matt found that NPRO temperature control input has an input resistance of 10kOhm.

Then we put a flat filter ( just a voltage divider made by a resistor of ~300kOhm and the input impedance ) with a gain of 0.03 for the temperature control to inject a relatively small signal, and we could get the lock with the pzt feedback and it.

In addition, to obtain more stable lock we then also tried to put an integration filter which can have more gain below 0.5Hz.

After some iterations we finally made a right filter which is shown in the attached picture and succeeded in obtaining stable lock.

 

 

 

Attachment 1: DSC_1402.JPG
DSC_1402.JPG
  2712   Wed Mar 24 15:59:59 2010 kiwamu, mattUpdateGreen Lockingleave PLL locked

Matt checked it in this morning and he found it's been locked during the night.

 

 

  5582   Fri Sep 30 05:35:42 2011 kiwamuUpdateLSClength fluctuations in MICH and PRCL

The MICH and PRCL motions have been measured in some different configurations.

According to the measurements :

      + PRCL is always noisier than MICH.

      + MICH motion becomes noisier when the configuration is Power-Recycled Michelson (PRMI).

The next actions are :

      + check the ASPD

      + check the demodulation phases

      + try different RFPDs to lock MICH

 


(Motivation)
 The lock of PRMI have been unstable for some reason.
One thing we wanted to check was the length fluctuations in MICH and PRCL.


(Measurement)
Four kinds of configuration were applied.
     (1) Power-recycled ITMX (PR-ITMX) locked with REFL33_I, acting on PRM.
     (2) Power-recycled ITMY (PR-ITMY) locked with REFL33_I, acting on PRM.
     (3) Michelson locked with AS55_Q, acting on BS.
     (4) Power-recycled Michelson locked with REFL33_I and AS55_Q, acting on PRM and BS.

In each configuration the spectrum of the length control signal was measured.
With the measured spectra the length motions were estimated by simply multiplying the actuator transfer function.
Therefore the resultant spectra are valid below the UGFs which were at about 200 Hz.
The BS and PRM actuator responses had been well-measured at AC (50 - 1000 Hz)
For the low frequency responses they were assumed to have the resonances at 1 Hz with Q of 5.
 

(Results)
The below plot shows the length noise spectra of four different configurations.
There are two things which we can easily notice from the plot.
    + PRCL (including the usual PRCL and PR-ITMs) is always noisier than MICH.
    + MICH became noisier when the power recycling was applied.
In addition to them, the MICH noise spectrum tended to have higher 3 Hz bump as the alignment gets improved.
In fact everytime when we tried to perfectly align PRMI it eventually unlocked.
I am suspecting that something funny (or stupid) is going on with the MICH control rather than the PRCL control.

noise.png

(Notes)
   BS actuator = 2.190150e-08 / f2
   PRM actuator = 2.022459e-08 /  f2
  5584   Fri Sep 30 08:40:02 2011 KojiUpdateLSClength fluctuations in MICH and PRCL

Tip-Tilts has almost no isolation up to 3Hz, and isolation of about 0.5 up to 10Hz.
They have vertical resonances at around 20Hz.

See Nicole's entry

Quote:

noise.png

 

  5638   Sat Oct 8 04:41:07 2011 kiwamuUpdateLSClength fluctuations in SRCL

For a comparison, the length fluctuation of Signal-Recycled ITMX (SRX) and ITMY (SRY) have been measured.

Roughly speaking the length motion of SRX and SRY are as loud as that of PRCL.

Some details about the measurement and data analysis can be found in the past elog entry (#5582).

In the process of converting the raw spectra to the calibrated displacements the SRM actuator was assumed to have a resonance at 1Hz with Q = 5.

length_noise.png

(Notes on SRX/Y locking)

     Sensor = REFL11_I
     Actuator = SRM
     Demod. phase = 40 deg
     SRCL_GAIN = 20
     UGF = 100 - 200 Hz
     Resonant condition = Carrier resonance
     Whitening gain = 0 dB
     ASDC = 360 counts

Quote from #5582

The MICH and PRCL motions have been measured in some different configurations.

      + PRCL is always noisier than MICH.

ELOG V3.1.3-