40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 292 of 339  Not logged in ELOG logo
ID Date Author Type Category Subjectup
  1418   Mon Mar 23 15:50:44 2009 ranaConfigurationLSCfilters deleted, lsc rebooted

The LSC time had gone too high. I deleted ~20 filters and rebooted. CPU time came down to 50 usec.

The filters all looked like old trash to me, but its possible they were used.

I didn't delete anything from the DARM, CARM, etc. banks but did from the PD and TM filter banks. You can always go back in time by using the

filter_archive/

  5355   Wed Sep 7 08:14:01 2011 steveUpdateSUSfinal OSEM check

All fine, except ITMX_sensor_UL's  60  counts deep hoop for an hour.

Attachment 1: finalcheck.jpg
finalcheck.jpg
Attachment 2: ITMX10min.jpg
ITMX10min.jpg
Attachment 3: finalsum.png
finalsum.png
  13704   Mon Mar 26 16:10:33 2018 KiraUpdatePEMfinal setup sketch

I made sketches of the final setup. There will be a box in the rack that contains both the heater circuit and the temperature sensor boards. One of them is in the loop while the other isn't. Instead of having many cables leading to the can, there will only be these three, though they can be made into a single wire. It will be connected to the can through a D-9 connector. The second attachment is what will be inside of the box, with all the major wires and components labeled.

-----

Edit: I've canged the layout to (hopefully) make the labels easier to read. I've also added in a cable to the ADC that reads out the voltage across the 1 ohm resistor. I also attached the circuit diagrams for the heater circuit and the temperature sensors. The one for the heater circuit was made by Kevin and I used the same design, except I have LM7818 and LM7918, since the 15V ones were not available at the time I made the circuit. 

In addition, all the wires leading to the can will all be part of one bundle of wires (I didn't clearly indicate it as such). There will be a total of 6 wires: two are needed for the wire to supply power to the heater and will have a LEMO connector on the rack end and two are needed for each temperature sensor, which will be attached to the board directly on the rack end. 

Also, we don't need two voltage regulators for each temperature circuit. We can just have one of each of LM7815 and LM7915 to supply +/- 15V to the boards.

Attachment 1: heater_1_new.png
heater_1_new.png
Attachment 2: heater_2_new.png
heater_2_new.png
Attachment 3: HeaterCircuit.pdf
HeaterCircuit.pdf
Attachment 4: temp_sensor.png
temp_sensor.png
  13759   Wed Apr 18 12:18:39 2018 KiraUpdatePEMfinal setup sketch

I've updated the sketches and added in front panels for the seismometer block and the 1U panel (attachments 3 and 4). There was an issue when it came to the panel on the block because the hole is only big enough for the cable that already exists there and there is no space to add in the D-9 connector. Not quite sure how to resolve this issue. Attachment 7 is the current panel on the seismometer block. Attachments 5 and 6 are the updated temperature circuit and the heater circuit.

The boxes will be located in the short racks at EX and EY to minimize cable length.

Attachment 1: heater_1_new.png
heater_1_new.png
Attachment 2: heater_2_new.png
heater_2_new.png
Attachment 3: 1U-panel.pdf
1U-panel.pdf
Attachment 4: EX-can-panel.pdf
EX-can-panel.pdf
Attachment 5: IMG_20180412_120427.jpg
IMG_20180412_120427.jpg
Attachment 6: HeaterCircuit.pdf
HeaterCircuit.pdf
Attachment 7: IMG_20180418_121115.jpg
IMG_20180418_121115.jpg
  13782   Tue Apr 24 09:10:20 2018 KiraUpdatePEMfinal setup sketch

I've attached the final sketch for the panel on the granite block.

Attachment 1: EX-can-panel_1.pdf
EX-can-panel_1.pdf
  13800   Mon Apr 30 15:36:18 2018 KiraUpdatePEMfinal setup sketch

I've attached a sketch of how the panel will be mounted. We should make a small rectangular box that would raise the panel from the block by 1 cm or so to allow the cables to fit into the hole in the block without getting bent. It also has to be airtight so maybe having a thin layer of rubber between the mount and block would be good.

Attachment 1: mount.png
mount.png
  13769   Thu Apr 19 12:23:30 2018 KiraUpdatePEMfinal setup sketch update

I've added in the dimensions to my sketch.

It seems like placing the two connectors right next to each other would allow both cables to just barely go through the hole in the block.

Quote:

Can you please add dimensions to the drawing, so we can see if things fit and what the cable lenghts need to be?

For the panel on the granite slab, we should use a thinner piece of metal and mount it with an offset so that the D-sub cable can be fished through the hole in the slab. The hole is wide enough for 2 cables, but not 2 connectors.

 

Attachment 1: heater_1_new.png
heater_1_new.png
Attachment 2: heater_2_new.png
heater_2_new.png
  13771   Thu Apr 19 18:23:51 2018 KiraUpdatePEMfinal setup sketch update

since we're just going from the short rack (not the tall rack) to the seismometer, can't we use a cable shorter than 45' ?

Quote:

I've added in the dimensions to my sketch.

the panel should be completely replaced like I described. We don't want to try to squeeze it in artificially and torque the wires. It just needs to be separated from the slab by a few more cm.

  13774   Fri Apr 20 15:07:45 2018 KiraUpdatePEMfinal setup sketch update

If we lay the cable along the floor then it should be around 6' to the current setup and about 20' to the actual seismometer.

Edit: 16 gauge wire should be good.

Quote:

since we're just going from the short rack (not the tall rack) to the seismometer, can't we use a cable shorter than 45' ?

 

  13760   Wed Apr 18 16:59:35 2018 ranaUpdatePEMfinal setup sketch: EX Seis

Can you please add dimensions to the drawing, so we can see if things fit and what the cable lenghts need to be?

For the panel on the granite slab, we should use a thinner piece of metal and mount it with an offset so that the D-sub cable can be fished through the hole in the slab. The hole is wide enough for 2 cables, but not 2 connectors.


Attached is a 8-day minute trend of the heater control signals, as well as the in-loop temperature sensor (which underestimates the true fluctuations; we really need an out-of-loop sensor attached to the can or seismometer).

You can see that since the last tuning (on the 13th), its been stable at the set point of 39 C with 8.5 - 10 W of heating power. Need to add the PID loop settings (all the sliders on the MEDM screen) to the frames so that we can help in diagnosing. Also, fix the spelling of "Celcisususs".

Attachment 1: Screen_Shot_2018-04-18_at_5.20.53_PM.png
Screen_Shot_2018-04-18_at_5.20.53_PM.png
  4925   Thu Jun 30 21:00:20 2011 kiwamuUpdateABSLfine alignment done

(Just a quick report)

The fine alignment of the ABSL laser injection was successfully done.

I was able to see the DRMI fringings at the AS camera. The ABSL beam is injected from the AS port, therefore what I saw on the camera was the reflection back from the interferometer.

 

(Things to be done)

 -  A beat-note setup on the PSL table.

 - Refinement of the mode matching. The beam spot on the AS camera is a bit bigger, so I should more tightly focus the injected beam.

Quote from #4917

 - Fine alignments will be done tomorrow

  6230   Fri Jan 27 05:21:43 2012 kiwamuUpdateGreen Lockingfine alignment of the Y end green setup

I did a fine alignment on the Y end green setup. The green light became able to be locked again.

Quote from #6227

 The alignment is finished after the realization that the 3rd steering mirror had to be adjusted too.

  13979   Mon Jun 18 11:12:23 2018 KiraSummaryPEMfinishing up work at the lab

Since I am finishing my job at the lab, I have stored all my electronics in a box (attachment 1) and placed it under the table in the control room where some other electronics are stored. The box contains the heater circuit box, two temperature sensor boards, one temperature sensor, a short power cable and +/- 15V supply cables. In the lab I left the wires for the current setup and tied them down to the wall so that they aren't in the way (attachment 2). I left the can as is and the other temperature sensor is still attached to the inside of the can. I have labeled the wires going from the sensor as 'in' and 'out'. I've also left the wires for the heater there as well (attachment 3). I turned off the PID control and deactivated the tmux session on megatron.

Thanks to Rana and the LIGO team for giving me the opportunity to work at the 40m on this project with the seismometer.

Attachment 1: IMG_20180618_101640.jpg
IMG_20180618_101640.jpg
Attachment 2: IMG_20180618_093920.jpg
IMG_20180618_093920.jpg
Attachment 3: IMG_20180618_093932.jpg
IMG_20180618_093932.jpg
  11976   Thu Feb 4 10:19:05 2016 SteveUpdatesafetyfire marshal inspection

Pasadena fire marshal inspected the lab today. No violation was found.

  11901   Wed Dec 23 16:15:47 2015 ranaOmnistructureALARMfire alarm

Fire alarm went off several minutes ago. Talked to security and they said there was no fire. It beeped twice again just now. No one has been working on the IFO today.

  12216   Mon Jun 27 15:26:03 2016 SteveOmnistructureALARMfire alarm test

The fire alarm came on around 15:05  for about 2-3 minutes. We all  left the lab and counted heads.  I called Paul Mackel x2646 (cell 626/ 890- 3259) at Fire Protection Services. He said that this alarm test was planned and we should of got an email notice. Perhaps I missed that notes.

Quote:

Fire alarm went off several minutes ago. Talked to security and they said there was no fire. It beeped twice again just now. No one has been working on the IFO today.

 

Attachment 1: fireAlarmTest.png
fireAlarmTest.png
  11452   Tue Jul 28 15:05:09 2015 SteveUpdatesafetyfire alarm test ? no

We just had fire alarm trigged avacuation of the 40m lab.

It turned out that the CES building second floor sensor caused this action.

Attachment 1: 6minGaps.png
6minGaps.png
  708   Mon Jul 21 15:52:22 2008 steveBureaucracySAFETYfire alarms test
The fire alarm test and evacuation drill was successfully completed at 13:45 Wednesday, July 16, 2008

Everybody was on time for the 40m meeting.
  8026   Thu Feb 7 17:24:13 2013 SteveUpdateSAFETYfire extinguishers checked

The fire department weighted and pressure checked our units today. Surprisingly they found one powder filled can. We can only use HALON  gas in the lab.

 

  9090   Fri Aug 30 08:08:29 2013 steveUpdatesafetyfire horns-flashers tested OK

 

 We had fire alarm tests  and evacuation drills at 1:30pm yesterday. All flashers and horns are functioning unbearably loud and bright including clean assembly room.

  14812   Thu Jul 25 14:28:03 2019 gautamConfigurationComputersfirewalld disabled for EPICS CA

I think rana did some more changes to this workstation to make it useful for commissioning activities - but the MEDM screens were still white blanks. The problem was that the firewalld wasn't disabled (last two steps of the KThorne setup wiki). I disabled it. Now donatella can run MEDM, ndscope and StripTool. DTT doesn't work to get online data because of a "Synchronization Error", I'm not bothering with this for now. I think Kruthi successfully demonstrated the fetching of offline data with DTT.

Attachment 1: donatellaCommissioning.png
donatellaCommissioning.png
  8661   Fri May 31 10:25:17 2013 steveUpdatesafetyfirst aid kits refilled

Quote:

 

                                                 Recommended correction list:

1,  refill- upgrade first aid boxes

2,  maintain 18" ceiling to bookshelf clearance so the ceiling fire sprinklers are not blocked: room 101

3,  label chilled water supply & return valves in IFO room

4,  calibrate bake room hoods annually

5,  update safety sign at fenced storage

 

              40m still to do list:

1,   clean and measure all safety glasses

2,   annual crane inspection is scheduled for 8am March 19, 1013

3,   make PSL encloser shelf earthquake proof

 

Do you see something that is not safe? Add it to this list please.

 

 

 

Restocked First Aid Kits Location:

Main entrance, room 100

Drill press - above N2 cylinders, room 103

Control room, next to fire extinguisher, room  102

Vertex-north wall, IFO room 104

ETMY - right on ends light switches, IFO room 104_ east end

ETMX - on vertical I-beam of crane, IFO room 104_south end

Behind 1X3 Rack, on south wall - under instrument breakers panel PC-1, IFO room 104

 

Last thing remaining to be fixed from 2013 Safety Audit:

replace book shelf with 83" height

 

 

Attachment 1: refilledFAidK.jpg
refilledFAidK.jpg
  6063   Fri Dec 2 20:16:41 2011 DenUpdatedigital noisefirst order transition

In order to verify our theory about coherence corruption in linear systems due to the line

if((new_hist < 1e-20) && (new_hist > -1e-20)) new_hist = new_hist<0 ? -1e-20: 1e-20;  

in the /opt/rtcds/caltech/c1/core/release/src/include/drv/fm10Gen.c in the iir_filter function I've changed -20 to other numbers and watched at the coherence input and output of the digital filter cheby1("LowPass", 3, 0.1, 0.5)cheby1("LowPass", 6, 1, 1.5). The sampling rate was 2K. The frequency responce of the filter presented in this figure.

freqz.pdf

The next plot shows psd and coherence of the signal for different numbers in the if-statement line : 1e-20 , 1e-25, 1e-100.

 trans.pdf

We can see that for present value coherence between input and output signals is small even for low frequencies. The psd of the output signal is also corrupted because at low frequencies it should have the same psd as input signal. For 1e-25 and 1e-100 we can see that coherence is close to 1 at low frequencies so if-statement does not work and we have a first order transition from bad to good filter performance with discontinious jump of coherence.

However, for 1e-25 and 1e-100 data is still corrupted by the round-off error. Lack of coherence for high frequencies can be explained by the fact that dtt tools use single precision for data analysis and output is too small to plot a right coherence. But the coherence is also not precisely 1 for low frequencies. Actually, it is 0.99 for this aggresive filter. We use double precision in the real-time code but still for such kinds of filters round-off error is present. As wrote Daniel Sigg for Cheby filter:  "You need a lot more digits than you may naively suspect. In the 8th order example, the output of each SOS is amplified by ~106. This regardless of the fact that the coefficients are all of order 1. If you require a level of 10-3 attenuation in the stop band, you would have lost 9 digits already. Then, add the fact that you have to do of order 104 subtractions to get from 16kHz to 12Hz, loosing another ~2 digits. On top, the high Q section is probably 10 worse than the others and you lost 12 digits. In a real example this may stack up even worse."

Next we need to figure out what effects does round-off error introduce in the performance of the interferometer.

  8238   Wed Mar 6 08:40:03 2013 SteveUpdateVACfirst scan of this pumpdown

 Pumpdown 75 - Maglev - day 12

Precondition: 66 days at atm, installed TIP-TILTs with coils that  replaced PZT-Jena input steering

 

Attachment 1: pd75m12d.png
pd75m12d.png
Attachment 2: pd75mRGA12d.png
pd75mRGA12d.png
  14329   Sun Dec 2 19:32:35 2018 ranaUpdateIOOfit times

need to vary start/stop times in fit to test for systematics

  11439   Thu Jul 23 15:41:41 2015 SteveUpdatePEMfitting ants with TERRO

Pasadena got 0.2" of  rain on Saturday. Temperatures were high with high humidity since than.The ants were back in the Control room east side benches.

We have started using TERRO Liquid Ant Baits in January 2015   This worked very well to this point.

Tree new packages were opened  yesterday and the ants are gone.

We can conclude that these baits must be replaced after 6 mounts.

The liquid baits contains BORAX  and it is safe.

 

  3478   Fri Aug 27 13:41:02 2010 kiwamuUpdateSUSfix watchdogs

 [Joe, Kiwamu]

We found that the vertex watchdogs were not correctly running.

After I powercycled c1susaux, the problem was fixed successfully.

 

The symptom: the watchdogs didn't disable the coil signal even when PD_VAR signals went larger than the threshold values PD_MAX_VAR.

Also we replaced the label by the correct name "C1SUSAUX" on a tag which was tied to the front end machine mounted on the new 1X5 rack.

  4691   Wed May 11 17:10:04 2011 kiwamuConfigurationElectronicsfixed : MC3 LL PD has no signal

[Valera / Kiwamu]

It was because of a loose connection. Pushing the connector solved the issue.

looseconnectionMC3.png

We really have to think about making reliable connections and strain reliefs.

Quote from #4685

Yesterday we found that MC3 OSEM LL PD did not have a sensible signal - the readback was close to zero and it was making MC move around. I disabled the PD LL so that the damping is done with just three face plus side PDs. There still no signal from MC3 LL PD today. It needs debugging.

 

  3788   Tue Oct 26 17:31:17 2010 yutaUpdateCDSfixed OPLEV stuff and MCL filters

(Joe, Yuta)

Background:

 We are currently working on getting rid of "white stuff" in MEDM screens.
 Today, we fixed OPLEV stuff, MCL filters, and time stamps.

What we did:
 1. Plugged in OPLEV cables to ADC2. (See this wiki page for wiring)

 2. Connected ADC2 and OPLEV in Simulink model and fixed MEDM screens for OPLEVs (Attached #1).

 3. Put MCL filters for BS,ITMX,ITMY,PRM,SRM.
  They don't need them, but just for getting rid of "white stuff."
  They are connected to the ground, so the outputs are always 0.

 4. Fixed "TIME_STRING"s in MEDM screens so that they show current time correctly.
  You only need to put text monitor with channel "C1:FEC-DCU_NODE_ID_TIME_STRING" into master files(DEFAULTNAME things) and run generate_master_screens.py.
  It will automatically sets DCU ID correctly!! (Great work, Joe!)
  See this wiki page for more info on making MEDM screens.

 5. Checked OPLEV for MC2 by pointing a laser pointer to QPD. (For MC2, OPLEV is just a transmission beam position monitor)
  Each quadrant looked like they are connected to the right channel numbers.

Plan:
 - figure out what C1:SUS-NAME_MODE_SW1 does and fix
 - fix Whitening, Dewhitening ON/OFF button in main MEDM screens, so that they switch every channels' filters
 - make a new screen for MC (like the old one C1IOO_ModeCleaner.adl)
 - create a new mark for new MEDM screens

Attachment 1: optlevMC2.png
optlevMC2.png
  3789   Tue Oct 26 21:27:02 2010 JenneUpdateCDSfixed OPLEV stuff and MCL filters

Since MC2_TRANS is, in fact, MC2 Transmission, and not an oplev at all (it's not red, and it's not a lever, although it does use a QPD), I propose that the name be changed to something sensical, since calling it an oplev is completely non-sensical.  The name change should happen sooner rather than later, to avoid confusion.

Quote:

(Joe, Yuta)

Background:

 Today, we fixed OPLEV stuff, MCL filters, and time stamps.

What we did:
 5. Checked OPLEV for MC2 by pointing a laser pointer to QPD. (For MC2, OPLEV is just a transmission beam position monitor)
  Each quadrant looked like they are connected to the right channel numbers.

 

  5454   Mon Sep 19 02:08:24 2011 kiwamuUpdateLSCfixed POP clipping

Actually the clipping of POP wasn't in the chamber but it was on the first lens on the optical bench.

So I repositioned the lens to avoid the clipping and now there are no clipping on POP.

Quote from #5445

We found that POP beam is clipped by the steering mirrors inside the tank.

 

  8550   Wed May 8 17:23:04 2013 JamieConfigurationCDSfixed direct IPC connection between c1als and c1mcs

As with the previous post, I eliminated and unnecessary hop through c1rfm for the c1als --> c1mcs connection for the ALS output to MC2 POS.

As a side note, we might considering piping the ALS signals into the LSC input matrix, elevating them to actual LSC error signals, which in some since they are.  It's just that right now we're routing them directly to the actuators without going through the full LSC control.

  3769   Sat Oct 23 03:36:05 2010 yutaUpdateCDSfixed filters for C1SUS, C1RMS, C1MCS

(Joe, Yuta)

Summary:

 This Monday, MC suspension damping got something wrong.
 We started to check filters and found that digital filters were wrong because of mis-conversion from old filter files to new files.
 We converted the file again, and with mutual understanding between Foton and us, we finally got correct filters(I hope!).

What we did:
 1. Merged filter files in old /cvs/cds/caltech/chans/ directory into C1SUS.txt(BS,ITMX,ITMY), C1RMS.txt(PRM,SRM), C1MCS.txt(MC123).

 2. Rebuilt the RT models in order to get a correct filter file header(they have list of filter modules).

 3. Concatenate the header with filter design part which we got from step1.

 4. Replaced 'N 2048' with 'N 16384'
   It replaces sampling rate of "XXSEN"s.

Basically, step1-4 was the same with what we did last time. We didn't changed the fitler coefficients, so Foton somehow changed the original filter design.
So, this time, we

 5. Deleted coefficients like we did on Tuesday (see elog #3774).

But Foton couldn't read the file correctly this time. Foton seemed to be unbeatable.
Even if we replaced the sampling rate, Foton kept saying 2048! (This is maybe because Foton's default value is 2048Hz. Everytime Foton notice some editting in the file, he destroys everything. He hates editting)
The problems were always associated with the sampling rate, so

 6. Got back to step4 and undo-ed the replacement.

 7. Foton could read it this time, so I changed the sampling rate one by one using Foton GUI.

 8. Checked filters using Foton's Bode Plot.(Not for all, but some that had problem before)

 9. Splitted SDSEN filters to SDSEN, SUSSIDE, and SDCOIL.

 10. Put some missing whitening filters, and 28HzELPs.
   BS, PRM, SRM didn't have any 28HzELP for SDCOIL.
   ITMX and ITMY SDCOIL had SimDW/InvDW which doesn't make sense(SIDEs don't have analog DW). So, I deleted and replaced with 28HzELP.

F2A issue:
 We failed in sending F2A filters to new filter files.
 These are a little bit complicated because TO_COIL_X_X filters were named ULPOS,URPOS etc before.
 Also, MC3 didn't have any F2As, so maybe we should but the same F2As as MC1/2.
 Note that every F2As are different, and TO_COIL matrix have UL,UR,LL,LR order(not same as INMATRIX).
 Also, SRM had f2pv instead of F2A!

Next work:
 - Check whole filters by actually measuring transfer function between SENs and COILs.
 - Damp MC suspentions, and lock MC.
 - Measure openloop TF and compare with the designed.


How do you read a Foton filter file:
 When you open up a Foton filter file, you see filters like this.

################################################################################
### modulename                                                               ###
################################################################################
# SAMPLING modulename samplingrate
# DESIGN   modulename n filterdesign
# DESIGN   modulename n filterdesign
###                                                                          ###
modulename   n xy z      v      w filtername                        gain    a1     a2    b1     b2
modulename   n xy z      v      w filtername                        gain    a1     a2    b1     b2
                                                                            a1     a2    b1     b2
n: filter number
 0 for FM1, 1 for FM2, ... , 9 for FM10

x: Input Switching setting
 1 Always On
 2 Zero History

y: Output Switching setting
 1 Immediately
 2 Ramp
 3 Input Crossing
 4 Zero Crossing

z: number of filters cascaded.

v: if y=2, (Ramp Time(sec))*(samplingrate)
   if y=3 or 4, Tolerance

w: (Timeout(sec))*(samplingrate)

 Note that v and w are changed when sampling rate is changed.

 Transfer function will be;
  H(1/z)=G*(1+b1/z+b2/z/z)/(1+a1/z+a2/z/z)
  z=exp(s/fs)

 where fs is the sampling frequency.

Reference:
 Kiwamu Izumi: "Notes about Digital Filters," http://tamago.mtk.nao.ac.jp/izumi/green/DigitalFilter.pdf

  3744   Wed Oct 20 01:22:18 2010 yutaUpdateCDSfixed filters for MCs, but no damping

(Rana, Yuta)

Summary:

 The damping servo for MCs has not been working since Monday.
 We already found that some filter coefficients were wrong.(see elog #3742)
 So, we fixed it (just for MCs now).
 But still doesn't work.

What we did:
Fixed the filters for MC suspensions;
 1. In /cvs/cds/rtcds/caltech/c1/chans/ directory, we replaced C1MCS.txt with ./filter_archive/c1mcs/C1MCS_101019_101927.txt.
  This is the archive of the filter bank for MC suspensions, when filter designs were correct(when we hadn't opened it with foton yet).

 2. Deleted all the filter coefficients.
   By using emacs magical C-<space> C-x r k.

 3. Loaded with foton to get correct coefficients.

 4. Reloaded coefficients for C1MCS.

We now have the correct filters, but the damping still doesn't work.

 5. To confirm that the filters are now correct and the model is working correctly, we measured the transfer function between ULSEN input and ULCOIL output and compared with the calculated TF from the filter bank file (Attachment #1) .

 6. To calculate the designed function, we used the following matlab scripts;
    /cvs/cds/caltech/users/rana/mat/utilities/FotonFilter.m        # takes the foton file and returns the TF
    /cvs/cds/caltech/users/rana/mat/utilities/readFilterFile.m    # reads the foton file
    /cvs/cds/caltech/users/yuta/scripts/sentocoil.m                   # calculates the total TF from SEN to COIL

Result:
 As you can see from the attachment, filters seem fine now. (The red line is the measured and the blue line is the calculated function in the plot)
 But the damping (for POS, PIT, YAW) still doesn't work. Why????

Next work:

 - see if coils are pushing correctly
 - push cable connectors further!
 - check input and output filters (whitening, dewhitening)
 - when fixed, use my fully updated QAdjuster.py for adjusting Q for multiple channels automatically!!

Attachment 1: Screenshot.png
Screenshot.png
  16986   Mon Jul 11 17:25:43 2022 TegaUpdateVACfixed obsolete reference bug in serial_XGS600 service

Koji noticed that the FRG sensors were not updating due to reference to an obsolete modbusIOC_XGS service, which was used temporarily to test the operation of the serial XGS sensor readout to EPICS. The information in this service was later moved into modbusIOC.service but the dependence on the modbusIOC_XGS.service was not removed from the serial_XGS600.service. This did not present any issue before the shutdown, probably bcos the obsolete service was already loaded but after the restart of c1vac, the obsolete service file modbusIOC_XGS.service was no longer available. This resulted in  serial_XGS600.service throwing a failure to load error for the missing obsolete modbusIOC_XGS service. The fix involved replacing two references to 'modbusIOC_XGS' with 'modbusIOC' in  /opt/target/services/serial_XGS600.service.

I also noticed that the date logged in the commit message was Oct 2010 and that I could not do a push from c1vac due to an error in resolving git.ligo.org. I was able to push the commit from my laptop git repo but was unable to do a pull on c1vac to keep it synced with the remote repo.

  3767   Fri Oct 22 23:29:48 2010 yutaUpdateCDSfixed output filter switching

(Joe, Yuta)

Summary:
 This evening, we fixed output filter analog/digital switchings in Simulink logic, but they didn't actually switched filters.
 That was because what we thought BO0 was BO2 and BO2 was BO0.
 We re-wired, re-labeled the cables.
 We checked it by using a LED(see "test configuration" in this wiki page).
 Also, we checked the Simulink connection by probing MAX333A in SOS Dewhitening Board located at 1X4-8-11B, which switches SIDE coils for all optics.

Next work:

 - Currently, FM10 in UL/UR/LR/LL/SDCOIL filters switch analog/digital, but let's make FM9 do the switching.(See schematic in elog #3758)
 - Check all the logic by measuring transfer functions(maybe single frequency measurement is enough. we can use tdsdmd etc to automate).

  3757   Thu Oct 21 21:35:16 2010 yutaUpdateCDSfixed whitening filter switches

(Joe, Yuta)

Summary:
 We found some mistakes in whitening filter switches yesterday(see elog #3748).
 One was the mis-connection between LL/LR and the second was the bit invert.
 So, we fixed it and checked that they are now switching correct by probing MAX333A.

Why we made mistake:
1. LL/LR mis-connection
  Simply, Simulink model was wrong.
  It occurred because UL/UR/LR/LL order is currently arbitrary and confusing.
  For now, we just swap the connection(Attached #1).
  In the future, we should fix all orders corresponding to the actual wiring order(UL > LL > UR > LR).

2. Bit invert
  It occurred from counter-intuitive output of Contec 32 BO(See Attachment #2(taken from this wiki page)).
  If "1," current goes from A to B, so MAX333A input will be 0V, which means analog WF is on.
  If "0," no current, so MAX333A input will be +15V, which means analog WF is off(bypassed).

Next work:
 - There are no digital WF in SDSEN inputs. So, we have to put them.
 - Dewhitening filters are totally messed up now. Switches are wrong, some optics have digital DW but some doesn't, some have 28Hz elliptic LPF.......
     We have to grand unify them.

Attachment 1: LLLR.png
LLLR.png
Attachment 2: Contec23BO.png
Contec23BO.png
  8991   Fri Aug 9 21:05:28 2013 KojiUpdateSUSfixed: SRM coils fine - problem with slow bias slider

Now the SRM Yaw bias in yaw is functional without any strage behavior.
The problem was found at the connector of the flat ribbon cable from the DAC to the cross connect.

I used the extender board to diagnose the SRM coil driver circuit at 1X4.
The UL coil input did not show any sign of voltage no matter how the bias slider was jiggled.

I opened the side panel of the rack and found the signal was absent at the cross connect which relays two flat ribbon cables
for the SRM coil driver. I checked the DAC output with a multimeter. All the bias outputs were OK at the DAC.

Then I opened the IDC connector at the DAC side of the crossconnect as the signal was already missing there.
I found that the flat ribbon cable was a half line shifted from the supposed location.
This resulted a short circuit of the DAQ +/- pins for the SRM UL coil.

I recrimped the connector and now the SRM Yaw slider is back.
This changed the nominal position of the SRM. The new slider values were saved.

  5664   Thu Oct 13 23:58:38 2011 KojiUpdateLSCfixing REFL165

I already have reported in this entry that REFL165 shows too high DC output which does not depend on the light level on the diode.
Today I removed REFL165 from the table and inspected it.

The diode has been burnt as shown in the first picture (left).
The window is smoked, and the photo sensitive surface has been removed from its base. It moves in the can.

The burnt diode was replaced to the new one.
The new one shows ~30% better capacitance of ~50pF
and I had to increase the inductance from 14nH (i.e. 15nH//220nH) to 18nH.
After some struggles to increase/decrease the stray inductance by moving the SMD capacitors a little, the resonance is reasonably tuned to 166MHz.

The comprehensive test will be performed shortly.

Attachment 1: PA131612.jpg
PA131612.jpg
Attachment 2: PA131618.jpg
PA131618.jpg
  9444   Thu Dec 5 12:20:09 2013 JenneUpdateCDSfixing c1mcs timing - go for it

Quote:

We still have c1mcs having regularly time-over. Can I remove the WFS->OAF connections temporarily?

 Yes.  I think that's a good idea, since we're not using them at this time, and we want c1mcs to behave.  Maybe make an elog note of which version is the first without them, so that we can conveniently find the model(s) in the svn?

  9309   Tue Oct 29 18:14:52 2013 JamieConfigurationComputer Scripts / Programsfixing python-matplotlib from LSCSOFT

Jenne just discovered an issue with the python-matplotlib package that I knew was coming but forgot about.

We pull packages from the LSCSOFT Debian "squeeze" archive, which is a convenient way for us to install LIGO data analysis software.  There are no LSCSOFT archives for Ubuntu, and Debian "squeeze" is the closest supported distribution to Ubuntu 10.04 "lucid", which is what we are using.

DASWG recently added python-matplotlib to the LSCSOFT squeeze archive.  The version they added (1.0.1-3) supersedes the version in lucid, so by default apt wants to install it.  However, the LSCSOFT version is compiled against a newer version of some standard libraries, so it won't function on our system and seg faults.

The solution (a solution) is to use apt "pinning" to pin the package to the lucid version that works.  I've added the following file on all the 10.04 workstations to prevent the package from upgrading to the LSCSOFT version:

controls@pianosa:~ 0$ cat /etc/apt/preferences.d/pin_python-matplotlib
Package: python-matplotlib
Pin: release a=lucid
Pin-Priority: 1000

 

  9310   Tue Oct 29 18:54:36 2013 JamieConfigurationComputer Scripts / Programsfixing python-matplotlib from LSCSOFT

Quote:
controls@pianosa:~ 0$ cat /etc/apt/preferences.d/pin_python-matplotlib
Package: python-matplotlib
Pin: release a=lucid
Pin-Priority: 1000 

I forgot that there were a couple of different matplotlib packages that all needed to be pinned.  To be safe I decided to just pin all packages to the lucid versions.  This will still allow us to install lscsoft packages that are not ubuntu, but it will always prefer packages from lucid instead.  Here's the new pinning file:

controls@pianosa:~ 0$ cat /etc/apt/preferences.d/pinning 
Package: *
Pin: release a=lucid
Pin-Priority: 1000
controls@pianosa:~ 0$ 

  7062   Tue Jul 31 20:59:35 2012 JamieUpdateCDSfixing up MEDM snapshots

I've started to try to fix all the old non-working medm snapshots.  I've made a new snapshot directory, and some new snapshot scripts to handle taking the snapshots, and view old ones.

Snapshots are now in:  /opt/rtcds/caltech/c1/medm/snap

There is one main script: /opt/rtcds/caltech/c1/medm/snap/snapcommands.  This is linked by:

  • updatesnap: update a snapshot
  • viewsnap: view most recent snapshot
  • viewold: view old snapshots

These commands take either the path to the screen in question, or it's relative path to the medm directory (/opt/rtcds/caltech/c1/medm).  The snapshots for a specific screen are stored in a directory specific to the screen, in a place relative to the snap directory that mimics the screens relative path to the overall medm directory.  So for instance, the snap directory for: 

/opt/rtcds/caltech/c1/medm/c1lsc/master/C1LSC_OVERVIEW.adl

is:

controls@rossa:~ 0$ find /opt/rtcds/caltech/c1/medm/snap/c1lsc/master/C1LSC_OVERVIEW/
/opt/rtcds/caltech/c1/medm/snap/c1lsc/master/C1LSC_OVERVIEW/
/opt/rtcds/caltech/c1/medm/snap/c1lsc/master/C1LSC_OVERVIEW/2012-08-01-02:21:34-UTC.png
/opt/rtcds/caltech/c1/medm/snap/c1lsc/master/C1LSC_OVERVIEW/2012-08-01-02:21:04-UTC.png
/opt/rtcds/caltech/c1/medm/snap/c1lsc/master/C1LSC_OVERVIEW/2012-08-01-02:21:23-UTC.png
/opt/rtcds/caltech/c1/medm/snap/c1lsc/master/C1LSC_OVERVIEW/2012-08-01-03:41:34-UTC.png
/opt/rtcds/caltech/c1/medm/snap/c1lsc/master/C1LSC_OVERVIEW/current.png
controls@rossa:~ 0$ 

The "current.png" is a link to the most recent snapshot, and is what you get when you "viewsnap".

Below is an example medm snippet of what could be used in screens to enable this functionality.  I've fixed up a couple of the screens, but there are a lot more that need to be updated.

"shell command" {
    object {
        x=666
        y=597
        width=40
        height=40
    }
    command[0] {
        label="Settings and Procedures"
        name="emacs"
        args="./help/IFO_ALIGN.txt &"
    }
    command[1] {
        label="View Snapshot"
        name="/opt/rtcds/caltech/c1/medm/snap/viewsnap"
        args="MISC/IFO_ALIGN.adl &"
    }
    command[2] {
        label="View old snapshots"
        name="/opt/rtcds/caltech/c1/medm/snap/viewold"
        args="MISC/IFO_ALIGN.adl &"
    }
    command[3] {
        label="Update Snapshot"
        name="/opt/rtcds/caltech/c1/medm/snap/updatesnap"
        args="MISC/IFO_ALIGN.adl &"
    }
    clr=14
    bclr=30
}
  3621   Wed Sep 29 15:34:46 2010 kiwamuUpdateGeneralfixing vertex crane
From Vertex crane -- fixing


Quote:

The vertex crane drive is overheating, it stopped functioning. Service man will be here tomorrow morning.

I crane was just turned  on for for may be  about 5 minutes. The vertical drive was fine for a while, but the horizontal did not worked at all.

The crane is tagged out again and the controller box is cooling down.

 

  10176   Thu Jul 10 15:57:00 2014 SteveUpdateSUSflow bench is turned off

Flow bench effect on oplev error signal     is here.

I turned off the south X-end flow bench.

  6845   Thu Jun 21 09:10:08 2012 steveUpdatePEMflow bench must be running all times

The south end flow bench HEPA filter should be run all times. You can turn it off for a measurement or two but remember we are storing clean optics there.

The zero count bench will reach  room particle count  ~ 10,000 in one minute.

  6846   Thu Jun 21 12:13:35 2012 JenneUpdatePEMflow bench must be running all times

Quote:

The south end flow bench HEPA filter should be run all times. You can turn it off for a measurement or two but remember we are storing clean optics there.

The zero count bench will reach  room particle count  ~ 10,000 in one minute.

 My bad.  I turned it off last night to see if it would help make the Xgreen more stable, and then when I woke up this morning I realized that I had forgotten to turn it back on.  Bad Jenne.

  8222   Mon Mar 4 17:32:26 2013 yutaBureaucracyGeneralflowchart for PRMI g-factor measurement

I made a very useful flowchart for the week. Our goal for the week is to measure g-factor of PRC in PRMI.

PRMIgfactorPlan.png

  9003   Tue Aug 13 11:04:44 2013 SteveUpdatePEMfluorecent lights

Our fluorecent lights became obsolete.  We'll have change fixtures over to some more energy efficient one. Do you have any recommendation regarding to less noise performer unit?

We may go this direction of LED fluorecent lamps ?

  6197   Fri Jan 13 17:40:38 2012 ZachUpdateRF Systemfoam box and temp controller taken off of PSL table

I stole the foam EOM box and the temperature controller circuit from the PSL table so that I could continue with the RAM measurements here at the ATF.

That is all.

ELOG V3.1.3-