40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 57 of 337  Not logged in ELOG logo
ID Date Author Type Category Subject
  14132   Fri Aug 3 19:02:11 2018 gautamUpdateASSX arm ASS recovery

[koji, gautam]

After I effected the series resistance change for ETMX, the X arm ASS didn't work (i.e. IR transmission would degrade if the servo was run). Today, we succeeded in recovering a functional ASS servo yes.

So both arms have working dither alignment servos now. But remember that the Y arm ASS gains have been set for locking the Y arm with MC2 as the actuator, not ETMY.

Details:

  • Koji pointed out that the demodulated signals from the ETM dither are only used to center the spot on the ETM, and that we should first run the servo with existing settings with the ETM pitch and yaw spot centering loops disabled.
    • This improved TRX level from ~0.8 to 1.1
  • Next, we tried increasing the LO amplitudes by x5 to account for the reduced actuation of the dither on ETMX
    • We then re-enabled the two loops that were earlier disabled.
    • This resulted in TRX degrading very quickly.
  • So we decided to try going back to the nominal LO gains, and reducing the gain of the two ETM spot centering loops.
    • This did the trick, TRX went from 1.1 --> ~1.23, which is the nominal maximum pre-vent value.
  • The snap file used to recover the correct settings to run the dither alignment servos have been updated, the old one has been backed up with today's datestamp.

We then tried to maximize GTRX using the PZT mirrors, but were only successful in reaching a maximum of 0.41. The value I remember from before the vent was 0.5, and indeed, with the IR alignment not quite optimized before we began this work, I saw GTRX of 0.48. But the IR dither servo signals indicate that the cavity axis may have shifted (spot position on the ITM, which is uncontrolled, seems to have drifred significantly, the Pitch signal doesn't stay on the StripTool scale anymore). So we may have to double check that the transmitted beam isn't falling off the GTRX DC PD.

  14131   Fri Aug 3 18:54:58 2018 gautamUpdateSUSGlitchy MC1

The wall StripTool indicated that the IMC wasn't too happy when I came in today. Specifically:

  • MC1 watchdog was tripped.
  • Even in the tripped state, MC REFL spot on the camera showed spot motion that was too large to be explained as normal seismic driven motion (i.e. with local damping supposedly disabled).
  • Strange excursions were observed in the MC1 shadow sensor signal levels as well, see Attachment #1 - negative values don't make any sense for this readout.

The last time this happened, it was due to the Sorensens not spitting out the correct voltages. This time, there were no indications on the Sorensens that anything was funky. So I just disabled the MCautolocker and figured I'd debug later in the evening.

However, around 5pm, the shadow sensor values looked nominal again, and when I re-enabled the local damping, the MC REFL spot suggested that the local damping was working just fine. I re-enabled the MCautolocker, MC re-locked almost immediately. To re-iterate, I did nothing to the electronics inside the VEA. Anyways, this enabled us to work on the X arm ASS (next elog).

Attachment 1: MC1_sensorAnomaly.png
MC1_sensorAnomaly.png
  14130   Fri Aug 3 16:27:40 2018 ranaUpdateSUSLow noise bias path idea

Bah! Too complex.

  14129   Fri Aug 3 15:53:25 2018 gautamUpdateSUSLow noise bias path idea

Summary:

The idea we are going with to push the coil driver noise contribution down is to simply increase the series resistance between the coil driver board output and the OSEM coil. But there are two paths, one for fast actuation and one that provides a DC current for global alignment. I think the simplest way to reduce the noise contribution of the latter, while preserving reasonable actuation range, is to implement a precision DC high-voltage source. A candidate that I pulled off an LT application note is shown in Attachment #1.

Requirements:

  • The series resistance in the bias path should be 10 k\Omega, such that the noise from this stage is dominated by the Johnson noise of said resistor, and hence, the current noise contribution is negligible compared to the series resistance in the fast actuation path (4.5 k\Omega).
  • Since we only really need this for the test masses, what actuation range do we want?
    • Currently, ETMY has a series resistance of 400\Omega and has a pitch DC bias voltage of -4 V. 
    • This corresponds to 10 mA of DC current.
    • To drive this current through 10 k\Omega, we need 100 V. 
    • I'm assuming we can manually correct for yaw misalignments such that 10mA of DC current will be sufficient for any sort of corrective alignment.
    • So +/- 120 V DC should be sufficient.
  • The current noise of this stage should be negligible at 100 Hz. 
    • The noise of the transistors and the HV supply should be suppressed by the feedback loop and so shouldn't be a significant contribution (I'll model to confirm).
    • The input noise of the LT1055 is ~20nV/rtHz at 100 Hz, while the Johnson noise of 10 k\Omega is ~13nV/rtHz so maybe the low-passing needs to be tuned, but I think if it comes to it, we can implement a passive RC network at the output to achieve additional filtering.
  • To implement this circuit, we need +/- 125V DC. 
    • At EX and EY, we have a KEPCO HV supply meant to be used for the Green Steering PZTs. 
    • I'm not sure if these can do bipolar outputs, if not, for temporary testing, we can transport the unit at EY to EX.

If all this seems reasonable, I'd like to prototype this circuit and test it with ETMX, which already has the high series resistance for the fast path. So I will ask Steve to order the OpAmp and transistors.

Attachment 1: LT1055_precOpAmp.pdf
LT1055_precOpAmp.pdf
  14128   Fri Aug 3 14:35:56 2018 gautamSummaryElectronicsEX AUX electronics power restored

Steve and I restored the power to the EX AUX electronics rack. The power strip on the lowest shelf of the AUX rack now goes to another power strip laid out vertically along the NW corner of 1X9. The EX green locks to the arm just fine now.

  14127   Thu Aug 2 23:09:25 2018 ranaSummaryComputersX Green "Mystery" solved

I'm going to guess that this was me: I was disconnecting some octopus power strip nonsense down there (in particular illuminators and cameras), so I might have turned off the AUX rack by mistake.

Quote:

I walked down to the X end and found that the entire AUX laser electronics rack isn't getting any power. There was no elog about this.

I couldn't find any free points in the power strip where I think all this stuff was plugged in so I'm going to hold off on resurrecting this until tomorrow when I'll work with Steve.

Quote:

The X arm green does not stay locked to the cavity - the alignment looks fine, and the green flashes are strong, but the lock does not hold. This shouldn't be directly connected to anything we did today since the Green PDH servo is entirely analog.

  14126   Thu Aug 2 20:54:18 2018 gautamSummaryComputersc1omc model looks stable

Actually, c1lsc had crashed again sometime last night so I had to reboot everything this morning. I used the reboot script again, but I increased the sleep time between trying to start up the models again so that I could walk into the VEA and power cycle the c1lsc expansion chassis, as this kind of frequent model crash has been fixed by doing so in the past. Sure enough, there have been no issues since I rebooted everything at ~1030 in the morning. 

The c1omc model itself has been stable as well, though of course, there is nothing in there at the moment. I may do a check of the newly installed DAC tomorrow just to see that we can put out a sine wave.

Steve has ordered the D-sub cabling that will allow us to route signals between AA/AI boards in 1X1/1X2 to the HV PZT electronics in the OMC rack. Things look setup for a measurement next week. Aaron will post a block diagram + photoz of what box goes where in the electronics racks.

  14125   Thu Aug 2 20:47:29 2018 gautamSummaryElectronicsX Green "Mystery" solved

I walked down to the X end and found that the entire AUX laser electronics rack isn't getting any power. There was no elog about this.

I couldn't find any free points in the power strip where I think all this stuff was plugged in so I'm going to hold off on resurrecting this until tomorrow when I'll work with Steve.

Quote:

The X arm green does not stay locked to the cavity - the alignment looks fine, and the green flashes are strong, but the lock does not hold. This shouldn't be directly connected to anything we did today since the Green PDH servo is entirely analog.

  14124   Thu Aug 2 16:30:08 2018 SteveUpdateTreasuretime capsule location

I 've just found this time capsule note from Nov. 26, 2000 by Kip Thorne:  LIGO will discover gravitational waves by Dec.31, 2007

Quote:

   Beautifully Done

   Chirp

  what is next?

Atm 3, Ron Drever could not celebrate with us because of health issues.

 

 

Attachment 1: time_capsule.JPG
time_capsule.JPG
  14123   Wed Aug 1 20:44:57 2018 gautamSummaryComputersc1omc model (re?)created

The main motivation behind adding a DAC card in c1ioo was to setup an RTCDS model for the OMC. Attachment #1 shows the new look CDS overview screen. Here is what I did.

Mostly, I followed instructions from when I setup the model for the EX green PZTs.


Simulink model:

The model is just a toy for now (CDS parameters, ADC block and 2 CDS filter modules). I leave it to Aaron to actually populate it, check functionality etc. The path to the model is /opt/rtcds/caltech/c1/userapps/release/isc/c1/models/c1omc.mdl. I am listing the parameters set on the CDS_PARAMETERS block:

  • host = c1ioo
  • site = c1
  • rate = 16k
  • dcuid = 27 (which I chose after making sure that this dcuid was not used on this list which I also updated by adding c1omc and moving c1imc to "old")
  • specific_cpu = 6 (again chosen after checking the available CPUs in the above list and confirming using the cset utility).
  • adc_Slave = 1
  • shmem_daq = 1
  • no_rfm_dma = 1
  • biquad = 1

Building and installing model:

Once the model was installed, I logged into c1ioo, and built and installed the models using the usual rtcds make and rtcds install instructions. Before starting the model, I edited /diskless/root.jessie/etc/rtsystab to allow c1omc to be run on c1ioo. Using sudo cset set, I verified that CPU #6 is no longer listed (if I understand correctly, the RTCDS system takes over the core).


MEDM:

To reflect all this on the MEDM CDS OVERVIEW screen, I just edited the screen.

  • Moved the orange explanation of bits over to the c1iscey panel to make space in the c1ioo panel.
  • Edited the macros to reflect the c1omc parameters.

DAQD:

Finally, I followed the instructions here to get the channels into frames and make all the indicators green. Went into fb and restarted the daqd processes. All looks good smiley. I'm going to leave the model running overnight to investigate stability. I forgot to svn commit the model tonight, will do it tomorrow.


The testing plan (at least initially) is to install the AA and AI boards from the OMC rack in 1X1/1X2. Then we will have short SCSI cables running from the ADC/DAC to these. The actual HV driving stages will remain in the OMC rack (NE corner of AS table).

@Steve, can we get 10 Male-Female D9 cables so that we can run them from 1X1/1X2 to the OMC rack?


Unrelated to this work: There were 2 crashes of the models on c1lsc, one ~6pm and one right now ~1030pm. The restart script brought everything back gracefully  yes...

Attachment 1: CDS_OVERVIEW_withOMC.png
CDS_OVERVIEW_withOMC.png
  14122   Wed Aug 1 19:41:15 2018 gautamSummaryComputersRTCDS recovery, c1ioo changes

[Gautam Koji]

After this work, we recovered the nominal RTCDS state. The main points were:

  1. We needed to restart the bind9 service on chiara such that the FEs knew their IP addresses upon reboot and hence, could get their root filesystems over NFS.
  2. We recovered suspension local damping, IMC locking and POX/POY locking with nominal arm transmission.

Some stuff that is not working as usual:

  1. The EX QPD is reporting strange transmission values - even with the PRM completely misaligned, it reports transmission of ~30. But we were able to lock the Xarm with the Thorlabs PD and revover transmission of ~1.15.
  2. The X arm green does not stay locked to the cavity - the alignment looks fine, and the green flashes are strong, but the lock does not hold. This shouldn't be directly connected to anything we did today since the Green PDH servo is entirely analog.

I made a model change in c1x03 (the IOP model on c1ioo) to add a DAC part. The model compiled, installed and started correctly, and looking at dmesg on c1ioo, it recognises the DAC card as what it is. Next step is to use a core on c1ioo for a c1omc model, and actually try driving some signals.

Note that the only change made to the c1ioo expansion chassis was that a DAC card was installed into the PCIe bus. The adaptor card which allows interfacing the DAC card to an AI board was already in the expansion chassis, presumably from whenever the DAC was removed from this machine.

*I think I forgot to restart optimus after this work...

Attachment 1: CDS_overview.png
CDS_overview.png
  14121   Wed Aug 1 16:23:48 2018 KojiSummaryComputersTransition of the main NFS disk on chiara

[Gautam Koji]

Taking the opportunity to shutdown c1ioo for adding a DAC card, we shutdown chiara and worked on moving of the main disk to the bigger home.

We shutdown most of the martian machines including the control machines, megatron, optimus, and nodus.

- Before shutting down chiara, we ran rsync to make the 4TB disk (used to be teh backup) and /cvs/cds synced.

sudo rsync -a --progress /home/cds/ /media/40mBackup

- Modified /etc/fstab

proc            /proc           proc    nodev,noexec,nosuid 0       0
# / was on /dev/sda1 during installation
UUID=972db769-4020-4b74-b943-9b868c26043a /               ext4    errors=remount-ro 0       1
# swap was on /dev/sda5 during installation
UUID=a3f5d977-72d7-47c9-a059-38633d16413e none            swap    sw              0       0
UUID="90a5c98a-22fb-4685-9c17-77ed07a5e000"    /media/40mBackup       ext4      defaults,relatime,commit=60       0         0
#fb:/frames      /frames nfs     ro,bg

UUID=92dc7073-bf4d-4c58-8052-63129ff5755b   /home/cds    ext4    defaults,relatime,commit=60    0   0

- Shutdown chiara. Put the 4TB disk in the chassis. We also installed a new disk (but later it turned out that it only has 2TB...)

- Restart the mahcine. This already made the 4TB disk mounted as /cvs/cds .

- Restart bind9 with DHCP for the diskless clients (cf. https://wiki-40m.ligo.caltech.edu/CDS/How_to_join_martian)

sudo service bind9 restart
sudo service isc-dhcp-server restart

- Looks like /etc/resolv.conf is automatically overwritten by a tool or something everytime we restart the machine!? I still don't know how to avoid this. (cf.  https://www.ctrl.blog/entry/resolvconf-tutorial). But at least for today we manually wrote /etc/resolv.conf

controls@chiara|backup> cat /etc/resolv.conf
# Dynamic
resolv.conf(5) file for
glibc resolver(3) generated by resolvconf(8)
#     DO NOT EDIT THIS FILE BY HAND -- YOUR CHANGES WILL BE OVERWRITTEN
nameserver 192.168.113.104
nameserver 131.215.125.1
nameserver 8.8.8.8

search martian

  14120   Tue Jul 31 22:50:18 2018 aaronUpdateOMCOMC Expected Refl Signal

I learned a lot about lasers this week from Siegman. Here are some plots that show the expected reflectivity off of the OMC for various mode matching cases.

The main equation to know is 11.29 in Siegman, the total reflection coefficient going into the cavity:

R=r-\frac{t^2}{r}\frac{g(\omega)}{1-g(\omega)}

Where r is the mirror reflectivity (assumed all mirrors have the same reflectivity), t is the transmissivity, and g is the complex round-trip gain, eq 11.18

g(\omega)=r_1r_2(r_3...)e^{-i\phi}e^{-\alpha_0p}

The second exponential is the loss; in Siegman the \alpha_0 is some absorption coecfficient and p is the total round trip length, so the product is just the total loss in a round trip, which I take to be 4*the loss on a single optic (50ppm each). \phi is the total round trip phase accumulation, which is 2\pi*detuning(Hz)/FSR. The parameters for the cavity can be found on the wiki.

I've added the ipynb to my personal git, but I can put it elsewhere if there is somewhere more appropriate. I think this is all OK, but let me know if something is not quite right.

Attachment 1: omcRefl.pdf
omcRefl.pdf
  14119   Tue Jul 31 08:17:55 2018 SteveUpdateSUSTrillium interface box was fixed,reinstalled & working

 

 

Attachment 1: all_OK.png
all_OK.png
  14118   Mon Jul 30 18:19:03 2018 KojiUpdateSUSTrillium interface box was fixed and reinstalled

The trillium interface box was removed from the rack.

The problem was the incorrect use of an under-spec TVS (Transient Voltage Suppression) diodes (~ semiconductor fuse) for the protection circuit.
The TVS diodes we had had the breakdown voltages lower than the supplied voltages of +/-20V. This over-voltage eventually caused the catastrophic breakdown of one of the diodes.

I don't find any particular reason to have these diodes during the laboratory use of the interface. Therefore, I've removed the TVS diodes and left them unreplaced. The circuit was tested on the bench and returned to the rack. All the cables are hooked up, and now the BRLMs look as usual.


Details

- The board version was found to be D1000749-v2

- There was an obvious sign of burning or thermal history around the components D17 and D14. The solder of the D17 was so brittle that just a finger touch was enough to remove the component.

- These D components are TVS diodes (Transient Voltage Suppression Diodes) manufactured by Littelfuse Inc. It is sort of a surge/overvoltage protector to protect rest of the circuit to be exposed to excess voltage. The specified component for D17/D14 was 5.0SMMDJ20A with reverse standoff voltage (~operating voltage) of 20V and the breakdown voltage of 22.20V(min)~24.50V(max). However, the spec sheet told that the marking of the proper component must be "5BEW" rather than "DEM," which is visible on the component. Some search revealed that the used component was SMDJ15A, which has the breakdown voltage of 16.70V~18.50V. This spec is way too low compared to the supplied voltage of +/-20V.

Attachment 1: P_20180730_173134.jpg
P_20180730_173134.jpg
Attachment 2: P_20180730_180151.jpg
P_20180730_180151.jpg
  14117   Mon Jul 30 16:11:54 2018 gautamUpdateSUSTrillium interface box is broken

[koji, steve, gautam]

We debugged this in the following way:

  1. Disconnect all fuses in the terminal blocks coming from the +/- 20 VDC Sorensens.
  2. Check that they are indeed isolated using DMM.
  3. Test blocks of fuses in order to identify where the problem is happening (i.e. plug fuses in, turn up Sorensen voltage knobs, look for current overload). We did things in the following order:
    • MC suspensions
    • BS, PRM and SRM
    • ITMY
    • ITMX
    • Trillium interface box.
  4. Turns out that the Trillium box is the culprit.
  5. Confirmed that the problem is in the trillium interface box and not in the seismometer itself by unplugging all cables leading out of the interface box, and checking that the problem persists when the box is powered on.

So for now, the power cable to the box is disconnected on the back end. We have to pull it out and debug it at some point.

Apart from this, megatron was un-sshable so I had to hard reboot it, and restart the MCautolocker, FSSslowPy and nds2 processes on it. I also restarted the modbusIOC processes for the PSL channels on c1auxex (for which the physical Acromag units sit in 1X5 and hence were affected by our work), mainly so that the FSS_RMTEMP channel worked again. Now, IMC autolocker is working fine, arms are locked (we can recover TRX and TRY~1.0), and everything seems to be back to a nominal state. Phew.

  14115   Mon Jul 30 11:05:44 2018 gautamUpdateSUSIFO SUS wonky

When I came in this morning:

  • PMC was unlocked.
  • Seis BLRMS were off scale.
  • ITMX OSEM LEDs were dark on the CRT monitor even though Sat Box was plugged in.

Checking status of slow machines, it looked like c1sus, c1aux, and c1iscaux needed reboots, which I did. Still PMC would not lock. So I did a burtrestore, and then PMC was locked. But there seemed to be waaaaay to much motion of MCREFL, so I checked the suspension. The shadow sensor EPICS channels are reporting ~10,000 cts, while they used to be ~1000cts. No unusual red flags on the CDS side. Everything looked nominal when I briefly came in at 6:30pm PT yesterday, not sure if anything was done with the IFO last night.

Pending further investigation, I'm leaving all watchdogs shutdown and the PSL shutter closed.

A quick look at the Sorensens in 1X6 revealed that the +/- 20V DC power supplies were current overloaded (see Attachment #1). So I set those two units to zero until we figure out what's going on. Possibly something is shorted inside the ITMX satellite box and a fuse is blown somewhere. I'll look into it more once Steve is back.

Attachment 1: IMG_7102.JPG
IMG_7102.JPG
  14114   Sun Jul 29 23:15:34 2018 poojaUpdateCamerasDeveloping CNN

Aim: To develop a convolutional neural network that resolves mirror motion from video.

Input : Previous simulated video of beam spot motion in pitch by applying 4 sine  waves of frquencies 0.2, 0.4, 0.1, 0.3 Hz  and amplitude ratios to frame size to be 0.1, 0.04, 0.05, 0.08 where random uniform noise ranging 0.05 has been added to amplitudes and frequencies. This is divided into train (0.4), validation (0.1) and test (0.5).

Model topology:

  • Number of filters = 2
  • Kernel size = 2
  • Size of pooling windows = 2
  •                                        ----->         Dense layer of 4 nodes  ---->    Output layer of 1 node 

         Activation:                      selu                                                  linear

Batch size = 32, Number of epochs = 128, loss function = mean squared error

Optimizer: Nadam ( learning rate = 0.00001, beta_1 = 0.8, beta_2 = 0.85)

Plots of CNN output & applied signal given in Attachment 1. The variation in loss value with epochs given in Attachment 2.

This needs to be further analysed with increasing random uniform noise over the pixels and by training CNN on simulated data of varying ampltides and frequencies for sine waves.

Attachment 1: conv_nn_varying_freq_amp_1.pdf
conv_nn_varying_freq_amp_1.pdf
Attachment 2: conv_nn_varying_freq_amp_2.pdf
conv_nn_varying_freq_amp_2.pdf
  14113   Sun Jul 29 20:03:02 2018 ranaUpdatePEMSeismometer temp control

While Shruti is re-building Kira's heater circuit, I looked up how to do one of these (i.e. what does a real EE say about how to build a current source?):

It turns out that there is an Analog Devices application note (AN-968) about this (as there usually is once we get tired of playing around and try to look up the right answer).

I've linked to the note and attached the recommended schematic for high current applications. We'll go ahead as is, but we'll make a PCB according to this App Note for the v3 circuit.

 

Attachment 1: Screen_Shot_2018-07-29_at_8.00.27_PM.png
Screen_Shot_2018-07-29_at_8.00.27_PM.png
  14112   Sun Jul 29 00:59:54 2018 KojiUpdateElectronicsCharacterization of Transimpedance Amplifier

You have this measurement problem when the IF bandwidth is larger than the measurement frequency. I suspect the IF bandwidth is 30kHz.

  14111   Sat Jul 28 22:16:49 2018 John MartynUpdate Characterization of Transimpedance Amplifier

Kevin and I meaured the transfer function of the photodiode circuit using the Jenne laser and agilent in the 40m lab. The attached figures depict our measured transfer function over the modulation frequency ranges of 30kHz-30MHz and 1kHz-30MHz when the power of the laser was set to 69 and 95 μW. These plots indicate a clear roll off frequency around 300 kHz. In addition, the plots beginning at 1kHz display unstable behavior at frequencies below 30kHz. I am not sure why there is such a sharp change in the transfer function around 30kHz, but I suspect this to be due to an issue with the agilent or photodiode. 

Attachment 1: PD_TF1.pdf
PD_TF1.pdf
Attachment 2: PD_TF2.pdf
PD_TF2.pdf
Attachment 3: PD_and_TIA_Transfer_Function_Measurements.zip
  14110   Sat Jul 28 00:45:11 2018 terra, sandrineSummaryThermal CompensationHeater measurements overview

[Sandrine, Koji, Terra]

Summary: We completed multiple scans at different heating powers for the reflector set up, observing unique HOM peak shifts of tens of kHz. We also observed HOM5 shifts with the cylinder set up. Initial Lorentzian fittings of the magnitude give tens of Hz resolution. I summarize the main week's work below. 

Set-up

Heater set-up is described in several previous elogs, but attachments #1 and #2 show the full heater set-up and wiring/pinouts in and out of vacuum, since we're all intimately aware of how confusing in-vacuum pinouts can be. We are not using the Sorenson power supply (as described in 14071); we just have the BKPrecision power supply 1735 sitting next to the ETMY rack and are manually going out to turn on/off. 

We've continued to use the scan setup described in elog 14086, which is run using /users/annalisa/postVent/AGfast.py. Step by step notes for setting up the scan, running the scans, and processing the scans are attached in notes.txt.

Inducing/witnessing HOMs

The aux input beam was already clipped and on wednesday (after Trans was centered, 14093) we also clipped the output aux beam with razor blade (angled vertically and horizontally, elog 14103) before PDA255; we clipped ~1/3 of the output beam. Attachment #3 shows before and after clipping output, where orange 'cold' == unclipped, black 'mean' == clipped (all in cold state). Up to HOM5 is visible. 

Measurements

Below is a summary of the available scan data. We also have cold (0A) scans CAR-HOM5 and full FSR scans for most configurations. 

Elliptic Reflector
current[A] voltage[V] power[W] scans
0.4 2 0.8 CAR-HOM3(x1)
0.5 3.4 1.7 CAR-HOM3(x1)
0.6 5 3.0 CAR-HOM3(x1)
0.8 9.4(9.7) 7.5(7.8) CAR-HOM5(>x5)
0.9 12 10.8 CAR-HOM5(x4)
1.09 17 18.5 CAR-HOM3

 

 

 

 

 

 

 

Cylinder + Lenses
current[A] voltage[V] power[W] scans
0.9 15 13.5 CAR-HOM5(odds x4)

We tried the cylinder set-up again tonight for the first time since inital try and can see shifts of HOM5 - see attachment #5; we haven't looked in detail yet, but it looks like odd modes are more effected, suggesting the ring heat pattern is off centered from the beam axis. 

Scan data is saved in the following format: users/annalisa/postVent/scandata/{reflector,cylinder}/{parsed,unparsed}/{CAR,HOM1,HOM2,HOM3,HOM4,HOM5}{_datetime}{_parsed,_unparsed}.{txt,pdf}

Minimum heating

On 7/26 we increased the power to the elliptical reflector heater in steps to find the minimum heater power required to see frequency shifts with our measurement setup. Lowest we can resolve is a shift in HOM3 with 1.7W (0.5A/3.4V). According to Annalisa's measurements in elog 14050, this would be something like 30-60 mW radiated power hitting the test mass. We only looked at CAR - HOM3 for this investigation; data for scans at 0.4A, 0.5A, 0.6A is available as indicated above.

Lorentizian Fitting

The Lorentzian fitting was done using the equation a + b / sqrt(1+((x-c)/d*2), where a = constant background, b = peak height above background, c = peak frequency, d = full width at half max. 

The fitting is still being edited and optimized. We will crop the data to zoom in around the peak more.

The Lorentzian fit of the magnitude shows ~10Hz of resolution. (See attachment 6 for the carrier at 8A and attachment 7 for HOM 1 at 9A)

We're working on fitting the full complex data.

 

 

Attachment 1: heater_setup.jpg
heater_setup.jpg
Attachment 2: heater_wiring.jpg
heater_wiring.jpg
Attachment 3: notes.txt
Notes for running scans:
1. when first turning on Agilent, set initial stuff
    > cd /users/annalisa/postVent/20180718
    > AGmeasure TFAG4395Atemplate.yml
2. tweak arm alignment and offset PLL
    > sitemap (then IFO --> ALIGN and also PSL --> AUX)
    > to increase 
3. make sure X-arm is misagligned (hit '! Misalign' button for ITMX, ETMX) 
3. run scan
    > python AGfast.py startfreq stopfreq points
... 36 more lines ...
Attachment 4: FSR_clipped.pdf
FSR_clipped.pdf
Attachment 5: cylinderHOM5.pdf
cylinderHOM5.pdf
Attachment 6: pt8A_CAR.pdf
pt8A_CAR.pdf
Attachment 7: pt9A_HOM1.pdf
pt9A_HOM1.pdf
  14109   Fri Jul 27 17:16:14 2018 SandrineUpdateThermal CompensationCopied working scripts for mode spectroscopy into new directory (modeSpec)

The scripts: AGfast.py, make HDF5.py, plotSpec_marconi.py, and SandrineFitv3.py were copied into the new directory modeSpec.

The path is: /opt/rtcds/caltech/c1/scripts/modeSpec

These scripts can still be found under Annalisa's directory under postVent.

  14108   Fri Jul 27 10:48:57 2018 SteveUpdateSUSBS oplev window

Yesterday I inspected this BS oplev viewport. The heavy connector tube was shorting to table so It was moved back towards the chamber. The connection is air tight with kapton tape temporarly.

 The beam paths are well centered. The viewport is dusty on the inside.

The motivation was to improve the oplev noise.

Attachment 1: BSOw_.jpg
BSOw_.jpg
Attachment 2: dustInsideBSO.jpg
dustInsideBSO.jpg
  14107   Fri Jul 27 02:30:51 2018 gautamUpdateGeneralGlitchy MC

Kevin and I saw some weird IMC / PEM BLRMS behaviour today - see Attached screenshot. Not sure what was happening with the IMC, but MCtrans was oscillating at ~3Hz for a good 20 minutes or so. I just killed the lock, and restarted MCautolocker on megatron. There was a strange feature in the 3-10Hz BLRMS around that time as well. All seems back to normal now...

Attachment 1: 38.png
38.png
  14106   Thu Jul 26 15:11:18 2018 SteveUpdateGeneral Viewports & coating of 2001

New optical quality BK-7 windows in 2001 [4 substrates ] AR coated R<0.75 % for 630-1064nm " Azury BLue" broadband : TRX, TRY, ITMY-Oplev &  ITMX-Oplev viewports.

The BS-Oplev and PRM-Oplev 10" CF with 5.38" diameter view was coated the same way. The window here is Corning 7056 Borosilicate

5 more BK-7 substrates were coated R <0.1% of 1064 nm "Golden Orange" Their location: IMC-IN, IFO-REF and OMC   The next vent we have to confirm optical quality window locations.

All other conflat flange viewports are 7056 Kovar sealed .

Technical notes of 2001 40m upgrade can be seen at LIGO-T010115- 00- R  ....page 14

Attachment 1: BK7window_Coatings.PDF
BK7window_Coatings.PDF BK7window_Coatings.PDF
  14105   Thu Jul 26 01:52:01 2018 terraUpdateThermal Compensationheater work update

Just a quick update: over the past few days we've taken (at least) 5 scans around each peak [carrier - HOM3] at 9.4V/0.8A, 4 scans around [carrier - HOM5] at 12V/0.9A hot state with the reflector setup. We also have (at least) 5 scans of carrier - HOM5 in cold state. I attach a rough overview of the peak magnitude shifts in the first attachment. Analysis ongoing. All data stored in annalisa/postVent/{date}

Initial shifts just based on rought peak placement in the meantime:

            [9.4V/0.8A]   [12V/0.9A]

HOM1    10 kHz         20 kHz

HOM2    18 kHz         28 kHz

HOM3     30 kHz        40 kHz

HOM4     N/A             26 kHz

HOM5     N/A             35 kHz

I also attach the heating thermal transient from today (12V/0.9A) as seen by the opLevs. We see a shorter time constant for pitch, longer for yaw, preceeded by a dip in yaw. Similar behavior yesterday for slightly less heating, though less pronounced pre-dip. The heater is offcentered on the optic horizontally; likely this is part of the induced yaw. The spikey stuff i removed is from people walking around inside during the transient.

I've left the heater and LSC off for the night. Heater off at 2:07 am local time.

Please don't touch the oplevs; we're taking a cool down measurement.

Attachment 1: OpLev_thermal_drift.pdf
OpLev_thermal_drift.pdf
Attachment 2: hotColdAll.pdf
hotColdAll.pdf
  14104   Wed Jul 25 22:46:15 2018 gautamConfigurationComputersNDS access from outside

After this work, I've been having some trouble getting data with Python NDS. Eventually, I figured out that the nds connection request has to be pointed at '131.215.115.200' (the address of the NAT router which faces the outside world), port 31200 (it used to work with 'nds40.ligo.caltech.edu' or '131.215.115.189'). So the following snippet in python allows a connection to be opened. Offline access of frame data via NDS2 now seems possible.

import nds2
conn = nds2.connection('131.215.115.200',31200)
Quote:
 

So far, ssh (22), web services (30889), and elog (8081, 8080) were tested. We also need to test megatron NDS port forwarding and rsync for nodus, too.Finally I turned off the firewall rules of shorewall on nodus as it is no longer necessary.

  14103   Wed Jul 25 14:45:59 2018 SandrineSummaryThermal CompensationETM Y Table AUX read out

Attached is a photo of the set up of the ETM Y table showing the AUX read out set up. 

Currently, the flip mount sends the AUX to the PDA255. Terra inserted a razor blade so the PDA255 will witness more HOMs. The laser is also sent to the regular PD and the CCD.

Attachment 1: EY_table_.JPG
EY_table_.JPG
  14101   Tue Jul 24 09:47:51 2018 gautamUpdateCamerasDeveloping neural networks on simulated video

I was thinking a little more about the way we are training the network for the current topology - because the network has no recurrent layers, I guess it has no memory of past samples, and so it doesn't have any sense of the temporal axis. In fact, Keras by default shuffles the training data you give it randomly so the time ordering is lost. So the training amounts to requiring the network to identify the center of the Gaussian beam and output that. So in the training dataset, all we need is good (spatial) coverage of the area in which the spot is most likely to move? Or is the idea to develop some tools to generate video with spot motion close to that on the ETM in lock, so that we can use it with a network topology that has memory? 

Quote:

This looks like good progress. Instead of fixed sines or random noise, you should generate now a time series for the motion which is random noise but with a power spectrum similar to what we see for the ETM pitch motion in lock. You can use inverse FFT to get the time series from the open loop OL spectra (being careful about edge effects)

  14100   Tue Jul 24 06:11:50 2018 ranaUpdateCamerasDeveloping neural networks on simulated video

This looks like good progress. Instead of fixed sines or random noise, you should generate now a time series for the motion which is random noise but with a power spectrum similar to what we see for the ETM pitch motion in lock. You can use inverse FFT to get the time series from the open loop OL spectra (being careful about edge effects).

Quote:

Aim: To develop a neural network that resolves mirror motion from video.

  14098   Mon Jul 23 09:58:52 2018 SteveSummaryVACRGA scan at day 6

 

 

Attachment 1: pd81-560Hz-d6.png
pd81-560Hz-d6.png
  14097   Sun Jul 22 14:01:07 2018 poojaUpdateCamerasDeveloping neural networks on simulated video

Aim: To develop a neural network that resolves mirror motion from video.

Since error was high for the same input as in my previous elog http://nodus.ligo.caltech.edu:8080/40m/14089

I modified the network topology by tuning the number of nodes, layers and learning rate so that the model fitted the sum of 4 sine waves efficiently, saved weights of the final epoch and then in a different program, loaded saved weights & tested on simulated video that's produced by moving beam spot from the centre of image by sum of 4 sine waves whose frequencies and amplitudes change with time.

Input : Simulated video of beam spot motion in pitch by applying 4 sine  waves of frquencies 0.2, 0.4, 0.1, 0.3 Hz  and amplitude ratios to frame size to be 0.1, 0.04, 0.05, 0.08. This is divided into train (0.4), validation (0.1) and test (0.5).

Model topology:

                                          Input               -->                  Hidden layer               -->                    Output layer                                  

                                                                                          8 nodes                                              1 node

Activation function:                                  selu                                             linear

Batch size = 32, Number of epochs = 128, loss function = mean squared error

Optimizer: Nadam ( learning rate = 0.00001, beta_1 = 0.8, beta_2 = 0.85)

Normalized the target sine signal of NN by dividing by its maximum value.

Plot of predicted output by neural network, applied input signal & residual error given in 1st attachment. These weights of the model in the final epoch have been saved to h5 file and then loaded & tested with simulated data of 4 sine waves with amplitudes and frequencies changing with time from their initial values by random uniform noise ranging from 0 to 0.05. Plot of predicted output by neural network, target signal of sine waves & residual error given in 2nd attachment. The actual signal can be got from predicted output of NN by multiplication with normalization constant used before. However, even though network fits training  & validation sets efficiently, it gives a comparatively large error on test data of varying amplitude & frequency.

Gautam suggested to try training on this noisy data of varying amplitudes and frequencies. The results using the same model of NN is given in Attachment 3. It was found that tuning the number of nodes, layers or learning rate didn't improve fitting much in this case.

 

 

Attachment 1: nn_simulation_2_normalized_mult_sin_nodes8_128epochs_lr0p00001_beta1_0p8_beta2_0p85_0p4train_0p1valid_marked.pdf
nn_simulation_2_normalized_mult_sin_nodes8_128epochs_lr0p00001_beta1_0p8_beta2_0p85_0p4train_0p1valid_marked.pdf
Attachment 2: nn_simulation_normalizedtarget_128epochs_mult_sin_load_wt_varyingtest_nodes8_lr0p00001_beta1_0p8_beta2_0p85_0p4train_0p1valid_marked.pdf
nn_simulation_normalizedtarget_128epochs_mult_sin_load_wt_varyingtest_nodes8_lr0p00001_beta1_0p8_beta2_0p85_0p4train_0p1valid_marked.pdf
Attachment 3: nn_simulation_2_normalized_varying_mult_sin_nodes8_128epochs_lr0p00001_beta1_0p8_beta2_0p85_0p4train_0p1valid_marked.pdf
nn_simulation_2_normalized_varying_mult_sin_nodes8_128epochs_lr0p00001_beta1_0p8_beta2_0p85_0p4train_0p1valid_marked.pdf
  14096   Sat Jul 21 14:03:19 2018 KojiSummaryThermal CompensationY arm locking

Ah. With MC2 feedback, we have about 3 times smaller "optical gain" for the ASS A2L. We have same dither, same actuator, but we need only 1/3 actuation of the MC2 compared to the ETMY case.
We had to reduce the ASS spot servo from 1 to 0.3 to make is stable, so this means that the ASS is really merginally stable.

  14095   Sat Jul 21 01:14:02 2018 gautamUpdateOMCPZT Jena driver board check

[Aaron, gautam]

We did a quick check of this board today. Main takeaways:

  • There are two voltages (HV pos and HV neg) that are output from this unit.
  • Presumably, these goto different piezoelectric elements, referenced to ground. Are there any spec sheets for these describing the geometry/threshold voltages?
  • The outputs are:
    • \mathrm{HV_{+}} = 10(V_{\mathrm{DAC}}+V_{\mathrm{offset}}), \mathrm{HV_{-}} = 10(-V_{\mathrm{DAC}}+V_{\mathrm{offset}})
    • So with V_{\mathrm{offset}} = 7.5 \mathrm{V}, we expect to be able to use +/- 7.5 V of DAC range.
  • The trim pot had to be adjusted to realize V_{\mathrm{offset}} = 7.5 \mathrm{V}​.
  • I assume 150V is some kind of damage threshold of the PZT, so there is no benefit to using 10V offset voltage (as this would result in 200 V at full range DAC voltages).

With the correct V_{\mathrm{offset}} = 7.5 \mathrm{V}, we expect 0V from the DAC to result in 0 actuation on the mirror, assuming that an equal 75V goes to 2 PZTs mounted diametrically opposite on the optic. Hopefully, this means we have sufficient range to scan the input pointing into the OMC and get some sort of signal in the REFL signal (while length PZT is being scanned) which indicates a resonance. 

We plan to carve out some IFO time for this work next week.

  14094   Sat Jul 21 01:06:49 2018 gautamSummaryThermal CompensationY arm locking

I implemented this today. For now, the LSC output matrix is set to actuate on MC2 for Y arm locking. As expected, the transmission was much more stable, and the PLL control signal RMS was also reduced by factor of ~3. MC2 control signal does pick up a large (~2000 cts) DC component over a few hours, so we need to relieve this periodically.

Now that we have a workable ASS for the Y arm as well, we should be able to have more confidence in returning to the same beam spot position on the ETM and staying there during a scan using this technique.

The main improvement to be trialled next in the scanning is to improve the speed of scanning. As things stand, my script takes ~2.5 seconds per datapoint. If we can cut this in half, that'd be huge. On Wednesday night, we were extraordinarily lucky to avoid MC3 glitching, EPICS/slow machine failures, and GPIB freezes. Today, the latter reared its head. Fortunately, since I'm dumping data to file for each datapoint, this means we at least have data till the GPIB freeze.

Quote:

For future measurements, we should consider locking the IMC length to the arm cavity - this would eliminate such alignment drifts, and maybe also make the PLL control signal RMS smaller. 


Not related to this work: Terra, Sandrine, Keerthana and I cleaned up the lab a bit today, and made better cable labels. Aaron and I have to clean up the OMC area a bit. Huge thanks to Steve for taking care of our mess elsewhere in the lab!

  14093   Fri Jul 20 22:53:15 2018 KojiUpdateASSAttempt to resurrect Yarm ASS

[Koji Gautam]

We managed to realize stable ASS configuration for Yarm. The transmission of 1.06~1.07 was recovered by introducing intentional beam spot offset in the horizontal direction towards the opposite side of the elliptic reflector. The end table optics were adjusted to have the spots about the center of the mirrors, lenses, and PDs/QPDs.


Preparation

- The Y arm was manually aligned with a given input axis. The transmission was ~0.8.
- Then, TT2 was moved in yaw such that it introduced the horizontal beam shift at the end. By moving the spot to the opposite side of the reflector. The transmission ~0.95 was obtained after patient alignment work.

- Went to the end table and checked the spots. The beam was not at the center of the last 1" lens for the Trans PDs. The beam steering was adjusted to have the spot nicely going through the lens and the mirrors. This made the transmission level to be ~1.05.

- The beam centering on the Trans PD was checked and adjusted.
- The beam centering on the RF BBPD for the arm scan was checked. The spot was too big for that PD. The lens was slightly moved away from the PD to make the spot on the BBPD small. Now the PD saw the plateu when the steering was scanned (i.e. the spot is small enough).

- With the Y arm locked with MC2, the servo gain needs to be 0.012 instead of nominal 0.015 with ETMY to prevent from servo oscilating.

ASS tuning

- First of all, only the bottom 4 loops out of total 8 loops were tuned. They are the servos for the beam alignment with regard to the caivty. The linearity and the zero crossings were checked with regard to the reference alignment. All of these 4 showed offsets that causes the servo running away. Don't know the reason of this offset, but it is freq dependent. Therefore the dither freqs were tuned to make the offset zeroed, and tuned the demod phases there. This kept the transmission as high as the reference (~1.05)

- This allowed us to play with the spot position a bit by tuning the caivty alignment. In the end, the transmission of ~1.08 was obtained. Using this alignment, A2L offset for ETMY Yaw was determined to be +17 (to make the error signal -17). This offset produces almost a beam radius (5mm) shifted on the end mirror towards the opposite direction of the reflector.

- The nominal servo setting made the spot servo running away. Gautam pointed out that this could be a gain hierarchy problem (i.e. the spot servos are too fast). We ended up reducing the gain of the servo from 1.0 to 0.3 to make the spot servo stable.

- All the ASS setting was stored in a new snap file "script/ASS/ASS-DITEHR_ON.snap". The previous snap was saved to "script/ASS/ASS_DITHER_ON_preVent201807.snap". This did not save the exc gains of the oscillators. Therefore "DITHER_ASS_ON.py" was modified to have the new exc gains (CLKGAIN). The old values are stored in the comments in this script.


Overall this is not an ideal situation as we don't know what is the actually cause of the offsets in the dither error signals. We expect to correct the beam clipping and the suspension sooner or later. Therefore, we will come back to the ASS again once the other issues are corrected.

Attachment 1: 02.png
02.png
  14092   Fri Jul 20 22:51:28 2018 KojiUpdateIOOIMC WFS path alignment

IMC WFS tuning

- IMC was aligned manually to have maximum output and also spot at the center of the end QPD.
- The IMC WFS spots were aligned to be the center of the WFS QPDs.
- With the good alignment, WFS RF offset and MC2 QPD offsets were tuned via the scripts.

  14091   Fri Jul 20 18:30:47 2018 JonConfigurationAUXRecommend to install AUX PZT driver

I recently realized that the PLL is only using about 20% of the available actuation range of the AUX PZT. The +/-10 V control signal from the LB1005 is being directly inputted into the fast AUX PZT channel, which has an input range of +/-50 V.

I recommend to install a PZT driver (amplifier) between the controller and laser to use the full available actuator range. For cavity scans, this will increase the available sweep range from +/-50 MHz to +/-250MHz. This has a unique advantage even if slow temperature feedback is also implemented. To sample faster than the timescale of most of the angular noise,  scans generally need to be made with a total sweep time <1 sec. This is faster than the PLL offset can be offloaded via the slow temperature control, so the only way to scan more than 100 MHz in one measurement is with a larger dynamic range.

  14090   Fri Jul 20 07:43:54 2018 SteveSummarySUSETMY

 

 

Attachment 1: ETMY_leveling.png
ETMY_leveling.png
Attachment 2: ETMY.png
ETMY.png
  14089   Thu Jul 19 18:09:17 2018 poojaUpdateCamerasUpdate in developing neural networks

Aim: To develop a neural network that resolves mirror motion from video.

Case 1:

Input : Simulated video of beam spot motion in pitch by applying 4 sine  waves of frquencies 0.2, 0.4, 0.1, 0.3 Hz  and amplitude ratios to frame size to be 0.1, 0.04, 0.05, 0.08

The data has been split into train, validation and test datasets and I tried training on neural network with the same model topology & parameters as in my previous elog (https://nodus.ligo.caltech.edu:8081/40m/14070)

The output of NN and residual error have been shown in Attachment 1. This NN model gives a large error for this. So I think we have to increase the number of nodes and learning rate so that we get a lower error value with a single sine wave simulated video ( but not overfitting) and then try training on linear combination of sine waves.

Case 2 :

Normalized the target sine signal of NN so that it ranges from -1 to 1 and then trained on the same neural network as in my previous elog with simulated video created using single sine wave. This gave comparatively lower error (shown in Attachment 2). But if we train using this network, we can get only the frequency of test mass motion but we can't resolve the amount by which test mass moves. So I'm unclear about whether we can use this.

Attachment 1: nn_simulation_mlt_sine_nodes4_lr0p00001_beta1_0p8_beta2_0p85_marked.pdf
nn_simulation_mlt_sine_nodes4_lr0p00001_beta1_0p8_beta2_0p85_marked.pdf
Attachment 2: nn_simulation_2_nodes4_target-1to1_marked.pdf
nn_simulation_2_nodes4_target-1to1_marked.pdf
  14088   Thu Jul 19 13:35:30 2018 SteveSummaryVACannuloses pumped

Roughing down the annuloses required closing V1 for 13 minutes

IFO is 2.2e-5 Torr

Attachment 1: AnsPumped.png
AnsPumped.png
  14087   Thu Jul 19 11:01:03 2018 SteveSummaryVACpd81 @ 2e-5 Torr

Cold cathode gauge just turned on.

Attachment 1: pd81@2days.png
pd81@2days.png
  14086   Thu Jul 19 04:44:09 2018 Annalisa, TerraSummaryThermal Compensationfrequency shift observed with heating!

Annalisa, Gautum, Koji, Terra

Summary: with the reflector setup, we measured a frequency shift of the first and second order modes! First looks of shifts show 1st HOM shift ~-10 kHz, 2nd HOM shift ~-18 kHz (carrier ~4 kHz). We saw no shift with the cylinder/lenses set up.

- - - - -

Tonight we modified the cavity scan setup: the LO is provided by the Marconi which, at the same time, is also used to scan the AUX laser frequency instead of the Agilent. In order to get rid of the free running noise between Marconi and Agilent, the Marconi frequency was scanned and, point by point, the Agilent center frequency was changed accordingly. In order to speed up the process, the whole procedure was automated. The script is called AGfast.py and can be found in /users/annalisa/postVent.

One thing that helped in improving the data quality of the phase information was to set the Agilent IF bandwidth @1kHz. Not yet clear why, but it was better than having a lower bandwidth. To be further investigated.

With this setup, we made some coarse scan of the full FSR and then we "zoomed" around the main peaks in order to increase the resolution and get a more precise information about the peak frequency.

Here are the frequency ranges that we scanned:

  • carrier - central frequency: 31.73MHz; range: [31.68MHz - 31.78MHz]
  • HOM1 - central frequency: 32.88MHz; range: [32.84MHz - 32.93MHz]
  • HOM2 - central frequency: 34.03MHz; range: [33.95MHz - 34.06MHz]
  • HOM3 - central frequency: 35.18MHz; range: [35.09MHz - 35.25MHz]

We powered the heater of the lenses setup @4:55 am at 14.4V and 0.9A. Then we slightly increased the power @5:05am and the final "hot state" configuration is with heater powered at 16V and 0.9A.

With this setup we couldn't see any frequency shift

Then, at around 6:30 am we turned on the reflector setup and we measured a frequency shift of the first and second order modes. First scans show 1st HOM shift ~10 kHz, 2nd HOM shift ~18 kHz. First attachment shows carrier hot/cold, second attachment shows HOM2 hot/cold. We started to get plauged by high seismic noise. Heaters turned off at 7:45 am. Lots of scans and actual analysis to go.


gautam: about the questionable plotting -

  • 10 faint (alpha~0.3) lines are individual measurements with the reflector doing its heating. (AG4395A, 0 span, single frequency measurements plotted together).
  • charcoal line, labelled mean, is the mean of the 10 above lines.
  • bright green ("Reference") is the mean of a coarse scan (cold ETM) overlaid for comparison. 
  • "cold" - self explanatory.

My personal favourite plot is Attachment #3, which is a 5 MHz scan (cold) to identify positions of the various peaks. The power of including phase information in the analysis is clear. The second FSR on the right edge of the plot is not as prominent as the first is because the arm transmission was degrading throughout the measurement. For future measurements, we should consider locking the IMC length to the arm cavity - this would eliminate such alignment drifts, and maybe also make the PLL control signal RMS smaller. 

Attachment 1: scanning_fine_2018-07-19-07-32-08_parsed.pdf
scanning_fine_2018-07-19-07-32-08_parsed.pdf
Attachment 2: scanning_fine_2018-07-19-06-57-47_parsed.pdf
scanning_fine_2018-07-19-06-57-47_parsed.pdf
Attachment 3: Yscan_scanning_parsed_2am.txt.pdf
Yscan_scanning_parsed_2am.txt.pdf
  14085   Thu Jul 19 01:56:25 2018 gautamSummaryVACAUX pump shutdown

[koji, gautam]

Per Steve's instructions, we did the following:

  • TP3fl pressure reading was 65 torr.
  • TP3 controller reported pumping current of ~0.18A, temperature of 24C.
  • We throttled the manual valve which was connecting the "AUX" pump to the TP3fl.
  • The TP3fl pressure went up to 330 torr.
  • TP3fl controller reported current of 0.22A, temperature of 24C.
  • After ~5mins, we shut the AUX pump off.
  • We have monitored it over the last 1hour, no red flags.
    • (Before stopping AUX RP)
      0:56AM TP3 I=0.18A, P=6W, 23degC, TP3FL: 66
    • 0:59AM TP3 I=0.22A, P=7W, 23degC, TP3FL: 336
    • 1:15AM TP3 I=0.21A, P=7W, 23degC, TP3FL: 320
    • 1:31AM TP3 I=0.21A, P=7W, 23degC, TP3FL: 310
    • 2:06AM TP3 I=0.21A, P=7W, 23degC, TP3FL: 301
    • 5:06AM TP3 I=0.21A, P=7W, 23degC, TP3FL: 275
  14084   Wed Jul 18 23:43:50 2018 KojiUpdateGeneralVent 80 recovery

Is the reflector too close to the beam and causing clipping?

Quote:

For unknown reasons, the Y arm ASS does not maximize TRY. So we are in the unfortunate situation of neither arm having a working ASS servo. To be worked on later.

Attachment 1: IMG_5868.JPG
IMG_5868.JPG
Attachment 2: IMG_5382.JPG
IMG_5382.JPG
  14083   Wed Jul 18 17:36:50 2018 SteveSummaryVACpumpdown 81 at 6 +9 hrs completed

IFO P1 6e-4 Torr,  manual gate valve is fully open

The annuloses will be pumped down tomorrow.

Valve configuration: vacuum normal, RGA and annuloses are not pumped

Quote:

The manual gate valve scan was clean. Atm1     TP1 was pumping on it overnight.

                                                Pumpdown continued to hand over the pumping to TP1 maglev turbo

V1 was opened at P1 400 mTorr  with manual gate at 3/4 turn open position as Magev at 560 Hz rotation.

This is the first time we pumping down from atm with one small "beer can" turbo  and throttled gate to control load on small turbo forepump

The 70 l/s turbo is operating at 50k RPM, 0.7 A and 31 C,  pumping speed  ~ 44 mTorr/h at 200-400 mTorr range.

Watching foreline pressures and current one can keep opening gate valve little by little the so the load is optimized. It is working but not fast.

Let's keep small turbo at 0.8 Amp and 32 C max at this pumpdown. 

Quote:

10:20PM

  • Opened VM2 to pump down the RGA section with TP1
  • Stopped rotary roughing pumps
    • Manually closed RV1
    • Closed V3
    • Stopped RP1 and RP3
    • Vented the RP hose

The P1 pressure is 380mTorr. I allowed Gautam to use the full PSL power (~1W).

Attachment 1: pd81completed.png
pd81completed.png
Attachment 2: pd81@30hrs.png
pd81@30hrs.png
  14082   Wed Jul 18 12:49:08 2018 SteveSummaryVACpumpdown 81 at 6 +4.5hrs

The manual gate valve scan was clean. Atm1     TP1 was pumping on it overnight.

                                                Pumpdown continued to hand over the pumping to TP1 maglev turbo

V1 was opened at P1 400 mTorr  with manual gate at 3/4 turn open position as Magev at 560 Hz rotation.

Two aux fans on to hold tubo temps TP1 & TP3 . Atm3

This is the first time we pumping down from atm with ONE small "beer can" turbo  and throttled gate valve to control load on small turbo forepump

The 70 l/s turbo is operating at 50k RPM, 0.7 A and 31 C,  pumping speed  ~ 44 mTorr/h at 200-400 mTorr range with aux drypump in the foreline of TP3

Watching foreline pressures and current one can keep opening gate valve little by little the so the load is optimized. It is working but not fast.

Let's keep small turbo at 0.8 Amp and 32 C max at this pumpdown. 

Quote:

10:20PM

  • Opened VM2 to pump down the RGA section with TP1
  • Stopped rotary roughing pumps
    • Manually closed RV1
    • Closed V3
    • Stopped RP1 and RP3
    • Vented the RP hose

The P1 pressure is 380mTorr. I allowed Gautam to use the full PSL power (~1W).

Attachment 1: manlGateScan.png
manlGateScan.png
Attachment 2: handing_over_Mag.png
handing_over_Mag.png
Attachment 3: TGVw2auxfans_.jpg
TGVw2auxfans_.jpg
  14081   Wed Jul 18 03:14:48 2018 AnnalisaUpdateGeneralVent 80 recovery

[Gautam, Johannes, Koji, Annalisa]

Tonight we increased the power of the PSL laser and we achieved the lock of both arms with high power.

The AUX beam alignment to the Y arm was recovered and the PLL restored (using the Marconi as LO).

We made a quick measurement of the phase noise and the results will be posted tomorrow.

The beam on the PSL has been blocked, as well as the AUX beam on the AS table. The Marconi has been switched off.


gautam:

  1. Before turning up PSL power, I placed a block in front of MC refl to avoid any PD burning. Replaced HR Y1 2" optic with the usual 10% reflective BS to direct MC REFL to the locking PD.
  2. Waveplate was rotated back to 180 deg (original position before the vent). After optimizing PMC transmission, I measured 1.05 W going into the IMC (pre-vent value was 1.07 W, prolly within power meter absolute accuracy).
  3. IMC autolocker restored to usual high power version on megatron.
  4. There seems to be some kind of vacuum interlock in effect that prevents me from opening the PSL shutter via EPICS - I had to toggle the position on the shutter controller under the table. After tonight's work, I returned the controller to the NC state, to avoid any further interference with this interlock code that may prevent pumping in the AM.
  5. PLL gain was re-adjusted to achieve maximum stability (judged by eye) of the beat-note in lock triggered on the Marconi LO signal. Alignment onto the NF beatPD was also tweaked to squeeze out as much beat as possible.
  6. The main objective tonight was to send AUX beam in, recover transmission beat, scan the AUX frequency, and resolve some peaks (MAX HOLD scanning technique, magnitude only for now, no phase info). Thanks to JE's expert fiber alignment and beatnote maximization, we achieved this yes. Annalisa will post a plot tmr. 
  7. For unknown reasons, the Y arm ASS does not maximize TRY. So we are in the unfortunate situation of neither arm having a working ASS servo. To be worked on later.
  14080   Tue Jul 17 22:25:41 2018 KojiSummaryVACpumpdown 81 at 6 hrs

10:20PM

  • Opened VM2 to pump down the RGA section with TP1
  • Stopped rotary roughing pumps
    • Manually closed RV1
    • Closed V3
    • Stopped RP1 and RP3
    • Vented the RP hose

The P1 pressure is 380mTorr. I allowed Gautam to use the full PSL power (~1W).

ELOG V3.1.3-