ID |
Date |
Author |
Type |
Category |
Subject |
14161
|
Tue Aug 14 00:50:32 2018 |
gautam | Update | ASS | X arm ASS still not quite right? | While working on the single arm alignment, I noticed that today, i was able to get the X arm transmission back to ~1.22, and the GTRX to 0.52. These are closer to the values I remember from prior to the vent. Running the dither alignment promptly degrades both the green and IR transmissions. Since the pianosa SL7 upgrade, I can't use the sensoray to capture images, but to me, the spot looks a little off-center in Yaw on ETMX in this configuration, I've tried to show this in the phone grab (Atm #2). Maybe indicative of clipping somewhere upstream of ITMX?
Anyways, I'm pushing onwards for now, something to check out in the daytime.
Quote: |
[koji, gautam]
After I effected the series resistance change for ETMX, the X arm ASS didn't work (i.e. IR transmission would degrade if the servo was run). Today, we succeeded in recovering a functional ASS servo .
We then tried to maximize GTRX using the PZT mirrors, but were only successful in reaching a maximum of 0.41. The value I remember from before the vent was 0.5, and indeed, with the IR alignment not quite optimized before we began this work, I saw GTRX of 0.48. But the IR dither servo signals indicate that the cavity axis may have shifted (spot position on the ITM, which is uncontrolled, seems to have drifred significantly, the Pitch signal doesn't stay on the StripTool scale anymore). So we may have to double check that the transmitted beam isn't falling off the GTRX DC PD.
|
|
Attachment 1: POXPOY.png
|
|
Attachment 2: IMG_7108.JPG
|
|
14160
|
Tue Aug 14 00:27:55 2018 |
gautam | Update | LSC | Locking prep | In preparation for attempting some DRMI locking, I did the following:
- Slow machine reboots for unresponsive c1psl, c1susaux and c1iscaux. The latter requried a manual burtrestore to recover the usual LSC PD whitening settings.
- Shuttered AUX laser (which was on Standby anyways) - we should really install a remotely controllable shutter for this on the AS table.
- Re-aligned PMC (half turn of knob in yaw, full turn in pitch) - IMC transmission 15,000cts ---> 15,600cts.
- Squished sat. box cables at ITMX and ETMX.
Not related to this work, but I turned the Agilent NA off since we aren't using it immediately. |
14159
|
Mon Aug 13 20:21:10 2018 |
aaron | Update | OMC | New DAC for the OMC | [aaron, gautam]
We finished up making the new c1omc model (screenshot attached).
The new channels are only four DAC for ASC into the OMC, and one DAC for the OMC length:
C1:OMC-ASC_PZT1_PIT
C1:OMC-ASC_PZT1_YAW
C1:OMC-ASC_PZT2_PIT
C1:OMC-ASC_PZT2_YAW
C1:OMC-PZT
The model compiles and we can change the channel values, so we are all set to do this OMC scan on the software side. |
Attachment 1: c1omcSCREENSHOT.png
|
|
14158
|
Mon Aug 13 17:20:07 2018 |
aaron | Configuration | Upgrade | Parts list for BHD | I've attached the diagram of what I mean.
There are a couple caveats and changes that would have to be made that are not included in this diagram, because they would be made on different tables.
- I moved MMT2, which means the other MMT optics would have to be adjusted to accomodate this. I checked out the configuration on the BS table and this seems doable with a small rotation of MMT1 and maybe PJ2.
- I don't know the best way to get the OMC REFL beam out... thoughts?
- This diagram is kind of crappy after my edits, someone should tell me how to avoid collapsing all layers when I open these layout diagrams in inkscape, because I ended up editing the layout in Acrobat instead, where the lack of object grouping caused a bunch of the optics to lose one or two lines along the way.
- I didn't include all beam paths explicitly but can if this looks like a good configuration.
- The optic that picks off the transmission through MMT2 will need to move a bit, but there was a clamp in the way; this should be a minor change.
- The optic just before the OMC needs to be moved up a bit.
- The optic after the signal OMMT should be changed to a PBS and translated a bit; this is where the LO and signal beams will combine
Gautam also had some questions about the BHD/OMC timeline and plan. I feel somewhat on shaky ground with the answers, but figured I'd post them so I can be corrected once and for all.
- Is the plan really to use the current OMC setup to make a homodyne measurement?
- I'm not sure where on the timeline the new OMC and BHD switchover are relative to each other. I have been imagining doing the swap to BHD before having a new OMC.
- I thought the current OMC resurrection plan was to do DC readout and not homodyne?
- I think the OMC resurrection plan is leading to DC readout, but when we switch over to BHD we'll be able to operate at the dark fringe. Is that right?
- Is it really possible to use our single OMC to clean both the LO and dark port beams? Isn't this the whole raging debate for A+?
- My understanding is yes, with the LO and DP in orthogonal polarizations. It's not clear to me why we expect to be able to do this while there is a debate for A+, perhaps our requirements are weaker. It is something I should calculate, I'll talk to Koji.
- A layout diagram would be really useful.
- Attached now.
- Where in the priority list does this come in?
- I am a leaf in the wind. I think this comes well after we have the OMC resurrected, we just want to get a sense for what parts we need so we can order them before the fiscal year closes.
Quote: |
That seems fine, I wasn't thinking of that beam. in that case could we just have a PBS directly behind MMT2 and send both beams to the same OMMT?
Alternatively we can move OM5 and the beam path OMPO-OMMTSM towards -y, then put the LO-OMMT parallel to the existing OMMT but displaced in +x... we'd have to move the existing OMC and BHD towards +x as well.
Quote: |
Can we use the leakage beam from MMT2 on the OMC table as the LO beam? I can't find the spec for this optic, but the leakage beam was clearly visible on an IR card even with the IMC locked with 100 mW input power so presumably there's enough light there, and this is a cavity transmission beam which presumably has some HOM content filtered out.
|
|
|
Attachment 1: BHD_layout.pdf
|
|
14157
|
Mon Aug 13 11:44:32 2018 |
gautam | Update | Computer Scripts / Programs | Patch updates on nodus | Larry W said that some security issues were flagged on nodus. So I ran
sudo yum upgrade --exclude=elog-3.1.3-2.el7.x86_64
on nodus. The exclude flag is because there were some conflicts related to that particular package. Hopefully this has fixed the problem. It's been a while since the last update, which was in January of this year.
controls@nodus|~> sudo yum history
Loaded plugins: langpacks
ID | Command line | Date and time | Action(s) | Altered
-------------------------------------------------------------------------------
29 | upgrade --exclude=elog-3 | 2018-08-13 11:36 | E, I, U | 136 EE
28 | install yum-utils | 2018-08-13 11:31 | Update | 1
27 | install nmap | 2018-06-29 01:57 | Install | 2
26 | install grace | 2018-05-31 16:52 | Install | 11
25 | install https://dl.fedor | 2018-05-31 16:51 | Install | 1
24 | install perl-Digest-SHA1 | 2018-05-31 15:34 | Install | 1
23 | install python-devel | 2018-01-13 15:33 | Install | 1
22 | install gcc | 2018-01-13 15:32 | Install | 6
21 | install git | 2018-01-12 18:11 | Install | 4
20 | update | 2018-01-12 18:01 | I, U | 39
19 | install motif | 2018-01-05 17:35 | Install | 3
18 | install sendmail sendmai | 2017-12-03 05:11 | Install | 6
17 | install vim | 2017-11-21 18:12 | Install | 3
16 | reinstall mod_dav_svn | 2017-11-21 17:40 | Reinstall | 1
15 | install mod_dav_svn | 2017-11-21 17:39 | Install | 1
14 | install subversion | 2017-11-21 15:36 | Install | 2
13 | -y install php | 2017-11-20 22:15 | Install | 4
12 | install links | 2017-11-20 19:10 | Install | 2
11 | install openssl098e.i686 | 2017-11-18 18:28 | Install | 1
10 | install openssl-libs.i68 | 2017-11-18 18:26 | Install | 11
history list |
14156
|
Mon Aug 13 09:56:23 2018 |
Steve | Update | SUS | ETMX trip follow-up | Here is an other big one
Quote: |
A brief follow-up on this since we discussed this at the meeting yesterday: the attached DV screenshot shows the full 2k data for a period of 2 seconds starting just before the watchdog tripped. It is clear that the timescale of the glitch in the UL channel is much faster (~50 ms) compared to the (presumably mechanical) timescale seen in the other channels of ~250 ms, with the step also being much smaller (a few counts as opposed to the few thousand counts seen in the UL channel, and I guess 1 OSEM count ~ 1 um). All this supports the hypothesis that the problem is electrical and not mechanical (i.e. I think we can rule out the Acromag sending a glitchy signal to the coil and kicking the optic). The watchdog itself gets tripped because the tripping condition is the RMS of the shadow sensor outputs, which presumably exceeds the set threshold when UL glitches by a few thousand counts.
|
|
Attachment 1: ETMXglitch.png
|
|
Attachment 2: ETMXgltch.png
|
|
14155
|
Sun Aug 12 10:59:34 2018 |
aaron | Configuration | Upgrade | Parts list for BHD | That seems fine, I wasn't thinking of that beam. in that case could we just have a PBS directly behind MMT2 and send both beams to the same OMMT?
Alternatively we can move OM5 and the beam path OMPO-OMMTSM towards -y, then put the LO-OMMT parallel to the existing OMMT but displaced in +x... we'd have to move the existing OMC and BHD towards +x as well.
Quote: |
Can we use the leakage beam from MMT2 on the OMC table as the LO beam? I can't find the spec for this optic, but the leakage beam was clearly visible on an IR card even with the IMC locked with 100 mW input power so presumably there's enough light there, and this is a cavity transmission beam which presumably has some HOM content filtered out.
|
|
14154
|
Fri Aug 10 16:43:50 2018 |
gautam | Configuration | Upgrade | Parts list for BHD | Can we use the leakage beam from MMT2 on the OMC table as the LO beam? I can't find the spec for this optic, but the leakage beam was clearly visible on an IR card even with the IMC locked with 100 mW input power so presumably there's enough light there, and this is a cavity transmission beam which presumably has some HOM content filtered out.
Quote: |
My current thought is to use the MC reflection, the beam that heads from MC1 to MCR1, as the LO beam
|
|
14153
|
Fri Aug 10 11:29:39 2018 |
aaron | Configuration | Upgrade | Parts list for BHD | I've started putting together a list of things we'll need to buy to do BHD readout. I'm still messing around with more detailed optics layouts, but wanted to get a list started here so people can let me know if I'm missing any big, obvious categories of goods.
My current plan makes minimal changes to the signal path going to the OMC, and tries to just get the LO beam into the OMC with minimal optics. I'm not thinking of any of the optics as suspended, and it requires several reflections of the LO beam, so probably this is not an excellent configuration, but it's a start for getting the parts list:
- My current thought is to use the MC reflection, the beam that heads from MC1 to MCR1, as the LO beam
- From MCR1, send the LO to a BS that directs it into an MMT, oriented along x (and lets us keep the MC refl PO)
- After the two MMT optics, the beam will be traveling along -x, and can be directed to a mirror that sends it towards -y to two steering mirrors that send it along -x then +x near the top of the table (perpendicular to the signal MMT.
- Then, send it to a PBS, which replaces the mirror directly after the signal MMT. This is where it combines
- Beam is then sent to the steering mirrors to bring it into the OMC
- In parallel, the signal beam is going through the same path it has now, including the pickoff beam, with the one change that we need a HWP somewhere before the PBS, and the PBS replaces the mirror directly after the MMT (and needs to move a bit closer to have the beam properly directed)
I started making a layout of this scheme, but it's probably not going to work so I'm going to make a quick layout of this more major modification instead:
- Both the MCR beam and the AS beam come in about parallel. Send each to a PO mirror.
- The PO mirror directs both beams into parallel MMT aligned along x
- From the MMT, each is directed to a pair of steering mirrors located where the OMC MMT is located now
- From the steering mirrors go to the PBS that combines the signal and LO
- Then to two more steering mirrors to get into the OMC, which may be moved towards +x
- From the OMC go to the BHD PBS
What we need
Optics
HWP for just before the LO combines with the signal
HWP for just before the signal combines with the LO (is this necessary?)
PBS to replace OM5 (combines the LO and the signal)
Two MMT optics
Two piezo-driven TT optics for steering the LO to the PBS
One additional non-piezo optic for between the LOMMT and the LO-TTs
One additional BS to get the LO into the MMT (and to let us still have the PO)
-1 optic—we pick up one mirror that we replace with the PBS
Optomechanics
2x HWP mounts
1x PBS mount
2x mounts for piezo-driven TT
2x MMT optic mounts—I didn’t take a close enough look at these during the vent to know what we need here
2x mounts for ordinary optics
9x clamps for optics mounts (maybe fewer if some are on blocks)
9x posts for optics mounts
Electronics
-
Additional TT driver
-
 HV supply for the new TTs
-
 Are the HWP actively controlled? We might need something to drive those?
-
 Do we have enough DAC/ADC channels?
Questions
These are mostly just miscellaneous
- What of these optics need to be suspended? If we need suspensions on all of the LO optics, including the MMT, I’m not sure we’re going to be able to fit all of this on the table as I envision it…..
- What if anything can we put out of vacuum (HWP for example)?
- Do we actually need two MMT?
|
14152
|
Fri Aug 10 01:10:56 2018 |
gautam | Update | LSC | Some vertex locking restored | For the first time after the whirlwind vent, I managed to lock the PRMI.
- First, I did POX/POY locking, dither aligned the arms to maximize TRX and TRY.
- Next, I misaligned the ETM and tested the Michelson locking
- Since we've lost ~70% of power on the AS55 PD, I set the whitening gain for AS55 I and Q channels to +6dB (old value was 0dB).
- worked alright. In this new config, the peak-to-peak Michelson fringe count is ~80 cts, while I reported ~60cts-pp a couple of months ago, so all seems good on that front.
- But the config script in the IFOconfigure MEDM screen somehow doesn't set the AS55_Q ----> MICH_A element in the LSC input matrix anymore.
- I edited the .snap file for this configuration to set the relevant matrix element EPICS channel to +1.0.
- I also edited the overall loop gain for this configuration from +30 to +2 (for bright fringe, use -2 for dark fringe).
- Feeling adventerous, I decided to try PRMI in the carrier resonant tuning (to be clear, PRCL on REFL11_I, MICH on AS55_Q).
- Finding the REFL spot on the camera took a while since the PRM has been macroscopically misaligned for the mode-scanning
- Went out to the table and centered the REFL beam onto REFL11 and REFL55 PDs - didn't need much tweaking, which is a good sign, since we shouldn't have screwed anything up on the symmetric side by any of the vent activities.
- Restored PRMI locking using the IFOconfigure MEDM screen - lock caught almost immediately.
- Ran the dither alignment servos for MICH and PRCL - BS needed a bit of encouragement to make the dark spot dark, but POP has been pretty stable over ~15mins.
- I didn't take any loop transfer functions, to do.
I don't have the energy to make a DRMI attempt tonight - but the signs are encouraging. I'd like to use the IFO in the next few days to try and recover DRMI locking. The main concern is that the optical path on the AS beam has changed by ~0.3m I estimate. So the demod phase for AS55 may need to be adjusted, but the change due to optical path length only should be ~10degrees so the DRMI locking with the old settings should still work. Perhaps we also want to scan the PRC and SRC with the phase information from the Trans/Refl transfer functions as well.
Don't want to jinx it, but the c1lsc FE models have been stable. Tomorrow, I'd like to re-enable c1cal, since it has some useful channels for NBing. Could c1daf/c1oaf which have significant amounts of custom C code be the culprits? |
Attachment 1: PRMIcarrier.png
|
|
14151
|
Thu Aug 9 22:50:13 2018 |
gautam | Update | CDS | AlignSoft script modified | After this work of increasing the series resistance on ETMX, there have been numerous occassions where the insufficient misalignment of ETMX has caused problems in locking vertex cavities. Today, I modified the script (located at /opt/rtcds/caltech/c1/medm/MISC/ifoalign/AlignSoft.py) to avoid such problems. The way the misalign script works is to write an offset value to the "TO_COIL" filter bank (accessed via "Output Filters" button on the Suspension master MEDM screen - not the most intuitive place to put an offset but okay). So I just increased the value of this offset from 250 counts to 2500 counts (for ETMX only). I checked that the script works, now when both ETMs are misaligned, the AS55Q signal shows a clean Michelson-like sine wave as it fringes instead of having the arm cavity PDH fringes as well .
Note that the svn doesn't seem to work on the newly upgraded SL7 machines: svn status gives me the following output.
svn: E155036: Please see the 'svn upgrade' command
svn: E155036: Working copy '/cvs/cds/rtcds/userapps/trunk/cds/c1/medm/MISC/ifoalign' is too old (format 10, created by Subversion 1.6)
Is it safe to run 'svn upgrade'? Or is it time to migrate to git.ligo.org/40m/scripts? |
Attachment 1: MichelsonFringing.png
|
|
14150
|
Thu Aug 9 12:40:14 2018 |
gautam | Update | SUS | ETMX trip follow-up | A brief follow-up on this since we discussed this at the meeting yesterday: the attached DV screenshot shows the full 2k data for a period of 2 seconds starting just before the watchdog tripped. It is clear that the timescale of the glitch in the UL channel is much faster (~50 ms) compared to the (presumably mechanical) timescale seen in the other channels of ~250 ms, with the step also being much smaller (a few counts as opposed to the few thousand counts seen in the UL channel, and I guess 1 OSEM count ~ 1 um). All this supports the hypothesis that the problem is electrical and not mechanical (i.e. I think we can rule out the Acromag sending a glitchy signal to the coil and kicking the optic). The watchdog itself gets tripped because the tripping condition is the RMS of the shadow sensor outputs, which presumably exceeds the set threshold when UL glitches by a few thousand counts. |
Attachment 1: ETMXglitch.png
|
|
14149
|
Thu Aug 9 12:31:13 2018 |
gautam | Update | CDS | CDS status update | The model seems to have run without issues overnight. Not completely related, but the MC1 shadow sensor signals also don't show any abnormal excursions to negative values in the last 48 hours. I'm thinking about re-connecting the satellite box (but preserving the breakout setup at 1X6 for a while longer) and re-locking the IMC. I'll also start c1ass on the c1lsc frontend. I would say that the other models on c1lsc (i.e. c1oaf, c1cal, c1daf) aren't really necessary for basic IFO operation.
Quote: |
As part of this slow but systematic debugging, I am turning on the c1lsc model overnight to see if the model crashes return.
|
|
14148
|
Thu Aug 9 02:12:13 2018 |
gautam | Update | COC | South East or West? | Summary:
For operating the SRC in the "Signal-Recycled" tuning, the SRC macroscopic length needs to be ~4.04m (compared to the current value of ~5.399m), assuming we don't do anything fancy like change the modulation frequencies and not transmit through the IMC. We're putting together a notebook with all the calculations, but today I was thinking about what the signal extraction path should be, specifically which chamber the SRM should be in. Just noting down the thoughts I had here while they're fresh in my head, all this has to be fleshed out, maybe I'm making this out to be more of a problem than it actually is.
Details:
- For the current modulation frequencies, if we want the reosnance conditions such that the f2 sideband is resonant in the SRC (but not f1, i.e. small Schnupp asymmetry regime) while the carrier is resonant in the arms (required for good sensing of the SRC length), the macroscopic length of the SRC needs to be changed to ~4.04m.
- Practically, this means that the folded SRC would only have one folding mirror (SR2).
- There is a shorter SRC length of ~1.something metres which would work, but that would involve changing the relative position between ITMs and BS (currently ~2.3m) so I reject that option for now.
- So the SR2 would be roughly where it is right now, ~20cm from the BS.
- The question then becomes, where do we direct the reflection from the SR2? We need an optical path length of ~1.5m from SR2. So options are
- ITMY table (East)
- ITMX table (South)
- IMC table (West)
- Moreover, after the SRM, we have to accommodate:
- Some kind of pickoff for in-air PDs.
- OFI.
- OMC MMT.
- OMC.
- Some kind of CBA (as of now I think going to the ITMY table is the best option):
Option |
Advantages |
Disadvantages |
ITMY |
- Easy to direct beam from BS/PRM chamber to the ITMY table (i.e. we don't have to worry too much about avoiding other optics in the path etc).
- Ease of access to chamber, ease of working in there.
- ITMY table probably has the most room to work out an OFI + OMC MMT + OMC solution.
|
- AS beam extraction to air will be more complicated, possibly have to do it on ITMY optical table.
- Not sure if the ITMY table can accommodate all of the output optics subsystems I listed above.
- Routing the LO beam to this table would be tricky I guess.
|
ITMX |
- Routing the LO beam for homodyne detection is probably easiest in this chamber.
- Allows for small AoI on folding mirror, reducing the impact of astigmatism.
|
- Pain to work in this chamber because of IMC tube.
- Steering beam from SR2 to ITMX table means threading the needle between PRM and PR3 possibly.
|
IMC |
- Probably allows the use of (almost) the entire existing OMC chamber for the output optics (OFI, OMC MMT, OMC).
|
- IMC table is crowded (2 SOS towers, several steering optics for the input beam, input faraday).
- Not sure what is the performance of the seismic isolation stacks on these tables vs the larger optical tables.
- Painful to work in these smaller chambers.
|
|
14147
|
Wed Aug 8 23:06:59 2018 |
gautam | Update | SUS | Another low noise bias path idea | Today while Rich Abbott was here, Koji and I had a brief discussion with him about the HV amplifier idea for the coil driver bias path. He gave us some useful tips, perhaps most useful being a topology that he used and tested for an aLIGO ITM ESD driver which we can adapt to our application. It uses a PA95 high voltage amplifier which differs from the PA91 mainly in the output voltage range (up to 900V for the former, "only" 400V for the former. He agrees with the overall design idea of
- Having a LN opamp with the HV amp inside the feedback loop for better voltage noise at low frequencies.
- Having a passive RC network at the output of the HV amp to filter out noise at high frequencies.
He also gave some useful suggestions like
- Using the front panel of the box that as a heatsink for the HV amps.
- Testing the stability of the nested opamp loop by "pinging" the output of the opamp with some pulses from a function generator and monitoring the response to this perturbation on a scope.
I am going to work on making a prototype version of this box for 5 channels that we can test with ETMX. I have been told that the coupling from side coil to longitudinal motion is of the order of 1/30, in which case maybe we only need 4 channels. |
14146
|
Wed Aug 8 23:03:42 2018 |
gautam | Update | CDS | c1lsc model started | As part of this slow but systematic debugging, I am turning on the c1lsc model overnight to see if the model crashes return. |
14145
|
Wed Aug 8 20:56:11 2018 |
Koji | Update | PSL | EOM measuement preparation | Rich and I worked on the EOM measurement. After the measurement, the setup was reverted to the nominal state
- AUX PLL mixer was restored to ZAD-6
- The PLL gain was restored to 3.10
- The main PSL marconi is connected to the freq generator again. Using the beat note, I've confirmed that the modulations are applied on the beam.
- The PSL HEPA was reduced from 100 to 30.
|
14144
|
Tue Aug 7 23:06:30 2018 |
Koji | Update | PSL | EOM measuement preparation | I was preparing for the aLIGO EOM measuement to be carried out tomorrow afternoon.
I did a few modifications to the PLL setup.
- The freq mixier in the PLL setup was replaced with ZP3 (level 7) from ZAD-6
- The PLL gain was reduced from 3.10 to 2.80 to prevent servo oscillation
- The main PSL marconi is connected to the PLL mixier and providing fixed 200MHz 8dBm.
- The main PSL modulation is off.
Tomorrow I am going to modulate the EOM with the AUX Marconi via an amplifier (probably)
Automated scripts (AGinit.py and AGmeas.py) are in /users/koji/scripts
I will revert the setup once the measurement is done tomorrow. |
14143
|
Tue Aug 7 22:28:23 2018 |
gautam | Update | CDS | More CDS woes | I am starting the c1x04 model (IOP) on c1lsc to see how it behaves overnight.
Well, there was apparently an immediate reaction - all the models on c1sus and c1ioo reported an ADC timeout and crashed. I'm going to reboot them and still have c1x04 IOP running, to see what happens.
[97544.431561] c1pem: ADC TIMEOUT 3 8703 63 8767
[97544.431574] c1mcs: ADC TIMEOUT 1 8703 63 8767
[97544.431576] c1sus: ADC TIMEOUT 1 8703 63 8767
[97544.454746] c1rfm: ADC TIMEOUT 0 9033 9 8841
Quote: |
Overnight, all models on c1sus and c1ioo seem to have had no stability issues, supporting the hypothesis that timing issues stem from c1lsc. Moreover, the MC1 shadow sensor readouts showed no negative values over a ~12hour period. I think we should just observe this for another day, in any case I don't think there is any urgent IFO related activity scheduled.
|
|
14142
|
Tue Aug 7 11:30:46 2018 |
gautam | Update | CDS | More CDS woes | Overnight, all models on c1sus and c1ioo seem to have had no stability issues, supporting the hypothesis that timing issues stem from c1lsc. Moreover, the MC1 shadow sensor readouts showed no negative values over a ~12hour period. I think we should just observe this for another day, in any case I don't think there is any urgent IFO related activity scheduled. |
14141
|
Mon Aug 6 20:41:10 2018 |
aaron | Update | DAQ | New DAC for the OMC | Gautam and I tested out the DAC that he installed in the latter half of last week. We confirmed that at least one of the channels is can successfully drive a sine wave (ch10, 1-indexed). We had to measure the output directly on the SCSI connector (breakout in the FE hard drive cabinet along the Y arm), since the SCSI breakout box (D080303) seems not to be working (wiring diagram in Gautam's elog from his SURF years).
I added some DAC channels to our c1omc model:
PZT1_PIT
PZT1_YAW
PZT2_PIT
PZT2_YAQ
And determined that when we go to use the ADC, we will initially want the following channels (even these are probably unnecessary for the very first scans):
TRANS_PD1
TRANS_PD2
REFL_PD
DVMDC (drive voltage monitor, DC level)
DVMAC ("", AC level, only needed if we dither the length)
I attach a screenshot of the model, and a picture of where the whitening/dewhitening boards should go in the rack. |
Attachment 1: OMCDACmdl.png
|
|
14140
|
Mon Aug 6 19:49:09 2018 |
gautam | Update | CDS | More CDS woes | I've left the c1lsc frontend shutdown for now, to see if c1sus and c1ioo can survive without any problems overnight. In parallel, we are going to try and debug the MC1 OSEM Sensor problem - the idea will be to disable the bias voltage to the OSEM LEDs, and see if the readback channels still go below zero, this would be a clear indication that the problem is in the readback transimpedance stage and not the LED. Per the schematic, this can be done by simply disconnecting the two D-sub connectors going to the vacuum flange (this is the configuration in which we usually use the sat box tester kit for example). Attachment #1 shows the current setup at the PD readout board end. The dark DC count (i.e. with the OSEM LEDs off) is ~150 cts, while the nominal level is ~1000 cts, so perhaps this is already indicative of something being broken but let's observe overnight. |
Attachment 1: IMG_7106.JPG
|
|
14139
|
Mon Aug 6 14:38:38 2018 |
gautam | Update | CDS | More CDS woes | Stability was short-lived it seems. When I came in this morning, all models on c1lsc were dead already, and now c1sus is also dead (Attachment #1). Moreover, MC1 shadow sensors failed for a brief period again this afternoon (Attachment #2). I'm going to wait for some CDS experts to take a look at this since any fix I effect seems to be short-lived. For the MC1 shadow sensors, I wonder if the Trillium box (and associated Sorensen) failure somehow damaged the MC1 shadow sensor/coil driver electronics.
Quote: |
Let's see how stable this configuration is. Onto some locking now...
|
|
Attachment 1: CDScrash.png
|
|
Attachment 2: MC1failures.png
|
|
14138
|
Mon Aug 6 09:42:10 2018 |
Koji | Summary | Computers | Transition of the main NFS disk on chiara | Follow up:
- At least it was confirmed that the local backup (4TB->2TB) is regularly running every morning.
- The 2TB disk was used up to 95%. To ease the size of the remaining space, I have further compressed the burt snapshot folders. (~2016). This released another 150GB. The 2TB is currently used up to 87%.
Prev
Filesystem 1K-blocks Used Available Use% Mounted on
/dev/sdc1 3845709644 1731391748 1918967020 48% /home/cds
/dev/sdd1 2113786796 1886162780 120249888 95% /media/40mBackup
Now
Filesystem 1K-blocks Used Available Use% Mounted on
/dev/sdc1 3845709644 1731706744 1918652024 48% /home/cds
/dev/sdd1 2113786796 1728124828 278287840 87% /media/40mBackup
|
14137
|
Mon Aug 6 09:34:02 2018 |
Steve | Update | VAC | RGA scan at day 20 |
|
Attachment 1: pd81d20.png
|
|
14136
|
Mon Aug 6 00:26:21 2018 |
gautam | Update | CDS | More CDS woes | I spent most of today fighting various CDS errors.
- I rebooted c1lsc around 3pm, my goal was to try and do some vertex locking and figure out what the implications were of having only ~30% power we used to have at the AS port.
- Shortly afterwards (~4pm), c1lsc crashed.
- Using the reboot script, I was able to bring everything back up. But the DC lights on c1sus models were all red, and a 0x4000 error was being reported.
- This error is indicative of some timing issue, but all the usual tricks (reboot vertex FEs in various order, restart the mx_streams etc) didn't clear this error.
- I checked the Tempus GPS unit, that didn't report any obvious problems (i.e. front display was showing the correct UTC time).
- Finally, I decided to shut down all watchdogs, soft reboot all the FEs, soft reboot FB, power cycle all expansion chassis.
- This seems to have done the trick - I'm leaving c1oaf disabled for now.
- The remaining red indicators are due to c1dnn and c1oaf being disabled.
Let's see how stable this configuration is. Onto some locking now... |
Attachment 1: CDSoverview.png
|
|
14135
|
Sun Aug 5 15:43:50 2018 |
gautam | Update | SUS | Another low noise bias path idea | OK, how about this:
- Attachment #1 shows the proposed schematic.
- It consists of a second order section with Gain x10 to map the +/-10V DC range of the DAC to +/- 100V DC such that we preserve roughly the same amount of DC actuation range.
- Corner frequency of the SOS is set to ~0.7 Hz. In hindsight, maybe this is more aggressive than necessary, we can tune this.
- DC gain is 20 dB (typo in the text where I say the DC gain is x15, though we could go with this option as well I think if we want a larger series resistance).
- A first order passive low-pass stage is added to filter out the voltage noise of the PA91, which dominates the output voltage noise (next bullet).
- Attachment #2 shows the transfer function from input to output
- The two traces compare having just a single SOS filtering stage vs the current topology of having two SOS stages.
- The passive output RC network is necessary in either case to filter the voltage noise of the PA91 OpAmp.
- For the DAC noise, I just assumed a flat noise level of
, I don't actually know what this is for the Acromag DACs.
- Attachments #3 shows a breakdown of the top 5 noise contributions.
- The PA91 datasheet doesn't give current noise information so I just assumed
, which was what was used for the PA85 in the existing opamp.lib file.
- The voltage noise is modelled as
, which seems to line up okay with the plot on Pg4 of the datasheet.
- So the model suggests we will be dominated by the voltage noise of the PA91.
- Attachment #4 translates the noise into current noise seen by the actuator.
- I add the Johnson noise contribution of the series resistance for this path, which is assumed to be
.
- For comparison, I add the filtered DAC noise contribution, and Johnson noise of the proposed series resistance in the fast path.
- For the bias path, we are dominated by the Johnson noise of the series resistor from ~60 Hz upwards.
- It's not quite fair to say that the Johnson noise of the resistance in the fast path dominates, the quadrature sum of fast and bais paths will be ~1.2 times of the former alone.
- Bottom line: we will be in the regime of total current noise of ~2.2 pA/rtHz, where I think Kevin's modeling suggests we can see some squeezing.
The question still remains of how to combine the fast and bias paths in this proposed scheme. I think the following approach works for prototyping at least:
- Remove the series resistance on the existing coil driver boards' bias path, hence isolating this from the coil.
- Route the DB15 output connector from the coil driver board (which is now just the fast actuation signals) into a sub-sattelite box housing the bias path electronics.
- Sum the two signals as it is done now, by simply having a conductor (PCB trace) merge the two paths after their respective series resistances.
In the longer term, perhaps the Satellite Box revamp can accommodate a bias voltage summation connector.
I have neglected many practical concerns. Some things that come to mind:
- Is it necessary to protect the upstream DAC from some potential failure of the PA91 in which the high voltage appears at the input?
- What is the correct OpAmp for this purpose? This chart on Apex's page suggests that PA15, PA85, PA91 and PA98 are all comparable in terms of drive capability, and the spec sheets don't suggest any dramatic differences. Some LIGO circuits use PA85, some use PA90, but I can't find any that use PA91. Perhaps Rana/Koji can comment about this.
- What kind of protection is necessary for the PA91 power?
- What is the correct way to do heat management? Presumably we need heatsinks, and in fact, there is a variant of the packaging style that has "formed" legs, which from what I can figure out, allow the heat sink plane on the PA91 to be parallel to the PCB surface. But I think the heat-sink wisdom suggests vertical fins are the most efficient (not sure if this holds if the PCB is inside a box though). What about the PCB itself? Are some kind of special traces needed?
- Can we use the current-limiting resistor feature on the PA91? The datasheet seems to advice against it for G>10 configurations, which is what we need, although our requirement is only at DC so I don't know if that table is applicable to this circuit.
- Are 3W resistors sufficient? I think we require only 10mA maximum current to preserve the current actuation range, so 100 V * 10mA = 1W, so 3W leaves some safety margin.
- All capacitors should be rated for 500 V per the datasheet.
|
Attachment 1: HV_Bias_schematic.pdf
|
|
Attachment 2: TF.pdf
|
|
Attachment 3: bias.pdf
|
|
Attachment 4: HVbias_currentNoise.pdf
|
|
14134
|
Sun Aug 5 13:45:00 2018 |
gautam | Update | SUS | ETMX tripped | Independent from the problems the vertex machine has been having (I think, unless it's something happening over the shared memory network), I noticed on Friday that the ETMX watchdog was tripped. Today, once again, the ETMX watchdog was tripped. There is no evidence of any abnormal seismic activity around that time, and anyways, none of the other watchdogs tripped. Attachment #1 shows that this happened ~838am PT today morning. Attachment #2 shows the 2k sensor data around the time of the trip. If the latter is to be believed, there was a big impulse in the UL shadow sensor signal which may have triggered the trip. I'll squish cables and see if that helps - Steve and I did work at the EX electronics rack (1X9) on Friday but this problem precedes our working there... |
Attachment 1: ETMX_tripped.png
|
|
Attachment 2: ETMX_tripped_zoom.png
|
|
14133
|
Sun Aug 5 13:28:43 2018 |
gautam | Update | CDS | c1lsc flaky | Since the lab-wide computer shutdown last Wednesday, all the realtime models running on c1lsc have been flaky. The error is always the same:
[58477.149254] c1cal: ADC TIMEOUT 0 10963 19 11027
[58477.149254] c1daf: ADC TIMEOUT 0 10963 19 11027
[58477.149254] c1ass: ADC TIMEOUT 0 10963 19 11027
[58477.149254] c1oaf: ADC TIMEOUT 0 10963 19 11027
[58477.149254] c1lsc: ADC TIMEOUT 0 10963 19 11027
[58478.148001] c1x04: timeout 0 1000000
[58479.148017] c1x04: timeout 1 1000000
[58479.148017] c1x04: exiting from fe_code()
This has happened at least 4 times since Wednesday. The reboot script makes recovery easier, but doing it once in 2 days is getting annoying, especially since we are running many things (e.g. ASS) in custom configurations which have to be reloaded each time. I wonder why the problem persists even though I've power-cycled the expansion chassis? I want to try and do some IFO characterization today so I'm going to run the reboot script again but I'll get in touch with J Hanks to see if he has any insight (I don't think there are any logfiles on the FEs anyways that I'll wipe out by doing a reboot). I wonder if this problem is connected to DuoTone? But if so, why is c1lsc the only FE with this problem? c1sus also does not have the DuoTone system set up correctly...
The last time this happened, the problem apparently fixed itself so I still don't have any insight as to what is causing the problem in the first place . Maybe I'll try disabling c1oaf since that's the configuration we've been running in for a few weeks. |
14132
|
Fri Aug 3 19:02:11 2018 |
gautam | Update | ASS | X arm ASS recovery | [koji, gautam]
After I effected the series resistance change for ETMX, the X arm ASS didn't work (i.e. IR transmission would degrade if the servo was run). Today, we succeeded in recovering a functional ASS servo .
So both arms have working dither alignment servos now. But remember that the Y arm ASS gains have been set for locking the Y arm with MC2 as the actuator, not ETMY.
Details:
- Koji pointed out that the demodulated signals from the ETM dither are only used to center the spot on the ETM, and that we should first run the servo with existing settings with the ETM pitch and yaw spot centering loops disabled.
- This improved TRX level from ~0.8 to 1.1
- Next, we tried increasing the LO amplitudes by x5 to account for the reduced actuation of the dither on ETMX
- We then re-enabled the two loops that were earlier disabled.
- This resulted in TRX degrading very quickly.
- So we decided to try going back to the nominal LO gains, and reducing the gain of the two ETM spot centering loops.
- This did the trick, TRX went from 1.1 --> ~1.23, which is the nominal maximum pre-vent value.
- The snap file used to recover the correct settings to run the dither alignment servos have been updated, the old one has been backed up with today's datestamp.
We then tried to maximize GTRX using the PZT mirrors, but were only successful in reaching a maximum of 0.41. The value I remember from before the vent was 0.5, and indeed, with the IR alignment not quite optimized before we began this work, I saw GTRX of 0.48. But the IR dither servo signals indicate that the cavity axis may have shifted (spot position on the ITM, which is uncontrolled, seems to have drifred significantly, the Pitch signal doesn't stay on the StripTool scale anymore). So we may have to double check that the transmitted beam isn't falling off the GTRX DC PD. |
14131
|
Fri Aug 3 18:54:58 2018 |
gautam | Update | SUS | Glitchy MC1 | The wall StripTool indicated that the IMC wasn't too happy when I came in today. Specifically:
- MC1 watchdog was tripped.
- Even in the tripped state, MC REFL spot on the camera showed spot motion that was too large to be explained as normal seismic driven motion (i.e. with local damping supposedly disabled).
- Strange excursions were observed in the MC1 shadow sensor signal levels as well, see Attachment #1 - negative values don't make any sense for this readout.
The last time this happened, it was due to the Sorensens not spitting out the correct voltages. This time, there were no indications on the Sorensens that anything was funky. So I just disabled the MCautolocker and figured I'd debug later in the evening.
However, around 5pm, the shadow sensor values looked nominal again, and when I re-enabled the local damping, the MC REFL spot suggested that the local damping was working just fine. I re-enabled the MCautolocker, MC re-locked almost immediately. To re-iterate, I did nothing to the electronics inside the VEA. Anyways, this enabled us to work on the X arm ASS (next elog). |
Attachment 1: MC1_sensorAnomaly.png
|
|
14130
|
Fri Aug 3 16:27:40 2018 |
rana | Update | SUS | Low noise bias path idea | Bah! Too complex. |
14129
|
Fri Aug 3 15:53:25 2018 |
gautam | Update | SUS | Low noise bias path idea | Summary:
The idea we are going with to push the coil driver noise contribution down is to simply increase the series resistance between the coil driver board output and the OSEM coil. But there are two paths, one for fast actuation and one that provides a DC current for global alignment. I think the simplest way to reduce the noise contribution of the latter, while preserving reasonable actuation range, is to implement a precision DC high-voltage source. A candidate that I pulled off an LT application note is shown in Attachment #1.
Requirements:
- The series resistance in the bias path should be
, such that the noise from this stage is dominated by the Johnson noise of said resistor, and hence, the current noise contribution is negligible compared to the series resistance in the fast actuation path ( ).
- Since we only really need this for the test masses, what actuation range do we want?
- Currently, ETMY has a series resistance of
and has a pitch DC bias voltage of -4 V.
- This corresponds to 10 mA of DC current.
- To drive this current through
, we need 100 V.
- I'm assuming we can manually correct for yaw misalignments such that 10mA of DC current will be sufficient for any sort of corrective alignment.
- So +/- 120 V DC should be sufficient.
- The current noise of this stage should be negligible at 100 Hz.
- The noise of the transistors and the HV supply should be suppressed by the feedback loop and so shouldn't be a significant contribution (I'll model to confirm).
- The input noise of the LT1055 is ~20nV/rtHz at 100 Hz, while the Johnson noise of
is ~13nV/rtHz so maybe the low-passing needs to be tuned, but I think if it comes to it, we can implement a passive RC network at the output to achieve additional filtering.
- To implement this circuit, we need +/- 125V DC.
- At EX and EY, we have a KEPCO HV supply meant to be used for the Green Steering PZTs.
- I'm not sure if these can do bipolar outputs, if not, for temporary testing, we can transport the unit at EY to EX.
If all this seems reasonable, I'd like to prototype this circuit and test it with ETMX, which already has the high series resistance for the fast path. So I will ask Steve to order the OpAmp and transistors. |
Attachment 1: LT1055_precOpAmp.pdf
|
|
14128
|
Fri Aug 3 14:35:56 2018 |
gautam | Summary | Electronics | EX AUX electronics power restored | Steve and I restored the power to the EX AUX electronics rack. The power strip on the lowest shelf of the AUX rack now goes to another power strip laid out vertically along the NW corner of 1X9. The EX green locks to the arm just fine now. |
14127
|
Thu Aug 2 23:09:25 2018 |
rana | Summary | Computers | X Green "Mystery" solved | I'm going to guess that this was me: I was disconnecting some octopus power strip nonsense down there (in particular illuminators and cameras), so I might have turned off the AUX rack by mistake.
Quote: |
I walked down to the X end and found that the entire AUX laser electronics rack isn't getting any power. There was no elog about this.
I couldn't find any free points in the power strip where I think all this stuff was plugged in so I'm going to hold off on resurrecting this until tomorrow when I'll work with Steve.
Quote: |
The X arm green does not stay locked to the cavity - the alignment looks fine, and the green flashes are strong, but the lock does not hold. This shouldn't be directly connected to anything we did today since the Green PDH servo is entirely analog.
|
|
|
14126
|
Thu Aug 2 20:54:18 2018 |
gautam | Summary | Computers | c1omc model looks stable | Actually, c1lsc had crashed again sometime last night so I had to reboot everything this morning. I used the reboot script again, but I increased the sleep time between trying to start up the models again so that I could walk into the VEA and power cycle the c1lsc expansion chassis, as this kind of frequent model crash has been fixed by doing so in the past. Sure enough, there have been no issues since I rebooted everything at ~1030 in the morning.
The c1omc model itself has been stable as well, though of course, there is nothing in there at the moment. I may do a check of the newly installed DAC tomorrow just to see that we can put out a sine wave.
Steve has ordered the D-sub cabling that will allow us to route signals between AA/AI boards in 1X1/1X2 to the HV PZT electronics in the OMC rack. Things look setup for a measurement next week. Aaron will post a block diagram + photoz of what box goes where in the electronics racks. |
14125
|
Thu Aug 2 20:47:29 2018 |
gautam | Summary | Electronics | X Green "Mystery" solved | I walked down to the X end and found that the entire AUX laser electronics rack isn't getting any power. There was no elog about this.
I couldn't find any free points in the power strip where I think all this stuff was plugged in so I'm going to hold off on resurrecting this until tomorrow when I'll work with Steve.
Quote: |
The X arm green does not stay locked to the cavity - the alignment looks fine, and the green flashes are strong, but the lock does not hold. This shouldn't be directly connected to anything we did today since the Green PDH servo is entirely analog.
|
|
14124
|
Thu Aug 2 16:30:08 2018 |
Steve | Update | Treasure | time capsule location | I 've just found this time capsule note from Nov. 26, 2000 by Kip Thorne: LIGO will discover gravitational waves by Dec.31, 2007
|
Attachment 1: time_capsule.JPG
|
|
14123
|
Wed Aug 1 20:44:57 2018 |
gautam | Summary | Computers | c1omc model (re?)created | The main motivation behind adding a DAC card in c1ioo was to setup an RTCDS model for the OMC. Attachment #1 shows the new look CDS overview screen. Here is what I did.
Mostly, I followed instructions from when I setup the model for the EX green PZTs.
Simulink model:
The model is just a toy for now (CDS parameters, ADC block and 2 CDS filter modules). I leave it to Aaron to actually populate it, check functionality etc. The path to the model is /opt/rtcds/caltech/c1/userapps/release/isc/c1/models/c1omc.mdl. I am listing the parameters set on the CDS_PARAMETERS block:
- host = c1ioo
- site = c1
- rate = 16k
- dcuid = 27 (which I chose after making sure that this dcuid was not used on this list which I also updated by adding c1omc and moving c1imc to "old")
- specific_cpu = 6 (again chosen after checking the available CPUs in the above list and confirming using the cset utility).
- adc_Slave = 1
- shmem_daq = 1
- no_rfm_dma = 1
- biquad = 1
Building and installing model:
Once the model was installed, I logged into c1ioo, and built and installed the models using the usual rtcds make and rtcds install instructions. Before starting the model, I edited /diskless/root.jessie/etc/rtsystab to allow c1omc to be run on c1ioo. Using sudo cset set, I verified that CPU #6 is no longer listed (if I understand correctly, the RTCDS system takes over the core).
MEDM:
To reflect all this on the MEDM CDS OVERVIEW screen, I just edited the screen.
- Moved the orange explanation of bits over to the c1iscey panel to make space in the c1ioo panel.
- Edited the macros to reflect the c1omc parameters.
DAQD:
Finally, I followed the instructions here to get the channels into frames and make all the indicators green. Went into fb and restarted the daqd processes. All looks good . I'm going to leave the model running overnight to investigate stability. I forgot to svn commit the model tonight, will do it tomorrow.
The testing plan (at least initially) is to install the AA and AI boards from the OMC rack in 1X1/1X2. Then we will have short SCSI cables running from the ADC/DAC to these. The actual HV driving stages will remain in the OMC rack (NE corner of AS table).
@Steve, can we get 10 Male-Female D9 cables so that we can run them from 1X1/1X2 to the OMC rack?
Unrelated to this work: There were 2 crashes of the models on c1lsc, one ~6pm and one right now ~1030pm. The restart script brought everything back gracefully ... |
Attachment 1: CDS_OVERVIEW_withOMC.png
|
|
14122
|
Wed Aug 1 19:41:15 2018 |
gautam | Summary | Computers | RTCDS recovery, c1ioo changes | [Gautam Koji]
After this work, we recovered the nominal RTCDS state. The main points were:
- We needed to restart the bind9 service on chiara such that the FEs knew their IP addresses upon reboot and hence, could get their root filesystems over NFS.
- We recovered suspension local damping, IMC locking and POX/POY locking with nominal arm transmission.
Some stuff that is not working as usual:
- The EX QPD is reporting strange transmission values - even with the PRM completely misaligned, it reports transmission of ~30. But we were able to lock the Xarm with the Thorlabs PD and revover transmission of ~1.15.
- The X arm green does not stay locked to the cavity - the alignment looks fine, and the green flashes are strong, but the lock does not hold. This shouldn't be directly connected to anything we did today since the Green PDH servo is entirely analog.
I made a model change in c1x03 (the IOP model on c1ioo) to add a DAC part. The model compiled, installed and started correctly, and looking at dmesg on c1ioo, it recognises the DAC card as what it is. Next step is to use a core on c1ioo for a c1omc model, and actually try driving some signals.
Note that the only change made to the c1ioo expansion chassis was that a DAC card was installed into the PCIe bus. The adaptor card which allows interfacing the DAC card to an AI board was already in the expansion chassis, presumably from whenever the DAC was removed from this machine.
*I think I forgot to restart optimus after this work...
|
Attachment 1: CDS_overview.png
|
|
14121
|
Wed Aug 1 16:23:48 2018 |
Koji | Summary | Computers | Transition of the main NFS disk on chiara | [Gautam Koji]
Taking the opportunity to shutdown c1ioo for adding a DAC card, we shutdown chiara and worked on moving of the main disk to the bigger home.
We shutdown most of the martian machines including the control machines, megatron, optimus, and nodus.
- Before shutting down chiara, we ran rsync to make the 4TB disk (used to be teh backup) and /cvs/cds synced.
sudo rsync -a --progress /home/cds/ /media/40mBackup
- Modified /etc/fstab
proc /proc proc nodev,noexec,nosuid 0 0
# / was on /dev/sda1 during installation
UUID=972db769-4020-4b74-b943-9b868c26043a / ext4 errors=remount-ro 0 1
# swap was on /dev/sda5 during installation
UUID=a3f5d977-72d7-47c9-a059-38633d16413e none swap sw 0 0
UUID="90a5c98a-22fb-4685-9c17-77ed07a5e000" /media/40mBackup ext4 defaults,relatime,commit=60 0 0
#fb:/frames /frames nfs ro,bg
UUID=92dc7073-bf4d-4c58-8052-63129ff5755b /home/cds ext4 defaults,relatime,commit=60 0 0
- Shutdown chiara. Put the 4TB disk in the chassis. We also installed a new disk (but later it turned out that it only has 2TB...)
- Restart the mahcine. This already made the 4TB disk mounted as /cvs/cds .
- Restart bind9 with DHCP for the diskless clients (cf. https://wiki-40m.ligo.caltech.edu/CDS/How_to_join_martian)
sudo service bind9 restart
sudo service isc-dhcp-server restart
- Looks like /etc/resolv.conf is automatically overwritten by a tool or something everytime we restart the machine!? I still don't know how to avoid this. (cf. https://www.ctrl.blog/entry/resolvconf-tutorial). But at least for today we manually wrote /etc/resolv.conf
controls@chiara|backup> cat /etc/resolv.conf
# Dynamic resolv.conf(5) file for glibc resolver(3) generated by resolvconf (8)
# DO NOT EDIT THIS FILE BY HAND -- YOUR CHANGES WILL BE OVERWRITTEN
nameserver 192.168.113.104
nameserver 131.215.125.1
nameserver 8.8.8.8
search martian
|
14120
|
Tue Jul 31 22:50:18 2018 |
aaron | Update | OMC | OMC Expected Refl Signal | I learned a lot about lasers this week from Siegman. Here are some plots that show the expected reflectivity off of the OMC for various mode matching cases.
The main equation to know is 11.29 in Siegman, the total reflection coefficient going into the cavity:

Where r is the mirror reflectivity (assumed all mirrors have the same reflectivity), t is the transmissivity, and g is the complex round-trip gain, eq 11.18

The second exponential is the loss; in Siegman the \alpha_0 is some absorption coecfficient and p is the total round trip length, so the product is just the total loss in a round trip, which I take to be 4*the loss on a single optic (50ppm each). \phi is the total round trip phase accumulation, which is 2\pi*detuning(Hz)/FSR. The parameters for the cavity can be found on the wiki.
I've added the ipynb to my personal git, but I can put it elsewhere if there is somewhere more appropriate. I think this is all OK, but let me know if something is not quite right. |
Attachment 1: omcRefl.pdf
|
|
14119
|
Tue Jul 31 08:17:55 2018 |
Steve | Update | SUS | Trillium interface box was fixed,reinstalled & working |
|
Attachment 1: all_OK.png
|
|
14118
|
Mon Jul 30 18:19:03 2018 |
Koji | Update | SUS | Trillium interface box was fixed and reinstalled | The trillium interface box was removed from the rack.
The problem was the incorrect use of an under-spec TVS (Transient Voltage Suppression) diodes (~ semiconductor fuse) for the protection circuit.
The TVS diodes we had had the breakdown voltages lower than the supplied voltages of +/-20V. This over-voltage eventually caused the catastrophic breakdown of one of the diodes.
I don't find any particular reason to have these diodes during the laboratory use of the interface. Therefore, I've removed the TVS diodes and left them unreplaced. The circuit was tested on the bench and returned to the rack. All the cables are hooked up, and now the BRLMs look as usual.
Details
- The board version was found to be D1000749-v2
- There was an obvious sign of burning or thermal history around the components D17 and D14. The solder of the D17 was so brittle that just a finger touch was enough to remove the component.
- These D components are TVS diodes (Transient Voltage Suppression Diodes) manufactured by Littelfuse Inc. It is sort of a surge/overvoltage protector to protect rest of the circuit to be exposed to excess voltage. The specified component for D17/D14 was 5.0SMMDJ20A with reverse standoff voltage (~operating voltage) of 20V and the breakdown voltage of 22.20V(min)~24.50V(max). However, the spec sheet told that the marking of the proper component must be "5BEW" rather than "DEM," which is visible on the component. Some search revealed that the used component was SMDJ15A, which has the breakdown voltage of 16.70V~18.50V. This spec is way too low compared to the supplied voltage of +/-20V. |
Attachment 1: P_20180730_173134.jpg
|
|
Attachment 2: P_20180730_180151.jpg
|
|
14117
|
Mon Jul 30 16:11:54 2018 |
gautam | Update | SUS | Trillium interface box is broken | [koji, steve, gautam]
We debugged this in the following way:
- Disconnect all fuses in the terminal blocks coming from the +/- 20 VDC Sorensens.
- Check that they are indeed isolated using DMM.
- Test blocks of fuses in order to identify where the problem is happening (i.e. plug fuses in, turn up Sorensen voltage knobs, look for current overload). We did things in the following order:
- MC suspensions
- BS, PRM and SRM
- ITMY
- ITMX
- Trillium interface box.
- Turns out that the Trillium box is the culprit.
- Confirmed that the problem is in the trillium interface box and not in the seismometer itself by unplugging all cables leading out of the interface box, and checking that the problem persists when the box is powered on.
So for now, the power cable to the box is disconnected on the back end. We have to pull it out and debug it at some point.
Apart from this, megatron was un-sshable so I had to hard reboot it, and restart the MCautolocker, FSSslowPy and nds2 processes on it. I also restarted the modbusIOC processes for the PSL channels on c1auxex (for which the physical Acromag units sit in 1X5 and hence were affected by our work), mainly so that the FSS_RMTEMP channel worked again. Now, IMC autolocker is working fine, arms are locked (we can recover TRX and TRY~1.0), and everything seems to be back to a nominal state. Phew. |
14115
|
Mon Jul 30 11:05:44 2018 |
gautam | Update | SUS | IFO SUS wonky | When I came in this morning:
- PMC was unlocked.
- Seis BLRMS were off scale.
- ITMX OSEM LEDs were dark on the CRT monitor even though Sat Box was plugged in.
Checking status of slow machines, it looked like c1sus, c1aux, and c1iscaux needed reboots, which I did. Still PMC would not lock. So I did a burtrestore, and then PMC was locked. But there seemed to be waaaaay to much motion of MCREFL, so I checked the suspension. The shadow sensor EPICS channels are reporting ~10,000 cts, while they used to be ~1000cts. No unusual red flags on the CDS side. Everything looked nominal when I briefly came in at 6:30pm PT yesterday, not sure if anything was done with the IFO last night.
Pending further investigation, I'm leaving all watchdogs shutdown and the PSL shutter closed.
A quick look at the Sorensens in 1X6 revealed that the +/- 20V DC power supplies were current overloaded (see Attachment #1). So I set those two units to zero until we figure out what's going on. Possibly something is shorted inside the ITMX satellite box and a fuse is blown somewhere. I'll look into it more once Steve is back. |
Attachment 1: IMG_7102.JPG
|
|
14114
|
Sun Jul 29 23:15:34 2018 |
pooja | Update | Cameras | Developing CNN | Aim: To develop a convolutional neural network that resolves mirror motion from video.
Input : Previous simulated video of beam spot motion in pitch by applying 4 sine waves of frquencies 0.2, 0.4, 0.1, 0.3 Hz and amplitude ratios to frame size to be 0.1, 0.04, 0.05, 0.08 where random uniform noise ranging 0.05 has been added to amplitudes and frequencies. This is divided into train (0.4), validation (0.1) and test (0.5).
Model topology:
- Number of filters = 2
- Kernel size = 2
- Size of pooling windows = 2
- -----> Dense layer of 4 nodes ----> Output layer of 1 node
Activation: selu linear
Batch size = 32, Number of epochs = 128, loss function = mean squared error
Optimizer: Nadam ( learning rate = 0.00001, beta_1 = 0.8, beta_2 = 0.85)
Plots of CNN output & applied signal given in Attachment 1. The variation in loss value with epochs given in Attachment 2.
This needs to be further analysed with increasing random uniform noise over the pixels and by training CNN on simulated data of varying ampltides and frequencies for sine waves. |
Attachment 1: conv_nn_varying_freq_amp_1.pdf
|
|
Attachment 2: conv_nn_varying_freq_amp_2.pdf
|
|
14113
|
Sun Jul 29 20:03:02 2018 |
rana | Update | PEM | Seismometer temp control | While Shruti is re-building Kira's heater circuit, I looked up how to do one of these (i.e. what does a real EE say about how to build a current source?):
It turns out that there is an Analog Devices application note (AN-968) about this (as there usually is once we get tired of playing around and try to look up the right answer).
I've linked to the note and attached the recommended schematic for high current applications. We'll go ahead as is, but we'll make a PCB according to this App Note for the v3 circuit.
|
Attachment 1: Screen_Shot_2018-07-29_at_8.00.27_PM.png
|
|
14112
|
Sun Jul 29 00:59:54 2018 |
Koji | Update | Electronics | Characterization of Transimpedance Amplifier | You have this measurement problem when the IF bandwidth is larger than the measurement frequency. I suspect the IF bandwidth is 30kHz. |
14111
|
Sat Jul 28 22:16:49 2018 |
John Martyn | Update | | Characterization of Transimpedance Amplifier | Kevin and I meaured the transfer function of the photodiode circuit using the Jenne laser and agilent in the 40m lab. The attached figures depict our measured transfer function over the modulation frequency ranges of 30kHz-30MHz and 1kHz-30MHz when the power of the laser was set to 69 and 95 μW. These plots indicate a clear roll off frequency around 300 kHz. In addition, the plots beginning at 1kHz display unstable behavior at frequencies below 30kHz. I am not sure why there is such a sharp change in the transfer function around 30kHz, but I suspect this to be due to an issue with the agilent or photodiode. |
Attachment 1: PD_TF1.pdf
|
|
Attachment 2: PD_TF2.pdf
|
|
Attachment 3: PD_and_TIA_Transfer_Function_Measurements.zip
|
|