40m QIL Cryo_Lab CTN SUS_Lab CAML OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m elog, Page 36 of 357  Not logged in ELOG logo
ID Date Author Type Category Subjectup
  13437   Tue Nov 21 11:37:29 2017 gautamUpdateOptical LeversBS OL calibration updated

I calibrated the BS oplev PIT and YAW error signals as follows:

  1. Locked X-arm, ran dither alignment servos to maximize transmission.
  2. Applied an offset to the ASC PIT/YAW filter banks. Set the ramp time to something long, I used 60 seconds.
  3. Monitored the X arm transmission while the offset was being ramped, and also the oplev error signal with its current calibration factor.
  4. Fit the data, oplev error signal vs arm transmission, with a gaussian, and extracted the scaling factor (i.e. the number which the current Oplev error signals have to be multiplied by for the error signal to correspond to urad of angular misalignment as per the overlap of the beam axis to the cavity axis.
  5. Fits are shown in Attachment #1 and #2.
  6. I haven't done any error analysis yet, but the open loop OL spectra for the BS now line up better with the other optics, see Attachment #3 (although their calibration factors may need to be updated as well...). Need to double check against OSEM readout during the sweep.
  7. New numbers have been SDF-ed.

The numbers are:

BS Pitch     15  /  130    (old/new)     urad/counts

BS Yaw       14  /  170    (old/new)     urad/counts

Quote:

I bet the calibration is out of date; probably we replaced the OL laser for the BS and didn't fix the cal numbers. You can use the fringe contrast of the simple Michelson to calibrate the OLs for the ITMs and BS.

 

Attachment 1: OL_calib_BS_PERROR.pdf
OL_calib_BS_PERROR.pdf
Attachment 2: OL_calib_BS_YERROR.pdf
OL_calib_BS_YERROR.pdf
Attachment 3: VertexOLnoise_updated.pdf
VertexOLnoise_updated.pdf
  13439   Tue Nov 21 16:28:23 2017 gautamUpdateOptical LeversBS OL calibration updated

The numbers I have from the fitting don't agree very well with the OSEM readouts. Attachment #1 shows the Oplev pitch and yaw channels, and also the OSEM ones, while I swept the ASC_PIT offset. The output matrix is the "naive" one of (+1,+1,-1,-1). SUSPIT_IN1 reports ~30urad of motion, while SUSYAW_IN1 reports ~10urad of motion.

From the fits, the BS calibration factors were ~x8 for pitch and x12 for yaw - so according to the Oplev channels, the applied sweep was ~80urad in pitch, and ~7urad in yaw.

Seems like either (i) neither the Oplev channels nor the OSEMs are well diagonalized and that their calibration is off by a factor of ~3 or (ii) there is some significant imbalance in the actuator gains of the BS coils...

Quote:
 

Need to double check against OSEM readout during the sweep.

 

Attachment 1: BS_oplev_sweep.png
BS_oplev_sweep.png
  5235   Mon Aug 15 10:02:27 2011 kiwamuUpdateSUSBS OSEM adjustment done

[Suresh / Kiwamu]

 Adjustment of the OSEMs on BS has been done.

All the bad suspensions (#5176) has been adjusted. They are waiting for the matrix inversion test.

  14541   Mon Apr 15 10:20:44 2019 gautamUpdateOptical LeversBS Oplev PIT was oscillating

The AS spot on the camera was oscillating at ~3 Hz. Looking at the Oplevs, the culprit was the BS PIT DoF. Started about 12 hours ago, not sure what triggered it. I disabled Oplev damping, and waited for the angular motion to settle down a bit, and then re-enabled the servo - damps fine now...

Attachment 1: BS_OL_oscillating.png
BS_OL_oscillating.png
  16912   Tue Jun 14 08:41:36 2022 JCUpdateOPLEV TablesBS Oplev Table Sketch

[JC]

Lately, I have been working on a 3d sketch of the BS OPLEV Table on SolidWorks. This is my progress so far, a few of the components I will have to sketch myself, such as the HeNe laser and photodiodes. This will just be a general layout of the HeNe laser, optics, and photodiodes.

Attachment 1: BS_OPLEV_Table.PNG
BS_OPLEV_Table.PNG
  17936   Mon Oct 30 18:04:46 2023 PacoUpdateSUSBS Oplev check

I measured the Oplev PIT/YAW OLTFs and adjusted the gains to get 0.1 to ~ 4 Hz feedback as is recommended in Gautam's thesis (Fig 3.3 and discussion thereafter). Attachments #1-2 show the measured OLTF and simple expected model using the foton filters and plant. Attachment#3 shows the final filter modules and gains after the changes were done.


Next steps:

  • Repeat for PRM, ITMs and ETMs
Attachment 1: BSOL_OLTF_measuredScreenshot_2023-10-30_18-11-12.png
BSOL_OLTF_measuredScreenshot_2023-10-30_18-11-12.png
Attachment 2: IDEAL_BSOL_OLTFScreenshot_2023-10-30_18-10-19.png
IDEAL_BSOL_OLTFScreenshot_2023-10-30_18-10-19.png
Attachment 3: FINALBSOL_screenScreenshot_2023-10-30_18-10-52.png
FINALBSOL_screenScreenshot_2023-10-30_18-10-52.png
  15749   Wed Jan 6 16:18:38 2021 gautamUpdateOptical LeversBS Oplev glitchy

As part of the hunt why the X arm IR transmission RIN is anomalously high, I noticed that the BS Oplev Servo periodically kicks the optic around - the summary pages are the best illustration of this happening. Looking back in time, these seem to have started ~Nov 23 2020. The HeNe power output has been degrading, see Attachment #1, but this is not yet at the point where the head usually needs replacing. The RIN spectrum doesn't look anomalous to me, see Attachment #2 (the whitening situation for the quadrants is different for the BS and the TMs, which explains the HF noise). I also measured the loop UGFs (using swept sine) - seems funky, I can't get the same coherence now (live traces) between 10-30 Hz that I could before (reference traces) with the same drive amplitude, and the TF that I do measure has a weird flattening out at higher frequencies that I can't explain, see Attachment #3.

The excess RIN is almost exactly in the band that we expect our Oplevs to stabilize the angular motion of the optics in, so maybe needs more investigation - I will upload the loop suppression of the error point later. So far, I don't see any clean evidence of the BS Oplev HeNe being the culprit, so I'm a bit hesitant to just swap out the head...

Attachment 1: missingData.png
missingData.png
Attachment 2: OLRIN.pdf
OLRIN.pdf
Attachment 3: BS_OL_P.pdf
BS_OL_P.pdf
Attachment 4: BS_OL_suppression.pdf
BS_OL_suppression.pdf
  6369   Wed Mar 7 04:08:48 2012 kiwamuUpdateSUSBS SIDE gain was too low

The BS SIDE damping gain seemed too low. The gain had been 5 while the rest of the suspensions had gains of 90-500.

I increased the gain and set it to be 80.

 

I did the "Q of 5" test by kicking the BS SIDE motion to find the right gain value.

However there was a big cross coupling, which was most likely a coupling from the SIDE actuator to the POS motion.

Due to the cross coupling, the Q of 5 test didn't really show a nice ring down time series. I just put a gain of 80 to let the Q value sort of 5.

I think we should diagonalize the out matrices for all the suspensions at some point.

  15510   Sat Aug 8 07:36:52 2020 Sanika KhadkikarConfigurationCalibration-RepairBS Seismometer - Multi-channel calibration

Summary : 

I have been working on analyzing the seismic data obtained from the 3 seismometers present in the lab. I noticed while looking at the combined time series and the gain plots of the 3 seismometers that there is some error in the calibration of the BS seismometer. The EX and the EY seismometers seem to be well-calibrated as opposed to the BS seismometer.

The calibration factors have been determined to be :

BS-X Channel: \small {\color{Blue} 2.030 \pm 0.079 }

BS-Y Channel: \small {\color{Blue} 2.840 \pm 0.177 }

BS-Z Channel: \small {\color{Blue} 1.397 \pm 0.182 }


Details :

The seismometers each have 3 channels i.e X, Y, and Z for measuring the displacements in all the 3 directions. The X channels of the three seismometers should more or less be coherent in the absence of any seismic excitation with the gain amongst all the similar channels being 1. So is the case with the Y and Z channels. After analyzing multiple datasets, it was observed that the values of all the three channels of the BS seismometer differed very significantly from their corresponding channels in the EX and the EY seismometers and they were not calibrated in the region that they were found to be coherent as well. 


Method :

Note: All the frequency domain plots that have been calculated are for a sampling rate of 32 Hz. The plots were found to be extremely coherent in a certain frequency range i.e ~0.1 Hz to 2 Hz so this frequency range is used to understand the relative calibration errors. The spread around the function is because of the error caused by coherence values differing from unity and the averages performed for the Welch function. 9 averages have been performed for the following analysis keeping in mind the needed frequency resolution(~0.01Hz) and the accuracy of the power calculated at every frequency. 

  1. I first analyzed the regions in which the similar channels were found to be coherent to have a proper gain analysis. The EY seismometer was found to be the most stable one so it has been used as a reference. I saw the coherence between similar channels of the 2 seismometers and the bode plots together. A transfer function estimator was used to analyze the relative calibration in between all 3 pairs of seismometers. In the given frequency range EX and EY have a gain of 1 so their relative calibration is proper. The relative calibration in between the BS and the EY seismometers is not proper as the resultant gain is not 1. The attached plots show the discrepancies clearly : 
  • BS-X & EY-X Transfer Function : Attachment #1
  • BS-Y & EY-Y Transfer Function : Attachment #2

          The gain in the given frequency range is ~3. The phase plotting also shows a 180-degree phase as opposed to 0 so a negative sign would also be required in the calibration factor. Thus the calibration factor for the Y channel of the BS seismometer should be around ~3. 

  • BS-Z & EY-Z Transfer Function : Attachment #3

The mean value of the gain in the given frequency range is the desired calibration factor and the error would be the mean of the error for the gain dataset chosen which is caused due to factors mentioned above.

Note: The standard error envelope plotted in the attached graphs is calculated as follows :

         1. Divide the data into n segments according to the resolution wanted for the Welch averaging to be performed later. 

         2. Calculate PSD for every segment (no averaging).

         3. Calculate the standard error for every value in the data segment by looking at distribution formed by the n number values we obtain by taking that respective value from every segment.

Discussions :

The BS seismometer is a different model than the EX and the EY seismometers which might be a major cause as to why we need special calibration for the BS seismometer while EX and EY are fine. The sign flip in the BS-Y seismometer may cause a lot of errors in future data acquisitions. The time series plots in Attachment #4 shows an evident DC offset present in the data. All of the information mentioned above indicates that there is some electrical or mechanical defect present in the seismometer and may require a reset. Kindly let me know if and when the seismometer is reset so that I can calibrate it again. 

Attachment 1: BS_X-EY_X.png
BS_X-EY_X.png
Attachment 2: BS_Y-EY_Y.png
BS_Y-EY_Y.png
Attachment 3: BS_Z-EY_Z.png
BS_Z-EY_Z.png
Attachment 4: timeseries.png
timeseries.png
  5648   Tue Oct 11 03:35:16 2011 kiwamuUpdateLSCBS actuator reponse at low frequency : measured

The response of the BS actuator in a low frequency regime has been measured.

After the measurement I did a coarse fit to see if the low frequency data agree with the high frequency response which I have measured two weeks ago (#5583)
So far it shows a good agreement with the high frequency data (see the plot below). Tomorrow I will do a serious fitting.
Once the calibration of BS is done, the low frequency responses of ITMs, PRM and SRM will be done by simply exciting BS and comparing them (maybe at a couple of frequency points around 0.1Hz).
BSactuator.png

(Measurement)

 + With free swinging MICH, the sensor (AS55_Q) was calibrated into counts/m.

     => The peak-peak counts was about 110 counts. So the sensor response is about 6.5x108 counts/m

 + Locked Michelson with AS55_Q and the signal was fedback to BS.

 + Set the UGF high enough so that the open loop gain below 10 Hz is greater than 1.

 + With DDT's swept sine measurement, C1:LSC-MICH_EXC was excited with a big amplitude of 40 counts.

 + Took a transfer function from C1:LSC-MICH_OUT to C1:LSC-MICH_EXC.

 + Calibrated the transfer function into m/counts by dividing it with the sensor response.

Quote from #5641
One possible reason is that my actuator responses are not so accurate below 1Hz.
I will measure the DC response of all the actuators and it will completely determine the shapes of the actuator responses except for the region around the resonance.

  5649   Tue Oct 11 15:14:50 2011 ranaUpdateLSCBS actuator reponse at low frequency : measured

Quote:

The response of the BS actuator in a low frequency regime has been measured.

 

This seems like an error prone method for DC responses due to the loop gain uncertainty. Better may be to use the fringe hopping method (c.f. Luca Matone) or the fringe counting method

  5656   Wed Oct 12 17:53:01 2011 kiwamuUpdateLSCBS actuator response : fitting done and histroy of delays

An update on calibration of the BS actuator : A fitting has been done.

(Fitting)

 I used LISO for fitting the complex transfer function.
Because the data points around 1 Hz didn't have big coherence a few data points, which had coherence of less than 0.9, were excluded.
Also the fitting of the Q-factor wasn't successful due to the lack of good data points around the resonance.So I left Q fixed to be 5 in the fitting.
 

(Fitting result)

G =  2.18060874008e-8 +/- 6.425e-10 (2.95%)
f0 =  1.0100491195 +/- 1.51e-2 (1.49%)  [Hz]
Q = 5 (fixed)
delay =  423.2753462089e-6 +/- 4.989e-6 (1.18%)  [sec]

bode_BSactuator.png

 

(History of delay)
 Because we have been observing several different amount of delays in different configurations, perhaps it is worth to summarize those numbers.
     description  delay [usec]   elog entry
       MICH lock (BS actuation)  423 this entry
       LSC feed forward path  127 #5218
       MICH lock (BS actuation)  600 #4638
      ALS on X arm (ETMX actuation)  330 #4196
      RFM (from c1lsc to c1sus)  125 #4153
      from ADC to DAC (all the front end machine)  38-110 #3961
      from ADC to DAC (c1sus)  124 #3838
     RFM (c1ioo and c1sus)   8-62 #3855

Quote from #5648
Tomorrow I will do a serious fitting.

  7350   Thu Sep 6 16:46:44 2012 JenneUpdateSUSBS aligned, target removed

Q and I aligned the BS such that we were hitting the center of ETMX. The ETMX cage does not have OSEM setscrew holes on the front, so it is not possible to put the targets that Steve made on this optic.  So, I put the freestanding ruler in front of the optic, with the edge of the ruler at the center (as viewed from above) of the optic.  Then Eric steered the BS until we were hitting the 5.5" mark, and roughly half of the beam was obscured by the ruler.

We then aligned ITMX such that the prompt reflection was colinear with the incoming beam. 

I checked the 2 spots through the BS, heading to the AS port.  (2 spots since MICH hasn't been locked / finely aligned yet).  They were being clipped on the 2nd output PZT.  I adjusted the knobs of the first output PZT to center the spots on the 2nd PZT.  Note that the output PZTs' power is still off, and has been off for some unknown length of time.  I had found them off when prepping for the vent a week or two ago.  So the current alignment depends on them staying off.  We don't really need them on until we're ready to employ our OMC.

The beams now look nicely unclipped on the AS camera, and we're aligning MICH.

  806   Wed Aug 6 22:19:07 2008 YoichiUpdateSUSBS alignment
Koji, Yoichi

We realized that we did not pay attention to the BS alignment while working on the alignment of the ITMX today. Because we were injecting the ALM laser (absolute length measurement laser) from the AS port, the ITMX alignment depends on the BS alignment.
The BS optical lever was not centered and the sum was about 2000cnt, which is low compared, for example, to the SRM oplev.
So we were not sure if the BS was in a good alignment or not.
So we decided to move the BS to center the QPD.
In doing so, we also moved the ITMX so that we do not lose the ALM laser beam coming back to the AS port.
When the BS oplev was centered, the sum of the QPD was still about 2000. So it was not far off centered.
After the tweaking, we were able to see some interference between the light reflected by the ITMY and ITMX at the AS port (actually this is the bright port for the ALM laser). By tweaking the ITMY, we were able to see Michelson fringes at the AS port.
If we believe the ALM laser alignment is still good after the vent, the ITMX, ITMY, BS and SRM should be now in a good alignment condition.
The OSEM values for the ITMX, BS, SRM seem to be ok (0.9+/-0.2). The ITMY LL is a bit low (~ 0.45).
  7392   Fri Sep 14 21:03:02 2012 DenUpdatePEMBS and AS tables

I've measured seismic and acoustic noise on BS and AS tables. It seems that horizontal motion of BS table is ~1.5-2 times more then AS table in the frequency range 5-50 Hz.

Edit by Den: this was POI table, not BS!

vert.png  horiz_4_50.png

acoustic.png      coh.png

  7395   Sat Sep 15 20:27:54 2012 DenUpdatePEMBS and AS tables

Quote:

I've measured seismic and acoustic noise on BS and AS tables. It seems that horizontal motion of BS table is ~1.5-2 times more then AS table in the frequency range 5-50 Hz.

Edit by Den: this was POI table, not BS!

 This seismic measurement is for BS and AS tables.

bs_as_h.jpg       bs_as_v.jpg

  7396   Mon Sep 17 07:57:58 2012 ranaUpdatePEMBS and AS tables for REFL change

 

 This was in response to my suggestion to move the REFL beam path to the table containing the BS/PRM Oplevs. From this seismic data it is clear that the BS table is no worse than the AS table, so we should plan to make the layout change during the next vent.

  6903   Mon Jul 2 18:27:25 2012 yutaUpdateGeneralBS and ITMX chambers closed

[Koji, Steve, Jamie, Jenne, Yuta]

We opened BS and ITMX chambers, took lots of photos, and closed them with heavy doors.
I turned off high voltage power supplies for PZTs and blocked PSL beam. We are ready for the pumping tomorrow.

Important photos we took:
  - positions of green optics at BS chamber, which was moved on the vent on Aug 2011
  - positions of PZT mirrors and cable connectors at BS chamber, which will be replaced with tip-tilts on the next vent
  - arrow on PR2 pointing HR (it was correct)
  - tried to take photos of clipping IR beam at BS OSEM holder from ITMX chamber
 
 We also took bunch of other photos.


Beam dump needed at BS chamber:
  We also checked some un-dumped beams at BS chamber. We need dumps;
  - behind MMT1, for unwanted transmitted beam
  - behind IPPOSSM3, for unwanted transmitted beam (IPPOSSM3 is the last mirror in BS chamber for IPPOS)

  7329   Fri Aug 31 17:26:54 2012 SteveUpdateGeneralBS camera

Steve and Eric

Placed pick up mirror for BS face and PRM back. I will ask Jamie to clamp it.

There will be an other camera set up to view the face of PRM

Attachment 1: BS_PRM.png
BS_PRM.png
  7772   Sat Dec 1 00:24:37 2012 Den, AyakaUpdateAlignmentBS chamber

Today at 11:13 AM the stack of invacuum BS table was kicked and IFO misaligned. We adjusted PZT2 voltage by ~20 V in yaw such that IPPOS was restored. Then we could lock arms.

push.png

  7773   Sat Dec 1 13:56:38 2012 ranaUpdateAlignmentBS chamber

 

 Whoever was working around the BS chamber at 11 AM on Friday should admit it now and take the punishment.

For those of you who like to do work on the interferometer without reporting it in the elog because you think that what you did doesn't affect anything, this is your example of how our time can be wasted by such laziness.

  7784   Tue Dec 4 18:53:50 2012 AyakaUpdateAlignmentBS chamber

BS chamber seemed to be kicked again around 10:00 am today.

I moved PZT mainly in YAW and locked both arms. I adjusted the beam to be almost on the center of both ETM by sights.

BSchamb1204.pdf

  3105   Wed Jun 23 12:52:35 2010 kiwamuUpdatePhotosBS chamber before cleaning up

  2800   Tue Apr 13 20:02:02 2010 KojiUpdateSUSBS chamber opened, PRM/SRM SOS removed from the table

Bob, Steve, and Koji

We opened North heavy door of the BS chamber in the afternoon.

In the evening, Koji worked on the PRM/SRM removal.

- Cleaned up the OPLEV mirrors to create some spaces near the door.

- Clamped PRM/SRM.

- Removed OSEMs. Made a record of the OSEMs. The record is on the wiki (http://lhocds.ligo-wa.caltech.edu:8000/40m/Upgrade_09/Suspensions)

- Found the SOSs are quite easy to remove from the table as they are shorter than the MOSs.

- Put a new Al sheet on a wagon. Put the SOSs on it. Wrapped them by the Al foils.

- Carried it to the clean room. They are on the right flow bench. Confirmed the wires are still fine.

- Closed up the chamber putting a light door.

Attachment 1: IMG_2384.jpg
IMG_2384.jpg
  3332   Fri Jul 30 12:46:38 2010 KojiConfiguration40m UpgradingBS chamber status
Attachment 1: BS_layout_Jul_29_2010.png
BS_layout_Jul_29_2010.png
  16574   Tue Jan 11 14:21:53 2022 PacoUpdateElectronicsBS feedthroughs and in-vac cables installed

[Paco, Yehonathan, Chub]

The BS chamber 10" flange with 4 DSUB-25 feedthroughs has been installed with the cables connected at the in-vac side. This is the second of two flanges, and includes 4 cables ordered vertically in stacks of 2 & 2 for [[LO2-1, LO2-2, PR3-1, PR3-2]] respectively.

  8257   Fri Mar 8 12:57:57 2013 AnnalisaUpdateABSLBS installed on ITMY table

 Sendhil and I installed the S polarized BS on the ITMY table to steer the NPRO beam through the AR wedge and align it to the POY beam. 

We took a shutter from the BSPRM table (which was not used) and a beam dump from the AS table (which was used by the auxiliary laser already removed and installed on the ITMY).

To do: do better alignment of the NPRO beam, maybe installing some iris after the BS and before the AS wedge, phase lock the two beams. 

  8258   Fri Mar 8 13:42:35 2013 JenneUpdateABSLBS installed on ITMY table

Re:  POY beam reduction.

We are able to lock the Yarm with the beam / gain as it is.  I had thought we might need to increase the DC gain in the whitening board by a factor of 2, but so far it's fine.

  10852   Mon Jan 5 12:42:09 2015 ericqUpdateSUSBS misbehaving

The BS was showing some excess motion. I think I've fixed it. Order of operations:

  • The DC PIT bias from previous ASS runs was at ~500, I zeroed this and aligned the BS to be centered on its oplev QPD with DC alignment sliders
  • I squished the gold box cables. This changed the alignment slightly, and brought the UR voltage back to a normal value. Excess motion still existed
  • I found that the the C1:SUS-BS_LRSEN filter had HOLD OUTPUT enabled. I turned it off. All seems well. 

I'm not sure how this might have gotten switched on...

  7714   Thu Nov 15 02:18:24 2012 DenUpdateModern ControlBS oplev

I've applied LQR feedback technique to BS oplev in pitch. I think the most inconvenient thing in using LQR controller is the amount of additional states created during cost function shaping. It requires 1 filter bank for each state. To avoid this I wrote state estimation code so all states are calculated inside one function.

On the plots below cost function and oplev feedback controller performance are shown.

lqr_cost.png    olpit_hg.png

  11199   Fri Apr 3 14:57:38 2015 manasaUpdateSUSBS oplev

The BS oplev has been misbehaving and kicking the optic from time to time since noon. The kicks are not strong enough to trip the watchdogs (current watchdog max counts for the sensors is 135).

I took a look at the spectrum of the BS oplev error in pit and yaw with both loops enabled while the optic was stable. There is nothing alarmingly big except for some additional noise above 4Hz.

I have turned the BS oplev servo OFF for now.

Attachment 1: BS_oplev_Apr3.png
BS_oplev_Apr3.png
  11200   Fri Apr 3 15:15:55 2015 SteveUpdateSUSBS oplev

I saw this kicking before  

Quote:

The BS oplev has been misbehaving and kicking the optic from time to time since noon. The kicks are not strong enough to trip the watchdogs (current watchdog max counts for the sensors is 135).

I took a look at the spectrum of the BS oplev error in pit and yaw with both loops enabled while the optic was stable. There is nothing alarmingly big except for some additional noise above 4Hz.

I have turned the BS oplev servo OFF for now.

 

  11201   Fri Apr 3 19:35:14 2015 JenneUpdateSUSBS oplev centered

I think that this happens when the beam gets too close to the edge of the QPD.  We see this regularly in the ETMs, if they've been kicked a bit, but not enough to trip the watchdogs.  I think it might be the step/impulse response of the RES3.3 filter, which rings for almost 20 seconds. 

Anyhow, I've just recentered the BS oplev.  It was at -21urad in pitch, and had more than 400 counts on the top two quadrants, but only about 100 counts on the bottom two.  Now it's around 300 counts on all 4 quadrants.

As a totally unrelated aside, I have installed texlive on Donatella, so that I could run pdflatex.

  4976   Fri Jul 15 16:14:00 2011 steveUpdateSUSBS oplev error signal spectra
Attachment 1: BS_opl_ersig.pdf
BS_opl_ersig.pdf
  11100   Thu Mar 5 10:32:58 2015 SteveUpdateSUSBS oplev servo turned off

The BS oplev servo was kicking up the BS. It was turned off

  4947   Wed Jul 6 16:44:37 2011 steve, kiwamuUpdateSUSBS oplev spectra

Healthy BS oplev

Attachment 1: BS.jpg
BS.jpg
  4963   Tue Jul 12 17:30:24 2011 steve,UpdateSUSBS oplev spectra

Quote:

Healthy BS oplev

 I repeated the BS oplev spectrum today and I do not understand why it does look different. I did it as Kiwamu describes it in entry#4948  The oplev servo was left ON!

Attachment 1: BS_oplev.jpg
BS_oplev.jpg
  4966   Thu Jul 14 09:38:50 2011 steve,UpdateSUSBS oplev spectra

Quote:

Quote:

Healthy BS oplev

 I repeated the BS oplev spectrum today and I do not understand why it does look different. I did it as Kiwamu describes it in entry#4948  The oplev servo was left ON!

 It is working today! Finally I repeated the BS spectra, that we did with Kiwamu last week

Attachment 1: BS_oplev.jpg
BS_oplev.jpg
  5622   Wed Oct 5 17:08:49 2011 steveUpdateSUSBS oplev spectra

Kiwamu and Steve,

The He/Ne oplev shows no coherece so relative intensity noise is not limiting factor for the oplev servo

Attachment 1: BSoplservON2.png
BSoplservON2.png
  14108   Fri Jul 27 10:48:57 2018 SteveUpdateSUSBS oplev window

Yesterday I inspected this BS oplev viewport. The heavy connector tube was shorting to table so It was moved back towards the chamber. The connection is air tight with kapton tape temporarly.

 The beam paths are well centered. The viewport is dusty on the inside.

The motivation was to improve the oplev noise.

Attachment 1: BSOw_.jpg
BSOw_.jpg
Attachment 2: dustInsideBSO.jpg
dustInsideBSO.jpg
  831   Wed Aug 13 17:00:59 2008 steveConfigurationSUSBS sat amp removed
The PRM sat amp is broken. Ben is working on it.
The BS sat amp was removed from the BS sus and it is used with the PRM in
order to damp it for wire stand-off alignment.
  226   Mon Jan 7 09:01:39 2008 steveUpdateSUSBS sus damping restored
The BS sus damping was lost at 8am Sunday morning.
Attachment 1: bssdl.jpg
bssdl.jpg
  8119   Wed Feb 20 19:48:16 2013 yutaUpdateAlignmentBS table oplev re-arranged

[Sendhil, Yuta]

After aligning IFO and putting the access connector on, we also centered IPANG/IPPOS and all oplevs (except SRM).
To avoid clipping of PRM/BS oplevs, we re-arranged oplev steering mirrors on BS table.

What we did:
  1. Checked IPANG comes out unclipped after putting on the access connector.
  2. Centered IPANG on its QPD.
  3. Checked oplevs beams for ITMX/ITMY centered on in-vac mirrors, and centered them on their QPDs.
  4. Checked IPPOS beam is centered on the mirrors inside BS chamber, and centered IPPOS on its QPD.
  5. Tweaked oplev mirrors on BS chamber to make PRM/BS oplev beam unclipped and centered on mirrors, and centered them on their QPDs. To avoid clipping of oplev beams in BS table, we re-arranged oplev steering mirrors on BS table (outside the vaccum).


Current status:
  QPD values, IFO_ALIGN/MC_ALIGN screens, OSEM values attached.

  - IR incident beam and IFO aligned
  - X/Y end green coming out to PSL table (in higher order modes)
  - IPANG/IPPOS available
  - All oplevs available
  - AS/REFL/POP cameras ready
  - access connector, ETMX/ETMY heavy doors on
  - ITMX/ITMX/BS heavy doors are not on
  - AS/REFL/POP PDs not centered
  - POX/POY/TRX/TRY not aligned
  - AS beam coming out of the OMC chamber low by ~ 1 beam diameter (my bad)


Tomorrow:
  - Align AS/REFL/POP PD and lock PRMI
  - Take pictures of ITMX/ITMY/BS stacks
  - Put heavy doors on ITMX/ITMY/BS chambers
  - Start pumping down

Attachment 1: IFOALIGN_QPDs_OSEMs.png
IFOALIGN_QPDs_OSEMs.png
  8120   Wed Feb 20 19:58:59 2013 ranaUpdateAlignmentBS table oplev re-arranged

Please confirm the SRM OL beam is not too bad and also find where the mis-aligned SRM puts its beam. WE want to be sure that there is not too much unwanted scattering from SRM into the PRFPMI.

  8915   Wed Jul 24 10:35:41 2013 SteveUpdateVACBS, ITMY doors are removed

Quote:

We will open the BS and ITMY doors first thing tomorrow morning. I plan to try to be in around 9 am. The first order of business will be to flip the folding mirrors that are not currently flipped (SR2, SR3, PR3).

 Jenne, Annalisa & Steve

Attachment 1: beforeDoorsOff.png
beforeDoorsOff.png
Attachment 2: particlecount10d.png
particlecount10d.png
  8914   Tue Jul 23 22:55:13 2013 JenneUpdateVACBS, ITMY doors to be opened in the morning

We will open the BS and ITMY doors first thing tomorrow morning. I plan to try to be in around 9 am. The first order of business will be to flip the folding mirrors that are not currently flipped (SR2, SR3, PR3).

  8393   Tue Apr 2 18:19:30 2013 JenneUpdateSUSBS, PRM oplev servos improved

 

[Gabriele, Jenne]

We have implemented 4Hz resonant gains for both PRM and BS yaw.  The filter was already in place for PRM Yaw, so we just turned it on, but we also copied the filter over to BS Yaw.  We also changed the 3.3Hz res gain and the ELP for the PRM servo to match the BS servo, since after implementing the 4Hz gain, PRM was still much noisier than BS.  Now the 2 servos match, and PRM is a little quieter.  We hope that tonight's locking might be a little more stable after this work.

PRM_servo_matches_BS.pdf

prm_bs_optical_levers_comparison.pdf

  14345   Tue Dec 11 18:20:59 2018 gautamUpdateOptical LeversBS/PRM HeNe is dead

I found that the BS/PRM OL SUM channels were reading close to 0. So I went to the optical table, and found that there was no beam from the HeNe. I tried power-cycling the controller, there was no effect. From the trend data, it looks like there was a slow decay over ~400000 seconds (~ 5 days) and then an abrupt shutoff. This is not ideal, because we would have liked to use the Oplevs as a DC alignment reference during the ventnoI plan to use the AS camera to recover some sort of good Michelson alignment, and then if we want to, we can switch out the HeNe.

*How can I export PDF from NDscope?

Attachment 1: BSOL_dead.png
BSOL_dead.png
  14404   Fri Jan 18 12:52:07 2019 gautamUpdateOptical LeversBS/PRM Oplev HeNe replaced

I replaced the BS/PRM Oplev HeNe with one of the heads from the SP table where Steve was setting up the OL RIN/pointing noise experiment. The old one was dead. The new one outputs 3.2 mW of power, I've labelled it with this number, serial number and date of replacement. The beam comes out of the vacuum chamber for both the BS and PRM, and the RIN spectra (Attachment #1) look alright. The calibration into urad and loop gains possibly have to be tweaked. Since the beam comes out of vacuum, I say that we shouldn't open the BS/PRM chamber for this vent - we don't have a proper plan for the in-air layout yet, so we can add this to the list of to-dos for the next vent.

I think we are down to our last spare HeNe head in the lab - @Chub, please look into ordering some more, the ITMX HeNe is going to need replacement soon.

Attachment 1: OLRIN_20190118.pdf
OLRIN_20190118.pdf
  17757   Sat Aug 5 01:46:01 2023 KojiUpdateOptical LeversBS/PRM/SRM Oplev dead

[Koij Hiroki]

While KA was working on the DAC issue, BS/PRM/SRM Oplev died.

It seems that the BS/PRM/SRM HeNe died and was replaced in 2019 (4yrs + 200 days ago) and 2021 Jan (2yrs + 209 days ago).

We have no energy to work on the HeNe replacement tonight. This needs to be done on Monday.

Tag: OPLEV oplev HeNe died dead

Attachment 1: Screenshot_2023-08-05_08-52-18.png
Screenshot_2023-08-05_08-52-18.png
Attachment 2: Screenshot_2023-08-05_08-56-26.png
Screenshot_2023-08-05_08-56-26.png
ELOG V3.1.3-