ID |
Date |
Author |
Type |
Category |
Subject |
13377
|
Thu Oct 12 07:56:33 2017 |
Steve | HowTo | Cameras | ETMX GigE side view at 50 deg of IR scattering | Telescope front lens to wall distance 25 cm, GigE camera lenght 6 cm and cat6 cable 2cm
Atm3, Existing short camera can has 16cm lenght to lexan guard on viewport. Available 2" od periscope tube lenght is 8cm. The one in use 16 cm long.
Note: we can fabricate a lite cover with tube that would accomodate longer telescope.
Can we calibrate the AR coated M5018-SW and compare it's performance agains the 2" periscope
Look at the Edmond Optics 3" od camera lens with AR
This lower priced 1" apeture Navitar lens can be an option too.
Atm1, Now I can see dust. This is much better. The focus is not right yet.
Atm2, Chamber viewport wiped and image refocused. Actually I was focusing on the dust.
Quote: |
I calculated a better lens solution for the ETMX side view with the simple python script that's attached. The camera is still not as close to the viewport as we would like, and now the front lens is almost all the up to the end of the tube. With a little more playing around there maybe a better way, especially if we expand the repertoire of focal lengths. Using Steve's wonderful camera fixture I put the beam spot in focus. I turned the camera sideways for better use of the field of view, and now the beam spot actually fills the center area of the beam, to the point where we probably don't want more magnification or else we start losing the tails of the Gaussian.
We'll take a serious of images tomorrow, and will have an estimate of the scatter loss by the end of tomorrow.
|
|
Attachment 1: Image__2017-10-11__15-29-52_15k400g.png
|
|
Attachment 2: Image__2017-10-12__15-50-18wipedRefocud2.png
|
|
Attachment 3: camCan16cm.jpg
|
|
13376
|
Thu Oct 12 01:50:11 2017 |
gautam | Update | LSC | AS55Q Dark Noise | I worked on the daughter board a little more in the evening. I have somehow managed to make the dark noise ~25% worse [Attachment #1].
- Earlier in the day, I had switched out both on-board AD797s for OP27. The latter has ~3x the input voltage noise, and LISO modeling suggests that this is the dominant contribution to the output voltage noise.
- There are some differences in the actual components with which the board is stuffed, and the schematic.
- After updating the LISO model, I expect to get an output voltage noise of ~50nV/rtHz. But I measured ~2x this value (measured with LO input of demod board driven, RFPD input terminated).
- While I had the board out, I replaced most of the installed thick-film resistors with thin film ones. For good measure, I also changed the AD829s.
After making all these changes, I re-installed the card in the eurocrate and repeated the measurement. The Q channel noise was close to the expected value (~50nV/rtHz), but the I channel is twice as noisy. I will continue this investigation tomorrow. |
Attachment 1: AS55_dark.png
|
|
13375
|
Thu Oct 12 01:03:49 2017 |
johannes | HowTo | Cameras | ETMX GigE side view | I calculated a better lens solution for the ETMX side view with the simple python script that's attached. The camera is still not as close to the viewport as we would like, and now the front lens is almost all the up to the end of the tube. With a little more playing around there maybe a better way, especially if we expand the repertoire of focal lengths. Using Steve's wonderful camera fixture I put the beam spot in focus. I turned the camera sideways for better use of the field of view, and now the beam spot actually fills the center area of the beam, to the point where we probably don't want more magnification or else we start losing the tails of the Gaussian.
We'll take a serious of images tomorrow, and will have an estimate of the scatter loss by the end of tomorrow.
|
Attachment 1: IMG_20171011_164549698.jpg
|
|
Attachment 2: Image__2017-10-11__16-52-01.png
|
|
Attachment 3: GigE_lens_position_helper.py.zip
|
13374
|
Wed Oct 11 19:31:32 2017 |
gautam | Update | LSC | AS55Q Dark Noise | I tried replacing the AD797s on the daughter board with OP27s, and saw no significant improvement in the electronics noise of the demod board. Note that according to LISO, in this configuration, the voltage noise of the Op27 is expected to dominate the total noise of the daughter board. Measurement condition was that the RFPD input was terminated, but the LO input was still being driven (SR785 input range is -50dBVpk for all traces, and the input ranging was set to "UpOnly"). Need to do a more systematic investigation to figure out where this excess noise is coming from. I will upload photos of the board later.
Quote: |
This supports the hypothesis that something is wonky on the daughter board, because the purple trace should only be the quad sum of the orange and green traces. I will pull it out and have a look.
|
|
Attachment 1: AS55Q_darkNoises.pdf
|
|
13373
|
Wed Oct 11 17:59:45 2017 |
rana | Update | Electronics | SR560 noise level | these are not the SR785 settings that you're looking for
Quote: |
Gautam and Steve,
All 3 show the same noise level ~80 nV / rt Hz at 1 kHz as shown. Batteries ordered to be replaced in the top 2
We'll do more measurement to see how can we get to 4 nV / rt Hz specification level.
|
To get low noise measurements on the SR785, you have to have the input range set to -50 dB, not +20 dB. Its not within the powers of commercial electronics ADCs to give you a 10 nV noise floor with +10 V input signals. The SR560 has an input referred noise of 5 nV/rHz, so the output noise should be 5e-9 x 500 = 2.5 uV/rHz. Your picture shows it giving 1 uV RMS, so you also need to use the PSD units. |
13372
|
Wed Oct 11 14:42:03 2017 |
gautam | Update | LSC | AS55Q Dark Noise | I keep adding traces to this plot, here is the most complete one I have now. Looks like the input noise to the D040179 (measured at "Q out" SMA jack of D990511 with RFPD input terminated) is ~10nV/rtHz. This supports the hypothesis that something is wonky on the daughter board, because the purple trace should only be the quad sum of the orange and green traces. I will pull it out and have a look.
Some other follow-ups on the questions raised at the meeting:
- Doesn't look like I've implemented thin film resistors on the input of the coil driver boards. De-whitening boards have the critical signal path resistors (judged as the ones with largest contribution as per LISO model) changed to thin film. Pictures are here.
- I think I didn't make a full elog of my demod board efficiency investigations, but from my notes and Attachment #4 of elog 12972, I calculated the gain in the signal path as the ratio of Vpp_out / Vpp_in.
Quote: |
I measured the output voltage noise of the Q output of the AS55 Demod Board with the PSL shutter closed, using the SR785 (see Attachment #1). The measured noise is consistent with the expected number of ~120nV/rtHz around 100Hz. I had measured the gain of this board from RFPD input to Q output to be ~5.1: so if the PD dark noise is 16nV/rtHz, this would be amplified to ~80nV/rtHz. Still a discrepancy of ~50%. I didn't measure the noise with the PD input terminated. Added the noise of the demod board output with the RFPD input terminated. The level of ~100nV/rtHz seems consistent with the actual PD dark noise being ~80nV/rtHz, as their quadrature sum is around 130nV/rtHz. Need to dig up the schematics for the demod board + daughter board, and check against LISO, to see if this is consistent with what is expected.
Also - I think I was using the wrong value of the DC power on the AS55 photodiode for shot noise calculations - 13mW was for REFL55, not AS55. I did a crude measurement of the power by sticking the Ophir power meter (filter removed) in front of the AS55 PD with the Michelson flashing around, and noticed the maximum value registered was ~1.2mW. So in the DRMI lock, there would be ~2.4mW, which is 10x lower than the value I was assuming. I've made the correction in the NB code, for the next time the plot is generated. A more rigorous measurement would involve sticking the Ophir in front of the AS110 PD during a DRMI lock. The light from the AS port is split by a 50-50 BS to the AS55 and AS110 PDs (so measuring at AS110 is a reasonable proxy for power at AS55), and the AS110 signals are not used for triggering in the DRMI lock, so this is feasible.
|
|
Attachment 1: AS55Q_darkNoises.pdf
|
|
13371
|
Wed Oct 11 10:29:43 2017 |
Steve | Update | Electronics | SR560 noise level | Gautam and Steve,
All 3 show the same noise level ~80 nV / rt Hz at 1 kHz as shown. Batteries ordered to be replaced in the top 2
We'll do more measurement to see how can we get to 4 nV / rt Hz specification level. |
Attachment 1: sr560.jpg
|
|
Attachment 2: sr560noise.jpg
|
|
13370
|
Tue Oct 10 22:04:06 2017 |
rana | Update | LSC | AS55Q Dark Noise | how about calibrate the DC channels so that you can just get the acutal power levels from the trend data? |
13369
|
Mon Oct 9 22:18:34 2017 |
gautam | Update | LSC | AS55Q Dark Noise | I measured the output voltage noise of the Q output of the AS55 Demod Board with the PSL shutter closed, using the SR785 (see Attachment #1). The measured noise is consistent with the expected number of ~120nV/rtHz around 100Hz. I had measured the gain of this board from RFPD input to Q output to be ~5.1: so if the PD dark noise is 16nV/rtHz, this would be amplified to ~80nV/rtHz. Still a discrepancy of ~50%. I didn't measure the noise with the PD input terminated. Added the noise of the demod board output with the RFPD input terminated. The level of ~100nV/rtHz seems consistent with the actual PD dark noise being ~80nV/rtHz, as their quadrature sum is around 130nV/rtHz. Need to dig up the schematics for the demod board + daughter board, and check against LISO, to see if this is consistent with what is expected.
Also - I think I was using the wrong value of the DC power on the AS55 photodiode for shot noise calculations - 13mW was for REFL55, not AS55. I did a crude measurement of the power by sticking the Ophir power meter (filter removed) in front of the AS55 PD with the Michelson flashing around, and noticed the maximum value registered was ~1.2mW. So in the DRMI lock, there would be ~2.4mW, which is 10x lower than the value I was assuming. I've made the correction in the NB code, for the next time the plot is generated. A more rigorous measurement would involve sticking the Ophir in front of the AS110 PD during a DRMI lock. The light from the AS port is split by a 50-50 BS to the AS55 and AS110 PDs (so measuring at AS110 is a reasonable proxy for power at AS55), and the AS110 signals are not used for triggering in the DRMI lock, so this is feasible.
|
Attachment 1: AS55Q_dark.pdf
|
|
13368
|
Mon Oct 9 11:55:01 2017 |
Koji | Update | LSC | DRMI Nosie Budget v3.0 | My last characterization of the AS55 PD was on Feb 2013. ELOG 8100
There I said the dark noise at the PD output was 16nV/rtHz. I don't have the measurement of the Voltage noise at the output of the demod board.
Note that the PD can only be limited by shot noise when the DC current is larger then 4mA. |
13367
|
Mon Oct 9 01:29:26 2017 |
gautam | Update | LSC | DRMI Nosie Budget v3.0 | Summary:
I spent this weekend doing a more careful investigation of the DRMI noise. I think I have some new information/insights. Attachment #1 is the noise budget (png attached because pdf takes forever to upload, probably some ImageMagick problem. The last attachment is a tarball of the PDF). Long elog, so here are the Highlights:
- Coil de-whitening does result in small improvement in noise in the 60-200Hz band.
- Above 200Hz, we seem to be limited by "Dark" noise. More on this below.
- The coupling from SRCL->MICH is the other limiting noise in the 60-200Hz band now.
Sensing Matrix Measurement:
- I rotated the AS55 demod phase from -42 degrees to -82 degrees, the idea being to get more of the MICH error signal in AS55_Q.
- Consequently, the MICH servo gain has been lowered from -0.035 to -0.021. Settings have been updated in the snap file used by the locking script.
- Seems to have worked.
- Attachment #2 is the measured sensing elements.
- One major source of uncertainty in these sensing element numbers is the actuator gains for PRM, SRM and BS. The coil driver electronics for the latter two have been modified recently, and for them, I am using numbers from this elog scaled by the expected factor as a result of removing the x3 gain in the de-whitening boards for SRM and BS.
MICH OLTF
- Measurement was done in lock using the usual IN1/IN2 method.
- Model made by loading the FOTON filters + assumed models for the BS pendulum and AA/AI filters in Matlab, and fitting to an overall gain + delay.
- Attachment #3 shows the agreement between measurement and model.
- The model was exported and used to invert in-loop signals to their out-of-loop counterparts in the noise budget.
DAC Noise
- I had claimed that turning on the coil de-whitening did not improve the MICH noise.
- This was not exactly true - I had only compared MICH noise with the BS de-whitening turned ON/OFF, while the ITM de-whitening was always on.
- Turns out that there is in fact a small improvement - see Attachment #4 (DTT crashes everytime I try to print a pdf, so png screenshot will have to do for now).
- I have also changed the way in which DAC noise is plotted in the Noise Budget code:
- I used to directly convert the measured voltage noise (multiplied by appropriate scalar to account for quadrature sum of 4 coils each in 3 optics) to displacement noise using the sensing measurement cts/m values.
- Now I convert the measured voltage noise first to current noise (knowing the series resistance), then to force noise (using the number 0.016 N/A per coil), then to displacement noise (assuming a mirror mas of 250g).
- Quadrature sum is again taken for 4 coils on 3 optics.
- I've also added the option to plot the DAC noise with the de-whitening filter TF applied (taking care that the maximum of filtered DAC noise / coil driver electronics noise is used at each frequency).
- So the major source of uncertainty in the calculated DAC noise is the assumed actuator gain of 0.016 N/A.
The DAC noise is not limiting us anywhere when the coil de-whitening is switched on.
Dark Noise
I think this is the major find.
- The dark noise spectrum is measured with:
- the PSL shutter closed
- the AS55 I and Q analog whitening filters (and corresponding digital de-whitening filters) engaged, to mimic the operating conditions under which the in-lock error signal is acquired.
- Comparing the blue and black traces, it is clear that turning on the analog whitening is having some effect on the dark noise.
- However, the analog whitening filters should suppress the ADC noise by ~30dB @ 100Hz - so assuming 1uV/rtHz, this would be ~30nV/rtHz @100Hz.
- But the measured noise seems to be ~5x higher, with 4*10^-4 cts/rtHz translating to roughly 120nV/rtHz.
- The photodiode dark noise is only 15nV/rtHz according to the wiki. Where is this measured?
So I don't understand the measured Dark Noise level, and it is limiting us at frequencies > 200Hz. Some busted electronics in the input signal chain? Or can the LSC demod daughter board gain of ~5 explain the observed noise?
Shot noise
- The DC power on AS55 photodiode was measured to be ~13mW with the SRM misaligned.
- This corresponds to ~100cts peak amplitude on the ASDC channel (derived from AS55 photodiode).
- In the DRMI lock, the ASDC level is ~200cts.
- I used these numbers, and equation 2.17 in Tobin's thesis, to calculate this curve.
Edit 1730 9 Oct: I had missed out the factor of 5 gain in the demod board in calculating the shot noise curve. Attachment #7 shows the corrected shot noise level. Explicitly:
, where is to convert shot noise in W to displacement units.
AUX coupling
This is the other find.
- While chatting with Gabriele, he suggested measuring the SRCL->MICH and PRCL->MICH cross couplings.
- I injected a signal in SRCL servo EXC channel, and adjusted amplitude till coherence in MICH_IN1 was good.
- The actual TF measured was MICH_IN1 / SRCL_IN1 (so units of cts/ct).
- My multiplying the in-lock PRCL and SRCL IN1 signals by these coupling coefficients (assumed flat in frequency for now, note that measurement was only made between 100Hz and 1kHz), I get the trace labelled "AUX coupling" in Attachment #1 (this is the quadrature sum for SRCL and PRCL couplings).
- Also repeated for PRCL -> MICH coupling in the same way.
- Measurements of these TFs and coherence are shown in Attachment #5 (again png screenshot because of DTT).
- However, there is no significant coherence in MICH/SRCL or MICH/PRCL in this frequency range.
This seems to be limiting us from saturating the dark noise once the coil de-whitening is engaged. But lack of coherence means the mechanism is not re-injection of SRCL/PRCL sensing noise? Need to think about what this means / how we can mitigate it.
OL A2L coupling
- I didn't measure these
- These couplings would have changed because I modified the Oplev loop shapes to allow engaging of coil de-whitening filters.
- But anyways, their effect will only be below 100Hz because I made the roll-offs steeper.
Still to measure (but not likely to be limiting us anywhere in the current state):
- Laser intensity noise -> MICH coupling (using AOM).
- Laser frequency noise -> MICH coupling (using CM board IN2).
- Oscillator noise (amplitude + phase) -> MICH coupling (using AM/FM input of Marconi).
I've also made several changes to the NB code - will push to git once I finish cleaning stuff up, but it is now much faster to make these plots and see what's what. |
Attachment 1: DRMI_NB.png
|
|
Attachment 2: DRMI1f_Oct8.pdf
|
|
Attachment 3: MICH_OLTF_model.pdf
|
|
Attachment 4: MICH_noises.png
|
|
Attachment 5: AUX_couplings.png
|
|
Attachment 6: C1NB_disp_40m_MICH_NB_2017-10-08.pdf.tar.gz
|
Attachment 7: MICH_NB_corrected.png
|
|
13366
|
Fri Oct 6 17:08:09 2017 |
Steve | Update | ALS | X End table beam traps corrected | There are no more double sided tape on this table.
|
Attachment 1: c1.jpg
|
|
Attachment 2: c2.jpg
|
|
Attachment 3: c3.jpg
|
|
Attachment 4: c4.jpg
|
|
13365
|
Fri Oct 6 12:56:40 2017 |
gautam | Summary | LSC | RTCDS NN | [gabriele, gautam]
Gabriele had prepared a C code implementation of his NN for MICH/PRCL state estimation. He had been trying to get it going on some of the machines in WB, but was unsuccessful. The version of RCG he was trying to compile and run the code on was rather dated so we decided to give it a whirl on our new RCG3.4 here at the 40m. Just noting down stuff we tried here:
- Code has been installed at /opt/rtcds/userapps/release/cds/c1/src/nn.This is to facilitate compilation by the RCG.
- There is also a simulink block diagram (x3tst.mdl) in there which we copied and pasted into c1pem. Changed the appropriate paths in the C-Code block to point to the location in the previous bullet point.
- c1pem was chosen for this test as it runs at 2k which is what the network is designed for.
- Since we were running the test on c1sus and expected the machine to crash, I shutdown all the watchdogs for the test.
- Code compiled without any problems (i.e. rtcds make c1pem and rtcds install c1pem executed successfully). There were some warning messages related to C-Code blocks but these are not new, they show up in all models in which we have custom C-code blocks.
- Unfortunately, the whole c1sus FE crashed when we tried rtcds restart c1pem.
- We tried a couple of more iterations, and attempted to monitor dmesg during the restart process to see if there were any clues as to why this wasn't working, but got nothing useful.
All models have been reverted to their state prior to this test, and everything on the CDS_OVERVIEW MEDM screen is green now. |
Attachment 1: post_NN_test.png
|
|
13364
|
Fri Oct 6 12:46:17 2017 |
gautam | Update | CDS | 40m files backup situation | Looks to have worked this time around.
controls@fb1:~ 0$ sudo dd if=/dev/sda of=/dev/sdc bs=64K conv=noerror,sync
33554416+0 records in
33554416+0 records out
2199022206976 bytes (2.2 TB) copied, 55910.3 s, 39.3 MB/s
You have new mail in /var/mail/controls
I was able to mount all the partitions on the cloned disk. Will now try booting from this disk on the spare machine I am testing in the office area now. That'd be a "real" test of if this backup is useful in the event of a disk failure.
Quote: |
The 4TB HGST drives have arrived. I've started the FB1 dd backup process. Should take a day or so.
|
|
13363
|
Fri Oct 6 00:25:45 2017 |
rana | Update | LSC | FS725 for Marconi stabilization | Steve, can you please connect this fan to the rack power and remove this extra power supply?
Quote: |
Re-arranged the DC bench supply on the shelf in the PSL enclosure, whose only purpose seems to be to supply 12V to a fan attached to the rear of the PSL NPRO controller. Seems to be a waste of space! The fan was momentarily disconnected but has since been reconnected and is spinning again.
|
|
13362
|
Thu Oct 5 18:40:27 2017 |
gautam | Update | LSC | FS725 for Marconi stabilization | [steve, gautam]
- We installed the FS725 on the shelf inside the PSL enclosure - see Attachment #1.
- We ran a long BNC cable (labelled "GPS 1pps" on both ends) from 1X7 to the PSL enclosure - this was to pipe the 1PPS signal from the GPS timing unit (EndRun Technologies Tempus LX) rear panel (50 ohm output according to the datasheet) to the 1PPS input of the FS725 (high impedance). See Attachments #2. Note that the 1pps output was already tee'd on the rear panel. One port of the tee was unused (this now goes to the FS725) while the other was going to the 1PPS input of the Master Timing Sequencer (D050239), so I decided that there was no need to tee the 1pps input of the FS725 with a 50ohm terminator. In a few minutes, the Rb standard indicated that it was locked to its internal reference, and also to the external 1pps input (see Attachment #1).
- We ran a long BNC cable (labelled "Rb 10MHz" on both ends) from the 10MHz output of the FS725 (50 ohm output impedance), in the PSL enclosure to the rear BNC "FREQ_STD IN/OUT" BNC connector of the Marconi (1kohm input impedance). Changed the frequency reference setting on the Marconi to "External Direct". The FS725 datasheet recommends terminating the load with a 50ohm inline terminator, I have not yet done this (see Attachment #3). Is it appropriate to use a Balun (FTB-1-1) here? This would avoid ground loops between the Marconi and the FS725, and also make the load seen by the FS725 50ohms.
- Found that there was an unused long cable from the PSL enclosure to the 1X2 electronics rack. We re-purposed this to drive the AOM driver via the DAC output in 1Y2. The cable is labelled "AOM driver" on both ends. This was to facilitate measurement of the coupling of laser intensity noise to AS55_Q in a DRMI lock.
- Removed 2 long cables between 1X7 and 1X2 that weren't connected to anything.
- Re-arranged the DC bench supply on the shelf in the PSL enclosure, whose only purpose seems to be to supply 12V to a fan attached to the rear of the PSL NPRO controller. Seems to be a waste of space! The fan was momentarily disconnected but has since been reconnected and is spinning again.
- Removed a couple of unused power cables from the mess on the shelf in the PSL enclosure. Also removed an unused Sony Video Squential Switcher YS-S6 from the PSL enclosure.
Quote: |
I've located the Stanford Research FS725 Rb reference unit. The question is where to put it. This afternoon Steve and I put it inside the little electronics rack next to 1X3, but in hindsight, this probably isn't such a great place for a timing reference as there are a bunch of Sorensen power supplies in there (and presumably the accompanying harmonics from these switching supplies).
The unit itself was repaired in 2015, and powering it on, it locked to the internal reference within a few minutes as prescribed in the manual.
|
|
Attachment 1: IMG_7617.JPG
|
|
Attachment 2: IMG_7619.JPG
|
|
Attachment 3: IMG_7618.JPG
|
|
13361
|
Thu Oct 5 13:58:26 2017 |
gautam | Update | CDS | 40m files backup situation | The 4TB HGST drives have arrived. I've started the FB1 dd backup process. Should take a day or so.
Quote: |
Edit: unmounting /frames won't help, since dd makes a bit for bit copy of the drive being cloned. So we need a drive with size that is >= that of the drive we are trying to clone. On FB1, this is /dev/sda, which has a size of 2TB. The HGST drive we got has an advertised size of 2TB, but looks like actually only 1.8TB is available. So I think we need to order a 4TB drive. |
|
13360
|
Thu Oct 5 11:46:15 2017 |
gautam | Update | CDS | slow machine bootfest | MC Autolocker was umnhappy because c1iool0 was unresponsive and hence it couldn't write to the "C1:IOO-MC_AUTOLOCK_BEAT" channel. I keyed the crate and IMC locked almost immediately. I'm moving this channel into the RTCDS model as we did for the IFO_STATE EPICS channel so that the autolocker isn't dependant on c1iool0 (which was the whole point of migrating the IFO-STATE variable anyways). I also commented out all of these channels in /cvs/cds/caltech/target/c1iool0/autolocker.db so that there aren't duplicate channels.
Quote: |
Eurocrate key turning reboots for c1susaux, c1auxex,c1auxey, c1iscaux, c1iscaux2 and c1aux. Usual precautions were taken for ITMX. Did burtrestore for c1iscaux andc1iscaux2 in order to restore the LSC PD whitening gains.
Un-related to this work: input pointing into PMC was tweaked as the PMC_REFL spot was pretty bright.
|
|
13359
|
Thu Oct 5 02:14:51 2017 |
gautam | Update | LSC | More DRMI coupling measurements - setup | In the end I decided to access the available spare DAC channels via the C1ASS model - for this purpose, I added a namespace block "TEST" in the C1ASS simulink model, which is a SISO block. Inside is just a single CDS filter module. My idea is to use the EXC of this filter module to inject excitations for measuring various couplings. Rather than have a simple testpoint, we also have the option of adding in some filter shapes in the filter module which could possibly allow a more direct read-off of some coupling TF. Recompiling the model went smooth - there was a crash earlier in the day which required me to hard-reboot c1lsc (and also restart all models on c1sus and c1ioo but no reboots necessary for those machines).
Note that to get the newly added channels to show up in the channel lists in DTT/AWGGUI etc, you need to ssh into fb1 and restart the daqd processes via sudo systemctl restart daqd_*. If I remember right, it used to be enough to do telnet fb 8088 followed by shutdown. This is no longer sufficient.
It took me a while to get the DRMI locking going again. The model restarts earlier in the evening had changed a bunch of EPICS channel settings (and out config scripts don't catch all of these settings). In particular, I forgot to re-enable the x3 digital gain for the ITMs, BS and SRM (necessitated by removing an analog x3 gain on the de-whitening boards). I was hesitant to spend time re-adjusting all damping / oplev loop gains because if we change the series resistor on the coil driver board, we will have to do this again. I also didn't want this arbitrary FM to be enabled in the SDF safe.snap. But maybe it's worth doing it anyways - if nothing it'll be good practise.
Once I hunted down all the setting diffs and tweaked alignment, the DRMI locks were pretty robust.
I had hoped to make some of these TF measurements tonight. But I realized I needed to look up a bunch of stuff in manuals/datasheets, and think about these measurements a little. I wasn't sure if the DW/AI board could drive a signal over 40m of BNC cabling so I added an SR560 (DC coupled, gain=1, low noise mode, 50ohm output used) to buffer the output. The Marconi's external modulation input is high impedance (100k) but for the AOM driver we want 50ohm. For the Marconi, the external input accepts 1Vrms max, while for the AOM driver, we want to drive a signal between 0V and 1V at most.
The general measurement setup is schematically shown in Fig 1. Questions to address:
- What happens if we apply a negative voltage to the input of the AOM driver? What is the damage threshold? Do we have to worry about SR560 offset level?
- Is there a way to dynamically adjust the offset in DTT such that we can have different amplitude signals at different frequencies (usually done by specifying an envelope in DTT) but still satisfy the requirement that the entire signal lie between 0-1V?
- For the Laser Intensity noise -> MICH coupling TF measurement, I guess we can use the AOM to inject an excitation, and measure the ratio of the response in MC_TRANS and in MICH_IN1. Then we multiply the in-loop MC_TRANS spectrum by the magnitude of this TF to get the Laser Intensity Noise contribution to MICH.
- The Laser Frequency Noise coupling should be negligible in MICH - but the measurement principle should be the same. Drive the AO input of the Mode Cleaner Servo board from the DAC, look at ratio of response in MICH_IN1 and MC_F. Multiply the DRMI in-lock MC_F spectrum by this TF.
- The oscillator noise seems more tricky to me (also Finesse modeling suggests this may be the most significant of the 3 couplings described in this elog, though I may just be computing the coupling in Finesse wrongly)
- I don't understand all the External Modulation options specified in the manual.
- DC? AC? FM? PM? AM? Need to figure out what is the right settings to use.
- I'm not sure how independent the various modulations will be - i.e. if I select PM, how much AM is induced as a result of me driving the EXT MOD input?
- What is the right level of excitation drive? I tried this a bunch of times tonight - set the PM range to 0.1rad (for the full scale 1Vrms sine wave input), but with an excitation of just a few counts, already saw non-lineaer coupling in MICH_IN1 which probably means I'm driving this too hard.
- This measurement needs a bit more algebra. We have an estimate of the Marconi phase noise from Rana (is this the right one to use?). But the "Transfer Function" we'd measure is cts in MICH_IN1 in response to counts to Marconi via the signal chain in Attachment #1. So we'd need to know (and divide out) the AI/DW board TF, and the Marconi's TF, which the datasheet suggests has a lower 3dB frequency of 100Hz (assuming SR560 and cable can be treated as flat).
- A simpler test may be to just hook up the Marconi to the Rb standard, and the Rb to 1pps from GPS, and look for a change in the MICH noise.
Am I missing something? |
Attachment 1: CB4709D0-3FA7-43E3-BC25-3CF4164E6C6A.jpeg
|
|
13357
|
Wed Oct 4 17:38:25 2017 |
gautam | Update | LSC | FS725 for Marconi stabilization | I've located the Stanford Research FS725 Rb reference unit. The question is where to put it. This afternoon Steve and I put it inside the little electronics rack next to 1X3, but in hindsight, this probably isn't such a great place for a timing reference as there are a bunch of Sorensen power supplies in there (and presumably the accompanying harmonics from these switching supplies).
The unit itself was repaired in 2015, and powering it on, it locked to the internal reference within a few minutes as prescribed in the manual. |
13356
|
Wed Oct 4 17:18:15 2017 |
gautam | Update | CDS | Free DAC channels in c1lsc | There are at least 5 free DAC channels (4 if you discount the one channel from these that I am hijacking) available in the 1Y2 electronics rack.
Jamie's nice wiring diagram shows the topology - the actual DAC card sits in 1Y3 inside the c1lsc expansion chassis (while the c1lsc frontend itself is in 1X4). The output of the DAC goes via SCSI to an interface box (D080303) and then to some dewhitening/AI boards (D000316). There are a total of 16 DAC channels available, out of which 8 are used for the TTs, 2 are used for the DAFI model, and one is labeleld "From c1ioo 1X2" (I don't know what this one is for). So I'm going to use some of these channels for measuring the coupling of oscillator noise and intensity noise to MICH in the DRMI lock.
The de-whitening/AI board seems to be old - it has 2x 800Hz Butterworth LPFs and no notch for the clock frequency, but maybe this doesn't matter for the tests I have in mind. The AI board available on 1X2 is more modern but routing the DAC channels from 1Y2 to it is going to be some work.
I'm going to add my testpoint to c1daf given that it seems to be the least critical model on c1lsc.
EDIT: testpoints added to c1daf don't show up in the list of available channels - there was some issue with this model while we were getting the new RTCDS going. So I'm moving my temporary testpoint to c1cal instead. |
13355
|
Tue Oct 3 19:39:10 2017 |
gautam | Update | CDS | slow machine bootfest | Eurocrate key turning reboots for c1susaux, c1auxex,c1auxey, c1iscaux, c1iscaux2 and c1aux. Usual precautions were taken for ITMX. Did burtrestore for c1iscaux andc1iscaux2 in order to restore the LSC PD whitening gains.
Un-related to this work: input pointing into PMC was tweaked as the PMC_REFL spot was pretty bright. |
13354
|
Tue Oct 3 01:58:32 2017 |
johannes | HowTo | Cameras | CCD calibration | Disclaimer: Wrong calibration factors! See https://nodus.ligo.caltech.edu:8081/40m/13391
The factors were indeed enormously off. The correct table reads:
Camera IP |
Calibration Factor CF |
192.168.113.152 |
85.8 pW*s |
192.168.113.153 |
78.3 pW*s |
I did subtract a 'dark' frame from the images, though not in the sense of your point 1, just an exposure of identical duration with the laser turned off. This was mostly to reduce the effect of residual light, but given similar initial conditions would somewhat compensate for the offset that pre-existing charge and electronics noise put on the pixel values. The white field is of course a difference story.
I wonder how close we can get to a white field by putting a thin piece of paper in front of the camera without lenses and illuminate it from the other side. A problem is of course the coherence if we use a laser source... Or we scrap any sort of screen/paper and illuminate directly with a strongly divergent beam? Then there wouldn't be a specular pattern.
I'm not sure I understand your point about the 1.5V/A. Just to make sure we're talking about the same thing I made a crude drawing:

The PD sees plenty of light at all times, and the 1.5V/uW came from a comparative measurement PD<-->Ophir (which took the place of the CCD) while adjusting the power deflected with the AOM, so it doesn't have immediate connection to the conversion gain of silicon in this case. I can't remember the gain setting of the PD, but I believe it was 0dB, 20dB at most. |
Attachment 1: gige_calibration.pdf
|
|
13353
|
Tue Oct 3 01:32:39 2017 |
gautam | Update | LSC | Laser intensity noise coupling to MICH (simulated) | GV Oct 6: This coupling is probably not correct - Finesse outputs TF magnitude in units of W/W, and not W/RIN.
Since I was foiled (by lack of DAC) in my attempt to measure the coupling of laser intensity noise to MICH in the DRMI (no arms) configuration, I decided to try understanding the effect with a simulation.
For this purpose, I used my DRMI Finesse model - this had mirror positions tuned for locking and photodiode demod phases tuned to give a sensing matrix model that wasn't too far from an actual measurement (within factor of a few). So the model seems okay for a first pass at estimating this coupling.
Measuring transfer functions in Finesse is straightforward - use the fsig command to modulate some quantity (in this case the input beam intensity), and use the pd2 detector to demodulate the effect of this modulation at the port of interest (in this case AS55_Q).
**Note that to apply a modulation to an input beam (i.e. Laser) in Finesse, the keyword for the "type" argument given to fsig is "amp" and not "amplitude" as the manual would had me believe. In fact, there seem to be quite a few such caveats. The best way to figure this out is to go to the pykat installation directory, find the file components.py, and look for the fsig_name for the component of interest. It is also indicated in the same file, via the canFsig argument, if that property of the component can be modulated for transfer function measurements.
Attachment #1 shows the result of such a sweep.
To estimate what the actual contribution to the displacement noise is, I used the DQ-ed MC transmission (recorded at 1024Hz) from the DRMI lock, computed the ASD using scipy.signal.welch, divided by the nominal MC transmission of ~15,000 counts to convert to RIN/rtHz. The RIN was then multiplied by the above calculated coupling function, and divided by the sensing matrix element for AS55_Q (in units of W/m) to give the curve shown in Attachment #2. If we believe the simulation, then Laser Intensity Noise shouldn't be the limiting noise between 10Hz-1kHz.
I will of course measure the actual coupling and see how it lines up with Attachment #1 - would be a nice additional validation of the Finesse model. I will also try using the Finesse model to estimate some other coupling transfer functions (e.g. Laser Frequency Noise, Oscillator Noise).
Quote: |
The absence of evidence is not evidence of absence.
|
|
Attachment 1: MICH_intensityNoiseCoupling.pdf
|
|
Attachment 2: MICH_intensityNoiseASD.pdf
|
|
13352
|
Mon Oct 2 23:16:05 2017 |
gautam | HowTo | Cameras | CCD calibration | Going through some astronomy CCD calibration resources ([1]-[3]), I gather that there are in general 3 distinct types of correction that are applied:
- Dark frames --- this would be what we get with a "zero duration" capture, some documents further subdivide this into various categories like thermal noise in the CCD / readout electronics, poissonian offsets on individual pixels etc.
- Bias frames --- this effect is attributed to the charge applied to the CCD array prior to the readout.
- Flat-field calibration --- this effect accounts for the non-uniform responsivity of individual pixels on the CCDs.
The flat-field calibration seems to be the most complicated - the idea is to use a source of known radiance, and capture an image of this known radiance with the CCD. Then assuming we know the source radiance well enough, we can use some math to back out what the actual response function of individual pixels are. Then, for an actual image, we would divide by this response-map to get the actual image. There are a number of assumptions that go into this, such as:
- We know the source radiance perfectly (I guess we are assuming that the white paper is a Lambertian scatterer so we know its BRDF, and hence the radiance, perfectly, although the work that Jigyas and Amani did this summer suggest that white paper isn't really a Lambertian scatterer).
- There is only one wavelength incident on the CCD.
- We can neglect the effects of dust on the telescope/CCD array itself, which would obviously modify the responsivity of the CCD, and is presumably not stationary. Best we can do is try and keep the setup as clean as possible during installation.
I am not sure what error is incurred by ignoring 2 and 3 in the list at the beginning of this elog, perhaps this won't affect our ability to estimate the scattered power from the test-masses to within a factor of 2. But it may be worth it to do these additional calibration steps.
I also wonder what the uncertainty in the 1.5V/A number for the photodiode is (i.e. how much do we trust the Ophir power meter at low power levels?). The datasheet for the PDA100A says the transimpedance gain at 60dB gain is 1.5 MV/A (into high impedance load), and the Si responsivity at 1064nm is ~0.25A/W, so naively I would expect 0.375 V/uW which is ~factor of 4 lower. Is there a reason to trust one method over the other?
Also, are the calibration factor units correct? Jigyasa reported something like 0.5nW s / ct in her report.
Camera IP |
Calibration Factor CF |
192.168.113.152 |
8.58 W*s |
192.168.113.153 |
7.83 W*s |
The incident power can be calculated as Pin =CF*Total(Counts-DarkCounts)/ExposureTime.
|
References:
[1] http://www.astrophoto.net/calibration.php
[2] https://www.eso.org/~ohainaut/ccd/
[3] http://www.astro.ufl.edu/~lee/ast325/handouts/ccd.pdf |
13351
|
Mon Oct 2 19:03:49 2017 |
gautam | Update | CDS | [Solved] c1ioo DC errors | This did the trick - I simply ran
sudo systemctl restart daqd_*
on FB1, and now all the CDS overview lights are green again.
I thought I had done this already, but I realize that I was supposed to restart the daqd processes on FB1 (which is where they are running) and not on c1ioo .
Thanks Jamie for the speedy resolution!
Quote: |
Quote: |
- This time the model came back up but I saw a "0x2000" error in the GDS overview MEDM screen.
- Since there are no DACs installed in the c1ioo expansion chassis, I thought perhaps the problem had to do with the fact that there was a "DAC_0" block in the c1x03 simulink diagram - so I deleted this block, recompiled c1x03, and for good measure, restarted all (three) models on c1ioo.
- Now, however, I get the same 0x2000 error on both the c1x03 and c1als GDS overview MEDM screens (see Attachment #1).
|
From page 21 of T1100625, DAQ status "0x2000" means that the channel list is out of sync between the front end and the daqd. This usually happens when you add channels to the model and don't restart the daqd processes, which sounds like it might be applicable here.
It looks like open-mx is loaded fine (via "rtcds lsmod"), even though the systemd unit is complaining. I think this is because the open-mx service is old style and is not intended for module loading/unloading with the new style systemd stuff.
|
|
Attachment 1: CDSoverview.png
|
|
13350
|
Mon Oct 2 18:50:55 2017 |
jamie | Update | CDS | c1ioo DC errors |
Quote: |
- This time the model came back up but I saw a "0x2000" error in the GDS overview MEDM screen.
- Since there are no DACs installed in the c1ioo expansion chassis, I thought perhaps the problem had to do with the fact that there was a "DAC_0" block in the c1x03 simulink diagram - so I deleted this block, recompiled c1x03, and for good measure, restarted all (three) models on c1ioo.
- Now, however, I get the same 0x2000 error on both the c1x03 and c1als GDS overview MEDM screens (see Attachment #1).
|
From page 21 of T1100625, DAQ status "0x2000" means that the channel list is out of sync between the front end and the daqd. This usually happens when you add channels to the model and don't restart the daqd processes, which sounds like it might be applicable here.
It looks like open-mx is loaded fine (via "rtcds lsmod"), even though the systemd unit is complaining. I think this is because the open-mx service is old style and is not intended for module loading/unloading with the new style systemd stuff. |
13349
|
Mon Oct 2 18:08:10 2017 |
gautam | Update | CDS | c1ioo DC errors | I was trying to set up a DAC channel to interface with the AOM driver on the PSL table.
- It would have been most convenient to use channels from c1ioo given proximity to the PSL table.
- Looking at the 1X2 rack, it looked like there were indeed some spare DAC channels available.
- So I thought I'd run a test by adding some TPs to the c1als model (because it seems to have the most head room in terms of CPU time used).
- I added the DAC_0 block from CDS_PARTS library to c1als model (after confirming that the same part existed in the IOP model, c1x03).
- Model recompiled fine (I ran rtcds make c1als and rtcds install c1als on c1ioo).
- However, I got a bunch of errors when I tried to restart the model with rtcds restart c1als. The model itself never came up.
- Looking at dmesg, I saw stuff like
[4072817.132040] c1als: Failed to allocate DAC channel.
[4072817.132040] c1als: DAC local 0 global 16 channel 4 is already allocated.
[4072817.132040] c1als: Failed to allocate DAC channel.
[4072817.132040] c1als: DAC local 0 global 16 channel 5 is already allocated.
[4072817.132040] c1als: Failed to allocate DAC channel.
[4072817.132040] c1als: DAC local 0 global 16 channel 6 is already allocated.
[4072817.132040] c1als: Failed to allocate DAC channel.
[4072817.132040] c1als: DAC local 0 global 16 channel 7 is already allocated.
[4073325.317369] c1als: Setting stop_working_threads to 1
- Looking more closely at the log messages, it seemed like rtcds could not find any DAC cards on c1ioo.
- I went back to 1X2 and looked inside the expansion chassis. I could only find two ADC cards and 1 BIO card installed. The SCSI cable labelled ("DAC 0") running from the rear of the expansion chassis to the 1U SCSI->40pin IDE breakout chassis wasn't actually connected to anything inside the expansion chassis.
- I then undid my changes (i.e. deleted all parts I added in the simulink diagram), and recompiled c1als.
- This time the model came back up but I saw a "0x2000" error in the GDS overview MEDM screen.
- Since there are no DACs installed in the c1ioo expansion chassis, I thought perhaps the problem had to do with the fact that there was a "DAC_0" block in the c1x03 simulink diagram - so I deleted this block, recompiled c1x03, and for good measure, restarted all (three) models on c1ioo.
- Now, however, I get the same 0x2000 error on both the c1x03 and c1als GDS overview MEDM screens (see Attachment #1).
- An elog search revealed that perhaps this error is related to DAQ channels being specified without recording rates (e.g. 16384, 2048 etc). There were a few DAQ channels inside c1als which didn't have recording rates specified, so I added the rates, and restarted the models, but the errors persist.
- According to the RCG runtime diagnostics document, T1100625 (which admittedly is for RCG v 2.7 while we are running v3.4), this error has to do with a mismatch between the DAQ config files read by the RTS and the DAQD system, but I'm not sure how to debug this further.
- I also suspect there is something wrong with the mx processes:
controls@c1ioo:~ 130$ sudo systemctl status mx
● open-mx.service - LSB: starts Open-MX driver
Loaded: loaded (/etc/init.d/open-mx)
Active: failed (Result: exit-code) since Tue 2017-10-03 00:27:32 UTC; 34min ago
Process: 29572 ExecStop=/etc/init.d/open-mx stop (code=exited, status=1/FAILURE)
Process: 32507 ExecStart=/etc/init.d/open-mx start (code=exited, status=1/FAILURE)
Oct 03 00:27:32 c1ioo systemd[1]: Starting LSB: starts Open-MX driver...
Oct 03 00:27:32 c1ioo open-mx[32507]: Loading Open-MX driver (with ifnames=eth1 )
Oct 03 00:27:32 c1ioo open-mx[32507]: insmod: ERROR: could not insert module /opt/3.2.88-csp/open-mx-1.5.4/modules/3.2.88-csp/open-mx.ko: File exists
Oct 03 00:27:32 c1ioo systemd[1]: open-mx.service: control process exited, code=exited status=1
Oct 03 00:27:32 c1ioo systemd[1]: Failed to start LSB: starts Open-MX driver.
Oct 03 00:27:32 c1ioo systemd[1]: Unit open-mx.service entered failed state.
- Not sure if this is related to the DC error though.
|
Attachment 1: c1ioo_CDS_errors.png
|
|
13348
|
Mon Oct 2 12:44:45 2017 |
johannes | Update | Cameras | Basler 120gm calibration | Disclaimer: Wrong calibration factors! See https://nodus.ligo.caltech.edu:8081/40m/13391
The two acA640-120gm Basler GigE-Cams have been calibrated. I used the collimated output of a fiber that carried the auxiliary laser light from the PSL table. With a non-polarizing beam splitter some of the light was picked off onto a PD, and I modified the RF amplitude of the AOM drive signal to vary the power coming out of the fiber. The fiber output was directed at a white paper, which was placed 1.06m from the front of the lens tube assembly, which is where the focal plane is. Using the Pylon Viewer App I made sure that the entirety of the beam spot was imaged onto the CCD. Since the camera sensor is 1/4" across, I removed the camera from the lens tube and instead placed the Ophir power meter head at the position of the sensor and measured the power reported versus PD voltage, which turned out to be 1.5 V/uW.
The camera was put back in place and I used the Pypylon package Gautam had stumbled upon to sweep the exposure time from 100us to 10ms at different light power settings including no laser light at all for background subtraction, and rather than keeping the full bitmap data for O(100s) of images I recorded only the quantities
- Pixel Max
- Pixel Sum
- Pixel Mean
- Pixel Standard Deviation
- Pixel Median
I performed this procedure for both the 152 and 153 cameras and plotted the pixel sum and the pixel max vs the exposure time. All the exposures were taken at a gain setting of 100, which is the smallest possible setting (out of 100-600). To obtain the calibration factor I use the input power Pin=75nW in the 'safe' region 1ms to 10ms where the pixel sum looks smooth and the CCD is reportedly not saturated.
Camera IP |
Calibration Factor CF |
192.168.113.152 |
8.58 W*s |
192.168.113.153 |
7.83 W*s |
The incident power can be calculated as Pin =CF*Total(Counts-DarkCounts)/ExposureTime. |
Attachment 1: calib_20170930_152.pdf
|
|
Attachment 2: calib_20170930_153.pdf
|
|
13347
|
Fri Sep 29 18:36:25 2017 |
gautam | Update | LSC | DAC noise measurement (again) | BS connections and damping restored.
Quote: |
I am running some more measurements of the DAC noise, for which I've shut down the BS watchdog. Some of the cables on the coil driver side have been disconnected.
I will restore these tomorrow.
|
|
13346
|
Fri Sep 29 11:16:52 2017 |
Steve | Update | ALS | Y End table corrected | The first Faraday isolater rejected beam path from the NPRO is fixed.
|
Attachment 1: ETMYf1.jpg
|
|
13345
|
Fri Sep 29 11:07:16 2017 |
gautam | Update | CDS | 40m files backup situation | The FB1 dd backup process seems to have finished too - but I got the following message:
dd: error writing ‘/dev/sdc’: No space left on device
30523666+0 records in
30523665+0 records out
2000398934016 bytes (2.0 TB) copied, 50865.1 s, 39.3 MB/s
Running lsblk shows the following:
controls@fb1:~ 32$ lsblk
NAME MAJ:MIN RM SIZE RO TYPE MOUNTPOINT
sdb 8:16 0 23.5T 0 disk
└─sdb1 8:17 0 23.5T 0 part /frames
sda 8:0 0 2T 0 disk
├─sda1 8:1 0 476M 0 part /boot
├─sda2 8:2 0 18.6G 0 part /var
├─sda3 8:3 0 8.4G 0 part [SWAP]
└─sda4 8:4 0 2T 0 part /
sdc 8:32 0 1.8T 0 disk
├─sdc1 8:33 0 476M 0 part
├─sdc2 8:34 0 18.6G 0 part
├─sdc3 8:35 0 8.4G 0 part
└─sdc4 8:36 0 1.8T 0 part
While I am able to mount /dev/sdc1, I can't mount /dev/sdc4, for which I get the error message
controls@fb1:~ 0$ sudo mount /dev/sdc4 /mnt/HGSTbackup/
mount: wrong fs type, bad option, bad superblock on /dev/sdc4,
missing codepage or helper program, or other error
In some cases useful info is found in syslog - try
dmesg | tail or so.
Looking at dmesg, it looks like this error is related to the fact that we are trying to clone a 2TB disk onto a 1.8TB disk - it complains about block size exceeding device size.
So if we either have to get a larger disk (4TB?) to do the dd backup, or do the backing up some other way (e.g. unmount /frames RAID, delete everything in /frames, and then do dd , as Jamie suggested). If I understand correctly, unmounting /frames RAID will require that we stop all the daqd processes for the duration of the dd backup
Quote: |
sudo dd if=/dev/sda of=/dev/sdb bs=64K conv=noerror,sync
in a tmux session on nodus (as I did for chiara and FB1 , latter backup is still running).
|
Edit: unmounting /frames won't help, since dd makes a bit for bit copy of the drive being cloned. So we need a drive with size that is >= that of the drive we are trying to clone. On FB1, this is /dev/sda, which has a size of 2TB. The HGST drive we got has an advertised size of 2TB, but looks like actually only 1.8TB is available. So I think we need to order a 4TB drive. |
13344
|
Fri Sep 29 09:43:52 2017 |
jamie | HowTo | CDS | pyawg |
Quote: |
I've modified the __init.py__ file located at /ligo/apps/linux-x86_64/cdsutils-480/lib/python2.7/site-packages/cdsutils/__init__.py so that you can now simply import pyawg from cdsutils . On the control room workstations, iPython is set up such that cdsutils is automatically imported as "cds ". Now this import also includes the pyawg stuff. So to use some pyawg function, you would just do (for example):
exc=cds.awg.ArbitraryLoop(excChan,excit,rate=fs)
One could also explicitly do the import if cdsutils isn't automatically imported:
from cdsutils import awg
pyawg- away!
Linking this useful instructional elog from Chris here: https://nodus.ligo.caltech.edu:8081/Cryo_Lab/1748
|
? Why aren't you able to just import 'awg' directly? You shouldn't have to import it through cdsutils. Something must be funny with the config. |
13343
|
Thu Sep 28 23:50:04 2017 |
gautam | Update | LSC | DAC noise measurement (again) | I am running some more measurements of the DAC noise, for which I've shut down the BS watchdog. Some of the cables on the coil driver side have been disconnected.
I will restore these tomorrow.
As Rana pointed out to me, one important fact to keep in mind w.r.t. DAC noise is that it can be non-linear. So the RMS of the DAC noise in a higher frequency band (say 60-100Hz) can be affected by the RMS of the requested DAC signal in some lower frequency band (say 10-20Hz). One test to see if this hypothesis can explain the difference @100Hz between the ITMX channels and BS channels I observed a couple of days ago is to see if the noise around 100Hz becomes lower when I enable a 20-40Hz bandstop in the digital signal chain. |
13342
|
Thu Sep 28 23:47:38 2017 |
gautam | Update | CDS | 40m files backup situation | The nodus backup too is now complete - however, I am unable to mount the backup disk anywhere. I tried on a couple of different machines (optimus , chiara and pianosa ), but always get the same error:
mount: unknown filesystem type 'LVM2_member'
The disk itself is being recognized, and I can see the partitions when I run lsblk, but I can't get the disk to actually mount.
Doing a web-search, I came across a few blog posts that look like the problem can be resolved using the vgchange utility - but I am not sure what exactly this does so I am holding off on trying.
To clarify, I performed the cloning by running
sudo dd if=/dev/sda of=/dev/sdb bs=64K conv=noerror,sync
in a tmux session on nodus (as I did for chiara and FB1 , latter backup is still running). |
13341
|
Thu Sep 28 23:32:38 2017 |
gautam | HowTo | CDS | pyawg | I've modified the __init.py__ file located at /ligo/apps/linux-x86_64/cdsutils-480/lib/python2.7/site-packages/cdsutils/__init__.py so that you can now simply import pyawg from cdsutils . On the control room workstations, iPython is set up such that cdsutils is automatically imported as "cds ". Now this import also includes the pyawg stuff. So to use some pyawg function, you would just do (for example):
exc=cds.awg.ArbitraryLoop(excChan,excit,rate=fs)
One could also explicitly do the import if cdsutils isn't automatically imported:
from cdsutils import awg
pyawg- away!
Linking this useful instructional elog from Chris here: https://nodus.ligo.caltech.edu:8081/Cryo_Lab/1748 |
13340
|
Thu Sep 28 11:13:32 2017 |
jamie | Update | CDS | 40m files backup situation |
Quote: |
After consulting with Jamie, we reached the conclusion that the reason why the root of FB1 is so huge is because of the way the RAID for /frames is setup. Based on my googling, I couldn't find a way to exclude the nfs stuff while doing a backup using dd, which isn't all that surprising because dd is supposed to make an exact replica of the disk being cloned, including any empty space. So we don't have that flexibility with dd. The advantage of using dd is that if it works, we have a plug-and-play clone of the boot disk and root filesystem which we can use in the event of a hard-disk failure.
- One option would be to stop all the daqd processes, unmount /frames, and then do a dd backup of the true boot disk and root filesystem.
- Another option would be to use rsync to do the backup - this way we can selectively copy the files we want and ignore the nfs stuff. I suspect this is what we will have to do for the second layer of backup we have planned, which will be run as a daily cron job. But I don't think this approach will give us a plug-and-play replacement disk in the event of a disk failure.
- Third option is to use one of the 2TB HGST drives, and just do a dd backup - some of this will be /frames, but that's okay I guess.
|
This is not quite right. First of all, /frames is not NFS. It's a mount of a local filesystem that happens to be on a RAID. Second, the frames RAID is mounted at /frames. If you do a dd of the underlying block device (in this case /dev/sda*, you're not going to copy anything that's mounted on top of it.
What i was saying about /frames is that I believe there is data in the underlying directory /frames that the frames RAID is mounted on top of. In order to not get that in the copy of /dev/sda4 you would need to unmount the frames RAID from /frames, and delete everything from the /frames directory. This would not harm the frames RAID at all.
But it doesn't really matter because the backup disk has space to cover the whole thing so just don't worry about it. Just dd /dev/sda to the backup disk and you'll just be copying the root filesystem, which is what we want. |
13339
|
Thu Sep 28 10:33:46 2017 |
gautam | Update | CDS | 40m files backup situation | After consulting with Jamie, we reached the conclusion that the reason why the root of FB1 is so huge is because of the way the RAID for /frames is setup. Based on my googling, I couldn't find a way to exclude the nfs stuff while doing a backup using dd, which isn't all that surprising because dd is supposed to make an exact replica of the disk being cloned, including any empty space. So we don't have that flexibility with dd. The advantage of using dd is that if it works, we have a plug-and-play clone of the boot disk and root filesystem which we can use in the event of a hard-disk failure.
- One option would be to stop all the daqd processes, unmount /frames, and then do a dd backup of the true boot disk and root filesystem.
- Another option would be to use rsync to do the backup - this way we can selectively copy the files we want and ignore the nfs stuff. I suspect this is what we will have to do for the second layer of backup we have planned, which will be run as a daily cron job. But I don't think this approach will give us a plug-and-play replacement disk in the event of a disk failure.
- Third option is to use one of the 2TB HGST drives, and just do a dd backup - some of this will be /frames, but that's okay I guess.
I am trying option 3 now. dd however does requrie that the destination drive size be >= source drive size - I'm not sure if this is true for the HGST drives. lsblk suggests that the drive size is 1.8TB, while the boot disk, /dev/sda, is 2TB. Let's see if it works.
Backup of chiara is done. I checked that I could mount the external drive at /mnt and access the files. We should still do a check of trying to boot from the LaCie backup disk, need another computer for that.
nodus backup is still not complete according to the console - there is no progress indicator so we just have to wait I guess.
Quote: |
Backups of the root filesystems of chiara and nodus are underway right now. I am backing them up to the 1 TB LaCie external hard drives we recently acquired.
We also wanted to do a backup of the root of FB1 - but I'm not sure if dd will work with the external hard drive, because I think it requires the backup disk size (for us, 1TB) to be >= origin disk size (which on FB1, according to df -h, is 2TB). Unsure why the root filesystem of FB is so big, I'm checking with Jamie what we expect it to be. Anyways we have also acquired 2TB HGST SATA drives, which I will use if the LaCie disks aren't an option.
|
|
13338
|
Thu Sep 28 06:35:44 2017 |
Chris | Update | LSC | DAC noise measurement (again) | Since you're monitoring two channels simultaneously, you could try subtracting them, as an alternative to carving out bandstops.
- Drive both channels with the same time series
- Tweak the filter module gains if needed to equalize the analog outputs
- Apply SR785 user math to subtract the two channels (or A-B mode of a single input, if you're not using that already?)
- Measure the residual, which should be the incoherent part containing DAC + coil driver noise
Subtraction can conceal certain annoying effects (like numerical noise or level crossing glitches) that remain coherent for two identical outputs. It might be worth experimenting with a differential offset or sinusoid, to try to break up that kind of coherence if it exists. |
13337
|
Wed Sep 27 23:44:45 2017 |
gautam | Update | ALS | Proposed PM measurement setup | Attachment #1 is a sketch of the proposed setup to measure the PM response of the EX NPRO. Previously, this measurement was done via PLL. In this approach, we will need to calibrate the DFD output into units of phase, in order to calibrate the transfer function measurement into rad/V. The idea is to repeat the same measurement technique used for the AM - take ~50 1 average measurements with the AG4395, and look at the statistics.
Some more notes:
- Delay line box is passive, just contains a length of cable.
- IQ Demodulation is done using an aLIGO 1U chassis unit, with the actual demod board electronics being D0902745
- The RF beatnote amplitude out of the IR beat PD is ~ -8dBm.
- The ZHL-3A amplifiers have gain of 24dB, so the amplified beat should be ~16dBm
- At the LSC rack, the amplified beat is split into two - one path goes to the LO input of D0902745 (so at most 13dBm), the other goes through the delay line.
- On the demod board, the LO signal is amplified with a AP1053, rated at 10dB gain, max output of 26dBm, so the signal levels should be fine for us, even though the schematic says the nominal LO level is 10dBm - moreover, I've ignored cable losses, insertion losses etc so we should be well within spec.
- The mixer is PE4140. The datasheet quotes LO levels of 17dBm for all the "nominal" tests, we should be within a couple of dBm of this number.
- There is no maximum value specified for the RF input signal level to the mixer on the datasheet, but I expect it to be <10dBm.
- We should park the beatnote around 30MHz as this should be well within the operational ranges for the various components in the signal chain.
|
Attachment 1: IMG_7609.JPG
|
|
13336
|
Wed Sep 27 22:25:21 2017 |
gautam | Update | LSC | DAC noise measurement (again) | Attachment #1: Summary of results of measurements made on Friday. There is a lot in this plot, here is a breakdown:
- I drove the excitation points of the coil output filter banks with raw time-series data downloaded during a DRMI lock with pyawg. Today during the meeting, Rana pointed out that we could just acquire median (as opposed to mean since the former is more immune to glitches during the averaging process) spectra during a lock, and then do the ifft in python to generate time series data for pyawg. Another advantage of doing it this way is that we don't need to store a large (~200MB in my case) file of 16k data for numerous channels. But since I already had this file, I decided against changing the methodology for this round of tests. Time series plots of the signals do not show any large glitches.
- The SR785 was used in dual channel mode to acquire spectra from 2 coil driver outputs simultaneously, in the interest of saving time. Input range was set to -32dbVpk, AC coupled, which was the smallest value that worked for the given signal profile. Spectra were taken from DC-200Hz, with 801 points, 25 averages. The DB15 output of the coil driver board was connected to a DB15 breakout board, and then a BNC->Pomona mini-grabber adapter was used to connect to the SR785 input. The newly acquired linear power supplies for the GPIB box mean that spurious 60Hz harmonics were not present.
- Initially, I had planned to enable various bandstops from 20Hz-200Hz, to get a more complete profile of the noise. But in the end, I only used two elliptic bandstops (6th order, 60dB stopband attenuation): 60-90Hz, for which data is plotted in red and 90-200Hz, for which data is plotted in green.
- I've used the same noise model as I used here, plotted in dashed grey (summed with SR785 noise at the above input range, with input terminated via 50ohm terminator) - but had to tweak the parameters to get the curve to line up with the measurement. It looks like there is considerable variation between DAC channels, and certainly between the ITMX channels and the BS channels as groups.
- I took the measurement in two conditions - with the coil de-whitening off (left column) and coil de-whitening on (right column). Note that the input to the excitation was acquired at the IN1 of the relevant filter bank, and since the de-whitening happens downstream of this, we don't have to do anything special.
- In the right column, I have also plotted the LISO modelled noise, which was shown to match well with the measured curve, admittedly only for one channel (for the coil driver alone, so I am not taking into account the noise of the de-whitening board - I will fix this once I dig up that data).
Some remarks:
- According to this measurement, the de-whitening filters are the same on the ITMX channels and BS channels. So I don't understand the difference in the right column for BS and ITMX channels.
- While there is considerable variation between channels and also between ITMX and BS, there is certainly >6dB of reduction in the DAC noise when the de-whitening is engaged. However, no improvement was seen in the MICH error signal spectrum between 60-300Hz. So we have to continue to investigate other noises that can explain the noise in that band.
- Also, the realized improvement in DAC noise by turning on the coil de-whitening seems marginal - the low pass has gain of ~-80dB at 100Hz, but we seem to be hitting some sort of electronics noise in all channels at the level of ~100nV/rtHz (assuming the actual DAC noise doesn't degrade significantly when the digital simulated de-whitening filter is engaged).
- It remains to do the test for the ITMY channels.
- It would be useful to visualize the incoherent sum of all these channels - this is what should go into the MICH displacement NB. To be added.
- I'm currently loading pyawg from my user directory. Need to figure out a place to put this and add it to $PATH.
Data + code for this plot will be attached later. |
Attachment 1: coilNoises.pdf
|
|
13335
|
Wed Sep 27 00:20:19 2017 |
gautam | Update | ALS | More AM sweeps | Attachment #1: Result of AM sweeps with EX laser crystal at nominal operating temperature ~ 31.75 C.
Attachment #2: Tarball of data for Attachment #1.
Attachment #3: Result of AM sweeps with EX laser crystal at higher operating temperature ~ 40.95 C.
Attachment #4: Tarball of data for Attachment #2.
Remarks:
- Confirmed that PDA 55 is in the "0dB" setting - the actual dial is unmarked, and has 5 states. I guessed that the left-most one is 0dB, and checked that if I twiddled the dial by one state to the right, the DC level on the scope increased by 10dB as advertized. Didn't check all the states.
- DC level is ~2.3V on a high-impedance scope. So it will be ~1.15V to a 50ohm load, which is what the DC block is. The inverse of this value is used to calibrate the vertical axis of the TF measurement to RIN/V.
- Input R (split RF source signal) attenuation: 20dB. Input A (PDA55 output) attenuation: 0dB.
- Main problem is still network hangups when trying to do many sweeps.
- Seems to persist even when I connect the GPIB box to one of the network switches - so don't think we can blame the WiFi.
- Need to explore possibility of speedup - takes >2hours to run ~50scans!
To-do:
- Overlay median and uncertainty plots for the two temp. settings. There is a visible diference in both the locations and depths/heights of various notches/peaks in the AM profile.
- Repeat test with a fast focusing lens to focus the beam more tightly on the PD active area to confirm that the measured AM is indeed due to the PZT drive and not from beam-jitter (presently, spot diameter is ~0.5x active area diameter, to eye).
- Get the PM data.
- Depending on what the PM data looks like, do a more fine-grained scan around some promising AM notches / PM peaks.
|
Attachment 1: TFAG4395A_26-09-2017_202344_FourSquare.pdf
|
|
Attachment 2: lowTemp.tgz
|
Attachment 3: TFAG4395A_26-09-2017_231630_FourSquare.pdf
|
|
Attachment 4: highTemp.tgz
|
13334
|
Tue Sep 26 22:11:08 2017 |
johannes | Update | Cameras | post-vent camera capture comparison | I configured the remaining GigE-Camera to work on the 40m network. We currently have 3 operational Basler cameras:
The 120gm's have been assigned the IPs 192.168.113.152 (was already configured) and 192.168.113.153 (freshly configured) and have been labeled accordingly. Note that it was not necessary to connect the out-of-the-box camera directly to a dedicated ethernet adapter whose IP was set manually to 169.254.0.XXX as pointed out in earlier posts - a few seconds after connecting the camera to the control room switch (with PoE adapter to power it) the camera showed up in the configuration software tool which is launched via
/opt/rtcds/caltech/c1/scripts/GigE/pylon5/bin/./IpConfigurator
and can be assigned a corrected, static IP.
We have a plethora of 2" tubes for the lens assembly, but not a great variety of focal lengths for 2" lenses. Present with the camera gear were two f=250 mm and one f=150 mm 2" lenses with a NIR broadband AR coating
To determine the lens positions relativ to the sensor I assumed that the camera we're setting up looks at its test mass from a distance of 1m. Using the two available focal lengths we can look for solutions which have reasonable lens separations <~10cm and suitable magnification. We primarily want to image the central mirror area onto a 1/4" sized sensor, which can be achieved with a magnification of ~1/8.

I chose a lens separation of 6cm, which gives a theoretical magnification of -.12 and a sensor-lens 2 distance of 7.95 cm. I placed the lenses accordingly in the tubes and checked the focusing with Gautam's help:

It's pretty close to what we would expect. We will do the calibration using the auxiliary laser on the PSL table. For this I temporarily routed a fiber from the PSL enclosure to the SP table. Since the main cable hole is sort of cramped it's going in through a gap near the ceiling instead.
|
Attachment 1: lens_distance.pdf
|
|
13333
|
Tue Sep 26 19:10:13 2017 |
gautam | Update | ALS | Fiber ALS setup neatened | [steve, gautam]
The Fiber ALS box has been installed on the existing shelf on the PSL table. We had to re-arrange some existing cabling to make this possible, but the end result seems okay (to me). The box lid was also re-installed.
Some stuff that still needs to be fixed:
- Power supply to ZHL amplifiers - it is coming from a table-top DC supply currently, we should hook these up to the Sorensens.
- We should probably extend the corrugated fiber protection tubing for the three fibers all the way up to the shelf.
Beat spectrum post changes to follow.
Quote: |
Is it better to mount the box in the PSL under the existing shelf, or in a nearby PSL rack?
Quote: |
Further characterization needs to be done, but the results of this test are encouraging. If we are able to get this kind of out of loop ALS noise with the IR beat, perhaps we can avoid having to frequently fine-tune the green beat alignment on the PSL table. It would also be ideal to mount this whole 1U setup in an electronics rack instead of leaving it on the PSL table
|
|
|
Attachment 1: IMG_7605.JPG
|
|
13332
|
Tue Sep 26 15:55:20 2017 |
gautam | Update | CDS | 40m files backup situation | Backups of the root filesystems of chiara and nodus are underway right now. I am backing them up to the 1 TB LaCie external hard drives we recently acquired.
I first initialized the drives by hooking them up to my computer and running the setup.app file. After this, plugging the drive into the respective machine and running lsblk, I was able to see the mount point of the external drive. To actually initialize the backup, I ran the following command from a tmux session called ddBackupLaCie:
sudo dd if=/dev/sda of=/dev/sdb bs=64K conv=noerror,sync
Here, /dev/sda is the disk with the root filesystem, and /dev/sdb is the external hard-drive. The installed version of dd is 8.13, and from version 8.21 onwards, there is a progress flag available, but I didn't want to go through the exercise of upgrading coreutils on multiple machines, so we just have to wait till the backup finishes.
We also wanted to do a backup of the root of FB1 - but I'm not sure if dd will work with the external hard drive, because I think it requires the backup disk size (for us, 1TB) to be >= origin disk size (which on FB1, according to df -h, is 2TB). Unsure why the root filesystem of FB is so big, I'm checking with Jamie what we expect it to be. Anyways we have also acquired 2TB HGST SATA drives, which I will use if the LaCie disks aren't an option.
|
13331
|
Tue Sep 26 13:40:45 2017 |
gautam | Update | CDS | NDS2 server restarted on megatron | Gabriele reported problems with the nds2 server again. I restarted it again.
update: had to do it again at 1730 today - unclear why nds2 is so flaky. Log files don't suggest anything obvious to me...
Quote: |
I was unable to download data using nds2. Gabriele had reported similar problems a week ago but I hadn't followed up on this.
I repeated steps 5-7 from elog 13161, and now it seems that I can get data from the nds2 servers again. Unclear why the nds2 server had to be restarted. I wonder if this is somehow related to the mysterious acromag EPICS server tmux session dropout.
|
|
13330
|
Mon Sep 25 17:56:33 2017 |
johannes | Update | Computer Scripts / Programs | transmitted power during lossmap | I had to do a reboot + burt restore of c1psl today. It was unresponsive and I couldn't get the PMC to lock. I also had to slightly realign the PMC, and the IMC was too misaligned for the autolocker to catch lock. Adjusting it manually, it was predominantly MC1 PIT that was off. The YARM locked on a 10 mode and had to be aligned manually as well.
I left a script running on Donnatella that tilts ETMX and thus moves the beam on ITMX. I'm monitoring the transmitted power to evaluate sane thresholds for the demodulation offsets in a lossmap measurement. The script will return the IFO to normal after it is done and will take <2 hours to complete (no real clue, but there's no way it takes longer than that for ~50 datapoints). |
13329
|
Sun Sep 24 20:47:15 2017 |
rana | Update | Computer Scripts / Programs | RF TF Uncertainties | I have made several changes to Craig's script for better pythonism. Its more robust with different libraries and syntaxes and makes a tarball by default (w/o a command line flag). These kinds of general util scripts will be going into a general use folder in the git.ligo.org/40m/ team area so that it can be used throughout the LSC.
I don't think we need/want a coherence calculation, so I have not included it. Usually, we use coherence to estimate the uncertainty, and here we are just plotting it directly from the dist of the sweeps so coherence seems superfluous. |
Attachment 1: TFAG4395A_21-09-2017_115547_FourSquare.pdf
|
|
Attachment 2: TFAG4395A_21-09-2017_115547.tgz
|
13328
|
Fri Sep 22 18:12:27 2017 |
gautam | Update | LSC | DAC noise measurement (again) | I've been working on setting up some scripts for measuring the DAC noise.
In all the DRMI noise budgets I've posted, the coil-driver noise contribution has been based on this measurement, which could be improved in a couple of ways:
- The measurement was made at the output of the AI board - we can make the measurement at the output of the coil driver board, which will be a closer reflection of the actual current noise at the OSEM coils.
- The measurement was made by driving the DAC with shaped random noise - but we can record the signal to the coils during a lock and make the noise measurement by using awg to drive the coil with this signal, with elliptic bandstops at appropriate frequencies to reveal the electronics noise.
- The IN1 signals to the coils aren't DQ-ed, but ideally this is the signal we want to inject into the coil_EXC channel for this measurement - so I re-locked the DRMI a couple of nights ago and downloaded the coil IN1 channel data for ~5mins for the ITMs and BS.
- AWGGUI supposedly has a feature that allows you to drive an EXC channel with an arbitrary signal - but I couldn't figure out how to get this working. I did find some examples of this kind of application using the Python awg packages, so I cobbled together some scripts that allows me to drive some channels and place elliptic bandstop filters as I required.
- I wasted quite a bit of time trying to implement these signals in Python using available scipy functions, on account of me being a DSP n00b
. When trying to design discrete-time filters, of course numerical precision errors become important. Initially I was trying to do everything in the "Transfer function (numerator-denominator)" basis, but as Rana pointed out, the way to go is using SOSs. Fortunately, this is a simple additional argument to the relevant python functions, after which elliptic bandstop filter design was trivial.
- The actual test was done as follows:
- Save EPICS PIT/YAW offsets, set them to 0, disable Oplev servos, and then shut down optic watchdog once the optic is somewhat damped. This is to avoid the optics getting a large kick when disconnecting the DB15 connector from the coil driver board output.
- Disconnect above-mentioned DB15 connector from the appropriate coil driver board output.
- Turn off inputs to coils in filter module EPICs screens. Since the full signal (local damping servo output + Oplev servo output + LSC servo output) to the coil during a DRMI lock will be injected as an excitation, we don't need any other input.
- Use scripts (which I will upload to a git repo soon) to set up the appropriate excitation.
- To measure the spectrum, I used a DB15 breakout board with test-points soldered on and some mini-grabber-to-BNC adaptors, in order to interface the SR785 to the coil driver output. We can use the two input channels of the SR785 to simultaneously measure two coil driver board output channels to save some time.
- Take a measurement of the SR785 noise (at the appropriate "Input Range" setting) with inputs terminated to estimate the analyzer noise floor.
- Just for kicks, I made the measurement with the de-whitening both OFF/ON.
I only managed to get in measurements for the BS and ITMX today. ITMY to be measured later, and data/analysis to follow.
The ITMX and BS alignments have been restored after this work in case anyone else wants to work with the IFO.
Some slow machine reboots were required today - c1susaux was down, and later, the MC autolocker got stuck because of c1iool0 being unresponsive. I thought we had removed all dependency of the autolocker on c1iool0 when we moved the "IFO-STATE" EPICS variable to the c1ioo model, but clearly there is still some dependancy. To be investigated. |
13327
|
Thu Sep 21 15:23:04 2017 |
gautam | Omnistructure | ALS | Long cable from LSC->IOO | [steve,gautam]
We laid out a 45m long BNC cable from the LSC rack to the IOO rack via overhead cable trays. There is ~5m excess length on either side, which have been coiled up and cable-tied for now. The ends are labelled "TO LSC RACK" and "TO IOO RACK" on the appropriate ends. This is to facilitate hooking up the output of the DFD for making a PM measurement of the AUX X laser. There is already a long cable that runs from the IOO rack to the X end. |
|