ID |
Date |
Author |
Type |
Category |
Subject |
16166
|
Fri May 28 10:54:59 2021 |
Jon | Update | CDS | Opto-isolator for c1auxey |
I have received the opto-isolator needed to complete the new c1auxey system. I left it sitting on the electronics bench next to the Acromag chassis.
Here is the manufacturer's wiring manual. It should be wired to the +15V chassis power and to the common return from the coil driver, following the instructions herein for NPN-style signals. Note that there are two sets of DIP switches (one on the input side and one on the output side) for selecting the mode of operation. These should all be set to "NPN" mode. |
Attachment 1: optoisolator.jpeg
|
|
16178
|
Thu Jun 3 17:15:17 2021 |
Yehonathan | Update | CDS | Opto-isolator for c1auxey |
As Jon wrote we need to use the NPN configuration (see attachments). I tested the isolator channels in the following way:
1. I connected +15V from the power supply to the input(+) contact.
2. Signal wire from one of the digital outputs was connected to I1-4
3. When I set the digital output to HIGH, the LED on the isolator turns on.
4. I measure the resistance between O1-4 to output(-) and find it to be ~ 100ohm in the HIGH state and an open circuit in the LOW state, as expected from an open collector output.
Unlike the Acromag output, the isolator output is not pulled up in the LOW state. To do so we need to connect +15V to the output channel through a pull-up resistor. For now, I leave it with no pull-up. According to the schematics of the HAM-A Coil Driver, the digital output channels drive an electromagnetic relay (I think) so it might not need to be pulled up to switch back. I'm not sure. We will need to check the operation of these outputs at the installation.
During the testing of the isolator outputs pull-up, I accidentally ran a high current through O2, frying it dead. It is now permanently shorted to the + and - outputs rendering it unusable. In any case, we need another isolator since we have 5 channels we need to isolate.
I mounted the isolator on the DIN rail and started wiring the digital outputs into it. I connected the GND from the RTS to output(-) such that when the digital outputs are HIGH the channels in the coil driver will be sunk into the RTS GND and not the slow one avoiding GND contamination. |
Attachment 1: Optical_Isolator_NPN_Input.png
|
|
Attachment 2: Optical_Isolator_NPN_Output.png
|
|
16181
|
Thu Jun 3 22:08:00 2021 |
Koji | Update | CDS | Opto-isolator for c1auxey |
- Could you explain what is the blue thing in Attachment 1?
- To check the validity of the signal chain, can you make a diagram summarizing the path from the fast BO - BO I/F - Acromag - This opto-isolator - the coil driver relay? (Cut-and-paste of the existing schematics is fine)
|
16182
|
Fri Jun 4 14:49:23 2021 |
Yehonathan | Update | CDS | Opto-isolator for c1auxey |
I made a diagram (Attached). I think it explains the blue thing in the previous post.
I don't know what is the grounding situation in the RTS so I put a ground in both the coil driver and the RTS. Hopefully, only one of them is connected in reality.
Quote: |
- Could you explain what is the blue thing in Attachment 1?
- To check the validity of the signal chain, can you make a diagram summarizing the path from the fast BO - BO I/F - Acromag - This opto-isolator - the coil driver relay? (Cut-and-paste of the existing schematics is fine)
|
|
Attachment 1: Optical_isolator_Wiring.pdf
|
|
16183
|
Fri Jun 4 17:46:25 2021 |
unYehonathan | Update | CDS | Opto-isolator for c1auxey |
I mounted the optoisolator on the DIN rail and connected the 3 first channels
C1:SUS-ETMY_UL_ENABLE |
C1:SUS-ETMY_UR_ENABLE |
C1:SUS-ETMY_LL_ENABLE
|
to the optoisolator inputs 1,3,4 respectively. I connected the +15V input voltage into the input(+) of the optoisolator.
The outputs were connected to DB9F-2 where those channels were connected before.
I added DB9F-1 to the front panel to accept channels from the RTS. I connected the fast channels to connectors 1,2,3 from DB9F-1 to DB9F-2 according to the wiring diagram. The GND from DB9F-1 was connected to both connector 5 of DB9F-2 and the output (-).
I tested the channels: I connected a DB9 breakout board to DB9F-2. I measured the resistance between the RTS GND and the isolated channels while switching them on and off. In the beginning, when I turned on the binary channels the resistance was behaving weird - oscillating between low resistance and open circuit. I pulled up the channels through a 100Kohm resistor to observe whether the voltage behavior is reasonable or not. Indeed I observed that in the LOW state the voltage between the isolated channel and slow GND is 15V and 0.03V in the HIGH state. Then I disconnected the pull up from the channels and measured the resistance again. It showed ~ stable 170ohm in the HIGH state and an open circuit in the LOW state. I was not able to reproduce the weird initial behavior. Maybe the optoisolator needs some warmup of some sort.
We still need to wire the rest of the fast channels to DBF9-3 and isolate the channels in DBF9-4. For that, we need another optoisolator.
There is still an open issue with the BI channels not read by EPICS. They can still be read by the Windows machine though. |
Attachment 1: 20210604_173420.jpg
|
|
16184
|
Sun Jun 6 03:02:14 2021 |
Koji | Update | CDS | Opto-isolator for c1auxey |
This RTS also use the BO interface with an opto isolator. https://dcc.ligo.org/LIGO-D1002593
Could you also include the pull up/pull down situations? |
16186
|
Sun Jun 6 12:15:16 2021 |
Jon | Update | CDS | Opto-isolator for c1auxey |
Since this Ocean Controls optoisolator has been shown to be compatible, I've gone ahead and ordered 10 more:
- (1) to complete c1auxey
- (2) for the upgrade of c1auxex
- (7) for the upgrade of c1susaux
They are expected to arrive by Wednesday. |
16187
|
Sun Jun 6 15:59:51 2021 |
Yehonathan | Update | CDS | Opto-isolator for c1auxey |
According to the BO interface circuit board https://dcc.ligo.org/D1001266, PCIN wires are connected to the coil driver and they are not pulled either way.
That means that they're either grounded or floating. I updated the drawing.
|
Attachment 1: Optical_isolator_Wiring.pdf
|
|
16199
|
Mon Jun 14 15:31:30 2021 |
Yehonathan | Update | CDS | Opto-isolator for c1auxey |
I checked the BI situation on the HAM-A coil driver. It seems like these are sinking BIs and indeed need to be isolated from the Acromag unit GND to avoid contamination.
The BIs will have to be isolated on a different isolator. Now, the wires coming from the field (red) are connected to the second isolator's input and the outputs are connected to the Acromag BI module and the Acromag's RTN.
I updated the wiring diagram (attached) and the wiring spreadsheet.
In the diagram, you can notice that the BI isolator (the right one) is powered by the Acromag's +15V and switched when the coil driver's GND is supplied. I am not sure if it makes sense or not. In this configuration, there is a path between the coil driver's GND and the Acromag's GND but its resistance is at least 10KOhm. The extra careful option is to power the isolator by the coil driver's +V but there is no +V on any of the connectors going out of the coil driver.
I installed an additional isolator on the DIN rail and wired the remaining BOs (C1:SUS-ETMY_SD_ENABLE, C1:SUS-ETMY_LR_ENABLE) through it to the DB9F-4 feedthrough. I also added DB9F-3 for incoming wires from the RTS and made the required connection from it to DB9F-4.
I tested the new isolated BOs using the Windows machine (after stopping Modbus). As before, I measure the resistance between pin 5 (coil driver GND) and the channel under test. When I turn on the BO I see the resistance drops from inf to 166ohm and back to inf when I turn it off. Both channels passed the test.
|
Attachment 1: Optical_isolator_Wiring.pdf
|
|
16203
|
Tue Jun 15 21:48:55 2021 |
Koji | Update | CDS | Opto-isolator for c1auxey |
If my understanding is correct, the (photo receiving) NPN transistor of the optocoupler is energized through the acromag. The LED side should be driven by the coil driver circuit. It is properly done for the "enable mon" through 750Ohm and +V. However, "Run/Acquire" is a relay switch and there is no one to drive the line. I propose to add the pull-up network to the run/acquire outputs. This way all 8 outputs become identical and symmetric.
We should test the configuration if this works properly. This can be done with just a manual switch, R=750Ohm, and a +V supply (+18V I guess). |
Attachment 1: Acromag_RTS_BI_config.jpg
|
|
16205
|
Wed Jun 16 17:24:29 2021 |
Yehonathan | Update | CDS | Opto-isolator for c1auxey |
I updated the wiring diagram according to Koji's suggestion. According to the isolator manual, this configuration requires that the isolator input be configured as PNP.
Additionally, when the switch in the coil driver is open the LED in the isolator is signaling an on-state. Therefore, we might need to configure the Acromag to invert the input.
There are the Run/Aquire channels that we might need to add to the wiring diagram. If we do need to read them using slow channels, we will have to pull them up like the EnableMon channels to use them like in the wiring diagram. |
Attachment 1: Optical_isolator_Wiring.pdf
|
|
16207
|
Wed Jun 16 20:32:39 2021 |
Yehonathan | Update | CDS | Opto-isolator for c1auxey |
I installed 2 additional isolators in the Acromag chassis. I set all the input channels to PNP. I ran the digital inputs (EnableMon channels) through these isolators according to the previous post.
I tested the digital inputs in the following way:
I connected an 18V voltage source to the signal wire under test through a 1Kohm resistor. I connected the GND of the voltage source to the RTN wire of the feedthrough. When the voltage source was connected, the LED on the isolator turned on and the EPICs channel under test was Enabled. When I disconnected the voltage source or shorted the signal wire to GND the LED on the isolator turned off and the EPICs channel showed a Disabled state. |
16243
|
Fri Jul 9 18:35:32 2021 |
Yehonathan | Update | CDS | Opto-isolator for c1auxey |
Following Koji's channel list review, we made changes to the wiring spreadsheet.
Today, I made the changes real in the Acromag chassis. I went through the channel list one by one and made sure it is wired correctly. Additionally, since we now need all the channels the existing isolators have, I replaced the isolator with the defective channel with a new one.
The things to do next:
1. Create entries for the spare coil driver and satellite box channels in the EPICs DB.
2. Test the spare channels. |
16244
|
Mon Jul 12 18:06:25 2021 |
Yehonathan | Update | CDS | Opto-isolator for c1auxey |
I edited /cvs/cds/caltech/target/c1auxey1/ETMYaux.db (after creating a backup) and added the spare coil driver channels.
I tested those channels using caget while fixing wiring issues. The tests were all succesful. The digital output channel were tested using the Windows machine since they are locked by some EPICs mechanism I don't yet understand.
One worrying point is I found that the differential analog inputs to be unstable unless I connected a reference to some stable voltage source unlike previous tests showed. It was unstable (but less) even when I connected the ref to the ground connectors on the power supplies on the workbench. This is really puzzling.
When I say unstable I mean that most of the time the voltage reading shows the right value, but occasionly there is a transient sharp volage drop of the order of 0.5V. I will do a more quantitative analysis tomorrow.
|
16263
|
Wed Jul 28 12:47:52 2021 |
Yehonathan | Update | CDS | Opto-isolator for c1auxey |
To simulate a differential output I used two power supplies connected in series. The outer connectors were used as the outputs and the common connector was connected to the ground and used as a reference. I hooked these outputs to one of the differential analog channels and measured it over time using Striptool. The setup is shown in attachment 3.
I tested two cases: With reference disconnected (attachment 1), and connected (attachment 2). Clearly, the non-referred case is way too noisy. |
Attachment 1: SUS-ETMY_SparePDMon0_NoRef.png
|
|
Attachment 2: SUS-ETMY_SparePDMon0_Ref_WithGND.png
|
|
Attachment 3: DifferentialOutputTest.png
|
|
16276
|
Wed Aug 11 12:06:40 2021 |
Yehonathan | Update | CDS | Opto-isolator for c1auxey |
I redid the differential input experiment using the DS360 function generator we recently got. I generated a low frequency (0.1Hz) sine wave signal with an amplitude 0.5V and connected the + and - output to a differential input on the new c1auxcey Acromag chassis. I recorded a time series of the corresponding EPICS channel with and without the common on the DS360 connected to the Ref connector on the Acromag unit. The common connector on the DS360 is not normally grounded (there is a few tens of kohms between the ground and common connectors). The attachment shows that, indeed, the analog input readout is extremely noisy with the Ref being disconnected. The point where the Ref was connected to common is marked in the picture.
Conclusion: Ref connector on the analog input Acromag units must be connected to some stable voltage source for normal operation. |
Attachment 1: SUS-ETMY_SparePDMon0_2.png
|
|
16635
|
Tue Feb 1 15:33:28 2022 |
Koji | Summary | BHD | Optomechanics configuration for the in-vacuum aux small optics |
I've summarized the optomechanics configuration for the in-vacuum aux small optics
It's not obvious here but the post for POP_SM4 is the stack of BA2V, Newport 9953, PLS-T238, LMR1V. The mirror is a CM254-750-E03 curved mirror. This should go on the suspension base. I hope I did not make a mistake there... |
Attachment 1: Invac_opto_mechanics_v2.pdf
|
|
16640
|
Wed Feb 2 18:55:58 2022 |
Koji | Summary | BHD | Optomechanics configuration for the in-vacuum aux small optics |
Here is more detail of the POP_SM4 mount assembly.
It's a combination of BA2V + PLS-T238 + BA1V + TR-1.5 + LMR1V + Mirror: CM254-750-E03
Between BA1V and PLS-T238, we have to do a washer action to fix the post (8-32) with a 1/4-20 slot. Maybe we can use a 1" post shim from thorlabs/newport.
Otherwise, we should be able to fasten the other joints with silver-plated screws we already have/ordered.
I think TR-1.5 (and a shim) has not been given to Jordan for C&B. I'll take a look at these. |
Attachment 1: Screen_Shot_2022-02-02_at_6.46.46_PM.png
|
|
16641
|
Wed Feb 2 21:16:23 2022 |
Koji | Summary | BHD | Optomechanics configuration for the in-vacuum aux small optics |
Asked Jordan to clean 2x 1.5" posts (0.5 dia) and a washer with 1" dia.
|
Attachment 1: PXL_20220203_050156031.jpg
|
|
15319
|
Wed May 6 00:31:09 2020 |
gautam | Update | ALS | Optomechanics during CARM offset reduction |
Summary:
The apparent increase in the ALS noise (witnessed in-loop, e.g. Attachment #2 here) during the CARM offset reduction may have an optomechanical origin.
Details:
- A simplified CARM plant was setup in Finesse - 3 mirror coupled cavity with PRM, ITM and ETM, 40m params for R/T/L used.
- For a sanity check, DC power buildup and coincident resonance of the PRC and arm cavity were checked. PRG and CARM linewidth also checks out, and scales as expected with arm losses.
- To investigate possible optomechanical issues - I cut the input power to 300 mW (I estimate 600 mW incident on the PRM), set a PRG of ~20, to mimic what we have right now.
- I drive the ITM at various CARM offsets, and measure the m/m transfer function to itself and the ETM.
- Attachement #1 shows the results.
Interpretation:
- ericq had similar plots in his thesis, but I don't think the full implications of this effect were investigated, the context there was different.
- The optomechanical resonance builds up at ~10 Hz and sweeps up to ~100 Hz as the CARM offset approaches zero, with amplification close to x100 at the resonance.
- What this means is that the arm cavity is moving by up to 100x the ambient seismically driven dispalcements.
- The EX/EY uPDH servos have considerable gain at these frequencies, and so the AUX laser frequency can keep up with this increased motion (to be quantified exactly what the increase in residual is).
- However, the ALS loop that maintains the frequency offset b/w the PSL and the AUX lasers is digitial, and only has ~20 dB gain at 30 Hz. - so the error signal for CARM control becomes noisier as we see.
- I speculate that the multiple peaky features in the in-loop error signal are a result of some dynamical effects which Finesse presumably does not simulate.
- The other puzzler is: this simulation would suggest that approaching the zero CARM offset from the other side (anti-spring) wouldn't have such instabilities building up. However, I am reasonably sure I've seen this effect approaching zero from both sides, though I haven't checked in the last month.
- Anyways, if this hypothesis is correct, we can't really take advantage of the ~8pm RMS residual noise performance of the IR ALS system sadly, because of our 250g mirrors and 800mW input power.
- Possible workarounds:
- High BW ALS - this would give us more gain at ~30 Hz and this wouldn't be a problem anymore really. But in my trials, I think I found that the IN2 gain on the CM board has to be inverted for this to work (the IN1 path and the IN2 path share a common AO path polarity, and we need the two paths to have the opposite polarity).
- Cut the input power - this would reduce the optomechanical action, but presumably the vertex locking becomes noisier. In any case, this isn't really practical without some kind of motorized/remote-controlled waveplate for power adjustment.
Update 415pm 5/6: Per the discussion at the meeting, I have now uploaded as Attachment #2 the force-->displacement (i.e. m/N) transfer functions. I now think these are appropriate units. For the ALS case, we could convert the m/N to Hz/N of extra frequency noise imprinted on the AUX laser due to the increased cavity motion. Is W/N really better here, since the mechanism is extra frequency noise on a beatnote, and there isn't really a PDH or DC error signal? |
Attachment 1: CARMplant.pdf
|
|
Attachment 2: CARMplant_force2disp.pdf
|
|
16953
|
Tue Jun 28 09:03:58 2022 |
JC | Update | General | Organizing and Cleaning |
The plan for the tools in 40m
As of right now, there are 4 tool boxes. X-end, Y-end, Vertex, and the main tool box along the X-arm. The plan is the give each toolbox a set of their own tools. The tools of X-end, Y-end, and Vertex toolboxes will be very similar containing the basic tools such as pliers, screwdrivers, allen ball drivers. Along with this, each tool box will have a tape measure, caliper, level, and other measuring tools we find convinient.
As for the new toolbox, I have done research and found a few good selections. The only problem I have ran into with this is the width of the tool box corresponding with the prices. The tool cabinet we have now is 41" wide. The issue I have is not in finding another toolbox of the same width, but for a similar price we can find a 54" wide tool cabinet. Would anyone be objected to making a bit more space for this?
How the tools will stay organized.
I the original idea I had was to use a specified color of electrical tape for each tool box. Then to wrap the corresponding tools tools with the same color tape. But it was brought to my attention that the electrical tape would become sticky over time. So, I think the using the label maker would be the best idea. with the labels being 'X' for X-end, 'Y' for Y-end, 'V' for vertex, and 'M' for main toolboxes.
An idea for the optical tables:
Anchal brought it up to me that it is a hassle to go back and forth searching for the correct sizes of Hex Keys and Allen Wrenches. The idea of a pouch on the outside of each optical table was mentioned so I brought this up to Paco. Paco also gave me the idea of a 3D printed stand we could make for allen ball drives. Does anyone have a preference or an idea of what would be the best choice and why?
A few sidenotes:
Anchal mentioned to me a while back that there are many cables that are laying on the racks that are not being used. Is there a way we could identify which ones are being used?
I noticed that when we were vented that a few of the chamber doors were leaning up against the wall and not on a wooden stand like others. Although, the seats for the chamber doors are pretty spacious and do not give us much clearance. For the future ones, could we make something more sleek and put the wider seats at the end chambers?
The cabinets along the Y-Arm are labelled, but do not correspond with all the materials inside or are too full to take in more items. Could I organize these?
|
5969
|
Mon Nov 21 15:47:58 2011 |
Mirko | Update | IOO | Osem loop shape |
[Jenne, Mirko]
To reiterate: We changed the OSEM loop shape for MC1-MC3. Below in black is the old loop shape, which simulated pendulum response in there. In red is the new loop shape.

The differences are due to extra filter in C1:SUS-MC?_SUSPOS module 6,7,9
6: Elliptical LP @ 2.5Hz
7: Inverse Chebychev HP @0.3Hz
8: 1st order LP @ 10Hz
This has the potential to be unstable, but is not. At some point these filters should be tuned further. |
13071
|
Fri Jun 16 23:27:19 2017 |
Kaustubh, Jigyasa | Update | Computers | Ottavia Connected to the Netgear Box |
I just connected the Ottavia to the Netgear box and its working just fine. It'll remain switched on over the weekend.
Quote: |
Kaustubh and I are going to enable the ethernet connection to Ottavia and secure the wiring now.
|
|
13065
|
Thu Jun 15 14:24:48 2017 |
Kaustubh, Jigyasa | Update | Computers | Ottavia Switched On |
Today, I and Jigyasa connected the Ottavia to one of the unused monitor screens Donatella. The Ottavia CPU had a label saying 'SMOKED''. One of the past elogs, 11091, dated back in March 2015, by Jenne had an update regarding the Ottavia smelling 'burny'. It seems to be working fine for about 2 hours now. Once it is connected to the Martian Network we can test it further. The Donatella screen we used seems to have a graphic problem, a damage to the display screen. Its a minor issue and does not affect the display that much, but perhaps it'll be better to use another screen if we plan to use the Ottavia in the future. We will power it down if there is an issue with it. |
13067
|
Thu Jun 15 19:49:03 2017 |
Kaustubh, Jigyasa | Update | Computers | Ottavia Switched On |
It has been working fine the whole day(we didn't do much testing on it though). We are leaving it on for the night.
Quote: |
Today, I and Jigyasa connected the Ottavia to one of the unused monitor screens Donatella. The Ottavia CPU had a label saying 'SMOKED''. One of the past elogs, 11091, dated back in March 2015, by Jenne had an update regarding the Ottavia smelling 'burny'. It seems to be working fine for about 2 hours now. Once it is connected to the Martian Network we can test it further. The Donatella screen we used seems to have a graphic problem, a damage to the display screen. Its a minor issue and does not affect the display that much, but perhaps it'll be better to use another screen if we plan to use the Ottavia in the future. We will power it down if there is an issue with it.
|
|
13068
|
Fri Jun 16 12:37:47 2017 |
Kaustubh, Jigyasa | Update | Computers | Ottavia Switched On |
Ottavia had been left running overnight and it seems to work fine. There has been no smell or any noticeable problems in the working. This morning Gautam, Kaustubh and I connected Ottavia to the Matrian Network through the Netgear switch in the 40m lab area. We were able to SSH into Ottavia through Pianosa and access directories. On the ottavia itself we were able to run ipython, access the internet. Since it seems to work out fine, Kaustubh and I are going to enable the ethernet connection to Ottavia and secure the wiring now.
Quote: |
It has been working fine the whole day(we didn't do much testing on it though). We are leaving it on for the night.
Quote: |
Today, I and Jigyasa connected the Ottavia to one of the unused monitor screens Donatella. The Ottavia CPU had a label saying 'SMOKED''. One of the past elogs, 11091, dated back in March 2015, by Jenne had an update regarding the Ottavia smelling 'burny'. It seems to be working fine for about 2 hours now. Once it is connected to the Martian Network we can test it further. The Donatella screen we used seems to have a graphic problem, a damage to the display screen. Its a minor issue and does not affect the display that much, but perhaps it'll be better to use another screen if we plan to use the Ottavia in the future. We will power it down if there is an issue with it.
|
|
|
9925
|
Wed May 7 23:09:06 2014 |
rana, jamie | Summary | Computer Scripts / Programs | Ottavia back on network |
After Jamie fixed the third party repo issue with Ottavia, he was able to upgrade it to Ubuntu 12.04 Precise Pangolin. But its network stopped working.
I tried to fix its issues by ifconfig and GUI, but what it really wanted was for me to put the network cable back into its eth0 slot. The eth1 network card appears not be working anymore.
All seems fine now. Next I will mount the shared user disk from linux1 and put in a .bashrc. |
2878
|
Tue May 4 14:57:53 2010 |
josephb | Update | Computers | Ottavia has moved |
Ottavia was moved this afternoon from the control room into the lab, adjacent to Mafalda in 1Y3 on the top shelf. It has been connected to the camera hub, as well as the normal network. Its cables are clearly labeled. Note the camera hub cable should be plugged into the lower ethernet port. Brief tests indicate everything is connected and it can talk to the control room machines.
The space where Ottavia used to be is now temporarily available as a good place to setup a laptop, as there is keyboard, mouse, and an extra monitor available. Hopefully this space may be filled in with a new workstation in the near future. |
11091
|
Tue Mar 3 04:43:12 2015 |
Jenne | Omnistructure | PEM | Ottavia is dead? |
After some searching, including help from 4 security guys (I think they don't have a lot to do at 4:30am :), we found that Ottavia is super warm, and smelled burn-y. She has been powered down and unplugged. Security guys may call Steve's desk to follow up later today. |
10535
|
Wed Sep 24 18:56:45 2014 |
ericq | Update | General | Ottavia slowness |
Ottavia was having some severe interaction latency today. Xorg was taking up >90% of the CPU, just sitting around. The machine was logged in to a desktop session with lots of graphical effects turned on. I changed the system default session to "gnome-fallback" in /etc/lightdm/lightdm.conf, which was already set as the default for controls, but wouldn't get chosen for the autologin that happens on boot.
Hopefully this helps ottavia stay usable... |
9144
|
Fri Sep 20 08:15:30 2013 |
Steve | Update | Computer Scripts / Programs | Ottavia, Rossa and Pianosa |
Ottavia, Rossa and Pianosa are running out of storage space. |
Attachment 1: Ottavia.png
|
|
Attachment 2: Rossa.png
|
|
Attachment 3: Pianosa.png
|
|
16791
|
Wed Apr 20 16:11:08 2022 |
Koji | Update | SUS | Outpur resistors updated for LO1 coil drivers / for SR2, LO2, AS1, and AS4 in progress |
[JC Koji]
To give more alignment ranges for the SUS alignment, we started updating the output resistors of the BHD SUS coil drivers.
As Paco has already started working on LO1 alignment, we urgently updated the output Rs for LO1 coil drivers.
LO1 Coil Driver 1 now has R=100 // 1.2k ~ 92Ohm for CH1/2/3, and LO1 Coil Driver 2 has the same mod only for CH3. JC has taken the photos and will upload/update an elog/DCC.
We are still working on the update for the SR2, LO2, AS1, and AS4 coil drivers. They are spread over the workbench right now. Please leave them as they're for a while.
JC is going to continue to work on them tomorrow, and then we'll bring them back to the rack. |
Attachment 1: IMG_0527.jpeg
|
|
Attachment 2: IMG_0528.jpeg
|
|
Attachment 3: IMG_0529.jpeg
|
|
Attachment 4: IMG_0530.jpeg
|
|
Attachment 5: IMG_0531.jpeg
|
|
Attachment 6: IMG_0532.jpeg
|
|
16795
|
Thu Apr 21 15:22:44 2022 |
Koji | Update | SUS | Outpur resistors updates for SR2, LO2, AS1, and AS4 done |
[JC Koji]
Quick report: JC has done all the mods for the coil driver circuit in the morning and we worked on the reinstallation of them in the afternoon.
I'll check the damping loops / sus servo settings. JC is going to make an ELOG entry and DCC updates for more precise record of the mods. |
6599
|
Thu May 3 19:52:43 2012 |
Jenne | Update | CDS | Output errors from dither alignment (Xarm) script |
epicsThreadOnceOsd epicsMutexLock failed.
Segmentation fault
Number found where operator expected at -e line 1, near "0 0"
(Missing operator before 0?)
syntax error at -e line 1, near "0 0"
Execution of -e aborted due to compilation errors.
Number found where operator expected at (eval 1) line 1, near "* * 50"
(Missing operator before 50?)
epicsThreadOnceOsd epicsMutexLock failed.
Segmentation fault
Number found where operator expected at -e line 1, near "0 0"
(Missing operator before 0?)
syntax error at -e line 1, near "0 0"
Execution of -e aborted due to compilation errors.
syntax error at -e line 1, at EOF
Execution of -e aborted due to compilation errors.
Number found where operator expected at (eval 1) line 1, near "* * 50"
(Missing operator before 50?)
epicsThreadOnceOsd epicsMutexLock failed.
status : 0
I am going to execute the following commands
ezcastep -s 0.6 C1:SUS-ETMX_PIT_COMM +-0,50
ezcastep -s 0.6 C1:SUS-ITMX_PIT_COMM +,50
ezcastep -s 0.6 C1:SUS-ETMX_YAW_COMM +,50
ezcastep -s 0.6 C1:SUS-ITMX_YAW_COMM +-0,50
ezcastep -s 0.6 C1:SUS-BS_PIT_COMM +0,50
ezcastep -s 0.6 C1:SUS-BS_YAW_COMM +0,50
hit a key to execute the commands above
|
16973
|
Wed Jul 6 15:28:18 2022 |
Tega | Update | SUS | Output matrix diagonalisation : F2P coil balancing |
git repo : https://git.ligo.org/40m/scripts/-/tree/main/SUS/OutMatCalc
local dir:
/opt/rtcds/caltech/c1/Git/40m/scripts/SUS/OutMatCalc
Here is an update on our recent attempt at diagonalization of the SUS output matrices. There are two different parts to this: the first is coil balancing using existing F2P code which stopped working because of an old-style use of the print function and the second which should now focus on the mixing amongst the various degrees of freedoms (dof) without a DC/AC split I believe. The F2P codes are now working and have been consolidated in the git repo.
TODO:
- The remaining task is to make it so that we only call a single file that combines the characterization code and filter generation code, preferably with the addition of a safety feature that restores any changed values in case of an error or interruption from the user. The safety functionality is already implemented in the output matrix diagonalization stem of the code, so we just need to copy this over.
- Improve the error minimization algorithm for minimizing the cross-coupling between the various dof by adjusting the elements of the output matrix.
Previous work
https://nodus.ligo.caltech.edu:8081/40m/4762
https://nodus.ligo.caltech.edu:8081/40m/4719
https://nodus.ligo.caltech.edu:8081/40m/4688
https://nodus.ligo.caltech.edu:8081/40m/4682
https://nodus.ligo.caltech.edu:8081/40m/4673
https://nodus.ligo.caltech.edu:8081/40m/4327
https://nodus.ligo.caltech.edu:8081/40m/4326
https://nodus.ligo.caltech.edu:8081/40m/4762 |
12536
|
Thu Oct 6 15:42:51 2016 |
Lydia | Update | SUS | Output matrix diagonalization |
Summary: At the 40m meeting yesterday, Eric Q. gave the suggestion that we accept the input matrix weirdness and adjust the output matrix by driving each coil individually so that it refers to the same degrees of freedom. After testing this strategy, I don't think it will work.
Yesterday evening I tested this idea by driving one ITMY coil at a time, and measuring the response of each of the free swing modes at the drive frequency. I followed more or less the same procedure as the standard diagonalization: responses to each of the possible stimuli are compared to build a matrix, which is inverted to describe the responses given the stimuli. For the input matrix, the sensor readings are the responses and the free swing peaks are the stimuli. For the output matrix, the sensors transformed by the diagonalized input matrix as the responses of the dofs which are compared, and the drive frequency peak associated with a coil output is the stimulus. However, the normalization still happens to each dof independently, not to each coil independently.
The output matrix I got had good agreement with the ITMY input matrix in the previous elog: for each dof/osem the elements had the same sign in both input and output matrices, so there are no positive feedback loops. The relative magnitude of the elements also corresponded well within rows of the input matrix. So the input and output matrices, while radically different from the ideal, were consistent with each other and referred to the same dof basis. So, I applied these new matrices (both input and output) to the damping loops to test whether this approach would work.
drive-generated output matrix:
UL UR LR LL SD
pit 1.701 -0.188 -2.000 -0.111 0.452
yaw 0.219 -1.424 0.356 2.000 0.370
pos 1.260 1.097 0.740 0.903 -0.763
sid 0.348 0.511 0.416 0.252 1.000
but 0.988 -1.052 0.978 -0.981 0.060
However, when Gautam attempted to lock the Y arm, we noticed that this change significantly impacted alignment. The alignment biases were adjusted accordingly and the arm was locking. But when the dither was run, the lock was consistently destroyed. This indicates that the dither alignment signals pass through the SUS screen output matrix. If the output matrix pitch and yaw columns refer instead to the free swing eigenmodes, anything that uses the output matrix and attempts to align pitch and yaw will fail. So, the ITMY matrices were restored to their previous values: a close to ideal input matrix and naive output matrix. We could try to change everything that is affected by the output matrices to be independent of a transformation to the free swing dof basis, and then implement this strategy. But to me, that seems like an unneccessary amount of changes with unpredictable consequences in order to fix something that isn't really broken. The damping works fine, maybe even better, when the input matrix is set by the output matrix: we define pitch, for example, to be "The mode of motion produced by a signal to the coils proportional to the pitch row of the naieve output matrix," and the same for the other dofs. Then you can drive one of these "idealized" dofs at a time and measure the sensor responses to find the input matrix. (That is how the input matrix currently in use for ITMY was found, and it seems to work well.)
|
12540
|
Fri Oct 7 20:56:15 2016 |
Koji | Update | SUS | Output matrix diagonalization |
I wanted to see what is the reason to have such large coupling between pitch and yaw motions.
The first test was to check orthogonality of the bias sliders. It was done by monitoring the suspension motion using the green beam.
The Y arm cavity was aligned to the green. The damping of ITMY was all turned off except for SD.
Then ITMY was misaligned by the bias sliders. The ITMY face CCD view shows that the beam is reasonably orthogonally responding to the pitch and yaw sliders.
I also confirmed that the OPLEV signals also showed reasonablly orthogonal responce to the pitch and yaw misalignment.
=> My intuition was that the coils (including the gain balance) are OK for a first approximation.
Then, I started to excite the resonant modes. I agree that it is difficult to excite a pure picth motion with the resonance.
So I wanted to see how the mixing is frequency dependent.
The transfer functions between ITMY_ASCPIT/YAW_EXC to ITMY_OPLEV_PERROR/YERROR were measured.
The attached PDFs basically shows that the transfer functions are basically orthogonal (i.e. pitch exc goes to pitch, yaw exc goes to yaw) except at the resonant frequency.
I think the problem is that the two modes are almost degenerate but not completely. This elog shows that the resonant freq of the ITMY modes are particularly close compared to the other suspensions.
If they are completely degenerate, the motion just obeys our excitation. However, they are slightly split. Therefore, we suffer from the coupled modes of P and Y at the resonant freq.
However, the mirror motion obeys the exitation at the off resonance as these two modes are similar enough.
This means that the problem exists only at the resonant frequencies. If the damping servos have 1/f slope around the resonant freqs (that's the usual case), the antiresonance due to the mode coupling does not cause servo instability thank to the sufficient phase margin.
In conclusion, unfortunately we can't diagnalize the sensors and actuators using the natural modes because our assumption of the mode purity is not valid.
We can leave the pitch/yaw modes undiagnalized or just believe the oplevs as a relatively reliable reference of pitch and yaw and set the output matrix accordingly.
The figures will be rotated later. |
Attachment 1: 161007_P.pdf
|
|
Attachment 2: 161007_Y.pdf
|
|
12354
|
Fri Jul 29 13:17:34 2016 |
Koji | Update | General | Oven |
While the air bake oven situation is being improved, how about to buy a cheepo toaster oven at Target, BestBuy, or anywhere?
We don't need precise temp control for the glue cure test. At LLO I saw that they are using cooking grade oven for this purpose.
(Of course, we should not use this oven for foods once it is used for epoxy)
I have a fryer temp sensor in my office on the freezer stole from the 40m long time ago. You should be able to measure high temp.
If you have such an oven, I'd love to borrow it for the OMC lab later, as I expect to work on epoxy bonding later. |
14073
|
Mon Jul 16 15:07:19 2018 |
Koji | Summary | VAC | Oven C vent |
[Steve Koji]
- Attachment1: Removed the thermal cap. Checked the temperature of the oven. It was totally cold.
- Attachment2: Confirmed the RGA section was isolated. The pumps for the RGA was left running.
- Attachment3: Closed the main valve. The pumps for the main volume was left running.
- Attachment4: Started removing the rid. This did not change the gause readings as they were isolated from the venting main volume.
- Attachment5: Opened the rid. Took the components out on a UHV foil bag. The rid was replaced but loosely held by a few screws with the old gasket, just to protect the frange and the volume from rough dusts. |
Attachment 1: P_20180716_141512.jpg
|
|
Attachment 2: P_20180716_141601.jpg
|
|
Attachment 3: P_20180716_141610.jpg
|
|
Attachment 4: P_20180716_141827.jpg
|
|
Attachment 5: P_20180716_143901.jpg
|
|
15528
|
Sat Aug 15 15:12:22 2020 |
Jon | Configuration | VAC | Overhaul of small turbo pump interlocks |
Summary
Yesterday I completed the switchover of small turbo pump interlocks as proposed in ELOG 15499. This overhaul altogether eliminates the dependency on RS232 readbacks, which had become unreliable (glitchy) in both controllers. In their place, the V4(5) valve-close interlocks are now predicated on an analog controller output whose voltage goes high when the rotation speed is >= 80% of the nominal setpoint. The critical speed is 52.8 krpm for TP2 and 40 krpm for TP3. There already exist hardware interlocks of V4(5) using the same signals, which I have also tested.
Interlock signal
Unlike the TP1 controller, which exposes simple relays whose open/closed states are sensed by Acromags, what the TP2(3) controllers output is an energized 24V signal for controlling such a relay (output circuit pictured below). I hadn't appreciated this difference and it cost me time yesterday. The ultimate solution was to route the signals through a set of new 24V Phoenix Contact relays installed inside the Acromag chassis. However, this required removing the chassis from the rack and bringing it to the electronics bench (rather than doing the work in situ, as I had planned). The relays are mounted to the second DIN rail opposite the Acromags. Each TP2(3) signal controls the state of a relay, which in turn is sensed using an Acromag XT1111.

Signal routing
The TP2(3) "normal-speed" signals are already in use by hardware interlocks of V4(5). Each signal is routed into the main AC relay box, where it controls an "interrupter" relay through which the Acromag control signal for the main V4(5) relay is passed. These signals are now shared with the digital controls system using a passive DB15 Y-splitter. The signal routing is shown below.

Interlock conditions
The new turbo-pump-related interlock conditions and their channel predicates are listed below. The full up-to-date channel list and wiring assignments for c1vac are maintained here.
Channel |
Type |
New? |
Interlock-triggering condition |
C1:Vac-TP1_norm |
BI |
No |
Rotation speed < 90% nominal setpoint (29 krpm) |
C1:Vac-TP1_fail |
BI |
No |
Critical fault occurrence |
C1:Vac-TP1_current |
AI |
No |
Current draw > 4 A |
C1:Vac-TP2_norm |
BI |
Yes |
Rotation speed < 80% nominal setpoint (52.8 krpm) |
C1:Vac-TP3_norm |
BI |
Yes |
Rotation speed < 80% nominal setpoint (40 krpm) |
There are two new channels, both of which provide a binary indication of whether the pump speed is outside its nominal range. I did not have enough 24V relays to also add the C1:Vac-TP2(3)_fail channels listed in ELOG 15499. However, these signals are redundant with the existing interlocks, and the existing serial "Status" readback will already print failure messages to the MEDM screens. All of the TP2(3) serial readback channels remain, which monitor voltage, current, operational status, and temperature. The pump on/off and low-speed mode on/off controls remain implemented with serial signals as well.
The new analog readbacks have been added to the MEDM controls screens, circled below:

Other incidental repairs
- I replaced the (dead) LED monitor at the vac controls console. In the process of finding a replacement, I came across another dead spare monitor as well. Both have been labeled "DEAD" and moved to Jordan's desk for disposal.
- I found the current TP3 Varian V70D controller to be just as glitchy in the analog outputs as well. That likely indicates there is a problem with the microprocessor itself, not just the serial communications card as I thought might be the case. I replaced the controller with the spare unit which was mounted right next to it in the rack [ELOG 13143]. The new unit has not glitched since the time I installed it around 10 pm last night.
|
Attachment 1: small_tp_signal_routing.png
|
|
Attachment 3: small_tp_signal_routing.png
|
|
Attachment 4: medm_screen.png
|
|
273
|
Sat Jan 26 02:34:34 2008 |
Andrey | Update | Computer Scripts / Programs | Overnight Measuremts in XARM |
I am running the program for measuring RMS of peaks in XARM tonight. I just started it, and it will run for about 9 hours until noon on Saturday. Please do not disturb the interferometer. Now the XARM is locked, it should stay locked over the night.
Andrey. |
12179
|
Tue Jun 14 19:37:40 2016 |
jamie | Update | CDS | Overnight daqd test underway |
I'm running another overnight test with new daqd software on fb1. The normal daqd process on fb has been shutdown, and the front ends are sending their signals to fb1.
fb1 is running separate data concentrator (dc) and frame writer (fw) processes, to see if this is a more stable configuration than the all-in-one framebuilder (fb) that we have been trying to run with. I'll report on the test tomorrow. |
2541
|
Fri Jan 22 02:54:06 2010 |
Alberto | Update | ABSL | Overnight measurement |
I'm leaving a measurement running overnight. It should be done in about one hour.
Tomorrow morning, If you need to use the interferometer, and you don't want to have the auxiliary beam going onto the dark port, you can turn down the flipping mirror and close the NPRO's mechanical shutter. |
2543
|
Fri Jan 22 14:40:49 2010 |
Alberto | Update | ABSL | Overnight measurement |
Quote: |
I'm leaving a measurement running overnight. It should be done in about one hour.
Tomorrow morning, If you need to use the interferometer, and you don't want to have the auxiliary beam going onto the dark port, you can turn down the flipping mirror and close the NPRO's mechanical shutter.
|
This is what I measured last night:

This is not a fit. It's just a comparison of the model with the data. |
191
|
Thu Dec 13 23:56:02 2007 |
Andrey | Configuration | Computer Scripts / Programs | Overnight measurements |
After my disease (fever, vomitting, nose problem, overall weakness) I returned to LIGO today for the first time after the weekend, and I am running the script for the XARM-measurements over this night.
So, suspension dumping gains should undergo changes in the interval from 1 to 10 in both ITMX and ETMX.
XARM has been of course locked.
I started running the script for the first time at about 10PM, but I realized after an hour and a half that my step of gain increase 0.2 was too shallow, too small to execute my program during one night. Therefore, I needed to terminate the program, change my program so that it increases the gain with increment 0.5, not 0.2, and started it again around midnight.
Going home.
P.S. The red pump that I borrowed from the lab (Steve's pump?) is back at its previous place. The tire-worker tells me that I absolutely need to change all four tires for almost 500 dollars. I regret a lot about that huge money loss. |
194
|
Mon Dec 17 23:42:08 2007 |
Andrey | Configuration | Computer Scripts / Programs | Overnight measurements in X-arm |
I am making overnight measurements this night (from Monday to Tuesday) in XARM.
The X-arm is now locked, and the values for suspension damping gain will be changed in the interval from 1 to 7 with the step 0.5 in both ITMX and ETMX.
This is the second, repeated measurement. The results of the first measurement from Saturday to Sunday night will be reported in the separate ELOG entry (sorry, I did not make an ELOG entry on Saturday evening about running the program overnight).
The very first attempt to run the script in the night from Thursday to Friday was not successful. |
9909
|
Mon May 5 19:26:43 2014 |
Jenne | Update | LSC | Override ability for whitening triggering |
Today I finally implemented a feature in the whitening triggering that I should have a long time ago: an override button.
Now, on each RFPD's phase rotation screen, there is a button to either allow triggering for that PD (both quads) or to be in manual mode.
If you are allowing triggering, they will behave as they have for the last ~year.....if any degree of freedom is using either quadrant of that PD, and that degree of freedom is triggered, then engage analog whitening and digital de-whitening.
If you chose manual mode, then you can engage or disengage the whitening as you please. (The analog whitening and digital de-whitening are still tied together). |
3344
|
Sun Aug 1 18:50:35 2010 |
Koji | Update | VAC | P1 = 0.48 torr |
I resumed the pumping today. Now the pressure is 0.48 torr. The pump was stopped.
Quote: |
I resumed the pumping from 19:00.
Now the valve RV1 is full open. But the pumping is really slow as we are using only one RP.
After 3hrs of pumping, P1 reached 1.2mmtorr but still we need 2hrs of pumping...
I stopped pumping at 22:30.
|
|
14844
|
Tue Aug 13 08:07:09 2019 |
gautam | Update | CDS | P1--->P2 |
This morning, I wanted to move the existing cables going to the P1 connectors of the iLIGO whitening boards to the P2 connector, to test the modifications made to allow whitening stage switching. Unfortunately, I found that the shrouds werent installed. Where can I find these? |
14845
|
Tue Aug 13 14:36:17 2019 |
gautam | Update | CDS | P1--->P2 |
As it turns out, only one extra shroud needed to be installed - I did this and migrated the cables for the 4 whitening boards from the P1 to P2 connectors. So until the new Acromag box is installed, we have no control over the whitening gains (slow channels), but do still have control over the whitening filter enable/disable (controlled by fast BIO). I am thinking about the easiest way to test the latter - I think the ambient PD dark noise level is too low to be seen above ADC noise even with the whitening enabled, and setting up drive signals to individual channels is too painful - maybe with +45dB of whitening gain, the (z,p) whitening filter shape can be seen with just PD/demod chain electroncis noise.
Quote: |
This morning, I wanted to move the existing cables going to the P1 connectors of the iLIGO whitening boards to the P2 connector, to test the modifications made to allow whitening stage switching. Unfortunately, I found that the shrouds werent installed. Where can I find these?
|
|