ID |
Date |
Author |
Type |
Category |
Subject |
1338
|
Thu Feb 26 00:36:53 2009 |
Yoichi | Summary | Computers | C1:LSC-TRX_OUT broken (and fixed later). |
Today, Kakeru tried to convert C1:LSC-TRX_OUT and C1:LSC-TRY_OUT to DAQ channels.
He edited C1LSC.ini in the chans/daq directory to add the channel but it did not work.
Then he reverted the file back to the original one.
But after we still could not access these channels from dataviewer nor tds tools.
We restarted daqd and tpman on fb40m, but the problem persisted. Even rebooting the whole fb40m did not help.
After inspecting the log file of daqd, it was clear that tpman was failing to create test points for those channels.
I rebooted c1daqawg and then restarted tpman and daqd on fb40m again.
This time, the problem went away. |
1337
|
Wed Feb 25 11:48:02 2009 |
Jenne | Update | PEM | Wiener filtering update - work on filtering some S5 DARM_CTRL data |
Quick update on my wiener filtering status:
Joe has been helping me get on the GRID, so I now have a grid certificate, and accounts on most/all of the clusters.
Joe also helped me get menkar to get S5 data so that I can do wiener filtering to the back-data.
I've been running the wiener filtering algorithm, and right now, it doesn't do anything to improve the DARM_CTRL data. I am confident that this is because something is funky in the wiener filtering algorithm somewhere. The indicator of this is that the wiener filtering calculation takes the same amount of time (~95 seconds) to calculate a filter for 64 seconds of data as for 1 hour of data (both for N = 2000 taps).
For reference, attached are my plots for the wiener filtering result for (1) 64 seconds of S5 data, and for (2) 3600 seconds of S5 data.
These plots were made using H1:DARM_CTRL as the signal to minimize, with 4 seismometers as the witness channels (EX_SEISX, EY_SEISY, LVEA_SEISX, LVEA_SEISY)
I'm working on figuring out what's going on with the filtering algorithm, and why it does work for C1:MC_L minimization, but does not work for H1:DARM_CTRL minimization.
|
Attachment 1: h1_DARM_64s_4seis_25Feb09.png
|
|
Attachment 2: h1_DARM_3600s_4seis_25Feb09.png
|
|
1336
|
Wed Feb 25 03:10:24 2009 |
Yoichi | Update | Locking | Locking status |
Rob, Yoichi, Kakeru, Kiwamu
Tonight, CARM -> MCL hand off was not stable. The MCF signal monotonically went up to +2V after CARM and MCL gain was turned down to zero.
This was repeatable and it only goes up (not down).
After a while, we found that putting sleep (~5sec) between the zeroing of CARM gain and MCL gain prevents this problem.
Handing off of CARM error signal from TR to PODC was also not robust.
It seems that the suitable gain changes every time.
tdsavg started to exit with errors. We rebooted fb40m.
When tdsavg returns an error, the cm_step script tries to write NaN into SPOB DC offset.
To prevent this, I put the tdsavg in a while loop which runs until tdsavg returns something other than NaN.
I was able to hand off to PODC several times, but could not proceed further because the IFO lost lock soon.
|
1335
|
Tue Feb 24 18:42:15 2009 |
pete | Update | Locking | mc board repair |
Peter, Yoichi
Last night:
Quote: | However, when we measured the MCL loop gain with several different AO path gains, the loop shape did not change at all. This led us to suspect the AO path may not be connected. The cabling from the common mode board to the MC board seemed ok. We tested the signal flow in the MC board using a signal generator and an oscilloscope. Then we found that a signal injected to the IN2 (AO path) does not reach to the TP1A (right after the boost stages), though the signal is visible in the OUT2 (monitor BNC right after the initial amplifier (B-amp) for the AO path). The signal from IN1 (MC REFL) can be observed at TP1A. This means something is broken between the B-amp and the sum-amp in the AO path. We will check the MC board tomorrow.
|
Today we examined the MC board. With the extension board in place everything seemed fine. Without the extension board we could replicate the problem. Jiggling the IN2 jack caused the injected signal observed at TP1A to come and go. These jacks are unfortunately mounted directly on the board. We traced the problem to a resistor in this path (R30) which looked fishy. We soldered on a new 2K resistor with OWC and it fixed the problem. |
1334
|
Tue Feb 24 02:23:40 2009 |
Yoichi | Update | Locking | Locking - MC board bad |
Rob, Yoichi, Alberto, Kiwamu, Kakeru
We found that the OMC alignment feedback was on for the POS X loop even though the OMC was not locked.
This caused the PZT mirror to be tilted in yaw a lot. This was probably the reason for the mysterious shift in the AS beam last week, because the AS RF beam is picked up after this PZT mirror.
Rob aligned the OMC and we re-centered the AS PDs and the CCD.
This changed the DARM RF gain, so we changed it from 3 to 1. This gain used to be -1. It is still not understood why the polarity was changed.
The MC length was changed ? We should check the sideband transmission.
After this, we reached to the arm power 4. But the IFO loses lock immediately after the moving zero is turned off.
At this stage, the CARM loop bandwidth is supposed to be high enough that the moving zero is no longer necessary.
However, when we measured the MCL loop gain with several different AO path gains, the loop shape did not change at all.
This led us to suspect the AO path may not be connected. The cabling from the common mode board to the MC board seemed ok.
We tested the signal flow in the MC board using a signal generator and an oscilloscope.
Then we found that a signal injected to the IN2 (AO path) does not reach to the TP1A (right after the boost stages), though the signal is visible in the OUT2 (monitor BNC right after the initial amplifier (B-amp) for the AO path). The signal from IN1 (MC REFL) can be observed at TP1A. This means something is broken between the B-amp and the sum-amp in the AO path.
We will check the MC board tomorrow. |
1333
|
Mon Feb 23 16:42:08 2009 |
josephb | Configuration | Cameras | Camera Beta Testing |
I've setup the GC650 camera (ID 32223) to look at the mode cleaner transmission. I've also added an alias to the camera server and client for this camera.
To use, type: "pserv1 &"on the machine you want to run the server on and "pcam1 &" on the machine you want to actually view the video. At the moment, this only works for the 64-bit Centos 5 machines, Rosalba, Allegra and Ottavia.
Note, you will generally want to start the client first (pcam1 or pcam2) to see if a server is already running somewhere. The server will complain that it can't connect to a camera if it already is in use.
I've setup the GC750 camera (ID 44026) to look at the the right most analog quad TV. This can be run by using "pserv2 &" and "pcam2 &".
If the image stops playing you can try starting and stoping the server to see if will start back up.
You can also try increasing or decreasing the exposure, to see if that helps. The increase and decrease buttons change the exposure by a factor of 2 for each press.
Lastly, the button "Read Epic Channel" reads in the current value from the channel: "C1:PEM-stacis_EEEX_geo" and uses it as the exposure value, in microseconds (in principle 10 to 1000000 should work).
For example, to exposre for 10000 microseconds, use "ezcawrite C1:PEM-stacis_EEEX_geo 10000" and press the "Read Epic Channel" button. |
1332
|
Mon Feb 23 11:07:01 2009 |
steve | Bureaucracy | SAFETY | Kiwamu receives safety training |
Osamu and Kiwamu received 40m safety training on Thursday, Feb 19, 2009
Kiwamu needs Osamu's close supervision at PSL enclosure and AP table
I hope they already read, understood and signed the laser SOP |
1331
|
Sun Feb 22 23:43:07 2009 |
caryn | Summary | General | temperature sensor |
Comparing PSL-FSS-RMTEMP and PEM-MC1-TEMPS
So, to compare temp channels, I made a plot of PSL-FSS_RMTEMP and PEM-MC1_TEMPS(the test temp sensor channel after converting from cts to degC). This plot begins about 2 months ago t_initial=911805130. The temperature channels look kinda similar but MC1-TEMPS (the temp sensor clamped to MC1,3 chamber) is consistently higher in temperature than FSS_RMTEMP. See compare_temperature_channels.png.
MC1-TEMPS isn't exactly consistent with FSS-RMTEMP. I attached a few plots where I've zoomed in on a few hours or a few days. See compare_temperature_channels_zoom1.pdf & compare_temperature_channels_zoom2.pdf
Change the room temperature, see what happens to the chamber temperature
A while ago, somebody was fiddling around with the room temperature. See compare_temperature_channels_zoom4.pdf. This is a plot of PEM-MC1_TEMPS and PSL-FSS_RMTEMP at t0=911805130. You can see the chamber heating up and cooling down in happy-capacitory-fashion. Although, the PSL-FSS_RMTEMP and the PEM-MC1_TEMPS don't really line up so well. Maybe, the air in the location of the MC1,3 chamber is just warmer than the air in the PSL or maybe there's an offset in my calibration equation.
Calibration equation for PEM-MC1-TEMPS
For the calibration (cts to degC) I used the following equation based on the data-sheet for the LM34 and some measurements of the circuit:
TEMPERATURE[degC]=5/9*(((-CTS/16384/451.9/1.04094)-(.0499*10^-3))/(20*10^-6)-35);
How does the chamber temperature compare with the air temperature?
It looks like the chamber may be warmer than the air around it sometimes.
I wanted to check the temperature of the air and compare it with the temperature the sensor had been measuring. So, at t=918855087 gps, I took the temp sensor off of the mc1-mc3 chamber and let it hang freely, close to the chamber but not touching anything. See compare_temperature_chamber_air.png. MC1_TEMPS increases in temperature when I am handling the temp-sensor and then cools down to below the chamber temperature, close to FSS_RMTEMP, indicating the air temperature was less than the chamber temperature.
When, I reattached temp sensor to the chamber at t=919011131 gps, the the temperature of the chamber was again higher than the temperature of the air. See compare_temperature_air2chamber.pdf.
Also, as one might expect, when the temp-sensor is clamped to the chamber, the temperature varies less, & when it's detached from the chamber, the temperature varies more. See compare_temperature_air_1day.pdf & compare_temperature_chamber_1day.pdf.
New temp-sensor power supply vs old temp-sensor power supply
The new temp-sensor is less noisy and seems to work OK. It's not completely consistent with PSL-FSS_RMTEMP, but neither was the old temp-sensor. And even the air just outside the chamber isn't the same temperature as the chamber. So, the channels shouldn't line up perfectly anyways.
I unplugged the 'old' temp-sensor power supply for a few hours and plugged in the 'new' one, which doesn't have a box but has some capacitors and and 2 more voltage regulators. The MC1_TEMPS channel became less noisy. See noisetime.png & noisefreq.pdf. For that time, the minute trend shows that with the old temp-sensor power supply the temp sensor varies +/-30cts and with the new power supply, it is more like +/-5cts (and Volt/16,384cts * 1degF/10mV --> apprx +/-0.03degF). So, it's less noisy.
I kept the new temp-sensor power supply plugged in for about 8 hours, checking if new temp sensor power supply worked ok. Compared it with PSL-FSS_RMTEMP after applying an approximate calibration equation. See ver2_mc1_rmtemp_8hr_appxcal.png.
Just for kicks
Measuring time constant of temp sensor when detached from chamber. At 918858981, I heated up the temp sensor on of the mc1-mc3 chamber with my hand. Took hand off sensor at 918859253 and let it cool down to the room temperature. See temperature_sensor_tau.pdf. |
Attachment 1: compare_temperature_channels.png
|
|
Attachment 2: compare_temperature_channels_zoom1.pdf
|
|
Attachment 3: compare_temperature_channels_zoom2.pdf
|
|
Attachment 4: compare_temperature_channels_zoom4.pdf
|
|
Attachment 5: compare_temperature_chamber_air.png
|
|
Attachment 6: compare_temperature_air2chamber.pdf
|
|
Attachment 7: compare_temperature_air_1day.pdf
|
|
Attachment 8: compare_temperature_chamber_1day.pdf
|
|
Attachment 9: noisetime.pdf
|
|
Attachment 10: noisefreq.pdf
|
|
Attachment 11: ver2_mc1_rmtemp_8hr_appxcal.pdf
|
|
Attachment 12: temperature_sensor_tau.pdf
|
|
1330
|
Fri Feb 20 19:31:16 2009 |
Yoichi | Update | LSC | MICH low gain problem |
Last night, we found that MICH UGF was too low. Even after re-aligning the PDs, it was still too low.
Today, I compared the UGFs of MICH and PRC when in the DRMI configuration locked with the single demod. signals.
In this configuration, MICH signal comes from REFL33Q and the PRC signal comes from REFL33I (the same PD).
The PRC UGF was about 100Hz whereas MICH was only ~10Hz.
Since they uses the same PD, the low gain is not caused by the PD.
I checked conlog history and confirmed there is no change in the MICH->BS path in the last few days.
I also checked the svn history of chans directory for changes in filters. Nothing problematic found.
Then I noticed that the susvme computers were overloaded.
This time, I rebooted all the FE computers just in case.
Then the MICH gain was somewhat recovered (by a factor of 3 or so). Don't know why.
I adjusted the DD_handoff script to set the MICH gain to 0.7 before the bounce-roll filter is engaged.
|
1329
|
Fri Feb 20 03:52:23 2009 |
Yoichi | Update | Locking | Locking Tonight |
Yoichi, Peter
Tonight, we had a problem with the DD hand off.
It failed when the RG filters of MICH for the bounce-roll modes are engaged.
The reason for the failure was that the MICH UGF was too low (~10Hz).
As in the Peter's elog entry, we found that the AS PDs are mis-centered.
Even after we fixed the centering, the MICH UGF was still too low. So we increased the MICH feedback gain by a factor of 10.
The reason for the gain decrease is unknown. It seems almost like the BS coils get weaker.
I checked the UGF of the BS OL loops. These are around 4Hz, so fine. We should check the HWP on the AP table tomorrow.
After the DD hand-off goes ok, the switching of DARM signal from DC to RF failed.
I found that the gain and the polarity of the RF signal were wrong.
AS166 is one of the PDs we found mis-centered (and re-centered). But how can you flip the sign of the signal ?
After this, the cm_step script goes until the activation of the moving zero, but fails when the arm power is increased to 0.7.
Also the ontoMCL script succeeds only 50% of the time. |
1328
|
Fri Feb 20 01:54:18 2009 |
Kakeru | Update | Computer Scripts / Programs | tdsdata might have a bug |
I found a strange jump of value in my data taken with tdsdata.
I couldn't find same jump in a playback of DataViewer, so I think this is a problem of tdsdata.
Be careful when you use tdsdata!
The attached file is an example of jumped data.
I try to get data with allegra and op440m, and both has same kind of jump.
(A downsampling or interpolation may be wrong.)
Rana said there is a fixed version of tdsdata in some PC, but 64bit linux may not have.
I try it tomorrow. |
Attachment 1: jumped_data.png
|
|
1327
|
Thu Feb 19 23:50:31 2009 |
pete | Update | Locking | aligned pd's on AP table |
Yoichi, Peter
While continuing our efforts to lock, we noticed the procedure failed at a point it had gotten past last night: turning on the bounce/roll filters in MICH, PRC, and SRC. We checked the MICH transfer function and noticed that the unity gain point was ~10 Hz, well below the bounce modes. We tried increasing the gain but found saturation, and Rob suggested that there could be misalignment on the AP table, which Steve worked on today. We went out and found two of the PDs (ASDD133 and AS166) to be badly misaligned probably due to a bumped optic upstream. We re-aligned.
|
1326
|
Thu Feb 19 22:40:33 2009 |
Kiwamu | Update | Electronics | PSL angle QPD |
I checked a broken QPD, which was placed for PSL angle monitor, and finally I cocluded one segment of the quadrant diode was broken.
The broken segment has a offset voltage of -0.7V after 1st I-V amplifier. It means the diode segment has a current offset without any injection of light.
Tomorrow I will check a new QPD for replacement.
Kiwamu IZUMI
|
1325
|
Thu Feb 19 16:29:43 2009 |
Yoichi | Update | Computers | Martian wireless router bad |
The Martian wireless router is dead.
I rebooted it several times, but it hangs up in a minute.
I will ask steve to buy a new one. |
1324
|
Thu Feb 19 11:51:56 2009 |
steve | Update | MOPA | HTEMP variation is too much |
The C1:PSL-MOPA_HTEMP variation is more than 0.5 C daily
Normally this temp stays well within 0.1 C
This 80 days plots shows that we have just entered this unstable region some days ago.
The control room temp set unchanged at 70 F, actual temp at ac-mon 69-70 with occasional peaks at 74 F
Water temp at chiller repeatedly around 20.6 C at 8 am
This should be rock solid at 20.00C +- 0.02C
|
Attachment 1: 80dhtemp.jpg
|
|
1323
|
Thu Feb 19 04:16:17 2009 |
Yoichi | Update | LSC | Locking status |
Rob, Yoichi
We checked the CM-MC cross over just before turning off the moving zero.
There was a slight bump in the gain of the MC_L loop at (I believe) the optical spring freq. (~400Hz) just below 0 dB. The phase margin there was very thin.
Removing the moving zero will increase the bump more and make the loop unstable.
Rob suggested to increase the AO gain a bit more.
To see if the AO path is really working, I connected the OUT2 of the MC board to a spare DAQ channel (C1:PEM-OSA_APTEMP).
I confirmed that the PO_DC signal is actually coming to the AO path input of the MC board.
I also hooked up the SR785 to the A excitation channel of the common mode board, so that we can measure the loop gain of the AO path.
After these preparation, the lock acquisition process became somewhat unstable. The ifo loses lock randomly at various places in the lock acquisition steps.
So, as of 4:00 am, I have not gotten a chance to try Rob's suggestion nor the TF measurement with SR785 yet.
I will continue the work tomorrow (i.e. tonight ??).
|
1322
|
Wed Feb 18 21:10:21 2009 |
rana | Update | IOO | MC Drumhead mode lost again |
In early December, Caryn and I noticed that the MC Drumhead mode was visible at the Qmon point of
the MC demod board using a spectrum analyzer and no external excitation of the MC mirrors. We then
started tracking the MC Drumhead modes.
Today I found that it is gone again. It also wasn't there when I looked for it in 2007. 
I looked at the MC error point spectrum and it seemed reasonable. Changing the gains in the MZ, ISS, PMC, & FSS
had no good effect on the noise spectrum.
The voltage noise above 10 kHz in the MC error point is increasing like ~f. I think that this means that
the leftover is the noise from the FSS. Below 10 kHz it is the noise of the VCO (10 mHz/rHz).
One possibility is that the high frequency noise changes with the mood of the NPRO. There should be no
frequency noise induced by the decay of the PA diode power. We can do an NPRO SLOW scan to see if there
is some kind of mode hop noise happening. |
1321
|
Wed Feb 18 21:03:22 2009 |
rana | Update | Cameras | ETMY Camera work not elogged! |
The control room video is showing us a false ETMY image. Who worked on the ETMY camera or video today??!! |
1320
|
Wed Feb 18 19:13:20 2009 |
rana | Configuration | Computers | SVN & MEDM & old medm files |
Allegra had a 2 year old version of SVN installed and CentOS (yum) couldn't upgrade it, so I did an 'svn remove subversion'
and then a 'svn install subversion' to get us up to the Dec '08 version (1.5.5) which is the latest stable.
I also removed all of the old ASS medm directories without backing them up. There's a new RCG script version which is
fixed so that it no longer dumps these old medm directories in there; there's no need since there's already an
medm archive area.
I also removed the medm/old/ directory, did an svn remove, and then copied it back. This is the only way I know of
removing something from the repository without removing it from the working directory. |
1319
|
Wed Feb 18 15:31:38 2009 |
steve | Bureaucracy | SAFETY | emergency back-up lights checked |
All 13 exit and emergency back-up lights and their batteries were checked.
Two of the batteries were replaced. |
1318
|
Wed Feb 18 03:25:25 2009 |
Yoichi | Update | Computers | medm directory back |
I restored the medm directory from the backup on the tape.
The directory had an svn property svn:ignore set and the value of the property included *.snap and *.req.
This resulted in the exclusion of those files from the repository.
I fixed this problem by changing the property of all the directories under /cvs/cds/caltech/medm.
After fixing several other svn problems, the current medm directory contents were checked in to the repository. |
1317
|
Wed Feb 18 03:17:40 2009 |
Yoichi | Update | LSC | Locking |
Yoichi, Kakeru,
Last night, the cm_step script failed at the hand-off of CARM error signal from TR_DC to PO_DC.
This was fixed by reducing the PO_DC gain by a factor of 2.
Currently the script fails when changing C1:LSC-DEMOD_GAIN to zero.
To be honest, I don't fully understand the purpose of this step. |
1316
|
Tue Feb 17 05:20:11 2009 |
Yoichi | Update | LSC | Locking |
Since we excluded *.snap and *.req files from the svn control in the medm directory and these were not restored by the svn co, the burt part of the align/mis-align scripts were not working correctly this evening. So I recreated .req files and cooked up some mis-aligned .snap files.
After some cut-and-try work, I was able to run the dither alignment scripts fine.
Due to the above mentioned delay, the locking work started around midnight.
Tonight, the DD hand-off was not robust. I spent sometime to optimize this.
After the optimization, the locking proceeded to the DC CARM/DARM control state stably.
The CARM->MCL hand-off failed because the LSC-MC offset button was off.
I added a line to turn on the button in the ontoMCL script.
Today, the offloadMCF script worked fine.
Next, the cm_step script stumbled on the "ENGAGERIZING" of the AO path.
I got a hunch that the AO path might not be connected to the MC board.
Indeed, OMC_OSC_FM was connected to the IN2 of the MC board. Looks like it was used for the optimization of the modulation frequencies.
Probably I had the hunch because I did it 
I was able to increase the arm power up to 3.9.
The script failed when it tried to switch the CARM signal from TR_DC to SPOB_DC.
I haven't tackled on this issue yet. |
1315
|
Mon Feb 16 23:09:52 2009 |
rana | Update | Electronics | MC Servo Board offset gone bad! |
The attached plot shows that someone broke the MC_SUM_MON channel around 10:30 AM this past Wednesday the 11th. This is the EPICS monitor of the MC error point.
Come forward now with your confession and I promise that I won't let Steve hurt you. |
Attachment 1: mcoff.png
|
|
1314
|
Mon Feb 16 22:58:51 2009 |
rana, yoichi | Configuration | SUS | Hysteresis in SUS from Misalignments |
WE wondered if there was some hysteresis in the SUS alignments. When we leave the optics misaligned for a
long time it seems to take awhile for the optic to settle down. Possibly, this is the slow deformation of
the wires or the clamps.
The attached PNG shows the plot of the bias sliders for a few days. You can see that we misalign some of the
optics much more than the others. This must be stopped.
Kakeru is going to use his nearly complete optical lever calibrations to quatify this by stepping the optics
around and measuring the effect in the optical lever. Of course, the misalignment steps will be too large to
catch on the OL, but he can calibrate the align-sliders into radians to handle this. |
Attachment 1: a.png
|
|
1313
|
Mon Feb 16 21:49:06 2009 |
Kakeru, Rana | Update | IOO | WFS |
We centerd the input of WFS QPD. |
1312
|
Mon Feb 16 20:47:48 2009 |
rana | Configuration | IOO | Mode Cleaner WFS Loop Gain change |
I found the MCWFS gain slider down at 0.012. In this state the UGFs are probably around 10-30 mHz
and so there is no reduction of seismic noise. It is mainly a DC alignment tool in this state.
We often have reduced the loop gain thusly, to prevent the dreaded "MCWFS eating CM loop gain" disease.
That disease is where there are CM loop instabilities at ~5-30 Hz because of loop cross-couplings
who's exact nature has never been understood (TBI).
Today, I implemented a 4th order, 7 Hz low pass (RLP7) into the loops and turned up the gain by a factor
of 30 to 0.3. In this state, the damping time constants seem to be ~0.5-2 seconds as shown in the first
PDF. I didn't have enough patience to do the interminable swept sine measurements down to 0.1 Hz.
The second PDF shows the Bode plot of the RLP7 filter compared to the pre-existing but unused ELP10.
The third PDF shows my estimate of the OLG TF. This is made by just putting a "Pendulum" filter into the
MCWFS bank and then plotting all the filters together using FOTON. The BLUE curve shows the old TF but
with the new high gain and the RED curve shows the new TF with the new gain.
With this new filter, I bet that we can get away with the higher WFS gain, but if there's any problem during the
handoff, the gain should be reverted to the low value.
In the 4th PDF file, I plot the spectra of 4 of MC2's control signals so that you can see what is bigger than what.
ASCPIT is the one that has the feedback from the WFS's in it. These are all just in units of counts and so to compare
them in some sort of displacement units you have to take into account the pitch moment of inertia, the mirror mass,
and the mis-centering of the beam from the center of rotation of MC2... |
Attachment 1: pmc-pzt-cal.pdf
|
|
Attachment 2: a.pdf
|
|
Attachment 3: olg.pdf
|
|
Attachment 4: mc2.pdf
|
|
1311
|
Mon Feb 16 16:26:29 2009 |
rob | Update | Computers | medm directory wiped on nodus |
Quote: |
Quote: | I accidentally did an 'rm -rf' on the medm directory in nodus, instead of on my laptop as was intended.
I then did an svn checkout. So everything should be current as of the last update, but I am sure that
we have not done a checkin on all of the latest screen enhancements. So...we may have to revert to the
Sunday morning tar to get the latest changes back. |
Indeed, some changes to the medm directory I made were lost.
It was my fault not to check-in those changes.
I asked Alan to restore the directory from the daily rsync backup.
However, the backup job executed this morning have already overwritten the previous (good) backup with the current (bad) medm directory, which Rana restored from the svn. Alan will ask Stuart and Phil if there is still older backup remaining somewhere.
Anyway, I realized that we should stop the backup cron job whenever you think you made a mistake on /cvs/cds/ directory to prevent unwanted overwriting.
The procedure is:
(1) Login to fb40m
(2) Type 'crontab -e'. Emacs will open up in the terminal.
(3) Comment out the backup job (insert # at the beginning of the line containing /cvs/cds/caltech/scripts/backup/rsync.backup ).
(4) Save the file (Ctrl-x Ctrl-s) and exit (Ctrl-x Ctrl-c).
I will post this information on the wiki. |
We should change the rsync script so that it does not delete stuff. Maybe it can keep deleted stuff for 6 months or something. |
1310
|
Mon Feb 16 15:54:07 2009 |
Yoichi | Update | Computers | medm directory wiped on nodus |
Quote: | I accidentally did an 'rm -rf' on the medm directory in nodus, instead of on my laptop as was intended.
I then did an svn checkout. So everything should be current as of the last update, but I am sure that
we have not done a checkin on all of the latest screen enhancements. So...we may have to revert to the
Sunday morning tar to get the latest changes back. |
Indeed, some changes to the medm directory I made were lost.
It was my fault not to check-in those changes.
I asked Alan to restore the directory from the daily rsync backup.
However, the backup job executed this morning have already overwritten the previous (good) backup with the current (bad) medm directory, which Rana restored from the svn. Alan will ask Stuart and Phil if there is still older backup remaining somewhere.
Anyway, I realized that we should stop the backup cron job whenever you think you made a mistake on /cvs/cds/ directory to prevent unwanted overwriting.
The procedure is:
(1) Login to fb40m
(2) Type 'crontab -e'. Emacs will open up in the terminal.
(3) Comment out the backup job (insert # at the beginning of the line containing /cvs/cds/caltech/scripts/backup/rsync.backup ).
(4) Save the file (Ctrl-x Ctrl-s) and exit (Ctrl-x Ctrl-c).
I will post this information on the wiki. |
1309
|
Mon Feb 16 14:12:21 2009 |
Yoichi | Update | LSC | FE system rebooted |
Quote: |
I can't restart the MC autolocker on c1susvme2 because it doesn't let me ssh in. I tried to reboot it a few times but it didn't work. Once you restart it, it becomes inaccessible and doesn't even respond to pinging. Although the controls for the MC mirrors are on.
The mode cleaner stays unlocked. |
MC autolocker runs on op340m, not on c1susvme2.
I restarted it and now MC locks fine.
Before that, I had to reboot c1iool0 and restore the alignment of the MC mirrors (for some reason, burt did not restore the alignment properly, so I used conlog). |
1308
|
Mon Feb 16 10:18:13 2009 |
Alberto | Update | LSC | FE system rebooted |
Quote: |
Quote: |
I didn't get a chance to do much testing since the sus controller (susvme1) went nuts. In retrospect, this could be due to something in the script, so maybe we should try a burt restore to Friday afternoon next time someone wants to look at it. |
I tried the burtrestore today, it didn't work. Also tried some switching of timing cables, and multiple reboots, to no avail. This will require some more debugging. We might try diagnosing the clock driver and fanout modules, the penteks, and we can also try rebooting the whole FE system. |
I rebooted the whole FE system and now c1susvme1 and c1susvme2 are back on.
I can't restart the MC autolocker on c1susvme2 because it doesn't let me ssh in. I tried to reboot it a few times but it didn't work. Once you restart it, it becomes inaccessible and doesn't even respond to pinging. Although the controls for the MC mirrors are on.
The mode cleaner stays unlocked. |
1307
|
Mon Feb 16 00:43:46 2009 |
rana | Update | Computers | medm directory wiped on nodus |
I accidentally did an 'rm -rf' on the medm directory in nodus, instead of on my laptop as was intended.
I then did an svn checkout. So everything should be current as of the last update, but I am sure that
we have not done a checkin on all of the latest screen enhancements. So...we may have to revert to the
Sunday morning tar to get the latest changes back. |
1306
|
Sun Feb 15 15:53:21 2009 |
Rob | Update | LSC | Locking status |
Quote: |
I didn't get a chance to do much testing since the sus controller (susvme1) went nuts. In retrospect, this could be due to something in the script, so maybe we should try a burt restore to Friday afternoon next time someone wants to look at it. |
I tried the burt restore today, it didn't work. Also tried some switching of timing cables, and multiple reboots, to no avail. This will require some more debugging. We might try diagnosing the clock driver and fanout modules, the penteks, and we can also try rebooting the whole FE system. |
1305
|
Sun Feb 15 09:35:00 2009 |
Yoichi | Update | LSC | Locking status |
Quote: |
I found a '$<' in the offloadMCF script. I don't know precisely what that construct means, but I think it caused the script to wait for input when it shouldn't.
|
'$<' acts like 'read' in csh. I might have put it in the offloadMCF script to debug the behavior of the script.
Sorry I probably forgot to remove it from the script when I left. |
1304
|
Sat Feb 14 16:53:26 2009 |
rob | Update | LSC | Locking status |
Quote: | Yoichi, Jenne, Alberto, Rob
Last night, the locking proceeded until the CARM -> MC_L hand-off.
However, the MC_F gets saturated (as expected) and the IFO loses lock soon after the hand-off.
So we need to offload MC_F.
We ran the offloadMCF script, but it did not work, i.e. just waiting for CARM mode.
Looks like an EPICS flag is not set right. |
I found a '$<' in the offloadMCF script. I don't know precisely what that construct means, but I think it caused the script to wait for input when it shouldn't. It probably got in there accidentally. We need to be careful when we're opening scripts just to look at how they work that we don't accidentally change them. I like to use the command 'less' for this purpose.
With this gone, the script worked properly, although the lock didn't last long. I don't know if the next stage in the process is failing or if it's just a bit too noisy in the afternoon. I didn't get a chance to do much testing since the sus controller (susvme1) went nuts. In retrospect, this could be due to something in the script, so maybe we should try a burt restore to Friday afternoon next time someone wants to look at it. |
1303
|
Sat Feb 14 16:15:19 2009 |
rob | Configuration | Computers | c1susvme1 |
c1susvme1 is behaving weirdly. I've restarted it several times but its computation time is hanging out around 260 usec, making it useless for suspension control and locking. I also found a PS/2 keyboard plugged in, which doesn't work, so I unplugged it. It needs to be plugged into a PS/2 keyboard/mouse Y-splitter cable. |
1302
|
Fri Feb 13 16:30:49 2009 |
steve | Configuration | General | status quo is disturbed |
I have been getting ready for the annual safety inspection in the past 2-3 days.
Meaning cleaning up and disturbing the status quo on the floor mostly under the optical tables and their surroundings.
For example: pd power supplies, He/Ne laser ps. and their positions.
BNC cables and ac power line positions can be different.
The new rule: no electronic equipment on the floor.
All electronic equipment were moved-placed into a plastic dish or tray. |
1301
|
Fri Feb 13 13:35:38 2009 |
Yoichi | Update | LSC | Locking status |
Yoichi, Jenne, Alberto, Rob
Last night, the locking proceeded until the CARM -> MC_L hand-off.
However, the MC_F gets saturated (as expected) and the IFO loses lock soon after the hand-off.
So we need to offload MC_F.
We ran the offloadMCF script, but it did not work, i.e. just waiting for CARM mode.
Looks like an EPICS flag is not set right. |
1300
|
Fri Feb 13 08:38:03 2009 |
steve | Update | IOO | MC2 damping restored |
|
1299
|
Thu Feb 12 18:35:10 2009 |
Kakeru | Configuration | PSL | PA current limitter |
I added a PA current limiter.
It is only a voltage devider (composed with 3.09k and 1.02k resiste) between DAC and PA current adjustment input.
The output range of DAC is +/- 10[V] and the conversion factor of PA current adjustment is 0.84[A/V] (measured value), so the PA current adjustment is limited +/- 2.1[A] ( 10[V]*1.02k/(1.02k+3.09k)*0.84[A/V] ).
Actually, the manual of the PA tells that the conversion factor is 0.25[A/V].
There is 3 possibility.
1) There are some mistakes in channels of digital system.
2) The PA manual is wrong.
2-1) The conversion factor of current adjustment is wrong.
2-2) The conversion factor of current monitor is wrong.
I measured the signal of current adjustment and current monitor directly, and confirm that they are consistent to the value monitord from MEDM.
Hence the PA manual must be wrong, but I don't know which factor is wrong (or both?).
If the suspect 2-2) is guilty, it means we adjust PA current with very small range.
This is a completly safety way, but a wast of resource.
Now, the slider to control current adjustment indicate the output of DAC.
I will improve this to indicate current adjustment input, but it takes some time for me to learn about EPICS.
|
1298
|
Thu Feb 12 17:43:33 2009 |
Yoichi | Update | LSC | SRC strangeness solved |
I found the problem with the DRMI lock I had last night was caused by the zero gain in the PD11_I filter.
I don't know how it happened but putting it back to 1.000 made the DRMI lock far more stable and AS166Q got more than 3000.
I also re-centered POY PD to remove the offset in the y-arm loop. The large power drops while y-arm is locked by itself were eliminated. |
1297
|
Thu Feb 12 14:39:07 2009 |
rana | Summary | General | Silicon Beam Dump test |
Yesterday evening, Ken Mailand and I tested the reflectivity of a piece of polished Silicon. Since Silicon has such a high thermalconductivity (compared to stainless and fused silica) and can take much more heat than black glass and should have a very good BRDF and should absorb most of the 1064 nm light if we hit it at Brewster's angle, we want to try it out in the first version high power, low scatterbeam dump. This dump will be a 'V' style dump like what Steve has tested nearly a year ago, but the incoming beam will first hit this piece of Silicon.
The pictures of the setup and the Silicon with the beam hitting it are here.
Brewster's angle for p-pol at 1064 nm is 74.2 deg (n = 3.53 @ 1064 nm). We set up a cube polarizer on the output of the ~1064 nm CrystaLaser. 144 mW got to the Si; the reflected beam was ~1.9-2.0 mW after tuning the angle for minimum power. Via eyeball and protractor it seems that we're at ~74 deg. So the reflectivity is < 1.5-2%. This is good enough; the reflected power will be < 1 W in all cases for eLIGO and that can be absorbed by the rest of the stainless V dump. The 35 W of heat in the silicon will be mostly gotten rid of through conduction into the attached heat sink fins.
This kind of dump would go into places like the PMC-REFL, MC-REFL, & IFO-REFL, where we occasionally need to take high power, but also are sensitive to backscatter. |
1296
|
Thu Feb 12 11:21:54 2009 |
Yoichi | Update | LSC | Locking effort resumed |
Last night, I restarted the locking work.
Quite some time was wasted by the disconnected REFL199 by Alberto for the cavity length measurement.
From now on, please put the interferometer back to the original state every day.
If possible, please refrain from changing the IFO settings (cabling, optics, etc).
It is also very important to always restore the full IFO alignment after you are done with your work.
While I was working on the optimization of the DD hand-off, the DRMI alignment got into a strange state.
Even when I did the whole dither alignment procedure from the beginning (from x-arm), the AS166Q did not go above 1000.
PRMI looks ok (SPOB goes above 1100). I could lock the DRMI but the lock position hops to other modes easily.
Manual tweaks of SRM did not help.
After running the whole alignment procedure several times in vain, I was too tired and went home.
I noticed that the single arm lock shows power drops again. There are some offsets in the arm lock loops.
This may have prevented the Michelson alignment from being optimal. I will check this today. |
1295
|
Wed Feb 11 23:51:53 2009 |
Kakeru | Update | PSL | PA current and laser output |
I attached a plot of ISS monitor PD and MOPA output to PA current.
The both end of PA current (26.0353[A] and 28.4144[A]) correspond to the slider value of -2.0 and 1.0 .
It looks that we must use MOPA with PA current below 27.5[A]. |
Attachment 1: PA_current_output.png
|
|
1294
|
Wed Feb 11 15:01:47 2009 |
josephb | Configuration | Computers | Allegra |
So after having broke Allegra by updating the kernel, I was able to get it running again by copying the xorg.conf.backup file over xorg.conf in /etc/X11. So at this point in time, Allegra is running with generic video drivers, as opposed to the ATI specific and proprietary drivers. |
1293
|
Wed Feb 11 11:39:17 2009 |
steve | Configuration | General | vac envelope ground |
The vacuum envelope was grounded from the IOO chamber to master ground north (under electric breaker panel L)
with 4GA 25ft red battery cable.
|
1292
|
Wed Feb 11 10:52:22 2009 |
Yoichi | Configuration | DAQ | C1:PEM-OSA_APTEMP and C1:PEM-OSA_SPTEMP disconnected |
During the cleanup of the lab. Steve found a box with two BNCs going to the ICS DAQ interface and an unconnected D-SUB on the floor under the AP table. It seemed like a temperature sensor.
The BNCs were connected to C1:PEM-OSA_APTEMP and C1:PEM-OSA_SPTEMP.
Steve removed the box from the floor. These channels can be now used as spare DAQ channels. I labeled those cables. |
1291
|
Wed Feb 11 07:28:25 2009 |
Yoichi | Update | PSL | PA current and laser output |
I think we should also plot the laser power at the MOPA output. The horizontal axis should be the absolute current value read from the PA current monitor channel, not the slider value.
This result is consistent with my hypothesis that the thermal effect is canceling the power change at low frequencies (see elog:1276).
But if it is really caused by thermal effect or not is still unknown.
I'd like to see a larger scan into the lower current region.
Quote: | I changed the PA current and measured laser output power (monitor PD signal).
The gain of ISS is 13dB
Attached figure is the relation of PA current and the average and standard diviation of laser output.
The average of output power decreas as current increase. It looks something is wrong with PA.
When current is -0.125, 0, 0.5, ISS become ocsilating. This looks to be changed from previous measurement.
I wrote matlab code for this measurement. The code is
/cvs/cds/caltech/users/kakeru/scripts/CS_evaluate.m
This function uses
/cvs/cds/caltech/users/kakeru/scripts/moveCS.m |
|
1290
|
Wed Feb 11 00:50:24 2009 |
caryn | Update | General | ants? |
So, near 2 of the trashcans in the control room and underneath a desk there are hundrends of ants. Is this normal? |
1289
|
Tue Feb 10 23:36:25 2009 |
Kakeru | Update | PSL | PA current and laser output |
I changed the PA current and measured laser output power (monitor PD signal).
The gain of ISS is 13dB
Attached figure is the relation of PA current and the average and standard diviation of laser output.
The average of output power decreas as current increase. It looks something is wrong with PA.
When current is -0.125, 0, 0.5, ISS become ocsilating. This looks to be changed from previous measurement.
I wrote matlab code for this measurement. The code is
/cvs/cds/caltech/users/kakeru/scripts/CS_evaluate.m
This function uses
/cvs/cds/caltech/users/kakeru/scripts/moveCS.m |
Attachment 1: PA_current_output.png
|
|