ID |
Date |
Author |
Type |
Category |
Subject |
11757
|
Thu Nov 12 10:22:33 2015 |
Steve | Update | PEM | possible vibration for 4days |
Building: San Pasqual walkway East to West
(Between Holliston & Wilson)
Date: Thursday 11-12-15 to Wednesday 11-18-15
Time: Between 6:00 a.m. and 4:00 p.m. each day
Notification: Possible Noise Vibration
Contact: Ken Lewis (626) 298-2037
* Plumbing contractor will be inspecting and water jetting Storm drains
Type of interruption: (Some vehicle noise and small vibrations limited to close surrounding area)
Areas effected: San Pasqual walkway from Holliston Street to Wilson)
Potential effects: storm drain loss of use
Reason for interruption: Storm drain cleaning in preparation for rainy season
|
11895
|
Mon Dec 21 14:31:41 2015 |
Steve | Update | PEM | Rat traps set |
Two mechanical and two sticdky traps were set to catch univited visitor.
Absolutely no food or food remains into inside garbage cans!!!!!!!!!!!!!!!!!!!!!!!!!
Quote: |
A small rat / large mouse just ran through the control room. Ugh.
|
|
Attachment 1: ratsNC.jpg
|
|
11909
|
Tue Jan 5 09:48:52 2016 |
Steve | Update | PEM | Rat trap moved |
Our janitor confirmed that Q was not hallucinating about this animal. The dropping size indicating a good size one in the IFO room.
One of the mechanical traps moved from the control room to the east arm, close to the " machine shop " door.
I'm going to get more traps.
Quote: |
Two mechanical and two sticdky traps were set to catch univited visitor.
Absolutely no food or food remains into inside garbage cans!!!!!!!!!!!!!!!!!!!!!!!!!
Quote: |
A small rat / large mouse just ran through the control room. Ugh.
|
|
|
11914
|
Wed Jan 6 08:09:04 2016 |
Steve | Update | PEM | 4.5 Mag Banning EQ |
IFO restored after 4.5 Mag Banning, Ca earthquake. |
Attachment 1: 4.5Mag__Banning.png
|
|
11928
|
Tue Jan 12 08:40:06 2016 |
Steve | Update | PEM | Air cond maintenance |
Air condition maintenance is happening. It should be done by 10am |
11949
|
Mon Jan 25 11:06:08 2016 |
Steve | Update | PEM | noisy fan belt replaced |
The noisy fan belt on the roof of the control room was finally replaced on Friday, Jan 22 2016
Quote: |
Air condition maintenance is happening. It should be done by 10am
|
|
11950
|
Tue Jan 26 15:07:47 2016 |
Steve | Update | PEM | accelerometers moved to ETMX |
Objective: measure the noise floor on the optical table and the floor so we can decide if the table needs better anchoring before swapping in
the larger optical table
The accelerometrs labeled as MC1 ( just north east of IOO chamber floor ) and MC2 ( north east leg of MC2 table floor ) were moved:
MC1 to the floor at the north west leg of optical table.
MC2 is in the north east corner of the optical table
Atm2 was taken after table leg bolts were tighed at 40 ft/lb
The spectrum looks similar to ETMY (Guralp for below 3 Hz ) except the Z direction
ETMY . |
Attachment 1: ETMX_floor_vs_optab.png
|
|
Attachment 2: ETMX_floor_vs_optabtop.png
|
|
11952
|
Wed Jan 27 08:17:38 2016 |
Steve | Update | PEM | ETMX floor vs table noise |
Quote: |
Objective: measure the noise floor on the optical table and the floor so we can decide if the table needs better anchoring before swapping in
the larger optical table
The accelerometrs labeled as MC1 ( just north east of IOO chamber floor ) and MC2 ( north east leg of MC2 table floor ) were moved:
MC1 to the floor at the north west leg of optical table.
MC2 is in the north east corner of the optical table
Atm2 was taken after table leg bolts were tighed at 40 ft/lb
The spectrum looks similar to ETMY except the Z direction
ETMY .
|
Conclusion: up to 20 Hz this set up is good. |
Attachment 1: ETMX_X_floor_vs_table.png
|
|
Attachment 2: ETMX_Y_floor_vs_table.png
|
|
Attachment 3: ETMX_Z_floor_vs_table.png
|
|
Attachment 4: MC1@ETMX.jpg
|
|
Attachment 5: MC2@1Winno.jpg
|
|
Attachment 6: MC2@ETMXtable.jpg
|
|
11954
|
Wed Jan 27 20:51:00 2016 |
rana | Update | PEM | ETMX floor vs table noise |
I didn't really appreciate this measurement until just now. IF you can save the DTT .xml file with all the traces in it (i.e. NOT just the plots), we should save this data for comparison plotting later. Perhaps Gautam can post the gzipped xml file for you into the log.
The accelerometers don't read any real noise below ~3 Hz, so we can't judge the difference down low, but this seems like a good measurement in the 5 - 100 Hz band. |
11955
|
Wed Jan 27 23:14:25 2016 |
gautam | Update | PEM | ETMX floor vs table noise |
Quote: |
I didn't really appreciate this measurement until just now. IF you can save the DTT .xml file with all the traces in it (i.e. NOT just the plots), we should save this data for comparison plotting later. Perhaps Gautam can post the gzipped xml file for you into the log.
The accelerometers don't read any real noise below ~3 Hz, so we can't judge the difference down low, but this seems like a good measurement in the 5 - 100 Hz band.
|
Unfortunately I had closed all the DTT windows that Steve had used for the earlier plots. So I took the spectra again - there may be minor differences given that this measurement was taken at ~11pm at night. Anyways, plots and the xml data file are attached.
|
Attachment 1: X_end_ACC_Data.png
|
|
Attachment 2: X_End_ACC_Data.xml.tar.gz
|
11959
|
Fri Jan 29 08:18:13 2016 |
Steve | Update | PEM | PEM |
Quote: |
Air condition maintenance is happening. It should be done by 10am
|
|
Attachment 1: PEM.png
|
|
11965
|
Mon Feb 1 09:16:32 2016 |
Steve | Update | PEM | Rat got cut |
We got it! Traps are removed.
|
Attachment 1: bingo.jpg
|
|
Attachment 2: bingo3.png
|
|
11966
|
Mon Feb 1 14:08:06 2016 |
Steve | Update | PEM | roof condtion is good |
It rained hard yesterday. We have not had this strong downpoor for years. We got 0.7" and the roof did not leak. |
12000
|
Fri Feb 19 15:12:38 2016 |
Koji | Summary | PEM | Guralp Health Check |
I measured the guralp raw outputs and the TFs using the handheld unit and an FFT analyzer.
[Setup]
The handheld unit was connected to each guralp with the same cable which is confirmed t be functional with the Yend Guralp.
The signal for Z, N, and E directions are obtained from the banana connectors on the handheld unit. Each direction has mass, low gain velocity, and high gain velocity output. The PSDs of the signals were measured with an FFT analyzer. The transfer function from the mass signal to the low/high gain signals were also measured for each direction.
The adjustment screw for the E output of the Xend does not work. I had to tilt the Xend Guralp using the leg screws to bring the E signal to zero.
[Result]
Attachment 1: Raw voltage PSD for all outputs
Attachment 2: Comparison of the low gain vel outputs
- All of the mass output show similar PSDs.
- Low gain velocity outputs shows somewhat similar levels. I still need to check if the output is really the ground velocity or not.
- High gain velocity outputs are either not high gain, broken, or not implemented.
- We need to calibrate the low gain output using signal injection, huddle test, or something else.
Attachment 3: TFs between each mass output and the low or high gain outputs
- TFs between the mass signal and the low vel signals show the similar transfer functions between the channels.
- The high gain outputs show low or no transfer function with regard to the mass signals.
|
Attachment 1: Guralp_Raw_PSD.pdf
|
|
Attachment 2: low_vel_comparison.pdf
|
|
Attachment 3: Guralp_TF.pdf
|
|
12031
|
Fri Mar 11 16:52:53 2016 |
Steve | Update | PEM | leaky roof |
Johannes found dripping water at the vac rack. It is safe. It is not catching anything. Actual precipitation was only 0.62" |
Attachment 1: leakyRoof.jpg
|
|
Attachment 2: leakyRoofA.jpg
|
|
12044
|
Wed Mar 23 15:23:12 2016 |
Steve | Summary | PEM | Guralps as connected |
We have one calibration sheet of GUR- B, from 26 June 2008, model CMG-T40-0008, sn T4157 at ETMY east, interface box input 1
I'm looking for calibration paper of GUR- A, model CMG-T40-0053, sn T4Q17 at ETMX south, interface box input 2
Quote: |
I measured the guralp raw outputs and the TFs using the handheld unit and an FFT analyzer.
[Setup]
The handheld unit was connected to each guralp with the same cable which is confirmed t be functional with the Yend Guralp.
The signal for Z, N, and E directions are obtained from the banana connectors on the handheld unit. Each direction has mass, low gain velocity, and high gain velocity output. The PSDs of the signals were measured with an FFT analyzer. The transfer function from the mass signal to the low/high gain signals were also measured for each direction.
The adjustment screw for the E output of the Xend does not work. I had to tilt the Xend Guralp using the leg screws to bring the E signal to zero.
[Result]
Attachment 1: Raw voltage PSD for all outputs
Attachment 2: Comparison of the low gain vel outputs
- All of the mass output show similar PSDs.
- Low gain velocity outputs shows somewhat similar levels. I still need to check if the output is really the ground velocity or not.
- High gain velocity outputs are either not high gain, broken, or not implemented.
- We need to calibrate the low gain output using signal injection, huddle test, or something else.
Attachment 3: TFs between each mass output and the low or high gain outputs
- TFs between the mass signal and the low vel signals show the similar transfer functions between the channels.
- The high gain outputs show low or no transfer function with regard to the mass signals.
|
|
Attachment 1: GUR_A.jpg
|
|
Attachment 2: GUR_B.jpg
|
|
Attachment 3: GUR_Interface_Box.jpg
|
|
12046
|
Thu Mar 24 08:20:52 2016 |
Steve | Update | PEM | Guralp-A calibration sheet |
All Guralp instruments and digitisers are provided with calibration documentation. Should you require a copy of calibration information for any product, email caldoc@guralp.com with the serial number of the product in the subject field and calibration information will be sent to you through email.
See data in the 40m wiki
|
12047
|
Fri Mar 25 19:17:28 2016 |
Nikhil | Update | PEM | Guralp Seismometers |
Calibration of Guralp Seismometers
Objective
- Estimate transfer functions of Guralp A ( near ETMX) and Guralp B ( near ETMY)
- Calibrate the instruments by estimating Velocity Sensitity Parameter
- Convert previously measured Voltage Spectrum to Velocity Spectrum
Instruments Used
- Guralp CMG-40 T Seimometers : Guralp A (Serial Number: T4Q17)
- Guralp CMG-40 T Seimometers : Guralp B (Serial Number: T4157)
- Guralp Handheld Control Unit (HCU)
- FFT Spectrum Analyzer: Model SR785: 2 Channel Dynamic Signal Analyzer
- Oscilloscope: TDS 3014B
- Function Generator: DS 345
Procedure & Results
Sinusoidal current of known frequency and amplitude was injected to the Seismometer calibration coil using signal generator and handheld control unit & corresponding Magnitude and Phase response were recorded. For Guralp B, system response was also estimated with a FFT Spectrum Analyzer.

Frequnecy Range: 0.1 Hz to 45 Hz.
Equivalent Input Velocity was derived from the Input Voltage measurements using the relation: v = V/ (2*pi*f*R*K) , V is the peak to peak Calibration Signal voltage, f is the calibration signal frequency, R is the calibration resistor and K is the feedback coil constant. [See Appendix for R & K values]
Velocity Sensitity at the required frequency is obtained by dividing the Output Response Voltage by the Equivalent Input Velocity.

The obtained Velocity Sensitivity is used to convert the recorded Volatge PSD to Velocity PSD as shown below. The obtained results are compared to gloabl high noise model [NHNM] and USGS New Low Noise Model [NLNM,Peterson 1993] which gives the lowest observed vertical seismic noise levels across the seismic frequency band. Plot legend NLNM shows both the high & low levels.
Guralp A [X Arm] Low Velocity Output

Guralp B [Y Arm] Low Velocity Output

DTT Power Spectrum
Both the Seismometers were connected to the 40 M Control and Data Acquisition System (CDS) and Power Spectrum was estimated for the Vertical, North/South & East/West Channels using Diagnostic Test Tool (DTT) software.

Comments
- The transfer function from Guralp A [ETMX] looks similar to that of Guralp B [ETMY] in both magnitude and phase but with a lower gain.
- Velocity Sensitivity of Guralp A is comparable to the value provided in the Calibration Data Sheet [~ 400] for all the channels [Vertical, North/South, East/West] after 1 Hz. For Guralp B, Velocity Sensitivity is a factor of 2.5 higher [all channels] than the specification [~ 400] after 1 HZ.Below 1 Hz Sensitivity drops down for both sensors. I am not ruling out a missing common factor in the calculation, but anyway, test shows that Guralp B has ~2.5 times better Velocity Sensitivity than Guralp A.
- The Calibrated Seismic Velocity Spectrum for Guralp B is within the Globally Observed High and Low Noise Seismic Spectrum while Guralp A's Spectrum is more noisier above 1 Hz [Anthropogenic Activity normally contributes the most in 1 Hz to 10 Hz frequency band].
- Concurrently acquired Power Spectrum using DTT [Diagnostic Test Tools] shows that Guralp A Spectrum behaves rather strangely. The system response seems to be completely different from the one we obtained locally using signal generator. While Guralp B functionality seems normal. One reason for this erratic beahvior might be faulty cables used for data acquisition from Guralp A. This needs to be verified.
Appendix
CMG-40T Guralp A Calibration Sheet
|
Velocity Output: V/m/s (Differential) |
Mass Position Output (Acceleration Output) : V/m/s^2 |
Feedback Coil Constant : Amp/m/s^2 |
Vertical |
2 x 400 |
19 |
0.00397 |
North/South |
2 x 398 |
23 |
0.00486 |
East/West |
2 x 401 |
23 |
0.00479 |
Calibration Resistor: 51000
CMG-40T Guralp B Calibration Sheet
|
Velocity Output: V/m/s (Differential) |
Mass Position Output (Acceleration Output) : V/m/s^2 |
Feedback Coil Constant : Amp/m/s^2 |
Vertical |
2 x 401 |
19 |
0.00408 |
North/South |
2 x 400 |
20 |
0.00421 |
East/West |
2 x 404 |
22 |
0.00466
|
Calibration Resistor: 51000 |
Attachment 3: VelSens_XArm_Guralp_A.png
|
|
Attachment 4: VelSens_XArm_Guralp_A.png
|
|
Attachment 5: VelSens_YArm_Guralp_B.png
|
|
Attachment 8: Vel_PSD_XArm_GurB_E.png
|
|
Attachment 9: Vel_PSD_XArm_GurB_N.png
|
|
Attachment 16: Guralp_PowerSpectrum.pdf
|
|
12048
|
Fri Mar 25 23:54:04 2016 |
rana | Update | PEM | Guralp Seismometers |
Something seems not right. The Guralp response should be flat in velocity from 0.05-30 Hz. Why is there any feature at 1 Hz? Saturation of some kind? |
12084
|
Wed Apr 20 14:36:45 2016 |
Steve | Update | PEM | dusty lab |
It is worth wiping table top covers. Use isopropanol soaked lint free wipes.
|
Attachment 1: AP_accm.jpg
|
|
Attachment 2: wiped_AP.jpg
|
|
Attachment 3: temp-partic_6months.png
|
|
Attachment 4: 3_years_PEM.png
|
|
12087
|
Fri Apr 22 13:58:13 2016 |
Steve | Update | PEM | leaky roof is fixed |
Dan sealed the leak today.
|
Attachment 1: leakyRoof_(2).jpg
|
|
12096
|
Thu Apr 28 08:49:47 2016 |
Steve | Update | PEM | possible noise sources schedule |
Building: Campus (see attached Map)
Date: Manhole 1 - May 3 through May 5
Manhole 2 – May 6 through May 10
Manhole 2 - May 16 through May 19
Manhole 3 – May 11 through May 19
Time: Noise: 7:00 a.m. To 5:00 p.m.
Access: 24 Hours a day
Interruption: Noise/Vehicular & Pedestrian Access
Storm Drain Manholes
*In order to repair 3 manholes associated with a large storm drain that runs north-south through the campus, work will take place at the
3 manholes shown on the map. This work will interrupt vehicular and pedestrian access on the paths adjacent to the manholes. Though the work at Manholes 1 and 2 will allow vehicular and pedestrian access around the manholes, the work at Manhole 3 will completely block the driveway running south from the southeast corner of Parking Lot 11. Noise will also be created by the repair
|
Attachment 1: Campus_B&W_Map-2.pdf
|
|
12209
|
Tue Jun 21 14:12:10 2016 |
Steve | Update | PEM | vacuum envelope wiped |
I wiped down the cranes with wet towel and Mario our janitor did the chamber tops with the tubes.
The optical tabels were not touched.
The outside temp peaked at 44 C yesterday
|
Attachment 1: 60dPEM.png
|
|
12318
|
Thu Jul 21 08:24:20 2016 |
Steve | Update | PEM | ants alart |
The heat drives the ants to the lab! Make sure the light doors are tight on the chambers.
|
Attachment 1: ants_Alart.jpg
|
|
12365
|
Wed Aug 3 14:52:37 2016 |
Steve | Summary | PEM | Guralps as connected |
Guralps as connected with pictures |
12383
|
Mon Aug 8 10:35:12 2016 |
Steve | Update | PEM | vent 78 day 38 |
Quote: |
RGA background scan
Quote: |
Vacuum Status: Chamber Open
All chamber annuloses are vented. Vac Monitor screen is not communicating with gauges. The valve position indicator are working.
RGA is pumped by Maglev through VM2
|
|
|
Attachment 1: Vent38days.png
|
|
Attachment 2: vent78d40RgaBg.png
|
|
12406
|
Fri Aug 12 21:26:28 2016 |
rana | Update | PEM | Mic Self Noise |
Seems to good to be true. Maybe you're over fitting? Please put all the traces on one plot and let us know how you do the parameter setting. You should use half the data for training the filter and the second half for doing the subtraction. |
12408
|
Mon Aug 15 12:23:56 2016 |
Praful | Update | PEM | Mic Self Noise |
I didn't have a separate training set and data set, so I think that's why the graphs came out looking too good. The units on the graphs are also incorrect, I was interpreting PSD as ASD. I haven't been able to get my Wiener filtering code working well- I get unreasonable subtractions like the noise being larger than the unfiltered signal, so Eric showed me this frequency-dependent calculation described here: https://dcc.ligo.org/LIGO-P990002
This seems to be working well so far:
freq1.pdf

freq2.pdf

freq3.pdf

Here's all the plots on one figure:
frequency_dependent.pdf

Let me know if this looks believable.
Quote: |
Seems to good to be true. Maybe you're over fitting? Please put all the traces on one plot and let us know how you do the parameter setting. You should use half the data for training the filter and the second half for doing the subtraction.
|
|
Attachment 1: freq1.png
|
|
Attachment 2: freq1.pdf
|
|
Attachment 4: freq2.pdf
|
|
Attachment 6: freq3.pdf
|
|
Attachment 8: frequency_dependent.pdf
|
|
12429
|
Mon Aug 22 16:33:32 2016 |
Steve | Update | PEM | the lab temp is up |
The IFO room temp is up a bit and it is coming down. The out side temp is not really high. |
Attachment 1: temp.png
|
|
12431
|
Mon Aug 22 18:35:16 2016 |
Praful | Update | PEM | the lab temp is up |
The temperature is decreasing slowly but is still above 24 C.

temp_plot.png
Quote: |
The IFO room temp is up a bit and it is coming down. The out side temp is not really high.
|
|
Attachment 1: temp_plot.png
|
|
Attachment 3: temp_plot.png
|
|
12498
|
Fri Sep 16 14:15:28 2016 |
steve | Update | PEM | partical counts |
South end flow bench and both clean room assembly flow benches measured zero counts for 0.3 and 0.5 micron size particales.
The counting efficiency of 0.5 micron is 100%
location
|
0.3 micron particles / cf min
|
0.5 micron particles / cf min
|
counter
MET
ONE
|
effect |
ITMY table |
35,000 |
3,000 |
#3 |
|
ETMY table |
35,000 |
3,000 |
#3 |
|
ITMX table |
24,000 |
2,400 |
#3 |
PSL HEPA at 50V |
ITMX wall mounted |
- |
2,500 |
#1 |
|
ETMX work bench |
5,000 |
600 |
#3 |
flow bench on |
The PSL HEPA performance was measured at the center of the table with MET ONE #3
Voltage Variac |
particles /cf min |
particles /cf min |
particle size |
0.3 micron |
0.5 micron |
60 |
0 |
0 |
50 |
10 |
0 |
40 |
1,400 |
110 |
20 |
1,400 |
100 |
|
Attachment 1: 80days.png
|
|
12538
|
Fri Oct 7 15:36:03 2016 |
Steve | Update | PEM | the lab temp is up |
The IFO room temp is up
Quote: |
The IFO room temp is up a bit and it is coming down. The out side temp is not really high.
|
|
Attachment 1: tempUp.png
|
|
Attachment 2: PEM4d.png
|
|
12558
|
Thu Oct 13 14:49:57 2016 |
Koji | Configuration | PEM | XLR(F)-XLR(M) cable took from the fibox to the Blue microphone |
[Gautam Koji]
XLR(F)-XLR(M) cable for the blue microphone is missing. Steve ordered one.
We found one in the fibox setup. As we don't use it during the vent, we use this cable for the microphone.
Once we get the new one, it will go to the fibox setup.
|
12582
|
Thu Oct 27 09:38:32 2016 |
Steve | Update | PEM | mouse |
We may have a mouse in the lab. Do not leave any food scrap in trash ! Traps will be set.
|
Attachment 1: mouse.jpg
|
|
12605
|
Tue Nov 8 08:51:59 2016 |
Steve | Update | PEM | mouse hole sealed |
This is where the mammal came though. It is reach able from room 108 CES
Quote: |
We may have a mouse in the lab. Do not leave any food scrap in trash ! Traps will be set.
|
|
Attachment 1: CES108rs.jpg
|
|
12629
|
Mon Nov 21 11:30:01 2016 |
Steve | Update | PEM | rat is cut and removed |
Last jump at rack Y2. |
Attachment 1: rat#2.png
|
|
12656
|
Fri Dec 2 10:57:08 2016 |
steve | Update | PEM | door seal replaced |
The south end door leaky weather seals replaced.
The aim is here to get some overpressure inside / outside so the lab partical count would not depend on outside condition.
|
Attachment 1: 1_year_PEM.png
|
|
12673
|
Thu Dec 8 07:56:05 2016 |
Steve | Update | PEM | EQ6.5m Northen CA |
No damage. ITMY is glitching, so it has not been damped.
|
Attachment 1: eq6.5FerndaleCA.png
|
|
Attachment 2: 16d_glitching-_trend.png
|
|
Attachment 3: EQ6.5_&4.7mFerndaleCa.png
|
|
12687
|
Thu Dec 29 10:24:56 2016 |
Steve | Update | PEM | EQ5.7mHawthorn NV |
Sus damping restored.
|
Attachment 1: eq5.7HawthorneNV.png
|
|
12696
|
Mon Jan 9 09:18:47 2017 |
Steve | Update | PEM | power glitch |
There was a power glitch last night around 1:15am
The vacuum was not effected.
PSL laser turned on, PMC locked, PSL shutter opened and MC locked.
IR lasers at the ends turned on.
East arm air cond turned on.
The computers are all done.
The last power glitch was at Nov 3, 2016
|
Attachment 1: MondayMorning.png
|
|
12711
|
Fri Jan 13 10:53:03 2017 |
Steve | Update | PEM | doors are fixed |
Control room to outside door was realigned.
It is self closing now.
Control room to IFO door lock optimized to soft closing.
All other doors lubricated by Alex of the key shop. |
12712
|
Fri Jan 13 14:18:28 2017 |
Steve | Update | PEM | air condition fixed |
The old control room AC has been stick in heating mode for about 2 months. It's thermostate and fan belt was finally replaced. It was calibrated and set to 71 F ( just behind 1X6 on west wall ) around 1pm.
Out belt; sad inside
at 4 pm Rana cried
It must be too tight. |
Attachment 1: PEM_120d.png
|
|
12743
|
Fri Jan 20 14:42:12 2017 |
Steve | Update | PEM | Caltech weather station |
We should be able to connect to this station |
12744
|
Fri Jan 20 17:12:37 2017 |
Steve | Update | PEM | EM172 mic is hooked up in the PSL |
Gautam and Steve,
It is hanging in the midle of the PSL enclosure. Wired to 1X1 to get plus and minus 15V through fuse. It's output is connected to FB c17 input.
GV: C17 corresponds to "MIC 1" in the PEM model. So the output is saved as "C1:PEM-MIC_1_OUT_DQ"
Quote: |
I added an EM172 to my soldered circuit and it seems to be working so far. I have taken a spectra using the EM172 in ambient noise in the control room as well as in white noise from Audacity. My computer's speakers are not very good so the white noise results aren't great but this was mainly to confirm that the microphone is actually working.

white_v_ambient.pdf
|
|
Attachment 1: EM172c.jpg
|
|
12789
|
Thu Feb 2 17:34:25 2017 |
rana | Update | PEM | EM172 mic is hooked up in the PSL |
I don't know if anyone looked at the time series (not trend) or spectrum of the Microphone after installation, but it looks bad and featureless to me. Is the Microphone broken?
This shows the spectrum from early this morning and again from tonight. You can see that it is bi-stable in its noise properties. This thing is busted; we're now removing it from the PSL so that it doesn' light it self on fire.

|
Attachment 1: MicFail.pdf
|
|
12790
|
Thu Feb 2 17:43:20 2017 |
gautam | Update | PEM | EM172 mic is hooked up in the PSL |
I had noticed something wonky with the microphone, but neglected to elog it. I had tested it after installation by playing a sine wave from my laptop and looking at the signal on the PSL table, it worked fine. But you can see in the attached minute trend plot that the signal characteristics changed abruptly ~half a day after installation, and never quite recovered.\
|
Attachment 1: Mic_broken.png
|
|
12836
|
Fri Feb 17 10:56:12 2017 |
steve | Update | PEM | particle counter moved into PSL enclousure |
The MET#1 particle counter was moved from CES wall at ITMX to PSL enclousure south west corner at 11am.
The HEPA filter speed at the Variac was turned down to 20V from 40
This counter pumps air for 1 minute in every 20 minutes. Soft foam in bags used to minimize this shaking as it is clamped.
|
Attachment 1: from_here.jpg
|
|
Attachment 2: to_here.jpg
|
|
Attachment 3: PSL_particles.png
|
|
12841
|
Tue Feb 21 10:08:35 2017 |
steve | Update | PEM | noisy morning |
Our new janitor Francisco is started working in IFO room today.
Quote: |
The MET#1 particle counter was moved from CES wall at ITMX to PSL enclousure south west corner at 11am.
The HEPA filter speed at the Variac was turned down to 20V from 40
This counter pumps air for 1 minute in every 20 minutes. Soft foam in bags used to minimize this shaking as it is clamped.
|
Large film crews are working just out side the north west corner of the lab. They started around ~ 5:30am Do not plan on working late tonight.
ETMX sus damping restored.
C1:PSL-FSS_RMTEMP, C1:PSL-PMC_PMCTRANSPD and C1:PEM-count_temp channels are not reading since Friday
|
Attachment 1: outside_activity.png
|
|
12844
|
Wed Feb 22 08:54:17 2017 |
steve | Update | PEM | PSL enclousure particle count |
ETMX sus damping recovered. PSL enclousure is dusty at 20V rotation speed. Rainy days as outside condition.
Quote: |
The MET#1 particle counter was moved from CES wall at ITMX to PSL enclousure south west corner at 11am.
The HEPA filter speed at the Variac was turned down to 20V from 40
This counter pumps air for 1 minute in every 20 minutes. Soft foam in bags used to minimize this shaking as it is clamped.
|
|
Attachment 1: dusty__PSL.png
|
|
12855
|
Tue Feb 28 08:04:48 2017 |
steve | Update | PEM | ETMX damping recovered |
ETMX sus damping recovered.
Note: The giant metal garbage container was moved from the south west corner of CES months ago.
Quote: |
ETMX sus damping recovered. PSL enclousure is dusty at 20V rotation speed. Rainy days as outside condition.
|
|
Attachment 1: ETMX.png
|
|