40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 225 of 339  Not logged in ELOG logo
IDup Date Author Type Category Subject
  11273   Tue May 5 10:40:05 2015 ericqHowToComputer Scripts / ProgramsHow to get a web page running on Nodus

How to get your own web page running on Nodus

  1. On any martian machine, put your stuff in /users/public_html/$MYPAGE/
  2. On Nodus, run: ln -s /users/public_html/$MYPAGE /export/home/
  3. Your site is now available at https://nodus.ligo.caltech.edu:30889/$MYPAGE/
  4. If you want to allow straight up directory listing to the entire internet, on Nodus run: sudoedit /etc/sites-available/nodus, and add the following lines towards the bottom:
<Directory /export/home/$MYPAGE>
    Options +Indexes
</Directory>
  11274   Tue May 5 16:02:57 2015 SteveUpdateVACVac Summery Channels with discription

As it was requested by the Bos.

It would be nice to read from the epic screen C1:Vac-state_mon.......Current State: Vacuum Normal, valve configuration

Quote:

       Channel        

                    Function                                              Description                                             Interlock         
             

C1:Vac-P1_pressure   

 Main volume of 40m interferro meter        P=Pirani gauge, Pressure range: ATM-760  to 1e-4 Torr at 3 mT close V1 and PSL shutter
C1:Vac-P2_pressure  Maglev foreline pressure

Maglev is the main pump of our vacuum system below 500 mTorr

It's long term pressure has to be <500 mTorr                  

 at 6 Torr close V1 
C1:Vac-P3_pressure  annuloses

 Each chamber has it's own annulos. These small volumes are indipendent from main volume.     Their  pressure  ranges are <5 mTorr at vac. normal valve configuration.

                                          
C1:Vac-CC1_pressure  IFO main volume

CC1=cold cathode gauge (low emmision), Pressure range: 1e-4 to 1e-10 Torr,

In vac- normal configuration CC1= 2e-6 Torr

at 1e-5 Torr close VM1
C1:Vac-CC4_pressure  RGA  pressure In vac-normal configuration CC1=CC4  
 C1:Vac-N2pres  valve's drive pneumatic    

The N2 supply is regulated to 60-80 PSI out put at the auto cylinder changer.

 at 55 PSI close V1, at 45 PSI close all 
 It  does not exist yet 2 N2 cylinder sum pressure

Each cylinder pressure will be measured before the regulator and summed for warning message to be send

at 1000 PSI

 

  11275   Fri May 8 08:16:46 2015 SteveUpdateLSCdrift in Y arm

Why is the Y arm drifting so much?

The " PSL FSS Slow Actuator Adjust " was brought back to range from 1.5 to 0.3 yesterday as ususual. Nothing else was touched.

I'm not sure if the timing scale is working correctly on theses summery plots. What is the definition of today?

The y-arm became much better as I noticed it at 5pm

 

Attachment 1: 1goodday.png
1goodday.png
Attachment 2: Yishappy.png
Yishappy.png
  11276   Fri May 8 14:30:09 2015 SteveUpdateVACCC1 cold cathode gauges are baked now

CC1s  are not reading any longer. It is an attempt to clean them over the weekend at 85C

These brand new gauges "10421002" sn 11823-vertical, sn 11837 horizontal replaced 11 years old 421 http://nsei.missouri.edu/manuals/hps-mks/421%20Cold%20Cathode%20Ionization%20Guage.pdf  on 09-06-2012 http://nodus.ligo.caltech.edu:8080/40m/7441

Quote:

 

We have two cold cathode gauges at the pump spool and one  signal cable to controller. CC1  in horizontal position and CC1 in vertical position.  

CC1 h started not reading so I moved cable over to CC1 v

 

Attachment 1: cc1bake95C.jpg
cc1bake95C.jpg
  11277   Sun May 10 13:54:41 2015 ranaHowToComputer Scripts / Programssummary page URL change

Also, EQ gave us a better (and not pwd protected) URL for the summary pages. Please replace your previous links with this new one:

https://nodus.ligo.caltech.edu:30889/detcharsummary/

  11278   Mon May 11 01:28:33 2015 ranaHowToComputer Scripts / Programssummary page URL change

Like Steve pointed out, the summary pages show that the y-arm transmission drifts a lot when locked. The OL summary page shows that this is all due to ITMY yaw.

Could be either that they coil driver / DAC is bad or that the suspension is poorly built. We need to dig into ITMY OL trends over long term to see if this is new or now.

Also, weather station needs a reboot. And does anyone know what the MC_F calibration is?

  11279   Mon May 11 12:17:19 2015 max isiHowToGeneralSummary pages

I have created a wiki page with introductory info about the summary page configuration: https://wiki-40m.ligo.caltech.edu/Daily summary help

We can also use that to collect tips for editing the configuration files, etc.

Quote:

I have set up new summary pages for the 40m: http://www.ligo.caltech.edu/~misi/summary/
This website shows plots (time series, spectra, spectrograms, Rayleigh statistics) of relevant channels and is updated with new data every 30 min.

The content and structure of the pages is determined by configuration files stored in nodus:/users/public_html/gwsumm-ini/ . The code looks at all files in that directory matching c1*.ini. You can look at the c1hoft.ini file to see how this works. Besides, a quick guide to the format can be found here http://www.ligo.caltech.edu/~misi/iniguide.pdf

Please look at the pages and edit the config files to make them useful to you. The files are under version control, so don’t worry about breaking anything.

Do let me know if you have any questions (or leave a comment in the pages).

 

  11280   Mon May 11 13:21:25 2015 manasaUpdateCDSc1lsp and c1sup not running

I found the c1lsp and c1sup models not running anymore on c1lsc (white blocks for status lights on medm).

To fix this, I ssh'd into c1lsc. c1lsc status did not show c1lsp and c1sup models running on it.

I tried the usual rtcds restart <model name> for both and that returned error "Cannot start/stop model 'c1XXX' on host c1lsc".

I also tried rtcds restart all on c1lsc, but that has NOT brought back the models alive.

Does anyone know how I can fix this??

c1sup runs some the suspension controls. So I am afraid that the drift and frequent unlocking of the arms we see might be related to this.

 

P.S. We might also want to add the FE status channels to the summary pages.

  11281   Mon May 11 13:26:02 2015 manasaUpdateIMCMC_F calibration

The last MC_F calibration was done by Ayaka : Elog 7823

Quote:

And does anyone know what the MC_F calibration is?

 

  11282   Mon May 11 14:08:19 2015 manasaUpdateCDSc1lsp and c1sup removed?

I just found out that c1lsp and c1sup models no more exist on the FE status medm screens. I am assuming some changes were done to the models as well.

Earlier today, I was looking at some of the old medm screens running on Donatella that did not reflect this modification. 

Did I miss any elogs about this or was this change not elogged??

Quote:

I found the c1lsp and c1sup models not running anymore on c1lsc (white blocks for status lights on medm).

To fix this, I ssh'd into c1lsc. c1lsc status did not show c1lsp and c1sup models running on it.

I tried the usual rtcds restart <model name> for both and that returned error "Cannot start/stop model 'c1XXX' on host c1lsc".

I also tried rtcds restart all on c1lsc, but that has NOT brought back the models alive.

Does anyone know how I can fix this??

c1sup runs some the suspension controls. So I am afraid that the drift and frequent unlocking of the arms we see might be related to this.

 

P.S. We might also want to add the FE status channels to the summary pages.

 

  11283   Mon May 11 15:15:12 2015 manasaUpdateGeneralRan ASS for arms

Arm powers had drifted to ~ 0.5 in transmission.

X and Y arms were locked and ASS'd to bring the arm transmission powers to ~1.

  11284   Mon May 11 18:14:52 2015 ranaUpdateIMCMC_F calibration

I saw that entry, but it doesn't state what the calibration is in units of Hz/counts. It just gives the final calibrated spectrum.

  11285   Tue May 12 08:51:08 2015 ericqUpdateCDSc1lsp and c1sup removed?
Quote:

was this change not elogged??

This is my sin.

Back in Febuary (around the 25th) I modified c1sus.mdl, removing the simulated plant connections we weren't using from c1lsp and c1sup. This was included in the model's svn log, but not elogged. blush

The models don't start with the rtcds restart shortcut, because I removed them from the c1lsc line in FB:/diskless/root/etc/rtsystab (or c1lsc:/etc/rtsystab). There is a commented out line in there that can be uncommented to restore them to the list of models c1lsc is allowed to run. 

However, I wouldn't suspect that the models not running should affect the suspension drift, since the connections from them to c1sus have been removed. If we still have trends from early February, we could look and see if the drift was happening before I made this change. 

  11286   Tue May 12 12:04:41 2015 manasaUpdateGeneralSome maintenance

* Relocked IMC. I guess it was stuck somewhere in the autlocker loop. I disabled autolocker and locked it manually. Autolocker has been reenabled and seems to be running just fine.

* The X arm has been having trouble staying locked. There seemed to be some amount of gain peaking. I reduced the gain from 0.007 to 0.006.

*  I disabled the triggered BounceRG filter : FM8 in the Xarm filter module.  We already have a triggered Bounce filter: FM6 that takes care of the noise at bounce/roll frequencies. FM8 was just adding too much gain at 16.5Hz. Once this filter was disabled the X arm lock has been much more stable. 
Also, the Y arm doesn't use FM8 for locking either.

 

  11287   Tue May 12 14:57:52 2015 SteveUpdateVACCC1 cold cathode gauges are baked now

Baking both CC1 at 85 C for 60 hrs did not help.

The temperature is increased to 125 C and it is being repeated.

Quote:

CC1s  are not reading any longer. It is an attempt to clean them over the weekend at 85C

These brand new gauges "10421002" sn 11823-vertical, sn 11837 horizontal replaced 11 years old 421 http://nsei.missouri.edu/manuals/hps-mks/421%20Cold%20Cathode%20Ionization%20Guage.pdf  on 09-06-2012 http://nodus.ligo.caltech.edu:8080/40m/7441

Quote:

 

We have two cold cathode gauges at the pump spool and one  signal cable to controller. CC1  in horizontal position and CC1 in vertical position.  

CC1 h started not reading so I moved cable over to CC1 v

 

 

  11288   Wed May 13 09:17:28 2015 ranaUpdateComputer Scripts / Programsrsync frames to LDAS cluster

Still seems to be running without causing FB issues. One thought is that we could look through the FB status channel trends and see if there is some excess of FB problems at 10 min after the hour to see if its causing problems.

I also looked into our minute trend situation. Looks like the files are comrpessed and have checksum enabled. The size changes sometimes, but its roughly 35 MB per hour. So 840 MB per day.

According to the wiper.pl script, its trying to keep the minute-trend directory to below some fixed fraction of the total /frames disk. The comment in the scripts says 0.005%,

but I'm dubious since that's only 13TB*5e-5 = 600 MB, and that would only keep us for a day. Maybe the comment should read 0.5% instead...

Quote:

The rsync job to sync our frames over to the cluster has been on a 20 MB/s BW limit for awhile now.

Dan Kozak has now set up a cronjob to do this at 10 min after the hour, every hour. Let's see how this goes.

You can find the script and its logfile name by doing 'crontab -l' on nodus.

 

  11289   Wed May 13 10:07:36 2015 ranaFrogsPEMGuralp breakout paddle

Reward being offered for the safe return of this thing:

  11290   Wed May 13 13:33:34 2015 SteveFrogsPEMGuralp breakout box recovered

COD_Sugar napolion is due to Steve:  Item delivered, model CMG-SCU-0013, sn G9536

Quote:

Reward being offered for the safe return of this thing:

 

  11291   Thu May 14 17:41:10 2015 ranaUpdatePEMweather station and Guralp maintenance

Today Steve and I tried to recenter the Guralps. The breakout box technique didn't work for us, so we just turned the leveling screws until we got the mass position outputs within +/-50 mV for all DoF as read out by the breakout box.

Some points:

  1. GUR1 is at the ETMY (E/W arm) and GUR2 is at the X-end (South arm)
  2. The SS containers are good and make a good seal.
  3. We had to replace the screws on the granite slab interface plate. The heads were too big to allow the connector to snap into place.
  4. The Guralps had been left way, way off level and the brass locking screws were all the way up. We locked them down after leveling today. Steve was blaming Cathy(?).
  5. The GUR1_Z channel now looks good - see the summary pages for the before and after behavior. My mistake; the low frequency is still as bad as before.
  6. GUR2 X/Y still look like there is no whitening or if the masses are stuck or the interface box is broken.
  7. When we first powered them up, a few of the channels of both seismometers showed 100-200 Hz oscillations. This has settled down after several minutes.

 

The attachment shows the 6 channels after our work. You can see that GUR2_X/Y still look deadish. I tried wiggling the cables at the interface box and powering on/off, but no luck. Next, we swap cables.

Tried to bring the weather station back to life, but no luck. The unit on the wall is alive and so is the EPICS IOC (c1pem1). But there is apparently no communication between them. telnet into c1pem and the error message repeating at the prompt is:

Weather Monitor Output: NO COMM

Might be related to the flaky connector situation that Liz and I found there a couple summers ago, but I tried jiggling and reseating that one with no luck. Looks like it stopped working around 8 PM on March 24, 2014. That's the same time as a ~30s power outage, so perhaps we just need some more power cycling? Tried hitting the reset button on the VME card for c1pem1, but didn't change anything.

Let's try power cycling that crate (which has c1pem1, c0daqawg, and some GPS receiver)...nope - no luck.

Also tried power cycling the weather box which is near the BS chamber on the wall. This didn't change the error message at the c1pem1 telnet prompt.

Attachment 1: GurPost_150514.png
GurPost_150514.png
Attachment 2: secretWeatherTrends.png
secretWeatherTrends.png
  11292   Fri May 15 16:18:28 2015 SteveUpdateVACVac Operation Guide

Vacuum Operation Guide is up loaded into the 40m-wiki. This is an old master copy. Not exact in terms of real action, but it is still a good guide of logic.

Rana has promissed to watch the N2 supply and change cylinder when it is empty. I will be Hanford next week.

  11293   Sat May 16 20:37:09 2015 ranaHowToCDSBypassing the CDSUTILS prefix issue

The CDSUTILS package has a feature where it substitutes in a C1 or H1 or L1 prefix depending upon what site you are at. The idea is that this should make code portable between LLO and LHO.

Here at the 40m, we have no need to do that, so its better for us to be able to copy and paste channel names directly from MEDM or whatever without having to remove the "C1:" from all over the place.

the way to do this on the command line is (in bash) to type:

export IFO=''


To make this easier on us, I have implemented this in our shared .bashrc so that its always the case. This might break some scripts which have been adapted to use the weird CDSUTILS convention, so beware and fix appropriately.

  11294   Sat May 16 21:05:24 2015 ranaUpdateGeneralsome status

1) Checked the N2 pressures: the unregulated cylinder pressures are both around 1500 PSI. How long until they get to 1000?

2) The IMC has been flaky for a day or so; don't know why. I moved the gains in the autolocker so now the input gain slider to the MC board is 10 dB higher and the output slider is 10 dB lower. This is updated in the mcdown and mcup scripts and both committed to SVN. The trend shows that the MC was wandering away after ~15 minutes of lock, so I suspected the WFS offsets. I ran the offsets script (after flipping the z servo signs and adding 'C1:' prefix). So far powers are good and stable.

3) pianosa was unresponsive and I couldn't ssh to it. I powered it off and then it came back.

4) Noticed that DAQD is restarting once per hour on the hour. Why?

5) Many (but not all) EPICS readbacks are whiting out every several minutes. I remote booted c1susaux since it was one of the victims, but it didn't change any behavior.

6) The ETMX and ITMX have very different bounce mode response: should add to our Vent Todo List. Double checked that the bounce/roll bandstop is on and at the right frequency for the bounce mode. Increased the stopband from 40 to 50 dB to see if that helps.

7) op340 is still running ! The only reason to keep it alive is its crontab:

op340m:SUS>crontab -l

07 * * * * /opt/rtcds/caltech/c1/burt/autoburt/burt.cron >> /opt/rtcds/caltech/c1/burt/burtcron.log
#46 * * * * /opt/rtcds/caltech/c1/scripts/general/scripto_cron /opt/rtcds/caltech/c1/scripts/PSL/FSS/FSSSlowServo > /cvs/cds/caltech/logs/scripts/FSSslow.cronlog 2>&1
#14,44 * * * * /cvs/cds/caltech/conlog/bin/check_conlogger_and_restart_if_dead
15,45 * * * * /opt/rtcds/caltech/c1/scripts/SUS/rampdown.pl > /dev/null 2>&1
#10 * * * *  /opt/rtcds/caltech/c1/scripts/general/scripto_cron /opt/rtcds/caltech/c1/scripts/MC/autolockMCmain40m >/cvs/cds/caltech/logs/scripts/mclock.cronlog 2>&1
#27 * * * * /opt/rtcds/caltech/c1/scripts/general/scripto_cron /opt/rtcds/caltech/c1/scripts/PSL/FSS/RCthermalPID.pl >/cvs/cds/caltech/logs/scripts/RCthermalPID.cronlog 2>&1

00 0 * * * /var/scripts/ntp.sh > /dev/null 2>&1
#00 4 * * * /opt/rtcds/caltech/c1/scripts/RGA/RGAlogger.cron >> /cvs/cds/caltech/users/rward/RGA/RGAcron.out 2>&1
#00 6 * * * /cvs/cds/scripts/backupScripts.pl
00 7 * * * /opt/rtcds/caltech/c1/scripts/AutoUpdate/update_conlog.cron
00 8 * * * /opt/rtcds/caltech/c1/scripts/crontab/backupCrontab

added a new script (scripts/SUS/rampdown.py) which decrements every 30 minutes if needed. Added this to the megatron crontab and commented out the op340m crontab line. IF this works for awhile we can retire our last Solaris machine.

8) To see if we could get rid of the wandering PCDRIVE noise, I looked into the NPRO temperatures: was - T_crystal = 30.89 C, T_diode1 = 21 C, T_diode2 = 22 C. I moved up the crystal temp to 33.0 C, to see if it could make the noise more stable. Then I used the trimpots on the front of the controller to maximimize the laseroutput at these temperatures; it was basically maximized already. Lets see if there's any qualitative difference after a week. I'm attaching the pinout for the DSUB25 diagnostics connector on the back of the box. Aidan is going to help us record this stuff with AcroMag tech so that we can see if there's any correlation with PCDRIVE. The shifts in FSS_SLOW coincident with PCDRIVE noise corresponds to ~100 MHz, so it seems like it could be NPRO related.

 

Attachment 1: 48.png
48.png
Attachment 2: 39.png
39.png
  11295   Sat May 16 21:40:29 2015 ranaUpdatePEMGuralp maintenance

Tried swapping cables at the Guralp interface box side. It seems that all of our seismic signal problems have to do with the GUR2 cable being flaky (not surprising since it looks like it was patched with Orange Electrical tape!! rather than proper mechanical strain relief).

After swapping the cables today, the GUR2 DAQ channels all look fine: i.e. GUR1 (the one at the Y end) is fine, as is its cable and the GUR2 analog channels inside the interface box.

OTOH, the GUR1 DAQ channels (which have GUR2 (EX) connected into it) are too small by a factor of ~1000. Seems like that end of the cable will need to be remade. Luckily Jenne is still around this week and can point us to the pinout / instructions. Looks like there could be some shorting inside the backshell, so I've left it disconnected rather than risk damaging the seismometer. We should get a GUR1 style backshell to remake this cable. It might also be possible that the end at the seismometer is bad - Steve was supposed to swap the screws on the granite-aluminum plate on Thursday; I'll double check.

Attachment 1: GurPost_150516.png
GurPost_150516.png
  11296   Sun May 17 23:46:25 2015 ranaUpdateASCIOO / Arm trends

Looking at the summary page trends from today, you can see that the MC transmission is pretty flat after I zeroed the MCWFS offsets. In addition, the transmission from both arms is also flat, indicating that our previous observation of long term drift in the Y arm transmission probably had more to do with bad Y-arm initial alignment than unbalanced ETMY coil-magnets.

Much like checking the N2 pressure, amount of coffee beans, frames backups, etc. we should put MC WFS offset adjustment into our periodic checklist. Would be good to have a reminder system that pings us to check these items and wait for confirmation that we have done so.

  11297   Mon May 18 09:50:00 2015 ericqUpdateGeneralsome status
Quote:

Added this to the megatron crontab and commented out the op340m crontab line. IF this works for awhile we can retire our last Solaris machine.

For some reason, my email address is the one that megatron complains to when cron commands fail; since 11:15PM last night, I've been getting emails that the rampdown.py line is failing, with the super-helpful message: expr: syntax error

  11298   Mon May 18 11:59:07 2015 ranaUpdateGeneralsome status

Yes - my rampdown.py script correctly ramps down the watchdog thresholds. This replaces the old rampdown.pl Perl script that Rob and Dave Barker wrote.

Unfortunately, cron doesn't correctly inherit the bashrc environment variables so its having trouble running.

On a positive note, I've resurrected the MEDM Screenshot taking cron job, so now this webpage is alive (mostly) and you can check screens from remote:

https://nodus.ligo.caltech.edu:30889/medm/screenshot.html

  11299   Mon May 18 14:22:05 2015 ericqUpdateComputer Scripts / Programsrsync frames to LDAS cluster
Quote:

Still seems to be running without causing FB issues.

I'm not so sure. I just was experiencing some severe network latency / EPICS channel freezes that was alleviated by killing the rsync job on nodus. It started a few minutes after ten past the hour, when the rysnc job started. 

Unrelated to this, for some odd reason, there is some weirdness going on with ssh'ing to martian machines from the control room computers. I.e. on pianosa, ssh nodus fails with a failure to resolve hostaname message, but ssh nodus.martian succeeds. 

  11300   Mon May 18 14:46:20 2015 manasaSummaryGeneralDelay line frequency discriminator for FOL error signal

Measuring the voltage noise and frequency response of the Analog Delay-line Frequency Discriminator (DFD)

The schematic and an actual photo of the setup is shown below. The setup was checked to be physically sturdy with no loose connections or moving parts.

The voltage noise at the output of the DFD was measured using an SR785 signal analyzer while simultaneously monitoring the signal on an oscilloscope.

The noise at the output of the DFD was measured for no RF input and at several RF input frequencies including the zero crossing frequency and the optimum operating frequency of the DFD (20MHz).

The plot below show the voltage noise for different RF inputs to the DFD. It can be seen that the noise level is slightly lower at the zero crossing frequency where the amplitude noise is eliminated by the DFD.

I also did measurements to obtain the frequency response of the setup as the cable length difference has changed from the prior setup. The cable length difference is 21cm and the obtained linear signal at the output of the DFD extends over ~ 380MHz which is good enough for our purposes in FOL. A cosine fit to the data was done as before. //edit- Manasa: The gain of SR560 was set to 20 to obtain the data shown below//

Fit Coefficients (with 95% confidence bounds):
       a =     -0.8763  (-1.076, -0.6763)
       b =       3.771  (3.441, 4.102)

Data and matlab scripts are zipped and attached.

Attachment 4: DFD.zip
  11301   Mon May 18 16:28:18 2015 ericqUpdateGeneralsome status
Quote:

4) Noticed that DAQD is restarting once per hour on the hour. Why?

It looks like daqd isn't being restarted, but in fact crashing every hour.

Going into the logs in target/fb/logs/old, it looks like at 10 seconds past the hour, every hour, daqd starts spitting out:

[Mon May 18 12:00:10 2015] main profiler warning: 1 empty blocks in the buffer                                     
[Mon May 18 12:00:11 2015] main profiler warning: 0 empty blocks in the buffer                                     
[Mon May 18 12:00:12 2015] main profiler warning: 0 empty blocks in the buffer                                     
[Mon May 18 12:00:13 2015] main profiler warning: 0 empty blocks in the buffer
...
***CRASH***

An ELOG search on this kind of phrase will get you a lot of talk about FB transfer problems. 

I noticed the framebuilder had 100% usage on its internal, non-RAID, non /frames/, HDD, which hosts the root filesystem (OS files, home directory, diskless boot files, etc), largely due to a ~110GB directory of frames from our first RF lock that had been copied over to the home directory. The HDD only has 135GB capacity. I thought that maybe this was somehow a bottleneck for files moving around, but after deleting the huge directory, daqd still died at 4PM. 

The offsite LDAS rsync happens at ten minutes past the hour, so is unlikely to be the culprit. I don't have any other clues at this point. 

  11302   Mon May 18 16:56:12 2015 ericqHowToCDSBypassing the CDSUTILS prefix issue
Quote:

export IFO=''

This makes things act weird:

controls@pianosa|MC 1> z avg 1 "C1:LSC-TRY_OUT"
IFO environment variable not specified.

  11303   Mon May 18 17:42:14 2015 rana, ericQUpdateGeneralsome status

Today at 5 PM we replaced the east N2 cylinder. The east pressure was 500 and the west cylinder pressure was 1000. Since Steve's elogs say that the consumption can be as high as 800 per day we wanted to be safe.

  1. We closed the black valve before the regulator and closed the valve on the cylinder.
  2. We unscrewed the brass fill line to the cylinder.
  3. We unchained the cylinder and put in the dolly (and attached the chains on there).
  4. We rolled in a fresh cylinder from outside using the red dolly (it should have chains).
  5. We put it in place, hooked up the chains, and screwed on the brass nozzle with the large adjustable wrench (need to put a non-adjustable here).
  6. Opened up the cylinder valve.
  7. Opened up the black valve.
  8. New east pressure reading is 2500 PSI. Regulated N2 pressure is 68 PSI.
Quote:

1) Checked the N2 pressures: the unregulated cylinder pressures are both around 1500 PSI. How long until they get to 1000?

 

  11304   Mon May 18 17:44:30 2015 ranaHowToCDSBypassing the CDSUTILS prefix issue

Too weird. I undid me changes. We'll have to make the C1: stuff work inside each python script.

Quote:
Quote:

export IFO=''

This makes things act weird:

controls@pianosa|MC 1> z avg 1 "C1:LSC-TRY_OUT"
IFO environment variable not specified.

 

  11305   Mon May 18 18:03:12 2015 ranaUpdateGeneralsome status

The c1cal model was maxing out its CPU meter so I logged onto c1lsc and did 'rtcds c1cal stop'. Let's see if this changes any of our FB / DAQD problems.

Attachment 1: CPUtrend.png
CPUtrend.png
  11306   Tue May 19 00:19:23 2015 ranaUpdateGeneralsome status

There's a few hours so far after today's c1cal shut off that the summary page shows no dropouts. I'm not yet sure that this is related, but it seems like a clue.

Attachment 1: Screen_Shot_2015-05-19_at_12.17.39_AM.png
Screen_Shot_2015-05-19_at_12.17.39_AM.png
  11307   Tue May 19 11:15:09 2015 ericqUpdateComputer Scripts / ProgramsChiara Backup Hiccup

Starting on the 14th (five days ago) the local chiara rsync backup of /cvs/cds to an external HDD has been failing:

caltech/c1/scripts/backup/rsync_chiara.backup.log:

2015-05-13 07:00:01,614 INFO       Updating backup image of /cvs/cds
2015-05-13 07:49:46,266 INFO       Backup rsync job ran successfully, transferred 6504 files.
2015-05-14 07:00:01,826 INFO       Updating backup image of /cvs/cds
2015-05-14 07:50:18,709 ERROR      Backup rysnc job failed with exit code 24!
2015-05-15 07:00:01,385 INFO       Updating backup image of /cvs/cds
2015-05-15 08:09:18,527 ERROR      Backup rysnc job failed with exit code 24!
...
 

Code 24 apparently means "Partial transfer due to vanished source files."

Manually running the backup command on chiara worked fine, returning a code of 0 (success), so we are backed up. For completeness, the command is controls@chiara: sudo rsync -av --delete --stats /home/cds/ /media/40mBackup

Are the summary page jobs moving files around at this time of day? If so, one of the two should be rescheduled to not conflict. 

  11308   Tue May 19 11:24:44 2015 ericqUpdateComputer Scripts / ProgramsNotification Scheme

Given some of the things we've facing lately, it occurs to me that we could be better served by having some sort of unified human-alerting scheme in place, for things like:

  • Local/offsite backup failures
  • Vaccumm system problems
  • HDD status for things like /frames/ and /cvs/cds/, whether the disks are full, or their SMART status indicates imminent mechanical failure

Currently, many of these things are just checked sporadically when it occurs to someone to do so, or when debugging random issues. Smoother IFO operation and peace of mind could be gained if we're confident that the relevant people are notified in a timely manner. 

Thoughts? Suggestions on other things to monitor, like maybe frontend/model crashes?

  11309   Tue May 19 11:50:52 2015 manasaUpdatePEMNo noticeable effect from M4.0 earthquake

There was an earthquake: M4.0 - 40km SSW of South Dos Palos, California 

No noticeable effects on the IFO. MC did not lose lock; however the arms did unlock.

  11310   Tue May 19 14:51:44 2015 ericqUpdateModern ControlBrushing up on Wiener Filtering

As part of preparing for the SURF projects this summer, I grabbed ~50 minutes of MCL and STS_1 data from early this morning to do a little MISO wiener filtering. It was pretty straightforward to use the misofw.m code to achieve an offline subtraction factor of ~10 from 1-3Hz. This isn't the best ever, but doesn't compare so unfavorably to older work, especially given that I did no prefiltering, and didn't use all that long of a data stretch.

Code and plot (but not data) is attached. 

Attachment 1: mclData.png
mclData.png
Attachment 2: mclWiener.zip
  11311   Tue May 19 16:18:57 2015 ericqUpdateGeneralcrons fixed

I wrapped rampdown.py in rampdown.sh, which is just these lines:

#!/bin/bash
source /ligo/cdscfg/workstationrc.sh
/opt/rtcds/caltech/c1/scripts/SUS/rampdown.py > /dev/null 2>&1

This is now what megatron's cron runs. It appears to be working.

I also added the workstationrc line to the n2 and chiara HDD checking scripts that run on nodus, which should resolve the issue from ELOG 11249

  11312   Tue May 19 17:03:34 2015 KojiUpdateCDSMXstream restart script working (beta)

AutoMX is resetting mx_stream every 5 minutes. Basically everytime AutoMX is called,
it resets mx_stream. Is the mx_stream stalling such often? Or the script is detecting false alerms?


> tail -200 /opt/rtcds/caltech/c1/scripts/cds/autoMX.log

Tue May 19 16:43:01 PDT 2015
LSC - FB bad. Runnning restart:
 * Stopping mx_stream ...                                                 [ ok ]
 * Starting mx_stream ...                                                 [ ok ]
Connection to c1sus closed.
 * Stopping mx_stream ...                                                 [ ok ]
 * Starting mx_stream ...                                                 [ ok ]
Connection to c1lsc closed.
 * Stopping mx_stream ...                                                 [ ok ]
 * Starting mx_stream ...                                                 [ ok ]
Connection to c1ioo closed.
 * Stopping mx_stream ...                                                 [ ok ]
 * Starting mx_stream ...                                                 [ ok ]
Connection to c1iscex closed.
 * Stopping mx_stream ...                                                 [ ok ]
 * Starting mx_stream ...                                                 [ ok ]
Connection to c1iscey closed.
0
Tue May 19 16:48:02 PDT 2015
LSC - FB bad. Runnning restart:
 * Stopping mx_stream ...                                                 [ ok ]
 * Starting mx_stream ...                                                 [ ok ]
Connection to c1sus closed.
 * Stopping mx_stream ...                                                 [ ok ]
 * Starting mx_stream ...                                                 [ ok ]
Connection to c1lsc closed.
ssh_exchange_identification: read: Connection reset by peer
 * Stopping mx_stream ...                                                 [ ok ]
 * Starting mx_stream ...                                                 [ ok ]
Connection to c1iscex closed.
 * Stopping mx_stream ...                                                 [ ok ]
 * Starting mx_stream ...                                                 [ ok ]
Connection to c1iscey closed.
0
Tue May 19 16:53:01 PDT 2015
LSC - FB bad. Runnning restart:
 * Stopping mx_stream ...                                                 [ ok ]
 * Starting mx_stream ...                                                 [ ok ]
Connection to c1sus closed.
 * Stopping mx_stream ...                                                 [ ok ]
 * Starting mx_stream ...                                                 [ ok ]
Connection to c1lsc closed.
 * Stopping mx_stream ...                                                 [ ok ]
 * Starting mx_stream ...                                                 [ ok ]
Connection to c1ioo closed.
 * Stopping mx_stream ...                                                 [ ok ]
 * Starting mx_stream ...                                                 [ ok ]
Connection to c1iscex closed.
 * Stopping mx_stream ...                                                 [ ok ]
 * Starting mx_stream ...                                                 [ ok ]
Connection to c1iscey closed.

  11313   Tue May 19 17:53:38 2015 ranaUpdateModern ControlBrushing up on Wiener Filtering

Good to see that misofw.m is still alive and well. Todo:

  1. Apply weighting to the filter via time domain SOS prefiltering of the signals (see Jenne's code from the LLO Global FF paper)
  2. Consider using some of the MC SUSPOS signals. Its not strictly legal, but I wonder if its responsible for out noise below 1 Hz.
  3. Steve was in the midst of hooking up the Wilcoxones to get better subtraction at 5-15 Hz, but couldn't find cables. We need to hunt them down in W Bridge.
  4. Attach one of the medium or big Lings to the MC2 chamber platform and see if we can do this in hardware without driving the suspension. WE want to use a DAC channel (from where?) and pipe it to the Ling using a medium high current drive. Might start with the 50 Ohm output of the SR560 and later use a BUF634 ala Crackle coil driver.
  11314   Tue May 19 18:38:33 2015 ranaUpdateCDSMXstream restart script working (beta)

Good catch; that was some seriously bad programming on my part. I had some undeclared variable garbage going on. I fixed it and re-implemented the script in CRON on megatron. The log file shows that it has detected no problems for the last several checks. I'll check it again tomorrow to see if its doing good or bad.

  11315   Tue May 19 18:55:12 2015 rana, ericQUpdateGeneralsome status

After one day the pressures are east/west = 2200/450 PSI

Quote:

 

  1. New east pressure reading is 2500 PSI. Regulated N2 pressure is 68 PSI.

 

  11316   Tue May 19 19:24:30 2015 ranaUpdatePEMSeismic BLRMS filters

I was wondering about the design of the BLRMS fitlers for the seismic channels since the STS ones seem to have so little gain compared to the Guralps.

Here are some plots of the Bode magnitude and impulse responses of the bandpass filters (before the low passing). There's a bunch of entries from Masha on this from her SURF summer. Can anyone comment on why they are all so different?

One of the old Masha entries speaks of designing the lowpass filter in an intelligent way: by adjusting the filter order until the power in the stopband is less than 1% of the power in the passband. Seems like we could do that for bandpass too. For now I have made the names reasonable and changed all of the BP filters to 4th order Butterworth.

Also, it turns out that the Vel2Vel (gain ~0.02) filters were mistakenly on in the STS BP filter banks. The GUR inputs have a gain to scale the counts to velocity, but the STS seem to already be in microns/sec (where is this gain?) so I turned off and deleted the Vel2Vel filters; in any case the gain should not be done seperately in each BP bank, but altogether before the BP filtering.

Attachment 1: BLRMS_BP.pdf
BLRMS_BP.pdf
Attachment 2: BLRMS_imp.pdf
BLRMS_imp.pdf
  11317   Wed May 20 03:08:27 2015 ranaUpdateGeneralsome status

I think that the real clue was that the dropouts are in some channels and not in others:

https://nodus.ligo.caltech.edu:30889/detcharsummary/day/20150519/psl/

As it turns out, the channel with no dropouts is the RAW PSL RMTEMP channel. All the others are the minute trends. So something is up with the trend making or the trend reading in the cluster.

Quote:

There's a few hours so far after today's c1cal shut off that the summary page shows no dropouts. I'm not yet sure that this is related, but it seems like a clue.

 

  11318   Wed May 20 11:41:59 2015 ericqUpdateGeneralsome status

West cyclinder is empty, east is at 2000 psi; regulated N2 pressure is 64psi. I'll replace the west one after the meeting.

  11319   Fri May 22 11:59:54 2015 ericqUpdateSUSDampRestore script problem

PRM watchdog tripped, but the damprestore.py script wouldn't run. 

It turns out the script tries to import some ezca stuff from /users/yuta (angry), which had been moved to /users/OLD/yuta (crying). 

I've moved the yuta directory back to /users/ until I fix the damprestore script. 

  11320   Fri May 22 12:09:57 2015 ranaUpdateSUSDampRestore script problem

I will move it back. We need to fix our scripts to not use any users/ libraries ever again.

Quote:

PRM watchdog tripped, but the damprestore.py script wouldn't run. 

It turns out the script tries to import some ezca stuff from /users/yuta (angry), which had been moved to /users/OLD/yuta (crying). 

I've moved the yuta directory back to /users/ until I fix the damprestore script. 

 

  11321   Fri May 22 18:09:58 2015 ericqUpdateComputer Scripts / ProgramsifoCoupling

I've started working on a general routine to measure noise couplings in our interferometers. Often this is done with swept sine measurements, but this misses the nonlinear part of the coupling, especially if the linear part is alreay reduced through some compensation or feedforward scheme. Rana suggested using a series of narrow band-limited noise injections. 

The structure I'm working on is a python script that uses the AWG interface written by Chris W. to create the excitations. Afterwards, I calculate a series of PSD estimates from the data (i.e. a spectrogram), and apply a two-sample, unequal variance, t-test to test for statisically significant increases in the noise spectra to try and evaluate the nonlinear contriubutions to the noise. I've started a git repository at github.com/e-q/ifoCoupling with the code. 

So far, I've tested one such injection of noise coupling from the ETMX oplev error point to the single arm length error signal. It's completely missing the user interface and structure to do a general series of measurements, but this is just organizational; I'm trying to get the math/science down first. 

Here's a result from today:

Median, instead of the usual mean, PSDs are used throughout, to reject outliers/glitches.

The linear part of the coupling can be estimated using the coherence / spectrum height in the excitation band, but I'm not sure what the best what to present/paramerize the nonlinear parts of each individaul excitation band's result is.

Also, I anticipate being able to write an excitation auto-leveling routine, gradually increasing the exctiation level until the excited spectrum is some amount  noisier than the baseline spectrum, up to some maximum amount configurable by the user. 

The excitation shaping could probably be improved, too. It's currently and elliptic + butterworth bandpass for a sharp edge and rolloff. 

I'm open to any thoughts and/or suggestions anyone may have!

Attachment 1: ETMX_PIT_L_coupling.png
ETMX_PIT_L_coupling.png
  11322   Sat May 23 22:43:10 2015 ericqUpdateGeneralifo recovery log

Running train-of-thought elog:


East N2 cylinder found empty, replaced. West is >2kpsi


Removed Yuta-specific code from damprestore. A grep for 'yuta' in the python files within /scripts/ shows some other occurances, but nothing that is really in use at this time. New feature of damprestore.py: remembers oplev status.


Ran LSCoffsets. 

WFS offsets relieved (all <20).

Adjusted FSS offset to minimize MC_FAST_MON

ASS ran (but the arm alignment has been astoundingly stable lately. I haven't touched it all this week)

ITMX is the only optic that got a correction over 20 counts. 

BS and *TM oplev spots look well centered, except for ITMX. 

I undid the gain reduction rana introduced because X ASS seemed to be really slow. It is currently fine in its older state. What's going on here?


Some network latency stuff is going on, even freezing up terminals when trying to write text files. This may (or may not) be correlated with the summary page rysnc jobs on nodus. It occurs to me that we have a DAQ ethernet network seperate from the martian network, but the frame transfers need to go through the martian network, since nodus is the only way out to the outside world. If we had a machine/gateway directly from the DAQ network to the caltech network, the martian network wouldn't get bogged down when frames are being uploaded


GTRY = 0.55, ok. Aligned GTRX to 0.52, also ok

Y beatnote was found easily. Have spent >30 minutes looking for X green beatnote. Typical FSS slow and X temperature ranges don't seem to be giving much. Will check the beat alignment with a scope, but if the beat is too high to begin with, it may not work...


I suspect a problem with the X Green BBPD

I could see the IR beatnote between the PSL and AUX X lasers at the input to the frequency counter. (I believe it is a real beatnote because it reacted as expected to the end temperature moving, and stabilizing the end laser to the arm). However, when placing the IR beatnote at a frequency which should've made the green beatnote visible on an analyzer and/or scope, no beatnote was found. I played with the beat alignment to no avail; the DC output of the BBPD behaved as expected, but I never saw anything in the RF output or on the control room analyzer. sad I also checked the beatnote signal chain by hooking up a 1mV 26MHz signal into the cable that hooks up to the RF out of the BBPD; the signal showed up clearly on the control room analyzer. 

I'm not sure what may have happened. ELOG 9996 may be related. 

I'm calling it a night. 