ID |
Date |
Author |
Type |
Category |
Subject |
52
|
Thu Nov 1 19:54:22 2007 |
Andrey Rodionov | Bureaucracy | Photos | Rana's photo |
|
Attachment 1: DSC_0120.JPG
|
|
11635
|
Tue Sep 22 16:52:36 2015 |
ericq | Summary | General | Random Notes |
Some things bouncing around my head that haven't made it to ELOG yet:
- Last week, Rana and I were investigating excess power line noise coming from the DFD demodulation. We put transformers on the green beat signals where they arrive at the LSC rack, to avoid connecting their signal ground from the PSL table to the LSC rack ground. This didn't help; it's unclear what the culprit is; maybe the demod board power board?
- Lately, when the interferometer loses lock, the Y arm will not lock on POY, or even flash its IR resonance. for a little while. The green beam can be locked, the X arm can be locked, and no excess angular noise is evident from glancing at the oplev XY plots. Mysterious.
- Sometimes, when writing new values to the C1LSC SDF table, the c1lscepics process dies (though the write is successful). This is highly annoying. This may have been adressed in some slightly newer RCG code.
- C1OAF is running with a big red NO SYNC message on its GDS screen. C1LSC has shown this too, but I think only when the SDF/epics crash happens.
- C1OAF also doesn't seem to properly load the "safe" SDF table when starting up, and errantly puts ones in every element in the static FF matrix. Be careful when restarting OAF!
|
2430
|
Thu Dec 17 23:27:23 2009 |
rana | Update | PEM | Ranger Noise: sim w. Rai FET box as readout |
I have started measuring the low frequency noise of the FET front end + LT1128 low noise preamp from Rai Weiss. It has a very low input current noise because its FET based, which is not surprising. It is also a fairly low voltage noise box - the best measured ones have an input referred noise of ~0.35 nV/rHz.
Today I measured the noise of the one we have down to 0.1 Hz. It looks like a good candidate for a Geophone readout (e.g. Ranger or GS-13 or perhaps the L-4C). Because I didn't thermally shield any of this stuff, the broadband noise is ~0.8 nV/rHz. The low frequency corner is ~15 Hz.
I attach the LISO simulation of the voltage noise referred to the input. The circuit is described in this entry.
We can probably do better than this if we package it a little better or give it time to warm up or use metal film resistors inside. Even as it is, however, it would allow us to reach the thermal noise of the Ranger (or GS-13) down to 0.1 Hz.
This should be ~1.5 or 2x better than the LT1012 based readout at 1 Hz and 10x better down at 0.1 Hz (c.f. T0900457). |
Attachment 1: ranger.pdf
|
|
363
|
Fri Mar 7 00:47:54 2008 |
rana | Configuration | PEM | Ranger SS-1 |
Yesterday evening around 7:30 PM, I changed the Ranger seismometer from a
vertical to a horizontal seismometer. To do this I followed the instructions
in the manual.
1) Lock it down.
2) Turn it sideways. Use the leveling screws to center the bubble level.
3) Carefully loosen the hanger rod and release slowly the tension to allow
the mass to recenter.
4) Look through the little viewhole next to the rod to make the white lines
line up. This means the mass is centered.
5) Look at the output on a scope. It should be freely moving with a ~1 sec.
period.
The attached plot shows the before and after spectra. |
Attachment 1: ss1.pdf
|
|
881
|
Mon Aug 25 15:50:18 2008 |
rana | Summary | PEM | Ranger SS-1 |
The manual for the Ranger SS-1 seismometer can be found on line here:
ftp://ftp.kmi.com/pub/software_manuals/300190/300190nc.pdf
and now in our 40m PEM Wiki page:
Ranger_SS-1
To calibrate it, we use the formula from the manual:
R_x
G_L = G_0 * ------------ = 149 +/- 3 V/(m/s)
R_x + R_c
where
G_0 = 340 V/(m/s) (generator constant)
R_x = 4300 Ohms (external damping resistor in Pomona box)
R_c = 5500 Ohms (internal coil resistance)
Then we have a gain of 200 in the SR560 so that gets us to ~30000 V/(m/s).
And then there's a DAQ conversion factor of the usual 2^16 cts / 4 V.
so the calibration constant is
G = 488 counts / (micron/sec)
in the ~1-50 Hz band |
1167
|
Tue Dec 2 19:18:10 2008 |
rana | Summary | PEM | Ranger SS-1 |
In entry http://dziban.ligo.caltech.edu:40/40m/881 and a follow up from Jenne I put in the Ranger calibration.
Since then, we've reduced the SR560 gain from 200 to 100 so the calibration factor is now:
1e-9 (m/s)/count and then 2 poles at 0 Hz, and a Q~1 zero pair at 1 Hz.
in DTT:
G = 1e-9
p = 0, 0
z = 0.7 0.7 |
391
|
Fri Mar 21 23:15:11 2008 |
rana | Configuration | PEM | Ranger SS-1: New Setup |
The Ranger seismometer has been in a bad state. Its output had been sent into a SR560 without any termination.
The seismometer is, internally, just a mass on a flexure with a magnet and a pickup coil for readout.
The damping of the system depends on the resistor hooked up across the coil. With the SR560 this is
the 1 Meg input impedance of it and so the mass is undamped.
I installed a 4300 Ohm resistor in there which seems to nearly critically damp it. However, this will not
allow us to reach the ultimate quantum noise limited performance. We will have to analyze the thermal, voltage,
and current noise to get that.
I then also increased the gain from 10 to 100 on the SR560. This should now make the front end noise of the
seismometer/SR560 close to equal to the noise of the PEM ADC. |
1171
|
Wed Dec 3 19:21:09 2008 |
rana | Configuration | PEM | Ranger move |
I looked at the Ranger signals. Somehow it has a relative transfer function of 'f' between it and the Guralp. Ranger
i.e. ------ ~ f
Guralp
which is strange since according to their manuals, they should both be giving us a voltage output which is proportional
to velocity. I checked that the Ranger only has a load resistor and then an SR560 low pass at 300 Hz. Jenne assures
me that the Guralp breakout box shouldn't have any poles either (to be double checked). Its a mystery.
We made sure that the SR560 now is DC coupled, G = 100, & 1-pole low pass at 300 Hz. I moved it over next to the Guralp
(went through the mass recentering procedure after forgetting to lock it before moving). It is behaving as it was
before.
Attached is a 2 page PDF of the comparisons. The 'MC1' channels are Guralp and 'MC2' is Ranger.
The second attachment compares our seismometers (in counts) with the LHO Guralp seismometers. There's no high frequency
rolloff there like what we see here so I bet a dollar that there's a pole in the Guralp box somewhere. |
Attachment 1: c.pdf
|
|
Attachment 2: wsnb.pdf
|
|
2194
|
Fri Nov 6 16:27:15 2009 |
Jenne | Update | PEM | Ranger moved |
The Ranger seismometer has been moved to ~the middle of the Mode Cleaner tube, and it's orientation has been changed to horizontal (using all of the locking/mass centering procedures). This is similar in orientation to the way things were back in the day when Rana and Matt had the OAF running nicely. |
2351
|
Fri Dec 4 18:54:03 2009 |
Jenne | Update | PEM | Ranger moved |
The Ranger was left in a place where it could be bumped during next week's activities (near the crawl-space to access the inside of the "L" of the IFO on the Yarm). It has been moved a meter or so to a safer place.
Also, so that Steve can replace the battery in the SR560 that is used for the Ranger, I swapped it out with one of the ones which already has a new, charged battery. All of the settings are identical. For posterity, I took a pic of the front panel before unplugging the old SR560. |
Attachment 1: RangerSeismometer_SR560settings_4Dec2009.JPG
|
|
1106
|
Sun Nov 2 21:37:22 2008 |
rana | Update | PEM | Ranger recovery |
The ranger signal has been bad since around 11 AM on Oct 25 (last Saturday). There are no elog
entries from that day, but I am quite sure that someone must have been working around the PSL
rack area.
It looks like what happened is that someone moved the chair with the monitor on it and/or the wooden
stool next to it. That put tension on the cable connecting the SR560 and the seismometer. The SR560
connector now seems loose and I think probably the cable ground wasn't connected. I swapped the
cable over to the "B" side of the SR560 and the ranger signal is now reasonable (very small offset
and normal seismic signal).
Please be careful when working around there. Everyone always says "I didn't do anything" or "it doesn't
effect anything".
We need to clean up the cabling around there in addition to running a new power cable for the RF amplifier
on the POY table.
I have also reduced its sample rate from 2048 to 512 Hz. The data are OK after 909640694.
I also increased the sample rate of AS_MIC from 2048 to 16384 Hz but that one seems to be broken
---->> the microphone seems to be either disconnected or broken. |
10067
|
Wed Jun 18 22:47:48 2014 |
ericq | Update | CDS | Raspberry pi added to martian network |
I set up a raspberry pi on the martian network, to be hooked up to a frequency counter for tracking ALS beatnotes.
The instructions at https://wiki-40m.ligo.caltech.edu/Martian_Host_Table are outdated, the name server configuration is now at /etc/bind/zones/martian.db, I need to remember to update the wiki soon.
In any case, the raspberry pi is called "domenica," is found at 192.168.113.107, and has the standard controls user, with /cvs/cds mounted in the same way as the control room machines.
Once I'm comfortable with the configuration of the pi, I'm going to take an image of the SD card that serves as its hard drive, so that we can just image new cards for future raspberry pis on the martian network if we ever want them. |
11965
|
Mon Feb 1 09:16:32 2016 |
Steve | Update | PEM | Rat got cut |
We got it! Traps are removed.
|
Attachment 1: bingo.jpg
|
|
Attachment 2: bingo3.png
|
|
14186
|
Tue Aug 28 15:29:19 2018 |
Steve | Frogs | PEM | Rat is cut |
The rat is cut by mechanical trap and it was removed from ITMX south west location.
A nagy kover patkanyt a fogo elkapta es megolte. |
Attachment 1: rat#2.png.png
|
|
11911
|
Tue Jan 5 15:48:16 2016 |
Steve | Update | safety | Rat trap locations |
Please look around when working close to these five locations. Use flashlights or leave lights on.
These mechanial traps are HAZARDOUS !
No visitors or tours till Monday, Jan 11 2016 |
Attachment 1: Control_Room.jpg
|
|
Attachment 2: EarmCESdoor.jpg
|
|
Attachment 3: 1Y1.jpg
|
|
Attachment 4: MC2.jpg
|
|
Attachment 5: XarmMaglev.jpg
|
|
11913
|
Tue Jan 5 17:19:57 2016 |
rana | Update | safety | Rat trap locations |
In the modern times, people use glue traps to catch rats instead of springs. They are less hazardous to people and don't spread rat fluid on the floor. |
11909
|
Tue Jan 5 09:48:52 2016 |
Steve | Update | PEM | Rat trap moved |
Our janitor confirmed that Q was not hallucinating about this animal. The dropping size indicating a good size one in the IFO room.
One of the mechanical traps moved from the control room to the east arm, close to the " machine shop " door.
I'm going to get more traps.
Quote: |
Two mechanical and two sticdky traps were set to catch univited visitor.
Absolutely no food or food remains into inside garbage cans!!!!!!!!!!!!!!!!!!!!!!!!!
Quote: |
A small rat / large mouse just ran through the control room. Ugh.
|
|
|
11895
|
Mon Dec 21 14:31:41 2015 |
Steve | Update | PEM | Rat traps set |
Two mechanical and two sticdky traps were set to catch univited visitor.
Absolutely no food or food remains into inside garbage cans!!!!!!!!!!!!!!!!!!!!!!!!!
Quote: |
A small rat / large mouse just ran through the control room. Ugh.
|
|
Attachment 1: ratsNC.jpg
|
|
11893
|
Sun Dec 20 23:23:54 2015 |
ericq | Update | ALARM | Rats. |
A small rat / large mouse just ran through the control room. Ugh. |
10657
|
Fri Oct 31 11:46:15 2014 |
manasa | Update | | Rattling HEPA : Eventually stops |
The PSL HEPA stopped working while it was running at 80%. I have closed the PSL enclosure.
Steve is working to fix this. |
6792
|
Mon Jun 11 16:08:58 2012 |
Jenne | Update | Environment | Rattling in the HEPA |
There is an intermittent rattling sound coming from the HEPA in the NE corner of the PSL table (right above the PMC, all of our input optics).
Steve says it might be a bad bearing, but he'll check it out in the morning and get it fixed. |
6793
|
Mon Jun 11 21:35:55 2012 |
Jenne | Update | Environment | Rattling in the HEPA |
Quote: |
There is an intermittent rattling sound coming from the HEPA in the NE corner of the PSL table (right above the PMC, all of our input optics).
Steve says it might be a bad bearing, but he'll check it out in the morning and get it fixed.
|
MC was having a hard time staying locked, with no discernable reason from the control room (i.e. no big seismic, no PMC PZT railing). The HEPA was on 100%, so I turned it down to 50% to hopefully reduce the rattling, if that was what was wrong. |
10103
|
Wed Jun 25 17:49:36 2014 |
Harry | Update | General | Razorblade Analysis Pt. 2 |
Reconfigured razorblade analysis setup on the PD table as per instructions. Used it to collect data to calculate beam waist with, analyses to follow.
See attached schematic for optical setup. |
Attachment 1: RazorbladeSetup.pdf
|
|
10083
|
Fri Jun 20 18:33:53 2014 |
Harry | Update | General | Razorblade Beam Analysis Setup |
Eric Q and I set up the optical configuration for razorblade beam analysis on SP table for future use.
It has been aligned, and will be in use on Monday.
The beam will be characterized for future characterization of optical fibers. |
10092
|
Tue Jun 24 13:04:49 2014 |
Harry, Manasa | Update | General | Razorblade Beam Analysis Setup |
Harry will update this elog with details about his beam waist measurements for the old NPRO on the SP table. |
10099
|
Wed Jun 25 09:17:33 2014 |
Harry, Manasa | Update | General | Razorblade Beam Analysis Setup |
Quote: |
Harry will update this elog with details about his beam waist measurements for the old NPRO on the SP table.
|
see http://nodus.ligo.caltech.edu:8080/40m/10098 for the update |
10098
|
Wed Jun 25 09:16:52 2014 |
Harry | Update | General | Razorblade Measurements |
Purpose
To use a razorblade to measure beam waist at multiple points along the optical axis, so as to later extrapolate the modal profile of the entire beam. This information will then be used to effectively couple AUX laser light to fibers for use in the frequency offset locking apparatus.
Data Acquisition
1) Step the micrometer-controlled razorblade across the beam at a given value of Z, along optical axis, in the plane orthogonal to it (arbitrarily called X).
2) At each value of X, record the corresponding output of a photodiode, (Thorlabs PD A55) here given in mV.
3) Repeat process at multiple points along Z
Analysis
Data from each iteration in the X were fitted to the error function shown below.
V(x) = A*(erf((x-m)/s)+c)
In the Y, they were fitted to:
V(x) = -A*(erf((x-m)/s)+c)
'A' corresponds to an amplitude, 'm' to a mean, 's' to a σ, and 'c' to an offset.
(Only because in Y measurements, the blade progressed toward eclipsing the beam, as opposed to in the X where it progressively revealed the beam.
These fits can be solved for x = (erf-1((V/A)-c)*s)+m1 which can be calculated at the points (Vmax/e2) and (Vmax*(1-1/e2)). The difference between these points will yield beam waist, w(z).
Conclusion
Calculations yielded waists of: X1=66.43um, X2=67.73um, X3=49.45um, Y1=61.20um, Y2=58.70, Y3=58.89
These data seem suspect, and shall be subjected to further analysis.
|
Attachment 1: 40m.zip
|
3409
|
Thu Aug 12 16:18:00 2010 |
Jenna | Update | Electronics | Rb clocks overnight |
I took a look at the data from the middle of the night to see if it was significantly quieter than the data from the day, but it doesn't seem to be. The plot shows data from yesterday around 12:30pm and from this morning around 2am. It's a bit quieter at low frequencies, but not by much. |
Attachment 1: rbcomp.pdf
|
|
4573
|
Wed Apr 27 17:38:01 2011 |
kiwamu | Update | Electronics | Re : AS55 demod board with new 90 deg splitter : healthy |

Figure.1 I-Q relative phase measurement as a function of LO power.
Blue curve : relative phase of AS55 that I have modified today (#4572).
Red curve : relative phase of AS11 that I had modified a week ago (#4554). Just for comparison.
The relative phase of AS55 agrees approximately what we expected according to the datasheet of PSCQ-2-51W. We expected 85 degree.

Figure.1 I-Q amplitude imbalance as a function of LO power.
From - 5 dBm to 5 dBm in LO power the imbalance is within 3 %.
But the precision of the measurement is also about 2 % (because I used an oscilloscope). Even so the imbalance is still good.
Quote from #4572 |
Some plots will be posted later.
|
|
4943
|
Wed Jul 6 02:12:36 2011 |
kiwamu | Update | SUS | Re : More normalization of all sus controllers |
[Jenne / Rana/ Kiwamu]
We found the 30 Hz high pass filters had lower gain than what they used to be at low frequcnies.
So we increased the gain of the high pass filters called '30:0.0' by a factor of 10 to have the same gain as before.
Now all the suspension shows some kind of damping. Needs more optimizations, for example Q-adjustments for all the suspensions...
Quote from #4942 |
This is getting closer, but with the whitening left OFF and the cts2um filter also OFF, none of the suspensions are working correctly.
|
|
4944
|
Wed Jul 6 10:35:35 2011 |
Jamie | Update | SUS | Re : More normalization of all sus controllers |
Quote: |
We found the 30 Hz high pass filters had lower gain than what they used to be at low frequcnies.
So we increased the gain of the high pass filters called '30:0.0' by a factor of 10 to have the same gain as before.
|
I'm not convinced that this is what you want to do, or at least I wouldn't do it this way. The "k" in the zpk filter was set such that the filter had unity gain above the high-pass cut-off frequency. For a 30 Hz high-pass the k needs to be a factor of 10 smaller than it would be for a 3 Hz high-pass to achieve this high frequency unity gain.
As it is now these HP filters have 20 dB of gain above 30 Hz. If the open loop transfer function needs to more gain I would have done that by adjusting the overall DC gain of the filter bank, not by increasing the gain in this one filter. Maybe you guys have been doing it differently, though. Or maybe I'm just completely off base. |
4946
|
Wed Jul 6 15:32:32 2011 |
Jamie | Update | SUS | Re : More normalization of all sus controllers |
So after talking to Kiwamu about it, I understand now that since the damping loops need all of this extra gain when the high-pass corner is moved up, it's more convenient to put that gain in the control filter itself, rather than having to crank the overall DC gain up to some inconveniently high value. |
4799
|
Thu Jun 9 12:21:07 2011 |
kiwamu | Update | PSL | Re : PMC needs help |
Nulling the slow actuation offset fixed the issue. Now PMC is back to normal.

The reflected beam on the CCD was quite symmetric (it looked very TEM00 mode !) for some reasons, I somehow suspected the mode matching to PMC.
One possibility I thought of was the laser temperature because it could change the laser spatial mode.
So I looked at the slow actuation offset on the FSS screen and found it was at -4.0 which sounds somewhat big.
Then I zeroed the offset by the slider and relocked PMC.
Then the spatial pattern of the reflected beam became usual (i.e. junk light looking) and the transmitted light wet up to 0.83 which is normal.
Quote from #4798 |
The PMC is losing power.
|
|
5075
|
Sun Jul 31 00:37:57 2011 |
kiwamu | Update | PSL | Re : PSL table work |
I think you made a simple mistake in your diagram -- the mixer must be replaced by a summer circuit. Otherwise you cannot do the PDH lock.
|
5316
|
Mon Aug 29 00:49:00 2011 |
kiwamu | Update | CDS | Re : fb down |
Fb is in a bad situation. It needs a MANUAL fsck to fix the file system.
HELP US, Jamieeeeeeeeeeee !!!
When Suresh and I connected a display and tried to see what was going on, the fb computer was in a file system check.
This was because Suresh did a hardware reboot by pressing a power button on the front panel.
Since the file checking took so long time and didn't proceed fast, we pressed the reset button and again the power button.
Actually the reset button didn't work (maybe ?) it just made some light indicators flashing.
After the second reboot the reboot message said that it needs a manual fsck to fix the file system. This maybe because we interrupted the file checking.
We are leaving it to Jamie because the fsck command would do something bad if unfamiliar persons, like us, do it.
In addition to it, the boot message was also saying that line 37 in /etc/fstab was bad.
We logged into the machine with a safe mode, then found there was an empty line in 37th line of fstab.
We tried erasing this empty line, but failed for some reasons. We were able to edit it by using vi, but wasn't able to save it. |
5317
|
Mon Aug 29 12:05:32 2011 |
jamie | Update | CDS | Re : fb down |
fb was requiring manual fsck on it's disks because it was sensing filesystem errors. The errors had to do with the filesystem timestamps being in the future. It turned out that fb's system date was set to something in 2005. I'm not sure what caused the date to be so off (motherboard battery problem?) But I did determine after I got the system booting that the NTP client on fb was misconfigured and was therefore incapable of setting the system date. It seems that it was configured to query a non-existent ntp server. Why the hell it would have been set like this I have no idea.
In any event, I did a manual check on /dev/sdb1, which is the root disk, and postponed a check on /dev/sda1 (the RAID mounted at /frames) until I had the system booting. /dev/sda1 is being checked now, since there are filesystems errors that need to be corrected, but it will probably take a couple of hours to complete. Once the filesystems are clean I'll reboot fb and try to get everything up and running again. |
5319
|
Mon Aug 29 18:16:10 2011 |
jamie | Update | CDS | Re : fb down |
fb is now up and running, although the /frames raid is still undergoing an fsck which is likely take another day. Consequently there is no daqd and no frames are being written to disk. It's running and providing the diskless root to the rest of the front end systems, so, so the rest of the IFO should be operational.
I burt restored the following (which I believe is everything that was rebooted), from Saturday night:
/opt/rtcds/caltech/c1/burt/autoburt/snapshots/2011/Aug/27/23:07/c1lscepics.snap
/opt/rtcds/caltech/c1/burt/autoburt/snapshots/2011/Aug/27/23:07/c1susepics.snap
/opt/rtcds/caltech/c1/burt/autoburt/snapshots/2011/Aug/27/23:07/c1iooepics.snap
/opt/rtcds/caltech/c1/burt/autoburt/snapshots/2011/Aug/27/23:07/c1assepics.snap
/opt/rtcds/caltech/c1/burt/autoburt/snapshots/2011/Aug/27/23:07/c1mcsepics.snap
/opt/rtcds/caltech/c1/burt/autoburt/snapshots/2011/Aug/27/23:07/c1gcvepics.snap
/opt/rtcds/caltech/c1/burt/autoburt/snapshots/2011/Aug/27/23:07/c1gfdepics.snap
/opt/rtcds/caltech/c1/burt/autoburt/snapshots/2011/Aug/27/23:07/c1rfmepics.snap
/opt/rtcds/caltech/c1/burt/autoburt/snapshots/2011/Aug/27/23:07/c1pemepics.snap
|
16715
|
Tue Mar 8 19:29:36 2022 |
Paco | Update | BHD | Re-balance of AS1 |
[Paco]
Installed AS1 in vacuum, near the center of the table, and installed the OSEMs. All OSEMS are "balanced" nominally, i.e. their shadow is at the halfway point optimum, but fine tuning is required, which I will attempt tomorrow after restoring the AS1 suspension screen settings. Today, I tried damping the SIDE DOF, but didn't succeed, although there was definitely some oscillating behaviour with high (> 5) gains on the damping, so I believe this is a matter of patience. For now, all OSEMs are looking ok, the SOS is in place, and hopefully it will soon be damped. |
16716
|
Wed Mar 9 09:35:26 2022 |
Paco | Update | BHD | Re-balance of AS1 |
[Paco]
AS1 is installed, OSEMs balanced, and the optic damped successfully. We should run the free swinging test overnight to validate this re-installation. |
3798
|
Wed Oct 27 16:15:35 2010 |
Suresh | Update | SUS | Re-glued magnet to the PRM |
Thanh and I re-glued the magnet to the PRM following the procedure outlined by Jenne
The PRM in the gluing fixture has been placed in the little foil house and left to cure for a day.
If all goes well the balancing the PRM will be done tomorrow.
|
11501
|
Wed Aug 12 22:33:36 2015 |
Ignacio | Update | IOO | Re-measured MC2 -> MCL TF |
Since I will need to do transfer function measurements in order to implement FF for the arms and the MC2's yaw and pitch channels, I decided to practice this by replicating the transfer function measurement Eric did for MC2 to MCL. I followed his procedure and the data that I aquired for the TF looked as shown below,

About five minutes of data were taken (0.05 Hz resolution, 25 averages) by injecting noise from 1 to 100 Hz. The TF coherence looked as below,

|
Attachment 1: bode_TF.png
|
|
Attachment 2: Coherence.png
|
|
16678
|
Thu Feb 24 18:05:58 2022 |
Yehonathan | Update | BHD | Re-susspension of AS1 |
{Yehonathan, Anchal, Paco}
Yesterday, Anchal and Paco removed AS1 from the vacuum chamber and moved it into the cleanroom. The suspension wires were cut and the AS1 optic was put on the table.
Two things were noticed:
1. One of the wires was not sitting inside the side block groove (attachment 1)
2. One of the face magnets was grossly tilted (attachment 2). Probably due to uneven polishing of the dumbbell.
We put new wires into the side blocks making sure they sit in their grooves and we removed the tilted magnet. A different, more straight magnet was picked from the remaining spare magnets. The dumbbell and adapter were cleaned from glue residues and a batch of glue was prepared.
In the process of gluing a different magnet was knocked off. We cleaned that magnet too. The 2 magnets were glued on the adapter.
Today I came and saw that the gluing failed completely. One of the magnets was completely away from its socket and the other one wasn't glued at all.
I prepared a new batch of glue and glued the two magnets. |
Attachment 1: signal-2022-02-24-173933_003.jpeg
|
|
Attachment 2: signal-2022-02-24-173933_002.jpeg
|
|
16694
|
Wed Mar 2 14:02:43 2022 |
Yehonathan | Update | BHD | Re-susspension of AS1 |
Yesterday, I rebuilt the OpLev setup in the cleanroom in order to suspend AS1. It took me a while to find all the necessary parts but I found them in the end.
The HeNe laser was placed on the optical table and turned on. The beam was aimed to bounce off a folding mirror to the SOS tower.
The beam's height was controlled by the HeNe laser stage and made to be 5+14/32". The beam from the folding mirror was made parallel to the table, first with an iris and then with the QPD connected to a scope.
Preparing the SOS tower for the suspension I noticed that the wire clamp is scratched on both sides from previous suspensions. I discarded that wire clamp but couldn't find the spares. Time ran out and I had to stop. |
16698
|
Thu Mar 3 17:09:46 2022 |
Paco | Update | BHD | Re-susspension of AS1 |
[Anchal, Paco]
Wire clamp spare was installed, furthermore AS1 was reinstalled on adapter, attached wire clamps, and cleaned using ionized air gun. Finally, we suspended it on the SOS tower and left it resting on the bottom earthquake stops; ready for balancing.
Quote: |
Yesterday, I rebuilt the OpLev setup in the cleanroom in order to suspend AS1. It took me a while to find all the necessary parts but I found them in the end.
The HeNe laser was placed on the optical table and turned on. The beam was aimed to bounce off a folding mirror to the SOS tower.
The beam's height was controlled by the HeNe laser stage and made to be 5+14/32". The beam from the folding mirror was made parallel to the table, first with an iris and then with the QPD connected to a scope.
Preparing the SOS tower for the suspension I noticed that the wire clamp is scratched on both sides from previous suspensions. I discarded that wire clamp but couldn't find the spares. Time ran out and I had to stop.
|
|
16710
|
Mon Mar 7 16:56:08 2022 |
Yehonathan | Update | BHD | Re-susspension of AS1 |
{Paco, Yehonathan}
We tried to roughly balance the adapter with two counterweights at the front, like with the other thin optics using an iris. As before, we couldn't get the beam above the iris hole no matter how much we inserted the counterweights into the adapter. We noticed that one of the side blocks is actually the one where the clearance for the wire was made on the wrong side. So there was clearance on both the up and bottom sides of the side block (see attachment 1).
Could this be the cause of the balancing issue? Running out of ideas on how to fix it we gave it a try and replaced it with a spare side block. We also found that the wire on the other side block was kinked so we replaced the wire on this one as well.
After inserting new wires into the side blocks, we hung the adapter on the winches and the beam was above the iris aperture! How could this tiny amount of missing mass make this much difference?
We were able to roughly balance the adapter.
We then tried to balance the roll of the adapter but accidentally knocked off the side magnet 😫.
We usually glue several side magnets together and they all together support the metallic plate on which the magnets are magnetically attached to. This time we had only one side magnet to glue so instead of trying to glue the magnet vertically we are trying to glue it horizontally using a flat surface and a stage to clamp it (attachments 2,3).
BTW, the HeNe was not working when we came into the cleanroom. We realized it was the old HeNe that we already determined to be broken but there was no sign on it. I attached a "BAD" sign on it and replaced it with the new HeNe. The OpLeve beam was realigned. All of this happened before all the things described above |
Attachment 1: signal-2022-03-07-171520_001.png
|
|
Attachment 2: signal-2022-03-07-172659_001.jpeg
|
|
Attachment 3: signal-2022-03-07-172659_003.jpeg
|
|
16711
|
Mon Mar 7 18:53:16 2022 |
Koji | Update | BHD | Re-susspension of AS1 |
Not sure if that small difference can cause the alignment inability. Particularly, the removed metal was just below the wire. This means that there is no misalignment effect at the first order.
Here is my idea:
You may be able to assist the alignment by adding washers on one side of the four holes to this "H" shaped parts. The holes are away from the center line, adding some weight definitely do some misalignment.
|
16714
|
Tue Mar 8 12:24:13 2022 |
Yehonathan | Update | BHD | Re-susspension of AS1 |
The gluing seemed to be successful. I assembled the side block with the magnet on the adapter. Paco helped me hang the adapter on the SOS tower.
The height and roll of the adapters were balanced (attachment 1,2).
The QPD was placed at the beam reflection. The beam was centered horizontally on the QPD and then measured vertically. The pitch DOF was balanced using the counterweights. The counterweight was locked. Balance was retained.
I tried to assemble the upper mirror clamp on the tower but for some reason, one of its tap holes was not able to accept screws. I gave it to Jordan for retapping. I measured the motion spectrum using the QPD connected to a scope (attachment 3).
Major peaks are at 668mHz, 942mHz, and 1029mHz.
|
Attachment 1: AS1_Roll_Balance.png
|
|
Attachment 2: AS1_Height_Balance.png
|
|
Attachment 3: FreeSwingingSpectra_new.pdf
|
|
3823
|
Fri Oct 29 14:06:12 2010 |
kiwamu | Update | Green Locking | Re: 80MHz VCO for green PLL : VCO calibration |
P.S. There is a document about the 80MHz VCO box. This may be helpful.
link to LIGO DCC |
14014
|
Mon Jun 25 19:14:02 2018 |
Udit | Summary | General | Re: A summary of the Tip-TIlt Mirror Holder design changes |
2. Weighted screw rod at the bottom for tilting the mirror-holder:
The screw length selected here (2") is not interfering with any part of the assembly.
The 'weights' I have here are just thumb nuts from Mcmaster, so their weight is fixed (1.65g each, btw).
Problem I'd like to solve: Find an assortment of weighted, symmetric nuts with caps on one end to fix position on shaft.
3. Set-screws on both side of wire clamp to adjust its horizontal position:
Thanks for pointing out the mismatch in travel distance of protrusion and clamp screws. To match them, the clamp screw slot now sticks out of the profile (by 1.5mm). The range of the clamp motion is +/- 3 mm.

Also, here's a screenshot of the slot in the mirror holder:

--
- Excluding the weighted screw rod assembly, the height gap between assembly COM and wire release point is 3.1 mm.
Quote: |
> 2. Weighted screw rod at the bottom for tilting the mirror-holder:
Too long. The design of the holder should be check with the entire assembly.
We should be able to make it compact if we heavier weights.
How are these weights fixed on the shaft?
Also can we have options for smaller weights for the case we don't need such a range?
Note the mass of the weights.
> 3. Set-screws on both side of wire clamp to adjust its horizontal position:
How much is the range of the clamp motion limited by the slot for the side screws and the slot for the protrusion? Are they matched?
Can you show us the design of the slot made on the mirror holder?
>>
Where is the center of mass (CoM) for the entire mirror holder assy and how much is the height gap between the CoM and the wire release points. Can you do this with 3/8" and 1/2" fused silica mirrors?
|
|
14015
|
Mon Jun 25 21:14:08 2018 |
Koji | Summary | General | Re: A summary of the Tip-TIlt Mirror Holder design changes |
3.
- Do we need this much of extended range of the clamp location? How much range will we need if we use either 3/8 or 1/4 inch mirrors?
- This slot on the mirror holder ring is not machinable.
About the CoM height
- Include the angle adjustment screw and adjust the wire releasing point to have comparable pitch resonant freq to the SOS suspension.
|