Wed Aug 13 03:03:37 2014 |
ericq | Update | General | Green beatnote troubles | [Jenne, Rana, ericq]
No luck locking tonight, as spent a while trying to figure out the complete absence of the green beatnotes. Long story short, we ended up having to adjust the pointing on the PSL table.
Unrelated to this, we also turned on the noise eater on the PSL laser because why not.
We hooked the BBPDs directly up to a 300MHz scope to try to see the beat as it happened. We witnessed a very strange intermittent ~800MHz oscillation on the Y BBPD, and weirder still, on both the RF and DC outputs of the PD, and the frequency was independent of the laser temperatures. This is to be investigated in the future, but was not related to the beat note state.
Some progress was made when we took some components out, and looked at the far field of the PSL-Ygreen overlap, and saw some misalignment, and corrected it. Putting the end laser temperature in the usual area allowed the beat note to be found, with the eventual amplitude of ~-40dBm directly out of the BBPD. The Y green alignment was pretty bad throughout, so this can be improved to bring the beat amplitude up. We should also check and make sure we're well aligned to the SHG with the PSL light. We're leaving the X beat for tomorrow, now knowing that we should be able to get it with careful alignment. |
10390
|
Thu Aug 14 18:31:45 2014 |
ericq | Update | LSC | LSC Modeling Update | Based on the game plan, I have created a slew of updated pretty plots about our signals and loops.
First: With measured arm losses, when do we start to see REFL DC dip? At what arm buildup powers?
I updated my MIST model with the arm losses I've measured (Y:130ppm, X:530ppm), and some measured transmissions from the wiki, vs. the design parameters, as I used to have. Here is the DC sweep plot which is now hanging up in the control room.

In this plot, I also calculated what MIST thinks the full arm power buildup will be as compared to our single arm locking, and I get something of order 200, rather than the 600 we've tossed around in discussions. Nothing else is very different in this plot from the old version; though the REFLDC dip is a little bit wider.
Now, here are some radiation-pressure inclusive sensing transfer functions, for the anti-spring case (which in Rob's day was easier to lock for unknown reasons):
 
 
Next: Include new AO path TFs into CM model Look at possibilities for engaging AO path
With these TFs, and the recently measured+fit new AO TF, here are the open loop gains of the slow, digital, SqrtInv-sensed MCL CARM and fast, analog, REFLDC-sensed AO CARM loops for the region of offsets we've achieved and a little lower. The slow digital loop includes the 1k LP that we have used in the past, in addition to the normal CARM filters. I still need to figure out the right sequence of ( offset reduction / crossover frequency motion / overall gain adjustment ) that gets the coupled cavity resonance solidly within the loop bandwidth.
|
10402
|
Fri Aug 15 14:35:57 2014 |
ericq | Update | LSC | TRY mystery offset gone | One question answered, but another raised. The offset came from LSC-TRY switching to the ETMY-QPD signal from ETMY-TRY (Hi gain pd).
BUT WHY

|
10404
|
Fri Aug 15 20:26:37 2014 |
ericq | Update | General | Game plan |
Quote: |
Q already did the tweak up of the PSL SHG crystal alignment. HE SHOULD ELOG ABOUT THIS. What was the final power of green that you got? Do we have any record of a previous measurement to compare to?
|
As Jenne mentioned, I did this.
Specifically, I first tweaked the mirror pointing the IR into the SGH in pitch and yaw to maximize the green power, and then adjusted the little set screws on the side of the SHG to maximize further. Power after the harmonic separator was of order 150uW. On the Y Green BBPD, I got ~48uW, instead of the 40uW Rana, Jenne, and myself saw the other night.
HOWEVER,
now that I look through old ELOGs, I find some posts by Kiwamu saying the power should be around 650uW, and that he was able to get 640uW out. So: I should do this again, systematically, more carefully, etc., etc. (Linked ELOG also states that optimum SHG temperature is alignment dependent...)
|
10405
|
Fri Aug 15 20:38:17 2014 |
ericq | Update | General | ELOG dump | A few things that I have neglected to ELOG yet:
-
scripts/offsets/LSCoffsets is a new script that uses ezcaservo to set FM offsets of our LSC PDs. It still warns about large changes, and lets you revert. It reads the FM gain to pick the right gain for the ezcaservo call.
-
MC refl DC was all over the place today, and has recently been "fuzzier" on the wall StripTool than I like. I touched the MC2 pointing a little bit, and the WFS seemed to find a sweet spot where the refl got steady back at around and under 0.5. I then ran "offload WFS" to try and stay there.
-
Incidentally, the PMC transmission drifted up to 0.81 at some point today. This is weird, since not too long ago, we were not able to reach this level even with careful alignment. This coincided with the MC power being back up to ~17k, and arms locking at around 0.95.
-
Last week I quickly tried cranking up the x-end green modulation frequency to ~1.3MHz (corresponding to a notch in the PZT AM response), and using a 550k lowpass on the mixer output, instead of a 70k, to try to buy more phase and increase the UGF. It didn't work. I didn't have a way to tune the mixer phase angle, and the mixer output was super noisy, but there were instants where I could convince myself that a mode was briefly locked to the arm... I'm going to do the Right Thing and characterize the loop properly, to figure out how to get at least 10kHz of control bandwidth out of these things.
|
10407
|
Mon Aug 18 18:33:57 2014 |
ericq | Update | Green Locking | Yarm Green PDH | So far today, I've been working with the Y-end green PDH locking. Using a SR560 to roll off the AG4395A output to take a loop measurement at the servo output, I measured the following OLG, and inferred the CLG from it. The SR560 really helped it getting good coherence without introducing a big offset that changes the optical gain, thus distorting the loop shape, etc. etc.

You would think this loop looks pretty good, 10k UGF, and 45 degrees of phase margin, gain peaking is sane, and pretty smooth slope. But, the thing still was flipping out of lock while I measured this.
I suspect shenanigans at >100k. This is motivated by the fact that I've seen some big noise in the error signal around 150k. I don't have a good noise plot right now, because I'm trying to get a scheme going where I stitch together a bunch of 1 decade spectra from the 4395, but the noise floor isn't consistent across each patch (even though the attenuation stays the same, and I confirmed I'm in "noise" mode). I'm working on a loop measurement up there, too, but I haven't been able to get the right filter/amplitude settings yet.
So, even though this plot is not totally correct (read: wrong and bad), I include it just for the sake of showing the big honking spike of noise at ~150K.

|
10409
|
Tue Aug 19 18:32:40 2014 |
ericq | Update | Green Locking | Yarm Green PDH | Heading to dinner, going to come back for more green fun, but here's a quick update:
Xarm Peak-to-Peak of the PDH signal in the mixer output is about 70mV when GTRX was about 0.4. The sideband-generating function generator has an output of 2V (forgot to note rms or pp)
Yarm Peak-to-Peak of the PDH signal in the mixer output is about 640uV when GTRX was about 0.71. The sideband-generating function generator has an output of 0.091V (forgot to note rms or pp)
The Yarm signal thus correspondingly has a waaay noisier trace. I would've had scope plots to show here, but the scope freaked out about how large my USB drive capacity was and refused to talk to it >:|
This suggests to me that our modulation depth for the Yarm may be much too small, and may be part of our problems with it. |
10413
|
Wed Aug 20 04:09:21 2014 |
ericq | Update | Green Locking | Xarm Green PDH | I've made a whole bunch of measurements on the Xarm green situation.
TL;DRs:
- GTRX was around 0.55 for all of the measurements tonight.
- Based on where I saw gain peaking in the CLG, it looked like UGF was 1-2kHz. I cranked the gain to 10kHz, ~20dB gain peaking followed, making it hard to measure. Currently sitting at 5kHz-ish.
- Measured CLG with AG4395A, calibrated for injection point response, inferred OLG.
- Took various PSDs, still need to calibrate into physically meaningful units.
Reasonable amounts of time were spent bending the AG4395 to my will; i.e. figuring out the calibration things Jenne and Rana did, finding the right excitation amplitude and profile that would leave the light steadily locked, and finding the right GPIB incantation for getting spectra in PSD units instead of power units. I'm nearing completion of a newer version of AG4395 scripts that have proper units, and pseudo-log spectra (i.e. logarithmically spaced linear sweeps)
Transfer functions
Here is too many traces on one plot showing parts of the OLTF for the x green PDH. One notable omission is the PD response (note to self:check model and bandwidth). The servo oddly seems to have a notch around 100k. My calibration for the CLG injection may not have been perfect, instead of flattening out at 0dB, I had 2dB residual. I tried to correct for it after the fact, assuming that certain regions were truly flat at 0dB, but I want to revisit it to be thorough. I found some old measurements of the Innolight PZT PM response, which claims to be in rad/V, and have included that on the plot.

In the end, the mixer and PZT response make it look like getting over 10kHz bandwidth may be tough. Even finding a good higher modulation frequency to be able to scoot the LP up would leave us with the sharp slope in the PZT phase loss, and could cause bad gain peaking. Maybe it's worth thinking about a faster way of modulating the green light?
Noise Spectra
Tomorrow morning, I'll calibrate all the noise spectra I have into real units. These include:
- In loop error signal and control signal spectra
- Mixer output spectrum when PD is dark, and when mixer input is terminated
- Servo out spectrum when PD is dark, and when servo input is terminated
However, looking at the floors, it occurs to me that I may have left the attenuation on the input too high, in an effort to protect the input the PDH box, which rails all the time when not locked to a 00 mode, sometimes even with the input terminated or open. It's kind of a pain that the agilent makes it really hard to see the data when you're in V/rtHz mode, because I should've caught this while measuring :/
I used a scope to capture a pdh signal happening, which will let me transform the mixer output into cavity motion. The control signal goes to the innolight PZT with a ~1MHz/V factor. Here are the uncalibrated plots, for now.


|
10414
|
Wed Aug 20 15:31:27 2014 |
ericq | Update | COC | Arm Loss Investigations Continue | [ericq, Gabriele]
Summary: After today's meeting, Gabriele and I looked into the arm loss situation, to see if we should really believe the losses that had been suggested by my previous measurements. We made some observations that we're not sure how to explain, and we're thinking about other ways to try and estimate the losses to corroborate previous findings.
We first looked to see if the ASS had some effective offset, leaving the alignment not quite right. Once ASS'd, we twiddled each arm cavity mirror in pitch and yaw to see if we could achieve higher transmission. We could not, so this suggested that ASS works properly.
We then looked at potential offsets in the Xarm loop. We found that an input offset of 25 counts increased the transmission, but only very slightly. With this offset adjusted, we confirmed the qualitative observation that locking/unlocking the xarm causes a much bigger change in ASDC than doing the same with the harm.
However, we noted that the ASDC data (which is the DC value of the AS55 RFPD) was quite noisy, hovering around 50 counts. Looking at the c1lsc model, we found that we were looking at direct ADC counts, so the signal conditioning was not so great. We went to the LSC rack and stole the SR560 that had been hooked up as a REFLDC offsetter, and used it to give ASDC a gain of 100, and a LP at 100Hz, since we only care about DC values. We then undid the gain in the input FM; and this calmed the trace down a fair bit. The effects due to each arm locking/unlocking was still consistent with previous observations.
At this point, we looked at the arm transmission and ASDC signals simultaneously. Normally, when misaligning a cavity, one would expect the reflected power to rise and the transmission to fall.
However, we saw that when misalignment the Yarm in yaw in either direction, or the Xarm in one direction, both the IR transmission and ASDC would fall. This initially made us think of clipping effects.
So, we checked out the AS beam situation on the AP table. On a card, the beam looks round as we could tell, and the beam spot on AS55 was nice and small. (We tweaked its steering a little bit in pitch to put it at the center of the "falling-off" points) The reflection and transmission falling effect remained.
At this point, we're not really sure what could be causing this effect. After the reflected beams recombine at the BS, the output path is common, so it's strange that this odd effect would be the same for both arms.
Lastly, we discussed other ways that we may be able to see if the Xarm really has ~500ppm loss. Since its transmission is ~1.4%, Gabriele estimated that we may be able to see a ~300Hz difference in the arm cavity pole frequency between the two arms, based on the modification of the cavity finesse due to loss. Since we don't currently have the AOM set up to inject intensity noise, we talked about using frequency noise injection to measure the arm cavity poles, though this would be coupled with the IMC pole, but this could hopefully be accounted for. |
10415
|
Wed Aug 20 16:10:43 2014 |
ericq | Update | Green Locking | Xarm Green PDH | A MIST simulation tells me that the green pdh horn-to-horn displacement is about 1.2nm, or ~18kHz. I used this, along with the scope trace attached to the previous post, to calibrate the mixer output at 193419 Hz per V. (EDIT: I was a little too hasty here. What I'm really after is the slope of the zero crossing, which turns out to be almost exactly twice my earlier naïve estimate. See later post for correct spectra)
For the control signal, I assumed a flat Innolight PZT PM response of 1MHz/V. ( Under 10kHz, it is indeed flat, and this is the region where the control signal is above the servo output noise in yesterday's measurements)
Here are all of the same spectra from last night, with the above calibrations.

Going off Jenne's earlier plot, it looks like the in-loop error signal RMS is ten times bigger than the CARM linewidth. |
10417
|
Wed Aug 20 21:09:16 2014 |
ericq | Update | Green Locking | Xarm Green PDH | I remeasured all of the noise spectra again today, making sure the input attenuation was as low as it could safely be. I also got a snap of the y green PDH signal; it's fairly larger than I saw the other day, which is good. I used this to calibrate the error signal voltage spectra.

Here are the noise traces for each arm. During these measurements GTRX was about .6, GTRY about 1.0 The Yarm noise doesn't look so good: the error signal is just barely above the mixer+lowpass output noise, and the RMS is plauged by 60Hz lines. (Is this related to what we see in IR TRY sometimes?)
 
Here are the arms error signals compared directly:

|
10420
|
Thu Aug 21 19:04:52 2014 |
ericq | Configuration | Green Locking | Gain changes on Green Y PDH | I found that the barrel of one the BNC to BNC connectors used for getting the output of the PDH servo box to the laser controller was touching the ETMY chamber. When I held it away, all of the 60Hz harmonics disappeared from the mixer output spectrum; this was pretty repeatable. This inspired me to replace the refl PD and PZT signal cables (which were 2 and 3 cables stitched together, respectively) with 20' long BNCs. I also cleaned up a lot of the routing of signal and power cables in the little rack, and moved the big T->DC Block->Attenuator combo off of the panel mount, because I didn't like how it was wiggling. It and the summing pomona box are sitting on top of the PDH box and function generator, instead of hanging freely.
All of the 60Hz harmonics were banished afterwards, and the green locked happily. 
This required me touching the Y end table, to remove the old cable and its cable ties, and putting the new one in. I don't think I did anything immediately apparently bad; the green and IR transmissions both are within nominal ranges.
I haven't had luck measuring the CLG yet, which I wanted to do to get and set the UGF before measuring the noises. However, here is a scope trace of the in-lock error signal, which compares quite favorably to the trace posted in the previous post; the scope indicates that the signal has 1/3 of the RMS that it did before I replaced the cables.

I hope to measure up the current status after I get back from dinner.
|
10430
|
Tue Aug 26 23:16:49 2014 |
ericq | Configuration | Green Locking | Gain changes on Green Y PDH | Yesterday I measured the spectra and OLTF of the Y-Arm green PDH, after the LO touch-up and 60Hz hunt from last week. I also went to lower frequencies with the SR785, but forgot to take some of the background spectra down there, so I don't have the full breakdown plots yet. Nevertheless, here is the improvement in the PDH error signal:

I also measured the OLTF (SR785 injection at the error signal, Auto level ref 5mV at channel 2, 10mV/s source ramping, 50mV max output)

As you can see, we have tons of phase margin. Flipping the local boost switch had no visible effect on the OLTF; we should change it to something that puts this surplus of phase to good use, and squash the error signal even more. Putting an integrator at 5kHz should still leave about 45 degrees phase margin at 10k. I've started making a LISO model of the PDH board from the DCC drawing, and then I'll inspect the boards individually to make sure I catch the homegrown modifications.
Data, and code used to generate the plots is attached. |
Attachment 3: newY.zip
|
10431
|
Tue Aug 26 23:46:55 2014 |
ericq | Update | IMC | WFS tuneup | I decided to see what I could do with the new WFS setup.
First, I adjusted the WFS digital demod angles. Once I ensured that the static MC alignment and DC alignment onto the WFS was good, I drove MC2 in pitch with the WFS output off. I then did the usual thing of making the Q peak at the excitation frequency go away. Here are the changes:
- WFS1 Q1: 7 -> -24 (-31)
- WFS1 Q2: 6.5 -> -9.5 (-16)
- WFS1 Q3: -6.5 -> -26.5 (-20)
- WFS1 Q4: 47 -> 30 (-17)
- WFS2 Q1: -51 -> -39 (-12)
- WFS2 Q2: -39 -> -21 (-18)
- WFS2 Q3: -32 -> -20 (-12)
- WFS2 Q4: -120 -> -108 (-12)
I then drove each MC mirror in pitch and yaw respectively, and measured the TF from excitation to the WFS signal (dB Magnitude, sign):
Mirror |
DoF |
WFS1 Pitch |
WFS1 Yaw |
WFS2 Pitch |
WFS2 Yaw |
MC1 |
Pit |
-67.7,+ |
-81.9,+ |
-58.9,- |
-83.7,+ |
|
Yaw |
-82.5,- |
-48.7,- |
-83.7,+ |
-112.3,- |
MC2 |
Pit |
-50.4,- |
-77.1,- |
-54.2,- |
-67.9,+ |
|
Yaw |
-82.1,- |
-52.9,+ |
-59.6,- |
-44.0,- |
MC3 |
Pit |
-59.7,- |
-97.3,+ |
-62.0,+ |
-83.9,- |
|
Yaw |
-78.0,+ |
-52.9,+ |
-67.3,+ |
-51.4,+ |
I looked through some old ELOG's of Suresh's and used similar logic to scripts/MC/WFS/wfsmatrix2.m to generate a new output matrix. (This involves creating a null sensing vector that is orthogonal to the measured ones, and inverting that matrix)
Old:
Pitch |
WFS1 |
WFS2 |
MC2T |
|
YAW |
WFS1 |
WFS2 |
MC2T |
MC1 |
-1 |
0.044 |
0 |
|
MC1 |
-1 |
-0.294 |
0 |
MC2 |
0.19 |
1 |
1 |
|
MC2 |
-0.26 |
-0.045 |
-1 |
MC3 |
0.5 |
-0.681 |
0 |
|
MC3 |
-.9 |
1 |
0 |
New:
Pitch |
WFS1 |
WFS2 |
MC2T |
|
YAW |
WFS1 |
WFS2 |
MC2T |
MC1 |
0.835 |
-1 |
0 |
|
MC1 |
-1 |
-0.229 |
0 |
MC2 |
-0.948 |
-0.433 |
1 |
|
MC2 |
0.317 |
-1 |
-1 |
MC3 |
-1 |
0.865 |
0 |
|
MC3 |
0.743 |
0.628 |
0
|
I had to flip a gain or two to keep things stable, then measured the WFS error signal spectra to see if this made anything better. The WFS1 spectra look better, but WFS2 not so much.

The loops would need a more thorough investigation, but for now, they're at least a little calmer. The MC is stabler than immediately after the upgrade, but there's still room for improvement.
|
10433
|
Wed Aug 27 18:03:47 2014 |
ericq | Update | Green Locking | uPDH Box Checkup | Quick post of plots and data; I'll fill in more detail tonight.
TL;DR: I pulled both green PDH boxes and made LISO models, compared TFs and noise levels.
Pictures of X and Y boards, respectively
 
TF comparison to LISO. (Normalized to coincide at 1Hz)

Noise comparison to LISO

To Do:
- Figure out why TFs were made differently. Check PM response curves of PZTs to see if they are fine, or need tweaking.
- Make boosts useful. Both are currently integrators with corners under 10Hz, which is already pretty suppressed.
- I just noticed that the X board is missing C25, which should be a 1uF cap on the positive power pin of the primary TF stage opamp. This should be inserted.
All data, EAGLE schematics, LISO source and plots in the attached zip.
|
Attachment 5: uPDH_Aug27.zip
|
10434
|
Thu Aug 28 01:41:03 2014 |
ericq | Update | LSC | Phase Tracker UGF normalization | We want both the X and Y phase trackers to have the same UGF, so that the X and Y ALS signals are subject to the same phase characteristics and can be nicely decoupled into CARM/DARM.
I've started implementing a simple normalization scheme that Koji suggested, namely, dividing the I output of the phase tracker by a low passed version of the Q output. (Since the I is servoed to zero, the radius of the error signal in the IQ plane is essentially equal to the Q value) I put some simulink logic into the IQLOCK library part that BEAT[XY]_FINE are instances of to switch the normalization on/off, and to protect from divide-by-zeros. I also exposed the switching and FM on the ALS screen.

I then tried using it, to mediocre results. I put a 10mHz LP in the filter module, found a Y-Arm beat, set the phase tracker gain to give me a 2kHz UGF, and then set the gain of the UGH normalization FM to turn the current average Q to unity.
I then moved the laser temperature around to get different beatnote locations/amplitudes, hoping that the phase tracker UGF would stay the same when the UGH normalization was on.
It did not.
It did, however, correct it in the right direction... more work will be done with this, to try and make it useful. There's also the unfortunate effect that locking/unlocking the green causes erratic phase tracker output, which messes with the input to the normalizing LP filter, so if one were to leave it switched on, wonky stuff would come out. I don't want to go overboard with triggering shenanigans before I even get it working in the first place, though. |
10437
|
Thu Aug 28 17:34:20 2014 |
ericq | Update | Green Locking | uPDH Box Checkup | I had noticed in the past, that the digital control signal monitor for the X end would saturate well before the ADC should saturate (C1:ALS-X_SLOW_SERVO_IN1, which is from the "output mon" BNC on the box). It turns out that there is some odd saturation happening inside the box itself.
In this scope trace, the servo input is being driven with a 0.02Vpp, 0.1Hz sine wave, gain knob at 1.0. This is bad.

Evan and I poked around the board, and discover that for some reason currently unknown to us, the variable gain amplifier (AD8336) can't reach its negative rail, despite the +-12V arriving safely at its power supply pins.
I also realized that the LF356 in the integrator stage in this box had been replaced with a LT1792 by Kiwamu in ELOG 4373. I've updated my schematic, and will upload both boxes' schematics to the DCC page Jenne created for them. (D1400293 and D1400294) |
10439
|
Fri Aug 29 23:53:35 2014 |
ericq | Update | Green Locking | uPDH Box Checkup | I had pulled out both X and Y servo boxes for inspection, put the Y box back, soldered in a missing op amp power capacitor on the X end box, and had not yet put back the X end box yet because of the saturation issue I was looking into. Otherwise nothing was changed at the ends; I didn't open the tables at all, or touch laser/SHG settings, just unplugged the servo boxes. |
10441
|
Tue Sep 2 20:02:06 2014 |
ericq | Update | Green Locking | uPDH Box Checkup | I narrowed down the saturation point in the X green PDH box to the preamp inside the AD8336, but there is still no clear answer as to why it's happening.
As per Jenne's request, I put the X end PDH box back for tonight's work. It locks, but we have an artificially low actuation range. With SR785, I confirmed a PDH UGF around 5k. Higher than that, and I couldn't reliably measure the UGF due to SR560 saturations. The analyzer is not currently in the loop.
Both arms lock to green, but I haven't looked at beatnotes today. |
10447
|
Wed Sep 3 20:38:40 2014 |
ericq | Update | Green Locking | uPDH Box Checkup | The traces were from the front panel output BNCs, but the VGA preamp exhibited this asymmetric saturation at its output.
In any case, I tried to replace the Xend box's AD8336 with a new one, and in doing so, did some irreparable damage to the traces on the board I was not able to get a new AD8336 into the board. There are some ATF ELOGs where Zach found the AD8336 noise to be bad at low frequencies (link), and its form factor is totally unsuitable for any design that may involve hand modification, since it doesn't even have legs, just tiny little pads. I suggest we never use it for anything in the future.
Instead, I've hacked on a little daughter board with an OP27 as an inverting op-amp with the gain resistor on the front panel as its feedback resistor, which can swing from 0 to x20 gain (the old gain setting was around 15dB=~x6). I've checked out the TF and output noise, and they look ok. The board can output both rails as well.
I don't really like this as a long term solution, but I didn't want to leave things in a totally broken state when I left for dinner. |
10449
|
Thu Sep 4 01:28:32 2014 |
ericq | Update | LSC | Recycling cavity lengths | Going off some discussion we had at lunch today, here is my current knowledge of the state of cavity lengths.
Acknowledging that Koji changed the sideband modulation frequency recently, the ideal cavity lengths are (to the nearest mm):
- Lprc = c / ( 4 * fmod) = 6.773 m
- Lsrc = c / ( 5 * fmod) = 5.418 m
We when last hand measured distances, after moving PR2, we found:
- Lprc = 6.752 m = 2.1 cm short
- Lsrc = 5.474 m = 5.6 cm long.
However, when I looked at the sideband splitting interferometrically, I found:
- Lprc = 6.759m = 1.4 cm short
This is only 5mm from the hand measured value, so we can believe that the SRC length is between 5 and 6 cm too long. I'm building a MIST model to try and see what this may entail. |
10450
|
Thu Sep 4 03:12:55 2014 |
ericq | Update | LSC | Green PDH box boosts | Jenne made her board modifications, and the measured TF agreed with the design. Alas, the green would not lock to the arm in this state.
I think that the reason is that the new TF does not have nearly as much low frequency gain as the old one, for a given UGF. Thus, for example, the 1Hz noise due to the pendulum resonance, has 30dB less loop gain suppressing it.

NEED MORE 
|
10453
|
Thu Sep 4 18:16:20 2014 |
ericq | Update | LSC | Recycling cavity lengths | Koji correctly points out that I naïvely overlooked various factors. With a similar analysis to the wiki page, I get:
- Ideal arm length of 37.795 m
- Ideal PRC length of 6.753 m
- Ideal SRC length of 5.399 m
This means that:
- The PRC, measured at 6.759m, is 6mm long.
- The SRC, measured at 5.474m, is 7.5 cm long
Next step is to see how this may affect our ability to sense, and thereby control, the SRC when the arms are going.
MIST simulations and plots are in the attached zip. |
Attachment 1: 2014-09-CavityLengths.zip
|
10457
|
Fri Sep 5 04:07:44 2014 |
ericq | Update | Green Locking | X end uPDH Box Replaced | Just a quick note, plots and data will come tomorrow:
I grabbed an unused uPDH board from the ATF (thanks Zach!), and re-stuffed almost the entire thing to match Jenne's latest schematic for the y end box. I also threw some 22uF caps on the regulators, as Koji did with the previous box, to eliminate some oscillations up in the high 10s of kHz. I replaced the tragedy of a box that I created on Wednesday with this new box. The arm locks pretty stably with the boost on, 30 degrees of phase margin with 10kHz UGF, and locks pretty darn reliably.
Now we should now have two nicely boosted PDH loops. I'll do a noise/loop breakdown again in the upcoming days. |
10460
|
Fri Sep 5 16:13:00 2014 |
ericq | Update | Green Locking | new X Green PDH measurements | I measured the noise spectra and loop TF of the green PDH with the newly stuffed board. Unfortunately, I never took the noise below 100Hz of the previous box, so we can't see what has happened to the overall RMS, or more specifically, the RMS due to the pendulum resonance. All of these plots are in the boosted state, as that is how we intend to use the box.
Here is the loop, which does not have quite as much margin as the y-arm, but 10dB of gain peaking is probably ok, since the RMS at 10s of kHz is not so important to ALS. (OL measured, CL inferred) We see the 1/f shape from 1k to 50k or so, and 1/f^2 under 1k, as desired.

Comparing in the in loop error signals, we see the effect from the increased gain from 100Hz to 10kHz. (Here is where I regret not looking at the low frequency spectrum two weeks ago)

Finally, here is the noise breakdown.

The error signal RMS is now dominated by the 1Hz peak. We have talked about using digital feedback for this, since we have the PDH error signal coming into an ADC, and can sum in a DAC signal into the servo output. This also lets us intelligently trigger a sub-10Hz boost once the PDH box locks itself. With a good boost, we maybe could bring the in-loop RMS of the error signal to under 1kHz.
Something odd that Rana brought to my attention, however, is that my measurement and calibration indicates an RMS of ~5kHz, but the cavity pole should be something like 18kHz. If this is true, how can we be seeing stable power? This maybe means that my calibration is too many Hz per Volt.
I performed the calibration by creating a MIST model of the arm, and generating the PDH error signal on a demodulated PD, I then find the slope of Hz per arbitrary error signal unit. Then, looking at a scope trace, I match up the horn-to-horn voltage to the horn-to-horn arbitrary error signal units, which lets me finally find Hz per error signal volt.
However, there is some qualitative difference in the shape between the simulated and observed error signals, namely, that the outer horns are larger than the inner horns in the real signal.
 
Does this matter? Is there something in my simulation that I can correct that would give a more accurate calibration?
Data, plots, code, attached. |
Attachment 6: xG_Sep5.zip
|
10464
|
Sat Sep 6 14:49:12 2014 |
ericq | Update | Green Locking | new X Green PDH measurements | I used a modulation depth of 0.3, which, if I recall correctly, is what we aimed for on the Y-arm when we adjusted the LO signal there. However, this is probably not the case for the X arm.
In any case, I found the bug in my RMS calculation. (I had forgotten to flip the x array in addition to the y array for the right-to-left integration, and had uneven bin spacing, so the integration bandwidths weren't correct...)
Here are the updated plots. The properly evaluated RMS is ~600Hz, which seems to mostly come in around 10k, so we may want to turn down the gain for less gain peaking in that region.

|
10468
|
Mon Sep 8 11:10:26 2014 |
ericq | Update | Computer Scripts / Programs | preparing vac system to reboot |
Quote: |
Q and Steve will follow elog 10028 entry to prepare the vacuum system for safe reboot
|
Here's the sequence of the morning so far:
- I aligned the IFO (IR arms with ASS, X green with PZTs, PRM with PRMI locked on REFL33)
- I closed the PSL shutter, and went inside to align PRM and both ITM oplevs (all others were within 10urad of zero in both directions)
- While aligning those oplevs, I noticed the smell of burnt electronics. We tracked it down to the +15V sorensen in the rack nearest the PSL table
- I claim the precipitating event was PSL shutter activity. If I recall correctly, the seismic rainbow traces went bonkers around the same time as the shutter was closed. There is a Guralp interface in the rack powered by the failed sorensen, so this would explain the erratic seismometer signals correlated with the power supply failure. We will look into potential shorts caused by the shutter. (Steve looked up the PMC trans and Guralp DQ channels, and confirmed the temporal coincidence of the events.)
- We shut off all of the sorensens so that electronics were not being driven asymmetrically.
- Steve and I secured the vacuum system for computer reboots, as referred to in Steve's elog. Some combination of Jenne, Rana and Manasa shut down the control room computers, and turned off the watchdogs.
- Manasa and I moved Chiara inside, next to Mafalda, along with its backup HDs. It has been labeled.
- Booted up control room machines, they came up happy.
- FB and front-ends didn't need reboot, for some lucky reason. Watchdogs came back happily, oplev spots didn't move noticeably.
The IFO is still down, as the PMC won't lock without the rack power, and we haven't pinned down the shorting mechanism. We don't want the replacement sorensen to immediately blow when plugged in. |
10472
|
Mon Sep 8 15:50:47 2014 |
ericq | Update | Computer Scripts / Programs | +15V PSL Sorensen replaced | We replaced the +15V sorensen at 1X1, and brought the power supplies back up symmetrically, and everything seems fine. I noted that a quarter turn counter-clockwise took the current limit down by one amp, so I set the knob to just letting 2.8A (the nominal current), and then added one half turn, shooting for ~4A current limit.
In doing so, we had to cut power to the c1psl VME box. It didn't come back happily. We had to do the chiara /etc/hosts things, like we did for c1auxexx, to get it back. |
10484
|
Wed Sep 10 02:52:05 2014 |
ericq | Update | CDS | OTTAVIA lost network connection | I checked chiara's tables, all seemed fine. I switched ethernet cables from the black one labelled "allegra," which seemed maybe fragile, for the teal one that may have been chiara's old ethernet cable. It's back on the network now; hopefully it lasts. |
10486
|
Wed Sep 10 02:59:42 2014 |
ericq | Update | LSC | Locking activities - nothing new :( | Some small things I did tonight which did little to nothing to help:
- I reset the offsets in the SQRTINV FMs to try and match the DC level of the ALS CARM error signal as best as possible, to avoid moving away from the set-point too much, as I was worried we were wandering into regions of too low optical gain.
- I turned off the WFS, and hand tweaked the MC alignment. The WFS loops / matrices definitely have some room for improvement, and I was worried that excess angular motion of the MC was coupling into CARM. MC refl is much calmer in the last ~1.5 hrs since I turned off the WFS.
My main concern with tonights situation was the huge low frequency fluctuations of TRY while CARM/DARM locked on ALS. We saw this being very smooth very recently, but when one arm is fluctuating by multiple line widths, it isn't surprising that locks aren't stable. I want to know why the out of loop stability is so unpredictable. |
10497
|
Fri Sep 12 00:28:04 2014 |
ericq | Update | LSC | Holy sensitivity, Batman! | I took a quick measurement of the ALS stability, using POX and POY as out of loop sensors, using a CARM calibration line to line POX and POY up to the calibrated PHASE_OUT channels at 503Hz.
- X arm RMS ~1kHz
- Could use more low frequency suppression
- Y arm RMS ~200Hz

|
10499
|
Fri Sep 12 03:49:57 2014 |
ericq | Update | LSC | Some more PRFPMI efforts | Since DRMI didn't get fully commissioned, I tried my hand at PRFPMI locking with the newly improved ALS performance.
ALS seemed reliable, I think my main limiting factor was the PRMI locking. We should set up a restore script for PRFPMI that is a superset of the ALS CARM DARM, because the current restore script doesn't put all the vertex settings back, so I was trying to lock for a while without the FM boosts on PRCL and MICH, which really hurt my stability.
Transitioning to SqrtInv works fine; a couple of times I've gotten to arm power of ~10, and have been able to sit there for a while as I set up excitation line comparisons with the CM board's REFLDC, but the PRC would always lose it before I did anything interesting.
The PRMI locks with a reasonable MICH offset, I found that adding a offset of 20 to 40 makes the AS spot visibly dimmer, and ASDC falls to ~0.05 from .1-.2.
I looked into adding a boost to the CARM loop after transitioning to sqrtInv, but we only have 30 degrees of margin, and the error signal is already fairly white, so there isn't much to do, really.
The ALS locking script is sporadically hanging a fair while, as well, which is strange. Otherwise, not much to report... |
10501
|
Fri Sep 12 12:00:59 2014 |
ericq | Update | LSC | DRMI locking | REFL11 I, as seen in digital land, is connected to the slow output of the CM board. I tuned the demod angle of the REFL11 demodulator board by cable length back in ELOG 9850. It would be good to check that the phase is still good. If the CM board gains are at 0dB, we should be able to used the digital angle adjustment as normal. |
10502
|
Fri Sep 12 14:11:17 2014 |
ericq | Update | LSC | DRMI locking | We need to get an interferometric estimation of the SRC length error / SRC sideband splitting, because if the 7.5cm hand-measured error is true, it looks like it might be hard to control the DRMI on 3F.
I did some DRMI sensing simulations, to get an idea if sensing matrix elements might change as the CARM offset changes. Last night, I tried just going to zero CARM offset on ALS, and was having problems keeping the PRMI locked on REFL33, so I wanted to confirm that it should at least work in theory.
Thus, I simulated what happens to the sensing matrix element in the vertex DoFs as the CARM offset is reduced, in both the PR and SR cases. I normalized all of the elements to PRCL at zero carm offset, to get an idea of what the good relative gains should be for MICH and SRCL.
In the end, there don't seem to be significant DC gain changes, or demod angle fluctuations, in either the PRFPMI or DRFPMI case, as the CARM offset changes, which is good.
However, the SRC length as hand-measured, seems to mess up the MICH angle in the DRFPMI case, and really lowers the SRCL signal amplitude.
To be fair, past efforts of simulating demodulation angles haven't always been borne out on the IFO, so we should still forge ahead experimentally until it becomes apparent that there is a real problem.
Here are the simulations for the IFO as-is:
(A note on the plots. Though they kind of look like Bodes, they're just the sensing element represented as a complex number in the I-Q plane,I being phase=0 and Q = 90)
  
All three signals are along the I axis in the DRMI case, which seems like it would be tough to control, since we only have 2 3F diodes... We've been using REFL33Q when PRMIing, which is simulated at around 45 deg; it should be easy to verify this empirically.
Here are the same plots with the SRC length corrected. Now MICH shows up mostly in the Q phase as desired in the DRMI case. SRCL in REFL165 also wins 20dB of optical gain, as well.
  
To drive the point home, here's a simulated scan of AS110 and REFL55 Q to show the effect of the measured length error:

|
10509
|
Tue Sep 16 14:26:45 2014 |
ericq | Update | PSL | Laser turned on |
Quote: |
Our janitor turned off the laser accidentally.
|
The PMC wasn't locking very happily after this. I tweaked the pointing onto the PMC REFL diode, to make sure it was centered, and touched the alignment into the PMC. I also reset the FSS Slow output to zero. It took a little while for the laser to settle in, for some reason, but the transmission is up at 0.80 now.
Tweaked MC2 pointing to get the MC transmission high enough to let WFS kick in, which nicely got the rest of the MC alignment done. After that, I offloaded the WFS into the MC suspensions.
Lastly, I ran the command that Rana posted in ELOG 10391, to set the FSS input offset (From -0.18 to -0.06) |
10514
|
Wed Sep 17 15:40:00 2014 |
ericq | Update | LSC | DRMI locking | I have not had any success the past two days in getting an interferometric measurement of the SRC length.
So, the question posed at today's meeting was: "How precisely do we need to change the SRC length to be able to lock the DRMI on 3F"
The two ways I could think to quantify this are:
- How much MICH -> [S,P]RCL cross coupling is ok?
- How much [S,P]RCL -> MICH cross coupling is ok?
REFL33 should have its phase set to put PRCL along I, and REFL165 should have SRCL along I, so the simulation result that matters is the angle of MICH in these planes. The cross couplings are then given by the appropriate trigonometric projections. In the following plots, I used 10% as the acceptable cross coupling in either direction.
Result:

Thus,
- To limit the MICH -> [S,P]RCL coupling to 10%, we must hit the ideal length within +- 1.2cm.
- To limit the SRCL -> MICH coupling to 10%, we must hit the ideal length within +- 2mm.
- It doesn't look like we can get the REFL33 angle totally to 90 degrees, REFL165 looks more promising.
Code (finesse + pykat + ipython notebook) and plots are attached. |
Attachment 2: drfpmiVertexSensing.zip
|
10520
|
Fri Sep 19 04:05:05 2014 |
ericq | Update | LSC | AO path partly engaged | More AO efforts. No huge news.
Came at AO from each side. For each sign, I lost lock just a few dB from the AO portion of the loop crossing unity gain. Both attempts were about arm powers of 1, which should correspond to ~300pm CARM offset, which I have simulated the crossover as possible with my current loop models (including latest MC loop). The gain steps were usually 6dB in between measurements.
Positive polarity on CM board screen:
I made it to +5 dB of the last plot here, but the 6th broke it open. Gains on CM In2, CM AO, and MC In2 were -6, -4, -2 on that last, lock breaking, step.

Negative polarity on CM board screen:
Lost it just 2dB above the last trace. Gains were -6, +1, -2 (So, overall 5dB higher than the other polarization)

Many things happened in between these two lock stretches, but I'm not sure what may or may not have affected things. They include:
- Jenne mentioned PRMI being fussy earlier in the evening. I adjusted REFL33 and POP22 angles during a PRMI lock, while CARM was held away with ALS. My simulations suggest that there are small changes to the 3F sensing when the arms are totally absent, but doing it at a finite CARM offset is closer to where we want it, it seems.
- I tried using REFL165Q for MICH, since my simulations suggest a better MICH/PRCL angle, which would stave off cross couplings. Lined up excitations, etc., but no luck.
- I measured the PRMI loops
- found PRCL to have ~200Hz UGF, 8dB gain peaking. Maybe a little high, but didn't seem terrible.
- MICH had UGF of around 20Hz, with the FM gain at 0.8. By the shape of the phase bubble, the loop seems designed for higher bandwidth. I raised the gain to 2.5 for a 70ishHz UGF, and called in FMs 7 and 9 for additional triggered boosts. Things seemed to stay locked pretty well.
- Lower excitation amplitude the second time around, measuring the AO loop. Looking at the CM output spectra, you can see the excitation wailing away; I wanted to avoid it.
The location of the CARM resonance peak lines up with my simulation, which is good, but there appears to be less phase than expected... I tried making sure that we don't have any whitening uncompensated for, but it looked ok. All my AO path loop model contains is the CM board TF (measured and fitted), the IMC seen as an actuator(measured and fitted), and the REFLDC optical TF (simulated in MIST). Maybe the DC path of whatever diode this is coming from needs to be included...
Discontinuities / glitches could be seen in the CM board fast output when MC board gains were changed, which isn't so nice. Incidentally, I notice now that each lock loss corresponded to a step of AO gain on the CM board. |
10527
|
Tue Sep 23 17:37:10 2014 |
ericq | Update | LSC | DRMI locking | Rather than using a CAD drawing, I used Gabriele's code from ELOG 9590 to try and judge if we could shorten the SRC by the appropriate length, without clipping the SR3-SR2 beam.
Specifically, I used these lines:
% Move SRM 7.5 towards SR2, parallel to beam
delta=75;
dAS = BS2-AS; % Vector from SRM to SR2
dASmag = sqrt(dAS(1)^2+dAS(2)^2);
dMove = delta*dAS/dASmag; % delta times unit vector
CS = CS+dMove;
draw_sos(CS, 180/pi*angles)
to help generate this plot:

As a reminder, Gabriele's code used the following logic:
- We know the nominal dimensions of all of the suspensions
- We hand measured various distances between features of the suspension structures. (Corner to corner)
- A global fit, minimizing the maximum error, reconstructed the positions of the suspensions.
- Beam positions assumed to be ideally aligned.
- Beam trajectories traced out, and optical path lengths estimated (taking into account changing indices of refraction due to flipped mirrors)
In my opinion, this is the best estimate of beam trajectory that we currently have.
Thus, from looking at the plot above, I claim we can correct the SRC length without clipping the beam by moving the SRM forward by the required 7.5cm.
Although the measured distance may be off on the order of a cm (since our PRC correction had a 0.5cm disagreement between interferometric and hand distance measurements), this will nevertheless markedly improve our 3F DRMI sensing, based on my previous ELOG.
Hence, given our discussions last week, Jenne and I will proceed to ready the interferometer for venting in the morning, by following the vent checklist.
Our sole objective for this vent is this move of the SRM.
Steve, please check the jam nuts, and begin the vent when you get in. Thanks! |
10534
|
Wed Sep 24 18:17:46 2014 |
ericq | Update | General | Alignment Restored | Interferometer alignment is restored
ASS has been run on each arm, recycling mirrors were aligned by overlapping on AS camera.
Notes:
- Mode cleaner alignment took some manual tweaking, locked fine around 1k counts. Still no autolocker.
- At this point, some light was visible on AS and REFL, which was a good sign regarding TTs.
- Used green light to align ETMs to support a green 00 mode.
- Ensured no recylcying flashes were taking place on AS camera and PRM face camera.
- Arms were locked using AS55, with the other ITM misalgined, for better SNR than PO[XY]. ASS brought arm powers to ~0.06, which is about what we would expect from 1k MC2 trans instead of 16k.
- ASS Yarm required debugging, see below.
- ETMX was getting kicks again. Top Dsub connector on the flange near the ground closer to the end table was a little loose. We should fasten it more securely.
- At this point, michelson alignment was good. Brought in PRM to see PRC flashes, REFL spot was happy. Brought in SRM to AS sppot.
- Saved all optic positions.
- Oplevs:
- PRMs new aligned state is falling off the QPD.
- ETMs and BS oplev centering are fine, rest are less good, but still on the detector.
ASS-RFM issue:
ETMY was not getting its ASC pitch and yaw signals. C1SCY had a red RFM bit (although, it still does now...)
I took a look at the c1rfm simulink diagram and found that C1RFM had an RFM block called C1:RFM-TST_ETMY_[PIT/YAW] and C1SCY had one called C1:TST-SCY_ETMY_[PIT/YAW].
It seems that C1TST was illegally being used in a real signal chain, and Jenne's recent work with c1tst broke it. I renamed the channels in C1RFM and C1SCY to C1:RFM-SCY_ETMY_[PIT/YAW], saved, compiled, installed, restarted. All was well.
There are still some in SCY that have this TST stuff going on, however. They have to do with ALS, it seems, but are SHMEM blocks, not RFM. Namely:
- C1:TST-SCY_TRY
- C1:TST-SCY_GLOBALPOS
- C1:TST-SCY_AMP_CTRL
|
10535
|
Wed Sep 24 18:56:45 2014 |
ericq | Update | General | Ottavia slowness | Ottavia was having some severe interaction latency today. Xorg was taking up >90% of the CPU, just sitting around. The machine was logged in to a desktop session with lots of graphical effects turned on. I changed the system default session to "gnome-fallback" in /etc/lightdm/lightdm.conf, which was already set as the default for controls, but wouldn't get chosen for the autologin that happens on boot.
Hopefully this helps ottavia stay usable... |
10540
|
Thu Sep 25 15:41:15 2014 |
ericq | Update | Computer Scripts / Programs | Rossa having a better day |
I think I found out why rossa was mad.
An apt-get update on the 18th downloaded kernel 2.6.32-65-generic, so 2.6.32-58-generic, which what was previously chosen as a working kernel, had moved down in the grub ordering.
It turns out the grub configuration accepts strings, so I changed it to GRUB_DEFAULT="Ubuntu, with Linux 2.6.32-58-generic", ran sudo update-grub, and Rossa now seems to boot happily. |
10541
|
Thu Sep 25 19:42:12 2014 |
ericq | Update | General | Vent progress | [q, Jenne, Manasa]
ELOG Outline
- Aligned arms
- Took a bunch of photos of ITMY chamber.
- Used Crystallaser to reverse trace POY beam path
- Realized real POY flashes were visible
- Tried adjusted POY steering to give us enough room to move SRM forward 3 in
- Not enough steering room
- Placed beam centering targets just before OM3 and after OM4, on far side of BS table
- Twisted SR2 by ~5 degrees
- Moved SRM laterally ~3 inches (beam had been hitting optic center, then suspension cage after SR2 twist)
- Moved SRM foward about 3 inches, adjusted angle for coarse retroreflection
- Measured distances between suspensions
- Tried running Gabriele's distance reconstruction code, results not looking so good; redundancy checks are off by ~1cm
- Started roughly repositioning OM1 and OM2 to get through beam targets
- Closed light doors
- ITMY pointing has gone bad, no green or IR resonance to be seen, still have IR and Green in Xarm, so TTs, BS are ok
|
10544
|
Fri Sep 26 12:25:34 2014 |
ericq | Update | General | Vent progress | I figured out that didn't change the initial guess for the fit routine in Gabriele's code. I also changed the fminsearch criteria to least squares fitting, instead of minimax. The consistency checks now look just as good as the previous time we did these kind of measurements, no disagreements bigger than 1.6mm.
Thus, the current estimate of the SRC length after yesterday's motions is 5402mm, where we desire 5399mm. So, we will try to move SRM 3mm closer to SR2, after confirming that we are not clipping the POY beam. After all that, we will level the table. |
10545
|
Fri Sep 26 16:10:14 2014 |
ericq | Update | General | Vent update | Today so far:
- I moved SRM forward by 3mm
- Then I leveled the ITMY table
- At this point, bringing the ITMY oplev beam back onto its QPD got me back to green locking and IR flashes
- AS and POY beams are both making it out to their tables, as seen by IR card. (Though not to their in-air optics)
Here's my quick brain dump of things to do before we can pump down (anyone see anything missing?):
- Check the clearance of the POY beam at the SRM cage
- Re-do distance reconstruction measurements, confirm desired SRC length
- Lock the SRM cage down fully (right now, has 2 clamps on, and one laying unused)
- Align SRM for SRC flashes
- Adjust SRM OSEM positions as needed
- Adjust SRM oplev beam path, measure lever arm for calibration
- Confirm beam spots on output mirrors in ITMY and BS chambers are ok
- Take pictures of ITMY chamber.
- Closeup checklist
|
10546
|
Fri Sep 26 17:13:39 2014 |
ericq | Update | General | Vent update |
Quote: |
- Check the clearance of the POY beam at the SRM cage
- Re-do distance reconstruction measurements, confirm desired SRC length
|
POY has >2 inches of clearance from the SRM cage. 
Distance reconstruction indicates an SRC length of 5399mm, which was exactly our target.  |
10549
|
Mon Sep 29 12:47:51 2014 |
ericq | Update | General | Vent update |
Quote: |
- Lock the SRM cage down fully (right now, has 2 clamps on, and one laying unused)
- Align SRM for SRC flashes
- Adjust SRM OSEM positions as needed
- Adjust SRM oplev beam path, measure lever arm for calibration
- Confirm beam spots on output mirrors in ITMY and BS chambers are ok
|
[Koji, ericq]
We have completed the above points; the ITMY table is still level.
Despite what the wiki says, the SRM LR OSEM open voltage is ~1.97V instead of ~1.64, so we shot for half of that.
The in-air steering of the SRM oplev return beam needs adjustment. I'll estimate the beam path length when I'm taking pictures and closing up.
Left to do:
- Now that AS is back on diode, lock arms and align everything. Confirm everyone's happiness.
- Take numerous pictures of ITMY chamber.
- Center oplevs
- Put doors on
- Close shutters
- Pump down
- Replace MC refl Y1 with the beamsplitter
- Turn PSL power back up
Related In-Air work:
- Fix POY steering
- Fix SRM oplev return steering
|
10550
|
Mon Sep 29 17:10:51 2014 |
ericq | Update | General | Vent update | Everything is aligned, AS and POY make it out of vacuum unclipped, OSEM readings look good.
I set up the SRM oplev, centered all oplevs.
Tomorrow, we just have to take pictures of the ITMY chamber before we put the heavy doors on. |
10551
|
Mon Sep 29 18:12:24 2014 |
ericq | Update | General | Vent update | I closed the PSL shutter as we didn't want to burn the mirror surface when we are not working. |
10552
|
Tue Sep 30 11:53:29 2014 |
ericq | Update | General | Vent update |
Photos have been taken of the ITMY chamber, and uploaded to picasa. Here's a slideshow:
|
10554
|
Tue Sep 30 17:26:18 2014 |
ericq | Update | LSC | New AO cable in place | I've installed a new 2pin lemo cable going from the CM servo out to in2 of the MC servo board, and removed the temporary BNC. I used some electrical tape to give the cable some thickness where the lemo head screws on to try to strain relieve the solder joints; hopefully this cable is more robust than the last.
I put an excitation into the CM board, and saw it come out of MC_F, so I think we're set. |
|