Wed Jun 11 15:30:55 2008 |
josephb | Configuration | Cameras | GC1280 |
The trial use GC1280 has arrived. This is a higher resolution CMOS camera (similar to the GC750). Other than higher resolution, it has a piece of glass covering and protecting the sensor as opposed to a plastic piece as used in the GC750. This may explain the reduced sensitivity to 1064nm light that the camera seems to exhibit. For example, the image averages presented here required a 60,000 microsecond exposure time, compared to 1000-3000 microseconds for similar images from the GC750. This is an inexact comparison, and the actual sensitivity difference will be determined once we have identical beams on both cameras.
The attached pdfs (same image, different angles of view) are from 200 averaged images looking at 1064nm laser light scattering from a piece of paper. The important thing to note is there doesn't seem to be any definite structure, as was seen in the GC750 scatter images.
One possibility is that too much power is reaching the CMOS detector, penetrating, and then reflecting back to the back side of the detector. Lower power and higher exposure times may avoid this problem, and the glass of the GC1280 is simply cutting down on the amount passing through.
This theory will be tested either this evening or tomorrow morning, by reducing the power on the GC750 to the point at which it needs to be exposed for 60,000 microseconds to get a decent image.
The other possibility is that the GC750 was damaged at some point by too much incident power, although its unclear what kind of failure mode would generate the images we have seen recently from the GC750. |
Attachment 1: GC1280_60000E_scatter_2d.pdf
|
|
Attachment 2: GC1280_60000E_scatter_3d.pdf
|
|
649
|
Tue Jul 8 21:46:38 2008 |
Yoichi | Configuration | PSL | GC650M moved to the PMC transmission |
I moved a GC650M, which was monitoring the light coming out of the PSL, to the transmission port of the PMC to see the transmitted mode shape.
It will stay there unless someone find other use of it.
Just FYI, you can see the picture from the control computers by the following procedure:
ssh -X mafalda
cd /cvs/cds/caltech/target/Prosilica/40mCode
./SampleViewer
Chose 02-2210A-06223 and click on the Live View icon. |
521
|
Thu Jun 5 13:35:23 2008 |
josephb | Configuration | Cameras | GC750 looking at 1064nm scattered light |
I've taken 200 images of the GC750 (CMOS) camera while holding it by hand up to a beam card (also held by hand) in the path of ~5mW of beam power. I then averaged the images to produce the fourth attached plot.
Rob has pointed out the image looks a lot like PCB traces. So perhaps we're seeing the electronics behind the CMOS sensor?
I repeated the same experiment with HeNe laser light (again scattered off a card). These show none of the detailed structure (just what looks to be a large reflection from the card moving around depending on how steady my hand was). These are the first 3 attached plots. So only 1064nm light so far sees these features.
As a possible solution, I did a quick and dirty calibration by dividing a previous PSL output beam by the 1064 average scatter light values. These produce the last attached pdf (with multiple images). The original uncalibrated image is on top, while the very simply calibrated image is on the bottom of each plot.
It seems as the effect may be power dependent (which could still be calibrated properly, but would take a bit more effort than simply dividing), as determined by looking at the edges of the calibrated plot. |
Attachment 1: GC750_HeNe_scatter_avg.pdf
|
|
Attachment 2: GC750_HeNe_scatter_avg2.pdf
|
|
Attachment 3: GC750_HeNe_scatter_avg3.pdf
|
|
Attachment 4: GC750_scatter_avg.pdf
|
|
Attachment 5: GC750_nitrogen_white.pdf
|
|
378
|
Fri Mar 14 12:06:29 2008 |
josephb | Configuration | Cameras | GC750 looking at ETMX while locked |
The GC750 (CMOS) is currently looking at the front of ETMX. Unfortunately, its being routed through a 10Mbit connection (which I will be purchasing a replacement for today), so getting it to send images to Mafalda/Linux 2 or 3 isn't working well, but by using a local gigabit switch and a laptop I can get sufficient speed for full images with the sample viewer.
The attached image is from a full 752x480 reslution with 10,000 microsecond exposure with the X-arm locked. Although it looks like I still need to work on the focusing. Will be switching the GC750 with the GC 650 (CCD) later today and comparing the resulting images. |
Attachment 1: ETMX_zoom_exp_10000_750.tiff
|
558
|
Tue Jun 24 17:12:10 2008 |
josephb, Eric | Configuration | Cameras | GC750 setup, 1X4 Hub connected, ETMX images |
The GC750 camera has been setup to look at ETMX. In addition, the new 1X4 rack mounted switch (131.215.113.200) has been connected via new cat6 cable to the control room hub (131.215.113.1?), thanks to Eric. The camera is now plugged into 1X4 rack switch and now has a gigabit connection to the control room computers as well as Mafalda (131.215.113.23).
By using ssh -X mafalda or ssh -X 131.215.113.23, then typing:
target
cd Prosilica/bin-pc/x86/
./Sampleviewer
A viewer will be brought up. By clicking on the 3rd icon from the left (looks like an eye) will bring up a live view.
Closing the view, and then cd ../../40mCode, and then running ./Snap --help will tell you how to use a simple code for taking .tiff images as well as setting things such as exposure length and size of image (in pixels) to send.
When the interferometer was set to an X-arm only configuration, we took two series of 200 images each, with two different exposure lengths.
Attached are three pdf images. The first is just a black and white single image, the second is an average of 100 images, and the third is the standard deviation of the 100 images. |
Attachment 1: GC750_ETMX_E30000_single.pdf
|
|
Attachment 2: GC750_ETMX_E30000_avg.pdf
|
|
Attachment 3: GC750_ETMX_E30000_std.pdf
|
|
10854
|
Mon Jan 5 20:17:26 2015 |
jamie | Configuration | CDS | GDS upgraded to 2.16.14 |
I upgraded the GDS and ROOT installations in /ligo/apps/ubuntu12 the control room workstations:
- GDS 2.16.14
- ROOT 5.34.18 (dependency of GDS)
My cursory tests indicate that they seem to be working:

Now that the control room environment has become somewhat uniform at Ubuntu 12, I modified the /ligo/cdscfg/workstationrc.sh file to source the ubuntu12 configuration:
controls@nodus|apps > cat /ligo/cdscfg/workstationrc.sh
# CDS WORKSTATION ENVIRONMENT
source /ligo/apps/ligoapps-user-env.sh
source /ligo/apps/ubuntu12/ligoapps-user-env.sh
source /opt/rtcds/rtcds-user-env.sh
controls@nodus|apps >
This should make all the newer versions available everywhere on login. |
7238
|
Tue Aug 21 00:02:05 2012 |
rana | Summary | Computer Scripts / Programs | GDS/DTT bug: 10 digit GPS times not accepted |
I've noticed that we're experiencing this bug which was previously seen at LHO. We cannot enter 10 digit GPS times into the time fields for DTT due to a limit in TLGEntry.cc, which Jim Batch fixed in September of last year. Seems like we're running an old version of the GDS tools.
I checked the Lidax tool (which you can get from the GDS Mainmenu). It does, in fact, allow 10 digit entries. |
13
|
Thu Oct 25 00:01:21 2007 |
rana | Software Installation | CDS | GEO DV => LIGO DV |
Martin Hewitson of GEO600 fame has modified the cool GEO DV
to work with the LIGO NDS system with some NDS advice from Rolf (who's over in Germany this week).
I've moved it onto the 40m CDS system and installed it on the AdhikariLab computer named 'django'. It worked immediately.
I modified the main .m file to include the 40m's NDS server. When you run it you have to include the path to the NDS
client written by Ben Johnson.
The attached is a screenshot of it working on a Mac; it looks as cool on Linux.
Its installed in /cvs/cds/caltech/apps/ligoDV/. In matlab you navigate to that directory and then
type addpath('/cvs/cds/caltech/apps/linux/UNIX_NDS_Client_beta2/') to add the NDS client.
On the Solaris machines, type type addpath('/cvs/cds/caltech/apps/solaris9/UNIX_NDS_Client_beta2/') instead.
Then type ligoDV to start it up. Then click away and have fun.
In the example I've selected C1:PEM-BS_ACC_EAST_Z and plotted its specgram.
 |
Attachment 1: Picture_1.png
|
|
144
|
Fri Nov 30 11:22:22 2007 |
ajw | Summary | CDS | GEO DV => LIGO DV |
Quote: | Martin Hewitson of GEO600 fame has modified the cool GEO DV
to work with the LIGO NDS system with some NDS advice from Rolf (who's over in Germany this week).
I've moved it onto the 40m CDS system and installed it on the AdhikariLab computer named 'django'. It worked immediately.
I modified the main .m file to include the 40m's NDS server. When you run it you have to include the path to the NDS
client written by Ben Johnson.
The attached is a screenshot of it working on a Mac; it looks as cool on Linux.
Its installed in /cvs/cds/caltech/apps/ligoDV/. In matlab you navigate to that directory and then
type addpath('/cvs/cds/caltech/apps/linux/UNIX_NDS_Client_beta2/') to add the NDS client.
On the Solaris machines, type type addpath('/cvs/cds/caltech/apps/solaris9/UNIX_NDS_Client_beta2/') instead.
Then type ligoDV to start it up. Then click away and have fun.
In the example I've selected C1:PEM-BS_ACC_EAST_Z and plotted its specgram.
 |
Download and installation instructions, as well as a few examples for use
can be found here (typical lsc username and password):
https://www.gravity.phy.syr.edu/dokuwiki/doku.php?id=ligodv:home
https://www.gravity.phy.syr.edu/dokuwiki/doku.php?id=ligodv:downloading_the_ligodv_software |
234
|
Thu Jan 10 13:45:52 2008 |
Pkp | Update | Cameras | GLIBC Error |
So, I have tried to compile the camera files which are in /cvs/cds/caltech/target/Prosilica/examples for the past 2 days now and have been unable to get rid of the following error. (specifically ListCameras.cpp, as it doesnt have any other libraries required, which unnecessarily complicates things)
../../bin-pc/x86/libPvAPI.so: undefined reference to `__stack_chk_fail@GLIBC_2.4'
collect2: ld returned 1 exit status
make: *** [sample] Error 1
I used to get this error on mafalda too, but I had fixed it by installing the latest version of the glibc libraries. Inspite of doing so on linux2, the error still persists. I suspect it had something to do with it being a FC3 machine. My own laptop, which also runs Ubuntu works fine too. The problem with these Ubuntu machines is that they dont let me set the packet sizes to 9 kb which is required by the camera. Linux2 does.
If anyone has any idea how to resolve this issue, please let me know.
Thanks
Pinkesh. |
158
|
Mon Dec 3 16:24:47 2007 |
tobin | HowTo | Computers | GNU screen |
GNU screen is a utility that can be quite handy for managing long-running psuedo-interactive terminal programs on remote machines. In particular, I think it might be useful in developing and testing "Matlab DMT" tools on Mafalda. |
142
|
Thu Nov 29 18:10:13 2007 |
Alberto | HowTo | Computer Scripts / Programs | GPIB Scripts |
I've spent a lot of time trying to configurate the GPIB-USB interface for the HP4195. After installing 1) the Agilent libraries, 2) the drivers, 3) the matlab Instrument Toolbox, 4) Jamie script, 5) Alice's script the computer can see the HP but still they can't 'talk' to each other.
I give up. I asked Alice Wang how she managed to get data. I'm not sure she used the GPIB interace. Rob said she might have used the old fashion floppy disks that we can't read anymore here.
I would really appreciate any suggestion by anyone who happened to have the same problems. |
143
|
Thu Nov 29 19:35:14 2007 |
rana | HowTo | Computer Scripts / Programs | GPIB Scripts |
Quote: | I've spent a lot of time trying to configurate the GPIB-USB interface for the HP4195. After installing 1) the Agilent libraries, 2) the drivers, 3) the matlab Instrument Toolbox, 4) Jamie script, 5) Alice's script the computer can see the HP but still they can't 'talk' to each other.
I give up. I asked Alice Wang how she managed to get data. I'm not sure she used the GPIB interace. Rob said she might have used the old fashion floppy disks that we can't read anymore here.
I would really appreciate any suggestion by anyone who happened to have the same problems. |
Alice and Jamie used the USB-GPIB interface. You should just try using the black laptop which already has this capability or ask Jamie Rollins
who actually knows something. |
3969
|
Mon Nov 22 20:01:56 2010 |
kiwamu | Configuration | Computers | GPIB box : took it out from HP3563A and connect it to AG4395A |
I disconnected the yellow GPIB box from the backside of HP3563A (classic analyzer),
and connected it to AG4395A (network analyzer), which is the official place for it. |
3120
|
Fri Jun 25 12:09:27 2010 |
kiwamu | Update | Computers | GPIB controller of HP8591E |
I've just stolen a GPIB controller, an yellow small box, from the spectrum analyzer HP8591E.
The controller is going to be used for driving the old spectrum analyzer HP3563A for a while.
Gopal and I will be developing and testing a GPIB program code for HP3563A via the controller.
Once after we get a new GPIB controller, it will be back to the original place, i.e. HP8591E.
--- GPIB controller ----
name: teofila
address: 131.215.113.106 |
1479
|
Mon Apr 13 18:57:03 2009 |
Alberto | Frogs | Computers | GPIB/ETH Interface Troubles |
I really don't understand why my programs that I used to use to get data from the HP Spectrum Analyzer and the Marconi frequency generator don't work anymore.
I spent hours trying to debug the code but I can't sort the problem out.
The main problem seem to be with the function recv from the socket library. Somehow it can't anymore get any data from the instruments. The thing I can't understand, though, is that if called directly from the python terminal it works fine!
In particular the problem is with the following lines in my code:
netSock.send("mkpk;mka?\n")
netSock.send("++read eoi\n")
tmp = netSock.recv(1024)
Tried a lot of tickering but it didn't work.
I attach the two scripts I've been using. One (sweepfrequencyPRC.py) calls the other (HP4395PRC.py).
They worked egregiously for weeks in the past. Don't know what happened since then. |
Attachment 1: sweepfrequencyPRC.py
|
## sweepfrequency.py [-f filename] [-i ip_address] [-a startFreq] [-z endFreq] [-s stepFreq] [-m numAvg]
#
## This script sweeps the frequency of a Marconi local oscillator, within the range
## delimited by startFreq and endFreq, with a step set by stepFreq. An arbitary
## signal is monitored on a HP8590 spectrum analyzer and the scripts records the
## amplitude of the spectrum at the frequency injected by the Marconi at the moment.
## The GPIB address of the Marconi is assumed to be 17, that of the HP Spectrum Analyzer to be 18
## Alberto Stochino, October 2008
... 53 more lines ...
|
Attachment 2: HP8590PRC.py
|
# This function provides the measuremeent of the peak amplitude on the spectrum analyzer
# HP8590 analyzer while sweeping the excitation frequency on the function generator.
#
# Alberto Stochino 2008
import re
import sys
import math
from optparse import OptionParser
from socket import *
... 70 more lines ...
|
1481
|
Tue Apr 14 12:10:11 2009 |
Alberto | Frogs | Computers | GPIB/ETH Interface Troubles |
Quote: |
I really don't understand why my programs that I used to use to get data from the HP Spectrum Analyzer and the Marconi frequency generator don't work anymore.
I spent hours trying to debug the code but I can't sort the problem out.
The main problem seem to be with the function recv from the socket library. Somehow it can't anymore get any data from the instruments. The thing I can't understand, though, is that if called directly from the python terminal it works fine!
In particular the problem is with the following lines in my code:
netSock.send("mkpk;mka?\n")
netSock.send("++read eoi\n")
tmp = netSock.recv(1024)
Tried a lot of tickering but it didn't work.
I attach the two scripts I've been using. One (sweepfrequencyPRC.py) calls the other (HP4395PRC.py).
They worked egregiously for weeks in the past. Don't know what happened since then.
|
This morning Joe looked at my code and made me notice that for some reason the query to the Spectrum Analyzer made by netSock.recv(1024) contained two answers. It was like the buffer contained the answer two different queries.
After some experiment I found that basically the GPIB interface wasn't switching from the "auto 1" to the "auto 0" mode as it should. I rewrote part of the code and that seemed have solved the problem.
Still don't understand why it used to work in the past and then it stopped. |
5994
|
Thu Nov 24 02:23:43 2011 |
Koji | Summary | General | GPIB<->WIFI |
GPIB<->WIFI on Agilent Network Analyzer 4395A was broken.
The connection was fixed by replacing and configuring the LINKSYS bridge.
=======
Kiwamu and I have identified that the LINKSYS bridge was guilty.
I knew that we had another one in the bathroom, I configured it.
HOW TO CONFIGURE IT
- You need a host with 192.168.1.*.
- Connect the host and the LINKSYS directly
- Open http://192.168.1.226/
- Select 40MARS and infrastructure mode
- Register it in the MAC filter of the 40MARS (see wiki and use your head)
The new unit works fine.
I checked the malfunctioning unit and the configuration was the same as the others.
The bad one detected the existence of 40MARS during the configuration, but it does not establish
the connection in the actual use. I am afraid that something is broken or some setting is lost
during the power supply shutdown. |
1070
|
Wed Oct 22 20:50:30 2008 |
Alberto | Omnistructure | Computers | GPS |
Today I measured the GPS clock frequency at the output of CLOCK_MON in a board on the same crate where the c1iool0 computer is located. The monitor was connected with a BNC cable to the 10MHz reference input of the frequency counter on top of that rack, where it was used to check the 166MHz coming from one of the Marconi.
The frequency was supposed to be 10MHz but I actually measured 8 MHz. I tracked down the GPS input cable to the board and it turned out to come from one of the 1Y7 rack. Here it was connected to a board with a display that was showing corrupted digits, plus some leds on the front panel were red.
I'm not sure the GPS reference is working properly. |
1133
|
Thu Nov 13 15:50:44 2008 |
Alberto | Configuration | General | GPS 10MHz clock |
The schematic of the 1Y7 rack that Alan pointed out (see attachment) don't represent anymore the actual rack.
First, with Yoichi we found that the GPS receiver for the 10 MHz is in a different position,
on the other side of the VME computer. It seems to output 1 kHz, which also appears in some modulated way.
This signal is then passed to a board on 1Y7 that seems make just copies of the signal. One of these goes
to an other board in 1Y6 that looks like a GPS receiver but has actually no GPs antenna input. Here it is
not connected to anything, but on its same crate is a the awg computer, so it might be providing the clock
to that by the crate.
We also checked the clock monitor output on the boards in the PSL that provides for the clock to the Penteks
and it seems that these are actually getting 4MHz. |
Attachment 1: rackstuff.pdf
|
|
12182
|
Wed Jun 15 10:19:10 2016 |
Steve | Configuration | CDS | GPS antennas... debugging |
Visual inspection of rooftop GPS antennae:
The 2 GPS antennas are located on the north west corner of CES roof. Their condition looks well weathered. I'd be surprised if they work.
The BNC connector of the 1553 head is inside of the conduit so it is more likely to have a better connection than the other one.
I have not had a chance to look into the "GPS Time Server" unit. |
Attachment 1: GPStimeServer.jpg
|
|
Attachment 2: GPSsn.jpg
|
|
Attachment 3: GPSantennas.jpg
|
|
Attachment 4: GPSantHead.jpg
|
|
Attachment 5: GPSantHead2.jpg
|
|
Attachment 6: GPSantCabels.jpg
|
|
13211
|
Tue Aug 15 16:32:42 2017 |
Jamie, Gautum | Update | CDS | GPS receiver apparently set to correct mode as "UTC" |
Quote: |
The setting when we found it was "GPS", which seems logical enough. However, when we switched it to "UTC" the time as shown on the front panel was correct, now with "UTC" vertically to the right of the time, and fb1 was then showing the correct GPS time.
|
From Keith Thorne:
In the GPS receiver, you are trying to match the IRIG-B output format that is created by the aLIGO IRIG-B Fanout. Since we have to prep the aLIGO IRIG-B Fanout every time there is a leap second coming, I would suspect that we are sending UTC to the IRIG-B receivers. Thus, the GPS receiver needs to be set to that mode.
Soooo, "UTC" is the correct mode for the GPS receiver. |
12167
|
Fri Jun 10 12:21:54 2016 |
jamie | Configuration | CDS | GPS receiver not resetting properly |
The GPS receiver (EndRun Technologies box in 1Y5? (rack closest to door)) seems to not coming back up properly after the reboot. The front pannel says that it's "LKD", but the "sync" LED is flashing instead of solid, and the time of year displayed on the front panel is showing day 6. The fb1 symmetricom driver and VME timing module are still both seeing day 299, though. So something may definitely be screwy with the GPS receiver. |
12200
|
Mon Jun 20 11:11:20 2016 |
Steve | Configuration | CDS | GPS receiver not resetting properly |
I called https://www.endruntechnologies.com/pdf/USM3014-0000-000.pdf and they said it's very likely just needs a software update. They will email Jamie the details.
Quote: |
The GPS receiver (EndRun Technologies box in 1Y5? (rack closest to door)) seems to not coming back up properly after the reboot. The front pannel says that it's "LKD", but the "sync" LED is flashing instead of solid, and the time of year displayed on the front panel is showing day 6. The fb1 symmetricom driver and VME timing module are still both seeing day 299, though. So something may definitely be screwy with the GPS receiver.
|
|
12201
|
Mon Jun 20 11:19:41 2016 |
jamie | Configuration | CDS | GPS receiver not resetting properly |
I got the email from them. There was apparently a bug that manifested on February 14 2016. I'll try to software update today.
http://endruntechnologies.com/pdf/FSB160218.pdf
http://endruntechnologies.com/upgradetemplx.htm |
16299
|
Wed Aug 25 18:20:21 2021 |
Jamie | Update | CDS | GPS time on fb1 fixed, dadq writing correct frames again |
I have no idea what happened to the GPS timing on fb1, but it seems like the issue was coincident with the power glitch on Monday.
As was noted by Koji above, the GPS time kernel interface was off by a year, which was causing the frame builder to write out files with the wrong names. fb1 was using DAQD components from the advligorts 3.3 release, which used the old "symmetricom" kernel module for the GPS time. This old module was also known to have issues with time offsets. This issue is remniscent of previous timing issues with the DAQ on fb1.
I noted that a newer version of the advligorts, version 3.4, was available on debian jessie, the system running on fb1. advligorts 3.4 includes a newer version of the GPS time module, renamed gpstime. I checked with Jonathan Hanks that the interfaces did not change between 3.3 and 3.4, and 3.4 was mostly a bug fix and packaging release, so I decided to upgrade the DAQ to get the new components. I therefore did the following
-
updated the archive info in /etc/apt/sources.list.d/cdssoft.list, and added the "jessie-restricted" archive which includes the mx packages: https://git.ligo.org/cds-packaging/docs/-/wikis/home
-
removed the symmetricom module from the kernel
sudo rmmod symmetricom
-
upgraded the advligorts-daqd components (NOTE I did not upgrade the rest of the system, although there are outstanding security upgrades needed):
sudo apt install advligorts-daqd advligorts-daqd-dc-mx
-
loaded the new gpstime module and checked that the GPS time was correct:
sudo modprobe gpstime
-
restarted all the daqd processes
sudo systemctl restart daqd_*
Everything came up fine at that point, and I checked that the correct frames were being written out. |
6725
|
Thu May 31 01:36:17 2012 |
yuta | Update | Green Locking | GREEN_TRX/GREEN_TRY PDs |
I did the cabling for monitoring DC transmission of the green beam from the end table.
The two PDs are called GREEN TRX and GREEN TRY, and the channel names are C1:GCV-GREEN_TRX and C1:GCV-GREEN_TRY.
The two signal from the PDs go to the ADC_0 card of the c1ioo computer.
Now, C1:GCV-GREEN_TRX/Y are actually connected to the respective PDs, but green beams are not hitting on the PD. We need two pickoff mirrors. |
9055
|
Thu Aug 22 21:16:47 2013 |
Manasa | Update | Green Locking | GTRY normalized |
The Y arm green transmission has been measuring in counts all along. I modified the gain in the ALS-TRY filter module to normalise the transmission.
Transmission has been normalised with GTRY = 1 corresponding to 600 counts. |
9056
|
Thu Aug 22 22:05:36 2013 |
rana | Update | Green Locking | GTRY normalized |
Meh. 600 counts is too weak. You should fix the electronics so that the maximized green laser transmission gives more like ~10000 counts. |
9675
|
Tue Feb 25 23:38:05 2014 |
rana, jenne | Update | PEM | GUR1 Z channel excess noise: oscillating Z channel |
Last night we noticed an excess in the GUR1Z seis BLRMS on the StripTool. It was in the 0.1 - 0.3 Hz band. The rumor in the control room was that "this kind of noise has been showing up at night recently".
AS it turns out, this was not some environmental noise around the 40m at night, but instead its some internal servo oscillation in the GUR1 Z channel. In the Guralp seismometers, each channel is a different mechanical sensor (unlike the STS or T240), so when a single channel gets noisy it doesn't always implicate the others.
My guess is that the oscillation came from the Z channel needing to be recentered. I power cycled the interface box just now. The oscillation had already gone away, but I thought this might reduce the excess noise. Maybe it did, but the effect is tiny. You can see in the oscillation reference that the low frequency noise is high, but in the new trace its still kind of high. Needs to be re-centered correctly with the paddle. Or add a centering button to the interface box. |
Attachment 1: gur1z.pdf
|
|
8567
|
Mon May 13 23:05:51 2013 |
Jenne | Update | PEM | GUR1 masses recentered |
[Evan, Jenne]
Evan brought the Guralp handheld readout paddle and cables back from the ATF (Zach is using GUR2 and one of the T240s for gyro stuff this week), and we recentered the GUR1 masses. N/S and Vert were okay (within 0.1 V), but E/W was at -0.5 V, so we set it at zero. We then plugged the Guralp back in, and turned on the readout box.
There isn't much of a change on the BLRMS on the wall, so it's possible that we weren't actually having any trouble anyway. |
6973
|
Fri Jul 13 13:02:52 2012 |
Masha, Yaakov | Update | PEM | GUR2 Fixed |
Yaakov and I investigated the GUR 2 problem. It turns out that the ADC channels that GUR 2 was plugged into, ADC channels 6 through 8 (on the actual ADC they are C7 through C9), did not correspond to the channels labelled "GUR 2" in the PEM, ADC channels 3 through 5. We modified them so that GUR 2 is now plugged into ADC channels 3 through 5 (on the ADC it's +1).
Before we discovered that this was the problem, we attempted to take the cover off of GUR 1 to check the gains, and discovered a stripped Allen screw on the side by the "Vertical" pot, which we removed.
Now the GUR 2 readout looks good, and we will give it more time to settle down before we take data.
|
15153
|
Fri Jan 24 17:14:01 2020 |
gautam | Update | ALS | Gain blocks installed |
Jordan will write up the detailed elog but in summary,
- Former +24V Sorensen in the AUX OMC power rack (south of 1X2) has been reconfigured to +12V DC.
- The voltage was routed to a bank of fusable terminal blocks on the NW corner of 1X1.
- An unused cable running to the PSL table was hijacked for this purpose.
- The ZHL-1010+ were installed on the upper shelf of the PSL table, the two gain blocks draw a total of ~600mA of current when powered.
Quote: |
I will install these at the next opportunity, so that we can get rid of the many attenuators in this path (the main difficulty will be sourcing the required +12V DC for operation, we only have +15V available near the PSL table).
|
|
15130
|
Fri Jan 17 18:02:21 2020 |
gautam | Update | ALS | Gain blocks packaged and characterized |
Summary:
- The ZHL-1010+ gain blocks acquired from MiniCircuits arrived sometime ago.
- I packaged them in a box prepared (Attachment #1).
- Their performance was characterized by me (Attachment #2 and #3).
The measurements are consistent with the specifications, and there is no evidence of compression at any of the power levels we expect to supply to this box (<0dBm).
Details:
These "gain blocks" were acquired for the purpose of amplifying the IR ALS beat signals before transmission to the LSC rack for demodulation. The existing ZHL-3A amplifiers have a little too much gain, since our revamp to use IR light to generate the ALS beat.
Attachment #4: Setups used to measure transfer functions and noise.
For the transfer function measurement, I chose to send the output of the amplifier to a coupler, and measured the coupled port (output port of the coupler was terminated with 50 ohms). This was to avoid saturating the input of the AG4395. The "THRU" calibration feature of the AG4395 was used to remove the effect of cabling, coupler etc, so that the measurement is a true reflection of the transfer function of OUT/IN of this box. Yet, there are some periodic ripples present in the measured gain, though the size of these ripples is smaller than the spec-ed gain flatness of <0.6dB.
For the noise measurement, the plots I've presented in Attachment #3 are scaled by a factor of sqrt(2) since the noise of the ZFL-500-HLN+ and the ZHL-1010+ are nearly identical according to the specification. Note that the output noise measured was divided by the (measured) gain of the ZFL-500-HLN+ and the ZFL-1010+ to get the input referred noise. The trace labelled "Measurement noise floor" was measured with the input to the ZFL-500-HLN+ terminated with 50ohms, while for the other two traces, the inputs of the ZHL-1010+ were terminated with 50ohms.
Raw data in Attachment #5.
I will install these at the next opportunity, so that we can get rid of the many attenuators in this path (the main difficulty will be sourcing the required +12V DC for operation, we only have +15V available near the PSL table). |
Attachment 1: photos.pdf
|
|
Attachment 2: gain.pdf
|
|
Attachment 3: noise.pdf
|
|
Attachment 4: measSchem.pdf
|
|
Attachment 5: zhl1010Data.zip
|
10418
|
Thu Aug 21 02:42:17 2014 |
rana, ericq | Configuration | Green Locking | Gain changes on Green Y PDH |
[rana, ericq]
We spent time trying to relieve the Yend green PDH of it troubles.
We realized that the mixer in the PDH setup (mini circuits ZAD-8+), wants 7dBm of LO to properly function. However, we use one function generators output, through a splitter, to give signals to the laser PZT and the mixer LO.
We don't want 7dBm of power hitting the laser PZT, though. The summing node that adds the servo output to the sideband signal was supposedly designed to do some of this attenuation. Rana measured that 10Vpp out of the function generator resulted in 20mVpp on the fast input to the NPRO, after the summing node. Hence, the 0.09V setting was only resulting in something like 0.2mV hitting the PZT. The PZT has something like 30 rad/V PM response, meaning we only had ~0.006 rad of modulation.
Now, the function generator is set to 2 Vpp, meaning 4 mVpp hitting the PZT, meaning ~0.12 radians of modulation. The mixer is now getting +7dBm on its LO, and the PDH traces look much cleaner. However, the PDH error signal is now something like 100mVpp, which is much bigger than the PDH board is designed for, so there is now a 10dB attenuator between the reflection PD DC block and the RF input to the mixer.
Here are screenshots of the Inmon channel (which has a gain of ~20) showing a sweep through some PDH signal, and the error signal while in green lock. Huge 60Hz harmonics are still observed.
 
Regarding these 60Hz issues, we need to make sure that we remove all situations where long BNCs are chained together with barrel connectors, or Ts are touching other ones. We also should glue or affix the pomona summing box to the shelf, so that its not just laying on the floor.
The concrete next step is to go fiddle with things, and see if we can get the 60Hz noise to go away, then measure the PDH loop and noises again. Hopefully, this should make the ALS much more reliable. |
10420
|
Thu Aug 21 19:04:52 2014 |
ericq | Configuration | Green Locking | Gain changes on Green Y PDH |
I found that the barrel of one the BNC to BNC connectors used for getting the output of the PDH servo box to the laser controller was touching the ETMY chamber. When I held it away, all of the 60Hz harmonics disappeared from the mixer output spectrum; this was pretty repeatable. This inspired me to replace the refl PD and PZT signal cables (which were 2 and 3 cables stitched together, respectively) with 20' long BNCs. I also cleaned up a lot of the routing of signal and power cables in the little rack, and moved the big T->DC Block->Attenuator combo off of the panel mount, because I didn't like how it was wiggling. It and the summing pomona box are sitting on top of the PDH box and function generator, instead of hanging freely.
All of the 60Hz harmonics were banished afterwards, and the green locked happily. 
This required me touching the Y end table, to remove the old cable and its cable ties, and putting the new one in. I don't think I did anything immediately apparently bad; the green and IR transmissions both are within nominal ranges.
I haven't had luck measuring the CLG yet, which I wanted to do to get and set the UGF before measuring the noises. However, here is a scope trace of the in-lock error signal, which compares quite favorably to the trace posted in the previous post; the scope indicates that the signal has 1/3 of the RMS that it did before I replaced the cables.

I hope to measure up the current status after I get back from dinner.
|
10430
|
Tue Aug 26 23:16:49 2014 |
ericq | Configuration | Green Locking | Gain changes on Green Y PDH |
Yesterday I measured the spectra and OLTF of the Y-Arm green PDH, after the LO touch-up and 60Hz hunt from last week. I also went to lower frequencies with the SR785, but forgot to take some of the background spectra down there, so I don't have the full breakdown plots yet. Nevertheless, here is the improvement in the PDH error signal:

I also measured the OLTF (SR785 injection at the error signal, Auto level ref 5mV at channel 2, 10mV/s source ramping, 50mV max output)

As you can see, we have tons of phase margin. Flipping the local boost switch had no visible effect on the OLTF; we should change it to something that puts this surplus of phase to good use, and squash the error signal even more. Putting an integrator at 5kHz should still leave about 45 degrees phase margin at 10k. I've started making a LISO model of the PDH board from the DCC drawing, and then I'll inspect the boards individually to make sure I catch the homegrown modifications.
Data, and code used to generate the plots is attached. |
Attachment 3: newY.zip
|
10107
|
Fri Jun 27 12:54:13 2014 |
Akhil | Update | Electronics | Gain plots of UFC-6000 for 0.1s Sampling Rate |
Finally, the 0.1s sampling rate of the frequency counter(FC) has been achieved. For this I had to :
Send in byte codes to set a particular range of the frequency counter.
I was digging in to find how exactly the circuit inside the frequency counter works and how the processor inside is able to read and write bytes through a HID-USB interface. I found out that the 'AutoRange' setting (which I have been using so far) has an independent multiplexing circuit which consumes some time(that varies with the drift in frequencies) and thus, the the processor waits for some specific time for this process and cannot reach the minimum 0.1 s sampling time. To mitigate this issue, I set the range bytes to the appropriate range of frequencies so that I can bypass the MUX delay. Here is the list of Range and frequencies for the FC:
Range 1: 1 - 40 MHz
Range 2: 40 - 190 MHz
Range 3: 190 - 1400 MHz
Range 4 : 1400 - 6000 MHz
I then took measurements for sampling time of 0.1 s at carrier frequencies of 5 MHz and 25 MHz from SRS DS345 and plotted the improvised gain plots(attached) to those in my previous elog(10070) with the same procedure mentioned before.
To do Next:
Plot the gain plots for higher carrier frequencies till range 3 using Marconi Function generator.
Write the data from FC into C1: ALS-Y_SLOW_SERVO1_OFFSET EPICS channel.
|
Attachment 1: Sampling-0.1s.pdf
|
|
Attachment 2: 12.png
|
|
10118
|
Tue Jul 1 21:20:37 2014 |
Akhil | Update | Electronics | Gain plots of UFC-6000 for 0.1s Sampling Rate |
The attached are the gain plots at carrier frequencies of 100 MHz, 500 MHz and 1 GHz measured using IFR 2023B (Marconi). |
Attachment 1: gain.png
|
|
Attachment 2: gain.pdf
|
|
10070
|
Thu Jun 19 12:10:18 2014 |
Akhil | Update | Electronics | Gain plots to Characterize UFC-6000 RF Frequency Counter |
The goal is to characterize the Mini-Circuits RF FC (Model UFC-6000) by plotting Gain Plots.
Work Done:
The sampling rate of the UFC-6000 RF FC is 1s (should look into making the sampling time smaller). So to satisfy Nyquist criterion, the maximum modulation frequency is 0.5 Hz beyond which aliasing effects are seen.
The measurements taken (mentioned in my previous elog) are used to plot Gain vs Modulation frequency for carrier frequencies of 5 MHz and 25 MHz.
Calculations:
A modulated signal can be represented as X(t)= A*sin (Fc*t+D*sin(Fm*t+phase1)) where Fc and Fm are carrier and modulation frequencies respectively and D is the modulation depth.
This signal Y(t) is input to the FC and the output frequencies of the FC are recorded.
Let the output of the FC is Y(t)= A'*sin(Fc*t+D'*sin(Fm'*t+phase2));
Gain = D'/D;
phase = phase2 - phase1;
D' is calculated by subtracting the carrier frequency from the output frequency and calculating the amplitude of the resulting fitted sine wave.
The phase can be calculated if the phase of the input is known(which will be done next).
Plotting the Bode plots would give response of the FC to modulation.
The plots generated will be used to estimate:
1) Transfer Function of FC to be known to build an FOL-PID loop.
2) Quantization noise from Power Spectral Density(PSD) vs Hz.
To Do Next:
1)Calculate the phase difference to complete the Bode plot. This would require interfacing of the ADC on raspberry pi.
2)Estimate the quantization noise from the digital output.
|
Attachment 1: GainVsFreq.png
|
|
10377
|
Wed Aug 13 17:37:43 2014 |
Jenne | Update | General | Game plan |

Here's the game plan for things that we need to do to get this IFO locked up.
Red is for things that should be done today, or tomorrow if they don't get finished today (eg. laser mode hopping temperature check). Orange is for things that will become red once the current red things are gone (eg. inferring the POP QPD gouy phase, and moving it to minimized PRM information). Green is for things that we'd like to do, but aren't high priority (eg. X green mode matching). Blue is for things that we should remember, but not plan on working on soon (eg. putting PZTs on the Yend table for green).
TODAY so far:
Q already did the tweak up of the PSL SHG crystal alignment. HE SHOULD ELOG ABOUT THIS. What was the final power of green that you got? Do we have any record of a previous measurement to compare to?
Q helped me install PDA55s on each of the lasers (I did the ends, he did the PSL) so that we could do the mode hop temperature check. For the Yend, I took the leakage transmission through the first Y1 steering mirror after the laser. This beam was dumped, so I replaced the dump with a PDA55. For the Xend, the equivalent mirrors are too close to the edge of the table, so I put in a spare Y1, and reflect most of the light to a beam dump. The leakage transmission then goes to a PDA55. Note that for both of these cases, no alignment of main laser path mirrors was touched, so we should just be able to remove them when we're through. For the PSL, I believe that Q took the rejected light from one of the PBSes before the PMC. He mentioned that he bumped something, so had to realign the beam into the PMC, but that he was able to get the transmission back up to 0.802, when we were seeing it in the mid 0.7's for the last several days.
The end temporary PDs are using the TRX / TRY cables, so we will be looking at the C1:LSC-TR[x,y] channels for the power of the end lasers. The PSL's temporary PD is connected to the PMC REFL cable. For the end PDs, since I had filter banks available, I shuttered the end lasers and removed the dark offset. I then changed the gains to 1, so the values are in raw counts. The usual transmission normalization gains are noted in one of the control room notebooks.
I did a slow ezcastep and ramped the temperature of all 3 lasers over about an hour. I'll write a separate elog about how that went. |
10382
|
Thu Aug 14 02:51:46 2014 |
Jenne | Update | General | Game plan |
[Jenne, Rana]
* Decided that earlier mode hop scan won't give us the information that we were hoping for. We need to think about where we can actually see the frequency change. Can we use the IR beatnote that we will soon have to do this? We'd only be able to scan one laser temp at a time, but that's okay. Leave, say, the PSL temperature alone, and scan one of the end laser temps. Using the PSL as the reference, we will be able to see if the frequency of the end laser goes crazy and jumpy as we pass through a certain temp. Then, repeat while holding the end laser constant and scan the PSL. Thoughts?
* Meditated on PSL oplev servo, but I need to make a Matlab script that can evaluate different loops according to a cost function based on elog 9690.
* Aligned IFO to IR, then greens to arms (got back to 0.9 for GTRY, but only about 0.5 for GTRX, with the PSL green shutter closed). Then aligned green beams on the PSL table, since the PSL green pointing had changed a bit from Q's crystal alignment tweak-up earlier today. Beatnotes are nice and big (see elog 10381 - The Yarm is the larger beatnote, and the Xarm is the smaller one.)
* Was not able to lock ALS comm/diff and hold long enough to get both arms to IR resonance. Also, saw that TRY's RIN was more than 50%(!!!). We took a look, and there seems to be much more low frequency noise than there was when the spectrum in the control room was taken for the multicolor metrology paper:

* Tried to balance the ALS comm/diff input matrix, with not a lot of success. First of all, it looks like the Xarm has overall about 10 times more noise! We were exciting MC2 in position (~88 Hz, about 130 counts I think), and then looking at DARM_IN1 for the peak. When DARM_IN1 was just one of the 2 ALS error signals (i.e. one matrix element set to zero), versus when both matrix elements were set to 1, we saw a factor of only about 3 in reduction of the peak height. We were hoping to have better cancellation of this pure CARM signal in the DARM channel. The Xarm green PDH loses lock every ~5 or 10 minutes, and when we relock it, this cancellation seems different, so we want to try again tomorrow when the ALS is locked on comm / diff, rather than just the free running ALS that we have now. Although, if the balance of the input matrix changes lock-to-lock, we may need to consider redoing the green PSL table layout so we get a pure DARM beatnote signal like they have at the sites.
* We want to change how the watch script for ALS works, although this is a low-priority task. Rather than looking at the control signal, we should maybe look at the sum of all the coil outputs, multiplied by a pendulum TF, and use that as a rough displacement sensor. We want to be careful of pushing too hard at low frequencies, but we want to allow higher frequency actuation without having the watch script shut things down.
* Also, I should put on the to-do list the revamp of the ALS find IR resonance script. |
10404
|
Fri Aug 15 20:26:37 2014 |
ericq | Update | General | Game plan |
Quote: |
Q already did the tweak up of the PSL SHG crystal alignment. HE SHOULD ELOG ABOUT THIS. What was the final power of green that you got? Do we have any record of a previous measurement to compare to?
|
As Jenne mentioned, I did this.
Specifically, I first tweaked the mirror pointing the IR into the SGH in pitch and yaw to maximize the green power, and then adjusted the little set screws on the side of the SHG to maximize further. Power after the harmonic separator was of order 150uW. On the Y Green BBPD, I got ~48uW, instead of the 40uW Rana, Jenne, and myself saw the other night.
HOWEVER,
now that I look through old ELOGs, I find some posts by Kiwamu saying the power should be around 650uW, and that he was able to get 640uW out. So: I should do this again, systematically, more carefully, etc., etc. (Linked ELOG also states that optimum SHG temperature is alignment dependent...)
|
7606
|
Wed Oct 24 11:49:07 2012 |
Jenne | Update | Alignment | Game plan for the day |
Jamie has arranged for phase map measurements this afternoon, so I will take the 6 dichroic LaserOptik optics over to Downs at 1:15 this afternoon.
Team Jamie+Nic will lead the effort to clamp down dog clamps as placement markers for all 4 in-vac passive TTs, and then pull all 4 TTs out of the chambers. They plus Den will move the TTs to the Cleanroom, and will start to install the new pitch alignment hardware.
When I return with the optics, we will install them in the TTs and re-balance them. Then we can put them back in the chambers and get back to work on alignment.
After we re-install the TTs, we will need to check the leveling of all 3 corner tables, just to be sure. |
10400
|
Fri Aug 15 13:29:31 2014 |
Jenne | Update | General | Game plan: 15 Aug |

The game plan graffle file is now in the 40m dropbox, so anyone can edit it. Please just make sure to keep the date in the top right corner accurate. |
10440
|
Tue Sep 2 16:22:24 2014 |
Jenne | Update | General | Game plan: 2 Sept |
Slightly updated Game Plan. Mostly, Q is continuing to check out the Xend PDH box saturation, and I am thinking on what our requirements are for ALS, and thus for the green PDH boxes.

|
10635
|
Thu Oct 23 13:08:55 2014 |
ericq | Update | General | Gap in local Chiara backups |
After the second of the two recent power outages, the outlet powering Chiara's external drive for local backups didn't come back. The modification to the backup script I made correctly identified that the drive wasn't mounted, and happily logged its absence and didn't try to stuff the internal drive with a copy of itself. However, I hadn't checked the logs to see if the backups were proceeding until today... maybe I should set up an email alert for these, too.
I plugged the external drive into a live outlet, and mounted the 40mBackup drive with: sudo udisks --mount /dev/sdc1 , which is a helpful command that puts the drive at /media/40mBackup as it should be, based on the drive label.
The /cvs/cds backup is now proceeding, to make up for lost time. |
4843
|
Mon Jun 20 17:58:00 2011 |
rana | Update | CDS | Gateway program killed |
There was a rogue, undocumented, gateway process running on NODUS since ~4 PM. This guy was broadcasting channels back into the Martian and causing lockups in the IOO controls. I did a kill -9 on its process.
Someone will pay for this. |
16794
|
Thu Apr 21 11:31:35 2022 |
JC | Update | VAC | Gauges P3/P4 |
[Jordan, JC]
It was brought to attention during yesterday's meeting that the pressures in the vacuum system were not equivalent althought the valve were open. So this morning, Jordan and I reviewed the pressure gauges P3 and P4. We attempted to recalibrate, but the gauges were unresponsive. Following this, we proceeded to connect new gauges on the outside to test for a calibration. The two gauges successfully calibrated at atmosperic pressure. We then proceeded to remove the old gauges and install the new ones. |
Attachment 1: IMG_0560.jpeg
|
|
Attachment 2: IMG_0561.jpeg
|
|
11494
|
Tue Aug 11 16:13:28 2015 |
Eve | Update | General | Gaussianity tests |
I’m working on a code to determine the Gaussianity of a PSD.
Motivation:
It can be difficult to distinguish between GW events and non-Gaussian noise, especially in burst searches. By characterizing noise Gaussianity, we can better recognize noise patterns and distinguish between GW events and noise.
What I did:
I analyzed an hour of S5 L1 data. First, I plotted a timeseries, just to see what I was working with. Then, I produced a PSD (technically, an ASD) for the timeseries using Welch’s method in Python.
I split the data segment into smaller time-chunks and then produced a PSD for each chunk. All PSDs were superimposed in one plot. Here’s a plot for 201 time-chunks of equal length:

For a specific frequency, I can view the spread in PSD value through the production of a histogram.
Results:
I’ve made histograms displaying varying PSD values for the 201 PSD plot at 100 Hz, 500 Hz, and 1kHz.



For Gaussian noise, an exponential decay plot is expected. I will continue this analysis by following the statistical method in Ando et al. 2003 to calculate specific values indicative of the Gaussianity of various distributions. I’ll then look at different periods of time in the S5 L1 data to find periods of time suggesting non-Gaussian behavior. |