40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 341 of 341  Not logged in ELOG logo
New entries since:Wed Dec 31 16:00:00 1969
ID Date Author Type Category Subjectup
  7718   Fri Nov 16 03:12:39 2012 Ayaka, DenUpdateGreen Lockingxarm locked

We aligned and locked xarm for green.

etmxf.png

  7719   Fri Nov 16 09:57:57 2012 JenneUpdateGreen Lockingxarm locked

Quote:

We aligned and locked xarm for green.

 

 That's really, really awesome!

 

  9971   Mon May 19 22:44:21 2014 denUpdatePEMxend sei isolation kit is set

Quote:

Yend seismometer isolation kit (elog 8461) hosts Guralp seismometer. I made a cable for inside connection, assembled the kit and relocated the instrument from its previous position at the yend inside the kit.

Seismometer is connected to the readout box and running.

    

 Xend internal cabling and external connector is ready. We are waiting for seismometer from Gyro lab. We still need to fix the pot with clamps after we put the instrument in.

We also need a long cable from Xend to the guralp readout box.

  7716   Thu Nov 15 21:52:48 2012 Den, AyakaUpdateGreen Lockingyarm locked

We aligned and locked Y arm for green:

  • installed camera on PSL to monitor green transmission
  • aligned green path on the ETMY table to see the beam on the PSL camera
  • misaligned ETMY and aligned ITMY to see reflected beam on REFL PD
  • installed green transmission PD on PSL
  • aligned ETMY and locked YARM to 00 mode

I've switched error channel cable to output monitor. Whitening filter is need for scattering measurements.

   etmyf.png  yarm.png

  9485   Wed Dec 18 03:29:48 2013 DenUpdateLSCyarm locked on mc

As a CM slow path test I locked free swinging yarm by actuating on MC length with bandwidth of 200 Hz. Crossover with AO is not stable so far.

I used xarm as an ool frequency noise sensor. MC2 violin mode is at 645 Hz, I have added a notch filter to LSC-MC2 bank.

Attachment 1: MC_ARM.pdf
MC_ARM.pdf
  9967   Sat May 17 14:48:06 2014 denUpdatePEMyend sei isolation kit is set

Yend seismometer isolation kit (elog 8461) hosts Guralp seismometer. I made a cable for inside connection, assembled the kit and relocated the instrument from its previous position at the yend inside the kit.

Seismometer is connected to the readout box and running.

IMG_1405.JPG    IMG_1406.JPG

  10131   Mon Jul 7 08:13:41 2014 steveUpdateGeneralyes,there was firework
Attachment 1: 8days.png
8days.png
  2808   Mon Apr 19 13:23:03 2010 josephbConfigurationComputersyum update fixed on control room machines

I went to Ottavia, and tried running yum update.  It was having dependancy issues with mjpegtools, which was a rpmforge provided package.  In order to get it to update, I moved the rpmforge priority above (a lower number) that of epel ( epel -> 20 from 10, rpmforge -> 10 to 20).  This resolved the problem and the updates proceeded (all 434 of them). yum update on Ottavia now reports nothing needs to be done.

I went to Rosalba and found rpmfusion repositories enabled.  The only one of the 3 repositories in each file enabled was the first one.

I then added priority listing to all the repositories on Rosalba.  I set CentOS-Base and related to priority=1.  I set CentOS-Media.repo priority to 1 (although it is disabled - just to head off future problems). I set all epel related to priorities to 20. I set all rpmforge related priorities to 10.  I set all rpmfusion related priorities to 30, and left the first repo in rpmfusion-free-updates and rpmfusion-nonfree-updates were enabled.  All other rpmfusion testing repositories were disabled by me.

I then had to by hand downgrade expat to expat-1.95.8-8.3.el5_4.2.x86_64 (the rpmforge version).  I also removed and reinstalled x264.x86_64.  Lastly I removed and reinstalled lyx. yum update was then run and completed successfully.

I installed yum-priorities on Allegra and made all CentOS-Base repositories priority 1. I similarly made the still disabled CentOS-Media priority 1.  I made all epel related repos priority 20.  I made all lscsoft repos priority=50 (not sure why its on Allegra and none of the other ones).  I made all rpmforge priorities 10.  I then ran "yum update" which updated 416 packages.

 

So basically all the Centos control room machines are now using the following order for repositories:

CentOS-Base > rpmforge > epel > (rpmfusion - rosalba only) > lscsoft (allegra only)

I'm not sure if rpmfusion and lscsoft are necessary, but I've left them for now.  This should mean "yum update" will have far fewer problems in the future.

 

  1360   Thu Mar 5 02:24:19 2009 ranaConfigurationComputersyum.repos.d

I added the following repos which I found on allegra to megatron and then did a 'yum install sshfs' on both machines:

allegra:yum.repos.d>l
total 28
-rw-r--r-- 1 root root  428 Feb 12 16:47 rpmforge.repo
-rw-r--r-- 1 root root  684 Feb 12 16:47 mirrors-rpmforge
-rw-r--r-- 1 root root 1054 Feb 12 16:47 epel-testing.repo
-rw-r--r-- 1 root root  954 Feb 12 16:47 epel.repo
-rw-r--r-- 1 root root  626 Feb 12 16:47 CentOS-Media.repo
-rw-r--r-- 1 root root 1869 Feb 12 16:47 CentOS-Base.repo
-rw-r--r-- 1 root root  179 Feb 12 16:47 adobe-linux-i386.repo

This also required me to import the rpmforge GPG key:

sudo rpm --import http://dag.wieers.com/rpm/packages/RPM-GPG-KEY.dag.txt

  1546   Tue May 5 09:22:46 2009 carynUpdatePEMzeros

For several of the channels on the PEM ADCU, zeros are occuring at the same time. Does anyone know why that might happen or how to fix it?

Attachment 1: zerotest2.png
zerotest2.png
Attachment 2: zerotest.png
zerotest.png
  6784   Thu Jun 7 12:49:50 2012 JamieUpdateComputerszita network configured

Added zita to the martian host table, and configured his network accordingly:

zita            A       192.168.113.217

https://wiki-40m.ligo.caltech.edu/Martian_Host_Table

  9368   Tue Nov 12 21:50:01 2013 ranaUpdate zita network configured

 

 I installed 'nfs-client' on zita (the StripTool terminal). It now has mounted all the shared disks, but still can't do StripTool since its a 32-bit machine and our StripTool is 64.

  6774   Wed Jun 6 22:25:05 2012 JamieUpdateComputerszita now running ubuntu 10.04

I'll finish the rest of the config to turn it into a full-fledged workstation tomorrow.

  15016   Wed Nov 6 17:45:34 2019 gautamUpdateLSC~

Here is a comparison of the response of various DoFs in our various RFPD sensors for two different CARM offsets. Even in the case of the smaller CARM offset of ~1kHz, we are several linewidths away from the resonance. Need to do some finesse modeling to make any meaningful statement about this - why is the CARM response in REFL11 apparently smaller for the smaller CARM offset?

If you mistrust my signal processing, the GPS times for which I ran the sensing lines are:

CARM offset = ~30kHz (arm transmission <0.02) --- 1257064777+5min

CARM offset = ~1kHz (arm transmission ~5) --- 1257065566+5min

Quote:

Summary:

There seems to be stronger-than-expected coupling between CARM and the 3f sensors. 

ELOG V3.1.3-