Below are the transfer functions measured between the angular (pit, yaw) motion of X arm mirrors and the ALSX error signal. The measurements were again made for 1Hz-30Hz.
The transfer functions are mostly flat.
ITMX P - 200-300 Hz/urad (beat freq = 45 MHz)
ITMX Y - 700-800 Hz/urad (beat freq = 27MHz)
ETMX P - 500-600 Hz/urad (beat freq = 56 MHz)
ETMX Y - 1000-2000 Hz/urad (beat freq = 62.5MHz)
Data xml files can be found in /users/manasa/data/140521/ |