Entry  Fri May 16 00:25:53 2014, rana, Update, PSL, PMC realign 
    Reply  Fri May 16 09:46:05 2014, Steve, Update, PSL, pointing monitoring 
Message ID: 9961     Entry time: Fri May 16 09:46:05 2014     In reply to: 9960
Author: Steve 
Type: Update 
Category: PSL 
Subject: pointing monitoring 

Quote:

 Tonight I noticed that the drop in PMC transmission was ~1V, more than the usual of ~0.5V from the daily drift.

While re-aligning on the table, I noticed that the misalignment was not from either of the steering mirrors; i.e. I has to walk them both to get the alignment back. This implies that the misalignment is generated far upstream. Maybe the the laser itself is moving. We need some updates from Steve's laser misalignment tracker.

I'd like to replace the paper target with IOO -QPD_POS so we can log it.