Entry  Thu May 1 17:14:36 2014, steve, Update, PSL, thin window AR measured 
    Reply  Fri May 2 10:38:29 2014, steve, Update, PSL, thin window AR measured 
Message ID: 9895     Entry time: Thu May 1 17:14:36 2014     Reply to this: 9902
Author: steve 
Type: Update 
Category: PSL 
Subject: thin window AR measured 

CVI broadband AR coating was measured at the PSL-enclosure table around 9-10am today. The 2W Innolight first PBS  S polarization beam was used with an other 1/2 wave plate and PBS.

W2-PW1-1004-C-633-1064-0   This 0.045" thick window has 0.7- 0.8 % reflected beam on each sides at 5 degrees of incidence, P polarization.

The specification is  R avg <0.5 % per surface at 0 degree

Rana wants The device would be useless with such a high R, but R 0.1% is OK so I will get V coating.