Entry  Wed Mar 26 14:54:32 2014, Koji, Summary, LSC, PRMIsb locked with REFL165I&Q again 
    Reply  Wed Mar 26 21:51:42 2014, ericq, Summary, LSC, PRMIsb locked with REFL165I&Q again 
       Reply  Fri Mar 28 17:22:55 2014, Koji, Summary, LSC, PRMIsb locked with REFL165I&Q again 
          Reply  Mon Mar 31 13:15:55 2014, manasa, Summary, LSC, Alignment update 
       Reply  Mon Mar 31 17:47:57 2014, ericq, Summary, LSC, MICH sensing oddities in REFL 3F REFL_33_traj.pdfREFL_165_traj.pdf
          Reply  Mon Mar 31 21:23:30 2014, Gabriele, Summary, LSC, MICH sensing oddities in REFL 3F 
Message ID: 9754     Entry time: Wed Mar 26 21:51:42 2014     In reply to: 9753     Reply to this: 9758   9767
Author: ericq 
Type: Summary 
Category: LSC 
Subject: PRMIsb locked with REFL165I&Q again 

Incidentally, while messing around with transfer functions and sensing matrix elements this evening, I was able to sideband lock straight onto REFL33 I&Q.  The settings were all identical to Koji's ELOG, with the following differences:

Input ports:
REFL33   WHTN: 30dB demod phase +125.5deg (tweaked from 135.5 to minimize MICH in I)

Input matrix:

REFL33I x +1.0 -> PRCL
REFL33Q x +3.0 -> MICH

Servo:
MICH OFS 0 / Gain 1/ Limitter ON (Oscillations occurred at 1.3)
PRCL OFS 0 / Gain -0.04 / Limitter ON

Output matrix:

MICH ITMX -1.0 / ITMY 1.0
PRCL PRM 1.0

 