Message ID: 975     Entry time: Mon Sep 22 12:06:58 2008
Author: rob 
Type: Update 
Category: SUS 
Subject: ITMY UL OSEM 


Last week I found the ITMY UL OSEM dead. I went around and checked the connections on the various flat ribbon cables
in the suspension control chain; pushing hard on the rack end of the long cable that goes from the sus electronics rack to the
ITMY sat amplifier fixed the problem. It's been fine since then.

NB: A visual inspection of the cable connection would not have revealed a problem. You just can't trust those flat
ribbon connectors with the hook latches.