Entry  Mon Feb 24 13:40:13 2014, Jenne, Update, CDS, Computer weirdness with c1lsc machine Screenshot-Untitled_Window-1.png
    Reply  Mon Feb 24 15:25:29 2014, Jenne, Update, CDS, Computer weirdness with c1lsc machine 
       Reply  Mon Feb 24 16:26:14 2014, Jenne, Update, CDS, NTP fell out of sync on front end machines - fixed 
Message ID: 9663     Entry time: Mon Feb 24 15:25:29 2014     In reply to: 9662     Reply to this: 9664
Author: Jenne 
Type: Update 
Category: CDS 
Subject: Computer weirdness with c1lsc machine 

The LSC machine isn't any better, and now c1sus is showing the same symptoms.  Lame.

The link lights on the c1lsc I/O chassis and on the fiber timing system are the same as all other systems.  On the timing card in the chassis, the light above the fibers was solid-on, and the light below blinks at 1pps. 

Koji and I power-cycled both the lsc I/O chassis, and the computer, including removing the power cables (after softly shutting down) so there was seriously no power.  Upon plugging back in and turning everything on, no change to the timing status.  It was after this reboot that the c1sus machine also started exhibiting symptoms. 