Message ID: 9465     Entry time: Fri Dec 13 13:28:07 2013
Author: Den 
Type: Update 
Category: LSC 
Subject: arm calibration template 

I have calibrated ETMX and ETMY actuators and added a template armSpectra.xml into /users/Templates directory.

Template shows control and error signals of both arms. Procedure is standard: calibrate control to meters and match error based on UGF measurement. XARM UGF: 200 Hz, YARM UGF 210 Hz.

Noise level at high frequencies (>100 Hz) for YARM is 3*10-15 and is factor of 3 better then for XARM. Servo gains are in the same ratio. I think there is less light on POX than on POY RF PD because I checked phase rotation and analog gain. I assume transimpedances are the same.

Attachment 1: armsCal.pdf  67 kB  | Hide | Hide all
armsCal.pdf