40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Wed Nov 27 18:31:26 2013, Koji, Summary, LSC, Adittion of the ALS error signals to the LSC input matrix 
    Reply  Wed Dec 4 15:43:13 2013, Jenne, Summary, LSC, Put LSC DAQ channels back 
Message ID: 9430     Entry time: Wed Nov 27 18:31:26 2013     Reply to this: 9440
Author: Koji 
Type: Summary 
Category: LSC 
Subject: Adittion of the ALS error signals to the LSC input matrix 

The Phase tracker outputs (= ALS X/Y error signals) are now conveyed to the LSC model.

Their entry points at the LSC model are C1:LSC-ALSX_IN1 and C1:LSC-ALSY_IN1.
They are connected to the signal matrix (28th and 29th signals) via signal conditioning filters (C1:LSC-ALSX and C1:LSC-ALSY).

The main LSC screen has not been updated. The conventional ALS servos are still remains as they were.

This renovation required the recompilation of c1als, c1rfm, and c1lsc. Two PCIe-RFM bridge paths were added resulting in
increase of the c1rfm timing budget from 38 to 44.

ELOG V3.1.3-