Message ID: 928
Entry time: Thu Sep 4 17:17:03 2008
|
Author: |
Yoichi |
Type: |
Update |
Category: |
IOO |
Subject: |
MC open loop TF |
|
|
I measured open loop transfer functions of the MC servo.
The UGF was about 30kHz. Since there was some gain margin at higher frequencies, I increased
the input gain of the MC servo board from 19dB to 22dB. Now the UGF is 40kHz and we have more
phase margin (~30deg). |
|
|