40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Mon Mar 4 16:46:31 2013, yuta, Update, SUS, oplev calibration for PRM PRM_PIT.pngPRM_YAW.png
    Reply  Tue Mar 5 17:09:30 2013, yuta, Update, SUS, oplev calibration for ITMY ITMY_PIT.pngITMY_YAW.png
       Reply  Tue Apr 2 15:58:40 2013, Jenne, Update, SUS, oplev calibration for ITMX, BS opl_itmx.pngopl_bs.png
          Reply  Tue Apr 2 16:13:10 2013, rana, Update, SUS, oplev calibration for ITMX, BS 
             Reply  Tue Apr 2 16:55:34 2013, Jenne, Update, SUS, oplev calibration online for ITMs, BS, PRM comparison_opl.pdf
                Reply  Tue Apr 2 18:19:30 2013, Jenne, Update, SUS, BS, PRM oplev servos improved PRM_servo_matches_BS.pdfprm_bs_optical_levers_comparison.pdf
                Reply  Tue Apr 30 22:40:57 2013, Jenne, Update, SUS, oplev XY-plots reflect new calibration 
                   Reply  Sun Sep 22 22:05:10 2013, rana, Update, SUS, oplev XY-plots reflect new calibration noise.png
                      Reply  Thu Oct 3 01:06:48 2013, rana, jenne, Update, SUS, oplev XY-plots reflect new calibration etmx_ol.png
                         Reply  Thu Oct 3 09:49:49 2013, Steve, Update, SUS, ETMY oplev laser is sick 
                      Reply  Tue Nov 5 19:55:12 2013, rana, Update, SUS, oplev XY-plots reflect new calibration 
Message ID: 9152     Entry time: Sun Sep 22 22:05:10 2013     In reply to: 8514     Reply to this: 9188   9351
Author: rana 
Type: Update 
Category: SUS 
Subject: oplev XY-plots reflect new calibration 

The ETMX oplev signal looks kind of dead compared to the ETMY. It has no features in the spectra and the SUM is pretty low.

I noticed that the cal fields are still set to 1. To get it close to something reasonable, I calibrated it vs. the SUSPIT and SUSYAW values by giving it a step in angle and using 'tdsavg' plus some arithmetic.

OLPIT =  45 urads/ count

OLYAW = 85 urads / count

These are very rough. I don't even know what the accuracy is on the OSEM based calibration, so this ought to be redone in the way that Jenne and Gabriele did before.

The attached image shows the situation after "calibration" of ETMX. This OL system needs some noise investigation.