Quote: |
[Jenne, Manasa]
While we were trying to relock the MC after Jenne put back the RF box, we found there was some mysterious motion in MC2. After spending time trying to figure out where this was coming from, the source was found to be at LOCKIN2 of MC2 suspension "The MC TICKLER" that was left enabled. This was turned OFF and MC locked just fine after that.
EDIT JCD: The Tickler should be disabled, if the autolocker is disabled.
|
Sounds like this was just incidental, since the MC locked fine also with the tickler enabled for weeks.
The tickle is disabled by the down script, but there's no way to correctly handle all possible button pushes. If you want to disable the autolocker for some reason you should run mcdown before trying to lock. This will set up things with the correct settings. |