[Manasa, Masyauki]
Today we measured OLTF of PDH green lock of x-arm again. In the previous measurement the excitation signal was injected at the PDH servo box output(elog 9044), but in this measurement we changed the injection point to the RFPD mixer output (just before the servo input).
We measured the OLTF with the servo gain of 6.5 and source amplitude of 5 mV for all frequency band. The measured UGF was 11 kHz and the phase margin was 48 degree.
Next that measurement, we tried to measure the power spectrum density of the error signal and feedback signal. But the alignment was not so good, so we aligned the green light injection point. Tomorrow we will continue the alignment and will measure the PSD.
attatchment1 - OLTF of PDH green lock with servo gain of 6.5

|