901
Entry time:
Fri Aug 29 15:01:45 2008
Author:
steve
Type:
Update
Category:
PSL
Subject:
MOPA_HTEMP in increasing
The laser chiller temp is 21.9C ( it should be 20.0C )
Control room temp 73F ok, no obvious block
Ops, there is a piece of paper blocking the intake of the chiller
This is a four day plot. The paper was blocking the air flow all day.
Attachment 1:
htcl.jpg