Mon Jul 22 17:17:30 2013, annalisa, Update, endtable upgrade, End table picture
|
Tue Jul 23 00:51:46 2013, Koji, Update, endtable upgrade, End table picture   
|
Tue Jul 23 19:37:08 2013, Annalisa, Update, endtable upgrade, End table picture 
|
Wed Sep 25 19:34:51 2013, rana, Update, SUS, Problems with ETMY Optical Lever
|
Thu Sep 26 11:00:51 2013, Steve, Update, SUS, Problems with ETMY Optical Lever
|
Thu Sep 26 21:55:08 2013, rana, Update, SUS, Problems with ETMY Optical Lever
|
|
Message ID: 8910
Entry time: Tue Jul 23 19:37:08 2013
In reply to: 8896
|
Author: |
Annalisa |
Type: |
Update |
Category: |
endtable upgrade |
Subject: |
End table picture |
|
|
Quote: |
The spot on the IPANG QPD was checked. The spot is higher than the center and South side of the lens.
Some photos are found below.
The spot on the IPANG steering mirrors in the ETMY chamber was also checked.
It is clipped at the top of the steering mirror. (See attachment 4)
So basically the spot is about 1" above the center of the mirror.
|
After the vent, the IPang spot position on the steering mirrors on the Yend table moved approximately by 1 inch down.
Inside the chamber, the spot position is in the center of the steering mirror. (difficult to take a picture because the PSL beam power has been reduced) |
|
|
|