I tried to retake POP QPD calibration data again today. The MC was mostly fine, but whenever the PRMI unlocked, both ITM watchdogs would trip. I'm not sure what was causing this, but the ITM alignment wasn't perfect after this kind of event, so I felt like I was continuously locking and realigning the arms to get the alignment back. Then, after turning on the ASC and tweaking up the PRM alignment for maximum POP110I signal, I had to recenter the QPD, so none of my previously taken data was useful. Frustrating. Also, I had recentered the PRMI-relevant oplevs, but I had these weird locklosses even with nicely centered oplevs.
I have given up for the daytime, and will come back to it if there's a spot in the evening when arm measurements aren't going on.
Here is the data from last week, and the data from today. The micrometer readings have been calibrated into mm, and I have fit a line to the linear-looking region. Obviously, for the Pitch calibration, I definitely need to take more data.



|