Mon Jun 3 18:49:44 2013, Gautam, Update, General, ITMx Oplev Fixed
|
Wed Jun 5 20:45:19 2013, gautam, Update, General, ITMx Oplev Fixed 
|
Wed Jun 26 19:02:38 2013, gautam, Update, General, ITMx Oplev (not quite?) Fixed  
|
Thu Jun 27 18:11:53 2013, gautam, Update, General, ITMx Oplev-POX looks beam okay
|
Thu Jun 27 23:01:39 2013, gautam, Update, General, ITMx Oplev-servo gains adjusted 
|
Fri Jun 28 13:10:28 2013, gautam, Update, General, ITMx Oplev-servo gains adjusted 
|
|
Message ID: 8770
Entry time: Thu Jun 27 18:11:53 2013
In reply to: 8758
Reply to this: 8777
|
Author: |
gautam |
Type: |
Update |
Category: |
General |
Subject: |
ITMx Oplev-POX looks beam okay |
|
|
Jenne just aligned the X arm and I got a chance to check the status of the POX beam coming out of the chamber. Turned the Oplev servo off so that the red beam could be blocked, turned all the lights off, and had a look at the beam in the vicinity of the mirror steering the Oplev-out beam to the QPD with an IR view-card. The beam is right now about half a centimeter from the pitch knob of the said mirror, so its not getting clipped at the moment. But perhaps the offending mirror can be repositioned slightly, along with the Oplev QPD such that more clearance is given to the POX beam. I will work this out with Steve tomorrow morning. |