Particle counts were measured inside the PSL enclosure at 40VAC variac setting.
Directly under HEPAs, on the east half of the optical table : 0.3 micron 20 particles / cu ft, 0.5 micron 0 p / cu ft, 0.7 micron 0 p / cu /ft
Away from the HEPAs, on the west edge of the op table: 0.3 micron 960 particles / cu ft, 0.5 micron 50 p / cu ft, 0.7 micron 20 p / cu ft
We may want to increase the RPM a little bit.
Checked our counters against recently calibrated Met One GT-526:
#1 is right on, #2 measured 25% less (it will go for calibration)
Flow bench at the south end : zero particle for all 3 sizes
IOO chamber particle logging measures 0.5 and 1 micron. Chamber opening limit is set to 10,000 particles max of 0.5 micron
At 10,000 p of 0.5 micron means ~100,000 - 180,000 p of 0.3 micron
We may have to lower the limit. |