40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Message ID: 876     Entry time: Mon Aug 25 10:51:06 2008
Author: steve 
Type: Update 
Category: PSL 
Subject: psl headtemp is coming down 
The laser water cooler was overflowing this morning.
I removed 500 cc water from the chiller.

The 4 days plot shows clearly:
that the capacity of the chiller is depending on the water level.
Overflowing water is a heat load for the chiller, so laser head temp goes up.
Attachment 1: ht4dw.jpg  78 kB  | Hide | Hide all
ht4dw.jpg