[Rana, Manasa]
ALS noise suppressed to 1KHz/rtHz. 1kHz RMS.
Plot 1: Scan of X arm by changing offset into Phase Tracker -> Xarm loop. Filter bank ramp time set to 120 s + using a 30 mHz low pass filter. IR beam is aligned to x arm, but not well.
Plot 2: ALS error signal with loop open (BLUE), closed with old filters (PURPLE), and with new, better boost (RED).
Plot 3: Bode plot of new boost (FM10), v. old, sad boost (1:50 pole:zero). RMS is now less than 1 kHz or ~50 pm. (in your face, Kiwamu!)
Changes made to the ALS servo:
1. C1ALS-TRX
ALS-TRX has been calibrated to read from 0-1 instead of counts in 1000 s. Calibration factor = 1/4500 = 0.00022
2. C1ALS_BEATX_FINE
Old antiwhitening filter has been removed. Added LPF at 1000Hz to remove glitches at high frequencies.
3. C1ALS-BEATX_FINE_PHASE
No changes made.
4. C1ALS-XARM
FIlter FM5 modified. 1000:1 changed to 3000:1
5. Offset for ALS scan were given through C1ALS_OFFSETTER1 with LPF50m enabled.
The filter modules of the servo were:



Next:
Check PZT out range for ALS. Figure out what the deal is with ALS SLOW servos.
Add DQ channels for ALS.
Automatic ALS up script (enable and disable phase tracker included).
|