40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Message ID: 874     Entry time: Mon Aug 25 10:07:35 2008
Author: Jenne 
Type: Update 
Category: PSL 
Subject: Numbers for the PMC servo board (Re: entry # 873) 
Jenne, Rana

These are the numbers that go along with Rana's entry #873:

The existing notch in the PMC servo is at 31.41kHz.

The power spectrum of the PMC has a peak at 14.683kHz when it is just sitting on the PSL table (no extra mass). When we put a pile of steel and aluminum (~20lbs) on top of the PMC, the body resonance moves to 14.622kHz, but is decreased by about 40 dB!

Rana has ordered a lead brick + foil that should arrive sometime this week. To complete the mechanical part of this installation, we need to extend the earthquake mounts around the PMC so that the lead brick can't fall off of the PMC onto the rest of the table.
ELOG V3.1.3-