40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Tue Oct 25 01:14:15 2011, Suresh, Update, IOO, C1IOO model modified to include new WFS lockin structure c1ioo20111025.png
    Reply  Tue Oct 25 16:24:58 2011, Suresh, Update, IOO, C1IOO model modified to include new WFS lockin structure 
       Reply  Tue Oct 25 18:50:22 2011, Suresh, Update, IOO, C1IOO model modified to include new WFS lockin structure 
          Reply  Wed Jun 19 15:07:20 2013, rana, Update, IOO, WFS debugging 
             Reply  Thu Jun 20 15:15:39 2013, rana, Update, IOO, WFS debugging Untitled.png
                Reply  Thu Jun 20 20:46:16 2013, gautam, Update, IOO, WFS debugging-QPD debugging MC2_QPD-map.png
                Reply  Tue Jul 9 20:09:53 2013, Koji, Update, IOO, WFS debugging 
                Reply  Sun Oct 6 23:43:07 2013, rana, Update, IOO, WFS debugging 
Message ID: 8733     Entry time: Thu Jun 20 15:15:39 2013     In reply to: 8724     Reply to this: 8735     8815   9210
Author: rana 
Type: Update 
Category: IOO 
Subject: WFS debugging 

Tried a bunch of stuff, but eventually just turned off the TRANS_QPD loops and loops are stable. Needs more debugging.

  1. Modified the on/off scripts so that the Integrators are no longer toggled. No reason to turn them off since we are clearing the filter bank histories.
  2. With QPD feedback OFF, I have lowered the overall gain by 15x so that its just drift control.
  3. Deleted unused / bad filters from the main filter banks.
  4. Gautam is going to debug the QPD with a red laser pointer and then elog.
  5. Jamie is checking out the MC Coil dewhitening logic to see if that's in a funny state.
Attachment 1: Untitled.png  419 kB  | Hide | Hide all
Untitled.png