Entry  Thu May 16 04:46:37 2013, Jenne, Update, LSC, Kiwamu's sensing matrix measurement script revived 
    Reply  Thu May 16 23:48:39 2013, Jenne, Update, LSC, Kiwamu's sensing matrix measurement script revived 
       Reply  Mon May 20 17:49:36 2013, Jenne, Update, LSC, PRMI sensing matrix - not high quality data PRMIsensMatQuickAnalysis.m.gz
          Reply  Mon May 20 18:47:47 2013, Koji, Update, LSC, PRMI sensing matrix - not high quality data 
             Reply  Tue May 21 14:48:08 2013, Jenne, Update, LSC, PRMI sensing matrix - not high quality data 
       Reply  Mon May 20 18:50:22 2013, Jenne, Update, LSC, Kiwamu's sensing matrix measurement script revived 
          Reply  Tue May 21 18:22:18 2013, Jenne, Update, LSC, Sensing matrix scripts modified to include actuator calibration 
             Reply  Wed May 22 00:08:19 2013, Koji, Update, LSC, Sensing matrix scripts modified to include actuator calibration 
                Reply  Wed May 22 18:24:19 2013, Jenne, Update, SUS, Violin mode survey 
                Reply  Wed May 22 20:50:26 2013, Jenne, Update, LSC, Sensing matrix scripts don't calculate correctly 
                   Reply  Thu May 23 14:45:08 2013, Jenne, Update, LSC, Sensing matrix scripts calculations make more sense now SensMat_JenneMeas_23May2013.pngSensMat_KojiMeas_23May2013.png
                      Reply  Thu May 23 21:45:51 2013, Jenne, Update, LSC, Sensing matrix scripts now check for lockloss 
                Reply  Thu May 23 19:09:15 2013, Jenne, Update, LSC, Sensing matrix scripts modified to include actuator calibration SensMat_21May2013.png
          Reply  Wed May 22 18:07:36 2013, Jenne, Update, LSC, Kiwamu's sensing matrix measurement script revived 
             Reply  Thu May 23 00:49:13 2013, Jenne, Update, LSC, LSC filters loaded 
Message ID: 8601     Entry time: Mon May 20 18:47:47 2013     In reply to: 8600     Reply to this: 8603
Author: Koji 
Type: Update 
Category: LSC 
Subject: PRMI sensing matrix - not high quality data 

For now forget about the demodulation phase and assume all of the ports are independent.
I want to know the numbers in the following format.

          PRCL     MICH   (unit: cnt/m)
REFL11I:  x.xxxEx x.xxxEx
REFL11Q:  x.xxxEx x.xxxEx
REFL33I:  x.xxxEx x.xxxEx
REFL33Q:  x.xxxEx x.xxxEx
REFL55I:  x.xxxEx x.xxxEx
REFL55Q:  x.xxxEx x.xxxEx
REFL165I: N/A     N/A
REFL165Q: N/A     N/A
AS55I:    x.xxxEx x.xxxEx
AS55Q:    x.xxxEx x.xxxEx

If you really want to resolve the TF phase difference between the I and Q  demod-signals,
you need to look at the transfer functions between the excitation and these ports.
We can't understand what is happening only from the single point measurement.