40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Mon Mar 4 16:46:31 2013, yuta, Update, SUS, oplev calibration for PRM PRM_PIT.pngPRM_YAW.png
    Reply  Tue Mar 5 17:09:30 2013, yuta, Update, SUS, oplev calibration for ITMY ITMY_PIT.pngITMY_YAW.png
       Reply  Tue Apr 2 15:58:40 2013, Jenne, Update, SUS, oplev calibration for ITMX, BS opl_itmx.pngopl_bs.png
          Reply  Tue Apr 2 16:13:10 2013, rana, Update, SUS, oplev calibration for ITMX, BS 
             Reply  Tue Apr 2 16:55:34 2013, Jenne, Update, SUS, oplev calibration online for ITMs, BS, PRM comparison_opl.pdf
                Reply  Tue Apr 2 18:19:30 2013, Jenne, Update, SUS, BS, PRM oplev servos improved PRM_servo_matches_BS.pdfprm_bs_optical_levers_comparison.pdf
                Reply  Tue Apr 30 22:40:57 2013, Jenne, Update, SUS, oplev XY-plots reflect new calibration 
                   Reply  Sun Sep 22 22:05:10 2013, rana, Update, SUS, oplev XY-plots reflect new calibration noise.png
                      Reply  Thu Oct 3 01:06:48 2013, rana, jenne, Update, SUS, oplev XY-plots reflect new calibration etmx_ol.png
                         Reply  Thu Oct 3 09:49:49 2013, Steve, Update, SUS, ETMY oplev laser is sick 
                      Reply  Tue Nov 5 19:55:12 2013, rana, Update, SUS, oplev XY-plots reflect new calibration 
Message ID: 8514     Entry time: Tue Apr 30 22:40:57 2013     In reply to: 8391     Reply to this: 9152
Author: Jenne 
Type: Update 
Category: SUS 
Subject: oplev XY-plots reflect new calibration 

Back when Gabriele was here, he and I implemented online calibration of the oplevs, into microradians.  A consequence of this is that all of the XY-plots on the medm screens were too small. 

I have gone through all the screens that I could think of (SUS_SINGLE, SUS_SINGLE_OPTLEV_SERVO, OPLEV_MASTER, OPLEV_SUMMARY, OPLEV_SUMMARY_SMALL_SCALE, IFO_OVERVIEW) and made the range of the XY-plots +/- 100, rather than the old scale of +/-1. 

I have also added red boxes behind the numbers on many (but not yet all) of these screens, so that you can see when (a) the oplev sum is too low, or (b) either the pit or yaw value is over 50 microradians. 

While I was putzing around on the IFO overview screen, I also made the oplev sum numbers clickable, with the related display being that optic's oplev servo screen.