[Eric, Riju]
Summary: Routing Fibers on AP table for Photo Diode Frequency Response Measurement System
Objective: We are to set-up one simultaneous transfer-function measurement system for all the RF-PDs present in 40m lab. A diode laser output is to be divided by 1x16 fiber splitter and to be sent to all the PDs through single-mode fiber. The transfer function of the PDs will be measured using network analyzer. The output of the PDs will be fed to network analyzer via one RF-switch.
Work Done So Far: We routed the fibers on AP table. Fibers from RF PDS - namely MC REFL PD, AS55, REFL11, REFL33, REFL55, REFL165, have been connected to the 1x16 fiber splitter. All the cables are lying on the table now, so they are not blocking any beam.
We will soon upload the schematic diagram of the set up.
Missing Component: Digital Fiber Power Meter, Thorlab PM20C
|