40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Tue Apr 23 16:31:13 2013, Eric, Configuration, , PD frequency response 
    Reply  Wed Apr 24 18:51:12 2013, Koji, Configuration, optical tables, PD frequency response 
Message ID: 8478     Entry time: Tue Apr 23 16:31:13 2013     Reply to this: 8487
Author: Eric 
Type: Configuration 
Category:  
Subject: PD frequency response 

[Eric, Riju]

Summary: Routing Fibers on AP table for Photo Diode Frequency Response Measurement System

Objective: We are to set-up one simultaneous transfer-function measurement system for all the RF-PDs present in 40m lab. A diode laser output is to be divided by 1x16 fiber splitter and to be sent to all the PDs through single-mode fiber. The transfer function of the PDs will be measured using network analyzer. The output of the PDs will be fed to network analyzer via one RF-switch.

Work Done So Far: We routed the fibers on AP table. Fibers from RF PDS - namely  MC REFL PD, AS55, REFL11, REFL33, REFL55, REFL165, have been connected to the 1x16 fiber splitter. All the cables are lying on the table now, so they are not blocking any beam.

We will soon upload the schematic diagram of the set up.

 

Missing Component: Digital Fiber Power Meter, Thorlab PM20C

 

 

ELOG V3.1.3-