Message ID: 8440     Entry time: Thu Apr 11 03:23:12 2013
Author: Den 
Type: Update 
Category: General 
Subject: MCL threshold 

MC down script is too slow to block MC_L when the cavity goes out of lock. As a result the loop strongly kicks MC2. We decided to make a threshold inside MCS model on MC TRANS that will block MC_L during lock loss. This is a lower threshold. Upper threshold can be slow and is implemented inside MC up script.

Fast threshold can be set inside MC2 POS. I did not correct MC2 top level medm screen as it is the same for all core optics.

Note: Fast trigger will also block ALS signal if MC loose lock.