Message ID: 8263
Entry time: Fri Mar 8 21:00:05 2013
In reply to: 8237
|
Author: |
Manasa |
Type: |
Update |
Category: |
PSL |
Subject: |
PMC fixed |
|
|
Quote: |
Mystery
1. Filter module (FM1) on PRCL and MICH show significant delay while enabling and disabling.
2. I tried to fix PMC alignment (PMC trans was 0.76). I was not able to get PMC trans more than 0.79.
PMC has been this way since yesterday.
3. MICH is still bright when locked (ASDC_OUT reads 0.08 for dark and 2.0 for bright). We suspect it is because of the AS55_I error offset that persists even after running LSCoffsets script.
4. PRMI shows some dither at 3Hz when locked.
|
[Koji, Manasa]
PMC is fixed with 0.84 in transmission. It was misaligned in pitch (fixing this increased PMC_trans to 0.822 from 0.773) and Koji also touched the wave plate and polarizer (changed PMC_trans to 0.845). |