Message ID: 8244     Entry time: Wed Mar 6 18:51:07 2013
Author: Annalisa 
Type: Update 
Category: Alignment 
Subject: Auxiliary laser installed for FSR and TMS measurement of the PRC 

We want to measure the g-factor of the PRC using the beat note of the main laser with an auxiliary NPRO laser.

We are going to phase lock the NPRO to the main laser (taking it from POY) and then we will inject the NPRO  through the AS edge of the ITMY.

Today Sendhil and I installed the auxiliary laser on the ITMY table moving it from the AS table.

We also installed the beam steering optics, except the BS which will launch the beam through the AR edge of the ITMY.

To do: install the BS, take the POY beam and mix it with the auxiliary laser on a photodiode to phase lock the two lasers, do better calculations for the mode matching optics to be used for the auxiliary laser beam.

Attachment 1: IMG_3-6-13.JPG  4.635 MB  | Hide | Hide all