40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Message ID: 8193     Entry time: Wed Feb 27 22:28:53 2013
Author: Brett 
Type: Update 
Category: SUS 
Subject: Global Damping Update 

 New excitation points were added after the global damping loops for more testing options. The updated c1sus.mdl model was re-committed to the svn. Two interesting simulink 'requirements' were found during this minor modification. First, excitation points must be placed on the top level of the diagram. If they are in a subsystem you will get compiling errors. Second, the excitation name must end in _EXC. It will compile OK if you don't do this, but the excitation points will not put out any excitations.

To do further investigation on the mysterious gain factor of 2.5 between the ETMY and ITMY POS damping loops, I measured TFs in the POS direction to the locked YARM signal for each. This provides an additional sensor, common to both, so we can see if the gain is coming from the actuation side or sensing side of the damping loops. The difference in these TFs is about 

2.895

So it seems the majority of the damping gain difference is on the actuation side with some small difference on the sensing side. In order to allow for the later splitting of YARM LSC control between ITMY and ETMY (global damping and the cavity control must be along the same coordinate system), I placed this gain of 2.95 in ITMY_LSC.

To get a first measure of the relative performance of global damping to local damping I measured some TFs between the sensor signal inputs and YARM. So first, while the cavity was still locked with just ETMY, I measured a TF between C1:SUS-ITMY_SUSPOS_EXC and C1:LSC-YARM_IN1. Second, I split the cavity control evenly between the ETMY and ITMY by adjusting C1:LSC-OUTPUT_MTRX. I turned off the local damping and turned on the common DOF global damping (called CARM at this point despite being on just one arm). I then repeated the same TF but driving from C1:SUS-GLOBAL_CARMDAMP_EXC.

The resulting TFs are displayed in the attached figure. The blue curve is then the TF from local damping sensor noise to YARM. The green is global damping sensor noise to YARM. The suppression between local to global is in red. The global damping curve is about 50 to 100 times lower (better) than local damping. This can probably be improved with further tuning to account for remaining differences between the ITMY and ETMY.

Note, the damping loop used in the filter modules for all of these is zpk(0,[15 15],1), with a gain of 30. This purposely has little high frequency filtering so it is easier to see the influence on YARM.

Attachment 1: DampNoise_to_YARM_fig_27Feb2013.png  39 kB  | Hide | Hide all
DampNoise_to_YARM_fig_27Feb2013.png
ELOG V3.1.3-